TECNIS SYMFONY EXTENDED RANGE OF VISION INTRAOCULAR LENS
P980040S065 · Johnson & Johnson Surgical Vision, Inc. · POE · Jul 15, 2016 · Ophthalmic
Device Facts
Record ID
P980040S065
Device Name
TECNIS SYMFONY EXTENDED RANGE OF VISION INTRAOCULAR LENS
Applicant
Johnson & Johnson Surgical Vision, Inc.
Product Code
POE · Ophthalmic
Decision Date
Jul 15, 2016
Decision
APPR
Regulation
21 CFR 886.3600
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
The TECNIS® Symfony Extended Range of Vision IOL, Model ZXR00, is indicated for primary implantation for the visual correction of aphakia, in adult patients with less than 1 diopter of pre-existing corneal astigmatism, in whom a cataractous lens has been removed. The lens mitigates the effects of presbyopia by providing an extended depth of focus. Compared to an aspheric monofocal IOL, the lens provides improved intermediate and near visual acuity, while maintaining comparable distance visual acuity. The Model ZXR00 IOL is intended for capsular bag placement only. The TECNIS® Symfony Toric Extended Range of Vision IOLs, Models ZXT150, ZXT225, ZXT300, and ZXT375, are indicated for primary implantation for the visual correction of aphakia and for reduction of residual refractive astigmatism in adult patients with greater than or equal to 1 diopter of preoperative corneal astigmatism, in whom a cataractous lens has been removed. The lenses mitigate the effects of presbyopia by providing an extended depth of focus. Compared to an aspheric monofocal IOL, the lenses provide improved intermediate and near visual acuity, while maintaining comparable distance visual acuity. The Model Series ZXT IOLs are intended for capsular bag placement only.
Device Story
TECNIS® Symfony IOLs are ultraviolet light-absorbing posterior chamber lenses designed for capsular bag placement to replace the natural crystalline lens. The device features a biconvex optic with a proprietary wavefront-designed aspheric or toric-aspheric anterior surface to correct corneal spherical aberration. The posterior optic incorporates an achromatic diffractive surface to correct chromatic aberration and an 'echelette' feature to extend the range of vision (far, intermediate, near). Toric models include cylinder axis marks for alignment with the steep corneal meridian to reduce residual astigmatism. The device is implanted by a surgeon during cataract surgery. The extended depth of focus provides patients with improved intermediate and near visual acuity compared to monofocal IOLs, reducing spectacle dependence. Patients may experience increased visual symptoms such as halos, starbursts, or glare compared to monofocal lenses, and potential reduction in contrast sensitivity.
Clinical Evidence
Prospective, multicenter, bilateral, randomized, subject/evaluator-masked clinical study (N=299) comparing Symfony IOL to monofocal control. Primary endpoints: monocular UCIVA and DCIVA at 66 cm (superiority). Results: Symfony showed statistically significant improvement (p<0.0001) in UCIVA (1.7 lines) and DCIVA (2.4 lines). Secondary endpoints: DCNVA at 40 cm (2.2 lines improvement, p<0.0001) and defocus range (≥0.5 D increase, p<0.0001). Safety: Serious adverse event rates were low (2.7%) and comparable to ISO SPE rates. Contrast sensitivity was reduced compared to monofocal control. Increased reports of halos, starbursts, and glare noted.
Technological Characteristics
Hydrophobic SENSAR soft acrylic material with polyethylene glycol surface treatment. Biconvex optic (6.0 mm diameter) with wavefront-designed aspheric/toric-aspheric anterior surface and achromatic diffractive posterior surface with echelette feature. TRI-FIX haptic design. UV cutoff at 10% transmittance (374-375nm). Refractive index 1.47 at 35°C. Ethylene oxide sterilization. Standards: ISO 11979-2, -3, -5, -6, -9; ISO 10993; ANSI Z80.30.
Indications for Use
Indicated for adult patients (≥22 years) with cataracts requiring phacoemulsification and posterior IOL implantation. Model ZXR00 for patients with <1.0 D pre-existing corneal astigmatism; Models ZXT150-ZXT375 for patients with ≥1.0 D preoperative corneal astigmatism. Contraindications: None known.
Regulatory Classification
Identification
An intraocular lens is a device made of materials such as glass or plastic intended to be implanted to replace the natural lens of an eye.
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
## I. GENERAL INFORMATION
| Device Generic Name: | Intraocular Lens |
| --- | --- |
| Device Trade Name: | TECNIS® Symfony Extended Range of Vision IOLs - Model ZXR00; Toric Models: ZXT150, ZXT225, ZXT300 and ZXT375 |
| Device Procode: | POE |
| Applicant's Name and Address: | Abbott Medical Optics Inc. 1700 East Saint Andrew Place Santa Ana, CA 92705 |
| Date(s) of Panel Recommendation: | None |
| Premarket Approval Application (PMA) Number: | P980040/S065 |
| Date of FDA Notice of Approval: | July 15, 2016 |
## II. INDICATIONS FOR USE
The TECNIS® Symfony Extended Range of Vision IOL, Model ZXR00, is indicated for primary implantation for the visual correction of aphakia, in adult patients with less than 1 diopter of pre-existing corneal astigmatism, in whom a cataractous lens has been removed. The lens mitigates the effects of presbyopia by providing an extended depth of focus. Compared to an aspheric monofocal IOL, the lens provides improved intermediate and near visual acuity, while maintaining comparable distance visual acuity. The Model ZXR00 IOL is intended for capsular bag placement only.
The TECNIS® Symfony Toric Extended Range of Vision IOLs, Models ZXT150, ZXT225, ZXT300, and ZXT375, are indicated for primary implantation for the visual correction of aphakia and for reduction of residual refractive astigmatism in adult patients with greater than or equal to 1 diopter of preoperative corneal astigmatism, in whom a cataractous lens has been removed. The lenses mitigate the effects of presbyopia by providing an extended depth of focus. Compared to an aspheric monofocal IOL, the lenses provide improved intermediate and near visual acuity, while maintaining comparable distance visual acuity. The Model Series ZXT IOLs are intended for capsular bag placement only.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 1
{1}
### III. CONTRAINDICATIONS
There are no known contraindications.
### IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the labeling for the TECNIS® Symfony Extended Range of Vision IOLs, Models ZXR00, ZXT150, ZXT225, ZXT300 and ZXT375.
### V. DEVICE DESCRIPTION
The TECNIS® Symfony Extended Range of Vision IOLs, non-toric lens model ZXR00 and toric lens models ZXT150, ZXT225, ZXT300, and ZXT375, are ultraviolet light-absorbing posterior chamber IOLs, which are intended to provide a continuous range of vision without compromise in distance vision. In addition, the toric IOLs compensate for corneal astigmatism.
TECNIS® Symfony Extended Range of Vision IOLs are designed to be positioned in the lens capsule to replace the optical function of the natural crystalline lens. **Table 1**, below describes the physical characteristics of the lenses. The biconvex optic incorporates a proprietary wavefront-designed aspheric or toric-aspheric anterior optic, designed to compensate for corneal spherical aberration. The anteriorly located, cylinder axis marks in the toric-aspheric optic denote the meridian with the lowest power and is to be aligned with the steep corneal meridian. The squared posterior edge of the aspheric and toric-aspheric anterior optic is designed to provide a 360-degree barrier and has a frosted design to reduce potential edge glare effects. The posterior optic of the TECNIS® Symfony Extended Range of Vision IOLs has a proprietary achromatic diffractive surface designed to correct chromatic aberration and a echelette feature to extend the range of vision, including far, intermediate, and near, while maintaining the corneal spherical aberration correction. TECNIS® Symfony IOLs are designed to have pupil-independent lens performance in any lighting condition.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 2
{2}
**Summary of Physical Characteristics**
| TECNIS® Symfony Extended Range of Vision IOLs | | | | | |
| --- | --- | --- | --- | --- | --- |
| Model Numbers | ZXR00 | ZXT150 | ZXT225 | ZXT300 | ZXT375 |
| Optic type | Biconvex Aspheric | Biconvex Aspheric Toric | | | |
| Optic/Haptic Material *Measured in Water | Hydrophobic SENSAR soft acrylic material with polyethylene glycol surface treatment UV cutoff at 10% Transmittance: 374nm* (5.00 diopter lens) 375nm* (34.00 diopter lens) | | | | |
| IOL Spherical Equivalent Power (Diopter) | +5.0 D to +34.0 D in +0.50 D increments | | | | |
| IOL Cylinder Power, Labeled (Diopter) | N/A | 1.50 D | 2.25 D | 3.00 D | 3.75 D |
| Corneal Plane, approximate (Diopter) | 0.0 | 1.03 D | 1.54 D | 2.06 D | 2.57 D |
| Index of Refraction | 1.47 at 35°C | | | | |
| Haptic Configuration | TRI-FIX design Modified C, integral with optic | | | | |
| Optic Diameter | 6.0 mm | | | | |
| Overall Length | 13.0 mm | | | | |
| Haptic Angle | No angulation but offset from optic body | | | | |
## **VI. ALTERNATIVE PRACTICES AND PROCEDURES**
There are several other alternatives for the correction of aphakia resulting from surgical cataract removal (i.e., for patients who have had a cataractous lens removed) Non-surgical options include eye glasses or contact lenses. Surgical options come in the form of intraocular lenses, which may be monofocal, multifocal, toric or accommodative, depending on the patient’s needs, expectations and lifestyle. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
## **VII. MARKETING HISTORY**
The TECNIS® Symfony Extended Range of Vision IOLs are currently commercially available in Australia, Canada, European Union, India, New Zealand, Singapore, and many other countries in Latin America, the Middle East-Africa region, and Asia Pacific. The lenses have not been withdrawn or recalled from any country for any reason related to safety or effectiveness.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 3
{3}
### **VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH**
Potential adverse effects (e.g., complications) associated with the use of the device include the following:
- Infection (endophthalmitis)
- Hypopyon
- IOL dislocation
- Cystoid macular edema
- Corneal edema
- Pupillary block
- Iritis
- Retinal detachment/tear
- Raised IOP requiring treatment
- Tilt and decentration requiring repositioning
- Residual refractive error resulting in secondary intervention.
- Increased visual symptoms (compared to a monofocal IOL) related to the optical characteristics the IOL, including
- Reduction in contrast sensitivity. This may cause difficulty when driving, e.g., some decrease in visibility distance for detection/identification of road warnings
- Bothersome stray-light artifacts such as halos, starbursts, or glare.
