SiteSelect is a disposable, single use diagnostic device used to obtain localized large core biopsies of breast tissue of a mammographic abnormality, identified by the placement of a needle/localization wire, which is suspect to be malignant. The SiteSelect device is intended to provide tissue for histological examination with partial or complete removal of imaged lesions. The scope of a histological abnormality is not able to be determined from its mammographic appearance. Therefore, it is essential that the tissue margins be examined for margin involvement and completeness of removal in cases where the tissue sample is not found to be benign. SiteSelect is to be used in conjunction with a stereotactic mammographic imaging system capable of determining position of lesion within breast tissue. This device is used in conjunction with the SiteSelect Instrument Accessories, which are mounted on a stereotactic table and used with a stereotactic mammographic imaging system.
Device Story
SiteSelect is a sterile, single-use disposable biopsy instrument for obtaining large core breast tissue samples. Used in conjunction with stereotactic mammographic imaging systems and table-mounted accessories; device targets mammographic abnormalities identified by localization needle/wire. Operation involves inserting device to lesion site; device obturates up to lesion, preserving intervening tissue. Specimen collection utilizes integrated garrote wire and disc blades to transect tissue. Trigger mechanism allows manual deployment. Output is a tissue specimen suitable for histological examination. Device assists clinicians in diagnosing breast lesions by providing tissue for margin assessment and completeness of removal. Benefits include accurate targeting and capture of adequate tissue for pathology.
Clinical Evidence
Clinical study compared 22mm SiteSelect (n=25) to 15mm SiteSelect (n=25). Endpoints included procedure time, incision length, blood loss, use of electrocautery, patient discomfort, follow-up tissue appearance, and specimen suitability for pathology. Results showed no significant variances between the two devices; both demonstrated accurate targeting and capture of adequate tissue for histological analysis.
Technological Characteristics
Disposable, single-use biopsy instrument. Components: localization needle, localization wire, stylet, cutting cannula, garrote wire, and disc blades. Materials: sterile, thermoformed blister packaging with Tyvek lid. Energy source: manual. Connectivity: none (mechanical). Sterilization: sterile. Form factor: 22mm diameter biopsy device.
Indications for Use
Indicated for patients with mammographically identified breast abnormalities suspected of malignancy requiring localized large core biopsy for histological examination.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
K011575
# AUG 1 7 2001
# 510(k) Summary of Safety and Effectiveness May 18, 2001
#### 1. Submitter
Julie Powell, Quality Assurance/Regulatory Affairs Vice President Imagyn Surgical, A Division of Imagyn Medical Technologies, Inc. 8850 M-89, P.O. Box 351 Richland, MI 49083 ph: (616) 629-5811 fax: (616) 629-9654 e-mail: julie.powell(@imagyn.com
#### 2. Name of device
Common name: Biopsy Instrument Trade name: SiteSelect Breast Biopsy Device Classification name: Instrument, Biopsy
## 3. Devices to which substantial equivalence is claimed
| 510(k)# | Device | Manufacturer |
|---------|---------------------------------|-----------------|
| K993936 | SiteSelect Breast Biopsy Device | Imagyn Surgical |
| K972328 | SiteSelect Breast Biopsy Device | Imagyn Surgical |
| K983296 | ABBI | USSC |
| K954628 | ABBI | USSC |
#### 4. Description of Device
SiteSelect Breast Biopsy System consists of the SiteSelect device and the Site Select table mount accessories. We have previously received clearance under 510(k) # K972328 for the 10mm and 15mm SiteSelect device. We have since received clearance under 510(k) #K993963 to expand the intended use. The Instrument Accessories have received clearance under 510(k) # K972328 and are currently on the market.
The SiteSelect device is provided sterile for single use only. It contains a Localization Needle, Localization Wire, a Stylet, and Cutting Cannula. The device is equipped with a Garrote Wire and Disc Blades to transect the specimen. The device is packaged in individual thermoformed blisters with peel off Tyvek® lid.
