P990041 · DiaSorin, Inc. · Mar 30, 2001 · Microbiology
Device Facts
Record ID
P990041
Device Name
DIASORIN ETI-AB-EBK PLUS ASSAY
Applicant
DiaSorin, Inc.
Decision Date
Mar 30, 2001
Decision
APWD
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
P990041 · Mar 30, 2001
DIASORIN ETI-AB-EBK PLUS ASSAY
DiaSorin, Inc.
Frozen repository samples; Archived serial samples from plasmapheresis donors
Retrospective clinical studies were conducted to assess the performance of the ETI-AB-EBK PLUS assay in detecting anti-HBe using archived samples from diverse patient populations and serial samples from individuals undergoing plasmapheresis.
Retrospective study; Repository samples; Archived serial samples; Hepatitis B
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective clinical studies; Retrospective analysis of frozen repository samples
Patients with chronic hepatitis B, serologically diagnosed HBV infection, general hospital patients; Sample Size: 650; Number of Sites: 4 (3 US labs, 1 DiaSorin Italy)
Not applicable for this study
Performance in detecting anti-HBe compared to reference serological marker patterns
Acute Serial Seroconversion Panels; Retrospective analysis of archived serial samples; Follow-up/Duration: Up to 355 days post-HBsAg detection
ETI-AB-EBK PLUS is an in vitro enzyme immunoassay (EIA) intended for use in the qualitative detection of total antibodies to hepatitis Be antigen (anti-HBe) in human serum or plasma (EDTA, citrate or heparin). The ETI-AB-EBK PLUS is intended for manual use and with the Biochem Immunosystems Labotech/ETI-LAB automated instrument. The detection of anti-HBe is indicative of seroconversion and progression to hepatitis B virus (HBV) clearance. A reactive test is presumptive laboratory evidence of HBV seroconversion. Further HBV serological marker testing is required to define the specific disease-state. The anti-HBe assay’s performance has not been established for the monitoring of HBV disease or therapy.
Device Story
ETI-AB-EBK PLUS is an enzyme immunoassay (EIA) for qualitative detection of total anti-HBe antibodies in human serum or plasma. Device uses polystyrene microwells coated with mouse monoclonal anti-HBeAg antibodies. Principle: competitive assay; patient specimen and recombinant HBeAg are incubated in coated wells; anti-HBe in specimen competes with coated antibody for recombinant HBeAg. Enzyme tracer (HRP-labeled anti-HBeAg) binds to remaining antigen-antibody complexes. Chromogen/substrate (TMB/H2O2) added; HRP reduces substrate, oxidizing TMB to blue color. Color intensity measured spectrophotometrically; inversely proportional to anti-HBe concentration. Used in clinical laboratories; manual or automated (Biochem Immunosystems Labotech/ETI-LAB). Results compared to cutoff value derived from calibrator. Assists clinicians in defining HBV disease state by identifying seroconversion.
Clinical Evidence
Clinical performance evaluated via prospective (n=136) and retrospective (n=650) studies. Prospective study showed 100% negative agreement across all HBV classifications. Retrospective study demonstrated high agreement with reference assays: 100% positive agreement for acute infection (54/54), 93.5% for chronic infection (29/31), and 100% negative agreement for recovery, vaccine response, and past infection. Analytical sensitivity is 0.20 PEI U/mL. Reproducibility studies (manual) showed satisfactory performance across sites, lots, and days. Plasma matrix equivalence (citrate, heparin, EDTA) confirmed.
Indicated for qualitative detection of total anti-HBe antibodies in human serum or plasma to identify HBV seroconversion and progression to viral clearance. Intended for use in patients undergoing HBV serological testing. Performance not established for monitoring disease or therapy.
Submission Summary (Full Text)
{0}
1
# Summary of Safety and Effectiveness
## 1. General Information
1.1. Device Generic Name: Immunoassay for antibodies to hepatitis Be antigen
1.2. Device Trade Name: ETI-AB-EBK PLUS
1.3. Applicant’s Name and Address:
DiaSorin
Via Crescentino
Saluggia (VC) 13040, Italy
1.4. U.S. Representative:
Sienna Partners, LLC
P.O. Box 103
Baldwin, MD 21013
1.5. PMA Number: P990041
1.6. Date of Panel Recommendation: January 20, 2000
1.7. Date of Notice of Approval to the Applicant: March 30, 2001
## 2. Indications For Use
ETI-AB-EBK PLUS is an *in vitro* enzyme immunoassay (EIA) intended for use in the qualitative detection of total antibodies to hepatitis Be antigen (anti-HBe) in human serum or plasma (EDTA, citrate or heparin). The ETI-AB-EBK PLUS is intended for manual use and with the Biochem Immunosystems Labotech/ETI-LAB automated instrument.
The detection of anti-HBe is indicative of seroconversion and progression to hepatitis B virus (HBV) clearance. A reactive test is presumptive laboratory evidence of HBV seroconversion. Further HBV serological marker testing is required to define the specific disease-state. The anti-HBe assay’s performance has not been established for the monitoring of HBV disease or therapy.
## 3. Device Description
3.1. Principle of The Assay
ETI-AB-EBK PLUS uses monoclonal antibodies to hepatitis Be antigen as the basis for this enzyme immunoassay. The assay is a competitive test based on the use of polystyrene microwells coated with mouse monoclonal antibodies to HBeAg and an enzyme tracer containing horseradish-peroxidase-labeled mouse monoclonal antibodies to HBeAg. The monoclonal antibodies are of the IgG1-k class and are directed to the secreted form of HBeAg.