Secondary surgical interventions include, but are not limited to:
- Lens repositioning (due to decentration, rotation, subluxation, etc.)
- Lens replacement
- Vitreous aspirations or iridectomy for pupillary block
- Wound leak repair
- Retinal detachment repair
- Corneal transplant
- Severe inflammation.
For the specific adverse events that occurred during the TECNIS® Symfony Extended Range of Vision IOL clinical study, please see the *Summary of Primary Clinical Studies* section below.
### **IX. SUMMARY OF NONCLINICAL STUDIES**
Preclinical studies performed on either parent devices or subject devices demonstrate the safety and effectiveness of the TECNIS® Symfony Extended Range of Vision IOLs. The results of these studies are summarized below.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 4
{4}
# A. Laboratory Studies
# Physicochemical Testing
The TECNIS® Symfony Extended Range of Vision IOLs, Models ZXR00, ZXT150, ZXT225, ZXT300, and ZXT375, use the same lens material as the material parent, monofocal analog, and multifocal analog; therefore, physicochemical and biological data for these associated lenses are deemed applicable to the subject devices. All physicochemical reports pertaining to the SENSAR soft acrylic material were previously submitted to FDA in 2007 as part of the 180-Day PMA Supplement for the material parent SENSAR® 1-Piece IOL (P980040/S015). P980040 served as the parent lens for the aforementioned submission to which an SSED is available. The physicochemical characterization of the TECNIS® Symfony IOL material met the requirements of ISO 11979-5, Ophthalmic Implants – Intraocular Lenses – Part 5: Biocompatibility and EN ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process. The physicochemical tests are summarized in Table 2. All acceptance criteria for physicochemical testing were met.
TABLE 2
Physicochemical Test Summary:
TECNIS® Symfony Extended Range of Vision IOLs,
Indicating Relationship to the SENSAR® AR40e IOL
| Physicochemical Tests | Results of Testing |
| --- | --- |
| Exhaustive Extraction | Equivalent to Model AR40e approved under P980040 |
| Leachables | Equivalent to Model AR40e approved under P980040 |
| Insoluble Inorganics | No hazardous components identified |
| Hydrolytic Stability | Stable to 5 years equivalent age |
| Photostability | Stable to 20 years equivalent age |
| Nd:YAG Laser | Equivalent to Model AR40e approved under P980040 |
Note: The SENSAR® AR40e IOL has the OptiEdge design and was approved in the same PMA (P980040) as the SENSAR® AR40 IOL, which has a rounded optic edge design.
# B. Animal Studies
# Biological Testing
The TECNIS® Symfony Extended Range of Vision IOLs, Models ZXR00, ZXT150, ZXT225, ZXT300, and ZXT375, are made of the same SENSAR soft acrylic material and
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 5
{5}
have the same manufacturing contact materials previously qualified with the material parent, the SENSAR® 1-Piece IOL, Model AAB00. With the exception of genotoxicity testing, all other biocompatibility tests conducted on Model AAB00 were previously submitted to FDA in 2007 (P980040/S015). The biocompatibility studies were performed in accordance with the requirements in ISO 10993, Biological Evaluation of Medical Devices, and 11979-5 Ophthalmic Implants – Intraocular Lenses – Part 5: Biocompatibility guidelines, to establish a complete profile of the IOL material. The results are summarized in Table 3. All acceptance criteria for biocompatibility were met.
TABLE 3
Biocompatibility Test Summary:
TECNIS® Symfony Extended Range of Vision IOL
| Biological Tests | Results of Testing |
| --- | --- |
| Cytotoxicity: MEM | Non-cytotoxic |
| Agar Diffusion solid & saline extract | Non-cytotoxic |
| Percent Inhibition of Cell Growth Method (%ICG) | Non-inhibitory to cell growth |
| Guinea Pig Maximization a. Saline Extract b. Sesame Oil Extract | Non-sensitizing Non-sensitizing |
| Non-ocular Implant Study (Six-Week Subcutaneous Implantation in Rabbits) | Passed |
| Six-Month Rabbit Intraocular Study | Passed |
| Genotoxicity Testing (Salmonella typhimurium and Escherichia coli reverse mutation assay) | Non-genotoxic, non-mutagenic |
| Genotoxicity Testing (Chromosomal aberration assay in Chinese hamster ovary [CHO] cells) | Non-clastogenic |
| Genotoxicity Testing (Mouse lymphoma mutagenesis assay ISO 10993-3) | Non-mutagenic under short and long exposure conditions |
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 6
{6}
### C. Additional Studies
#### Dimensional, Optical, and Mechanical Testing
Dimensional, optical, and mechanical tests were conducted on finished, sterilized, TECNIS® Symphony Extended Range of Vision IOLs to verify the conformance to applicable sections of ISO 11979-2, *Ophthalmic Implants-Intraocular Lenses-Part 2: Optical Properties and Test Methods*, and ISO 11979-9, *Ophthalmic Implants-Intraocular Lenses-Part 9: Multifocal Intraocular Lenses*; ISO 11979-3, *Ophthalmic Implants-Intraocular Lenses-Part 3: Mechanical Properties and Test Methods*; and ANSI Z80.30, *American National Standard for Ophthalmics: Toric Intraocular Lenses*. As part of mechanical assessment, folding and insertion testing was also performed to verify recovery of lens properties (e.g., optical, etc.) following simulated insertion. Here, the TECNIS® Symphony Extended Range of Vision IOLs passed all predetermined requirements established in the aforementioned standards where applicable and internal product specifications. **Table 4** summarizes the results of the dimensional, optical and mechanical testing.
**TABLE 4**
**Dimensional, Optical and Mechanical Test Requirements Summary:**
**TECNIS® Symphony Extended Range of Vision IOLs**
| Preclinical Testing Requirement | Acceptance Criteria Based on What Standard? | Result |
| --- | --- | --- |
| **Optical Requirements** | | |
| Dioptric Power (D) | 0 ≤ D ≤ 15: ±0.3D 15 < D ≤ 25: ±0.4D 25 < D ≤ 30: ±0.5 D 30 < D: ±1.0D | Passed |
| Cylinder Power [C] (toric TECNIS® Symfony IOLs only) | 0 < C ≤ 2.5: ±0.3D 2.5 < C ≤ 4.5: ±0.4D 4.5 < C: ±0.5D | Passed |
| Image Quality | Greater or equal to 0.43 or 70% of maximal theoretical MTF value | Passed |
| Axis Orientation Mark(s) (toric TECNIS® Symfony IOLs only) | Combined angular errors of the cylindrical axis mark and any deviation from orthogonality between the meridians of highest and lowest dioptric power within ±5° | Passed |
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 7
{7}
| Preclinical Testing Requirement | Acceptance Criteria Based on What Standard? | Result |
| --- | --- | --- |
| Spectral Transmittance | % T > 90% at 600nm % T=10% at ~380nm | Passed |
| **Mechanical Requirements** | | |
| Overall Diameter | 13.00 ± 0.20mm | Passed |
| Vault Height | ± 0.25mm from nominal | Passed |
| Sagitta | 0.64 ± 0.35 mm | Passed |
| Clear Optic Diameter | >4.50mm | Passed |
| Optic Body Diameter | ± 0.10mm | Passed |
| Axial Displacement in Compression | 0.11 mm ± 0.062mm | Passed |
| Optic Decentration | Mean + 2SD= 0.076mm < 0.60mm (10% clear optic) | Passed |
| Optic Tilt | Mean + 2SD = 2.98° < 5° | Passed |
| Angle of Contact | 42° ± 1.0° | Passed |
| Compression Force and Decay | 0.39 mN ± 0.05 | Passed |
| Dynamic Fatigue Durability | No breakage or damage after 250,000 cycles of haptic compression | Passed |
| Surgical Manipulation | The IOL manufacturer shall provide evidence that the loops of an IOL design are capable of withstanding surgical manipulations without failure. | Passed |
| Surface and Bulk Homogeneity | The IOL shall be essentially free from defects | Passed |
| **Recovery of Properties Following Simulated Surgical Manipulation** | | |
| Dioptric Power | 0 ≤ D ≤ 15: ±0.3D 15 < D ≤ 25: ±0.4D 25 < D ≤ 30: ±0.5 D 30 < D: ±1.0D | Passed |
| Cylinder Power (toric TECNIS® Symfony IOLs only) | 0 < C ≤ 2.5: ±0.3D 2.5 < C ≤ 4.5: ±0.4D 4.5 < C: ±0.5D | Passed |
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 8
{8}
| Preclinical Testing Requirement | Acceptance Criteria Based on What Standard? | Result |
| --- | --- | --- |
| Axis Orientation Mark(s) (toric TECNIS® Symfony IOLs only) | Combined angular errors of the cylindrical axis mark and any deviation from orthogonality between the meridians of highest and lowest dioptic power within ±5° | Passed |
| Image Quality | Greater or equal to 0.43 or 70% of maximal theoretical MTF value | Passed |
| Overall Diameter | 13.00 ± 0.20mm | Passed |
| Sagitta | 0.64 ± 0.35 mm | Passed |
| Surface and Bulk Homogeneity | The IOL shall be essentially free from defects | Passed |
#### Microbiology, Sterilization, and Shelf Life Adoption / Testing
The lens material and platform, geometry, dimensions, manufacturing method, materials and equipment, sterilization method, and packaging materials and configuration of the TECNIS® Symfony Extended Range of Vision IOLs are the same as those of its monofocal analog, the TECNIS® 1-Piece IOL, Model ZCB00 (P980040/S015). Therefore, stability, packaging integrity, and transport stability data supporting this reference lens model was used to support the subject lenses. The TECNIS® Symfony Extended Range of Vision IOLs will be labeled with a 5-year shelf life.
The lens material/product configuration, packaging configuration, and load density used for the TECNIS® Symfony Extended Range of Vision IOLs are the same as those used for the approved and validated reference monofocal lens. Sterilization validations performed for this associated lens model are deemed applicable to the subject lenses. The ethylene oxide sterilization cycle was validated for the monofocal analog and assures a minimum sterility assurance level of 10⁻⁶. The TECNIS® Symfony Extended Range of Vision IOLs were successfully adopted into this validated cycle per the appropriate standard operating procedures and passed all acceptance criteria for bioburden, bacterial endotoxin, and sterilant residual levels.
