A. General Indications: Trauma procedures of the midface or craniofacial skeleton. Specific Indications: 1). Comminuted fractures of the naso-ethmoidal infraorbital areas. 2). Comminuted fractures of the frontal sinus wall: 3). Pediatric midface or craniofacial trauma 4). Lefort (I, II, III) fractures. 5). Orbital floor fractures. 6).Fractures of the maxilla, zygoma, zygomatic arch, orbital rim, nasal, ethmoid, and lacrimal bones. 7). Trauma of the craniofacial skeleton including: frontal, temporal , sphenoid, and occipital bones. B. General Indications: Reconstructive procedures of the midface or craniofacial skeleton.. Specific Indications: 1). Infant craniofacial surgery (i.e., craniosynostosis, congentital malformations, trauma, etc.). 2). Lefort (I, II, III) osteotomies. 3). Tumor reconstruction in midface or craniofacial procedures. 4). Bone graft procedures in the midface or craniofacial skeleton. 5). Pediatric reconstructive procedures. 6). Reconstructive procedures of the craniofacial skeleton including: frontal, parietal, temporal, sphenoid, and occipital bones. 7). Craniotomy flap fixation.
Device Story
Resorbable bone fixation system comprising porous/non-porous sheets, tacks, and screws; manufactured from polylactic acid. Sheets are malleable when heated to ~55°C, allowing 3D anatomical conformation; cut to size by surgeon using scissors or Power Pen. Used in midface/craniofacial trauma and reconstructive surgery. System includes manual instruments (drivers, drill bits, taps) for installation. Provides structural stabilization of bone fractures or grafts; facilitates healing via resorbable material. Surgeon-operated in clinical/OR settings.
Clinical Evidence
Bench testing only. Mechanical testing demonstrated that the device's mechanical properties are substantially equivalent to the predicate devices (MacroPore PX Pediatric System and MacroPore Protego System) under intended use conditions.
Indicated for trauma and reconstructive procedures of the midface and craniofacial skeleton in pediatric and adult patients, including comminuted fractures, Lefort fractures, orbital floor fractures, craniosynostosis, congenital malformations, tumor reconstruction, and craniotomy flap fixation.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
{0}------------------------------------------------
# SEP 1 8 2001
#### ADMINISTRATIVE INFORMATION
Manufacturer Name:
Official Contact:
K012412
MacroPore, Inc. 6740 Top Gun Street San Diego, CA 92121
Kenneth K. Kleinhenz Director of Regulatory Affairs Telephone (858) 458-0900 Fax (858) 458-0994
#### DEVICE NAME
Classification Name:
Trade/Proprietary Name:
Plate, Bone
MacroPore FX, PS, NS, LP
#### ESTABLISHMENT REGISTRATION NUMBER
2031733
### DEVICE CLASSIFICATION AND PRODUCT CODE
As shown in 21CFR 872.4760 - Bone Plates are intended to stabilize fractured bone structures in the oral cavity and are classified as Class II. Bone Plates have been assigned Product Code JEY.
{1}------------------------------------------------
# INTENDED USE
A. General Indications: Trauma procedures of the midface or craniofacial skeleton.
Specific Indications:
1). Comminuted fractures of the naso-ethmoidal infraorbital areas.
2). Comminuted fractures of the frontal sinus wall:
- 3). Pediatric midface or craniofacial trauma
4). Lefort (I, II, III) fractures. ·
5). Orbital floor fractures.
6).Fractures of the maxilla, zygoma, zygomatic arch, orbital rim, nasal, ethmoid, and lacrimal bones. 7). Trauma of the craniofacial skeleton including: frontal, temporal , sphenoid, and occipital bones.
B. General Indications: Reconstructive procedures of the midface or craniofacial skeleton.. Specific Indications:
1). Infant craniofacial surgery (i.e., craniosynostosis, congentital malformations, trauma, etc.).
- 2). Lefort (I, II, III) osteotomies.
- 3). Tumor reconstruction in midface or craniofacial procedures.
- 4). Bone graft procedures in the midface or craniofacial skeleton.
- 5). Pediatric reconstructive procedures.
- 6). Reconstructive procedures of the craniofacial skeleton including: frontal, parietal, temporal, sphenoid, and occipital bones.
- 7). Craniotomy flap fixation.
# DEVICE DESCRIPTION
# Design Characteristics
MacroPore FX, PS, NS, LP is a resorbable bone fixation system composed of various sized porous sheets, non-porous sheets, and associated fixation tacks and screws manufactured from poly lactic acid. The MacroPore FX, PS, NS, LP are composed of MacroPore FX, PS, NS, LP Protective Sheets and MacroPore FX and LP Screws and Tacks. MacroPore Protective Sheets can be cut with scissors to the desired shape and size. The MacroPore Power Pen can also be used to cut or shape the MacroPore Plates and Protective Sheets to the desired shape or size.