{1}
In the assay procedure, patient specimens or controls and neutralizing solution (recombinant HBeAg) are incubated with an incubation buffer in antibody-coated microwells. If anti-HBe is present in a specimen or control, it competes with the antibody coated on the microwell for the recombinant HBeAg. Excess sample and recombinant HBeAg are removed by a wash step, and the enzyme tracer is then added to the microwells and allowed to incubate. The enzyme tracer binds to any antigen-antibody complexes present in the microwells. The quantity of enzyme tracer that binds to the solid phase via recombinant HBeAg and the subsequent enzyme activity are inversely related to the presence of anti-HBe in the specimen or control. Excess enzyme tracer is removed by a wash step, and a chromogen/substrate solution is added to the microwells and allowed to incubate. If a sample does not contain anti-HBe, the bound enzyme (horseradish peroxidase) chemically reduces the substrate peroxide, which concurrently oxidizes the chromogen tetramethylbenzidine (TMB) to a blue color. The blue color turns to yellow after addition of the stop solution. If a sample contains anti-HBe, the microwell will be colorless after the chromogen/substrate solution is added and will remain colorless after the stop solution is added. Color intensity, which is measured spectrophotometrically, is inversely related to the presence of anti-HBe. Absorbance value readings for patient specimens are compared to a cutoff value determined from the mean absorbance value of the calibrator.
## 3.2. Kit Configuration and Components
For detection of antibodies to HBeAg, the ETI-AB-EBK PLUS system is comprised of the following:
Coated Strips
Microwells coated with mouse monoclonal antibodies to HBeAg.
Enzyme Tracer
Horseradish peroxidase-labeled mouse monoclonal antibody to HBeAg, buffer, protein stabilizers.
Preservative: 0.2% ProClin 300.
Tracer Diluent
Buffer, protein stabilizers.
Preservative: 0.2% ProClin 300.
Calibrator (Human)
Human serum/plasma non-reactive for all known HBV markers.
Preservative: 0.2% ProClin 300.
Negative Control (Human)
Human serum/plasma non-reactive for all known HBV markers.
Preservative: 0.2% ProClin 300.
Positive Control
HBeAg (recombinant DNA), protein stabilizers.
Preservative: 0.2% ProClin 300.
Neutralizing Solution (Recombinant HBeAg)
HBeAg, buffer, human serum/plasma, protein stabilizers.
Preservative: 0.2% ProClin 300
Incubation Buffer
Buffer, protein stabilizers, an inert blue dye.
2
{2}
Preservative: 0.2% ProClin 300.
Wash Buffer (concentrate)
Buffer, detergents, preservatives.
Chromogen/Substrate
Tetramethylbenzidine/hydrogen peroxide system.
Stop Solution
1N sulfuric acid
Strip Sealers
Plate Sealers
Pouch Sealer
4. Contraindications
None
5. Warnings and Precautions
For in vitro diagnostic use only.
Warnings and precautions for users of the ETI-AB-EBK PLUS assay are stated in the product labeling.
6. Alternative Practices and Procedures
Several methods are available for detecting anti-HBe in human serum or plasma. Methods commonly used include immunodiffusion, counterimmunoelectrophoresis, passive hemagglutination, radioimmunoassay, and enzyme immunoassay
7. Marketing History
The ETI-AB-EBK-PLUS kits have never been marketed in the US or outside the US.
8. Potential Adverse Effects of the Device on Health
Failure of the product to perform as indicated or human error in use of the product may lead to a false result. A false result cannot be considered a patient or public health concern because the overall hepatitis B marker pattern will indicate the correct disease-state interpretation.
9. Summary of Preclinical Studies
9.1. Comparison of Labotech Instrumentation with the Manual Assay
An instrument application study was conducted at DiaSorin, Saluggia Italy, to evaluate the performance of the ETI-AB-EBK PLUS assay on the Biochem Immunosystems Labotech/ETI-LAB, an automated microplate processing instrument, compared to the manual analysis. The Paul-Ehrlich-Institut (PEI) Standard, 12 serum samples near the ETI-AB-EBK PLUS cutoff and samples from the clinical trials (21 suspected hepatitis B patients and 19 apparently healthy adults) were tested in parallel manually and on the Labotech.
3
{3}
Serial dilutions of the PEI Standard were prepared in fetal calf serum to obtain a panel ranging from high concentration to below the analytical sensitivity of the assay. The diluted Standard samples were tested in duplicate, one run per day for three days both manually and on the Labotech. Due to the requirement that assay cutoff be established for each plate, reproducibility was evaluated based on specimen absorbance-to-cutoff ratios (S/CO) rather than absolute absorbance values. In this assay, the S/CO is inversely related to reactivity. The $95\%$ confidence intervals were established for the S/CO values of each point of the Standard-referenced curve and therefore the analytical endpoint sensitivity was defined (first dilution with $\mathrm{S / CO} < 0.9$ ). A graph summarizing these results is presented below:

The 12 samples near the cutoff were tested in triplicate, one run per day for three days both manually and on the Labotech. The samples from the clinical trials were tested in singlet in one run on one day, both manually and on Labotech. The mean, the standard deviation and the coefficient of variation (CV%) of the S/CO values were computed by the different components of variability for each of the tested specimens. A summary of the data is presented in the following table.
