HOMMED SENTRY III PATIENT MONITOR SYSTEM WITH CARD READER
Applicant
Hommed, LLC
Product Code
DQA · Cardiovascular
Decision Date
Apr 9, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The HomMed Sentry III Patient Monitor System with Card Reader is intended for in home and/or healthcare facility applications under physician orders. The Card Reader with the HomMed Sentry III Patient Monitor System allows patients to use multiple Sentry III monitors or multiple patients to use the same Sentry III monitor. The use of the system is to allow retrospective review of certain patient physiological functions. The HomMed Sentry III Patient Monitor System with Card Reader can measure and display patient data including noninvasive blood pressure, pulse rate, oral temperature, oximetry, and weight. Additionally, the patient vital signs data can be transmitted to a central review station via a pager network with a backup landline telephone modem for telephone communication with the central pager network if necessary.
Device Story
Portable vital signs monitor for home/clinical use; measures NIBP, pulse rate, SpO2, oral temperature, weight, and optional glucometry/spirometry. Inputs: NIBP cuff, pulse oximetry probe, electronic thermometer, electronic scale. Device acquires, displays, and transmits data via internal communication module over Skytel Pager Network to central station. Enables retrospective review/analysis by healthcare professionals. Card reader allows multi-user/multi-monitor flexibility. Benefits: facilitates remote patient monitoring, enabling physician oversight of physiological trends. Operated by patients or healthcare providers.
Clinical Evidence
Bench testing only. Testing included EMC, electrical safety, mechanical durability, and environmental (temperature/humidity) testing. Performance confirmed consistent with FDA reviewer's guides for respiratory devices and electronic thermometers. No clinical trial data provided.
Technological Characteristics
System includes NIBP monitor, pulse oximeter, electronic thermometer, electronic scale, and card reader. External medical-grade power supply. Standards: EN 60601-1 (Safety), IEC 601-1-2 (EMC), ISO 10993-5, 10-11 (Biocompatibility). Connectivity via internal communication module to pager network/landline modem.
Indications for Use
Indicated for pediatric to adult patients in home or clinical settings for monitoring vital signs including NIBP, pulse rate, SpO2, oral temperature, weight, and optional glucometry/spirometry. Requires physician order.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
HomMed Sentry III Patient Monitor System (K014025)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary
# HomMed Sentry III Patient Monitor System
Contact:
Herschel Peddicord, President HomMed, LLC 19275 West Capitol Dr., Suite 200 Brookfield . WI 53045 262 783-5440 Voice 262 783-5441 Fax
HomMed Sentry III Patient Monitor System with Card Reader Trade Name:
Patient Vital Signs Monitor with Card Reader Common Name:
Classification Name: Oximeter
Substantial Equivalence is Claimed to:
HomMed Sentry III Patient Monitor System
The HomMed Sentry III Patient Monitor System with Card Reader is a portable patient Device Description: vital signs monitoring system which allows patients to use multiple Sentry III monitors or multiple patients to use the same Sentry III monitor. The system measures noninvasive blood pressure, pulse rate, oral temperature, oximetry, and weight. ไท addition, the system has optional glucometry and spirometer measuring capabilities. The Sentry III Patient Monitor with Card Reader acquires the patient vital signs data and displays it. The data can also be transmitted via the communication system through the Skytel Pager Network to a central station for storage with retrospective display and analysis. The Card Reader with the HomMed Sentry III Patient Monitor System allows patients to use multiple monitors or multiple patients to use the same Sentry III monitor.
The HomMed Sentry III Patient Monitor System with Card Reader is intended for in Indications for Use: home and/or healthcare facility applications under physician orders. The Card Reader with the HomMed Sentry III Patient Monitor System allows patients to use multiple Sentry III monitors or multiple patients to use the same Sentry III monitor. The use of the system is to allow retrospective review of certain patient physiological functions. The HomMed Sentry III Patient Monitor System with Card Reader can measure and display patient data including noninvasive blood pressure, pulse rate, oral temperature, oximetry, and weight. Additionally, the patient vital signs data can be communicated to a central review station via a pager network with a backup landline telephone modem for telephone communication with the central pager network if necessary.
{1}------------------------------------------------
The HomMed Sentry III Patient Monitor System with Card Reader provides a noninvasive blood pressure (NIBP) monitor for measurements of a patient's systolic, diastolic, and mean arterial (MAP) blood pressures; pulse oximeter, acquires a pulse rate using an oximeter: oral temperature via an electronic thermometer; weight from an electronic scale. All data collected from these functions as well as optional glucometry and spirometry is sent through an internal communication module.
The device will provide fast, reliable measurements on patients ranging from children (pediatrics) to adults when using the appropriate blood pressure cuff. The pulse oximetry works with Sentry III Patient Monitor System pulse oximetry probes provided by HomMed, providing Sp02 and pulse rate on all patients from pediatric to adult. The electronic thermometry requires use of the Welch Allyn oral thermometry probe and probe covers. It provides only oral temperature information. The device is intended for use in the patient home and/or clinical environments by the patient as prescribed by or on orders by a physician with the information transmitted to a central viewing station where healthcare professionals can review the data.
