K021135 · Ge Medical Systems Information Technologies · HGM · May 9, 2002 · Obstetrics/Gynecology
Device Facts
Record ID
K021135
Device Name
MODEL 2120IS SERIES MATERNAL/FETAL MONITOR
Applicant
Ge Medical Systems Information Technologies
Product Code
HGM · Obstetrics/Gynecology
Decision Date
May 9, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.2740
Device Class
Class 2
Indications for Use
The 2120is Maternal/Fetal Monitoring System is intended for monitoring fetal and maternal vital signs: fetal heart rate; fetal heart rate high/low/poor signal quality alarms; maternal uterine activity; heart/pulse rate, blood pressure, nondiagnostic maternal ECG and %SpO2. Optional fetal movement detection is available.
Device Story
Corometrics 2120is is a maternal/fetal monitor for clinical use during antepartum, labor, and delivery. Inputs include fetal heart rate (Doppler ultrasound or FECG), uterine activity (intrauterine pressure catheter or tocotransducer), maternal non-invasive blood pressure, maternal pulse oximetry, and maternal ECG. Device processes signals to provide real-time monitoring, alarms for fetal heart rate, and bed-to-bed surveillance via Ethernet networking. It functions as a bedside terminal for the QS system. Clinicians use output data to assess maternal and fetal well-being. Benefits include continuous surveillance of vital signs to support clinical decision-making during pregnancy and childbirth.
Clinical Evidence
No clinical data. Bench testing only, including requirements specification review, code inspections, software and hardware testing, safety testing, environmental testing, and final validation.
Technological Characteristics
Maternal/fetal monitor with configurations for Doppler ultrasound, FECG, intrauterine pressure catheter, tocotransducer, NIBP, pulse oximetry, and MECG. Features Ethernet connectivity for bed-to-bed surveillance and integration with QS system. Software-based monitoring system.
Indications for Use
Indicated for non-invasive and invasive monitoring of the fetus during the antepartum period, labor, and delivery (fetal heart rate, uterine activity, fetal movement). Indicated for monitoring maternal vital signs (non-invasive blood pressure, functional oxygen saturation of arterial hemoglobin, heart/pulse rate, and non-diagnostic ECG) as an adjunct to clinical assessment. Not intended for neonatal or pediatric blood pressure monitoring.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
Predicate Devices
Corometrics Model 2120is Maternal/Fetal Monitor (K012718)
Masimo SET® 2000 Pulse Oximeter and LNOP® series of sensors and cables (K990966)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness
| Date: | April 8, 2002 |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems Information Technologies<br>61 Barnes Park Road North<br>Wallingford, CT 06492 USA |
| Contact Person: | Joelle Neider<br>Regulatory Affairs Specialist<br>GE Medical Systems Information Technologies<br>Phone: 203-949-8232<br>Fax: 414-918-8113 |
| Device: Trade Name: | Corometrics Model 2120is Maternal/Fetal Monitor |
| Common/Usual Name: | Maternal/Fetal Monitor |
| Classification Names: | 21 CFR 884.2740 System, Monitoring, Perinatal |
| Predicate Devices: | K012718 Corometrics Model 2120is Maternal/Fetal Monitor<br>K990966 Masimo SET® 2000 Pulse Oximeter and LNOP® series of<br>sensors and cables |
| Device Description: | The 2120is Maternal/Fetal Monitoring System consists of the following<br>features/options that can be available in multiple configurations:<br><ul><li>fetal heart rate (via Doppler Ultrasound of FECG)</li><li>maternal uterine activity (via intrauterine pressure catheter or<br/>tocotransducer</li><li>fetal movement detection</li><li>maternal non-invasive blood pressure (clinician prompted or<br/>automatic)</li><li>maternal pulse oximetry</li><li>maternal heart/pulse rate (MECG) and ECG waveform "snapshot"</li></ul> |
| | The 2120is is a full-featured maternal/fetal monitor. The device is also<br>capable of providing bed-to-bed surveillance when multiple 2120is<br>monitors are connected via an Ethernet network. The monitor is also<br>capable of acting as a bedside terminal to the QS system. In this<br>capacity, the 2120is monitor provides the features of a standard PC QS<br>bedside terminal. |
| Intended Use: | The 2120is Maternal/Fetal Monitoring System is intended for monitoring<br>fetal and maternal vital signs: fetal heart rate; fetal heart rate<br>high/low/poor signal quality alarms; maternal uterine activity;<br>heart/pulse rate, blood pressure, nondiagnostic maternal ECG and<br>%SpO2. Optional fetal movement detection is available. |
| Technology: | The Corometrics Model 2120is Maternal/Fetal Monitor employs the<br>same fundamental scientific technology as the predicate devices. |
8
{1}------------------------------------------------
KOZ1135 page 4/2
The Model 2120is Maternal/Fetal Monitor complies with the voluntary Test Summary: The Model 212016 Maternair to of this submission. The following Standards as detailou in Society of the development of the Model 2120is:
- Requirements specification review .