{4}
| | Manual | | | Labotech/ETI-LAB | | |
| --- | --- | --- | --- | --- | --- | --- |
| Analytical Endpoint Sensitivity (0.1 PEI U/mL) | Mean | W/R %CVa | D/D %CV | Mean | W/R %CV | D/D %CV |
| S/CO [95% CI]b | 0.85 [0.76 – 0.94] | 8.5 | 8.4 | 0.76 [0.70 – 0.82] | 5.4 | 6.2 |
| 12 Cutoff Samples S/CO Range of S/CO | 0.89 0.66 – 1.05 | 6.8 | 6.2 | 0.84 0.64 – 1.03 | 7.0 | 8.7 |
| Clinical Samples: | | | | | | |
| Suspected Hepatitis B Range of S/CO | Negative: 1.11 - >3.5 (7/21) Equivocal: N/A (0/21) Positive: 0.00 – 0.87 (14/21) | | | Negative: 1.24 - >3.5 (7/21) Equivocal: 0.98 (1/21) Positive: 0.01 – 0.86 (13/21) | | |
| Healthy Adults Range of S/CO | Negative: 1.53 – 2.12 (19/19) Equivocal: N/A (0/19) Positive: N/A (0/19) | | | Negative: 1.61 – 2.07 (19/19) Equivocal: N/A (0/19) Positive: N/A (0/19) | | |
a %CVs were calculated using specimen absorbance-to-cutoff ratios (S/CO) which normalized the data plate-to-plate
b $95\%$ CI = 95% Confidence Interval; W/R = within-run; D/D = day-to-day
No reproducibility testing with the Labotech instrument was conducted. As part of the conditions of approval agreement, DiaSorin will furnish FDA results from a reproducibility study using the Labotech instrument. Until that condition is met, a statement will be placed in the labeling.
"Assay reproducibility using the Labotech has not been established. If the Labotech is used, the user should establish appropriate assay reproducibility in accordance with NCCLS EP5-A, Evaluation of Precision Performance of Clinical Chemistry Devices."
## 9.2. Analytical Sensitivity
The analytical sensitivity of the assay was evaluated using single point serial dilutions of a standard preparation from the Paul-Ehrlich-Institut (PEI), Germany. The analytical sensitivity of the assay (last positive dilution) was determined to be 0.20 PEI U/mL.
## 9.3. Potential Cross-Reacting Substances
Serum samples were obtained from patients belonging to a number of different disease categories listed below. Of the 535 potentially interfering samples, 451 (84%) were negative and 84 (16%) were positive by ETI-AB-EBK PLUS. Among the 84 positive samples, 71 were confirmed positive by reference testing, 4 were equivocal by reference testing, and 2 were QNS for reference testing. Some equivocal samples were not repeated due to volume constraints. Although there were 13 samples that were positive by ETI-AB-EBK PLUS but negative by reference testing, no pattern of interference was seen in any specific disease category. Other hepatitis B markers or additional investigation (as explained in the table footnotes) indicated no interference in the ETI-AB-EBK PLUS result. Disease category was
{5}
determined by serological testing. The following table shows the summary of cross-reactivity testing:
Cross-Reactivity Study Results
| DISEASE CATERGORY | N | ETI-AB-EBK PLUS Negative and Equivocal samples | ETI-AB-EBK PLUS Positive samples | % Confirmed Positive By Additional Testing |
| --- | --- | --- | --- | --- |
| Acute EBV infection | 16 | 16 | 0 | - |
| Acute CMV infection | 20 | 20 | 0 | - |
| Acute HSV infection | 10 | 9 | 1 | 100% (1/1) |
| Acute Toxoplasma infection | 18 | 16 | 2 | 50% (1/2)a |
| Acute parvovirus B19 infection | 5 | 5 | 0 | - |
| HTLV-I/II infection | 50 | 43 | 7 | 100% (7/7) |
| Syphilis | 26 | 24 | 2 | 50% (1/2)b |
| HCV infection | 50 | 47 | 3 | 67% (2/3)c |
| HDV infection | 20 | 4 | 16 | 100% (16/16) |
| HIV infection | 50 | 41 | 9 | 100% (9/9) |
| Acute HAV infection | 50 | 45 | 5 | 60% (3/5)d |
| Past HAV infection | 50 | 33e | 17 | 82% (14/17)f |
| Rheumatoid factor (RF) + | 40 | 36 | 4 | 25% (1/4)g |
| Autoimmune disease | 19 | 18 | 1 | 100% (1/1) |
| Systemic lupus erythematosus (SLE) | 11 | 10 | 1 | 100% (1/1) |
| Autoimmune hepatitis | 5 | 5 | 0 | - |
| Myeloma | 20 | 17 | 3 | 100% (3/3) |
| Hypergammaglobulinemia | 20 | 19 | 1 | 100% (1/1) |
| Influenza vaccine | 5 | 5 | 0 | - |
| Elevated liver enzymes | 10 | 8 | 2 | 100% (2/2) |
| Non-viral liver disease | 30 | 20h | 10 | 80% (8/10)i |
| E. coli infection | 10 | 10 | 0 | - |
| TOTAL | 535 | 451 (84%) | 84 (16%) | 85% (71/84) |
a Positive on repeat testing with reference method and anti-HBc positive
b QNS for reference testing; marker pattern indicates recovery; both positive and negative anti-HBe results acceptable
{6}
7
c Anti-HBe equivocal by reference method; samples was anti-HBc and anti-HBe positive, indicating recovery; both positive and negative anti-HBe results acceptable
d One sample QNS for reference testing; one sample equivocal by reference testing, but was anti-HBc and anti-HBs positive, indicating recovery – both positive and negative anti-HBe results acceptable
e One sample was DiaSorin equivocal
f Two samples were anti-HBc and anti-HBs positive, indicating recovery; both positive and negative anti-HBe results are acceptable.
g One sample was positive on repeat reference testing; one sample was equivocal by reference testing and other markers gave uninterpretable pattern
h Two samples were DiaSorin equivocal
i One sample was repeat equivocal by reference testing and other markers gave uninterpretable pattern; one sample was positive on repeat testing with reference method and anti-HBc positive
Very high concentrations of the monoclonal antibody (approx. 83 times higher than used in the assay) have been shown to react with HBcAg.
