P210036 · Baxter Healthcare Corporation · LMG · May 19, 2023 · General, Plastic Surgery
Device Facts
Record ID
P210036
Device Name
PerClot® Polysaccharide Hemostatic System
Applicant
Baxter Healthcare Corporation
Product Code
LMG · General, Plastic Surgery
Decision Date
May 19, 2023
Decision
APPR
Regulation
21 CFR 878.4490
Device Class
Class 3
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
P210036 · May 19, 2023
PerClot® Polysaccharide Hemostatic System
Baxter Healthcare Corporation
Published clinical literature; Unpublished OUS clinical studies; Post-market surveillance data
The sponsor utilized OUS clinical studies and published literature to provide supplemental evidence of safety and effectiveness across various surgical disciplines (ENT, cardiovascular, gynecological, orthopedic, etc.) and to support the device's long-term safety profile based on over 10 years of commercial use.
OUS clinical studies; Published literature; Post-market surveillance; Supplemental clinical evidence
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Published Literature (9 articles); 8 clinical trials and 1 systematic review
Patients undergoing ENT, cardiovascular/vascular, head and neck, plastic, and orthopedic surgery; Sample Size: 255
Not applicable for this study
Safety and effectiveness of PerClot in various surgical applications
Unpublished European and Chinese Studies; Clinical trials
Patients undergoing various surgical procedures (ENT, Gynecological, Coronary, Spinal, General, Hepatic); Sample Size: 267
Not applicable for this study
Safety and effectiveness
Indications for Use
PerClot® Absorbable Hemostatic Powder is indicated in surgical procedures (except neurological and ophthalmic) as an adjunctive hemostatic device to assist when control of suture line bleeding or capillary, venous and arteriolar bleeding by pressure, ligature, and other conventional procedures is ineffective or impractical. The PerClot 20cm Extender Tip and PerClot 38cm Laparoscopic Tip are intended for the application of PerClot Absorbable Hemostatic Powder onto surgical wound surfaces consistent with the product labeling in open and laparoscopic procedures, respectively.
Device Story
PerClot is an absorbable hemostatic powder derived from purified plant starch. It is supplied in a bellows delivery system with optional extender or laparoscopic tips. During surgery, the physician applies the powder directly to bleeding wound surfaces. The granules absorb water, forming a gelled adhesive matrix that acts as a mechanical barrier and promotes the physiological clotting cascade by accumulating platelets, red blood cells, and coagulation proteins. The matrix is enzymatically degraded by alpha-amylase, glucoamylase, and macrophages, typically within 96 hours. The device is used in open and laparoscopic procedures to assist in achieving hemostasis when conventional methods like pressure or ligature are insufficient. By providing an effective adjunctive hemostatic barrier, the device helps control bleeding, potentially reducing blood loss and operative time.
Clinical Evidence
Evidence includes a prospective, multicenter, randomized, active-controlled pivotal trial (CLOT Trial, n=324) comparing PerClot to Arista in cardiac, general, and urological surgeries. Primary endpoint: hemostasis at 7 minutes. PerClot demonstrated comparable performance to Arista. Secondary endpoints (hemostasis at 5 and 12 minutes) also showed comparable results. Safety profile was similar between arms, with no statistically significant differences in adverse events, re-operations, or blood loss. Supplemental evidence includes 7 unpublished studies (n=267) and 9 published articles (n=255) supporting safety and efficacy across various surgical specialties.
Technological Characteristics
Composed of biocompatible, non-pyrogenic, purified plant starch granules. Delivery system includes a bellows and optional malleable applicator tips (20cm Extender, 38cm Laparoscopic). Sterilization: E-Beam (bellows/standard tip) and Ethylene Oxide (accessory tips). Complies with ISO 10993-1:2018 for biocompatibility. MR Safe. No animal/human components, phthalates, or natural rubber latex.
Indications for Use
Indicated for patients undergoing open or laparoscopic surgical procedures (excluding neurological and ophthalmic) to assist in controlling suture line, capillary, venous, or arteriolar bleeding when conventional methods are ineffective. Contraindicated for injection or placement into blood vessels due to risk of embolization and death.
Regulatory Classification
Identification
An absorbable hemostatic agent or dressing is a device intended to produce hemostasis by accelerating the clotting process of blood. It is absorbable.
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
I. GENERAL INFORMATION
Device Generic Name: Hemostatic Agent
Device Trade Name: PerClot® Polysaccharide Hemostatic System
Device Product Code: LMG
Applicant’s Name and Address: Artivion, Inc. (formerly CryoLife, Inc.)
1655 Roberts Boulevard NW
Kennesaw, GA 30144
Date(s) of Panel Recommendation: None
Premarket Approval Application Number: P210036
Date of FDA Notice of Approval: 5/19/2023
II. INDICATIONS FOR USE
PerClot® Absorbable Hemostatic Powder
PerClot® Absorbable Hemostatic Powder is indicated in surgical procedures (except neurological and ophthalmic) as an adjunctive hemostatic device to assist when control of suture line bleeding or capillary, venous and arteriolar bleeding by pressure, ligature, and other conventional procedures is ineffective or impractical.
Accessory Tips
The PerClot® 20cm Extender Tip and PerClot 38cm Laparoscopic Tip are intended for the application of PerClot Absorbable Hemostatic Powder onto surgical wound surfaces consistent with the product labeling in open and laparoscopic procedures, respectively.
III. CONTRAINDICATIONS
Do not inject or place PerClot into blood vessels such as artery or vein as potential for embolization and death may exist.
IV. WARNINGS AND PRECAUTIONS
PMA 210036: Summary of Safety and Effectiveness Data
Page 1
{1}
The warnings and precautions can be found in the PerClot® Absorbable Hemostatic Powder labeling and the PerClot® accessory Tips labeling
## V. DEVICE DESCRIPTION
PerClot Absorbable Hemostatic Powder (PerClot) is a medical device composed of absorbable polysaccharide granules. The PerClot System is comprised of three (3) components: 1) PerClot (granules/powder); 2) PerClot Delivery System; and the optional 3) PerClot Accessory Tips. The PerClot Delivery System includes a bellows and a standard applicator tip. Refer to the image of the PerClot Delivery System in Figure 1. The PerClot granules are biocompatible, non-pyrogenic and derived from purified plant starch. The granules do not contain any human or animal components. PerClot granules have a molecular structure that rapidly absorbs water, forming a gelled adhesive matrix that provides a mechanical barrier against bleeding and results in the accumulation of platelets, red blood cells, and coagulation proteins (thrombin, fibrinogen, etc.). The gelled adhesive matrix thus promotes the normal, physiological clotting cascade. PerClot granules are enzymatically degraded by alpha-amylase and glucoamylase and by macrophages. Based on preclinical studies, absorption occurs within 96 hours. Absorption is dependent on the amount of material applied on the wound and the site of use.
PerClot is supplied as a bellows pre-loaded with hemostatic powder. The bellows contains at least 1, 3, or 5 grams of material.
- 1g: 1.4-1.6 grams
- 3g: 3.4-3.7 grams
- 5g: 5.4-5.8 grams
An applicator tip is also supplied for application of the hemostatic powder. Contents of the PerClot package are supplied sterile for single-patient use only. Not made with natural rubber latex.
PerClot must be stored between 5°C and 25°C.

Figure 1: PerClot Delivery System (Assembled)
PMA 210036: Summary of Safety and Effectiveness Data
Page 2
{2}
The PerClot Accessory Tips, which include the PerClot 20cm Extender Tip and PerClot 38cm Laparoscopic Tip, are sterile, single-use, malleable tips, designed to deliver PerClot Absorbable Hemostatic Powder onto surgical wound surfaces consistent with the PerClot Absorbable Hemostatic Powder product labeling. Refer to scaled images of the optional applicator tips in Figures 2 and 3.

Figure 2: PerClot 20cm Extender Tip – Optional applicator tip available separately

Figure 3: PerClot 38cm Laparoscopic Tip – Optional applicator tip available separately
## VI. ALTERNATIVE PRACTICES AND PROCEDURES
A variety of existing alternative practices for achieving surgical hemostasis exist today. For bleeding along a suture line, the most common conventional method is the application of additional sutures. Direct pressure compression (with and without gauze or sponge) remains a common surgical technique to assist the body in creating its own platelet plug or fibrin clot, depending on the severity of the injury and bleed. Staples and ligating clips can also be used as mechanical methods for achieving hemostasis. Other alternative methods employed when the body’s normal coagulation process cannot function properly include heat/energy-based methods such as electrocautery, ultrasonic devices and lasers. When these conventional methods are impractical or ineffective, there are topical agents on the market ranging from cellulose, gelatin and collagen-based products to active thrombin-based agents and chemical sealants.
PerClot Absorbable Hemostatic Powder, like other topical agents, can be used as an adjunctive agent during a variety of procedures where it may not be prudent or possible to
PMA 210036: Summary of Safety and Effectiveness Data
{3}
achieve surgical hemostasis with other means. Because it is plant-based, PerClot does not pose the risks associated with the use of hemostats manufactured from human- or animal-derived components.
# VII. MARKETING HISTORY
PerClot Absorbable Hemostatic Powder has been on the market outside the United States since 2008 when it was first CE-marked, manufactured, and distributed by Starch Medical (SMI) (San Jose, CA). The device consists of the PerClot Powder (hemostatic granules) and a delivery system, which includes a bellows containing the powder and an applicator tip. PerClot Accessory Tips (20cm Extender Tip and 38cm Laparoscopic Tip) are sold separately.
# PerClot Distributed Outside the United States (OUS)
Between September 28, 2010 and May 20, 2021, Artivion has distributed PerClot worldwide to OUS regions. Table 1 provides the countries where the device has been distributed. Adverse events occurring with the commercial PerClot device OUS are collected for evaluation. The device has not been withdrawn from any market for reasons of safety or effectiveness.
