MacroPore Protective Sheet (MPS) is intended for use in trauma and reconstructive procedures in the midface and craniofacial skeleton: 1. Comminuted fractures of the naso-ethmoidal and infraorbital areas 2. Comminuted fractures of the frontal sinus wall 3. Trauma of the midface or craniofacial skeleton 4. Reconstructive procedures of the midface or craniofacial skeleton. The system is not intended for use in the mandible and/or for full load bearing procedures.
Device Story
MacroPore Protective Sheet (MPS) is an absorbable, macroporous implant in sheet form used for fixation of non-load bearing bone defects in the craniofacial skeleton. Fabricated from poly(L-lactide-co-D,L-lactide) 70:30, the sheet is thermoplastic when heated to 55°C, allowing surgeons to cut and conform it three-dimensionally to bone defects. It can be used as a flat sheet or rolled into a tube, often in conjunction with resorbable screws and tacks for stabilization. The device is intended for use by surgeons in clinical settings to provide structural support during healing. By providing a stable, resorbable scaffold, the device facilitates bone reconstruction and fracture fixation without the need for permanent metallic implants, potentially reducing long-term complications associated with foreign body presence.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by design, material, and intended use comparisons to predicate devices. Animal studies were referenced to characterize the degradation profile of the poly(L-lactide-co-D,L-lactide) 70:30 material.
Technological Characteristics
Absorbable, macroporous sheet made of poly(L-lactide-co-D,L-lactide) 70:30. Thickness: 500-1000 microns. Pore size: 500-2000 microns. Thermoplastic (softens at 55°C). Sterilization: Ethylene Oxide per ANSI/AAMI/ISO 11135-1994. Non-load bearing.
Indications for Use
Indicated for patients requiring trauma or reconstructive procedures in the midface and craniofacial skeleton, specifically for comminuted fractures of the naso-ethmoidal, infraorbital, and frontal sinus wall areas. Contraindicated for use in the mandible or for full load-bearing applications.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
JUL 3 0 1998
# MacroPore Protective Sheet (Protego™ System)
Image /page/0/Picture/3 description: The image shows a sequence of handwritten digits and a forward slash. The sequence starts with what appears to be the letter 'K', followed by the numbers '9', '7', '2', '9', then a forward slash '/', and finally the number '3'. The digits are written in a bold, dark font, making them stand out against the background.
June 2, 1998
# ADMINISTRATIVE INFORMATION
| Manufacturer Name: | MacroPore, Inc.<br>3252 Holiday Court, Suite 223<br>La Jolla, CA 92037 |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Christopher J. Calhoun<br>Telephone (619) 458-0900<br>Fax (619) 458-0994 |
| Representative/Consultant: | Floyd G. Larson<br>Pacific Materials and Interfaces<br>4329 Graydon Road<br>San Diego, CA 92130<br>Telephone (619) 792-1235<br>FAX (619) 792-1236 |
| DEVICE NAME | |
| Classification Name: | Plate, fixation, bone |
Trade/Proprietary Name: MacroPore Protective Sheet (Protego™ System)
Common Name:
Plating System
#### ESTABLISHMENT REGISTRATION NUMBER
MacroPore, Inc. has not yet obtained an Establishment Registration Number.
#### DEVICE CLASSIFICATION
In the Federal Register of July 2, 1982 [FR 47 page 29052], FDA proposed that bone fixation systems be classified as Class II, as shown in 21 CFR 888.3030. Although the listed Classification Name in 21 CFR is "single/multiple component metallic bone fixation appliances and accessories," the LactoSorb Trauma Plating System, a system manufactured from resorbable polymers and a predicate device for this submission, has been cleared under an expanded definition of that Classification Name. The LactoSorb device was reviewed by
{1}------------------------------------------------
the General and Plastic Surgery Device Section of the Surgical and Rehabilitation Devices Panel, and has been assigned Product Code 87HRS. We propose that MacroPore Protective Sheet (MPS), the subject of this Notification, be classified as shown above.
# CONFORMANCE WITH PERFORMANCE STANDARDS
No performance standards have been established under Section 514. Voluntary standards with which the MacroPore Protective Sheet complies include American National Standards Institute (ANSI)/Association for the Advancement of Medical Instrumentation (AAMI)/ISO 11135-1994, Medical Devices - Validation and Routine Control of Ethylene Oxide Sterilization.
