The Masimo SET® 2000 Pulse Oximeter and accessories is intended for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, and neonatal patients in hospitals, hospital-type facilities, mobile, and home environments.
Device Story
Portable, standalone pulse oximeter; monitors SpO2 and pulse rate. Inputs: light absorption signals from two LEDs via LNOP series sensors. Operation: monitor calculates ratio of oxygenated to total hemoglobin based on differential light absorption; displays plethysmographic waveform, SpO2, pulse rate, alarm limits, and trends. Used in hospitals, mobile, and home settings by qualified personnel. Output: visual/audible alarms and real-time physiological data. Benefits: provides early warning of patient deoxygenation; enables continuous monitoring during motion and low perfusion states. Device includes connecting cables and disposable/reusable sensors.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological similarity to the predicate device.
Technological Characteristics
Portable standalone monitor (4.26" x 9.75" x 9.75", 6.5 lbs). Power: 100-230 Vac or internal sealed gel cell battery (2-hour life). Sensing: dual-LED light absorption via LNOP series sensors. Connectivity: PC series connecting cables. Software: embedded firmware for signal processing and calculation of SpO2/pulse rate. Sterilization: not specified.
Indications for Use
Indicated for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate in adult, pediatric, and neonatal patients. Suitable for use during motion and no-motion conditions, and for well or poorly perfused patients in hospitals, hospital-type facilities, mobile, and home environments. Contraindicated for use as an apnea monitor.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Masimo SET MS-1P Pulse Oximeter and accessories (K973887)
Submission Summary (Full Text)
{0}------------------------------------------------
JUN 7 1999
Section 1 - Executive Summary
## Masimo SET® 2000 Pulse Oximeter
## 510(k) Summary
| Submitted By: | Masimo Corporation<br>2852 Kelvin Ave<br>Irvine, CA 92614-5826<br>(949) 250-9688<br>FAX (949) 250-9686 |
|-----------------------------------|--------------------------------------------------------------------------------------------------------|
| Contact: | James J. Cronin<br>Vice President, Regulatory Affairs/Quality Assurance |
| Trade Name: | Masimo SET® 2000 Pulse Oximeter and accessories |
| Common/Classification Name: | Oximeter (74DQA)<br>Transducer and Electrode Cable (including connector) (74DSA) |
| Substantially Equivalent Devices: | Masimo SET MS-1P Pulse Oximeter and accessories<br>510(k) Number - K973887 |
## Reason for Submission
Premarket notification for Masimo SET® 2000 Pulse Oximeter and accessories, a New Device, seeking authority to market the device under Section 510(k) as a device that is substantially equivalent to the Masimo SET® MS-1P Pulse Oximeter and accessories
## Intended Use of Device
The Masimo SET® 2000 Pulse Oximeter and accessories is intended for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, and neonatal patients in hospitals, hospital-type facilities, mobile, and home environments.
## Indications for Use
The Masimo SET® 2000 Pulse Oximeter and accessories are indicated for the continuous noninvasive of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor). The Masimo SET® 2000 Pulse Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
## Contraindications For Use:
The Masimo SET® 2000 Pulse Oximeter and accessories are contraindicated for use as apnea monitors.
The Masimo LNOP® Series of Disposable Sensors are contraindicated for patients that exhibit allergic reactions to addesive tape. The sensors must be removed and repositioned every eight (8) hours and if indicated by circulatory condition or skin integrity reapplied to a different monitoring site.
The Masimo LNOP® Reusable sensor is contraindicated for use . It is not intended for long-term monitoring. It must be removed and repositioned every 4 hours and if indicated by circulatory condition or skin integrity reapplied to a different monitoring site.
0018
{1}------------------------------------------------
# Section 1 - Executive Summary
## Device Description
The Masimo SET® 2000 pulse oximeter and accessories is a portable stand alone device, connecting cable, and oximenty sensors to nonivasively calculate the functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate.
The monitor consists of a screen that displays the pulse plethysmographic waveform, the pulse rate, SpO2 value, the high and low SpO, and pulse rate alarm limits, alarms, trends and status messages.