Table 1: PerClot OUS Distribution History
| Country/Geography | | |
| --- | --- | --- |
| Australia | Jordan | Philippines |
| Austria | Kuwait | Poland |
| Bahrain | Lebanon | Qatar |
| Belgium | Libya | Russia |
| Chile | Malaysia | Russian Federation |
| Colombia | Malta | Saudi Arabia |
| Costa Rica | Mauritius | Singapore |
| Cyprus | Mexico | South Africa |
| Denmark | Morocco | South Korea |
| Dominican Republic | Myanmar | Spain |
| Ecuador | Netherlands | Sweden |
| Finland | New Zealand | Switzerland |
| France | Nigeria | Thailand |
| Germany | Norway | Turkey |
| India | Oman | United Kingdom |
| Indonesia | Pakistan | Uruguay |
| Ireland | Palestine | United Arab Emirates |
| Israel | Panama | Vietnam |
| Italy | Peru | |
PMA 210036: Summary of Safety and Effectiveness Data
{4}
PMA 210036: Summary of Safety and Effectiveness Data
Page 5
# VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Below is a list of the potential adverse effects (e.g., complications) associated with the use of the device.
- Hypophosphatemia
- Pleural Effusion
- Anemia
- Hypotension
- Constipation
- Abdominal Pain
- Nausea
- Hyperglycemia
- Atelectasis
- Hypokalemia
- Acute Kidney Injury
- Dyspnea
- Leukocytosis
- Atrial Fibrillation
- Fever
- Pericardial Effusion
- Compression of Tubular Ducts, Vessels and Nerves with Swelling Causing Obstruction or Neuropathy
For the specific adverse events that occurred in the clinical study, please see section X below.
# IX. SUMMARY OF NONCLINICAL STUDIES
## A. LABORATORY STUDIES
Multiple non-clinical studies to support the safety and efficacy of PerClot and the Accessory Tips, including design verification and validation, pre-clinical open and laparoscopic animal studies, biocompatibility, shelf life, packaging testing and sterilization, and endotoxin validation and testing was conducted.
### Bench Testing
An overview of the non-clinical testing is provided in Table 2 through Table 4 that demonstrates that the PerClot System components meet the design specifications and are safe and effective for their intended clinical use.
Table 2: Summary of Bench Testing – PerClot Granules
{5}
PMA 210036: Summary of Safety and Effectiveness Data
Page 6
| Study | Test Purpose | Acceptance Criteria/Assessment Criteria | Results/Conclusion |
| --- | --- | --- | --- |
| The Determination of PerClot Polysaccharide Hemostatic System’s Gel Permeability Through a Suture Line | To determine if any of the PerClot gel could penetrate through the sutured incision | PerClot not detected on iodine-stained filter
Positive and negative controls behave as expected | PASS
Powdered and gelled PerClot does not permeate through suture lines on pressurized femoral popliteal arteries or saphenous veins |
| Evaluation of Potential Permeation of PerClot Polysaccharide Hemostatic System through an Arterial Puncture | To determine if, after the application of the PerClot Polysaccharide Hemostatic System onto an 18-gauge arterial puncture site, the PerClot permeated into the femoral artery | PerClot not detected on iodine-stained filter (PerClot does not seep through 18- gauge puncture into the femoral artery)
Negative control behaves as expected | PASS
Stained filters showed “No PerClot” for all test samples
Stained blood vessels showed “No PerClot” for all test samples
Some PerClot was visible around the puncture site in the bovine fatty tissue used to wrap the vessel
Powdered and gelled PerClot does not permeate beyond the fatty tissue in a pressurized 18-gauge arterial puncture |
| Testing the Homogeneity of PerClot | To confirm that PerClot Absorbable Hemostatic Powder is a homogeneous product | n=10 samples: ≤ L1% 15%
n=30 samples, if required: ≤ L1% (15%) and no individual content of any dosage unit is less than (1-L2*0.01)M nor more than (1+L2*0.01)M where L2=25.0 | PASS
The results of the n=10 samples yielded a degree of substitution average (label claim) of 0.349
The resulting acceptance value was 4.397% which is less than the acceptance maximum
The PerClot powder is homogenous |
{6}
| Study | Test Purpose | Acceptance Criteria/Assessment Criteria | Results/Conclusion |
| --- | --- | --- | --- |
| Evaluation of PerClot Performance When Left Open to the Operating Room Environment | To determine if the functional performance of the PerClot granules degrades over time when the PerClot bellows with the attached delivery applicator tips are exposed to the operating room (OR) environment | Rate of Water Absorption:
Mass absorbed at 10 sec must be ≥1.100g
Mass absorbed at 60 sec must be ≥2.900g
Total Water Absorbance:
All samples must have a swelling capacity ≥19 | PASS
At 10 seconds, all samples passed rate of water absorption testing with a range of 1.991g – 2.362g
At 60 seconds, all samples passed rate of water absorption testing with a range of 3.370g – 4.350g
All samples absorbed ≥19 times their weight
All samples were able to dispense at least 5g of PerClot granules with a range of 5.14g – 5.41g |
PMA 210036: Summary of Safety and Effectiveness Data
Page 7
{7}
Table 3: Summary of Bench Testing – PerClot Delivery System
| Study | Test Purpose | Acceptance Criteria/ Assessment Criteria | Results/Conclusion |
| --- | --- | --- | --- |
| Design Validation for Containing PerClot in the Primary Container and Dispensing the Marketed Volume | To validate that PerClot Absorbable Hemostatic Powder can be contained inside the primary container during agitation, and the amount of dispensed PerClot granules is greater than or equal to the marketed volume | After agitation, each primary container must be able to deliver > 5.0g of PerClot granules | PASS
Minimum amounts dispensed:
IDE: 5.07g
PMA: 5.04g
Both IDE and PMA primary containers and delivery tips delivered ≥5.0 g of PerClot granules after conditioning via agitation indicating the primary containers can contain and dispense the marketed weight of granules |
PMA 210036: Summary of Safety and Effectiveness Data
{8}
| Study | Test Purpose | Acceptance Criteria/ Assessment Criteria | Results/Conclusion |
| --- | --- | --- | --- |
| Design Validation for PerClot Ease of Removal from Packaging, Assembly, and Use | To validate that PerClot Absorbable Hemostatic Powder meets the user needs: 1) the delivery system must be easy to assemble, 2) must dispense PerClot granules, and 3) PerClot granules are ready to use off the shelf | All scaled testing requirements must be scored with an overall average rating of 3 (Satisfactory) or higher
If the bellow and delivery tip cannot be presented aseptically or any sample receives and overall average rating less than 3, the sample will be considered an “Incomplete Task” and evaluated by the team | PASS
No “Incomplete Tasks” and user needs met.
All samples had an overall average rating of 3.0 or greater, ranging from 4.0 to 5.0 in the IDE group and 3.0 to 5.0 in the PMA group
All primary operating functions were easily recognizable by the users
The average time to complete opening the pouches and fitting the delivery tip to the bellow was 33 seconds |
| Evaluation of PerClot Coverage When Dispensed Using the New Primary Container and Delivery Tip | To evaluate the coverage of the PerClot granules when they are dispensed using the new primary container and delivery tip | PerClot primary containers must be able to deliver 3.36-4.48g of PerClot granules and cover a 4cm x 4cm application site between 3-4 mm in depth | PASS
The dispensed weights ranged from 3.43g to 4.36g
The dispensed heights (thickness) ranged from 3mm to 4mm
All samples covered the 16cm² application site |
PMA 210036: Summary of Safety and Effectiveness Data
{9}
Table 4: Summary of Bench Testing – PerClot Accessory Tips
| Study | Test Purpose | Acceptance Criteria/ Assessment Criteria | Results/Conclusion |
| --- | --- | --- | --- |
| Design Validation for PerClot Ease of Removal from Packaging, Assembly, and Use | To validate that the PerClot Accessory Tips meet the system requirements and user interface specifications related to the preparation, use and disposal | All scaled testing requirements must be scored with an overall average rating of 3 (Satisfactory) or higher
If the laparoscopic tip cannot be presented aseptically or any sample receives and overall average rating less than 3, the sample will be considered an “Incomplete Task” and will be evaluated by the team | PASS
No “Incomplete Tasks” and all user needs met
All samples had an overall average rating of 3.0 or greater, ranging from 3.8 to 5.0
All primary operating functions were easily recognizable by the users |
| Evaluation of PerClot Coverage When Dispensed Using the 20cm Extender Tip | To evaluate the coverage of the PerClot granules when dispensed using the 20cm Extender Tip | PerClot Extender Tips must be able to deliver 3.36-4.48g of PerClot granules and cover a 4cm x 4cm application site between 3-4 mm in depth | PASS
The dispensed weights ranged from 3.37g to 4.32g
The dispensed heights (thickness) ranged from 3mm to 4mm
All samples covered the 16cm² application site |
| Evaluation of PerClot Coverage When Dispensed Using the 38cm Laparoscopic Tip | To evaluate the coverage of the PerClot granules when dispensed using the 38cm Laparoscopic Tip | PerClot Extender Tips must be able to deliver 3.36-4.48g of PerClot granules and cover a 4cm x 4cm application site between 3-4 mm in depth | PASS
The dispensed weights ranged from 3.46g to 4.12g
The dispensed heights (thickness) ranged from 3mm to 4mm
All samples covered the 16cm² application site |
PMA 210036: Summary of Safety and Effectiveness Data
{10}
| Study | Test Purpose | Acceptance Criteria/ Assessment Criteria | Results/Conclusion |
| --- | --- | --- | --- |
| Evaluation of the Flexibility and Rigidity of the PerClot 38cm Laparoscopic Tip | To verify that the PerClot 38cm Laparoscopic Tip has been designed such that the extruded tube portion of the assembly is rigid enough to prevent kinking during positioning of the tip to the site of the bleeding while allowing enough flexibility to allow tip advancement through a 5mm trocar | The 38cm laparoscopic tip must be able to be successfully inserted into and through the 5mm trocar
At each location:
• the extruded tube and the distal tip of the 38cm laparoscopic tip must not kink during positioning to occlude the delivery of the PerClot granules
• PerClot granules must be able to be dispensed through the 38cm laparoscopic tip
The 38cm laparoscopic tip must be able to be successfully removed through the 5mm trocar in a single piece (i.e., distal tip is still attached to the extruded tubing)
After removal from the trocar, the distal tip must not be visually deformed such that the distal tip would occlude delivery of the PerClot granules
The amount of PerClot dispensed must be ≥ the marketed volume of the bellow used | **PASS**
The 38cm laparoscopic tip was able to be successfully inserted into and through the 5mm trocar
At each location, the extruded tube and the distal tip of the 38cm laparoscopic tip did not kink during positioning
At each location, PerClot granules were able to be dispensed through the 38cm laparoscopic tip
The 38cm laparoscopic tip was able to be successfully removed through the 5mm trocar in a single piece
After removal from the trocar, the distal tip was not visually deformed
The amount of PerClot dispensed was ≥ the marketed volume of the bellow used (5g) with a range of 5.15g to 5.75g |
PMA 210036: Summary of Safety and Effectiveness Data
Page 11
{11}
# Biocompatibility Testing
Biological evaluation was conducted in compliance with ISO 10993-1:2018, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. To evaluate the biological safety of the PerClot Absorbable Hemostatic Powder, consideration was given to the following: type of patient contact and intended clinical use; potential hazards associated with the materials of construction, the history of clinical use of the materials of construction, manufacturing process, the results of biocompatibility and chemical characterization testing performed, and the clinical history of the PerClot Absorbable Hemostatic Powder. The testing demonstrated that the PerClot Absorbable Hemostatic Powder is biocompatible as summarized in Table 5 below.
