K983296 · United States Surgical, A Division of Tyco Healthc · KNW · Jun 4, 1999 · Gastroenterology, Urology
Device Facts
Record ID
K983296
Device Name
AUTO SUTURE ABBI SYSTEM
Applicant
United States Surgical, A Division of Tyco Healthc
Product Code
KNW · Gastroenterology, Urology
Decision Date
Jun 4, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
To be used when stereotactically localized, large diameter, breast biopsies, identified by the placement of a needle localization wire, are desired for diagnostic sampling of a mammographic abnormality that is suspicious for malignant disease (i.e. usually BIRADS class 4 or 5). The ABBI* device is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of a histologic abnormality cannot be determined from its mammographic appearance. Therefore, the extent of removal of a mammographic abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled mammographic abnormality or the sampled tissue is not histologically benign, it is essential that the tissue margins be examined for completeness of removal. The device is available in 5, 10, 15 and 20mm diameters. Device diameter should be matched to desired diameter of tissue sample for diagnosis.
Device Story
System uses stereotactic mammography to target non-palpable breast abnormalities; utilizes disposable biopsy device with circular scalpel and electrocautery snare to excise tissue samples. Used in clinical settings by physicians; device diameter (5-20mm) matched to lesion size. Provides tissue for histologic examination; aids in diagnosis of malignant disease. Benefits include ability to perform partial or complete removal of imaged abnormality for diagnostic purposes.
Clinical Evidence
Prospective, multi-centered, non-randomized study of 106 patients. Primary endpoint: 100% success rate for histological diagnosis (80.2% benign, 19.8% malignant). Secondary endpoint: 100% success in removing radiographically marked abnormalities; 96.2% complete removal verified by specimen and post-operative mammography.
Technological Characteristics
Disposable biopsy device; circular scalpel and electrocautery snare; diameters 5, 10, 15, 20mm. Materials compliant with ISO 10993-1.
Indications for Use
Indicated for patients requiring diagnostic sampling of mammographically suspicious breast abnormalities (typically BIRADS 4 or 5) that have been stereotactically localized via needle wire. Not contraindicated by age/gender, though intended for breast tissue biopsy.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
# Auto Suture* ABBI* System
K983296
SE 6-4-99
#### 510(k)_Summary of Safety and Effectiveness IX.
United States Surgical Corporation SUBMITTER: 150 Glover Avenue Norwalk, CT 06856
Christopher A. Graham CONTACT PERSON:
June 4, 1999 DATE PREPARED:
CI.ASSIFICATION NAME: Instrument, Biopsy
Instrument, Biopsy COMMON NAME:
Auto Suture* ABBI+ System PROPRIETARY NAME:
Auto Suture* ABBI* System (K963825). PREDICATE DEVICES:
The Auto Suture* ABBI* System uses stereotactic DEVICE DESCRIPTION: mammography to target non-palpable breast abnormalitics for biopsy. The system includes a disposable biopsy device, which uses a circular scalpel and elcctrocautcry snare to excise a diagnostic sample for diagnosis.
To be used when stereotactically localized, large diameter, INTENDED USE: breast biopsies, identified by the placement of a needle localization wire, are desired for diagnostic sampling of a mainmographic abnormality that is suspicious for malignant disease (i.e. usually BIRADS class 4 or 5). The ABBI* device is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.
> The extent of a histologic abnormality cannot be appearance. mammographic determined from its Therefore, the extent of removal of a mammographic abnormality docs not predict the extent of removal of a histologic abnormality, c.g., malignancy. When the sampled mammographic abnormality or the sampled tissue is not histologically benign, it is essential that the tissue margins be examined for completeness of removal.
> The device is available in 5, 10, 15 and 20mm diameters. Device diameter should be matched to desired diameter of tissue sample for diagnosis.
{1}------------------------------------------------
## Auto Suture* ABBI* System
#### MATERIALS: All component materials of the Auto Suture* ABBI* System are comprised of materials which are in accordance with ISO Standard # 10993-1.
CLINICAL SUMMARY: In a multi-centered, non-randomized, prospective study, 106 patients under went breast biopsy procedures using the Auto Suture* ABBI* System to evaluate the device for use in the removal of mammographic abnormalities. The Auto Suture* ABBI* System was evaluated on the following endpoints:
- Primary endpoint: 106 of the 106 (100%) complete . procedures were successful for a histological diagnosis. 85 patients (80.2%) had benign lesions and 21 patients (19.8%) had malignant lesions. Of the 21 patients with malignant lesions, 7 (33.3%) had negative margins, 11 (52.4%) had positive margins, and 3 (14.3%) were undetermined. Consistent with the standard of care for excisional biopsies, all patients with malignancies were treated with a therapoutic modality after the diagnostic ABBI* procedure.
- Secondary endpoint: 106 of the 106 (100%) . procedures were successful, in terms of removing a mammographically visible abnormality that had been radiographically marked for the purpose of diagnosis. 102 of the 106 (96.2%) procedures were completely removed, as verified by the presence of the Icsion in the specimen radiograph and by its absence in the patient's post-operative mammogram of the biopsy site.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized wing-like shapes and a flowing tail.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 4 1999
Mr. Christopher A. Graham Senior Associate, Regulatory Affairs United States Surgical Corporation 150 Glover Avenue Norwalk, Connecticut 06856
Re: K983296 Trade Name: Auto Suture ABBI System Regulatory Class: II Product Code: KNW Dated: May 10, 1999 Received: May 12, 1999
Dear Mr. Graham:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
### Page 2 - Mr. Christopher A. Graham
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
tocottefa
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
IV. Indications For Use:
1 - - - -
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
510(k) Number (if known):
Name: Auto Suture* ABBI* System
Indications For-Use:
To be used when stereotactically localized, large diameter, breast biopsies, identified by the placement of a needle localization wire, are desired for diagnostic sampling of a mammographic abnormality that is suspicious for malignant diseasc (i.e. usually BIRADS class 4 or 5). The ABBI* device is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.
The extent of a histologic abnormality cannot be determined from its mammographic appearance. Therefore, the extent of removal of a mammographic abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled mammographic abnormality or the sampled tissue is not histologically benign, it is essential that the tissue margins be examined for completeness of removal.
The device is available in 5, 10, 15 and 20mm diameters. Device diameter should be matched to desired diameter of tissue sample for diagnosis.
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Evaluation (ODE)
Prescription Use: OR Over-The-Counter Use:
(Per 21 CFR §801.109)
(Division Sign-Off)
Division of General Restorative Devices K983296
510(k) Number
510(k) K983296
Additional Information
May 20, 1999
United States Surgical
* Trademark of United States Surgical
** Trademark name not yet determined
Page 3 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.