ADVIA Centaur Anti-HBc Total (HBcT2) and Atellica IM Anti-HBc Total (HBcT2)

P210019 · Siemens Healthcare Diagnostics, Inc. · SEI · Jul 27, 2022 · Microbiology

Device Facts

Record IDP210019
Device NameADVIA Centaur Anti-HBc Total (HBcT2) and Atellica IM Anti-HBc Total (HBcT2)
ApplicantSiemens Healthcare Diagnostics, Inc.
Product CodeSEI · Microbiology
Decision DateJul 27, 2022
DecisionAPRL
Regulation21 CFR 866.3173
Device ClassClass 2

Indications for Use

ADVIA Centaur® HBc Total 2 (HBcT2) assay The ADVIA Centaur HBc Total 2 (HBcT2) assay is an in vitro diagnostic immunoassay for use in the qualitative determination of total antibodies to the core antigen of the hepatitis B virus (HBV) in human adult serum and plasma (EDTA, lithium heparin, and sodium heparin) using the ADVIA Centaur XP and ADVIA Centaur XPT systems. This assay can be used as an aid in the diagnosis of adults with acute or chronic hepatitis B virus (HBV) infection, and in the determination of the clinical status of HBV-infected individuals in conjunction with other HBV serological markers, for the laboratory diagnosis of HBV disease associated with HBV infection. This assay can also be used as an aid in the differential diagnosis in individuals displaying signs and symptoms of hepatitis in whom etiology is unknown. This assay is not intended for screening donors of blood or blood products or human cells, tissues, and cellular and tissue-based products (HCT/Ps). ADVIA Centaur® HBc Total 2 Quality Control (HBcT2 QC) The ADVIA Centaur® HBc Total 2 (HBcT2) Quality Control material is for in vitro diagnostic use for monitoring the performance of the ADVIA Centaur HBc Total 2 (HBcT2) assay using the ADVIA Centaur systems. Atellica IM® HBc Total 2 (HBcT2) assay The Atellica IM HBc Total 2 (HBcT2) assay is an in vitro diagnostic immunoassay for use in the qualitative determination of total antibodies to the core antigen of the hepatitis B virus (HBV) in human adult serum and plasma (EDTA, lithium heparin, and sodium heparin) using the Atellica IM Analyzer. This assay can be used as an aid in the diagnosis of adults with acute or chronic hepatitis B virus (HBV) infection and in the determination of the clinical status of HBV-infected individuals in conjunction with other HBV serological markers, for the laboratory diagnosis of HBV disease associated with HBV infection. This assay can also be used as an aid in the differential diagnosis in individuals displaying signs and symptoms of hepatitis in whom etiology is unknown. This assay is not intended for screening donors of blood or blood products or human cells, tissues, and cellular and tissue-based products (HCT/Ps). Atellica IM® HBc Total 2 Quality Control (HBcT2 QC) The Atellica IM HBc Total 2 (HBcT2) Quality Control material is for in vitro diagnostic use for monitoring the performance of the Atellica IM HBc Total 2 (HBcT2) assay using the Atellica IM systems.

Device Story

The ADVIA Centaur/Atellica Anti-HBcT2 assay is an in vitro diagnostic immunoassay used in clinical laboratories to detect total antibodies (IgG and IgM) against the hepatitis B core antigen (HBcAg). The device utilizes a 2-wash antigen sandwich immunoassay format. Patient serum or plasma is incubated with a solid phase consisting of streptavidin-coated microparticles and biotinylated recombinant HBc antigen, followed by a lite reagent containing acridinium ester-labeled recombinant HBc antigen and anti-human IgG Fab monoclonal antibody. The system measures chemiluminescence (Relative Light Units, RLUs), which is directly proportional to the amount of anti-HBc antibodies present. Results are reported as Index values, where ≥ 1.0 is reactive. The assay is performed on the ADVIA Centaur XP/XPT or Atellica IM analyzers. Healthcare providers use these results in conjunction with other HBV serological markers to diagnose HBV infection, determine disease status, and guide clinical management, including antiviral therapy decisions. The assay benefits patients by providing accurate diagnostic information to facilitate appropriate monitoring and treatment, potentially reducing disease burden and transmission.

Clinical Evidence

Clinical study evaluated 1,595 prospective subjects (at-risk or symptomatic) across three US sites. Performance compared to FDA-approved reference anti-HBcT assay. Overall PPA was 97.7-98.2% and NPA was 98.1-98.6% across platforms. Subgroup analysis included pregnant women (N=193) and various HBV disease classifications (acute, chronic, recovery, etc.). Precision and reproducibility were established per CLSI EP05-A3. Analytical sensitivity determined using WHO 1st International Standard (0.28 IU/mL at cutoff). No cross-reactivity observed with common viral antibodies or endogenous substances.