The PerClot Granules are categorized according to ISO 10993-1:2018 as implant medical device with prolonged ≥24 hours but not >30 days) contact with tissue/bone and blood based on its intended clinical use. The assessment of the PerClot granules is separate from that of the delivery system. The PerClot Delivery System is categorized according to ISO 10993-1:2018 as an externally communicating device with limited ≤24 hours) contact with tissue. The PerClot Standard Applicator Tip and Accessory Tips are categorized according to ISO 10993-1:2018 as an externally communicating device with limited ≤24 hours) contact with both tissue and blood path, indirect. Table 5 identifies the biological endpoints the results for test articles that underwent biocompatibility testing.
PMA 210036: Summary of Safety and Effectiveness Data
Page 12
{12}
Table 5: Summary of Biocompatibility Testing
| Endpoint | Applicable ISO 10993 Standard | Result |
| --- | --- | --- |
| PerClot Granules | | |
| Physical and Chemical Information | ISO 10993-1: Evaluating and Testing within a Risk Management Process | PASS
Negligible toxicological risk |
| Cytotoxicity | ISO 10993-5: Tests for in vitro cytotoxicity | PASS
Non-cytotoxic |
| Sensitization | ISO 10993-10: Tests for irritation and skin sensitization | PASS
Non-sensitizer |
| Intracutaneous Irritation | ISO 10993-10: Tests for irritation and skin sensitization | PASS
Non-irritant |
| Material Mediated Pyrogenicity | ISO 10993-11: Tests for systemic toxicity | PASS
Non-pyrogenic |
| Acute System Toxicity | ISO 10993-11: Tests for systemic toxicity | PASS
Non-toxic |
| Sub-Acute Systemic Toxicity | ISO 10993-11: Tests for systemic toxicity | PASS
Non-toxic
Non-irritant |
| Implantation | ISO 10993-6: Tests for local effects after implantation | PASS
At 28 days +/-2 days:
Ranked Reactivity Score was 0.0 for the liver/serosa and parenchyma
The Test Article was characterized as a non-irritant and was nonreactive as compared to the Control Article |
PMA 210036: Summary of Safety and Effectiveness Data
{13}
| Hemocompatibility | ISO 10993-4: Selection of tests for interactions with blood | PASS
Non-hemolytic |
| --- | --- | --- |
| Genotoxicity | ISO 10993-3: Tests for genotoxicity, carcinogenicity, and reproductive toxicity | PASS
Non-genotoxic |
| PerClot Delivery System | | |
| Cytotoxicity | ISO 10993-5: Tests for in vitro cytotoxicity | PASS
Non-cytotoxic |
| Sensitization | ISO 10993-10: Tests for irritation and skin sensitization | PASS
Non-sensitizer |
| Intracutaneous Irritation | ISO 10993-10: Tests for irritation and skin sensitization | PASS
Non-irritant |
| Material Mediated Pyrogenicity | ISO 10993-11: Tests for systemic toxicity | PASS
Non-pyrogenic |
| Acute System Toxicity | ISO 10993-11: Tests for systemic toxicity | PASS
Non-toxic |
| Hemocompatibility | ISO 10993-4: Selection of tests for interactions with blood | PASS
Non-hemolytic |
| PerClot Standard Tip and Accessory Tips | | |
| Cytotoxicity | ISO 10993-5: Tests for in vitro cytotoxicity | PASS
Non-sensitizer |
| Sensitization | ISO 10993-10: Tests for irritation and skin sensitization | PASS
Non-sensitizer |
| Intracutaneous Irritation | ISO 10993-10: Tests for irritation and skin sensitization | PASS
Non-irritant Non-pyrogenic |
| Material Mediated Pyrogenicity | ISO 10993-11: Tests for systemic toxicity | PASS
Non-pyrogenic |
PMA 210036: Summary of Safety and Effectiveness Data
Page 14
{14}
| Acute System Toxicity | ISO 10993-11: Tests for systemic toxicity | PASS
Non-toxic |
| --- | --- | --- |
| Hemocompatibility | ISO 10993-4: Selection of tests for interactions with blood | PASS
Hemolytic index of 0.2%; Non-hemolytic |
## B. ANIMAL STUDIES
Pre-clinical animal studies were conducted to support the safety and efficacy of PerClot and the Accessory Tips for their intended use in open and laparoscopic procedures. These studies demonstrate the safety and efficacy of the PerClot System and establish that benefit of its use outweighs the risk in open and laparoscopic procedures.
An overview of the non-clinical animal studies is provided in Table 6.
Table 6: Summary of Non-Clinical Animal Studies
| Study | Study Design | Acceptance Criteria/Endpoint(s) | Results/Conclusion |
| --- | --- | --- | --- |
| A GLP Preclinical Evaluation of the Effectiveness and Safety of a Polysaccharide Hemostat to Achieve Hemostasis of an Abdominal Aorta Lesion in Pig | Swine (pigs)
Total Used: n=14 (13 + 1 backup)
Test:
PerClot Polysaccharide Hemostatic System (PerClot Granules and Delivery System)
Control:
Gelfoam Plus
n=1 Acute* treated with test
n=7 survival treated with test (including 1 backup)
n=6 survival treated with control | The primary efficacy endpoint was hemostasis of the treatment site within 5 minutes after hemostat application
Any scoring or acceptance criteria related to histopathology are described separately in the individual study reports | Hemostasis:
All animals in both the PerClot and control groups achieved the primary efficacy endpoint of hemostasis by 5 minutes
PerClot was as safe as the control article when implanted chronically for a minimum of 14 days
PerClot was absorbed by all animals according to histology, while the control article had some residual material at the time of necropsy
The implant of neither PerClot or control article initiated an immune |
PMA 210036: Summary of Safety and Effectiveness Data
{15}
| Study | Study Design | Acceptance Criteria/Endpoint(s) | Results/Conclusion |
| --- | --- | --- | --- |
| | *Acute animal not included in the safety assessments of the test or control article | | response |
| A GLP Preclinical Evaluation of the Effectiveness and Safety of a Polysaccharide Hemostat to Achieve Hemostasis of a Kidney Lesion in a Pig | Swine (pigs)
Total Used: n=15 (12+ 3 backup)
Test:
PerClot Polysaccharide Hemostatic System (PerClot Granules and Delivery System)
Control:
Gelfoam Plus
n=9 survival treated with test (including 3 backups)
n=6 survival treated with control | The primary efficacy endpoint was hemostasis of the treatment site within 5 minutes after hemostat application
Any scoring or acceptance criteria related to histopathology are described separately in the individual study reports | Hemostasis:
PerClot:
5 minutes: 78%
PerClot was still effective within 10 minutes of application
Control:
5 minutes: 83%
PerClot was as safe as the control article when implanted chronically for a minimum of 14 days
PerClot was absorbed by all animals according to histology, while the control article still had some animals that had residual material at the time of necropsy
The implant of neither PerClot nor control article initiated an immune response
Clinical pathology (blood testing and animal morbidity and mortality) did not indicate a negative response to the implant of either PerClot or the control article |
| A GLP Preclinical Evaluation of the Effectiveness and Safety of a Polysaccharide | Swine (pigs)
Total Used: n= 19 (17 + 2 backup)
Test: | The primary efficacy endpoint was defined as the achievement of hemostasis by 5 minutes
Any scoring or | Hemostasis:
PerClot:
1 minute: 44%
3 minutes: 56% |
PMA 210036: Summary of Safety and Effectiveness Data
Page 16
{16}
PMA 210036: Summary of Safety and Effectiveness Data
Page 17
| Study | Study Design | Acceptance Criteria/Endpoint(s) | Results/Conclusion |
| --- | --- | --- | --- |
| Hemostat to Achieve Hemostasis of a Liver Lesion in a Pig | PerClot Polysaccharide Hemostatic System (PerClot Granules and Delivery System)
**Control:**
Gelfoam Plus
n=13 survival treated with test (including 2 backups)
(Total 13 animals with test and 16 wounds)
n=6 survival treated with control | acceptance criteria related to histopathology are described separately in the individual study reports | 5 minutes: 69%
10 minutes: 81%
Control:
1 minute: 83%
3 minutes: 100%
5 minutes: 100%
PerClot was as safe as the control article when implanted chronically for a minimum of 14 days
PerClot was absorbed by all animals according to histology, while some animals implanted with the control article had residual material present at the time of necropsy
There was a mild to moderate foreign body response noted by histology in the control animals
The implant of neither PerClot nor control had an immune response in regard to the clinical pathology evaluation |
| A GLP Evaluation of the Effectiveness of Two Plant Based Hemostats in a Pig Bleeding Model | **Swine (pigs)**
Total Used: n= 15
n= 3 bleeding group
n=6 treated with test
n=6 treated with control
**Pre-determined sized liver injuries of the following dimensions were** | The primary endpoint was the achievement of hemostasis by 5 minutes after Hemostat application; and secondary endpoints were hemostasis at the site of application evaluated at 1,3,7, and 10 minutes | **Hemostasis:**
PerClot:
1 minute: 52.78%
3 minutes: 72.22%
5 minutes: 88.89%
7 minutes: 94.45%
10 minutes: 97.23% |
{17}
PMA 210036: Summary of Safety and Effectiveness Data
Page 18
| Study | Study Design | Acceptance Criteria/Endpoint(s) | Results/Conclusion |
| --- | --- | --- | --- |
| | used throughout the study.
