REUSE CLAIM FOR THE BARD ELECTROPHYSIOLOGY SURELINK, TEMPLINK, AND TEMPLINK M EXTENSION CABLES
K013598 · C.R. Bard, Inc. · DSA · Jan 11, 2002 · Cardiovascular
Device Facts
Record ID
K013598
Device Name
REUSE CLAIM FOR THE BARD ELECTROPHYSIOLOGY SURELINK, TEMPLINK, AND TEMPLINK M EXTENSION CABLES
Applicant
C.R. Bard, Inc.
Product Code
DSA · Cardiovascular
Decision Date
Jan 11, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2900
Device Class
Class 2
Indications for Use
The SureLink extension cable is indicated for use during electrophysiology studies in conjunction with the appropriate electrode catheter. This cable may be reused subject to the cleaning and sterilization restrictions herein. The TempLink extension cable when used in conjunction with a thermistor configured Stinger/Stinger S catheter is indicated for use during cardiac ablation with set power to 50W. The TempLink M extension cable when used in conjunction with a thermocouple configured Stinger M/Stinger SM catheter is indicated for use during cardiac ablation with set power to 50W. These cables may be reused subject to the cleaning and sterilization restrictions herein.
Device Story
SureLink, TempLink, and TempLink M are extension cables connecting diagnostic or ablation catheters to recording equipment. Cables provide additional length (7-10 feet) and allow the recorder junction box to remain outside the sterile field. Operated by clinicians in electrophysiology labs. Cables transmit electrical signals between catheters and recorders. Subject device introduces a reuse claim, supported by cleaning and sterilization protocols (EO, steam autoclave, Sterrad). Failure of the cable results in physician inconvenience without direct patient impact, as cables are not intravascular and have no patient body contact.
Clinical Evidence
Bench testing only. Performance data provided to demonstrate that cables continue to perform as intended after the specified number of cleanings and resterilizations.
Technological Characteristics
Extension cables for EP catheters. Materials: copper wire (internal conductor). Connectivity: electrical interface between catheters and recording equipment. Sterilization: validated for EO, steam autoclave, and Sterrad. Design: 7-10 foot length. Non-patient-contact accessory.
Indications for Use
Indicated for patients undergoing electrophysiology studies (SureLink) or cardiac ablation (TempLink/TempLink M) when used with compatible electrode catheters. TempLink requires thermistor-configured Stinger/Stinger S catheters; TempLink M requires thermocouple-configured Stinger M/Stinger SM catheters. Ablation power limited to 50W.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
{0}------------------------------------------------
Bard Electrophysiology C.R. Bard, Inc. 55 Technology Drive Lowell, MA 01851 (978) 441-6202 BardEP@CRBARD.COM KO 3598 9.115
JAN 1 1 2002
BARD
### VI. 510(k) SUMMARY: FOR THE SURELINK, TEMPLINK, AND TEMPLINK M EXTENSION CABLES
This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990.
A. Submitter's Information
Name:
Address:
Phone: Fax: Email address: Contact Person: Bard electrophysiology Division of C.R. Bard, inc. 55 Technology Drive Lowell, MA 01851 978-323-2216 (Direct Line) 978-323-2222 deborah.herrington@crbard.com Deborah L. Herrington Manager, Regulatory Affairs October 30, 2001
Date of Preparation:
#### B. Device Name
Trade Name:
Common/Usual Name: Classification Name:
SureLink Extension Cable TempLink Extension Cable TempLink M Extension Cabie Electrode cable (including connector) Patient Transducer and Electrode Cable (including connector)
C. Predicate Device Names(s)
> SureLink Extension Cable TempLink Extension Cable TempLink M Extension Cable
510(k) Premarket Notification for the Reuse Claim for the SureLink, TempLink, and TempLink M Extension Cables Bard electrophysiology Division of C.R. Bard, Inc.
Image /page/0/Picture/19 description: The image shows the logo for Bard Electrophysiology. The logo consists of a stylized heart shape enclosed within a square on the left, followed by the word "Bard" in a bold, sans-serif font. Below "Bard" is the word "electrophysiology" in a smaller, italicized font. The logo is black and white.
