K013424 · Hand Innovations, Inc. · HTY · Dec 20, 2001 · Orthopedic
Device Facts
Record ID
K013424
Device Name
INTRAMEDULLARY FIXATION SYSTEM FOR THE HAND
Applicant
Hand Innovations, Inc.
Product Code
HTY · Orthopedic
Decision Date
Dec 20, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Intramedullary Fixation System is indicated for the fixation of extra-articular fractures of the long bones of the hand including the metacarpals and the proximal and middle phalanges. The optional locking device may be used to minimize axial and rotational motion of the implantable nail during healing.
Device Story
Intramedullary Fixation System (IFS) for hand fractures; sterile, single-use, disposable. Components: slotted awl assembly, implantable nail handle assembly, exchange guide, bend tube. Procedure: awl creates metacarpal access; nail advanced distally; nail trimmed and bent 90° at insertion site; remains implanted ~6 weeks; percutaneously removed. Optional locking device (stainless steel cannula/polymeric handle) minimizes axial/rotational motion; advanced through cortical perforation into metaphysis; ratchet engagement provides tactile feedback. Used by surgeons in clinical settings to stabilize fractures, facilitating healing.
Clinical Evidence
Bench testing only. Nine bench tests performed including stiffness/yield, cutting geometry comparison, and drilling tests. Materials selected based on long history of biocompatibility and compliance with ASTM standards (F138, F899, F86, F366). No clinical data presented.
Indicated for fixation of extra-articular fractures of long bones of the hand, specifically metacarpals, proximal phalanges, and middle phalanges.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Intramedullary Fixation System for the Hand (K991064)
Reference Devices
K-wire (MicroAire Surgical Equipment)
Submission Summary (Full Text)
{0}------------------------------------------------
510(k), Intramedullary Fixation System for the Hand (IFS)
Fixation System for the Hand
Submitted by
DEC 2 0 2001
Hand Innovations, Inc. 8905 SW 87th Avenue Suite 100 Miami, FL 33176-2227 Phone: (305) 412-8010 Fax: (305) 412-8060
| Contact Person: | Al Weisenborn |
|----------------------|---------------------------------------------------------------------------|
| Device Trade Name: | Intramedullary Fixation System for the Hand |
| Common Name: | K-Wire |
| Classification Name: | Smooth or threaded metallic bone fixation fastener, per 21 CFR § 888.3040 |
#### Identification of a Legally Marketed Predicate Device
The Intramedullary Fixation System for the Hand is substantially equivalent to the Intramedullary Fixation System for the Hand that is legally manufactured by Medcanica and marketed pursuant to premarket notification K991064.
#### Device Description
The Intramedullary Fixation System for the Hand is a sterile, single use, disposable device that is delivered non-toxic. The Intramedullary Fixation System for the Hand consists of the Slotted Awl Assembly, the Implantable Nail Handle Assembly, and Exchange Guide and Bend Tube.
Prior to use the implantable nail assembly is nested in the slotted awl assembly. The slotted awl assembly has a trocar point. The implantable nail has a blunt point that is positioned just behind the trocar point of the slotted awl. The sharp point of the slotted awl assembly is passed through a small incision. A hole is drilled into the metacarpal bone by twisting the assembled handles back and forth. After gaining access to the intramedullary space, the slotted awl handle is held stationary while the implantable nail is then advanced distally from the base of the metacarpal bone.
The awl handle is then withdrawn and removed for advancement of the implantable nail. The implantable nail is then cut adjacent to the nail handle. Using the bending tube end of the exchange guide and bend tube the implantable nail is bent to 90° with the apex of the
{1}------------------------------------------------
bend at the implantable nail insertion site. The nail is trimmed so that the end is below the skin. The small piece remaining will facilitate removal of the implantable nail subsequent to healing. The implantable nail will remain implanted for approximately six weeks. Upon healing of the fracture, the implantable nail is percutaneously removed.
In the event that it is desired to reform the implantable nail or implant a smaller nail, this may be accomplished without loosing access to the medullary canal. The exchange guide is advanced along the implantable nail into the medullar space. Once the medullar space is accessed, the nail is removed. Another nail may be placed into the medulla by inserting it into the groove of the exchange guide. After the nail has been inserted into the medullar space, remove the exchange guide.