Each test system had 6 different sizes of injuries each created on their liver. Each test system had one of each of these injuries created on their liver:
• 0.3 cm depth by 1.0 cm diameter
• 0.5 cm depth by 1.0 cm diameter
• 0.3 cm depth by 1.5 cm diameter
• 0.5 cm depth by 1.5 cm diameter
• 0.5 cm depth by 2.0 cm diameter
• 1.0 cm depth by 2.0 cm diameter
Test:
PerClot Polysaccharide Hemostatic System (PerClot Granules and Delivery System)
Control:
Surgicel Original Absorbable Hemostat | | Control:
1 minute: 27.78%,
3 minutes: 44.45%,
5 minutes: 91.67%
7 minutes: 91.67%,
10 minutes: 97.23%
For the 10-minute bleeding evaluation group, total blood loss throughout the entire evaluation period ranged from 1.22g to 163.18g
When the collected 30 second fluid mass was 10.0g or less, both PerClot and control were at least 90% effective for the achievement of the primary endpoint of hemostasis by 5 minutes
All three bleeding severity methods (hemoglobin concentration, mass and bleed scoring) were used to assess the blood loss from the pre-determined sized liver injuries
The bleeding severity assessment methods did not predict time to hemostasis. There were no strong relationships between bleeding severity and the amount of PerClot or control applied |
| Assessment of Blood Glucose Levels During Degradation of Starch Based | New Zealand White Rabbits
Total used n=15
n=6 received the test article | Evaluation of blood glucose levels | Mean baseline blood glucose levels were similar for all three groups (138.56 mg/dL, 147.33 mg/dL, and 139.33 mg/dL for PerClot, |
{18}
| Study | Study Design | Acceptance Criteria/Endpoint(s) | Results/Conclusion |
| --- | --- | --- | --- |
| Hemostatic Agents in a Rabbit Model | n=6 received the control article
n=3 were operated on as sham controls
**Test:**
PerClot Hemostatic system (PerClot Granules and Delivery System)
**Control:**
Arista AH (Arista) – P050038 (also referenced as HemoStase) Absorbable Hemostatic Particles | | Control, and sham groups, respectively)
Blood glucose levels 1 hour postoperatively were significantly higher than baseline blood glucose for all three groups (p<0.05), with glucose levels remaining significantly higher at 2 hours postoperatively for animals treated with PerClot or Control
The 1 hour and 2 hour glucose levels in animals treated with either PerClot or Control were significantly higher than glucose levels in sham operated control animals with control treated animals having higher glucose levels than PerClot treated animals 2 hours postoperatively
By 48 hours, glucose levels for all three groups were similar and remained so for the remainder of the study
At time of necropsy, no remaining article was grossly observed in any of the animals treated with either PerClot or control
Overall, no adverse events were noted in response to implanted PerClot for any animals for the duration of the study |
| Determination of Systemic Blood | New Zealand White Rabbits
Total used n=15 | Evaluation of blood glucose levels | Baseline blood glucose levels were similar for all |
PMA 210036: Summary of Safety and Effectiveness Data
Page 19
{19}
PMA 210036: Summary of Safety and Effectiveness Data
Page 20
| Study | Study Design | Acceptance Criteria/Endpoint(s) | Results/Conclusion |
| --- | --- | --- | --- |
| glucose Levels During Degradation of Starch Based Hemostatic Agents in a Rabbit Model | n=6 received the test article
n=6 received the control article
n=3 were operated on as sham controls
**Test:**
PerClot Hemostatic system (PerClot Granules and Delivery System)
**Control:**
Arista (also referenced as HemoStase) Absorbable Hemostatic Particles | | three groups (153.61 mg/dL, 148.17 mg/dL, and 141.89 mg/dL for PerClot,
Control, and Sham groups, respectively) and all three treatment groups had mean baseline blood glucose levels within normal range
Blood glucose levels 1 hour post-operatively were significantly higher than baseline blood glucose for all three groups (p<0.05), with glucose levels remaining relatively high at 2 hours post-operatively for animals treated with PerClot or Control
Glucose levels of PerClot and Control animals at 2 hours were almost equivalent while slightly higher than glucose levels in Sham operated control animals
By 4 hours, glucose levels for all three groups were similar and remained normal for the remainder of the study.
It was concluded that known structural differences in the polysaccharide molecules and the rates of breakdown of the two products did not result in elevated systemic blood glucose levels during the degradation of PerClot and control in a rabbit model
As all groups experienced brief elevation of blood |
{20}
PMA 210036: Summary of Safety and Effectiveness Data
Page 21
| Study | Study Design | Acceptance Criteria/Endpoint(s) | Results/Conclusion |
| --- | --- | --- | --- |
| | | | glucose levels at 1 and 2 hours post-surgery, postoperative stress and excitement most likely attributed to those slight elevations
This study demonstrated that implantation of starch based hemostatic agents into the body do not significantly alter systemic blood glucose levels during degradation for extended periods of time and that degradation of the particles occurs within 96 hours after application |
| Laparoscopic Application of the PerClot Absorbable Hemostatic Powder (5g) in a Swine Hemorrhagic (Liver) Model | Swine (Pigs)
Total Used = 8 (7 + 1 backup)
n=4 implanted with the test article
n=3 implanted with the control article
n=1 termination prior to procedure start
**Test:**
PerClot Hemostatic Powder (PerClot Granules and Delivery System)
PerClot 38cm Laparoscopic tip (Accessory Tip)
**Control:**
Arista Absorbable Hemostatic Powder
Arista FlexiTip Applicator, | Acceptance criteria included but was not limited to the following:
**Safety & Efficacy**
-Rating of Yes or No on End Use Rating Scale with success defined as a %
-Evidence of clinically significant events
-Complete cessation of bleeding (hemostasis)
-Average score of 2 or below for adhesions
**Usability**
-Average Rating of 3 or above in the end-user rating scale | The PerClot Absorbable Hemostatic Powder (5g) and PerClot 38cm Laparoscopic Tip met the safety and efficacy criteria set forth in the protocol for comparison with the Arista Absorbable Hemostatic Powder (5g) and FlexiTip Applicator, 38cm in a healthy swine model over 28 days from treatment.
All PerClot and control animals had easily removable (score 1) adhesions associated with the liver wounds created for the surgical model. The adhesions were mainly from the liver wound sites to the omentum and there were no adhesions observed remote from the surgical site. There were no findings of concern in the non-target tissues (non-treated regions of the |
{21}
PMA 210036: Summary of Safety and Effectiveness Data
Page 22
| Study | Study Design | Acceptance Criteria/Endpoint(s) | Results/Conclusion |
| --- | --- | --- | --- |
| | 38cm | | liver, hepatic lymph nodes, kidneys, spleen, pancreas, or lungs).
The observed easily removable adhesions are expected consequence due to prior experimental manipulations and therefore do not represent a clinically significant safety concern.
There was no evidence of significant (Grade 3 or above) adhesion formation found in any of the study animals regardless of assignment to receive PerClot or Control Articles and therefore the acceptance criteria was met.
Unintended dispersion (spilling remote from the target surgical site) and tip leakage was evaluated; 96.8% of the PerClot treated animals met the success criteria for this endpoint, vs 91.6% of the control treated animals.
In conclusion, this study demonstrated that the PerClot Absorbable Hemostatic Powder (5g) and PerClot 38cm Laparoscopic Tip did not introduce any new safety risks when compared to the predicate Arista Absorbable Hemostatic Powder (5g) and FlexiTip Applicator, 38cm. |
{22}
PMA 210036: Summary of Safety and Effectiveness Data
Page 23
# C. ADDITIONAL STUDIES
## Shelf-life Testing
Real-time testing conducted in support of devices' shelf-life has demonstrated that the product and packaging meet the required specifications for the stated life of the product.
Real-Time stability testing (PerClot Granules) and Real-Time and Accelerated stability testing (optional applicator tips) is ongoing on test articles following sterilization. Interim Real-Time stability data available at time of PMA approval establishes a shelf-life of 1 year for PerClot Granules when stored between 5°C and 25°C. Accelerated aging studies have established a shelf-life of 3 years for the optional applicator tips, when stored between 5°C and 25°C.
Sterile barrier system integrity testing was completed in compliance with ISO 11607-1: 2019, Packaging for terminally sterilized medical devices – Part 1: Requirements for materials, sterile barrier systems and packaging systems.
All package integrity testing conducted for the PerClot System and PerClot Accessory Tips demonstrate that these devices are safe and effective for their intended use in open and laparoscopic procedures.
## Sterilization
The PerClot filled bellows, snap-fit adaptor and cap and the standard Applicator Tip are terminally sterilized using an Electron Beam (E-Beam) irradiation in accordance with ISO 11137, Sterilization of Health Care Products: Requirements for Validation and Routine Control of - Radiation Sterilization.
The Accessory Tips (20cm Extender Tip and 38cm Laparoscopic Tip) are sterilized via Ethylene Oxide in accordance with ISO 11135: Sterilization of health-care products — Ethylene oxide — Requirements for the development, validation and routine control of a sterilization process for medical devices.
All sterilization testing conducted for the PerClot System and PerClot Accessory Tips demonstrate that these devices meet a sterility assurance level (SAL) of 10⁻⁶ and are safe and effective for their intended use in open and laparoscopic procedures.
## Other Essential Non-Clinical Testing and Assessments
Based on the results of these tests the following is concluded: the PerClot System is MR Safe; the PerClot System does not contain materials of animal origin, the PerClot Absorbable Hemostatic Powder does not contain phthalates; specifically, the components of the PerClot System do not pose a risk to the patient from exposure to phthalates; and the PerClot System is not made with natural rubber latex. The PerClot system does not pose a risk of significant adhesions, if per the IFU, once hemostasis is achieved, excess PerClot is removed from the site of application by irrigation and aspiration and inaccurate delivery, leakage or dispersion of PerClot Absorbable Hemostatic Powder is not likely.