{1}------------------------------------------------
### Device Description/Indications for Use D.
# Description
The SureLink, TempLink, and TempLink M Extension Cables are accessories that are used in conjunction with the appropriate diagnostic or ablation catheter. Cables, when initially introduced, provided an alternate means of electrically connecting EP catheters to recorders versus the catheter connecting directly to the recorder. The use of cables allows the catheter to remain in the sterile field, the recorder junction box doesn't have to enter the sterile field, and the cable provides more working length. The cables are offered in several lengths ranging from 7 to 10 feet in length.
## Indications
# SureLink Extension Cables
The SureLink extension cable is indicated for use during electrophysiology studies in conjunction with the appropriate electrode catheter. This cable may be reused subject to the cleaning and sterilization restrictions herein.
# TempLink/TempLink M Extension Cables
The TempLink extension cable when used in conjunction with a thermistor configured Stinger/Stinger S catheter is indicated for use during cardiac ablation with set power to 50W.
The TempLink M extension cable when used in conjunction with a thermocouple configured Stinger M/Stinger SM catheter is indicated for use during cardiac ablation with set power to 50W.
These cables may be reused subject to the cleaning and sterilization restrictions herein.
> 510(k) Premarket Notification for the Reuse Claim for the SureLink, TempLink, and TempLink M Extension Cables Bard electrophysiology Division of C.R. Bard, Inc.
(013598
{2}------------------------------------------------
### Technological Characteristics/Performance Data Summary ய்
The "510(k) Substantial Equivalence Decision-Making Process (Detailed)" decision tree (CDRH 510(k) Manual 92-4158) was utilized to make a determination of substantial equivalence. This decision tree is depicted in Figure IV-1 with the decision points relevant to the SureLink, TempLink, and TempLink M cables indicated by highlighting. The answers to the following questions at the indicated decision points in Figure IV-1 lead to a determination of substantial equivalence.
## New Device is Compared to a Marketed Device?
Yes. The SureLink, TempLink, and TempLink M Extension cables with the reuse claim are compared to their no reuse claim counterparts. The only change to the indications for use statement is the addition of the reuse claim subject to the cleaning and sterilization restrictions provided in the instructions for use.
## Does the New Device have the Same Indications Statement?
No. The indication statement for the SureLink cables, subject of this 510(k), and that of the predicate device are similar. Both are indicated for use during electrophysiology studies in conjunction with the appropriate electrode catheter. The SureLink cable, subject of this 510(k), is additionally indicated for reuse subject to the cleaning and sterilization restrictions included in the instructions for use (IFU).
The indication statement for the TempLink cables, subject of this 510(k), and that of the predicate device are similar. When used in conjunction with a thermistor configured Stinger / Stinger S catheter the TempLink cable is indicated for use during cardiac ablation with set power to 50W.
The indication statement for the TempLink M cable, subject of this 510(k), and that of the predicate device are similar. When used in conjunction with a thermocouple configured Stinger M catheter the TempLink M cable is indicated for use during cardiac ablation with set power to 50W.
> 510(k) Premarket Notification for the Reuse Claim for the SureLink, TempLink, and TempLink M Extension Cables Bard electrophysiology Division of C.R. Bard, Inc.
{3}------------------------------------------------
The TempLink/TempLink M cables, subject of this 510(k), are additionally indicated for reuse subject to the cleaning and sterilization restrictions included in the IFU.
# Do the Differences Alter the Intended Therapeutic/Diagnostic/etc. Effect ?
No. The cables with the reuse claim continue to be used as originally described. These cables, like their respective predicate devices, are not placed intravascularly and have no patient body contact. These cables are accessory devices that continue to be used in conjunction with their respective therapeutic or diagnostic catheters.
