ARM BLOOD PRESSURE MONITOR, MODELS KP-6821, KP-6822
K014204 · K-Jump Health Co., Ltd. · DXN · Mar 5, 2002 · Cardiovascular
Device Facts
Record ID
K014204
Device Name
ARM BLOOD PRESSURE MONITOR, MODELS KP-6821, KP-6822
Applicant
K-Jump Health Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Mar 5, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The Arm BPM is intended to measure the systolic, diastolic, and pulse rate (heart rate) by using an inflating cuff which is wrapped around the arm. The device is indicated for use in adults.
Device Story
Device measures systolic/diastolic blood pressure and pulse rate using oscillometric method; utilizes inflatable arm cuff, semiconductor sensor, internal air pump, and LCD display. Operated by user or clinician; cuff wrapped around upper arm; device inflates/deflates automatically to capture pressure oscillations. Output displayed on LCD; assists in monitoring cardiovascular status. Intended for adult use.
Clinical Evidence
Clinical trial conducted per ANSI/AAMI SP10A-1996 standard. 100 patients tested using both an Ausculatory Mercurial Sphygmomanometer and the Arm BPM to compare blood pressure and heart rate measurements.
Technological Characteristics
Non-invasive oscillometric blood pressure monitor. Components: inflatable cuff (Velcro), semiconductor sensor, internal air pump, battery, LCD. Standards: EN 60601-1-2, EN 55011, IEC 801-2, IEC 801-3, EN 1060-1, EN 1060-3, AAMI/ANSI SP10A-1996.
Indications for Use
Indicated for use in adults to measure systolic/diastolic blood pressure and pulse rate via an upper arm cuff.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
A & D Engineering Inc. Arm Blood Pressure Monitor Model UA-779 (K014204)
K-Jump Health Co., Ltd. Wristwatch Blood Pressure Monitor Model KP-6120 (K014204)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) SUMMARY
K-Jump's Arm Blood Pressure Monitor, Models KP-6821 and KP-6822 series
K014204
p. 1/2
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
MAR 5 2002
Daniel Tseng K-Jump Health Co., Ltd. No. 56 Wu Kung 5th Road Wu Ku Industrial Park Taipei Hsien Taiwan Phone: +886 2 22991378 Facsimile: +886 2 22331386
Date Prepared: November 5, 2001
### Name of Device and Name/Address of Sponsor
Arm Blood Pressure Monitor, Model KP-6821 and KP-6822
K-Jump Health Co., Ltd. No. 56 Wu Kung 5th Road Wu Ku Industrial Park Taipei Hsien Taiwan Phone: 011 886 2 22991378 Facsimile: 011 886 2 22331386 Contact Person: Daniel Tseng
| Common or Usual Name | Blood Pressure Monitor |
|----------------------|---------------------------------------------------|
| Classification Name | System, Measurement, Blood Pressure, Non-Invasive |
| Predicate Device | I. A & D Engineering Inc. |
| | Arm Blood Pressure Monitor Model UA-779 |
| | II. K-Jump Health Co., Ltd. |
| | Wristwatch Blood Pressure Monitor Model KP-6120 |
#### Intended Use
The Arm BPM is intended to measure the systolic, diastolic, and pulse rate (heart rate) by using an inflating cuff which is wrapped around the arm. The device is indicated for use in adults.
{1}------------------------------------------------
## Technological Characteristics
The Arm BPM is designed to measure the systolic, diastolic, and pulse rate (heart rate) of an individual. The device consists of an inflatable cuff that is wrapped around the arm and held in place with Velcro™, an LCD display, a semiconductor sensor, an internal air pump, a battery power source, and keys for operation.
1014204 p. 2/2
## Performance Data
The Arm BPM complies with EN 60601-1-2 (1993); EN 55011 (1991), IEC 801-2 (1991); IEC 801-3; EN 1060-1 and EN 1060-3 and the AAMI/ANSI SP10A-1996 standard. "Electronic or Automatic Sphygmomanometer."
K-Jump conducted a clinical trial in accordance with the ANSI/AAMI SP10A-1996 "Electronic or Automatic Sphygmomanometer" standard. One hundred patients had their blood pressure and heart rate tested with both an Ausculatory Mercurial Sphygmomanometer and the Arm BPM.
#### Substantial Equivalence
The Arm BPM is substantially equivalent to the A & D Engineering Inc.'s Arm BPM Model UA-779 and K-Jump's WristWatch BPM Model 6120. The Arm BPM and its predicate device are all noninvasive blood pressure monitors. Both the Arm BPM and A & D's Arm BPM take readings from the Upper Arm using the oscillometric method. The Arm BPM and A & D UA-779 have the same intended use, principles of operation and virtually identical technological characteristics. With the exception of minor variations in pressure measurement ranges, minimum operational relative humidity, minimum storage temperature and memory capacity, the devices are technologically identical. These minor differences do not raise any new issues of safety or effectiveness because both devices comply with the AAMI/ANSI SP10A-1996 standard "Electronic or Automated Sphygmomanometer".". Thus, the Arm BPM is substantially equivalent to the A & D's UA-779 and the WristWatch BPM. Further, the Arm BPM substantial equivalence is demonstrated by conformance to the ANSI/AAMI SP10A-1996 standard.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a black and white circular logo. The logo features the symbol of the Department of Health and Human Services (HHS) in the center. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 5 2002
K-Jump Health Co., Ltd. c/o Mr. Jonathan S. Kahan, Esq. Regulatory Counsel Hogan & Hartson L.L.P. 555 Thirteenth Street, N.W. Washington, DC 20004
Re: K014204
Trade Name: Arm Blood Pressure Monitor, Models KP-6821, KP-6822 Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: December 21, 2001 Received: December 21, 2001
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass stated in the encreated of the enactment date of the Medical Device Amendments, or to commerce prior to May 20, 2017, 11:24 accordance with the provisions of the Federal Food, Drug, de rices that have been recuire approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, mererere, maxis of the Act include requirements for annual registration, listing of general controlly profices of tactice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (600 a00 royals. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2 - Mr. Jonathan S. Kahan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that i Drines Internations administered by other Federal agencies. You must or any I catal button and including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI It Fat 607); adoning (21 OFF Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter will and in yourse FDA finding of substantial equivalence of your device to a legally promative noticated device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of additionally 21 301) 594-4646. Additionally, for questions on the promotion and advertising of Compinance at (301) 59 + + the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Oiner general millimantes turers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
signature
Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use Form
510(k) Number (if known): _
Device Name: K-JUMP Health Co., Ltd. Arm Blood Pressure Monitor, Models KP-6821 and KP-6822 series
Indication for Use:
The K-Jump Arm Blood Pressure Monitor, Model KP-6821 AND KP-6822 are intended to measure the systolic, diastolic, and pulse rate (heart rate) by using an inflating cuff which is wrapped around the upper arm. The device is indicated for use in adults.
(PLEASE DO NOT WRITE BELOW THIS LINE---CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Cardiovascular & Respiratory Devices
510(k) N:
2014 207
.
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.