The Hygia Health Services Reprocessed Huntleigh Foot Wrap operates in the identical manner as the predicate device, the Huntleigh Foot Wrap Garment. It is designed to apply compression to a patient's plantar plexus for the prevention of both deep vein thrombosis (DVT) as well as the treatment of edema secondary to venous insufficiency. The devices are used in both the home and institutional settings on patient populations for which a leg or calf compression device would not be applicable.
Device Story
Compressible limb sleeve device; placed around patient's plantar arch; connected to external pneumatic controller. Device consists of brushed nylon, polyester foam lining, and internal polymer coating for airtight bladder; PVC inflation/deflation tube; polyethylene hook fasteners. Controller provides intermittent pneumatic compression; inflates/deflates bladder in predetermined cycles. Compression of plantar plexus collapses veins, increasing venous pressure to eject blood upward; deflation allows venous refill. Used in home and institutional settings; operated by clinicians or patients. Reprocessed via thermal kill pasteurization to enable multi-patient use. Output is mechanical compression; assists in DVT prevention and edema treatment.
Clinical Evidence
Bench testing only. Comparative bench testing assessed functional and operational performance, including inflation/deflation timing sequences, safety parameters, and control connectivity. Results demonstrated substantial equivalence to the predicate device.
Technological Characteristics
Brushed nylon garment with polyester foam lining; internal polymer coating for bladder; PVC tubing; polyethylene hook fasteners. Operates via intermittent pneumatic compression using ambient air. Powered by external controller. No software/electronics in the garment itself.
Indications for Use
Indicated for patients requiring enhanced blood circulation in the lower extremities to prevent deep vein thrombosis, reduce wound healing time, treat venous leg ulcers, reduce edema from venous insufficiency, and decrease compartmental pressures. Contraindicated for patients with congestive heart failure, known/suspected DVT, severe arteriosclerosis/ischemic vascular disease, or local leg conditions (dermatitis, gangrene, recent skin grafts, untreated infected wounds).
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Hygia Health Services. On the left is a black square with three white concentric circles inside. To the right of the square is the word "HYGIA" in large, bold, black letters, with the words "HEALTH SERVICES" in smaller letters underneath.
K012651
8-1
APR 1 2 2002
# 510(k) Summary
## Summary
## Substantial Equivalence Summary for the Hygia Health Services Reprocessed Huntleigh Foot Wrap.
In accordance with 21 CFR Part 807.92, this summary is submitted by:
Hygia Health Services, Inc. 2800 Milan Court Suite 259 Birmingham, Alabama 35211
Date: August 6, 2001
】. Contact Person
> Geoff M. Fatzinger Director, Compliance and Regulatory Affairs (205) 943-6670
#### 2. Name of Device
Classification Name: Compressible Limb Sleeve Common or Usual Name: Intermittent Pneumatic Compressible Limb Sleeve Review Panel: Cardiovascular Classification: Class II Proprietary Name: Hygia Health Services Reprocessed Huntleigh Foot Wrap.
### HYGIA HEALTH SERVICES
M | L A N | C O U R T 2 8 0 0 2 0 5 - 9 4 3 - 6 6 7 0 V O | C E ・ 2 0 5 - 9 4 3 - 6 6 7 1 F A X
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains a logo for Hygia Health Services. The logo features a stylized symbol on the left, resembling interconnected circles within a square. To the right of the symbol, the word "HYGIA" is written in a bold, sans-serif font, and below it, the words "HEALTH SERVICES" are written in a smaller, lighter font.
#### Predicate Device 3.
Classification Name: Compressible Limb Sleeve Intermittent Pneumatic Compressible Limb Sleeve Common Name: Class II Classification: Huntleigh Foot Wrap Garment Proprietary Name:
#### Statement of Substantial Equivalence 4.
The Hygia Health Services Reprocessed Huntleigh Foot Wrap employs no new technology other than the method used to reprocess the garment in order to allow the device to be utilized more than once. The Hygia Health Services Reprocessed Huntleigh Foot Wrap is substantially equivalent to the Huntleigh Foot Wrap Garment in that the basis of operation for both of the devices is the intermittent inflation of a single bladder, which is placed around the patient's plantar arch. The sleeves are then connected to a controller. Inflation of the device is accomplished using ambient air, and a controller cycle that functions to alternately inflate and deflate the sleeves in a predetermined manner and interval.
The Hygia Health Services Reprocessed Huntleigh Foot Wrap is substantially equivalent in function, operating parameters, and intended use to the Huntleigh Foot Wrap Garment that is currently commercially available and in distribution. The predicate device, the Huntleigh Foot Wrap, is marked for "single-patient use only". Hygia Health Services does not change the device in any way except to render the device "reusable" by placing it through a scientifically validated thermal kill pasteurization process. The Hygia Health Services HLD protocol does not alter the device's efficacy, safety, composition, or intended use.