{23}
PMA 210036: Summary of Safety and Effectiveness Data
Page 24
# X. SUMMARY OF PRIMARY CLINICAL STUDIES
PerClot has an extensive clinical history which includes the clinical trial conducted under IDE G110072 in the US, additional clinical studies conducted Outside-the-US (OUS), clinical literature from the SMI-marketed product, and over 10 years of commercial use OUS with a safe history. The clinical history of reported scientific literature and unpublished clinical trials, including the IDE trial, totals more than 650 patient treatments over a 12-year period with positive clinical outcomes. The studies supporting the safety and efficacy of PerClot include:
- A prospective, multicenter, multidisciplinary, randomized, controlled (Arista) clinical IDE investigation (CLOT Trial; G110072) for subjects undergoing open elective cardiac, general, or urological surgical procedures, with a total of 324 subjects randomized to PerClot (n=161) or Arista (n=163) subjects at 19 US centers.
- Four unpublished clinical studies conducted in Europe totaling 119 PerClot subjects and three unpublished clinical studies conducted in China totaling 148 PerClot subjects. Surgical areas covered in these studies are discussed in Section XI.
- Nine published literature articles (8 clinical trials and 1 systematic review) totaling 255 PerClot subjects. Surgical areas covered in these studies are discussed in Section XI.
- Over 10 years of marketing history in more than 50 countries with a safe history.
# Summary of Primary Clinical Study (CLOT Trial)
A clinical trial was conducted to establish a reasonable assurance of safety and efficacy of PerClot as an adjunctive hemostatic device to assist when control of suture line bleeding or capillary, venous and arteriolar bleeding by pressure, ligature, and other conventional procedures is ineffective or impractical, excluding neurological and ophthalmic surgeries, in the US under IDE G110072. Data from this study forms the primary clinical basis for the PMA submission.
# A. Study Design
Patients were treated and enrolled between March 2015 and January 2019. The database for this PMA reflected data collected through February 2021 and included 324 patients. There were 24 investigational sites with 19 sites enrolling patients into the study.
The CLOT Investigation, a pivotal study of the hemostatic agent PerClot, was designed as a prospective, multicenter, multidisciplinary, randomized, active-controlled (Arista) clinical investigation for subjects undergoing open elective cardiac, general, or urological surgical procedures. The primary objective of this investigation was to demonstrate non-inferiority in the achievement of hemostasis of the treated bleeding site at 7 minutes in subjects receiving PerClot compared to subjects receiving Arista in multiple surgical disciplines when used as an adjunct to conventional means of achieving hemostasis such as pressure or ligature. For
{24}
this trial, hemostasis was defined as complete cessation of bleeding and was assessed at the time of surgery by the treating physician.
Subjects were randomized in a 1:1 ratio to PerClot or Arista with randomization stratified by bleeding severity (severity score 1 vs. 2) and therapeutic area (cardiac, general, or urology).
The study was sponsored by Artivion, Inc. and was conducted in compliance with United States Food and Drug Administration (FDA) regulations 21 CFR Parts 50, 54, 56, and 812, and also International Council for Harmonization (ICH) E6 Good Clinical Practices (GCP). Artivion collaborated with Contract Research Organizations (CRO) for study management, data management, site activation, DSMB facilitation, site monitoring, and data verification.
A DSMB with independent medical experts and an expert statistician operated throughout the course of the study to review unblinded data regarding trial conduct, patient selection, and safety and efficacy endpoints. These data were 100% source data monitored throughout the course of the study.
The clinical trial used Arista Absorbable Hemostatic Particles as the active control device. Arista is an ideal comparative device for the following reasons:
- The two products are based on the same raw materials (potato starch) and do not contain any pharmaceutical components, such as thrombin or fibrinogen.
- The two products are used under the same conditions and have the similar principles of operation.
- The products are similar in terms of clinical, biological and technical characteristics.
- Arista is a legally marketed alternative with similar indications for use.
The study enrolled subjects undergoing an elective surgical procedure in one of three different therapeutic areas (general, cardiac, or urology) with bleeding categorized by a bleeding severity score of 1 or 2. The bleeding site area and bleeding severity were rigorously defined and measured in each case prior to determination of subject eligibility preceding enrollment. The bleeding anatomic site could be no more than 25 cm² and the area for hemostatic application could be no more than 47 cm².
The bleeding severity score was defined by ranges of bleeding flux which was measured and calculated by weighing dedicated gauze with known absorption properties before and after applying the gauze for a measured period of time (see Table 7). Subjects with ooze (>0.000040-0.0056 g/(cm² s)) or slight bleeding (>0.0056-0.013 g/(cm² s)) were eligible, but subjects with moderate, severe, or life-threatening bleeding were not. These rigorous measurement procedures provide assurance that the bleeds treated in the trial met eligibility criteria and are accurately classified.
PMA 210036: Summary of Safety and Effectiveness Data
Page 25
{25}
Table 7: Bleeding Severity Score Definition
| Bleeding Severity Score | Bleeding Flux (g/cm² per second) |
| --- | --- |
| 0=No Bleeding | 0-0.000040 |
| 1=Ooze | >0.000040-0.0056 |
| 2=Slight Bleeding | >0.0056-0.013 |
| 3=Moderate Bleeding | >0.013-0.041 |
| 4=Severe Bleeding | >0.041-0.063 |
| 5=Life-Threatening Bleeding | >0.063 |
1. Subject Inclusion and Exclusion Criteria
Patient selection criteria were based on the inclusion and exclusion criteria listed below.
Preoperative Inclusion Criteria
- Subject is undergoing one of the following open elective cardiac, general, or urological surgical procedures:
- Cardiac procedure (Epicardium);
- Cardiac procedure (Aortic Anastomosis or Aortotomy Suture Line);
- Liver resection;
- Total splenectomy;
- On-clamp partial nephrectomy; or
- Radical nephrectomy.
- Subject is willing and able to give prior written informed consent for investigation participation; and
- Subject is > 22 years of age.
Preoperative Exclusion Criteria
- Subject with known sensitivity to starch or starch-derived materials;
- Subject who has a clinically significant coagulation disorder or disease, defined as a platelet count <100,000 per microliter, International Normalized Ratio >1.5, or a PTT more than 1.5 times outside the laboratory’s normal reference range;
PMA 210036: Summary of Safety and Effectiveness Data
Page 26
{26}
- Subject who used corticosteroids (excluding inhalers, eye-drops, and dermatologic corticosteroids) within 6 weeks prior to surgery;
- Subject who has been treated with an investigational product and has not completed the entire follow-up period for that investigational product;
- Subject who is pregnant (as confirmed by a pregnancy test), planning on becoming pregnant during the follow-up period, or actively breast-feeding; and
- Subject with poor blood glucose control as per glycosylated hemoglobin > 9%.
## Intraoperative Inclusion Criteria
- Subject is undergoing one of the following elective procedures:
- Cardiac procedure (Epicardium);
- Cardiac procedure (Aortic Anastomosis or Aortotomy Suture Line);
- Liver resection;
- Total splenectomy;
- On-clamp partial nephrectomy; or
- Radical nephrectomy
- Subject in whom all visible vessels or suture holes, greater than or equal to 2mm in diameter have been ligated, or suture line gaps greater than or equal to 2mm have been ligated;
- Subject in whom there is bleeding at the specified area for each surgical procedure after any applicable conventional means for hemostasis are attempted as specified by the intraoperative protocol;
- Subject in whom the anatomic site is equal to or less than 25cm²;
- Subject in whom the anatomic application site is equal to or less than 47cm²; and
- Subject in whom the bleeding flux from the identified bleeding site is >0.000040[g/(cm²·s)] and ≤0.013[g/(cm²·s)].
## Intraoperative Exclusion Criteria
- Subject undergoing a cardiac procedure in which there is no aortic anastomosis or aortotomy suture line to evaluate using the bleeding severity scale (i.e., not for treatment at the distal coronary artery bypass graft anastomosis);
PMA 210036: Summary of Safety and Effectiveness Data
Page 27
{27}
- Subject in whom any major intraoperative bleeding incidences during the surgical procedure occurred (i.e., subject with assignment of an American College of Surgeons Advanced Trauma Life Support Hemorrhage Class of II, III, or IV Hemorrhage);
- Subject who has an active or potential infection at the surgical site, or whose surgical wound is defined as a wound classification of CO (Contaminated) or D (Dirty or Infected) based upon the Center for Disease Control and Prevention’s wound classification system;
- Subject who has undergone platelet receptor GP IIb/IIIa antagonist therapy less than 48 hours prior to surgery.
## 2. Patient Follow-up Schedule
Follow-up occurred intra-operatively, post-operatively and at discharge (up to 14 days post-randomization) and 6-weeks post-randomization. Oncology subjects had a survival assessment at 24 months post-randomization. Five patients enrolled under the original protocol version had follow-up at 1 and 3 months rather than 6 weeks post-randomization.
## 3. Clinical Endpoints
The primary efficacy endpoint was the proportion of patients with complete hemostasis at 7 minutes post-application. The primary analysis was a non-inferiority comparison with a margin of 10% in order to demonstrate that the PerClot was clinically comparable or better than Arista with respect to the primary efficacy endpoint. The primary hypothesis is that the proportion of subjects achieving hemostasis of the first treated lesion by 7 minutes in the PerClot subjects is no more than 10% lower than the proportion of subjects achieving hemostasis within 7 minutes in Arista subjects:
$$
\mathrm{H}_0: \mathrm{P}_{\text{PerClot}} < \mathrm{P}_{\text{Arista}} - 10\%
$$
$$
\mathrm{Ha}: \mathrm{P}_{\text{PerClot}} \geq \mathrm{P}_{\text{Arista}} - 10\%
$$
Non-inferiority comparison of the hemostasis rate between PerClot and Arista was performed using the Farrington-Manning method in the As Treated population. The primary non-inferiority endpoint was met if the one-sided p-value < 0.025. A total sample size of 324 was planned to provide >80% power for this test assuming a 90% hemostasis rate for both PerClot and Arista and a 5% attrition.
