K973883 · Bio-Logic Systems Corp. · GWQ · Mar 5, 1998 · Neurology
Device Facts
Record ID
K973883
Device Name
BIO-LOGIC CEEGRAPH 128-CHANNEL RECORDING SYSTEM
Applicant
Bio-Logic Systems Corp.
Product Code
GWQ · Neurology
Decision Date
Mar 5, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Bio-logic Ceegraph family is intended for use in the recording and analysis of routine EEG tests. Typical EEG tests are 20-30 minutes in duration, but the Ceegraph system can also be used for longer tests, including continuous long-term EEG monitoring with patient video. EEG testing is intended for use whenever it is necessary to measure and record the electrical activity of a patient's brain by attaching multiple electrodes at various locations on the scalp. The 128 Channel Data Recording system has a similar intended use to that of the 32 channel Ceegraph recording system (predicate device). It can be used for patients of all ages, from newborn infants through adults, to and including geriatric patients.
Device Story
The Bio-logic Ceegraph 128-Channel Recording System is a digital EEG recorder designed to capture electrical brain activity. It utilizes up to eight 16-channel analog boards to provide patient isolation and signal amplification, accepting inputs from electrode arrays attached to the patient's scalp. The system converts analog signals to digital data via A/D converters and transmits this data to a host computer via a high-speed Ethernet interface. It is intended for use by clinicians in clinical settings for routine EEG tests (20-30 minutes) or long-term monitoring. The system provides the clinician with recorded EEG data for analysis, facilitating the diagnosis and monitoring of neurological conditions. It benefits patients by allowing for high-density (up to 128 channels) brain activity recording, which can improve the resolution and diagnostic capability of EEG studies compared to lower-channel systems.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including EMI/EMC testing (CISPR 11, EN 61000 series, MIL-STD-461C), current leakage tests (ANSI/AAMI), dielectric withstand tests (IEC 601-1), environmental/temperature testing, and data integrity verification to ensure digital output accurately represents analog patient inputs.
Technological Characteristics
128-channel digital EEG recorder; microprocessor-based control; 16-channel analog boards for signal amplification and patient isolation (optical isolators); Ethernet connectivity (10 base-T); transformer-coupled isolation; low-voltage DC power; metal enclosure. Complies with IEC 601-1 and various EMI/EMC standards (CISPR 11, EN 61000 series).
Indications for Use
Indicated for recording and analysis of routine EEG tests and continuous long-term EEG monitoring with patient video. Applicable to all patient populations, including newborn infants, children, adults, and geriatric patients, whenever measurement of electrical brain activity via scalp electrodes is required.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
K973883
Bio-logic One Bio-logic Plaza Mundelein, Illinois 60060-3700 1-800-323-8326 Fax: 847-949-8615
MAR - 5 1998
# SECTION 2: SUMMARY AND CERTIFICATION
## 510(K) SUMMARY
## SAFETY AND EFFECTIVENESS SUMMARY
Safety and effectiveness information concerning this Modification to Ceegraph for 128-Channel Recording is summarized below.
Because this is not a CLASS III device, the special certification defined for this section is not required.
PREPARED BY: Bio-logic Systems Corp One Bio-logic Plaza Mundelein, IL 60060
(847)-949-5200 TELEPHONE:
CONTACT PERSON: Norman E. Brunner
DATE ON WHICH THE SUMMARY WAS PREPARED: October 13, 1997
NAME OF DEVICE: Bio-logic CEEGRAPH 128-Channel Recording System.
Digital EEG Recorder. COMMON NAME:
Electroencephalograph (per CFR 882.1400). CLASSIFICATION NAME:
PREDICATE DEVICE: Bio-logic Ceegraph Digital EEG System , 510(k) #K933233.
## DESCRIPTION OF THE DEVICE:
The 128-Channel data collection system consists of a metal enclosure approximately 13.5' x 7.375' x 5.25" in size, weighing approximately 10.9 (when fully-equipped to handle all 128 channels). It can be configured to perform from 16 to 128 channel collection, depending on the hardware installed in the box. Power to the box is supplied with an external medical-grade power supply, which supplies regulated 12 volts and 5 volts DC to a connector at the rear of the box. Communication to the host computer is performed through a standard Ethernet interface connector capable of running at data rates up to 10 MHz (10 base-T). Four patient connectors are provided on the front of the box, each capable of interfacing to 32 channels of patient input. A sync connector is provided at the rear of the box, to provide synchronization and sampling rate information to external devices.
{1}------------------------------------------------
The system consists of a microprocessor board, a digital interface board, and up to eight (8) 16-channel analog boards connected to the system through a backplane interconnect board. The analog boards are identical to the 16-channel analog boards used in the patient connection module (headbox) of the predicate device. These boards provide patient isolation and signal amplification. Boards are connected in pairs to one of the four external connectors on the front of the unit. These 50-pin external connectors are identical to the auxiliary connector on the present headbox of the predicate device. This allows the use of existing patient connection hardware, such as electrode arrays and the 32-channel" electrode connection panel called the "quick disconnect box". The microprocessor board contains program and data memory and control functions for reading the analog data, converting it to digital, and communicating it to the host computer through the Ethernet cable. The digital interface board contains the interface to the A/D converters and the communications hardware.
