Cal-Chex CD Plus is used to calibrate the WBC, RBC, Hgb, MCV, PLT and MPV (MPV on CD4000 only) parameters on the Abbott Cell-Dyn 3200 and 4000 hematology instruments.
Device Story
Cal-Chex CD Plus is a stabilized suspension of human red blood cells, white blood cells, and platelets in 3.0 mL glass vials. It serves as a calibrator for Abbott Cell-Dyn 3200 and 4000 hematology analyzers. The device is used by laboratory personnel to calibrate instrument parameters (WBC, RBC, Hgb, MCV, PLT, MPV). It utilizes a formulation optimized for systems employing Multiple Angle Polarized Scatter Separation (MAPSS) technology. By providing a stable, reproducible reference material, it ensures the accuracy of hematology instrument measurements, facilitating reliable clinical diagnostic results for patients.
Clinical Evidence
Bench testing only. Three studies were conducted: run-to-run reproducibility, comparison to whole blood calibration, and long-term/open-vial stability. Results confirmed consistent reproducibility and stability for the entire product dating, supporting equivalence to whole blood calibration.
Technological Characteristics
Suspension of stabilized human red blood cells, white blood cells, and platelets. Packaging: 3.0 mL glass vials with injection-molded polypropylene screw-top caps in PVC clamshells. Principle: Calibration reference material for hematology analyzers. Optimized for MAPSS (Multiple Angle Polarized Scatter Separation) technology.
Indications for Use
Indicated for the calibration of WBC, RBC, Hgb, MCV, PLT, and MPV parameters on Abbott Cell-Dyn 3200 and 4000 hematology instruments in a clinical laboratory setting.
Regulatory Classification
Identification
A calibrator for cell indices is a device that approximates whole blood or certain blood cells and that is used to set an instrument intended to measure mean cell volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC), or other cell indices. It is a suspension of particles or cells whose size, shape, concentration, and other characteristics have been precisely and accurately determined.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
{0}------------------------------------------------
APR - 4 2002
KC20469
## 510(k) Summary of Safety and Effectiveness
| 510(k) Submitter: | Streck Laboratories, Inc.<br>PO Box 45625<br>Omaha, NE 68145 |
|-------------------|--------------------------------------------------------------|
|-------------------|--------------------------------------------------------------|
March 21, 2002 Date Prepared:
Names of Device:
Cal-Chex CD Plus Trade Name: Assaved hematology calibrator Common Name: Calibrator for hemoglobin or hematocrit measurement (§864.8165) Classification Names: Calibrator for platelet counting (§864.8175) Hematology quality control mixture (§864.8625) 81KRZ, 81KRY and 81GLQ Classification Numbers:
Cal-Chex manufactured by Streck Laboratories, Inc. Predicate Devices:
Description: Cal-Chex CD Plus is a suspension of stabilized human red blood cells, human white blood cells, and human platelets packaged in glass vials containing 3.0 mL volumes. Closures are injection molded polypropylene screw-top caps. The vials are packaged in a PVC clamshell.
- Intended Use: Cal-Chex CD Plus is intended to be used as a calibrator for the calibration of WBC, RBC, Hgb, MCV, PLT and MPV (MPV on CD4000 only) on the Abbott Cell-Dyn 3200 and 4000 hematology instruments.
- Comparison with Predicate Devices: Like Cal-Chex CD Plus is a product utilized to calibrate hematology instrumentation for specified parameters.
Unlike Cal-Chex, Cal-Chex CD Plus utilizes a formulation that has been optimized for use as a calibrator on Abbott Cel-Dyn® Systems employing Multiple Angle Polarized Scatter Separation (MAPSS™) technology.
Discussion of Tests: Three studies of Cal-Chex CD Plus were conducted:
- I. Run-to-Run Reproducibility and Comparison to Whole Blood
- II. Long Term and Open Vial Stabilities;
- III. Verification as a calibrator. Study results showed Cal-Chex CD Plus to be consistently reproducible, substantially equivalent to whole blood calibration methods and stable for the entire product dating.
- Study results show Cal-Chex CD Plus to be Conclusions Drawn From Tests: consistently reproducible and stable for the entire product dating. Cal-Chex CD Plus is a safe and effective alternative to Whole Blood Calibration when utilized as instructed in the product package insert.
{1}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, creating a sense of depth and unity.
## APR = 4 2002
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Paul Kittelson Quality Assurance/Regulatory Affairs Streck Laboratories, Inc. 7002 South 109th Street La Vista, Nebraska 68128
k020469 Re: Trade/Device Name: Cal-Chex CD Plus Regulation Number: 21 CFR § 864.8150 Regulation Name: Calibrator for Cell Indices Product Code: KRX Regulatory Class: II Regulation Number: 21 CFR § 864.8165 Regulation Name: Calibrator for Hemoglobin or Hematocrit Measurement Product Code: KRZ Regulatory Class: II Regulation Number: 21 CFR § 864.8175 Regulation Name: Calibrator for Platelet Counting Product Code: KRY Regulatory Class: II Regulation Number: 21 CFR § 864.8185 Regulation Name: Calibrator for Red Cell and White Cell Counting Product Code: KSA Regulatory Class: II Dated: February 8, 2002 Received: February 12, 2002
## Dear Mr. Kittelson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{2}------------------------------------------------
Page 2 - Mr. Paul Kittelson
or any Federal statutes and regulations administered by other Federal agencies. You must of any i citization and regisments, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI It Fat 607); adoming (21 CFR Part 820); and if applicable, the electronic forth in and quant) 1978. more (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will and w yourse finding of substantial equivalence of your device to a legally prematication of the results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of additionally 21 Of It Far 0588. Additionally, for questions on the promotion and advertising of Compinatee at (301) 597 - 1 the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). I other general information on your responsibilities under the Act may be obtained from the Other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
> Sincerely yours, Steven
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Page 1 of 1
510(k) Number:
K020469
Cal-Chex CD Plus
Device Name:
Indications For Use:
Cal-Chex CD Plus is used to calibrate the WBC, RBC, Hgb, MCV, PLT and MPV (MPV on CD4000 only) parameters on the Abbott Cell-Dyn 3200 and 4000 hematology instruments.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use V (Per 21 CFR 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Josephine Buntin
inical Laboratory Devices
510(k) Number K020469
Date:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.