## 9.4. Interfering Substances
DiaSorin ETI-AB-EBK PLUS assay was evaluated for interference by testing the following substances listed below. Testing was performed using matched pairs of negative donor serum and negative donor serum spiked with a high titer HBeAg samples to obtain a result near the cutoff. None of the compounds at the levels indicated were found to interfere with the clinical interpretation of the assay in serum. No interference was found with bilirubin in plasma (EDTA, heparin or citrate), testing for interference with hemoglobin and triolein was not performed in plasma.
| Compound | Concentration | |
| --- | --- | --- |
| Bilirubin | 0.35 mmol/L | 20 mg/dL |
| Hemoglobin | 0.06 mmol/L | 100 mg/dL |
| Triolein | 33.9 mmol/L | 3000 mg/dL |
The ETI-AB-EBK PLUS assay was also evaluated for possible interference from heterophilic anti-mouse antibodies (HAMA). A dilutional panel was used, consisting of 21 samples prepared from a stock pool of high positive human serum. The HAMA concentrations in the samples ranged from 0 to $2975.5\mathrm{ng / mL}$, as determined by a HAMA ELISA. In a direct non-competitive assay, such as ETI-AB-EBK PLUS, interference would manifest as false positive results. No interference was seen in that all 21 dilutions were negative by the ETI-AB-EBK PLUS assay, and supported by positive anti-HBe results.
## 9.5. Stability Studies
### 9.5.1. Kit Stability
Stability studies were performed on 3 different ETI-AB-EBK PLUS kit lots. At specified intervals from time of kit release, performance of the kits was evaluated testing the Calibrator, Negative and Positive Controls, and Q.C. sera panel according
{7}
to the instructions for use. The kit must meet established acceptance criteria. The obtained stability data demonstrate that the kit performance is acceptable for at least 11 months. On the basis of the stability results, a shelf life of 11 months has been established for the kit.
## 9.5.2. Working Enzyme Tracer Stability
The Enzyme Tracer was diluted with the Tracer Diluent to obtain the working Enzyme Tracer according to the instructions for use. After 7 days from dilution, the performance of the kit was evaluated, according to the instructions for use, testing various specimens with freshly prepared working Enzyme Tracer and the 7-days old working Enzyme Tracer. The kit must meet established acceptance criteria. The tests on the working Enzyme Tracer demonstrate that the performance of the kit is acceptable when the 7 day-diluted Enzyme Tracer is used. The working Enzyme Tracer can be used for one week if stored at 2-8 °C.
## 9.5.3. Working Wash Buffer Stability
The Wash Buffer concentrate was diluted with deionized water according to the instruction for use to obtain the working Wash buffer. After 7 days from dilution, the performance of the kit was evaluated by testing various specimens with a freshly prepared working Wash buffer and the 7-days-old working Wash buffer, according to the instructions for use. The kit must meet established acceptance criteria. The tests on the working Wash Buffer demonstrate that the performance of the kit is acceptable when the 7 day-old working Wash Buffer is used. The working Wash Buffer can be used for one week if stored at 2-8 °C.
## 9.6. Common Reagents Interchangeability Study
Studies were performed to demonstrate that the lots of some components included with ETI-AB-EBK PLUS kit and common to all kits of ETI-PLUS line (Wash Buffer, Chromogen/Substrate, Stop Solution), can be exchanged with other lots of the same component produced for the ETI-PLUS line (interchangeability). Three lots of Wash Buffer, Chromogen/Substrate and Stop Solution were combined with one lot of ETI-AB-EBK PLUS; the three combinations were then tested with various samples. Regression analysis was applied to the Optical Densities of the samples. The regression analyses for the three studies gave slopes close to 1.0, with low intercepts and excellent correlation values. These results indicate that the use of different lots of Wash Buffer, Chromogen/Substrate and Stop Solution with the same ETI-AB-EBK PLUS lot gave equivalent results with samples distributed over the range of reactivity, confirming their interchangeability.
## 9.7. Reproducibility
Manual Assay: Intra-assay, inter-assay, inter-lot, and inter-site variability studies were carried out on the ETI-AB-EBK PLUS kit to test the variability within runs, between runs, between days, between kit lots, and between test sites. Variability was measured on a panel of ten sera that included negative, borderline, and positive samples. Three ETI-AB-EBK PLUS kit lots were tested at three independent test sites. Due to the requirement that assay cutoff be established for each plate, reproducibility was evaluated based on specimen
8
{8}
absorbance to cutoff ratios (S/CO) rather than absolute absorbance values. The results of that study are tabulated below showing the reproducibility of the assay to be satisfactory.
