K014062 · Sterling Medivations, Inc. · FPA · Jan 17, 2002 · General Hospital
Device Facts
Record ID
K014062
Device Name
SIMPLICITY EASY ACCESS INFUSION SET
Applicant
Sterling Medivations, Inc.
Product Code
FPA · General Hospital
Decision Date
Jan 17, 2002
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Simplicity Easy Access Infusion Set is to provide a means for infusion and/or injection of fluids into the body below the surface of the skin when attached to an external pump or syringe.
Device Story
Simplicity Easy Access Infusion Set facilitates subcutaneous fluid delivery; connects to external pumps or syringes via threaded luer connector. Device comprises AISI 304 stainless steel indwelling needles, infusion hub for disconnection, and polyethylene connecting tubing. Used by patients or clinicians to deliver continuous or intermittent medication flow. Indwelling needles inserted below skin surface; hub/connector seal prevents leakage. Benefits include reliable access for fluid administration.
Clinical Evidence
Bench testing only. Compliance demonstrated via ISO 10555 (catheter requirements), ISO 594-1/2 (luer fittings), ISO 9626 (needle tubing), ISO 11135/11138 (sterilization), ISO 10993-1 (biocompatibility), and FDA LAL endotoxin testing guidelines.
Technological Characteristics
Materials: AISI 304 stainless steel needles, polyethylene tubing. Design: Infusion set with threaded luer connector, indwelling needle, and hub. Standards: ISO 10555, ISO 594-1/2, ISO 9626, ISO 11135, ISO 11138, ISO 11607, ISO 8535, ISO 10993-1. Sterilization: Ethylene oxide.
Indications for Use
Indicated for patients requiring subcutaneous infusion or injection of fluids via an external pump or syringe.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Maersk Medical Pureline™ Comfort™ Infusion Set (K972135)
Submission Summary (Full Text)
{0}------------------------------------------------
## JAN 1 7 2002
K014062
Sterling Medivations, Inc. 25285 La Loma Drive Los Altos Hills, CA 94022 650-949-0470 (voice) 650-949-0342 (fax
**510(k) SUMMARY**
Date Submitted: December 7, 2001
| Submitter: | Sterling Medivations, Inc. 25285 La Loma Drive, Los Altos Hills, CA 9 |
|------------|-----------------------------------------------------------------------|
| | Company Phone 952-473-7971, Company Fax 952-473-4758 |
- Joel Douglas, Chief Technology Officer Contact: Sterling Medivations, Inc. Applicant Phone 408-297-9474 ext. 0 voice mail 21 or 650-949-0470 Applicant Fax 408-297-9473 or 650-949-0342
Trade Name of Device: Simplicity™ Easy Access Infusion Set for infusion and/or injection of fluids into the body below the surface of the skin when attached to an external pump or syringe.
| Common Name of Device: | Intravascular administration set. |
|------------------------|--------------------------------------|
| Classification Name: | Percutaneous intravascular catheter. |
Medical Pureline™ Comfort™ Infusion Set, K972135. Predicate Device:
Description of the New Device: Sterling Medivations Inc.'s ("SMI") Simplicity Easy Access Infusion Set is designed for infusion and/or injection of fluids into the body below the surface of the skin when attached to an external pump or syringe.
The Simplicity Easy Access Infusion Set proposed for commercial distribution is similar in all significant respects to the existing Maersk Medical Pureline™ Comfort™ Infusion Set, K972135 and it has the same intended use.
The device consists of four main parts: (1) infusion catheters made from AISI 304 stainless steel, (2) an infusion hub that provides the patient the capability of disconnecting the connecting tube from the infusion catheter, (3) a connecting tube and (4) a threaded reservoir connector.
The Simplicity Easy Access Infusion Set is an infusion administration set, connecting to a medicine reservoir or syringe that is placed in an external infusion device and inserted below the surface of the skin of a patient. The SMI Simplicity Easy Access Infusion Set may be used with any infusion device that delivers continuous or intermittent flow.
The administration set attaches to the reservoir/syringe by means of a threaded luer connector, and under the surface of the skin in the patient through an indwelling needles formed from AISI 304 stainless steel. The connecting tubing is made from a polyethylene tube.