These tests were conducted in accordance with the following standards and United States Pharmacopoeial chapters:
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 9
{9}
- ANSI/AAMI/ISO 11135-1, Sterilization of Healthcare Products – Ethylene Oxide – Part 1: Requirements for Development, Validation, and Routine Control of a Sterilization Process
- ISO 10993-7, Biological evaluation of medical devices – Part 7: Ethylene oxide sterilization residuals
- 34/29 USP 2011, Bacterial Endotoxin Testing
- ISO 11979-6, Ophthalmic Implants – Intraocular Lenses – Part 6: Shelf-life and transport stability
The results of the sterilization, packaging, shelf life and transport stability studies are summarized in Table 5.
TABLE 5
Microbiology, Sterilization, and Shelf Life Testing / Adoption:
TECNIS® Symfony Extended Range of Vision IOLs
| Test | Results |
| --- | --- |
| Lens material/product configuration | Same |
| Packaging material and configuration | Same |
| Sterilization load configuration and density | Same |
| Shelf life | 5 years |
| Package integrity | Equivalent |
| Transport stability | Equivalent |
| Bioburden | Passed |
| Bacterial Endotoxin | Passed |
| Ethylene Oxide Residuals | Passed |
Toric Calculator Software Validation
A software validation was performed for the TECNIS® Symfony Toric Calculator according to FDA's guidance documents, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, dated May 11, 2005, General Principles of Software Validation: Final Guidance for Industry and FDA Staff, dated January 11, 2002, and Content of Premarket Submissions for Management of Cybersecurity in Medical devices – Guidance for Industry and Food and Drug Administration Staff, dated October 2, 2014. The TECNIS® Symfony Toric Calculator was developed using an appropriate software development program. The hazard analysis addressed all identified hazards applicable to both patient and user. These procedures help assure that the TECNIS® Symfony Toric Calculator will operate in a manner described in the specifications when used according to its instructions for use.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 10
{10}
## **X. SUMMARY OF PRIMARY CLINICAL STUDIES**
Data from a recent clinical study of the TECNIS® Symfony Extended Range of Vision IOL, Model ZXR00 (non-toric), and data from other relevant prior clinical studies are included to support the safety and effectiveness of the TECNIS Symfony IOLs, Model ZXR00, and TECNIS Symfony Toric IOLs, Models ZXT150, ZXT225, ZXT300, and ZXT375.
As stated above in the nonclinical section, the applicant leveraged data/information from various parent lenses. The TECNIS Multifocal IOL (Model ZM900) approved under P980040 served as the optical parent whereby the posterior optic design of the TECNIS® Symfony Extended Range of Vision IOLs (non-toric and toric) was derived from that of the TECNIS Multifocal IOL. The SENSAR 1-Piece IOL, Model AAB00 which was approved under P980040/S015 served as the material and mechanical parent to support the safety and effectiveness of the 1-piece platform and SENSAR acrylic material used for the TECNIS® Symfony Extended Range of Vision IOLs (non-toric and toric models). Although preclinical data/information was leveraged, a clinical study was warranted to assess the change to the optic to support the extended depth of focus claim.
The toric toric models (ZXT150, ZXT225, ZXT300, and ZXT375) involved imposing the toric feature from the toric design parents (P980040/S039: TECNIS Toric 1-Piece IOLs (Models ZCT150, ZCT225, ZCT300 and ZCT400Model ZCT400) onto the anterior optic surface of the TECNIS® Symfony Extended Range of Vision IOL, Model ZXR00 (non-toric). Since the study for Model ZXR00 established safety and the applicant has approved toric parent IOLs, additional clinical data was not warranted as the only difference is in cylinder powers.
### **A. Study Design**
Subjects were treated between August 2014 and June 2015. The database for this Panel-Track PMA Supplement reflected data collected through June 2015 and included 299 implanted subjects. There were 15 investigative sites in the U.S.
The clinical study of the TECNIS® Symfony Extended Range of Vision IOL, Model ZXR00, was a prospective, 6-month, multicenter, bilateral, randomized, subject/evaluator-masked clinical investigation designed to evaluate at least 135 subjects bilaterally implanted with the Symfony IOL, and 135 subjects bilaterally implanted with the monofocal TECNIS® 1-Piece IOL, Model ZCB00. The monofocal control IOL is a legally-marketed alternative with similar indications for use, except that it does not provide an extended depth of focus and is not intended to provide improved vision at intermediate and near distances.
Statistical analyses were frequentist. For the key effectiveness analyses the hypothesis tests were to demonstrate superiority over the control group. In addition, an independent Data Safety Monitoring Board (DSMB), comprised of two ophthalmologists, one statistician, and one facilitator, was used to monitor all safety endpoints.
#### **1. Clinical Inclusion and Exclusion Criteria**
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 11
{11}
Enrollment in the TECNIS® Symfony Extended Range of Vision IOL clinical study was limited to subjects who met the following inclusion criteria:
- Age 22 years or greater
- Bilateral cataracts for which phacoemulsification extraction and posterior IOL implantation were planned for both eyes
- Preoperative best corrected distance visual acuity (BCDVA) of 20/40 Snellen or worse with or without a glare source
- Potential for postoperative BCDVA of 20/30 Snellen or better
- Preoperative corneal astigmatism of 1.00 D or less in both eyes
Subjects were not permitted to enroll in the TECNIS® Symfony Extended Range of Vision IOL clinical study if they met any of the following exclusion criteria:
- Requiring an intraocular lens power outside the available range of +16.0 D to +28.0 D
- Use of systemic or ocular medications that may affect vision or likely to impact pupil dilation or iris structure
- Acute, chronic, or uncontrolled systemic or ocular disease or any illness that, in the opinion of the investigator, would increase the operative risk or confound the outcome(s) of the study (e.g., poorly-controlled diabetes)
- History of ocular trauma, prior refractive or other ocular surgery or subjects expected to require retinal laser treatment or other surgical intervention
- Presence of ocular pathology other than cataract such as:
- Corneal abnormalities (including irregular astigmatism)
- Pupil abnormalities
- Strabismus or any inability to focus, or fixate for prolonged periods of time
- Capsule or zonule abnormalities
- Glaucomatous changes
- Intraocular inflammation
- Known pathology that may affect visual acuity and/or is predicted to cause future acuity losses to a level of 20/30 or worse (e.g., macular degeneration)
- Inability to achieve keratometric stability for contact lens wearers
- Desire for monovision correction
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 12
{12}
## 2. Follow-up Schedule
All subjects were evaluated according to the follow-up schedule below:
**Clinical Study Visit Schedule**
| Visit | Exam | Eyes Evaluated | Visit Window |
| --- | --- | --- | --- |
| 1 | Preoperative Exam | Both Eyes | Within 45 days prior to 1^{st} surgery |
| 2 | Operative | 1^{st} Eye | 0-45 days following preoperative exam |
| 3 | Postop 1 (1 day) | 1^{st} Eye | 1-2 days postoperative |
| 4 | Postop 2 (1 week)* | 1^{st} Eye | 7-14 days postoperative |
| 5 | Operative | 2^{nd} Eye | Within 1 month after 1^{st} eye surgery |
| 6 | Postop 1 (1 day) | 2^{nd} Eye | 1-2 days postoperative |
| 7 | Postop 2 (1 week) | 2^{nd} Eye | 7-14 days postoperative |
| 8 | Postop 3 (1 month) | Both Eyes | 30-60 days postop from 2^{nd} eye surgery |
| 9 | Postop 4 (6 months) | Both Eyes | 120-180 days postop from 2^{nd} eye surgery |
\* Postop 2 for the first eye was to be completed prior to surgery on the second eye.
Preoperative study procedures included informed consent, ocular history, potential visual acuity, uncorrected and best corrected distance visual acuity, slit-lamp and fundus examinations and determination of the subject meeting all inclusion/exclusion criteria. Following determination of the first study eye, subjects were randomized and underwent cataract surgery and IOL implantation; all eyes were targeted for emmetropia. Subject and testing evaluators were masked throughout the study as to which lens type each subject received.
Postoperative study procedures included uncorrected and best corrected distance visual acuity, uncorrected and distance corrected intermediate visual acuity at 66 cm, uncorrected, distance corrected and best corrected (with add) near visual acuity at 40 cm, defocus testing (sub-study), contrast sensitivity, keratometry, intraocular pressure, pupil size, slit-lamp and fundus examinations, assessment of spontaneously-reported optical visual symptoms, subject questionnaires (spectacle wear, directed optical visual symptoms, etc.), and assessment of complications and adverse events. Adverse events and complications were recorded at all visits.
The key time points are shown below in the tables summarizing safety and effectiveness.
## 3. Clinical Endpoints
With regards to safety:
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 13
{13}
- The primary co-endpoints were the rates of serious adverse events vs. the historical control safety and performance endpoints (SPE) rates in ISO 11979-7 (2006).
- Success criterion for each type of event was a rate not statistically greater for the Symfony than the historical control rate.
- A secondary endpoint was monocular (primary eye) best corrected distance contrast sensitivity under mesopic conditions with and without glare at 12 cycles per degree (cpd). (This was related to a potential claim of non-inferiority compared to monofocal control.)
- Success criterion was statistical non-inferiority, based upon a non-inferiority margin of 0.15 log units.
- Other safety endpoints included:
- Mean monocular (primary eye) BCDVA compared to the monofocal control IOL.
- Success criterion was statistical non-inferiority for Symfony compared to the control, with a non-inferiority margin of 1 line (logMAR).
- Visual symptoms via patient-reported outcomes (PRO) instrument.
- Success criterion: no specific criterion
- Other contrast sensitivity outcomes under mesopic, mesopic with glare and photopic with glare conditions
- Success criterion: no specific criterion
With regards to effectiveness:
- The primary effectiveness co-endpoints were monocular (primary eye) uncorrected intermediate visual acuity (UCIVA) and distance corrected intermediate visual acuity (DCIVA) under photopic conditions at 66 cm.
- For each of these, the success criteria were statistical superiority of the mean Symfony acuity compared to the control; plus the proportion of Symfony eyes achieving at least 20/25 had to be at least 50%, and at least 25% higher than the control.
- Secondary endpoints were:
- Mean monocular (primary eye) diopters of depth of focus from optical infinity. (Sub-study of eyes achieving at least BCDVA of 20/25 consistently.)
- Success criteria were statistical superiority over the control, plus a mean difference of ≥ 0.50 D.
- The proportion of subjects who reported wearing glasses or contacts "none of the time" or "a little of the time" for overall spectacle wear via binocular questionnaire response,
- Success criteria were at least 50% of Symfony eyes achieving this outcome, and the proportion had to be at least 25% higher than the control.