- Code inspections .
- Software and hardware testing .
- Safety testing .
- Environmental testing .
- Final validation .
Conclusion:
The results of these measurements demonstrated that the Model The Tesults of these Thousarentions SET Pulse Oximetry) is as safe, as effective, and performs as well as the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission to promote health, well-being, and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAY 0 9 2002
Mr. Joelle Neider Corporate Regulatory Affairs GE Medical Systems Information Technologies 61 Barnes Park Road North P.O. Box 333 WALLINGFORD CT 06492-0333 Re: K021135
Trade/Device Name: Corometrics Model 2120is Maternal/Fetal Monitor Regulation Number: 21 CFR 884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II
Product Code: 85 HGM Dated: April 8, 2002 Received: April 9, 2002
Dear Mr. Neider:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 510(k) Number (if known):
### 510(k) filed on April 8, 2002
Device Name:
Corometrics Model 2120is Maternal/Fetal Monitoring System
Indications For Use:
#### Fetal Surveillance | .
A Corometrics 2120is Maternal/Fetal Monitoring System is used for non-invasive and invasive monitoring of the fetus during the antepartum period as well as throughout labor and delivery (i.e. fetal heart rate and uterine activity monitoring). Fetal movement detection and fetal heart rate alarm options (user selectable high/low and poor signal quality alarms) are available.
11 Maternal Monitoring
> A Corometrics 2120is Maternal/Fetal Monitoring System is intended for monitoring maternal vital signs to help assess maternal well-being. The vital signs which can be measured with these monitor configurations are summarized as follows:
NOTE: Maternal vital signs provided by the monitor should only be used as an adjunct in patient assessment in conjunction with clinical signs and symptoms.
Blood Pressure: The monitor is intended for use in the non-invasive monitoring of maternal blood pressure (NBP). This monitor is not intended for use in the neonatal or pediatric blood pressure monitoring.
Pulse Oximetry. The monitor is intended for use in the non-invasive monitoring of maternal functional oxygen saturation of arterial hemoglobin (MSpO2).
Heart/Pulse Rate. The monitor is intended for use in the non-invasive monitoring of the maternal heart/pulse rate. Additionally, an MECG waveform "snapshot" may be displayed and printed.
NOTE: Only the maximum configuration provides both maternal heart rate and pulse rate data.
Bed-to-bed surveillance is available when 2120is Series Monitors are networked together.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div style="display:inline-block; vertical-align:top;">/</div> | OR | Over-The-Counter Use |
|----------------------|----------------------------------------------------------------|----|--------------------------|
| (Per 21 CFR 801.109) | | | (Optional Format 1-2-96) |
Nancy Broadon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
| 510(k) Number | K021135 |
|---------------|---------|
|---------------|---------|
31
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.