A secondary endpoint was the complete hemostasis at 5 minutes post-application, also using a non-inferiority comparison with 10% non-inferiority margin. An additional hemostasis assessment was performed at 12 minutes to check whether hemostasis was maintained at 5 minutes after the primary endpoint assessment time of 7 minutes. If the primary endpoint was met, the secondary endpoint was evaluated and was met if the one-sided p-value < 0.025.
PMA 210036: Summary of Safety and Effectiveness Data
Page 28
{28}
The primary efficacy analysis was performed in several analysis populations. The intention-to-treat (ITT) analysis population included all randomized subjects and subjects were analyzed according to the study arm to which they were randomized. The ITT population consisted of all 324 patients that were randomized with 161 PerClot and 163 Arista subjects. The As Treated (AT) analysis population included all subjects who were randomized and treated with either PerClot or Arista, with subjects analyzed according to the treatment group for the treatment received (even if this differed from randomization assignment). The Per Protocol (PP) analysis population included all subjects who were randomized and treated with either PerClot or Arista and had no major protocol deviations. Due to the primary endpoint being a non-inferiority comparison, study success is based on the non-inferiority hypothesis being met in the AT analysis population.
The study was designed to assess safety through reporting on adverse events, including their relationship to the therapeutic device, seriousness, and severity. Safety and efficacy were further characterized by prospective data collection of key supplemental safety parameters including: operative time, maintenance of hemostasis 5 minutes after primary endpoint assessment at 7 minutes, alternate means used to achieve hemostasis, estimated blood loss, units of blood transfused, re-operation, and hospitalization time. The safety assessment was performed over a period of 6 weeks after the surgery to provide sufficient time for bleeding complications to manifest. Additionally, all-cause mortality assessments through 24 months were completed for oncology patients.
## B. Accountability of PMA Cohort
In total, the study consented 582 subjects and enrolled 324 subjects across 19 Sites. Study attrition and withdrawal were low in the study with 13 subjects 4% exiting the study early. Seven subjects discontinued participation prematurely for a reason other than death (2 lost to follow-up, 2 withdrawn after randomization before any hemostatic agent was applied, 1 voluntarily withdrew, 1 unable to return for an in-person visit due to remaining in a long term care facility, and 1 cancelled 6-week appointment with the transplant team and followed up with their primary care physician. Six subjects died during study follow-up through 6 weeks (1 PerClot and 5 Arista). Compliance with mortality reporting through up to 24 months was 99.4% (177/178).
Randomization for the trial was stratified by bleeding severity score and therapeutic area to help ensure balance within these sub-categories. The therapeutic areas included General Surgery (N=155), Urology (N=83) and Cardiac Surgery (N=86). Subjects enrolled had a bleeding severity score of 1 in approximately 57% of cases and a bleeding severity score of 2 in approximately 43% of cases. As shown in Table 8, approximately 50% of the subjects were undergoing a general surgery procedure, with approximately 25% in each of the other therapeutic areas (cardiac and urology). Use in cardiac cases included the application of the hemostatic agents to a surgical graft. The study was designed to pool data across the therapeutic areas and two bleeding severity scores. It was not powered to compare the relative safety and efficacy of PerClot and Arista within the individual arms of the study.
PMA 210036: Summary of Safety and Effectiveness Data
Page 29
{29}
Table 8: Randomization Strata*
| Variable | Summary Statistics | All Subjects (N=324) | PerClot (N=161) | Arista (N=163) |
| --- | --- | --- | --- | --- |
| Bleeding Severity Score | | | | |
| 1 | % (n/N) | 57.1% (185/324) | 57.8% (93/161) | 56.4% (92/163) |
| 2 | % (n/N) | 42.9% (139/324) | 42.2% (68/161) | 43.6% (71/163) |
| Therapeutic Area | | | | |
| Cardiac | % (n/N) | 26.5% (86/324) | 26.7% (43/161) | 26.4% (43/163) |
| General | % (n/N) | 47.8% (155/324) | 46.6% (75/161) | 49.1% (80/163) |
| Urology | % (n/N) | 25.6% (83/324) | 26.7% (43/161) | 24.5% (40/163) |
*Comparability of the PerClot and control groups were assessed using Fisher's exact tests of proportions for categorical variables
# C. Study Population Demographics and Baseline Parameters
Patients were 59 years old on average, $62.7\%$ male, and $88.9\%$ white. No statistically significant differences were observed between study arms. No statistically significant differences were observed between arms on the height, weight, temperature or systolic blood pressure of enrolled patients (see Table 9).
Malignant tumor was the most common indication for surgery $(33.0\%)$ . Cardiovascular disease $(26.2\%)$ and metastatic tumor $(22.5\%)$ were also common. Together, these three indications accounted for $81.8\%$ of all surgeries with each of the remaining indications accounting for $< 10\%$ . Surgery at a site that had been previously operated on was fairly common at $18.5\%$ . No statistically significant differences were observed between the PerClot and Arista arms.
Table 9: Physical Exam*
| Variable | Summary Statistics | All Subjects (N=324) | PerClot (N=161) | Arista (N=163) |
| --- | --- | --- | --- | --- |
| Height (cm) | Mean ± SD (n) | 173.0 ± 9.88 (324) | 173.3 ± 10.09 (161) | 172.7 ± 9.69 (163) |
| | Median (Range) | 172.7 (140, 208) | 172.7 (140, 196) | 172.7 (147, 208) |
PMA 210036: Summary of Safety and Effectiveness Data
{30}
| Variable | Summary Statistics | All Subjects (N=324) | PerClot (N=161) | Arista (N=163) |
| --- | --- | --- | --- | --- |
| Weight (kg) | Mean ± SD (n)
Median (Range) | 86.5 ± 20.15 (324)
84.5 (39, 150) | 86.0 ± 19.90 (161)
83.5 (46, 135) | 87.0 ± 20.44 (163)
84.8 (39, 150) |
| Temperature (°C) | Mean ± SD (n)
Median (Range) | 36.6 ± 0.40 (319)
36.6 (35, 38) | 36.6 ± 0.42 (160)
36.6 (35, 38) | 36.6 ± 0.38 (159)
36.6 (35, 37) |
| Systolic (mmHg) | Mean ± SD (n)
Median (Range) | 133.3 ± 19.78 (323)
131.0 (86, 196) | 135.3 ± 18.48 (160)
133.0 (94, 186) | 131.4 ± 20.86 (163)
130.0 (86, 196) |
| Diastolic (mmHg) | Mean ± SD (n)
Median (Range) | 76.3 ± 11.92 (323)
76.0 (42, 120) | 77.7 ± 11.98 (160)
78.0 (42, 120) | 74.9 ± 11.75 (163)
75.0 (47, 113) |
*Comparability of the PerClot and control groups were assessed using t-tests of means or Wilcoxon tests for continuous factors and Fisher's exact tests of proportions for categorical variables
Use of blood modifiers, insulin, blood sugar lowering medications, or reversal drugs was fairly common in the enrolled population (66.4%). Current smoking was uncommon (10.2%), while former smoking was more common (41.4%). Approximately 1 in 5 patients (20.7%) had Type II diabetes and a similar percentage (17.9%) had undergone chemotherapy within the past 6 months.
## D. Safety and Efficacy Results
### 1. Effectiveness Results
#### Primary Endpoint
In the AT population, the observed PerClot hemostasis rate at 7 minutes post-application is 90.6% versus 92.0% in Arista, a difference of -1.4% for the three therapeutic areas pooled. However, poolability of data across therapeutic areas and across sites was assessed and the pooling of data across the three therapeutic areas was not supported by the data due to variability in treatment difference between PerClot and Arista across therapeutic areas. Therefore, primary efficacy data was not pooled across the three therapeutic areas for hypothesis testing. Hemostasis rate at 7 minutes post-application by therapeutic area and treatment group is presented in Table 10.
The difference between PerClot and Arista varies across therapeutic area with a range from -9.3% (urology) to 16.7% (cardiac). Comparisons between PerClot and Arista
PMA 210036: Summary of Safety and Effectiveness Data
{31}
within each individual therapeutic area were not statistically powered for formal noninferiority assessments.
PMA 210036: Summary of Safety and Effectiveness Data
Page 32
Table 10: Hemostasis at 7 minutes: By therapeutic area – AT population
| Therapeutic Area | PerClot % (n/N) | Arista % (n/N) |
| --- | --- | --- |
| General | 93.3% (70/75) | 100.0% (80/80) |
| Cardiac | 85.7% (36/42) | 69.0% (29/42) |
| Urology | 90.7% (39/43) | 100.0% (40/40) |
Secondary Endpoint
Hemostasis at 5 minutes post-Application
In the AT population, PerClot demonstrated a comparable hemostasis rate at 5 minutes post-application versus Arista. Hemostasis rate at 5 minutes post-application by therapeutic area and treatment group is presented in Table 11.
Table 11: Hemostasis at 5 minutes: By therapeutic area – AT population
| Therapeutic Area | PerClot %(n/N) | Arista %(n/N) |
| --- | --- | --- |
| General | 92.0% (69/75) | 90.0% (72/80) |
| Cardiac | 83.3% (35/42) | 64.3% (27/42) |
| Urology | 93.0% (40/43) | 97.5% (39/40) |
The difference between PerClot and Arista by therapeutic area varies with a range from - 4.5% (urology) to 19% (cardiac). Comparisons between PerClot and Arista within each individual therapeutic area were not statistically powered.
Additional Assessment of Hemostasis
Maintained Hemostasis at 12 minutes post-Application
An additional hemostasis assessment was performed at 12 minutes to check whether hemostasis was maintained at 5 minutes after the primary endpoint assessment time of 7
{32}
minutes. This hemostasis assessment was defined as a supplemental safety measure to assess re-bleeding with results between PerClot and Arista more consistent across therapeutic areas. Hemostasis maintenance results by therapeutic area are provided at 12 minutes (Table 12). The 7 minute time point stands out as having the most variable results across therapeutic areas. Achieving hemostasis at 7 minutes that is not maintained through 12 minutes is of limited clinical value. Accounting for the 6 cases (2 general surgery and 4 urology) in which Arista did not maintain hemostasis from 7 to 12 minutes and the 1 case (1 general surgery) in which PerClot did not maintain hemostasis, the hemostasis results at 12 minutes demonstrate that the results of the PerClot across the three therapeutic areas are reasonably consistent, with poor performance of Arista in cardiac epicardial cases being the primary reason for variation in performance across therapeutic areas.