- Monocular (primary eye) distance corrected near visual acuity (DCNVA) at 40 cm.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 14
{14}
- Success criteria were statistical superiority of the mean Symfony acuity compared to the control; plus the proportion of Symfony eyes achieving at least 20/40 had to be at least 50%, and at least 25% higher than the control.
- Another endpoint was tolerance to residual refractive error as assessed by mean monocular (primary eye) uncorrected distance visual acuity (UCDVA) for eyes with ≥0.50 D of absolute residual manifest refraction spherical equivalent (MRSE).
o Success criterion was statistical demonstration of superiority of Symfony over the control.
All clinical endpoints were evaluated at 6 months postoperatively. Because the TECNIS® Symfony IOL is a modification of approved IOLs, conclusions regarding device effectiveness and safety are also substantiated from the results of the studies of the parent IOLs.
The primary and secondary study endpoints were analyzed in three population groups: safety (all implanted eyes), intent-to-treat with data imputation for missing values (ITT; included all implanted eyes and all subjects not necessarily implanted), and per-protocol (subjects/eyes without any protocol deviations). Safety endpoints and all “other” endpoints were primarily analyzed using only the safety population. For consistency, results presented in this summary are predominantly from the safety population, unless otherwise indicated.
#### 4. Statistical Methods
The TECNIS® Symfony Extended Range of Vision IOL, Model ZXR00, was compared to the monofocal control IOL (monofocal TECNIS 1-Piece IOL, Model ZCB00) for most safety and effectiveness endpoints. The primary analysis group consists of first eyes implanted or binocular data as appropriate; second-eye and all-eye outcomes are considered supportive. One-sided, two-sample t-tests with an alpha level of 0.025 were used for the primary endpoints of monocular UCIVA and DCIVA and the secondary endpoints of monocular mean diopters of defocus and monocular DCNVA. A logistic regression for imputed data or Fisher’s Exact test (for no data imputation) with a one-sided alpha of 0.025 was used for binocular overall spectacle wear. Nonparametric statistics with a lower limit of a 90% confidence interval were used for the contrast sensitivity endpoint. Hierarchical methods were used to adjust for multiple statistical comparisons for these endpoints. The ITT population was the primary analysis population used for the primary and secondary study endpoints with missing data imputed using multiple imputation methods.
Complications and adverse events, as well as the proportion of first eyes achieving 20/40 or better BCDVA, were compared to ISO SPE rates for the Symfony lens group using an Exact test based on the binomial distribution. Monocular, first-eye, mean BCDVA and contrast sensitivity for the Symfony lens group were compared to the control lens group and analyzed using non-inferiority methods.
The sample size was justified based on the primary study endpoints of monocular UCIVA and DCIVA, and the requirements for contrast sensitivity testing. With at least 135 evaluable
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 15
{15}
subjects in each lens group, this study had greater than 90% power to detect a 0.7 lines or greater difference in mean visual acuity between the Symfony and control lens groups for UCIVA and DCIVA.
### **B. Accountability of PMA Cohort**
At the time of database lock, of 324 patients enrolled (consented) in the PMA study, 299 (of approximately 300 intended per protocol) were implanted with a study lens in at least one operative eye across 15 U.S. investigative sites. Of 299 subjects implanted, 148 were in the Symfony group and 151 were in the monofocal group. Only one monofocal subject was not implanted bilaterally due to illness and subsequent death. A total of 147 Symfony and 148 monofocal control subjects (total N=295) were available for analysis at the completion of the study, the 6-month postoperative visit.
Subject accountability is presented in **Table 7** for Symfony and **Table 8** for monofocal control first eyes (subjects). Subject compliance was excellent at 6 months for an overall percent accountability of 98.7% (295/299). Only four subjects (1.3%; 4/299) were unavailable at 6 months; two were discontinued due to death and two were lost-to-follow-up due to reasons unrelated to vision.
**Symfony First Eyes Accountability (N=148)**
| Subject Status | 1 Day | | 1 Week | | 1 Month | | 6 Months | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | n | % | n | % | n | % | n | % |
| **Available for Analysis** | **148** | **100** | **148** | **100** | **148** | **100** | **147** | **99.3** |
| --In Interval (included in analysis) | 148 | 100 | 145 | 98.0 | 140 | 94.6 | 146 | 98.6 |
| --Out of Interval (included in analysis) | 0 | 0.0 | 3 | 2.0 | 8 | 5.4 | 1 | 0.7 |
| **Missing Subjects** | **0** | **0.0** | **0** | **0.0** | **0** | **0.0** | **1** | **0.7** |
| --Discontinued | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
| --Missed visit | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
| --Not seen but accounted for | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
| --Lost-to-follow-up | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 1^{a} | 0.7 |
| **Active** | **0** | **0.0** | **0** | **0.0** | **0** | **0.0** | **0** | **0.0** |
| --Active (not yet in visit interval) | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
| --In interval or Past interval (form not yet received) | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
$^{a}$ One subject was unable to be contacted.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 16
{16}
**Monofocal First Eyes Accountability (N=151)**
| Subject Status | 1 Day | | 1 Week | | 1 Month | | 6 Months | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | n | % | n | % | n | % | n | % |
| **Available for Analysis** | **151** | **100** | **150** | **99.3** | **150** | **99.3** | **148** | **98.0** |
| --In Interval (included in analysis) | 151 | 100 | 148 | 98.0 | 147 | 97.4 | 148 | 98.0 |
| --Out of Interval (included in analysis) | 0 | 0.0 | 2 | 1.3 | 3 | 2.0 | 0 | 0.0 |
| **Missing Subjects** | **0** | **0.0** | **1** | **0.7** | **1** | **0.7** | **3** | **2.0** |
| --Discontinued | 0 | 0.0 | 0 | 0.0 | 1^{b} | 0.7 | 2^{b,c} | 1.3 |
| --Missed visit | 0 | 0.0 | 1^{b} | 0.7 | 0 | 0.0 | 0 | 0.0 |
| --Not seen but accounted for | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
| --Lost-to-follow-up | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 1^{a} | 0.7 |
| **Active** | **0** | **0.0** | **0** | **0.0** | **0** | **0.0** | **0** | **0.0** |
| --Active (not yet in visit interval) | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
| --In interval or Past interval (form not yet received) | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
$^{a}$ One subject was lost-to-follow-up due to illness (Parkinson’s Disease).
$^{b}$ One subject died due to cancer approximately 1 month following the first-eye implant; the second eye was not implanted.
$^{c}$ One subject died due to stroke following the 1-month visit.
### **C. Study Population Demographics and Baseline Parameters**
The demographics of the study population are typical for a randomized, prospective, multicenter clinical study performed in the U.S.
**Table 9** presents the demographic data for the Symfony group vs. the monofocal control group. Subject demographics were similar between the Symfony and monofocal control groups. The mean age was 68.0 years for both lens groups, females represented more than half of both lens groups and most subjects in both lens groups were White.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 17
{17}
**Demographics**
| | Symfony N=148 | Monofocal N=151 |
| --- | --- | --- |
| Age (years) | N | 148 | 151 |
| Mean | 68.0 | 67.9 |
| Std | 7.5 | 7.9 |
| Median | 68 | 69 |
| Min | 46 | 47 |
| Max | 86 | 88 |
| Not Reported | 0 | 0 |
| Age Group | <60 | 15 (10.1%) | 24 (15.9%) |
| 60–69 | 70 (47.3%) | 61 (40.4%) |
| 70–79 | 52 (35.1%) | 56 (37.1%) |
| ≥80 | 11 (7.4%) | 10 (6.6%) |
| Not Reported | 0 - | 0 - |
| Sex | Male | 57 (38.5%) | 65 (43.0%) |
| Female | 91 (61.5%) | 86 (57.0%) |
| Not Reported | 0 - | 0 - |
| Race | African American | 4 (2.7%) | 16 (10.6%) |
| American Indian/Alaska Native | 0 (0.0%) | 2 (1.3%) |
| Asian | 1 (0.7%) | 3 (2.0%) |
| White | 143 (96.6%) | 130 (86.1%) |
| Not Reported | 0 - | 0 - |
| Ethnicity | Hispanic/Latino | 5 (3.4%) | 4 (2.6%) |
| Not Hispanic/Latino | 143 (96.6%) | 147 (97.4%) |
| Not Reported | 0 - | 0 - |
| Iris Color | Blue/Gray | 47 (31.8%) | 47 (31.1%) |
| Brown/Black | 63 (42.6%) | 63 (41.7%) |
| Green/Hazel | 38 (25.7%) | 41 (27.2%) |
| Not Reported | 0 - | 0 - |
%=n/N(Total) excluding not reported
$^{a}$ P value from 2-sided 2-sample t-test
$^{b}$ P value from 2-sided Fisher's exact test
**Table 10** presents key ocular baseline parameters of target spherical equivalent and preoperative keratometric cylinder.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 18
{18}
**Mean Target Spherical Equivalent and Preop Keratometric Cylinder**
| | IOL | N | Mean | Std. Dev. | Median | Min. | Max. |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Target Spherical Equivalent (D) | Symfony | 148 | -0.203 | 0.148 | -0.200 | -0.76 | 0.21 |
| | Monofocal | 151 | -0.192 | 0.152 | -0.190 | -0.49 | 0.32 |
| | Difference | - | -0.012 | - | - | - | - |
| Keratometric Cylinder (D) | Symfony | 148 | 0.505 | 0.248 | 0.490 | 0.00 | 1.00 |
| | Monofocal | 151 | 0.532 | 0.221 | 0.500 | 0.00 | 1.00 |
| | Difference | - | -0.026 | - | - | - | - |
$^{a}$ P Value from two-sided two-sample t-test.
## **D. Safety and Effectiveness Results**
### **1. Safety Results**
The analysis of safety was based on the safety cohort of 299 implanted subjects: 148 Symfony subjects (148 bilaterally implanted) and 151 monofocal subjects (150 bilaterally implanted).
#### Adverse Events that occurred in the PMA clinical study
Overall, 2.7% (4/148) of Symfony subjects experienced serious adverse events during the study and none (0%; 0/148) experienced device-related or unanticipated events. The incidence rates of cumulative serious adverse events for Symfony eyes are presented in **Table 11**. The incidence rates for the Symfony IOL compared favorably to the specified ISO SPE rates, as the observed rates for Symfony were within or not statistically significantly higher than the specified ISO SPE rates (primary safety endpoint). In addition, there were no secondary surgical interventions related to the optical properties of the Symfony IOL. Secondary surgical intervention events for the Symfony IOL are specified in **Table 12**.