Clinical Site Reproducibility Study
| ID# | | Number of tests per sample | Mean S/CO | Within-run %CV* | Between-runs %CV | Between-lots %CV | Between-days %CV | Between-sites %CV | Total |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| S01 | High | 108 | 20.03 | 6.37 | 6.29 | 10.07 | 7.30 | 5.14 | 14.47 |
| S02 | High | 108 | 10.83 | 4.16 | 5.07 | 11.56 | 7.34 | 9.44 | 15.69 |
| S03 | High | 108 | 6.83 | 4.47 | 7.16 | 11.78 | 7.09 | 6.73 | 17.18 |
| S04 | Low | 108 | 3.84 | 3.80 | 8.25 | 14.42 | 6.97 | 5.79 | 17.12 |
| S05 | Equiv | 108 | 0.80 | 6.08 | 6.79 | 8.51 | 6.38 | 3.79 | 13.03 |
| S06 | Low | 108 | 2.61 | 5.35 | 5.99 | 9.72 | 7.92 | 9.80 | 14.35 |
| S07 | Low | 108 | 2.11 | 3.55 | 10.49 | 10.44 | 7.47 | 10.18 | 17.97 |
| S08 | Low | 108 | 1.96 | 5.08 | 11.86 | 10.47 | 6.86 | 7.30 | 16.18 |
| S09 | Equiv | 108 | 1.67 | 6.27 | 11.85 | 10.47 | 4.04 | 4.04 | 19.19 |
| S10 | Neg | 108 | 0.06 | 20.20 | 36.73 | 36.18 | 15.26 | 29.61 | 89.32 |
* %CVs were calculated using specimen absorbance-to-cutoff ratios (S/CO) which normalized the data plate-to-plate
# 9.8. Plasma Reproducibility
A plasma reproducibility study was conducted at DiaSorin, Saluggia Italy, to evaluate the performance of the manual ETI-AB-EBK PLUS assay on serum versus a variety of plasma types. The plasma types evaluated were citrate, heparin and EDTA. Sample sets of matched serum/multiple plasma were used in the study. A sample set was prepared by spiking the same high positive sample into each of the matrices (serum and plasmas) resulting in a total of four specimens per set around the cutoff. Several high positive samples were used in the preparation of the 12 different near-cutoff sample sets. Six matched serum/multiple plasma samples sets were tested in triplicate in each run; thus there were two runs per day for three days, all tested in a manual mode. Due to the requirement that assay cutoff be established for each plate, reproducibility was evaluated based on specimen absorbance to cutoff ratios (S/CO) rather than absolute absorbance values. The mean, the standard deviation and the coefficient of variation $(\mathrm{CV}\%)$ of the S/CO values were computed by the different components of variability for each of the tested specimens. The $95\%$ confidence intervals were established for the S/CO values of all serum samples and each plasma type. A summary of the data is presented in the following table. The study supports the use of plasma specimens in the ETI-AB-EBK PLUS assay.
{9}
| | SERUM | | CITRATE | HEPARIN | EDTA |
| --- | --- | --- | --- | --- | --- |
| Mean (S/CO) | 0.99 | | 0.99 | 1.02 | 1.04 |
| 95% CI* | [0.95 - 1.03] | | [0.95 – 1.03] | [0.98 – 1.06] | [1.00 – 1.08] |
| W/R %CV** | 5.9% | | 4.8% | 5.4% | 4.5% |
| D/D %CV | 9.3% | | 7.7% | 7.2% | 8.5% |
| Total %CV | 10.4% | | 8.6% | 8.6% | 8.9% |
| Between matrix %CV | | 7.9% | | | |
| Across matrix total %CV | | 11.7% | | | |
* CI = 95% Confidence Interval; W/R = within-run; D/D = day-to-day.
** %CVs were calculated using specimen absorbance-to-cutoff ratios (S/CO) which normalized the data plate-to-plate
## 9.9. Acute Serial Seroconversion Panels
One hundred twenty-four (124) archived serial samples from nine individuals were tested for the appearance of anti-HBeAg. Most (8/9) of these individuals were defined as being acutely infected by the appearance of HBsAg and HBeAg with the subsequence appearance of IgM anti-HBc, total HBc, anti-HBe, and anti-HBs. One individual had detectable HBsAg but did not have detectable HBeAg in any specimen. However, this individual did seroconvert for anti-HBe.
The specimens were collected from individuals undergoing plasmapheresis for further manufacturing purposes. Three individuals were found to be infected with HBV during the first plasmapheresis and others became infected with HBV during subsequent plasmapheresis. It is unknown how long these three initially HBsAg reactives were infected prior to the first plasmapheresis. All nine individuals underwent sequential plasmapheresis after becoming HBV infected. However, the timing of subsequent plasmapheresis varied from individual to individual. The specimens draw times were normalized to represent the day that HBsAg was first detected by a FDA licensed assay as Day 0. For the remaining specimens, draw days ranged from day 0 (HBsAg first detected) through day 355 post day 0. Since all panels did not contain the same draw day, sample results were grouped within day intervals (e.g., days 0, 1-10, 11-20, etc., representing days since first detection of HBsAg).
The results are summarized in the following table and graph. All specimens were reactive for anti-HBe by the DiaSorin assay after day 90. In the graph below the graph for the reference HBsAg percent reactive has been overlaid for reference.