The MacroPore Plate and Protective Sheets are fully malleable when heated to approximately 55°C (for example, by the use of sterile hot water), and thus can be conformed three dimensionally to most any anatomical orientation. The MacroPore FX, PS, NS, LP bone fixation system includes a selection of resorbable screws, tacks, and associated manual instruments. Tacks range in size from 1.5mm to 2.0mm in outer diameter. Screws range in size from 2.0mm in outer diameter. The MacroPore Plates and Protective Sheets come in various sizes ranging from 0.5mm in thickness according to the region to be treated. The MacroPore Protective Sheets range in size from as small as 20mm x 20mm to as large as 120mm x 120mm. The MacroPore Protective Sheet is provided with and without macroporous holes. The macroporous holes range in size from 500 microns to 3,000 microns in diameter. All configurations are to be within a mass of 18 grams of polymer.
Various manual instruments (PowerTack Driver, StarBurst Screw drivers, Tack Pusher, taps, drill bit, etc.) are used in conjunction with the MacroPore FX, PS, NS, LP bone fixation system to assist in the installation process.
{2}------------------------------------------------
### Material Composition
The MacroPore FX, PS, NS, LP are fabricated from polylactic acid (PLA).
#### In Vitro Testing
Mechanical testing of the MacroPore FX, PS, NS, LP System demonstrates that the device is substantially equivalent to the predicate. Test results indicate that the mechanical properties of the MacroPore FX, PS, NS, LP are substantially equivalent to the mechanical properties of the predicate devices: MacroPore PX Pediatric System and MacroPore Protego System (plates, and screws) under indication for use conditions.
#### EQUIVALENCE TO MARKETED PRODUCT
The MacroPore FX, PS, NS, LP shares indications and design principles with the following predicate devices which have been determined by FDA to be substantially equivalent to pre-amendment devices: MacroPore PX Pediatric System and the MacroPore Protective Sheet (Protego System) Class II medical devices that were cleared for marketing in the United States under K002207, K972913 respectively.
#### Indications For Use
The MacroPore FX, PS, NS, LP and the predicate devices share identical indications for use with the MacroPore PX Pediatric System predicate. Additionally, the MacroPore FX, PS, NS, LP devices have substantially equivalent indications for use with the MacroPore Protego Ssytem as they are both indicated for the trauma and reconstructive procedures in the midface and craniofacial skeleton.
#### Design and Materials
Both the MacroPore FX. PS. NS. LP and the predicate devices are manufactured from bioabsorbable materials under substantially equivalent conditions. The physical designs of MacroPore FX, PS, LP and the predicate devices are substantially equivalent, consisting of a plates, screws, tacks, and protective sheets. The mechanical characteristics of the MacroPore FX, PS, NS, LP and the predicate devices are substantially equivalent to the predicate device under indications for use conditions. The material used in the MacroPore FX, PS, NS, LP is also substantially equivalent to the predicate devices as they are resorbable polylactide and polyglycolide polymers. In addition to physical characteristics, both the predicate device and the MacroPore FX, PS, NS, LP can be cut and molded to specific shapes and sizes by the end user.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.
Public Health Service
SEP 1 8 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Kenneth K. Kleinheinz Director of Regulatory Affairs MacroPore, Incorporated 6740 Top Gun Street San Diego, California 92121
Re: K012413
Trade/Device Name: MacroPore FX, PS, NS, LP Regulation Number: 872.4760 Regulation Name: Plate Bone Regulatory Class: II Product Code: JEY Dated: July 27, 2001 Received: July 30, 2001
Dear Mr. Kleinhenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements
{4}------------------------------------------------
# Page 2 - Mr. Kleinhenz
of the Act or any Federal statutes and regulations administered by other Federal agencies. of the Act of ally I ederal states and registered to the limited to: registration Touring (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice alla listing (21 CF R Furt 067), assetses (QS) regulation (21 CFR Part 820); and if requirents as sectorin in allo qualify adiation control provisions (section 531-542 of the Act; 21): CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your 510(k) This letter will anow you to begal finding of substantial equivalence of your device to a premarket notification. The PDF results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 If you desire specific as for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and Compliance at (301) 59 - e, please contact the Office of Compliance at (301) 594-4639. advertibing or your the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the notification (210) 10 00197) - Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
L. Clatrand
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
K012413
# Device Name: MacroPore FX, PS, NS, LP
#### Indications for Use:
A. General Indications: Trauma procedures of the midface or craniofacial skeleton. Specific Indications:
1). Comminuted fractures of the naso-ethmoidal infraorbital areas.
2). Comminuted fractures of the frontal sinus wall.
3). Pediatric midface or craniofacial trauma
4). Lefort (I, II, III) fractures.
5). Orbital floor fractures.
5). Oronal noor fractares. 7). Trauma of the craniofacial skeleton including: frontal, parietal, temporal , sphenoid, and occipital bones.
B. General Indications: Reconstructive procedures of the midface or craniofacial skeleton.. Specific Indications:
1). Infant craniofacial surgery (i.e., craniosynostosis, congentital malformations, trauma, etc.).
- 2). Lefort (I, II, III) osteotomies.
3). Tumor reconstruction in midface or craniofacial procedures.
4). Bone graft procedures in the midface or craniofacial skeleton.
- 5). Pediatric reconstructive procedures.
- 6). Reconstructive procedures of the craniofacial skeleton including: frontal, parietal, temporal, sphenoid, and occipital bones.
- 7). Craniotomy flap fixation.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
Susan Pump OR
Over-The-Counter Use
(Division Sign-Off) Division of Dental, Infection Control, and General Hospita 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.