The incidence rates of persistent serious adverse events at 6 months for Symfony eyes also compared favorably to the ISO SPE (safety and performance endpoint) rates (**Table 13**). There were no persistent medical complications/adverse events present at 6 months for Symfony first eyes. There was one case of cystoid macular edema (0.7%; 1/147) for Symfony second eyes at 6 months, however, this rate was not statistically significantly higher than the ISO SPE rate of 0.5% ( $p=0.5238$ ). As there were no other persistent medical complications/adverse events at 6 months, all other persistent event rates for Symfony eyes were below the ISO SPE persistent rates (corneal edema 0.3%; iritis 0.3%; and, raised IOP requiring treatment 0.4%).
The safety of the device for the correction of aphakia was not based on this sample alone, but rather on all available data for the device (and parent IOLs) to date. The safety data
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 19
{19}
from this study were for confirmatory purposes. The adverse events and complications observed in this study were typical for similar types of IOLs.
**Serious Adverse Events for Symfony Eyes**
| Cumulative Medical Complications/ Adverse Events | ISO SPE Rate | First Eyes N=148 | | Second Eyes N=148 | | All Eyes N=296 | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | n | % | n | % | n | % |
| Cystoid macular edema | 3 | 1 | 0.7 | 1 | 0.7 | 2 | 0.7 |
| Hypopyon | 0.3 | 0 | 0 | 1^{b} | 0.7^{b} | 1^{f} | 0.3^{f} |
| Endophthalmitis | 0.1 | 0 | 0 | 1^{c} | 0.7^{c} | 1^{g} | 0.3^{g} |
| Lens dislocated from posterior chamber | 0.1 | 0 | 0 | 0 | 0 | 0 | 0 |
| Pupillary block | 0.1 | 0 | 0 | 0 | 0 | 0 | 0 |
| Retinal detachment | 0.3 | 0 | 0 | 0 | 0 | 0 | 0 |
| Eyes with secondary surgical intervention | 0.8 | 0 | 0 | 2^{d} | 1.4^{e} | 2 | 0.7 |
| -- Device related | NA | 0 | 0 | 0 | 0 | 0 | 0 |
| -- Not device related | NA | 0 | 0 | 2^{d} | 1.4 | 2 | 0.7 |
| Other: | | | | | | | |
| - Pupillary capture | NA | 0 | 0.0 | 1 | 0.7 | 1 | 0.3 |
$^{a}$ Per ISO 11979-7: 2006/Amd. 1:2012 (E) ophthalmic Implants – Intraocular Lenses (Part 7): The SPE rate is the safety and performance endpoint.
$^{b}$ Incidence rate for “Hypopyon-Second Eye” is not statistically significantly higher than ISO SPE rate ( $p = 0.3590$ ) using 1-sided exact test.
$^{c}$ Incidence rate for “Endophthalmitis-Second Eye” is not statistically significantly higher than ISO SPE rate ( $p = 0.1376$ ) using 1-sided exact test.
$^{d}$ Treatment injections for endophthalmitis (Subject 1314) and CME (Subject 1425).
$^{e}$ Incidence rate for secondary surgical interventions is not statistically significantly higher than ISO SPE rate ( $p = 0.3318$ ) using 1-sided exact test.
$^{f}$ Incidence rate for “Hypopyon--All Eyes” (0.34) is not statistically significantly higher than ISO SPE rate ( $p = 0.5891$ ) using 1-sided exact test.
$^{g}$ Incidence rate for “Endophthalmitis-All Eyes” is not statistically significantly higher than ISO SPE rate ( $p = 0.2563$ ) using 1-sided exact test.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 20
{20}
**Secondary Surgical Interventions for Symfony Eyes**
| Secondary Surgical Interventions^{a}: Not Device-Related | First Eyes N=148 | | Second Eyes N=148 | | All Eyes N=296 | |
| --- | --- | --- | --- | --- | --- | --- |
| | n | % | n | % | n | % |
| Treatment injections for medical complications: | 0 | 0 | 2 | 1.4 | 2 | 0.7 |
| - Cystoid macular edema | 0 | 0 | 1 | 0.7 | 1 | 0.3 |
| - Endophthalmitis (with AC tap) | 0 | 0 | 1 | 0.7 | 1 | 0.3 |
%=n/N(Total)
$^{a}$ All SSIs were treatments for SAEs; there were no SSIs as the original event.
Rates of serious adverse events were similar to those seen in other IOL studies. There is no reason to believe that the adverse events rates for the Symfony lens differ substantially from those of the relevant parent IOLs from which the Symfony is an optical modification.
**6-Month Persistent Serious Adverse Events for the Symfony IOL Group**
| Persistent Medical Complications/ Adverse Events | ISO SPE Rate^{a} | First Eyes N=148 | | Second Eyes N=148 | | All Eyes N=296 | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | n | % | n | % | n | % |
| Corneal edema | 0.3 | 0 | 0 | 0 | 0 | 0 | 0 |
| Cystoid macular edema | 0.5 | 0 | 0.0 | 1 | 0.7^{b} | 1 | 0.3 |
| Iritis | 0.3 | 0 | 0 | 0 | 0 | 0 | 0 |
| Raised IOP requiring treatment | 0.4 | 0 | 0 | 0 | 0 | 0 | 0 |
| Other: | | | | | | | |
| - Pupillary capture | NA | 0 | 0.0 | 1 | 0.7 | 1 | 0.3 |
$^{a}$ Per ISO 11979-7: 2006/Amd. 1:2012 (E) ophthalmic Implants – Intraocular Lenses (Part 7): The SPE rate is the safety and performance endpoint.
$^{b}$ Incidence rate for “Cystoid Macular Edema-Second Eye” is not statistically significantly higher than ISO SPE rate ( $p = 0.5238$ ) using 1-sided exact test.
### Contrast Sensitivity
Monocular best corrected distance contrast sensitivity testing was performed using the Vector Vision ETDRS light box and contrast sensitivity charts under three lighting conditions: mesopic with glare, mesopic without glare, and photopic with glare. Hypothesis tests were conducted using the Hodges-Lehmann method, using a pre-assigned score for subjects who could not see the reference pattern. This may introduce potential bias, which would tend to cause underestimation of the difference in contrast sensitivity between the arms. An alternative analysis method that avoids this bias
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 21
{21}
is a simple comparison of the medians of the two arms. Median contrast scores for the Symfony IOL group were reduced compared to the monofocal control group under each lighting condition and spatial frequency (Table 14), particularly at the higher spatial frequencies. In addition, the secondary study endpoint hypothesis test for median contrast sensitivity at the 12 cpd spatial frequency under mesopic conditions with and without glare failed to demonstrate non-inferiority, as the lower 90% confidence intervals (CI) of the median differences were below the criterion of -0.15 log units (-0.165 log units under mesopic without glare conditions and -0.265 log units under mesopic with glare conditions; ITT population). Differences between Symfony and control medians at 12 cpd were -0.170 log units under mesopic without glare conditions and -0.320 log units under mesopic with glare conditions. No statistically significant difference in contrast sensitivity across pupil size groups was observed; however, the sample size may not have been sufficient to detect differences for subgroup analyses.
TABLE 14
Monocular Best Corrected Distance Contrast Sensitivity at 6 Months
| Spatial Frequency | Lens Model | N | Mesopic Without Glare | | | Mesopic With Glare | | | Photopic With Glare | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Median^{a} | Subjects who did not see the reference pattern | | Median^{a} | Subjects who did not see the reference pattern | | Median^{a} | Subjects who did not see the reference pattern | |
| | | | | n | % | | n | % | | n | % |
| **1.5 cpd** | Symfony | 146 | 1.520 | 0 | 0.0 | 1.520 | 0 | 0.0 | Not tested | | |
| | Control | 147 | 1.595 | 1 | 0.7 | 1.520 | 1 | 0.7 | Not tested | | |
| | Difference | | -0.075 | | | 0.00 | | | Not tested | | |
| **3.0 cpd** | Symfony | 146 | 1.415 | 0 | 0.0 | 1.445 | 1 | 0.7 | 1.560 | 0 | 0.0 |
| | Control | 147 | 1.490 | 3 | 2.0 | 1.490 | 3 | 2.0 | 1.705 | 1 | 0.7 |
| | Difference | | -0.075 | | | -0.045 | | | -0.145 | | |
| **6.0 cpd** | Symfony | 146 | 1.380 | 16 | 11.0 | 1.380 | 19 | 13.0 | 1.700 | 4 | 2.7 |
| | Control | 147 | 1.540 | 6 | 4.1 | 1.550 | 7 | 4.8 | 1.840 | 5 | 3.4 |
| | Difference | | -0.160 | | | -0.170 | | | -0.140 | | |
| **12.0 cpd** | Symfony | 146 | 0.910 | 38 | 26.0 | 0.760 | 44 | 30.1 | 1.325 | 12 | 8.2 |
| | Control | 147 | 1.080 | 23 | 15.6 | 1.080 | 28 | 19.0 | 1.540 | 9 | 6.1 |
| | Difference | | -0.170 | | | -0.320 | | | -0.215 | | |
| **18.0 cpd** | Symfony | 146 | Not tested | | | Not tested | | | 0.885 | 14 | 9.6 |
| | Control | 147 | Not tested | | | Not tested | | | 1.100 | 8 | 5.4 |
| | Difference | | Not tested | | | Not tested | | | -0.215 | | |
cpd = Cycles per degree
$^{a}$ In log units.
Distance High-Contrast Photopic Visual Acuity
Distance visual acuities were tested using 100% Early Treatment Diabetic Retinopathy Study (ETDRS) charts at 4.0 meters under photopic conditions (85 cd/m²).
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 22
{22}
**Table 15** presents mean monocular distance visual acuities at 6 months for Symfony and monofocal control first eyes. Mean UCDVA and BCDVA outcomes were comparable between IOL groups at 20/25 and 20/20, respectively. Comparison of mean monocular BCDVA to the control lens was one of the key safety endpoints. Non-inferiority testing showed the lower limit of the 90% confidence interval (CI) of the mean difference in BCDVA between IOL groups to be less than 1.0 line (LogMAR -0.036; -0.4 lines) between groups, indicating that the Symfony IOL is non-inferior to the control lens in providing best corrected distance visual acuity. It was hypothesized that Symfony-implanted subjects would have greater “tolerance to refractive error.” This was evaluated by trying to demonstrate that for eyes with residual manifest spherical equivalent ≥0.50 D at 6 months, the Symfony arm had statistically superior UCDVA compared to the control. Results did not confirm that Symfony eyes had greater “tolerance to refractive error.”