Table 12: Hemostasis Maintenance at 12 minutes: By therapeutic area – AT population
| Therapeutic Area | PerClot %(n/N) | Arista %(n/N) |
| --- | --- | --- |
| General | 92.0% (69/75) | 97.5% (78/80) |
| Cardiac | 85.7% (36/42) | 69.0% (29/42) |
| Urology | 90.7% (39/43) | 90.0% (36/40) |
These findings suggest that the consistently good performance of PerClot across therapeutic areas at 5, 7, and 12 minutes provides evidence that PerClot affords therapeutic benefit that is anticipated to be comparable to Arista in repeated future use of the product.
2. Safety Results
The safety profile of both adjunctive hemostatic devices was comparable with no type of adverse event or severity of adverse event having a statistically significant higher rate in the PerClot arm compared to Arista. Furthermore, all supplemental safety measures further supported the conclusion of comparable safety and effectiveness with no differences in total operative time, use of alternate means to achieve hemostasis, total estimated blood loss, total units of blood transferred, need for re-operation, or total days hospitalized. In the study, there were 30 re-operations that occurred in 27 subjects. Re-operations for bleeding were assessed to determine if the bleeding occurred at the hemostatic agent application site.
There was a total of 6 deaths reported during study follow-up through 6 weeks. Of these deaths, 5 occurred in the Arista arm and 1 in the PerClot arm. At 24 months for the oncology patients, the Kaplan-Meier estimate of survival was 76% in the PerClot arm versus 74% in the Arista arm. Device deficiencies were rare in the clinical study occurring in <1% of distributed PerClot devices.
PMA 210036: Summary of Safety and Effectiveness Data
Page 33
{33}
There were no unanticipated adverse device effects (UADEs) reported in the study. Non-serious AEs were common with 66.7% of subjects having 1 or more non-serious adverse event. Serious AEs were less common with 27.2% of subjects having 1 or more serious adverse event. Given that all patients in the study were undergoing a surgical procedure, a substantial number of events and a wide variety of types of events were expected to occur (see Table 13 and Table 13.1).
There was also a total of 34 device-related adverse events consisting of 17 SAEs and 17 non-SAEs. These events occurred in 21 separate subjects (21/324=6.5%). All 34 events were categorized as "Possibly related" with choices of "Possibly related", "Probably related" and "Definitely related" for the certainty with which an AE was related to the product.
For the PerClot cohort, a total of 21 adverse events (AEs) in 12 different subjects "Possibly related" to PerClot were reported in the study (see Table 14). Anemia, thromboembolic event, and pleural effusion were the only three types of AEs that had more than one occurrence. The number of device-related AEs were also comparable between the study arms. The observed rates of device-related adverse events by arm were 3.7% PerClot and 3.1% Arista for non-serious adverse events related to the device and 3.7% PerClot and 2.5% Arista for serious adverse events related to the device (Table 13). Therefore, a total of 7.5% (12/161) PerClot and 5.5% (9/163) Arista subjects experienced an adverse event, either serious or non-serious, that was "Possibly related" to the hemostatic device during the study.
Table 13: Overall Adverse Event Classification Summary*
| | # Events (# Subjects, % Subjects) | | |
| --- | --- | --- | --- |
| Adverse Event Category | All Subjects (N=324) | PerClot (N=161) | Arista (N=163) |
| Non-Serious AE | 909
216, 66.7%) | 457
114, 70.8%) | 452
102, 62.6%) |
| Non-Serious Device Related AE | 17
11, 3.4%) | 10
6, 3.7%) | 7
5, 3.1%) |
| Serious AE | 152
88, 27.2%) | 69
44, 27.3%) | 83
44, 27.0%) |
| Serious Device Related AE | 17
10, 3.1%) | 11
6, 3.7%) | 6
4, 2.5%) |
*Comparison of safety between treatment arms was based on Fisher's exact tests for the proportion of subjects with serious device-related adverse events, unanticipated adverse device effects, and frequent adverse events (≥ 5% overall incidence
PMA 210036: Summary of Safety and Effectiveness Data
Page 34
{34}
Table 13.1. Reported Adverse Event Summary (rates>5%).
| | # Events (#Subjects, % Subjects) | | |
| --- | --- | --- | --- |
| | All Events | | |
| Adverse Event Category | All Subjects (N=324) | PerClot (N=161) | Arista (N=163) |
| Hypophosphatemia | 46 (45, 13.9%) | 22 (22, 13.7%) | 24 (23, 14.1%) |
| Pleural Effusion | 38 (38, 11.7%) | 20 (20, 12.4%) | 18 (18, 11.0%) |
| Anemia | 32 (32, 9.9%) | 19 (19, 11.8%) | 13 (13, 8.0%) |
| Hypotension | 32 (31, 9.6%) | 12 (12, 7.5%) | 20 (19, 11.7%) |
| Constipation | 29 (28, 8.6%) | 16 (16, 9.9%) | 13 (12, 7.4%) |
| Abdominal Pain | 28 (28, 8.6%) | 14 (14, 8.7%) | 14 (14, 8.6%) |
| Nausea | 27 (27, 8.3%) | 13 (13, 8.1%) | 14 (14, 8.6%) |
| Hyperglycemia | 26 (26, 8.0%) | 14 (14, 8.7%) | 12 (12, 7.4%) |
| Atelectasis | 25 (24, 7.4%) | 15 (14, 8.7%) | 10 (10, 6.1%) |
| Hypokalemia | 24 (23, 7.1%) | 12 (11, 6.8%) | 12 (12, 7.4%) |
| Acute Kidney Injury | 21 (21, 6.5%) | 11 (11, 6.8%) | 10 (10, 6.1%) |
| Dyspnea | 21 (21, 6.5%) | 12 (12, 7.5%) | 9 (9, 5.5%) |
| Leukocytosis | 21 (21, 6.5%) | 9 (9, 5.6%) | 12 (12, 7.4%) |
| Atrial Fibrillation | 19 (19, 5.9%) | 10 (10, 6.2%) | 9 (9, 5.5%) |
| Fever | 17 (17, 5.2%) | 12 (12, 7.5%) | 5 (5, 3.1%) |
| Total | 909 (216, 66.7%) | 457 (114, 70.8%) | 452 (102, 62.6%) |
PMA 210036: Summary of Safety and Effectiveness Data
{35}
Table 14: Adverse Event Categories with Device-related Adverse Events – PerClot (Treatment)
| | All AEs in Treatment Arm Subjects (N=161)
# Events (# Subjects, % Subjects) | | | |
| --- | --- | --- | --- | --- |
| Adverse Event Category | Not related | Possibly related | Probably related | Definitely related |
| Anemia | 16 (16, 9.9%) | 3 (3, 1.9%) | 0 | 0 |
| Thromboembolic Event | 0 | 3 (3, 1.9%) | 0 | 0 |
| Pleural Effusion | 21 (20, 12.4%) | 2 (2, 1.2%) | 0 | 0 |
| Hyperglycemia | 13 (13, 8.1%) | 1 (1, 0.6%) | 0 | 0 |
| Hypoxia | 6 (6, 3.7%) | 1 (1, 0.6%) | 0 | 0 |
| INR Increased | 6 (6, 3.7%) | 1 (1, 0.6%) | 0 | 0 |
| Respiratory Failure | 2 (2, 1.2%) | 1 (1, 0.6%) | 0 | 0 |
| Sepsis | 2 (2, 1.2%) | 1 (1, 0.6%) | 0 | 0 |
| Activated Partial Thromboplastin Time Increased | 0 | 1 (1, 0.6%) | 0 | 0 |
| Distributive Shock | 0 | 1 (1, 0.6%) | 0 | 0 |
| Gastric Perforation | 0 | 1 (1, 0.6%) | 0 | 0 |
| Hematoma Infection | 0 | 1 (1, 0.6%) | 0 | 0 |
| Implant Site Fluid Collection | 0 | 1 (1, 0.6%) | 0 | 0 |
| Pericardial Tamponade | 0 | 1 (1, 0.6%) | 0 | 0 |
| Perihepatic Fluid Collection | 0 | 1 (1, 0.6%) | 0 | 0 |
| Pneumonia | 0 | 1 (1, 0.6%) | 0 | 0 |
In addition to the deaths reported as part of safety reporting during study follow-up, there were additional deaths reported at the 24-month survival assessment for oncology subjects. As stated previously, there were 184 of 324 randomized subjects that were oncology subjects. The Kaplan-Meier estimate of survival at 24 months was 76% (20 deaths) in the PerClot arm versus 74% (25 deaths) in the Arista arm (see Figure 4). The difference in the survival rate by arm was small, with Arista patients having similar survival compared to PerClot patients. The study was not specifically designed or powered to compare survival.
PMA 210036: Summary of Safety and Effectiveness Data
Page 36
{36}
Supplementary safety information is provided in Table 15. A total of 322 of the 324 enrolled subjects were assessed for supplementary safety. Overall, all supplemental safety endpoints were comparable across PerClot and Arista study arms, providing evidence of comparable safety performance. None of the re-operations were due to bleeding at the hemostatic agent application site for either product. PerClot performance with vascular grafts was studied in the CLOT Trial $(n = 6)$ , but this subset was not statistically powered. All patients met the primary and secondary endpoints, and no device related AEs were observed in this subset of subjects.