The indwelling needles formed from AISI 304 stainless steel are introduced below the surface of the skin. The connector needle is attached to the indwelling catheter. This seal on the connector needle mates with the indwelling catheter hub that forms a seal that permits the infusion of medication without leakage. The with and me work is made from AISI 304 stainless steel and it is connecting tubing. The connecting tubing. The connector tubing proximal end is attached to a threaded luer connector for attachment to the medicine reservoir.
Intended Use of the New Device: The intended use of the Simplicity Easy Access Infusion Set is to provide a means to for infusion and/or injection of fluids.into the body below the surface of the skin when attached to an external pump or syringe.
{1}------------------------------------------------
## Comparison of the Technological Features of the New Device and Predicate Device:
The Simplicity Easy Access Infusion Set proposed for commercial distribution is similar in all significant respects I the cumption Maersk Medical Pureline™ Comfort™ Infusion Set, K972135 and it has the same intended use.
The materials and manufacturing processes are substantially equivalent, the labeling is substantially equivalent and it has the same intended use as the Maersk Medical Pureline Comfort Infusion Set.
The differences that exist between the new and predicate device are as follows:
- The new device has a connecting tube of Polyethylene and the predicate device has a connecting tube of 1. co-extruded connecting tube Polyethylene id and PVC OD.
- The new device has multiple stainless steel indwelling needles formed from AISI 304 stainless steel and 2. the predicate device has an indwelling catheter form from polytetrafluorthylene (PTFE).
Performance Data Supporting Substantial Equivalence: To prove substantial equivalence the Simplicity Easy Access Infusion Set meets the catheter requirements of:
- CDRH 21 C.F.R. Section 880.5440 Intravascular administration set .
- ISO 10555 Sterile, single use intravascular catheters (Part 1: General Requirements) .
- ISO 10555 Sterile, single use intravascular catheters (Part 5: Peripheral Catheters). *
- ISO 11135:1994 Medical devices -- Validation and routine control of ethylene oxide sterilization .
- ISO 11138-2:1994 Sterilization of health care products -- Biological indicators -- Part 2: Biological indicators . for ethylene oxide sterilization.
- ISO 594-1:1986 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical . equipment - Part 1: General requirements,
- ISO 594-2:1998 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical . equipment -- Part 2: Lock fittings
- ISO 9626: 1991 Stainless Steel needle tubing for the manufacture of medical devices. .
- ISO 11607: 1997 Packaging for terminally sterilized medical devices. .
- ISO 8535: 1991 Sterile single use syringes, with or without needle, for insulin. .
- FDA Guidelines on validation of the Limulus Amebocyte Lysate (LAL) Test as an end product endotoxin test . for human and animal parenteral drugs, biological products, and medical devices.
- ODE Blue Book Memorandum #K90-1. �
- ISO 10993-1, Biological evaluation of medical devices Part 1: Evaluation and testing. .
And the design process adhered to is the Center for Devices and Radiological Health. DESIGN CONTROL GUIDANCE FOR MEDICAL DEVICE MANUFACTURERS. This Guidance relates to FDA 21 CFR 820.30 and Sub-clause 4.4 of ISO 9001. This is substantially equivalent to the predicate device.
Signed,
Joel S. Douglas Chief Technology Officer
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white circular logo. The logo features the symbol of the Department of Health and Human Services (HHS) in the center, which is a stylized caduceus with a triple helix. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 7 2002
Mr. Joel S. Douglas Chief Technology Officer Sterling Medivations, Incorporated 25285 La Loma Drive Los Altos Hills, California 94022-4583
Re: K014062
Trade/Device Name. Simplicity Easy Access Infusion Set Regulation Number: 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: December 7, 2001 Received: December 10, 2001
Dear Mr. Douglas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Douglas
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
K014062
510(k) Number (if known):
Device Name: Simplicity Easy Access Infusion Set
Indications For Use:
The intended use of the Simplicity Easy Access Infusion Set is to provide a means ♥ for infusion and/or injection of fluids into the body below the surface of the skin when attached to an external pump or syringe.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Patara Cuscente
(Division Sign-Off)
Division of Dental, Infection Control, and General Hospital Devices 510(lk) Number _______________________________________________________________________________________________________________________________________________________________ 014062
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.