**Monocular and Binocular Distance Visual Acuity at 6 Months**
| Distance Visual Acuity | Lens Group | Monocular | | | | Binocular | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | N | Mean | Snellen Line | Std. Dev. | N | Mean | Snellen Line | Std. Dev. |
| | | | LogMAR | Equivalent | LogMAR | | LogMAR | Equivalent | LogMAR |
| **Uncorrected** | Symfony | 147 | 0.114 | 20/25 | 0.142 | 147 | 0.034 | 20/20 | 0.106 |
| | Control | 148 | 0.088 | 20/25 | 0.149 | 148 | 0.013 | 20/20 | 0.118 |
| | **Difference** | | | **-0.3 lines** | | | | **-0.2 lines** | |
| **Best Corrected** | Symfony | 147 | -0.021 | 20/20 | 0.082 | 147 | -0.045 | 20/20 | 0.077 |
| | Control | 148 | -0.040 | 20/20 | 0.093 | 148 | -0.075 | 20/16 | 0.081 |
| | **Difference** | | **-0.020^{a}** | **-0.2 lines** | | | | **-0.3 lines** | |
$^{a}$ 90% Confidence Interval around mean difference: [-0.036; -0.003]
**Table 16** presents the distribution of monocular distance visual acuity results for Symfony and monofocal control first eyes at 6 months. As all eyes were best -case, the proportion of first eyes achieving monocular best corrected distance visual acuity (BCDVA) of 20/40 or better for the Symfony lens group (100.0%) was above the ISO Safety and Performance Endpoint (SPE) rates for overall (92.5%) and best-case (96.7%) BCDVA 20/40 or better. The distribution of binocular distance visual acuity results for Symfony and monofocal control subjects at 6 months are presented in **Table 17**.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 23
{23}
**Monocular Distance Visual Acuity at 6 Months**
| Visual Acuity LogMAR (Snellen) | Symfony N=147 | | | | Monofocal Control N=148 | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Uncorrected | | Best Corrected | | Uncorrected | | Best Corrected | |
| | n | % | n | % | n | % | n | % |
| 0.0 (20/20) or better | 57 | 38.8 | 123 | 83.7 | 70 | 47.3 | 131 | 88.5 |
| 0.1 (20/25) or better | 96 | 65.3 | 144 | 98.0 | 106 | 71.6 | 143 | 96.6 |
| 0.2 (20/32) or better | 129 | 87.8 | 147 | 100.0 | 126 | 85.1 | 146 | 98.6 |
| **0.3 (20/40) or better** | **142** | **96.6** | **147** | **100.0** | **139** | **93.9** | **148** | **100.0** |
| 0.4-0.6 (20/50-20/80) | 4 | 2.7 | 0 | 0.0 | 9 | 6.1 | 0 | 0.0 |
| 0.7 (20/100) or worse | 1 | 0.7 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
**Binocular Distance Visual Acuity at 6 Months**
| Visual Acuity LogMAR (Snellen) | Symfony N=147 | | | | Monofocal Control N=148 | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Uncorrected | | Best Corrected | | Uncorrected | | Best Corrected | |
| | n | % | n | % | n | % | n | % |
| 0.0 (20/20) or better | 92 | 62.6 | 137 | 93.2 | 106 | 71.6 | 141 | 95.3 |
| 0.1 (20/25) or better | 134 | 91.2 | 145 | 98.6 | 125 | 84.5 | 146 | 98.6 |
| 0.2 (20/32) or better | 143 | 97.3 | 147 | 100.0 | 142 | 95.9 | 148 | 100.0 |
| **0.3 (20/40) or better** | **147** | **99.3** | **147** | **100.0** | **148** | **100.0** | **148** | **100.0** |
| 0.4-0.6 (20/50-20/80) | 1 | 0.7 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
| 0.7 (20/100) or worse | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
Note that low contrast acuties were not evaluated in this study.
#### Optical/Visual Symptoms
The most commonly reported directed symptoms at 6 months based on a questionnaire (**Table 18**) were halos, starbursts, and glare for both IOL groups; halos and starbursts were reported with increased bother in the Symfony group compared with the monofocal group.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 24
{24}
**Experience/Bother with Visual Symptoms over the Past 7 Days at 6 Months**
| | | Symfony N=147 | | Monofocal N=148 | |
| --- | --- | --- | --- | --- | --- |
| | | n | % | n | % |
| **Halos** | None | 60 | 40.8 | 105 | 70.9 |
| | Bother a little bit | 46 | 31.3 | 24 | 16.2 |
| | Bother somewhat | 18 | 12.2 | 13 | 8.8 |
| | Bother quite a bit | 13 | 8.8 | 2 | 1.4 |
| | Very bothered | 10 | 6.8 | 4 | 2.7 |
| **Starbursts** | None | 62 | 42.2 | 110 | 74.3 |
| | Bother a little bit | 42 | 28.6 | 18 | 12.2 |
| | Bother somewhat | 18 | 12.2 | 12 | 8.1 |
| | Bother quite a bit | 13 | 8.8 | 2 | 1.4 |
| | Very bothered | 12 | 8.2 | 6 | 4.1 |
| **Glare^{b}** | None | 62 | 42.8 | 85 | 57.4 |
| | Bother a little bit | 53 | 36.6 | 35 | 23.6 |
| | Bother somewhat | 12 | 8.3 | 18 | 12.2 |
| | Bother quite a bit | 10 | 6.9 | 5 | 3.4 |
| | Very bothered | 8 | 5.5 | 5 | 3.4 |
| **Streaks of Light^{b}** | None | 122 | 84.7 | 126 | 85.1 |
| | Bother a little bit | 11 | 7.6 | 10 | 6.8 |
| | Bother somewhat | 5 | 3.5 | 7 | 4.7 |
| | Bother quite a bit | 2 | 1.4 | 1 | 0.7 |
| | Very bothered | 4 | 2.8 | 4 | 2.7 |
| **Occlusions (Shadows)^{b}** | None | 139 | 95.2 | 140 | 94.6 |
| | Bother a little bit | 4 | 2.7 | 4 | 2.7 |
| | Bother somewhat | 1 | 0.7 | 2 | 1.4 |
| | Bother quite a bit | 0 | 0.0 | 0 | 0.0 |
| | Very bothered | 2 | 1.4 | 2 | 1.4 |
| **Sensitivity to Light^{b}** | None | 65 | 44.5 | 75 | 50.7 |
| | Bother a little bit | 53 | 36.3 | 38 | 25.7 |
| | Bother somewhat | 15 | 10.3 | 18 | 12.2 |
| | Bother quite a bit | 9 | 6.2 | 6 | 4.1 |
| | Very bothered | 4 | 2.7 | 11 | 7.4 |
| **Poor Low Light Vision^{b}** | None | 60 | 41.1 | 60 | 40.5 |
| | Bother a little bit | 64 | 43.8 | 56 | 37.8 |
| | Bother somewhat | 13 | 8.9 | 21 | 14.2 |
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 25
{25}
| | Symfony N=147 | | Monofocal N=148 | |
| --- | --- | --- | --- | --- |
| | n | % | n | % |
| Bother quite a bit | 7 | 4.8 | 8 | 5.4 |
| Very bothered | 2 | 1.4 | 3 | 2.0 |
%=n/N (total) excluding not reported
$^{a}$ None includes “did not experience symptom” and “experienced symptom but not bothered”.
$^{b}$ “Not Reported” - Two Symfony subjects did not respond to the glare question, three Symfony subjects did not respond to the streaks of light question, and one Symfony subject did not respond to the occlusion question, to the sensitivity to light question and to the poor low light vision question.
Optical/visual symptoms spontaneously reported by subjects to investigators (non-directed reports) are noted in **Table 19**. Note that symptoms rates based upon reports directly to investigators are generally not considered as reliable as those collected using questionnaires. The rates reported to physicians are typically lower than the rates reported by patients on a questionnaire specifically asking about their experience (directed reports; **Table 18**).
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 26
{26}
**Spontaneous (Non-Directed$^{a}$) Reports of Ocular Symptoms at 6 Months**
| Ocular Symptoms | Symfony N=147 | | Control N=148 | |
| --- | --- | --- | --- | --- |
| | n | % | n | % |
| ***Image Quality*** | | | | |
| Blurred vision | 25 | 17.0 | 35 | 23.6 |
| Overall | 6 | 4.1 | 8 | 5.4 |
| Distance | 9 | 6.1 | 3 | 2.0 |
| Intermediate | 1 | 0.7 | 2 | 1.4 |
| Near | 13 | 8.8 | 26 | 17.6 |
| ***Optical/Visual*** | | | | |
| Halos | 24 | 16.3 | 2 | 1.4 |
| Mild | 9 | 6.1 | 1 | 0.7 |
| Moderate | 11 | 7.5 | 0 | 0.0 |
| Severe | 4 | 2.7 | 1 | 0.7 |
| Night Glare | 4 | 2.7 | 0 | 0.0 |
| Mild | 1 | 0.7 | 0 | 0.0 |
| Moderate | 3 | 2.0 | 0 | 0.0 |
| Severe | 0 | 0.0 | 0 | 0.0 |
| Starbursts | 13 | 8.8 | 2 | 1.4 |
| Mild | 6 | 4.1 | 1 | 0.7 |
| Moderate | 5 | 3.4 | 1 | 0.7 |
| Severe | 2 | 1.4 | 0 | 0.0 |
| Night vision difficulty (overall) | 4 | 2.7 | 0 | 0.0 |
| ***Sensation*** | | | | |
| Dryness | 12 | 8.2 | 16 | 10.8 |
%=n/N (Total)
Note: Includes reports of optical/visual symptoms common to traditional multifocal IOLs (halos, night glare, starbursts, and night vision difficulties) as well as any findings reported with an incidence of 10% or more at 6 months.
$^{a}$ Non-directed, spontaneously-reported subject responses were obtained from the open-ended question “Are you having any difficulties with your eyes or vision?”.
The rates of subjects expressing some desire to have lenses removed or replaced due to visual symptoms or other problems with vision are shown in **Table 20**.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 27
{27}
**Would Want to Have Lens(es) Removed and Replaced due to Visual Symptoms**
| | | Symfony N=147 | | Monofocal N=148 | |
| --- | --- | --- | --- | --- | --- |
| | | n | % | n | % |
| Lens removed and replaced | Yes | 5^{a} | 3.4 | 13^{a} | 8.8 |
| | No | 119 | 81.0 | 108 | 73.0 |
| | NA^{b} | 23 | 15.6 | 27 | 18.2 |
%=n/N(Total) excluding not reported.