# New Device Has Same Intended Use and May Be "Substantialy Equivalent"
# Does New Device Have the Same Technological Characteristics, e.g., Design, Materials, etc.?
No. The SureLink, TempLink, and TempLink M cables are similar to their predicate devices. The design of each cable remains unchanged. The internal tail component was originally made from tinsel wrapped kevlar, but was replaced with standard copper wire that is easier to work with in manufacturing, more readily available, and found equivalent to the kevlar material. With respect to the reuse claim for EO, steam autoclave, and Sterrad sterilization, data to support this claim is provided in the Appendices.
# Could the New Characteristics Affect Safety or Effectiveness?
No. The addition of a reuse claim could effect safety and/or effectiveness for a number of devices. However, these cables are accessories to their respective catheters and are not placed intravascularly, nor do they have patient body contact. Regarding effectiveness, failure of the cable to not perform as intended would result in physician inconvenience and have no negative impact on the patient.
### Are the Descriptive Characteristics Precise Enough to Ensure Equivalence?
No. Although the design of the cables has not changed, and the materials of construction are similar (change in intemal tail component from tinsel wrapped kevlar to copper wire ), data are required to support the reuse claim.
> 510(k) Premarket Notification for the Reuse Claim for the SureLink, TempLink, and TempLink M Extension Cables Bard electrophysiology Division of C.R. Bard, Inc.
ી 3598
{4}------------------------------------------------
# Are Performance Data Available to Assess Equivalence?
Yes. Performance data is provided in the Appendices with respect to the reuse claim.
## Performance Data Demonstrate Equivalence?
Yes. The performance data demonstrates that the SureLink, TempLink, and TempLink M cables continue to perform as intended after the specified number of cleanings and resterilizations as described in the instructions for use.
Therefore, the SureLink, TempLink, and TempLink M cables with the reuse claim are "Substantially Equivalent" to their respective predicate devices.
## SUBSTANTIALLY EQUIVALENT DETERMINATION:
The determination of substantial equivalence is based on the decision tree found in Figure IV-1, the safety information provided in Section IV.C, the effectiveness results regarding cable performance summarized in Section IV, and the actual data provided in the Appendices. Refer to the table under the tab "Appendices" for the actual information provided.
> 510(k) Premarket Notification for the Reuse Claim for the SureLink, TempLink, and TempLink M Extension Cables Bard electrophysiology Division of C.R. Bard, Inc.
KO13598
9.5/5
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three wavy lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 1 2002
Ms. Deborah L. Herrington Manager. Regulatory Affairs C.R. Bard, Inc. Bard Electrophysiology 55 Technology Drive Lowell, MA 01851
Re: K013598
Trade Name: Reuse Claim for the Bard Electrophysiology SureLink, TempLink, and TempLink M Extension Cables Regulation Number: 21 CFR 870.2900 Regulation Name: Patient Transducer and Electrode Cable Regulatory Class: Class II (two) Product Code: DSA Dated: October 30, 2001 Received: October 31, 2001
Dear Ms. Herrington:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
# Page 2 - Ms. Deborah L. Herrington
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
hs. X. O'Dell
ram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# D. INDICATIONS FOR USE
K013598
SureLink Extension Cables Device Name:
### Indications for Use:
indications for USC.
The SureLink extension cable is indicated for use during electrophysiology studies in The Sulecins extension cable is indicator for ass authig cross in process project to the cleaning and sterilization restrictions herein.
TempLink/TempLink M Extension Cables Device Name:
### Indications for Use:
indications for Osc.
The TempLink extension cable when used in conjunction with a thermistor configured The Tempelline exceller cable when as a se during cardiac ablation with set power to 50W.
The TempLink M extension cable when used in conjunction with a thermocouple configured The TempEllik in Oxlondon Cable Wee during cardiac ablation with set power to 50W.
These cables may be reused subject to the cleaning and sterilization restrictions herein.
Contraindications: Contraindications for the SureLink, TempLink, and TempLink M Extension cables.
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Prescription Use**
(Per 21 CFR 801.109)
Over-the-Counter Use _
Division of Cardiovascular & Regulatory Devices
510(k) Number K013398
OR
510(k) Premarket Notification for the Reuse Claim for the SureLink, TempLink, and TempLink M Extension Cables Bard electrophysiology Division of C.R. Bard, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.