An optional locking device may be used to minimize rotation of the implantable nail. The device consists of a pointed stainless steel cannula mounted to a polymeric handle. After the implantable nail is bent to a 90-degree angle, the locking sleeve is positioned over the end of the implantable nail and manually advanced downward through the cortical perforation and into the metaphysis. The locking device is then advanced until tactile feedback confirms ratchet engagement. The locking device may be further advance to the desired depth. When resistance is felt, the locking nail is impacted into its final position with a few sharp taps. The nail and locking sleeve are simultaneously trimmed.
# Intended Use
The Intramedullary Fixation System is indicated for the fixation of extra-articular fractures of the long bones of the hand including the metacarpals and the proximal and middle phalanges. The optional locking device may be used to minimize axial and rotational motion of the implantable nail during healing.
# Summary of Technological Characteristics
Twelve (12) technological characteristics of the Intramedullary Fixation System for the Hand were compared to the K-wire manufactured by MicroAire Surgical Equipment and found to be equivalent.
### Summary of Performance Data
The Intramedullary Fixation System for the Hand meets the requirements of the following recognized consensus standards.
- ASTM F138 97, Standard Specification for Wrought 18 Chromium-14 Nickel-. 2.5 Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673)
- ASTM F899 95, Standard Specification for Stainless Steel Billet, Bar and Wire . for Surgical Instruments
- ASTM F86 91, Standard Practice for Surface Preparation and Marking of Metal-. lic Surgical Implants
{2}------------------------------------------------
- ASTM F366 82 (Reapproved 1993), Standard Specification for Fixation Pins . and Wires
Additionally, nine (9) bench tests were performed to establish the safety and efficacy of the device.
Bench testing of both devices also demonstrated equivalence. Testing included stiffness/vield. cutting geometry comparison, and drilling test.
The tissue/bone contact materials of the device have been carefully selected for their long history of biocompatibility. The materials meet the requirements of the previously referenced recognized consensus standards.
Since the Intramedullary Fixation System for the Hand meets the requirements of the stated standards and embodies technological characteristics essentially identical to the predicate device, we believe the device is safe and effective and performs as well as or better than the predicate device. The Intramedullary Fixation System for the Hand was designed utilizing design controls compliant with the Quality System Regulation. The Intramedullary Fixation System for the Hand will be manufactured per specifications and good practices that ensure the device is safe and effective for its intended use.
K013424/
page 30 of 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three stripes representing the feathers. The eagle is facing to the left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 0 2001
Mr. Al Weisenborn Hand Innovations, Inc. 8905 SW 87th Avenue, Suite 100 Miami, Florida 33176-2227
Re: K013424
Trade/Device Name: Intramedullary Fixation System for the Hand Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HTY Dated: October 9, 2001 Received: October 16, 2001
Dear Mr. Weisenborn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require with the proval of a proval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Al Weisenborn
This letter will allow you to begin marketing your device as described in your Section This letter will and if your your the FDA finding of substantial equivalence of your J ro(k) promation noarketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 If you dosire uposition in the diagnostic devices), please contact the alle additionally 21 Of Crick 659. Additionally, for questions on the promotion Onloo of Coing of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, (Misbranding by reference to prematice) 4057. Thiso, promo note are rogg). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 6382041 or (301) 4436597 or at its Internet address HYPERLINK http://www.fda.gov/cdrh/dsma/dsmamain.html http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
signature
Celia M. Witten, Ph.D., M.D Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# The Indications for Use
Page 1 of 1
KO1 3424 510(k) Number (if known):
Intramedullary Fixation System for the Hand Device Name:
Indications for Use:
The Intramedullary Fixation System is indicated for the fixation of extra-articular fractures of the long bones of the hand including the metacarpals and the proximal and middle phalanges. The pars and one ing device may be used to minimize optional rotational motion of the implantable nail during healing.
signature
Division Sign-Off Division of General, Restorative and Neurological Devices
510(k) Number.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use_
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.