$^{a}$ One Symfony subject (0.7%; 1/147) and one monofocal subject (0.7%; 1/148) indicated a desire to have the lenses removed/replaced and the investigator determined the subject reason(s) to be related to optical lens design, i.e., a potential secondary surgical intervention.
$^{b}$ NA = NOT APPLICABLE, did not experience any visual symptoms
Based upon the safety analyses provided for this study, and the safety results of the parent IOLs, of which the Symfony is only an optical modification, FDA review finds that there is reasonable assurance of device safety.
## 2. Effectiveness Results
The analysis of effectiveness was based on 147 evaluable subjects for Symfony and 148 evaluable subjects for the control group at the 6-month time point. Key effectiveness outcomes are presented in **Tables 21 to 29** and **Figures 2 to 5**.
### Intermediate High-Contrast Photopic Visual Acuities
Intermediate visual acuities were tested using 100% ETDRS near charts at a fixed test distance of 66 cm, with and without distance correction, under photopic (85 cd/m$^{2}$) lighting conditions.
The primary effectiveness endpoints of improved monocular, uncorrected and distance corrected intermediate visual acuity (UCIVA and DCIVA) at 66 cm were achieved for the Symfony IOL. As shown in **Table 21**, there were statistically significant improvements (**p<0.0001**; ITT population) in mean UCIVA and DCIVA at 6 months in favor of the Symfony lens with improvements of 1.7 and 2.4 lines, respectively. Additionally, as shown in **Table 22**, there were clinically significant improvements in favor of the Symfony IOL with 76.9% (113/147) and 70.1% (103/147) of Symfony eyes achieving UCIVA and DCIVA of 20/25 or better, respectively, compared to 33.8% (50/148) and 13.5% (20/148) of monofocal eyes. These intermediate visual acuity results demonstrate the effectiveness of the Symfony IOL to provide improved intermediate vision compared to the monofocal control lens. Binocular distribution results are presented in **Table 23**.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 28
{28}
**Mean Monocular and Binocular Uncorrected and Distance Corrected Intermediate Visual Acuity at 66 cm at 6 Months**
| Visual Acuity^{a} | Lens Group | Monocular | | | | Binocular | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | N | Mean LogMAR | Snellen Line Equiv. | Line Change vs. Control | N | Mean LogMAR | Snellen Line Equiv. | Line Change vs. Control |
| Uncorrected | Symfony | 147 | 0.087^{a} | 20/25 | 1.7 lines | 147 | 0.002 | 20/20 | 1.3 lines |
| | Control | 148 | 0.256^{a} | 20/40 | | 148 | 0.134 | 20/25 | |
| Distance Corrected | Symfony | 147 | 0.104^{a} | 20/25 | 2.4 lines | 147 | 0.032 | 20/20 | 1.9 lines |
| | Control | 148 | 0.342^{a} | 20/40 | | 148 | 0.227 | 20/32 | |
$^{a}$ The primary study endpoints are uncorrected and distance corrected intermediate VA for first eyes. Symfony had significantly better mean UCIVA and DCIVA compared to Control with **p<0.0001** (from one-sided two-sample t-test).
**Monocular Uncorrected and Distance Corrected Intermediate Visual Acuity at 66 cm at 6 Months**
| Monocular Intermediate Visual Acuity LogMAR (Snellen) | Symfony | | | | Monofocal Control | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Uncorrected N=147 | | Distance Corrected N=147 | | Uncorrected N=148 | | Distance Corrected N=148 | |
| | N | % | n | % | n | % | n | % |
| 0.0 (20/20) or better | 60 | 40.8 | 51 | 34.7 | 19 | 12.8 | 7 | 4.7 |
| **0.1 (20/25) or better** | **113** | **76.9** | **103** | **70.1** | **50** | **33.8** | **20** | **13.5** |
| 0.2 (20/32) or better | 136 | 92.5 | 133 | 90.5 | 81 | 54.7 | 47 | 31.8 |
| 0.3 (20/40) or better | 145 | 98.6 | 143 | 97.3 | 103 | 69.6 | 79 | 53.4 |
| 0.4-0.6 (20/50-20/80) | 2 | 1.4 | 4 | 2.7 | 45 | 30.4 | 63 | 42.6 |
| 0.7 (20/100) or worse | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 6 | 4.1 |
**Binocular Uncorrected and Distance Corrected Intermediate Visual Acuity at 66 cm at 6 Months**
| Binocular Visual Acuity | Symfony | | Control | |
| --- | --- | --- | --- | --- |
| | Uncorrected | Distance Corrected | Uncorrected | Distance Corrected |
| 20/20 or better | 74.8% | 61.9% | 31.1% | 8.1% |
| **20/25 or better** | **96.6%** | **92.5%** | **60.1%** | **35.1%** |
| 20/32 or better | 100.0% | 100.0% | 83.1% | 62.8% |
| 20/40 or better | 100.0% | 100.0% | 91.9% | 79.7% |
| 20/50-20/80 | 0.0% | 0.0% | 8.1% | 20.3% |
| 20/100 or worse | 0.0% | 0.0% | 0.0% | 0.0% |
| **Total** | **147** | **147** | **148** | **148** |
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 29
{29}
Binocular UCIVA and DCIVA results are presented in **Tables 24 and 25**. Mean binocular results improved by 1-2 lines over monocular testing, although the differences between groups remained comparable (**Table 26**). The proportions of subjects achieving 20/25 or better binocularly were higher for the Symfony IOL vs. control IOL for binocular UCIVA and DCIVA (**Table 24**).
All statements apply only to high-contrast photopic visual acuities, as low-contrast intermediate visual acuities were not assessed in this study.
**Mean Binocular Uncorrected and Distance Corrected Intermediate Visual Acuity**
| Binocular Intermediate Visual Acuity | IOL | N | Mean LogMAR | Snellen Line | Std. Dev. |
| --- | --- | --- | --- | --- | --- |
| | | | | Equivalent | |
| **Uncorrected** | Symfony | 147 | 0.002 | 20/20 | 0.085 |
| | Monofocal | 148 | 0.134 | 20/25 | 0.142 |
| | **Difference** | | **0.132** | **1.3 lines** | - |
| **Distance Corrected** | Symfony | 147 | 0.032 | 20/20 | 0.086 |
| | Monofocal | 148 | 0.227 | 20/32 | 0.140 |
| | **Difference** | | **0.195** | **1.9 lines** | - |
#### Near High-Contrast Photopic Visual Acuities
Near visual acuities were tested using 100% ETDRS near charts at a fixed test distance of 40 cm, with and without distance correction, under photopic (85 cd/m$^{2}$) lighting conditions.
The secondary effectiveness endpoint of improved monocular, distance corrected near visual acuity (DCNVA) at 40 cm was achieved for the Symfony IOL. As shown in **Table 25**, there was a statistically significant improvement (**p<0.0001**; ITT population) in mean monocular DCNVA at 6 months in favor of the Symfony lens with an improvement of 2.2 lines. Additionally, as shown in **Table 27**, there were clinically significant improvements in favor of the Symfony IOL with 61.9% (91/147) of Symfony eyes achieving DCNVA of 20/40 or better compared to 16.2% (24/148) of monofocal eyes. Monocular uncorrected distance visual acuity results (UCNVA) were comparable to DCNVA results. These near visual acuity results demonstrate the effectiveness of the Symfony IOL to provide substantial near vision improvement compared to the monofocal control lens. **Table 26**, shows the mean binocular uncorrected and distance corrected near visual acuity at 40 cm at 6 months
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 30
{30}
**Mean Monocular Uncorrected and Distance Corrected Near Visual Acuity**
| Monocular Near Visual Acuity | IOL | N | Mean LogMAR | Snellen Line Equivalent | Std. Dev. |
| --- | --- | --- | --- | --- | --- |
| **Uncorrected** | Symfony | 147 | 0.241 | 20/32 | 0.142 |
| | Monofocal | 148 | 0.459 | 20/63 | 0.183 |
| | **Difference** | | **0.218** | **2.2 lines** | - |
| **Distance Corrected** | Symfony | 147 | 0.323^{a} | 20/40 | 0.146 |
| | Monofocal | 148 | 0.544^{a} | 20/63 | 0.175 |
| | **Difference** | | **0.221^{a}** | **2.2 lines** | - |
$^{a}$ Statistically significant improvement, **p<0.0001** (1-sided 2-sample t-test) 90% Confidence Interval around mean difference [0.184; 0.258].
**Mean Binocular Uncorrected and Distance Corrected**
| Visual Acuity | Lens Group | Binocular | | | |
| --- | --- | --- | --- | --- | --- |
| | | N | Mean LogMAR | Snellen Line Equiv. | Line Change vs. Control |
| Uncorrected | Symfony | 147 | 0.146 | 20/25 | 1.8 lines |
| | Control | 148 | 0.328 | 20/40 | |
| Distance Corrected | Symfony | 147 | 0.229 | 20/32 | 2.0 lines |
| | Control | 148 | 0.426 | 20/50 | |
**Monocular Uncorrected and Distance Corrected Near Visual Acuity**
| Monocular Near Visual Acuity LogMAR (Snellen) | Symfony | | | | Monofocal Control | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Uncorrected | | Distance Corrected | | Uncorrected | | Distance Corrected | |
| | N | % | n | % | n | % | n | % |
| 0.0 (20/20) or better | 14 | 9.5 | 5 | 3.4 | 0 | 0.0 | 0 | 0.0 |
| 0.1 (20/25) or better | 42 | 28.6 | 16 | 10.9 | 1 | 2.7 | 1 | 0.7 |
| 0.2 (20/32) or better | 82 | 55.8 | 49 | 33.3 | 26 | 17.6 | 5 | 3.4 |
| **0.3 (20/40) or better** | **119** | **81.0** | **91** | **61.9** | **46** | **31.1** | **24** | **16.2** |
| 0.4-0.6 (20/50-20/80) | 28 | 19.0 | 54 | 36.7 | 80 | 54.1 | 83 | 56.1 |
| 0.7 (20/100) or worse | 0 | 0 | 2 | 1.4 | 22 | 14.9 | 41 | 27.7 |
Binocular UCNVA and DCNVA results are presented in **Tables 28 and 29**. Mean binocular results improved by 1 - 2 lines over monocular testing, although the differences
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 31
{31}
between groups remained comparable (Table 28). The proportions of subjects achieving 20/40 or better binocularly were higher for the Symfony IOL vs. control IOL for binocular UCNVA and DCNVA, respectively (Table 29).