#### PACKAGING/LABELING/PRODUCT INFORMATION
MacroPore Protective Sheet will be packaged in an Ethylene Oxide (EtO) sterilizable package consisting of double heat sealed pouches, each pouch composed of a 48 gage polyester/0.002 LDPE top web and an uncoated 10739 Tyvek bottom web. Sterilization will be accomplished by ethylene oxide treatment for 6-6.5 hours at 7-9 psig in 10% EtO/90% HCFC. Sterilization will be validated using ANSVAAMI/ISO 11135-1994, Medical Devices -Validation and Routine Control of Ethylene Oxide Sterilization. The sterility assurance level (SAL) that MacroPore intends to meet for the MacroPore Protective Sheet is 104. The device is not represented to be "pyrogen free."
#### INTENDED USE
MacroPore Protective Sheet (MPS) is intended for use in trauma and reconstructive procedures in the midface and craniofacial skeleton:
- 1. Comminuted fractures of the naso-ethmoidal and infraorbital areas
- 2. Comminuted fractures of the frontal sinus wall
- 3. Trauma of the midface or craniofacial skeleton
- 4. Reconstructive procedures of the midface or craniofacial skeleton.
The system is not intended for use in the mandible and/or for full load bearing procedures.
#### DEVICE DESCRIPTION
#### Design Characteristics
MacroPore Protective Sheet is an absorbable, macroporous implant in sheet form manufactured from polylactic acid (PLA). The purpose of the sheet is to provide fixation of non-load bearing bone defects in the cranio-facial area of the mammalian skeletal system.
{2}------------------------------------------------
MPS can be cut with scissors, is thermoplastic when heated to 55 C (for example, by the use of sterile hot water) and thus can be conformed three dimensionally to any bone defect. It can be rolled into a tube or used as a flat sheet. It can be used in conjunction with internal bone fixation devices such as plates and screws which can be also used to fixate the MPS and prevent dislocation. The system includes a selection of resorbable screws and tacks.
MPS is provided in sheets of 40 x 40 mm and in premanufactured shapes and can be provided in other sizes as needed for particular surgical procedures. Its thickness ranges from 500 microns to 1000 microns according to the defect to be treated. The pore size ranges from 500 microns to 2000 microns with pores distributed uniformly throughout the sheet in an offset or aligned pattern.
#### Material Composition
MPS is fabricated from poly(L-lactide-co-D,L-lactide) 70:30.
## EQUIVALENCE TO MARKETED PRODUCT
MacroPore Protective Sheet shares indications and design principles with the following predicate devices, each of which has been determined by FDA to be substantially equivalent to pre-amendment devices: LactoSorb Trauma Plating System from W. Lorenz, a unit of Biomet, Inc. (K960988, K955729), Tilghman Titanium Mesh Skeletal Orbit Liner from TiMesh, Inc. (K922308), Motech Titanium Mesh from Biedermann MOTECH GmbH (K900138), and Leibinger Titanium DynamicMesh.
#### Intended Use
The indications for use of MacroPore Protective Sheet (MPS) are not new indications in that they are the same as or are included in those for the predicate devices. MPS is intended for use in the craniofacial skeleton for fracture fixation in surgical reconstruction. Specific indications are:
for use in trauma and reconstructive procedures in the midface and craniofacial skeleton (comminuted fractures of the naso-ethmoidal and infraorbital areas. comminuted fractures of the frontal sinus wall, trauma of the midface or craniofacial skeleton, reconstructive procedures of the midface or craniofacial skeleton).
The system is not intended for use in the mandible and/or for full load bearing procedures.
The predicate devices share intended uses with MPS as follows: The LactoSorb Trauma Plating System and the titanium mesh products have identical intended uses with regard to fracture fixation to those of MPS with a thickness of 1000 microns.
{3}------------------------------------------------
#### Design and Materials
The design and functional characteristics of MacroPore Protective Sheet and the predicate devices are similar. The physical designs of MPS and the LactoSorb device both consist of a thin sheet with perforations. The LactoSorb Trauma Plating System is made from a copolymer of polylactic acid and polyglycolic acid and has low crystallinity. The composition and processing of LactoSorb devices are selected to permit 70% of its strength to be maintained in vivo for 6-8 weeks (sufficient time to facilitate fracture fixation), yet to degrade and be cleared from the body within 9 to 15 months. The poly(L-lactide-co-D,Llactide) 70:30 of which MPS is made has been shown in animal studies to degrade more slowly, and to be fully degraded after 36 months. Both Lacrosorb and MPS are essentially amorphous. It is believed that degradation of an amorphous matrix is less likely than that of a crystalline matrix to leave microscopic undegraded crystalline regions. Such material could cause osteolysis in the same way as does the polyethylene debris that results from wear of articulating surfaces.