Technological Characteristics

Antigen sandwich chemiluminescent immunoassay. Solid phase: streptavidin-coated microparticles with biotinylated recombinant HBc antigen. Lite reagent: acridinium ester-labeled recombinant HBc antigen and anti-human IgG Fab monoclonal antibody. Energy source: chemiluminescent reaction (Acid/Base reagents). Platforms: ADVIA Centaur XP/XPT and Atellica IM analyzers. Connectivity: System-integrated. Calibration: Two-point, lot-specific Master Curve.

Indications for Use

Indicated for qualitative determination of total antibodies to hepatitis B virus (HBV) core antigen in human adult serum and plasma. Used as an aid in diagnosis of acute or chronic HBV infection, determination of clinical status of HBV-infected individuals, and differential diagnosis of hepatitis of unknown etiology. Not for blood/tissue donor screening.

Regulatory Classification

Identification

A hepatitis B virus (HBV) antibody assay is identified as an in vitro diagnostic device intended for prescription use in the detection of antibodies to HBV in human serum, plasma, or other matrices, and as a device that aids in the diagnosis of HBV infection in persons with signs and symptoms of hepatitis and in persons at risk for hepatitis B infection. Results from assays may be qualitative or quantitative, such as quantitative anti-HBs. In addition, results from an anti-HBc IgM (IgM antibodies to core antigen) assay indicating the presence of anti-HBc IgM are indicative of recent HBV infection. Anti-HBs (antibodies to surface antigen) assay results may be used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or when vaccination status is unknown. The assay is not intended for screening of blood, plasma, cells, or tissue donors. The assay is intended as an aid in diagnosis in conjunction with clinical findings and other diagnostic procedures.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The labeling required under § 809.10(b) of this chapter must include: (i) A prominent statement that the assay is not intended for the screening of blood, plasma, cells, or tissue donors. (ii) A detailed explanation of the principles of operation and procedures for performing the assay. (iii) A detailed explanation of the interpretation of results. (iv) Limitations, which must be updated to reflect current clinical practice and disease presentation and management. The limitations must include statements that indicate: (A) When appropriate, performance characteristics of the assay have not been established in populations of immunocompromised or immunosuppressed patients or other special populations where assay performance may be affected. (B) Detection of HBV antibodies to a single viral antigen indicates a present or past infection with hepatitis B virus, but does not differentiate between acute, chronic, or resolved infection. (C) The specimen types for which the device has been cleared, and that use of the assay with specimen types other than those specifically cleared for this device may result in inaccurate assay results. (D) Diagnosis of hepatitis B infection should not be established on the basis of a single assay result but should be determined by a licensed healthcare professional in conjunction with the clinical presentation, history, and other diagnostic procedures. (E) A non-reactive assay result may occur early during acute infection, prior to development of a host antibody response to infection, or when analyte levels are below the limit of detection of the assay. (F) Results obtained with this assay may not be used interchangeably with results obtained with a different manufacturer's assay. (v) For devices intended for the quantitative detection of HBV antibodies (anti-HBs), in addition to the special controls listed in paragraphs (b)(1) and (2) of this section, labeling required under § 809.10(b) of this chapter must include: (A) The assay calibrators' traceability to a standardized reference material that FDA has determined is appropriate ( *e.g.,* a recognized consensus standard) and the limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), linearity, and precision to define the analytical measuring interval.(B) Performance results of the analytical sensitivity study testing a standardized reference material that FDA has determined is appropriate ( *e.g.,* a recognized consensus standard).(2) Design verification and validation must include the following: (i) Detailed device description, including all parts that make up the device, ancillary reagents required but not provided, an explanation of the device methodology, and design of the antigen(s) and capture antibody(ies) sequences, rationale for the selected epitope(s), degree of amino acid sequence conservation of the target, and the design and composition of all primary, secondary and subsequent standards used for calibration. (ii) Documentation and characterization ( *e.g.,* supplier, determination of identity, and stability) of all critical reagents (including description of the antigen(s) and capture antibody(ies)), and protocols for maintaining product integrity throughout its labeled shelf life.(iii) Risk analysis and management strategies, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on assay performance. (iv) Final release criteria to be used for manufactured assay lots with appropriate evidence that lots released at the extremes of the specifications will meet the identified analytical and clinical performance characteristics as well as stability. (v) Stability studies for reagents must include documentation of an assessment of real-time stability for multiple reagent lots using the indicated specimen types and must use acceptance criteria that ensure that analytical and clinical performance characteristics are met when stability is assigned based on the extremes of the acceptance range. (vi) All stability protocols, including acceptance criteria. (vii) When applicable, analytical sensitivity of the assay that is the same or better than that of other cleared or approved assays. (viii) Analytical performance studies and results for determining the limit of blank (LoB), limit of detection (LoD), cutoff, precision (reproducibility), including lot-to-lot and/or instrument-to-instrument precision, interference, cross reactivity, carryover, hook effect, seroconversion panel testing, matrix equivalency, specimen stability, reagent stability, and cross-genotype antibody detection sensitivity, when appropriate. (ix) For devices intended for the detection of antibodies for which a standardized reference material (that FDA has determined is appropriate) is available, the analytical sensitivity study and results testing the standardized reference material. Detailed documentation of that study and its results must be provided, including the study protocol, study report, testing results, and all statistical analyses. (x) For devices with associated software or instrumentation, documentation must include a detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review. (xi) Detailed documentation of clinical performance testing from a clinical study with an appropriate number of HBV reactive and non-reactive samples in applicable risk categories and conducted in the appropriate settings by the intended users. Performance must be analyzed relative to an FDA cleared or approved HBV antibody assay or a comparator that FDA has determined is appropriate. Additional relevant patient groups must be validated as appropriate. The samples must include prospective (sequential) samples for each identified specimen type and, as appropriate, additional characterized clinical samples. Samples must be sourced from geographically diverse areas. (3) For any HBV antibody assay intended for quantitative detection of anti-HBV antibodies, the following special controls, in addition to those special controls listed in paragraphs (b)(1) and (2) of this section, also apply: (i) Detailed documentation of the metrological calibration traceability hierarchy to a standardized reference material that FDA has determined is appropriate. (ii) Detailed documentation of the following analytical performance studies conducted, as appropriate to the technology, specimen types tested, and intended use of the device, including upper and lower limits of quantitation (UloQ and LloQ, respectively), linearity using clinical samples, and an accuracy study using the recognized international standard material.