All statements apply only to high-contrast photopic visual acuities, as low-contrast near visual acuities were not assessed in this study.
**Table 28**
**Mean Binocular Uncorrected and Distance Corrected Near Visual Acuity**
**at 40 cm at 6 Months**
| Binocular Near Visual Acuity | IOL | N | Mean LogMAR | Snellen Line Equivalent | Std. Dev. |
| --- | --- | --- | --- | --- | --- |
| **Uncorrected** | Symfony | 147 | 0.146 | 20/25 | 0.112 |
| | Monofocal | 148 | 0.328 | 20/40 | 0.167 |
| | **Difference** | | **0.181** | **1.8** | - |
| **Distance Corrected** | Symfony | 147 | 0.229 | 20/32 | 0.114 |
| | Monofocal | 148 | 0.426 | 20/50 | 0.159 |
| | **Difference** | | **0.197** | **2.0 lines** | - |
**TABLE 29**
**Binocular Uncorrected and Distance Corrected Near Visual Acuity**
**at 40 cm at 6 Months**
| Binocular Near Visual Acuity LogMAR (Snellen) | Symfony | | | | Monofocal Control | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Uncorrected | | Distance Corrected | | Uncorrected | | Distance Corrected | | |
| | N=147 | N | % | N=147 | N=148 | N | % | N=148 | N |
| | N | | | n | | n | | n | |
| 0.0 (20/20) or better | 32 | 21.8 | 12 | 8.2 | 7 | 4.7 | 2 | 1.4 | |
| 0.1 (20/25) or better | 81 | 55.1 | 35 | 12.8 | 19 | 12.8 | 7 | 4.7 | |
| 0.2 (20/32) or better | 124 | 84.4 | 77 | 52.4 | 50 | 33.8 | 19 | 12.8 | |
| **0.3 (20/40) or better** | **141** | **95.9** | **133** | **90.5** | **93** | **62.8** | **51** | **34.5** | |
| 0.4-0.6 (20/50-20/80) | 6 | 4.1 | 14 | 9.5 | 48 | 32.4 | 87 | 58.8 | |
| 0.7 (20/100) or worse | 0 | 0.0 | 0 | 0.0 | 7 | 4.8 | 10 | 6.8 | |
### Defocus Testing
Defocus curve testing was performed on a subset of subjects from each lens group at 8 sites at the 6-month study exam to evaluate monocular and binocular best corrected distance visual acuity defocus curves and any effects of pupil size. Defocus testing was performed using the electronic Freiburg Visual Acuity and Contrast Test (FrACT). Mean monocular defocus range for which acuity was 20/32 or better was a secondary study endpoint; the primary analysis group included eyes that achieved BCDVA of 20/25 or better using both ETDRS and FrACT. Monocular results were also analyzed for three pupil size ranges: ≤2.5 mm; >2.5 mm and <4.0 mm; and ≥4.0 mm. The defocus secondary effectiveness endpoint was met, with >0.5 D of increased range of focus
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 32
{32}
(p<0.0001; ITT population) of 20/32 or better visual acuity for Symfony subjects vs. monofocal control subjects.
**Figures 1 and 2** present the monocular defocus curve for the Symfony and monofocal control groups with mean values and error bars for confidence intervals and standard deviations, respectively, while **Figure 3** represents the binocular defocus curves for the Symfony and monofocal groups (with mean values and error bars for confidence intervals). **Figure 4** presents monocular defocus curves by pupil size for the Symfony group. Mean monocular visual acuities were 20/32 or better for the Symfony group through intermediate defocus values of -1.5 D (66 cm); mean binocular acuities were 20/32 or better for the Symfony group through -2.0 D (50 cm). Both the monocular and binocular defocus curves demonstrate that visual acuity monotonically decreased while maintaining a 1-2 line acuity difference over the monofocal group through 4.0 D of defocus. Visual inspection of the defocus curves yielded an improvement in the range of defocus with visual acuity of 20/32 or better in favor of the Symfony IOL by approximately 1 D. When monocular results were analyzed by pupil size, no appreciable pupil size effect was observed. Because visual acuity improves in monofocal subjects with pupil sizes ≤ 2.5 mm, the improvements in depth of focus between Symfony and monofocal groups are less pronounced in this subset of subjects. Some individual eyes showed drops in acuity below 20/32 between far and intermediate/near distances that are believed to be related to measurement noise when using the FrACT automated test system used in the study.
**FIGURE 1**
**Monocular Defocus Curves at 6 Months**
**Symfony and Monofocal Control**
**(with 95% Confidence Intervals)**

P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 33
{33}
**FIGURE 2**
**Monocular Defocus Curves at 6 Months**
**Symfony and Monofocal Control**
**(with Error Bars Representing 1 Standard Deviation)**

**FIGURE 3**
**Binocular Defocus Curves at 6 Months**
**Symfony and Monofocal Control**

P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 34
{34}
**FIGURE 4**
**Monocular Defocus Curve at 6 Months By Pupil Size**
**Symfony Subjects**

### Overall Spectacle Wear
Spectacle wear and other related items were assessed using a self-reported, binocular subjective questionnaire: the Patient Reported Spectacle Independence Questionnaire (PRSIQ). This questionnaire was developed and evaluated following the US FDA guidance document “Patient-Reported Outcomes Measures: Use in Medical Product Development to Support Labeling Claims,” dated December 2009. Although the questionnaire was not determined to be a psychometrically valid assessment of the concept of spectacle independence, data showed that the Symfony IOL achieved the secondary effectiveness endpoint of reduced overall spectacle wear compared to the control monofocal IOL.
The spectacle wear secondary effectiveness endpoint is based on the proportion of subjects who reported wearing glasses or contacts “none of the time” or “a little of the time” for overall vision, collected from a single question in the PRSIQ. **Figure 5** presents the frequency of overall spectacle wear for bilaterally implanted subjects at 6 months. There was a statistically significantly higher (p<0.0001; modified ITT population) proportion of subjects in the Symfony group compared to the monofocal group who reported wearing glasses “none of the time” or “a little of the time” (secondary effectiveness endpoint). Clinical significance was achieved with 85% (125/147) of Symfony subjects vs. 59.9% (88/148) of control subjects reporting wearing glasses “none of the time” or “a little of the time” for overall vision. Additionally, 62.6% (92/147) of Symfony subjects vs. 32.0% (47/148) of monofocal subjects indicated wearing glasses or contacts “none of the time” for overall vision.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 35
{35}
**Overall Spectacle Wear at 6 Months**

$^{a}$ “None of the time” and “A little of the time” combined; 1-sided Fisher’s exact test.
FDA review found the analyses on the key effectiveness endpoint as providing valid scientific evidence concerning the effectiveness of the device in providing improved intermediate and near visual acuity, while maintaining comparable distance high contrast visual acuity.
### 3. Subgroup Analyses
The following preoperative characteristics were evaluated for potential association with outcomes: age, sex, ethnicity, iris color and race. No statistically significant differences were found for age, sex, ethnicity or iris color. Although, there was a statistically significant difference with respect to race, with more non-White subjects in the monofocal control group than in the Symfony group, primary and secondary effectiveness endpoints were not significantly different among races. As expected, best corrected distance visual acuity was related to age, with younger subjects having a larger proportion of eyes at 20/20 or better. Outcomes for intermediate visual acuity, near visual acuity, depth of focus, and contrast sensitivity were consistent across demographic population and among study sites.
### 4. Financial Disclosure
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 18 investigators of which none were full-time or part-time employees of the sponsor and 4 had disclosable financial interests/arrangements as defined in 21 CFR 54.2(a), (b), (c) and (f) and described below:
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 36
{36}
• Compensation to the investigator for conducting the study where the value could be influenced by the outcome of the study: 0 investigators
• Significant payments of other sorts: 4 investigators
• Proprietary interest in the product tested held by the investigator: 0 investigators
• Significant equity interest held by investigator in sponsor of covered study: 0 investigators.
The applicant has adequately disclosed the financial interest/arrangements with clinical investigators. Statistical analyses were conducted by FDA to determine whether the financial interests/arrangements had any impact on the clinical study outcome. The information provided does not raise any questions about the reliability of the data.
# XI. SUMMARY OF SUPPLEMENTAL CLINICAL INFORMATION
. Two feasibility studies comparing the non-toric TECNIS® Symfony IOL to a monofocal control were conducted. No issues regarding device safety or lack of effectiveness were raised by the results from these studies.
Please note that Tables 17-38 in the labeling are incorporated by reference from previous approvals. As a result, these tables were not repeated within this SSED. Below identifies which tables are associated with each submission:
Tables 17-26: P980040/S039
Tables 27-35: P080010
Tables 36-38: P980040/S015
P980040/S015 was a 180-day supplement; therefore, for convenience these tables are provided below and referenced as Tables 29-31. Here, a total of 123 subjects were enrolled and implanted with the SENSAR 1-Piece IOL, Model AAB00. In the study population, 56.9% of subjects were female and 43.1% were male; 93.5% were Caucasian, 4.1% were Black and 2.4% were Asian. The best corrected distance visual acuity results for the “best case” subjects at 1 year postoperatively are provided in Table 30. In addition the results compared to the FDA Grid values (historical control) are presented in Table 31. All the associated adverse events for this study is included in Table 32.
P980040/S065: FDA Summary of Safety and Effectiveness Data
Page 37
{37}
# **TABLE 30**
**Best Corrected Distance Visual Acuity (Snellen Equivalent) at 1 Year**
**Best Case Subjectsᵃ (N = 110)**
| Age Group | N | 20/20 or Better | | 20/25 to 20/40 | | 20/50 to 20/100 | | 20/125 or Worse | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | n | % | N | % | n | % | n | % |
| < 60 | 11 | 11 | 100.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
| 60-69 | 35 | 29 | 82.9 | 6 | 17.1 | 0 | 0.0 | 0 | 0.0 |
| 70-79 | 46 | 39 | 84.8 | 7 | 15.2 | 0 | 0.0 | 0 | 0.0 |
| ≥ 80 | 18 | 14 | 77.8 | 4 | 22.2 | 0 | 0.0 | 0 | 0.0 |
| **TOTALᵇ** | **110** | **93** | **84.5** | **17** | **15.5** | **0** | **0.0** | **0** | **0.0** |
ᵃ Excludes subjects with macular degeneration at any time during the study.
ᵇ Includes three subjects who experienced a Nd:YAG posterior capsulotomy.
# **TABLE 31**
**Best Corrected Distance Visual Acuity (Snellen Equivalent) at 1 Year**
**…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.