### Comparison with Predicate Devices:
This HomMed Sentry III Patient Monitor System allows uncomplicated measurement and remote monitoring of patient vital signs including weight utilizing the existing technologies of the predicate device, HomMed Sentry III Patient Monitor System.
### Determination of Substantial Equivalence:
The performance of each component of the HomMed Sentry III Patient Monitor System with Card Reader has been confirmed to be equivalent to the predicate device HomMed Sentry III Patient Monitor System. In addition, the HomMed device continues to utilize an external medical grade power supply ensuring the continued protection and safety for the patient vital signs monitor, scale, and card reader and communication module.
#### Compliance to Standards and Regulations:
The HomMed Model Sentry III Patient Monitor System with Card Reader complies with the following national and international standards:
Safety
EN 60601-1 IEC 601-1-2 ISO 10993-5,10-11
Medical Electrical Safety EMC Compliance Biocompatibility
Performance Data:
The HomMed Sentry III Patient Monitor System with Card Reader utilizes the HomMed Sentry III Patient Monitor System within the environments for which Sentry III is marketed. The Sentry III Patient Monitor System with Card Reader performs consistent with guidelines and standards found in the FDA reviewer's guides for respiratory devices and electronic thermometers. EMC, electrical, mechanical durability, safety (operator and patient), and temperature/humidity testing has been completed demonstrating compliance with applicable standards. The test results demonstrated that the Sentry III System with Card Reader is in compliance with the guidelines and standards referenced in the FDA reviewer's guides and that it performed within its specifications and functional requirements.
{2}------------------------------------------------
The HomMed Sentry III Patient Monitor System performance is consistent with the HomMed Sentry III System performance and additional testing has been done on the HomMed Sentry III Patient Monitor System with Card Reader assuring compliance with applicable electrical, safety and healthcare standards. Thus it is the HomMed position that the HomMed Sentry III Patient Monitor System with Card Reader performs as well as the legally marketed predicate device.
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of FDA regarding patient monitors.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 00 Corporate Boulevard ockville MD 20850
HomMed, LLC c/o Tommie J. Morgan, Ph.D. President Morgan Consultants, Inc. 2018 North Durham Drive Houston, TX 77008
Re: K014025
Trade Name: HomMed Sentry III Patient Monitor System with Card Reader Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II (two) Product Code: DQA Dated: March 6, 2002 Received: March 8, 2002
Dear Dr. Morgan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Tommie J. Morgan, Ph.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I toure of active a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I occur bakated and sequirements, including, but not limited to: registration and listing (21 Comply with an the Fee 811 (21 CFR Part 801); good manufacturing practice requirements as set OFF Part 877) regulation (21 CFR Part 820); and if applicable, the electronic form and quinn control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I mb letter will and the FDA finding of substantial equivalence of your device to a legally premated notificated on "ceresults in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). registeral information on your responsibilities under the Act may be obtained from the Oixision of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K014025
510(k) Number (if known):
Device Name: HomMed Sentry III Patient Monitor System with Card Reade
## Indications for Use
The HomMed Sentry III Patient Monitor System with Card Reader is designed to measure Patient Vital Signs in the home by patients or in clinical environments by health care providers. The HomMed Sentry, III Patient Monitor System with Card Reader is available with physicians' orders only. The HomMed Sentry III Patient Monitor System with Card Reader allows patients to use multiple Sentry III monitors or multiple patients to use the same Sentry III monitor.
The HomMed Sentry III Patient Monitor System with Card Reader measures the following Non-Invasive Blood Pressures (Systolic, Diastolic and Mean Arterial parameters: Pressure), Functional Oxygen Saturation (%SpO2), Peripheral Pulse Rate (PPR), Pulse Strength, Oral Temperature and Patient Weight via an external scale. The HomMed Sentry III Patient Monitor System with Card Reader's optional, compatible devices extends those measurements to glucometer and spirometer monitoring. The patient parameter data is collected and displayed by the HomMed Sentry III Patient Monitor System with Card Reader. Data can be transmitted via the communication module to a central station where the patient data can be viewed and analyzed.
(PLEASE DO NOT WRITE BELOW THIS LINE - - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH/Office of Device Evaluation (ODE)
| Division of Cardiovascular & Respiratory Devices |
|--------------------------------------------------|
|--------------------------------------------------|
| 510(k) Number | K014023 |
|---------------|---------|
|---------------|---------|
| Prescription Use<br>Per 21 CFR 801.109) | X |
|-----------------------------------------|---|
|-----------------------------------------|---|
OR
| Over-the-Counter Use | |
|----------------------|--|
|----------------------|--|
HomMed, LLC C:\My Documents\Sentry H1+\FDA 510k Submission SH1.11-25-01.doc
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.