## INTENDED USE:
The Bio-logic Ceegraph family is intended for use in the recording and analysis of routine EEG tests. Typical EEG tests are 20-30 minutes in duration, but the Ceegraph system can also be used for longer tests, including continuous long-term EEG monitoring with patient video. EEG testing is intended for use whenever it is necessary to measure and record the electrical activity of a patient's brain by attaching multiple electrodes at various locations on the scalp. The 128 Channel Data Recording system has a similar intended use to that of the 32 channel Ceegraph recording system (predicate device). It can be used for patients of all ages, from newborn infants through adults, to and including geriatric patients.
PATIENT POPULATION: Adults, children and infants.
## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS:
The 128-Channel Data Recording System uses substantially the same technical design approach as that of the predicate device patient connection module (headbox). The predicate device employs a microprocessor-based digital control board, a patient interface board, and two 16-channel analog boards. The same analog boards are used in the 128-Channel system, which can accommodate up to eight of them. The digital board is an updated design incorporating a more powerful microprocessor and more memory, and the digital control. interfacing and communications functions are contained on two boards instead of one, as in the predicate device. The method of communication to the host computer is through high-speed Ethernet cable, whereas the predicate device uses an RS-422 serial communication interface.
Both the predicate device and the 128-Channel System require only low-voltage DC power for operation. Patient isolation is provided through the use of optical isolators on the interface board. In addition, the 128-Channel System connects to the host computer through a transformer-coupled Ethernet connector, which provides a second level of isolation.
{2}------------------------------------------------
## SUMMARY OF NON-CLINICAL TESTING:
The following is a list of tests performed on the 128-Channel Data Recording System to demonstrate that the performance of the system is equivalent to that of the predicate device in terms of safety and effectiveness, and that the additional features provide utility and product performance which exceeds that of the predicate device. All tests were completed satisfactorily with no adverse reports.
- 1. EMI/EMC Testing per FDA Reviewer's Guidance (November, 1993):
- a.) CISPR 11 Conducted and CISPR 11 A Radiated Emissions
- b.) ENV 50140 Radiated Susceptibility Test
- c.) EN 61000-4-4 Transient Susceptibility Test
- d.) EN 61000-4-5 Surge Susceptibility Test
- e.) EN 61000-4-6 and CS114 Conducted Immunity Tests
- f.) RE101 Magnetic Emissions Test
- g.) RS101 and 60 Hz Magnetic Susceptibility Tests
- h.) EN 61000-4-8 Voltage Fluctuations Test
- i.) Reviewer's Guide Quasi-static Fields Test
- i.) Military Standard 461C CS02 Test
- k.) EN 60555-2 Power Line Harmonics Test Prescan
- Current Leakage Tests per ANSI/AAMI, Safe Current Limits for Electromedical Apparatus. 2.
- 3. Dielectric Withstand Test per FDA Reviewer' Guidance and IEC 601-1.
- Power Line Variations per FDA Reviewer's Guidance. 4.
- Power Line Transients per FDA Reviewer's Guidance. રું.
તમારી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, આંગણ
- Temperature testing per FDA Reviewer's Guidance. 6.
- 7. Electrostatic Discharge Tests (ESD) per FDA Reviewer's Guidance and EN60601-2.
- 8. Data Integrity Tests to demonstrate that digital data collected is a correct representation of the analog signals that are applied at the patient inputs.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR -- 5 1000
Mr. Norman E. Brunner Vice-President of R&D Bio-logic Systems Corporation One Bio-logic Plaza 60060-3700 Mundelein, Illinois
Re: K973883 Bio-logic Ceegraph 128-Channel Recording Trade Name: System Regulatory Class: II Product Code: GWO January 26, 1998 Dated: Received: January 27, 1998
Dear Mr. Brunner:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with ....... the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{4}------------------------------------------------
Page 2 - Mr. Brunner
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
#### K973883 510(k) Number (if known):
#### Device Name: BIO-LOGIC MODIFICATION TO CEEGRAPH FOR 128 CHANNEL DATA RECORDING
Indications For Use:
The Bio-logic Ceegraph family is indicated for use in the recording and analysis of routine EEG tests. Typical EEG tests are 20-30 minutes in duration. but the Ceegraph system can also be used for longer tests. including continuous lone-term EEG monitoring with patient video. EEG testing is indicated for use whenever it is necessary to measure and record the electrical activity of a patient's brain by attaching multiple electrodes at various locations on the scalp.
The 128 Channel Data Recording system has a similar intended use to that of the 32 channel Ceegraph recording system.
It can be used for patients of all ages, from newborn infants through adults, to and including geriatric patients.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
tocallos
n Sign-Off) Divisio Division of General Restorative Devices 973885 510(k) Number ..
**Prescription Use**
(Per 21 CFR 801.109)
OR
Over-The-Counter Use _______
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.