{10}
Acute Serial Panel Results
| Day Range | Number Specimens | DiaSorin anti-HBe Reactive | % Positive |
| --- | --- | --- | --- |
| 0 | 9 | 1 | 11.1% |
| 1-10 | 10 | 1 | 10.0% |
| 11-20 | 13 | 3 | 23.1% |
| 21-30 | 9 | 3 | 33.3% |
| 31-40 | 10 | 5 | 50.0% |
| 41-50 | 6 | 5 | 83.3% |
| 51-60 | 10 | 8 | 80.0% |
| 61-70 | 9 | 6 | 66.7% |
| 71-80 | 6 | 4 | 66.7% |
| 81-90 | 9 | 7 | 77.8% |
| 91-100 | 10 | 10 | 100.0% |
| 101-110 | 6 | 6 | 100.0% |
| 111-120 | 4 | 4 | 100.0% |
| 121-130 | 4 | 4 | 100.0% |
| 131-140 | 3 | 3 | 100.0% |
| 141-150 | 2 | 2 | 100.0% |
| 151-160 | 0 | 0 | NA |
| 161-170 | 1 | 1 | 100.0% |
| 171-180 | 0 | 0 | NA |
| 181-190 | 1 | 1 | 100.0% |
| 191-200 | 1 | 1 | 100.0% |
| 355 | 1 | 1 | 100.0% |

{11}
9.10. Expected Values Study
The 236 prospective samples used in the expected values study for the DiaSorin ETI-AB-EBK PLUS assay were from patients who were sent to the laboratory for HBV testing. Of those, 100 (42%) were frozen and 136 (58%) were fresh. The patients represented Florida, Georgia, Pennsylvania, California, Utah, and the southeastern US. The group was 69% (162/236) female and 31% (74/236) male; the ethnicity of the patients was unspecified. The ages ranged from 5 to 88 years old. The percent DiaSorin ETI-AB-EBK PLUS positive results observed in these samples was 11%.
The table below summarizes the percent DiaSorin ETI-AB-EBK PLUS positive and negative results by gender and age range. There were 6 samples for which gender and age were not reported; they were all positive. There were 6 samples for which age was not reported, 2 were from females and 4 were from males; all were negative. These 12 results were not included in the table.
Expected Values Summary
| | DiaSorin ETI-AB-EBK PLUS | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | + | | - | | E | | TOTAL |
| Age Range | Gender | n | % | n | % | N | % | |
| 0-9 | F | 0 | 0% | 2 | 100% | 0 | 0% | 2 |
| | M | 0 | 0% | 0 | 0% | 0 | 0% | 0 |
| 10-19 | F | 2 | 12% | 15 | 88% | 0 | 0% | 17 |
| | M | 0 | 0% | 2 | 100% | 0 | 0% | 2 |
| 20-29 | F | 3 | 6% | 47 | 92% | 1 | 2% | 51 |
| | M | 1 | 8% | 12 | 92% | 0 | 0% | 13 |
| 30-39 | F | 3 | 6% | 46 | 94% | 0 | 0% | 49 |
| | M | 2 | 12% | 15 | 88% | 0 | 0% | 17 |
| 40-49 | F | 5 | 25% | 15 | 75% | 0 | 0% | 20 |
| | M | 5 | 36% | 9 | 64% | 0 | 0% | 14 |
| 50-59 | F | 0 | 0% | 5 | 100% | 0 | 0% | 5 |
| | M | 2 | 25% | 6 | 75% | 0 | 0% | 8 |
| 60-69 | F | 0 | 0% | 3 | 100% | 0 | 0% | 3 |
| | M | 0 | 0% | 2 | 100% | 0 | 0% | 2 |
| 70-79 | F | 1 | 10% | 9 | 90% | 0 | 0% | 10 |
| | M | 0 | 0% | 5 | 100% | 0 | 0% | 5 |
| 80-89 | F | 0 | 0% | 3 | 100% | 0 | 0% | 3 |
| | M | 0 | 0% | 3 | 100% | 0 | 0% | 3 |
| TOTAL | | 24 | 11% | 199 | 88% | 1 | <1% | 224 |
{12}
13
# 10. Summary of Clinical Studies
## 10.1. Clinical Sample Testing
### 10.1.1 Prospective Samples
A study of 136 prospective specimens was conducted. These specimens represented individuals who were sent to the laboratory for hepatitis testing. Specimens were collected at a reference laboratory and assayed at the California clinical trial site. The patients were 86% (117/136) female and 14% (19/136) male. The ages ranged from 5 to 77 years old, with 3 specimens not specified. The study (testing) sites were blinded to the previous specimen categorization. All testing was performed by the manual ETI-EBK PLUS procedure. Specimens were characterized by testing with six HBV serological markers (HBsAg, HBeAg, IgM anti-HBc, total anti-HBc, anti-HBe, anti-HBs) using FDA-licensed or approved assays. Testing with these assays followed the FDA-licensed or approved procedure, including confirmation by neutralization of repeatably reactive HBsAg samples.
## Results by Specimen Classification
After study completion all samples were assigned a specimen classification based on the patterns of the six HBV serological markers established by the reference assays. Based on these serological marker patterns, the samples were categorized into the HBV classifications described in the table below. There were 5 unique HBV marker patterns observed in the DiaSorin ETI-AB-EBK PLUS prospective clinical studies.
| Characterization Based On Single Point Specimen | HBsAg | HBeAg | IgM anti-HBc | Total anti-HBc | anti-HBe | anti-HBs | n |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Chronic Infection | + | - | - | + | + | - | 1 |
| Recovery | - | - | - | + | + | + | 2 |
| Past Infection | - | - | - | + | - | + | 4 |
| | - | - | - | + | - | - | 4 |
| HBV Vaccine Response | - | - | - | - | - | + | 38 |
| Not Previously Infected with HBV | - | - | - | - | - | - | 87 |
Based on the above classifications the DiaSorin anti-HBe results for the prospective samples were compared to a reference assay's anti-HBe results. The following table shows this comparison and percent agreement with 95% confidence intervals with the reference HBeAg assay.