{4}------------------------------------------------
# 510(k) Summary
# SUMMARY: TABLE OF SUBSTANTIAL EQUIVALENCE
MacroPore Protective Sheet is substantially equivalent to the LactoSorl Liner, Motech Titanium Surgical Mesh and Titanium Dynamic Mesh ir
| | Subject Device | |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | MacroPore Protective Sheet | LactoSorb Trauma<br>System<br>(K960988, K955) |
| INTENDED<br>USE | For use in trauma and<br>reconstructive procedures in the<br>midface and craniofacial skeleton<br>1. Comminuted fractures of the<br>naso-ethmoidal and infraorbital<br>areas<br>2. Comminuted fractures of the<br>frontal sinus wall<br>3. Trauma of the midface or<br>craniofacial skeleton<br>4. Reconstructive procedures of the<br>midface or craniofacial skeleton | For use in trauma and<br>reconstructive proce<br>the midface and crani<br>skeleton<br>1. Comminuted frac<br>the naso-ethmoic<br>infraorbital areas<br>2. Comminuted frac<br>the frontal sinus<br>3. Trauma of the m<br>craniofacial skele<br>4. Reconstructive p<br>of the midface or<br>craniofacial skele |
| DESIGN | Sheet of 500 and 1000 microns<br>thickness, size 60 x 80 mm or as<br>required, with pores of 500 to 2000<br>microns diameter arranged in a<br>uniform offset or aligned pattern | Sheets of 500 and 1000<br>microns thickness, size<br>25 mm to 50 x 50 mm with<br>pores of 1500 to 2000<br>microns diameter arranaged in<br>a uniform aligned pattern |
| MATERIAL | poly(L-lactide-co-D,L-lactide)<br>70:30 (Resomer® LR 708),<br>amorphous | poly(L-lactide-co-glycolide)<br>82:18, low crystallinity |
{5}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
1
# ง Trauma Plating System, Tilghman Titanium Mesh Skeletal Orbit n the following respects:
| Predicate Devices | | | |
|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Plating<br>(729) | Tilghman Titanium<br>Mesh Skeletal Orbit<br>Liner (K922308) | Motech Titanium<br>Surgical Mesh<br>(K900138) | Titanium Dynamic<br>Mesh |
| F<br>Hures in<br>iofacial<br>tures of<br>al and<br>tures of<br>wall<br>dface or<br>iton<br>rocedures<br>ton | For use in orbital<br>bony defect repair/<br>reconstruction | For use in reinforcing<br>weak bony tissues<br>such as in acetabular<br>cups, ... and<br>reconstruction of<br>oral-maxillo-cranio-<br>facial surgeries | For defect-bridging<br>reconstruction of<br>non-loadbearing bony<br>structures |
| 100<br>zes 25 x<br>n, with<br>0<br>anged in<br>tern | Sheets formed to<br>various shapes | Sheets of 1000<br>microns thickness,<br>formed to various<br>shapes | Sheets of 300 to 600<br>microns thickness,<br>sizes of 40 x 40 mm<br>to 90 x 90 mm, with<br>open hole pattern,<br>including screw holes<br>of 1.0 mm to 2.3 mm |
| (colide)<br>ty | Titanium | Titanium | Titanium |
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three heads facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.
JUL 3 0 1998
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MacroPore, Inc. c/o Pacific Materials and Interfaces Mr. Floyd G. Larson 4329 Graydon Road San Diego, California 92130
Re: K972913 Trade Name: Macropore Protective Sheet (Protego System) Regulatory Class: II Product Code: HRS and HWC Dated: June 5, 1998 Received: June 8, 1998
Dear Mr. Larson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that. through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{7}------------------------------------------------
Page 2 - Mr. Floyd G. Larson
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
### Device Name: MacroPore Protective Sheet
Indications for Use:
MacroPore Protective Sheet (MPS) is intended for use in trauma and reconstructive procedures in the midface and craniofacial skeleton:
- 1. Comminuted fractures of the naso-ethmoidal and infraorbital areas
- 2. Comminuted fractures of the frontal sinus wall
- 3. Trauma of the midface or craniofacial skeleton
- 4. Reconstructive procedures of the midface or craniofacial skeleton.
The system is not intended for use in the mandible and/or for full load bearing procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Prescription Use Over-Theounter Use
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.