{1}------------------------------------------------
#### 5. Intended Use
SiteSelect is a disposable, single use diagnostic device used to obtain localized large core biopsies of breast tissue of a mammographic abnormality, identified by the placement of a needle/localization wire, which is suspect to be malignant. The SiteSelect device is intended to provide tissue for histological examination with partial or complete removal of imaged lesions. The scope of a histological abnormality is not able to be determined from its mammographic appearance. Therefore, it is essential that the tissue margins be examined for margin involvement and completeness of removal in cases where the tissue sample is not found to be benign.
SiteSelect is to be used in conjunction with a stereotactic mammographic imaging system capable of determining position of lesion within breast tissue.
This device is used in conjunction with the SiteSelect Instrument Accessories, which are mounted on a stereotactic table and used with a stereotactic mammographic imaging system.
#### 6. Device compared to predicate device
The 22mm SiteSelect is substantially equivalent to like devices in commercial distribution and currently marketed by Imagyn Surgical and United States Surgical Corporation (USSC).
Imagyn Site Select and USSC devices are all used to obtain biopsies of the breast. These devices are used in conjunction with instrument accessories, which are used to mount the devices on a stereotactic table. The devices are used with stereotactic mammographic imaging systems, such as Fischer or Lorad prone stereotactic tables.
The difference between SiteSelect and USSC ABBI devices is that the ABBI device removes tissue in route to and including the target area. The SiteSelect device obturates up to the lesion, leaving intact the tissue in route to the target area.
The difference between the 22mm SiteSelect and currently marketed SiteSelect devices is the 22mm device has disc blades along with a garrote wire to transect tissue. The other difference is the method of trigger deployment. The triggers are pulled away from patient instead of being pushed forward
The construction, materials, and sterilization of the 22mm SiteSelect device is similar to the Imagyn Surgical SiteSelect devices currently on the market.
K011575
{2}------------------------------------------------
K011575
### 7. Clinical Performance Data
The 22mm SiteSelect clinical data was compared to data gathered on the current 15mm SiteSelect device. A total of 50 procedures (25-22mm and 25-15mm) were completed. The data in the following categories from both devices were comparable with no significant variances between the two:
- Patient population .
- Procedure variables-procedure time, incision length, blood loss, . use of electrocautery
- Discomfort level during the procedure ●
- Breast tissue appearance at follow up visit ●
- Specimens were suitable for pathological analysis .
## 8. Conclusions drawn from clinical study
The 22mm SiteSelect device was shown to be safe for its intended use. In all clinical cases the device had accurate targeting of selected tissue and capture of adequate tissue specimen for histological analysis.
The clinical study showed the 22mm SiteSelect device was comparable to the 15mm SiteSelect device with no significant variances between the two.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three swooping lines representing its wings.
# AUG 1 7 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Julie Powell Vice President, Quality Assurance/Regulatory Affairs Imagyn Surgical A Division Of Imagyn Medical Technologies, Inc. 8850 M-89 P.O. Box 351 Richland, Michigan 49083
Re: K011575
Trade/Device Name: SiteSelect Breast Biopsy Device Regulation Number: 876.1075 Regulatory Class: II Product Code: KNW Dated: May 18, 2001 Received: May 22, 2001
Dear Ms. Powell:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
#### Page 2 - Ms. Julie Powell
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
Mark McMillan
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Ver/3-4/24/96
Applicant: Imagyn Surgical
5109(k) Number (if known):
Device Name: SiteSelect Breast Biopsy Device
#### Indications for Use:
SiteSelect® is a disposable, single use diagnostic device used to obtain localized large core biopsies of breast tissue of a mammographic abnormality, identified by the placement of a needle/localization wire, which is suspect to be malignant. The SiteSelect device is intended to provide tissue for histological examination with partial or complete removal of imaged lesions. The scope of a histological abnormality is not able to be determined from its mammographic appearance. Therefore, it is essential that the tissue margins be examined for margin involvement and completeness of removal in cases where the tissue sample is not found to be benign.
SiteSelect is to be used in conjunction with a stereotactic mammographic imaging system capable of determining position of lesion within breast tissue.
This device is used in conjunction with the SiteSelect Instrument Accessories, which are mounted on a stereotactic table and used with a stereotactic mammographic imaging system.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE) (Per 21 CFR 801.1091)
Murk N. Milbersen
vision of General, Restorative and Neurological Devices 1011575
510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.