Figure 4: Kaplan-Meier Plot of Survival in Oncology Subjects Through 24 months
Table 15: Supplemental Safety Summary*
| Variable | Summary Statistics | All Subjects (N=324) | PerClot (N=161) | Arista (N=163) |
| --- | --- | --- | --- | --- |
| Total Operative Time (minutes) | Mean ± SD (n) Median (Range) | 257.2 ± 124.61 (322) 233.5 (59, 733) | 254.2 ± 126.40 (160) 234.5 (61, 692) | 260.2 ± 123.13 (162) 233.5 (59, 733) |
| Hemostasis at 5 minutes (per investigator) | % (n/N) | 87.6% (282/322) | 90.0% (144/160) | 85.2% (138/162) |
| Hemostasis at 7 minutes (per investigator) | % (n/N) | 91.3% (294/322) | 90.6% (145/160) | 92.0% (149/162) |
| Hemostasis Maintained at additional 5 minutes | % (n/N) | 89.1% (287/322) | 90.0% (144/160) | 88.3% (143/162) |
PMA 210036: Summary of Safety and Effectiveness Data
{37}
| Variable | Summary Statistics | All Subjects (N=324) | PerClot (N=161) | Arista (N=163) |
| --- | --- | --- | --- | --- |
| Used Alternate Means to Achieve Hemostasis | % (n/N | 10.2% (33/322) | 10.0% (16/160) | 10.5% (17/162) |
| Stitches | % (n/N | 1.9% (6/322) | 1.9% (3/160) | 1.9% (3/162) |
| Manual pressure applied | % (n/N | 1.6% (5/322) | 1.9% (3/160) | 1.2% (2/162) |
| Other hemostatic device | % (n/N | 4.7% (15/322) | 3.8% (6/160) | 5.6% (9/162) |
| Other | % (n/N | 4.7% (15/322) | 6.3% (10/160) | 3.1% (5/162) |
| Total Estimated Blood Loss (mL) | Mean ± SD (n)
Median (Range) | 477.6 ± 546.76 (322)
300 (0, 4500) | 471.4 ± 476.91 (160)
350 (0, 2200) | 483.6 ± 609.38 (162)
300 (0, 4500) |
| Total Units Blood Transfused | Mean ± SD (n)
Median (Range) | 0.4 ± 0.94 (322)
0 (0, 4) | 0.4 ± 0.93 (160)
0 (0, 4) | 0.4 ± 0.95 (162)
0 (0, 4) |
| Reoperation | % (n/N | 8.4% (27/322) | 10.6% (17/160) | 6.2% (10/162) |
| Total Hospitalization Time (days) | Mean ± SD (n)
Median (Range) | 6.8 ± 5.96 (318)
5 (1, 66) | 7.0 ± 6.58 (158)
5.5 (2, 66) | 6.6 ± 5.29 (160)
5 (1, 37) |
*Comparability of the PerClot and control group subjects were assessed using t-tests of means or Wilcoxon tests for continuous factors and Fisher's exact tests of proportions for categorical variables
## Study Summary
PerClot Absorbable Hemostatic Powder demonstrated effectiveness as an adjunctive hemostatic device. For the primary endpoint, PerClot demonstrated comparable performance versus Arista. Comparability was also demonstrated for the secondary endpoint with a complete hemostasis rate at 5 minutes. Hemostasis for both groups was maintained at a high rate through the 12-minute assessment for both arms. PerClot and Arista showed comparable safety and efficacy for both mild (Severity Score 1) and moderate (Severity Score 2) bleeds, for men and women, and all races.
The safety profile of both adjunctive hemostatic devices was comparable with no type of adverse event or severity of adverse event having a statistically significant higher rate in the PerClot arm compared to Arista. Furthermore, all supplemental safety measures further supported the conclusion of comparable safety and effectiveness with no differences in total operative time, use of alternate means to achieve hemostasis, total
PMA 210036: Summary of Safety and Effectiveness Data
{38}
estimated blood loss, total units of blood transferred, need for re-operation, or total days hospitalized.
The results of the CLOT Trial of PerClot Absorbable Hemostatic Powder demonstrated safety and efficacy as an adjunctive hemostatic device by demonstrating comparable performance versus Arista.
## 3. Pediatric Extrapolation
In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population.
## E. Financial Disclosure
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 110 investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data.
## XI. SUMMARY OF SUPPLEMENTAL CLINICAL INFORMATION
### Additional European and Chinese Studies
Additional clinical information for PerClot is sourced from outside the United States (OUS) clinical studies (Table 16).
Table 16: Total Available Clinical Data for PerClot
| Source | Number of Studies / Reference Articles | Total PerClot |
| --- | --- | --- |
| European Studies | 4 | 119 |
| China Studies | 3 | 148 |
| Published Literature | 9 | 255 |
| Total | | 522 |
Four European (n=119 PerClot) and Three Chinese (n=148 PerClot) clinical trials are available totaling 267 patients. These studies cover a wide range of therapeutic areas as shown in Table 17.
PMA 210036: Summary of Safety and Effectiveness Data
Page 39
{39}
Table 17: Additional Clinical Studies Using PerClot
| Study | Surgical Area | # of PerClot |
| --- | --- | --- |
| Studies in Europe | | |
| Roberts D., “A post-market surveillance study evaluating the safety and effectiveness of PerClot to control mild bleeding in subjects undergoing endoscopic sinus surgery.”1 | ENT | N=12 |
| Ponce J. and Scambia G., “A prospective, multi-center, randomized, safety and effectiveness study of PerClot compared to usual care when used during gynecological procedures.”2 | Gynecological surgery | N=44 (20 laparoscopic) |
| Mair H., et al., “Use of PerClot, a plant-based polysaccharide hemostat, for bleeding control of the sternum in high risk patients.”3 | Coronary surgery requiring median sternotomy | N=21 |
| Busch F. “Safety and effectiveness of PerClot an absorbable adhesive forming hemostatic microporous polysaccharide in spinal surgery.”4 | Head and Neck surgery (Spinal) | N=42 patients N=127 bleeding events |
| Studies in China | | |
| Southwest Hospital Tangdu Hospital “Clinical Trial of Absorbable Polysaccharides Haemostatic.”5 | Surgical trauma | N=44 |
| Third Military Medical University Tangdu Hospital “Absorbable Polysaccharide Hemostatic Material”6 | General surgery | N=89 |
| Fourth Army Medical University Lu J. and Wang Q. “Clinical evaluation of hemostatic performance in hepatic surgery.”7 | Splenectomy for portal hypertension | N=15 |
| Total | | N=267 |
# Additional Published Scientific Literature
A literature search was conducted to identify any published clinical data related to PerClot. In total, 9 articles were identified (8 clinical trials and 1 systematic review). This real-world clinical evidence (RWE) included multiple therapeutic areas such as ENT surgery, cardiovascular / vascular surgery, head and neck surgery, plastic surgery, and orthopedic surgery. In total, 255 PerClot subjects are included in published literature (Table 18).
Table 18: Therapeutic Areas and Subjects in Published Literature
| Therapeutic Area | Articles (n) | Subjects (n) |
| --- | --- | --- |
| ENT surgery (Pagella8, Van Ahnen9) | 2 | 12+30 (42) |
| Head and Neck Surgery (Saragozzi10, Van Ahnen11) | 1 | 12+30 (42) |
| Gynecology (Saragozzi12, Van Ahnen12) | 1 | 12+30 (42) |
| Osteology (Saragozzi13, Van Ahnen13) | 1 | 12+30 (42) |
| Osteology (Saragozzi14, Van Ahnen14) | 1 | 12+30 (42) |
PMA 210036: Summary of Safety and Effectiveness Data
{40}
PMA 210036: Summary of Safety and Effectiveness Data
| Cardiovascular / vascular surgery (Janczak^{10}, Tscholl^{11}) | 2 | 26+51 (77) |
| --- | --- | --- |
| Head and neck surgery (Rao^{12}) | 1 | 57 |
| Laparoscopic abdominal surgery (Duran^{13}, Puchkov^{14}) | 2 | 31+16 (47) |
| Plastic surgery (Malik^{15}) | 1 | Not reported |
| Orthopedic surgery (Aktas^{16}) | 1 | 32 |
| Total | 9 | 255 |
## Supplemental Clinical Information - Laparoscopic
Several supplemental clinical studies have been conducted OUS to support the assessment of PerClot when administered laparoscopically. As shown in Table 19, PerClot was used successfully in 67 procedures from these three OUS studies, further supporting the laparoscopic clinical outcomes.
Table 19: PerClot Laparoscopic OUS Post-market Usage
| Reference | Surgical Area | # of Laparoscopic PerClot Uses |
| --- | --- | --- |
| Ponce J. and Scambia G., “A prospective, multi-center, randomized, safety and effectiveness study of PerClot compared to usual care when used during gynecological procedures.”^{2} | Gynecological surgery | N=20 |
| Duran, et al., “Comparative study of different means of hemostasia from the liver bed in the course of cholecystectomies complicated laparoscopic.”^{13} | Laparoscopic abdominal surgery (Cholecystectomy) | N=31 |
| Puchkov, et al. “Methods of treatment of the stump of the adrenal gland.”^{14} | Laparoscopic abdominal surgery (Adrenalectomy) | N=16 |
| TOTAL | | N=67 |
Published literature shows that PerClot is used in a wide range of real-world applications. No unexpected adverse events were reported in the clinical studies. PerClot was effective in achieving hemostasis with a positive safety profile compared to traditional closure methods. This evidence supports the continued use of PerClot at a physician's discretion in any procedure where demands for hemostasis fits the device parameters of PerClot and no known contra-indications are known.
Page 41
{41}
PMA 210036: Summary of Safety and Effectiveness Data
Page 42
# XII. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION
In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the General and Plastic Surgery Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
# XIII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES
## A. Efficacy Conclusions
The PerClot Polysaccharide Hemostatic System (PerClot System) consisting of the PerClot Absorbable Hemostatic Powder (PerClot), a containment and delivery system inclusive of all applicator tips, packaging and labeling has demonstrated a comparable hemostasis rate at 5 and 7 minutes when compared to the Arista control arm. Further, the hemostasis maintenance rate in both the treatment (PerClot) and control (Arista) arms at 12 minutes was similar to the hemostasis rate observed at 7 minutes.
## B. Safety Conclusions
The risks identified and assessed for the PerClot System are based on nonclinical laboratory and animal studies, as well as data collected in clinical studies conducted to demonstrate the safety and efficacy of the device in support of PMA approval. The safety profile of the PerClot Absorbable Hemostatic Powder was found to be similar to that of the control Arista with regard to the rate and types of Adverse Events and Serious Adverse Events. In addition, there were no unan…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.