The monitor contains the electronic hardware and software that receives and calculates the signals from the LED's to determine the functional oxygen saturation of arterial hemoglobin (SpO-) and pulse rate and provides for the connection to the connecting cable.
The PC series connecting cables connects the monitor to the oximetry sensors and transfers LED drive power and the calibration drive to the oximetry sensors from the monitor receives the detector signal from the oximetry sensor.
The LNOP® series of oximetry sensors measure the light absorption of blood from two light emitting diodes (LED's). Oxygen saturated blood absorbs light differently as compared to unsaturated blood. The amount of light absorbed by the blood is used to calculate the ratio of oxygenated hemoglobin to total hemoglobin in arterial blood.
## Configurations
The Masimo SET® 2000 pulse oximeter is available in one configuration as a portable stand alone pulse oximeter that is 4.26 inches high, 9.75 inches wide, 9.75 inches deep, and weighs 6.5 lbs. The unit is powered either with a voltage input of 100-230 Vac, 47 - 63 Hz or with a sealed gel cell battery with an operating time of 2 hours and a charge time of less 8 hours.
The PC series of connecting cables is available in one configuration and three lengths, 4 feet, 8 feet and 12 feet.
The LNOP® series of oximetry sensors is available in five configurations: a single use oximetry sensor intended for adults and pediatrics greater than 30 kg; a single use oximetry sensor intended for pediativs and small adults greater than 10 kg and less than 50 kg, a single use oximetry sensor intended for neonates less than 10 kg with good skin integrity; a single use oximetry sensor intended for neonates with poor skin integrity less than 1 Kg; and a reusable oximetry sensor intended for adults and pediatrics greater than 30 kg.
0019
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of a human figure, with three overlapping profiles suggesting a sense of community and support. The overall design is simple and conveys a sense of health and well-being.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 7 1999
Mr. James J. Cronin Masimo Corporation 2852 Kelvin Avenue Irvine, CA 92614-5826
Re: K990966 Masimo SET® 2000 Pulse Oximeter and Assessories Regulatory Class: II (two) Product Code: 74 DQA Dated: March 22, 1999 Received: March 23, 1999
Dear Mr. Cronin:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Paqe 2 - Mr. James J. Cronin
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callehan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Section 3 - Indications for Use
#### 510(k) Number (if known):
K990966
#### WARNINGS
Explosion hazard. Do not use the Masimo SET® 2000 Pulse Oximeter in the presence of flammable anesthetics.
A pulse oximeter should NOT be used as an apnea monitor.
A pulse oximeter should be considered an early warning device. As a trend towards patient deoxygenation is indicated, blood samples should be analyzed by a laboratory co-oximeter to completely understand the patient's condition.
If an alarm condition occurs while the alarm silence period is set to off, only alarm indications will be visual displays and symbols related to the alarm condition.
The Masimo SET® 2000 Pulse Oximeter is to be operated by qualified personnel only.
Electric shock hazard. Covers to be removed only by technically qualified service personnel. There are no userserviceable parts.
To ensure patient electrical isolation, connect only to other equipment with electronically isolated circuits.
Do not connect to an electrical outlet controlled by a wall switch. Note:
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
Do not place the monitor or external power supply in any position that might cause it to fall on the patient. Do not lift the monitor by the power supply cord or patient cable, use only the handle on the monitor.
Do not use the Masimo SET® 2000 Pulse Oximeter or Masimo LNOP® series of sensors during magnetic resonance imaging (MR) scanning. Induced current could potentially cause burns. The Masimo SET® 2000 Pulse Oximeter may affect the MRI image, and the MRI unit may affect the accuracy of the oximetry measurements.
Interfering Substances: Carboxyhemoglobin may erroneously increase readings. The level of increase is approximately equal to the amount of carboxyhemoglobin present. Dyes, or any substance containing dyes, that change usual arterial pigmentation may cause erroneous readings.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
| 510(k) Number | K990966 | | |
|------------------------------------------|---------|----------------------|-------------------------------|
| Prescription Use<br>(Per 21 CFR 801.109) | or | Over-The-Counter Use | 0022 (Optional Format 1-2-96) |
6)
Section 3 Page 2
്രാ
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.