Biowy Multi-Chamber Freezing Bag Set

BK261343 · Biowy Corporation Lake Forest, CA 92630 · LPZ · Apr 29, 2026

Device Facts

Record IDBK261343
Device NameBiowy Multi-Chamber Freezing Bag Set
ApplicantBiowy Corporation Lake Forest, CA 92630
Product CodeLPZ
Decision DateApr 29, 2026
DecisionSESE
Device ClassClass U

Indications for Use

The Biowy Freezing Bag Set is intended for a single sequence of freezing, storage (in Liquid Nitrogen (-196°C)), and subsequent thawing (at +37°C) of hematopoietic progenitor cells.

Device Story

Sterile plastic flexible container system; used for freezing, storage (liquid nitrogen, -196°C), and thawing (+37°C) of hematopoietic progenitor cells. System includes multi-chamber freezing bag, transfer/process bag, supernatant collecting bag, extension tubing, and connectors. Used in clinical/laboratory settings for cell processing. Provides containment for cryopreservation; ensures integrity of biological samples during temperature transitions.

Clinical Evidence

Bench testing only. Verification included mechanical strength, biocompatibility, and durability testing in liquid nitrogen freeze/thaw cycles with acceptable results.

Technological Characteristics

Materials: Freezing bag (EVA), accessory bags (PVC), tubing/connectors (PVC, ABS, PP). Design: Multi-chamber flexible container with integrated tubing and accessory bags. Capacity: 10-35ml. Sterilization: Sterile. Principle: Passive containment for cryopreservation.

Indications for Use

Indicated for the freezing, storage, and thawing of hematopoietic progenitor cells.

Predicate Devices

Submission Summary (Full Text)

{0} # Special 510(k) ## 510(k) Summary | **510(k) Owner:** | Biowy Corporation 27031 Vista Terrace Lake Forest, California 92630 Telephone No. 1-949-305-8211 | | --- | --- | | **Primary Contact:** | Arthur Lu Biowy Corporation Telephone: 1-949-305-8211 Fax: 1-866-506-5094 Email: alu@biowy.com | | **Date prepared:** | March 30, 2026 | | **Device name:** | | | Trade name: | Biowy Multi-Chamber Freezing Bag Set | | Common Name: | Freezing Bag | | Classification name: | Container, Frozen Donor Tissue Storage | | FDA Product Code: | LPZ | | Classification: | Unclassified | | **Predicate Device:** | Biowy Multi-Chamber Freezing Bag (BK251308) | **Device Description:** The Biowy multi-chamber freezing bag set is a sterile plastic flexible container intended only for a single sequence of freezing, storage (in Liquid Nitrogen (-196°C)), and subsequent thawing (at +37°C) of hematopoietic progenitor cells. The freezing bag product is comprised of a multi-chamber freezing bag with twist-off ports, extension tubing and connectors with a transfer/process bag and a supernatant collecting bag. **Indications for Use:** The Biowy Freezing Bag Set is intended for a single sequence of freezing, storage (in Liquid Nitrogen (-196°C)), and subsequent thawing (at +37°C) of hematopoietic progenitor cells. ### Technical Characteristics The Biowy Multi-Chamber Freezing Bag Set is substantially equivalent to the predicate in indications for use, design, technology/principal of operation, materials and performance. Differences between the devices do not raise any significant issues of safety and effectiveness. | | **Predicate** | **Subject Device** | | --- | --- | --- | | | **Biowy Multi-Chamber Freezing Bag (BK251308)** | **Biowy Multi-Chamber Freezing Bag Set** | | **Indications for Use** | The Biowy Freezing Bags are intended for a single sequence of freezing, storage (in Liquid Nitrogen (-196°C)), and subsequent thawing (at +37°C) of hematopoietic progenitor cells. | The Biowy Multi-Chamber Freezing Bags are intended for a single sequence of freezing, storage (in Liquid Nitrogen (-196°C)), and subsequent thawing (at +37°C) of hematopoietic progenitor cells. | | Specimen | Hematopoietic progenitor cells | Same as the predicate | | Materials | Freezing bag: EVA | Freezing bag: EVA (same as the predicate) | {1} | | Accessory bag: Mixing bag (EVA) Extension tubing/connectors: PVC, ABS, PP | Accessory bags: Transfer/process bag and supernatant collecting bag (PVC, different from the predicate) Extension tubing/connectors: PVC, ABS, PP (same as the predicate) | | --- | --- | --- | | Design | A sterile multi-chamber EVA freezing bag with integrated tubing and a mixing bag. | A sterile multi-chamber EVA freezing bag with integrated tubing, a transfer/process bag and a supernatant collecting bag. | | Freeze capacity | 10-35ml | 10-35ml | # **Nonclinical Performance Data:** Verification of device performance was established with acceptable results from the following non-clinical tests: mechanical strength, biocompatibility, durability in freeze/thaw studies in liquid nitrogen. # **Conclusions:** The Biowy Multi-Chamber Freezing Bag Set is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...