{13}
14
# Prospective Samples Comparison
| | – | Reference anti-HBe |
| --- | --- | --- |
| Reference Serology Classification | – | DiaSorin anti-HBe |
| Chronic infection | 1 | 1 |
| Recovery | 2 | 2 |
| Past Infection | 8 | 8 |
| HBV vaccine response | 38 | 38 |
| Not previously infected with HBV | 87 | 87 |
| Grand Total | 136 | 136 |
# Prospective Samples Agreement Rates
Chronic Infection Positive agreement with reference assay results = NA (0/0)
95% CI = NA
Negative agreement with reference assay results = 100% (1/1)
95% CI = 2.5 to 100%
Recovery Positive agreement with reference assay results = NA (0/0)
95% CI = NA
Negative agreement with reference assay results = 100% (2/2)
95% CI = 15.8 to 100%
HBV Vaccine Response Positive agreement with reference assay results = NA (0/0)
95% CI = NA
Negative agreement with reference assay results = 100% (38/38)
95% CI = 90.8 to 100%
Past Infection Positive agreement with reference assay results = NA (0/0)
95% CI = NA
Negative agreement with reference assay results = 100% (8/8)
95% CI = 63.1 to 100%
Not Previously Infected Positive agreement with reference assay results = NA (0/0)
95% CI = NA
Negative agreement with reference assay results = 100% (87/87)
95% CI = 98.5 to 100%
## 10.1.2. Retrospective Samples
Retrospective studies were carried out at three clinical laboratories in the United States (California, Missouri, and Minnesota) and at DiaSorin (Italy) to assess the performance of the ETI-AB-EBK PLUS assay in detecting anti-HBe. The study set included 650 frozen repository samples (the majority of which were purchased from commercial vendors) from the following populations:
{14}
- patients with chronic hepatitis B infection (diagnosed by a physician as having HBsAg positive for greater than 6 months) – 111 frozen repository samples;
- patients with serologically diagnosed hepatitis B infection (acute, chronic, asymptomatic, convalescent, etc.) – 82 frozen repository samples;
- patients sent to the laboratory for hepatitis B testing – 100 frozen repository samples;
- a general hospital patient population – 357 frozen repository samples.
The specimens represented Midwestern (2%), Southeastern (25%), Western (13%), and Northeastern US (2%), outside of the US (1%), and unspecified (57%). The group was 44% (287/650) female, 42% (270/650) male, and 14% (93/650) unspecified. Approximately 13% (84/650) were Caucasian, 4% (27/650) were African American, <1% (5/650) were Hispanic, <1% (3/650) were Asian, and 82% (531/650) were unspecified. The ages ranged from 5 to 98 years old, with 131 specimens not specified.
The study (testing) sites were blinded to the previous specimen categorization. All testing was performed by the manual ETI-AB-EBK PLUS procedure. Specimens were characterized by testing with six HBV serological markers (HBsAg, HBeAg, IgM anti-HBc, total anti-HBc, anti-HBe, anti-HBs) using FDA-licensed or approved assays. Testing with these assays followed the FDA-licensed or approved procedure with the exception of the HBsAg assay at two of the three sites. At these sites, the majority of specimens that were initially HBsAg-positive were repeated in duplicate, however the repeatedly reactive specimens were not confirmed by the licensed HBsAg confirmation assay at the two sites. Therefore, true HBsAg result was determined in one of three ways: 1) confirmed by reference assay neutralization during clinical trials, 2) based on a statement by the attending physician that HBsAg was positive for greater than 6 months, or 3) information provided by the vendor regarding confirmatory testing performed at their location or by the material source facility
## Results by Specimen Classification
After study completion all samples were assigned a specimen classification based on the patterns of the six HBV serological markers established by the reference assays. Based on these serological marker patterns, the samples were categorized into the HBV classifications described in the table below. There were 35 unique HBV marker patterns observed in the ETI-AB-EBK PLUS retrospective clinical studies.
15
{15}
| Characterization Based On Single Point Specimen | HBsAg | HBeA g | IgM anti-HBc | Total anti-HBc | anti-HBe | anti-HBs | n |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Acute infection | + | + | + or I | + | - | - | 52 |
| | + | - | + or I | + | + | - | 4 |
| | + | - | - | - | - | - | 2 |
| | + | + | - | - | - | - | 2 |
| Chronic Infection | + | - | - | + | + | - | 82 |
| | + | + | - | + | - | - | 21 |
| | + | - | - | + | - or I | - | 23 |
| | + | + | + | + | - | + | 4 |
| | + | + | - or I | + | - | + | 2 |
| | + | - | - | + | + | + | 2 |
| | + | + | - | + | + or I | + | 2 |
| | + | + | + | + | + | + | 1 |
| | + | + | - | + | + | - | 1 |
| | + | - | - | + | - | + | 1 |
| Recovery | - | - | - | + | + or I | + | 40 |
| | - | - | - | + | + | - | 6 |
| | - | - | + | + | + | - | 2 |
| | - | - | + or I | + | + | + | 2 |
| Past Infection | - | - or I | - | + | - | + | 12 |
| | - | - | - | + | - | - | 9 |
| HBV Vaccine Response | - | - | - | - | - | + | 20 |
| Not Previously Infected with HBV | - | - | - | - | - | - | 343 |
| Uninterpretable | - | + or I | - | - | - | - | 13 |
| | - | + | - | + | - | + | 2 |
| | - | + | - | + | + | + | 1 |
| | - | I | - | + | - | - | 1 |
*I = indeterminate result
Based on the above classifications the ETI-AB-EBK PLUS results for the retrospective samples were compared to a reference assay's results as determined by the methods above. The following table shows this comparison and percent agreement with $95\%$ confidence intervals with the reference anti-HBe results.
