The Geodesic EEG System is intended to measure and record the electrical activity in of the patient's brain. It can be used on adults, children, and infants.
Device Story
Digital EEG system; acquires, stores, and displays physiological signals. Input: electrical brain activity via Geodesic Sensor Net (dense array of 32-256 scalp electrodes). Components: scalp electrodes, amplifiers (Net Amps), central processing unit, and Net Station software. Operation: computer-controlled; electrodes applied to scalp without abrasion; signals amplified and digitized. Output: operator-selected waveform displays on PC monitor. Used in clinical settings; operated by trained personnel. Software provides system control and data storage; no automated data analysis performed. Benefits: rapid electrode application, even distribution across head, and high-density recording capability.
Clinical Evidence
No clinical data submitted. Bench testing only, including safety (CAN/CSA C22.2 No. 601.1-M90, UL 2601-1), electromagnetic compatibility (EN60601-1-2), and electroencephalograph performance standards (IEC 60601-2-26). Biocompatibility testing performed per ISO 10993.
Technological Characteristics
Digital EEG system; 32-256 channel dense sensor array. Components: Geodesic Sensor Net (non-abrasive electrodes), Net Amps (software-controlled amplifiers), PC-based workstation. Connectivity: standard PC peripherals. Biocompatibility: ISO 10993. Safety standards: IEC 60601-2-26, CAN/CSA C22.2 No. 601.1-M90, UL 2601-1. EMC: EN60601-1-2.
Indications for Use
Indicated for measuring and recording brain electrical activity in adults, children, and infants.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Predicate Devices
Bio-Logic Ceegraph 128-Channel Recording System (K973883)
{0}------------------------------------------------
SEP 1 9 2001
Ko/2079
## 510(k) SUMMARY
Submitter's name: Electrical Geodesics, Inc. 1850 Millrace Drive Eugene, OR 97403 541-687-7962
Date summary prepared:
June 25, 2001
Device name:
Proprietary name: Common or usual name: Classification name:
Geodesic EEG System™ EEG machine Electroencephalograph, 84 GWO Class II, 21 CFR 882.1400.
The product includes the Geodesic Sensor Net® which has the following identifiers:
| Common name: | EEG electrode |
|----------------------|-----------------------------------------------------------|
| Classification name: | Cutaneous electrode, 84 GXY<br>Class II, 21 CFR 882.1320. |
Legally marketed device for substantial equivalence comparison:
For the Geodesic EEG System it is the Bio-Logic Ceegraph 128-Channel Recording System submitted by Bio-Logic Systems Corporation and cleared for marketing under 510(k) *K973883. For the Geodesic Sensor Net, the predicate device is the Electro-Cap VII System submitted by Electro-Cap Inc. and cleared for marketing under 510(k) *K780045.
Description of the device:
The Geodesic EEG System is a digital electroencephalography system (EEG) with a dense sensor array of 32 to 256 channels. Like existing digital EEG systems, the Geodesic EEG System is computer controlled and capable of acquiring, storing, and displaying data. It includes scalp conductive electrodes, amplifiers, a central processing unit, and software. The Geodesic Sensor Net is a dense array of scalp electrodes designed to allow rapid application in an even distribution across the head. The Net Amps " consists of multiple amplifiers for physiological signals that are fully software controlled. Net Station® is the software package that provides control of the Geodesic EEG System, digital data storage, and operator-selected waveform displays. The software does not perform any data analysis. Additional components of the system are an articulated arm with extended cable, rack system, various cables, standard components of personal computer (monitor, keyboard, mouse), electrolyte solution, and disinfectant.
The Geodesic EEG System is a new device that has not previously been submitted to FDA. It has features similar to other digital EEG devices on the market.
{1}------------------------------------------------
Intended use of device:
The Geodesic EEG System is intended to measure and record the electrical activity in of the patient's brain. It can be used on adults, children, and infants.
Technological characteristics:
The technological characteristics of the Geodesic EEG System are similar to those of other digital EEG systems, including the predicate device, the Ceegraph 128-Channel Recording System. Each product is an EEG machine that is software controlled, can accommodate a variable number of electrodes, and can acquire, display, and record EEG data. Some differences in electrical parameters are described. Differences in the software include that the products use different operating systems and that Net Station does not analyze the data. Another difference is that the Geodesic EEG System has dedicated electrodes, the Geodesic Sensor Net.
The Geodesic Sensor Net is compared to the Electro-Cap VII System. Each product consists of a structure that links a number of electrodes so that they can be easily applied to the patient. The Geodesic Sensor Net uses a geodesic array of electrodes with equal distribution across the head. The Electro-Cap uses the 10-20 array. The Geodesic Sensor Net can accommodate a larger number of electrodes. Finally, the Geodesic Sensor Net does not require scalp abrasion for use.
Testing conducted:
Testing was conducted to ensure compliance with international standards related to electroencephalographs. The general safety standards used were: CAN/CSA C22.2 No. 601.1-M90 including Supplement 1 and Amendment 2 and UL Std. No. 2601-1 (2nd Edition). The electromagnetic compatibility standard was EN60601-1-2(1993). The electroencephalograph standards were IEC 60601-2-26 and CAN/CSA C22.2 No. 601.2.26. The biocompatibility standard was ISO 10993. The Geodesic EEG System passed all testing.
Performance testing:
Comparative performance testing and clinical evaluations were not submitted as part of this 510(k).
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting a sense of community and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure, emphasizing the department's role and national scope.
Public Health Service
SEP 1 9 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Electrical Geodesics, Inc. c/o Mr. Robert S. McQuate R. S. McOuate & Associates, Inc. 3636 E. Columbine Drive Phoenix, Arizona 85032
Re: K012079
Trade/Device Name: Geodesic EEG System™ Regulation Number: 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: GWQ Dated: June 29, 2001 Received: July 3, 2001
Dear Mr. McQuate:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Robert S. McQuate
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Walker, M.D
Celia M. Witten, Ph.D., M.D.
f
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Geodesic EEG System™ 510(k) Notification Page 4
## Indications for Use Statement
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device name: Geodesic EEG System™
Indications for Use:
The Geodesic EEG System is intended to measure and record the electrical activity of the patient's brain. It can be used on adults, children, and infants.
(Please do not write below this line)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <img alt="check mark" src=""/> | OR | Over-The-Counter Use | ______ |
|------------------|--------------------------------|----|----------------------|--------|
|------------------|--------------------------------|----|----------------------|--------|
K012079
![signature]()
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
| 510(k) Number | K012079 |
|---------------|---------|
|---------------|---------|
4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.