#### Description of the Device న.
The Hygia Health Services Reprocessed Huntleigh Foot Wrap Garment is a compressible limb device that is placed around the patient's foot and when attached to an approved controller, provides intermittent pneumatic compression. The garment is constructed out of brushed nylon with a polyester foam lining. The single bladder is constructed out of the garment itself using an internal polymer coating that provides an airtight surface. The garment does not use an inflation chamber of poly vinyl chloride (PVC). The inflation/deflation tube is however, composed of PVC. The hook fasteners are made of polyethylene. The tube terminates in a snap lock connector. The garments are placed around the foot at the area of the plantar arch and secured with a hook and loop fastener. As the garment compresses the plantar plexus, the veins collapse longitudinally, which increases the venous pressure thus ejecting the blood upward. After compression. the devices deflate allowing the veins to refill and bring oxygenated blood to the
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Hygia Health Services. The logo consists of a black square with three white concentric circles on the left, followed by the word "HYGIA" in large, bold, black letters. Below the word "HYGIA" are the words "HEALTH SERVICES" in smaller, thinner, black letters.
lower limbs. The controller predetermines the inflation and deflation sequence. The pressure of compression is determined by the controller and is adjusted by altering the readout on the controller.
#### Intended Use of Device 6.
The Hygia Health Services Reprocessed Huntleigh Foot Wrap operates in the identical manner as the predicate device, the Huntleigh Foot Wrap Garment. It is designed to apply compression to a patient's plantar plexus for the prevention of both deep vein thrombosis (DVT) as well as the treatment of edema secondary to venous insufficiency. The devices are used in both the home and institutional settings on patient populations for which a leg or calf compression device would not be applicable.
#### Technological Characteristics 7.
The technological characteristics of the Hygia Health Services Reprocessed Huntleigh Foot Wrap are identical to the original Huntleigh Foot Wrap Garment in overall design, materials, energy source, mode of operation, and performance characteristics.
#### Performance Data 8.
Nonclinical Tests- Comparative bench testing was utilized to assess and prove similarity of function between the Hygia Health Services Reprocessed Huntleigh Foot Wrap and the predicate device, the Huntleigh Foot Wrap Garment. All tests found that functional and operational performance characteristics including round that renotional acontrol, timing sequence, and both safety and operational parameters used when connected to a controller were substantially equivalent.
Test Conclusions- Nonclinical test results of the Hygia Health Services Reveocessed Huntleigh Foot Wrap indicated substantial equivalence in all aspects to the predicate device, the Huntleigh Foot Wrap Garment.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## APR 1 2 2002
Mr. Geoff M. Fatzinger, BS MS Director, Compliance and Regulatory Affairs c/o Hygia Health Services, Inc. 2800 Milan Court, Suite 259 Birmingham, AL 35211
K012651 Re:
Trade Name: Hygia Health Services Reprocessed Huntleigh Foot Wrap Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II (two) Product Code: JOW Dated: February 7, 2002 Received: February 8, 2002
Dear Mr. Fatzinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
## Page 2 - Mr. Geoff M. Fatzinger, BS MS
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quians by some (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter wifi anow your to FDA finding of substantial equivalence of your device to a legally prematicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of Compliation at (301) 29 - - t the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). regulation onlined, "Thisoranang or esponsibilities under the Act may be obtained from the Oiner general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Nekatuk
onna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications For Use
Applicant: Hygia Health Services, Inc.
510(k) Number:
Device Name: Hygia Health Services Reprocessed Huntleigh Foot Wrap
Indications For Use:
The Hygia Health Services Reprocessed Huntleigh Foot Wrap is designed to enhance The Trygia Iteann Ber Hees Repressed arterioles. It is to be used by patient's in both the checation of brood in the versas a non-invasive therapeutic method to prevent:
- . Deep vein thrombosis
- Reduce wound healing time .
- Treat and assist healing of venous leg ulcers .
- Reduce edema caused by venous insufficiency in the lower extremities
- Decrease compartmental pressures
# PRECAUTIONS AND CONTRAINDICATIONS
Contraindications:
Wraps may not be recommended for patients with the following:
- Congestive heart failure I.
- Known or suspected deep vein thrombosis 2.
- Severe arteriosclerosis or other ischemic vascular disease 3.
- Any local leg condition in which the wrap would interfere such as 4. dermatitis, gangrene, recent skin graft, or untreated infected wounds
4-1
{6}------------------------------------------------
### Indications for Use
### Precautions:
-
- One must ensure that the wrap is applied properly. 1. .
- One must ensure that the wrap is correctly connected to the pump and that 2. the connection is secure.
- If numbness, tingling, or leg pain is experienced by the patient, the wrap 3. should be removed.
Division of Cardiovascular & Respiratory Devices
510(k) Number 261265/
**Prescription Use**
(Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.