{16}
Retrospective Samples Comparison
| | - | | + | | I$ | Reference HBeAg |
| --- | --- | --- | --- | --- | --- | --- |
| Reference Serology Classification | - | + | - | + | - | DiaSorin HBeAg |
| Acute infection | 5 | 1 | 0 | 54 | 0 | 60 |
| Chronic infection | 107 | 1 | 2 | 29 | 0 | 139 |
| Recovery | 50 | 0 | 0 | 0 | 0 | 50 |
| HBV vaccine response | 20 | 0 | 0 | 0 | 0 | 20 |
| Not previously infected with HBV | 343 | 0 | 0 | 0 | 0 | 343 |
| Past Infection | 20 | 0 | 0 | 0 | 1 | 21 |
| Uninterpretable | 0 | 0 | 9 | 1 | 7 | 17 |
| Total | 545 | 2 | 11 | 84 | 8 | 650 |
$ Indeterminate result
## Retrospective Samples Agreement Rate
### Acute Infection
Positive agreement with reference assay results = 100% (54/54)
95% CI = 93.4 to 100%
Negative agreement with reference assay results = 83.3% (5/6)
95% CI = 35.9 to 99.6%
### Chronic Infection
Positive agreement with reference assay results = 93.5% (29/31)
95% CI = 78.6 to 99.2%
Negative agreement with reference assay results = 99.1% (107/108)
95% CI = 95.0 to 100%
### Recovery
Positive agreement with reference assay results = NA (0/0)
95% CI = NA
Negative agreement with reference assay results = 100% (50/50)
95% CI = 92.9 to 100%
### HBV Vaccine Response
Positive agreement with reference assay results = NA (0/0)
95% CI = NA
Negative agreement with reference assay results = 100% (20/20)
95% CI = 83.2 to 100%
### Not Previously Infected
Positive agreement with reference assay results = NA (0/0)
95% CI = NA
Negative agreement with reference assay results = 100% (343/343)
95% CI = 98.9 to 100%
### Past Infection
Positive agreement with reference assay results = NA (0/0)
95% CI = NA
Negative agreement with reference assay results = 100% (20/20)
95% CI = 83.2 to 100%
17
{17}
Uninterpretable
Positive agreement with reference assay results = 10.0% (1/10)
95% CI = 0.3 to 44.5%
Negative agreement with reference assay results = NA (0/0)
95% CI = NA
## 11. Conclusions Drawn from Studies
The study data demonstrates that acceptable performance is obtained with the DiaSorin ETI-AB-EBK PLUS assay when testing specimens collected in serum and plasma. The DiaSorin assay shows acceptable within-run, between-run, between-day, site-to-site, and lot-to-lot reproducibility. The quality control procedures described in the package insert are appropriate to assure accurate assay performance. The data from this study provide reasonable assurance that the DiaSorin ETI-AB-EBK PLUS assay is safe and effective for its stated purpose when used as instructed in the package insert. The DiaSorin ETI-AB-EBK PLUS assay can be stored up to 11 months at 2-8°C.
## 12. Panel Recommendations:
The Microbiology Advisory Panel met on January 20, 2000, to consider the safety and effectiveness of the ETI-AB-EBK PLUS assay. The panel recommended approval subject to the following conditions.
- Conduct additional studies for the immunity claim by testing individuals immediately after receiving the complete series of three vaccinations with the hepatitis B virus vaccines and three to nine months later.
- Provide more data on acute/chronic infections in high-risk populations such as those individuals that are infected with HIV, sexually transmitted diseases, and those patients that are immunosuppressed.
- Collect more data on patients meeting the standard definition for chronicity, i.e., >6 months of infection.
## 13. CDRH Decision
CDRH concurred with the Panel’s recommendation. DiaSorin Inc. has provided some additional data to address some of the Panel’s issues and those issues not fully resolved were addressed with labeling restrictions and the requirement of postapproval studies. The two postapproval studies were:
- Within 6 months of this approval, DiaSorin should submit a reproducibility study for the Biochem Immunosystems Labotech/ETI-Lab automated instrument.
- To address the concerns made by the Panel regarding the retrospective nature of the clinical studies, within 2 years of this approval, DiaSorin should submit the results of an additional prospective clinical study. We suggest that this study involve individuals that may be considered representative of an U.S. population, i.e., similar prevalence of HBV disease and serotypes.
18
{18}
The applicant's manufacturing facility was found to be in compliance with the Quality Systems Regulation (21 CFR 820).
CDRH issued an approval order on March 30, 2001.
## 14. Approval Specifications
Directions for use: See Labeling
Conditions of Approval: CDRH Approval of this PMA is subject to full compliance with the conditions described in the approval order.
19
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.