Submission Summary (Full Text)

{0} SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED) I. GENERAL INFORMATION Device Generic Name: Antibodies to Hepatitis B Core Antigen Assay Device Trade Name: ADVIA Centaur® HBc Total 2 (HBcT2) ADVIA Centaur® HBc Total 2 Quality Control (HBcT2 QC) Atellica IM® HBc Total 2 (HBcT2) Atellica IM® HBc Total 2 Quality Control (HBcT2 QC) Device Procode: LOM Applicant’s Name and Address: Siemens Healthcare Diagnostics Inc 511 Benedict Ave Tarrytown, NY 10591 Date of Panel Recommendation: None Premarket Approval Application (PMA) Number: P210019 Date of FDA Notice of Approval: July 27, 2022 II. INDICATIONS FOR USE ADVIA Centaur® HBc Total 2 (HBcT2) assay The ADVIA Centaur HBc Total 2 (HBcT2) assay is an in vitro diagnostic immunoassay for use in the qualitative determination of total antibodies to the core antigen of the hepatitis B virus (HBV) in human adult serum and plasma (EDTA, lithium heparin, and sodium heparin) using the ADVIA Centaur XP and ADVIA Centaur XPT systems. This assay can be used as an aid in the diagnosis of adults with acute or chronic hepatitis B virus (HBV) infection, and in the determination of the clinical status of HBV-infected individuals in conjunction with other HBV serological markers, for the laboratory diagnosis of HBV disease associated with HBV infection. This assay can also be used as an aid in the differential diagnosis in individuals displaying signs and symptoms of hepatitis in whom etiology is unknown. This assay is not intended for screening donors of blood or blood products or human cells, tissues, and cellular and tissue-based products (HCT/Ps). ADVIA Centaur® HBc Total 2 Quality Control (HBcT2 QC) The ADVIA Centaur® HBc Total 2 (HBcT2) Quality Control material is for in vitro diagnostic use for monitoring the performance of the ADVIA Centaur HBc Total 2 (HBcT2) assay using the ADVIA Centaur systems. PMA P210019: FDA Summary of Safety and Effectiveness Data {1} The performance of the ADVIA Centaur HBcT2 Quality Control material has not been with any other anti-HBc Total assay. ## Atellica IM® HBc Total 2 (HBcT2) assay The Atellica IM HBc Total 2 (HBcT2) assay is an in vitro diagnostic immunoassay for use in the qualitative determination of total antibodies to the core antigen of the hepatitis B virus (HBV) in human adult serum and plasma (EDTA, lithium heparin, and sodium heparin) using the Atellica IM Analyzer. This assay can be used as an aid in the diagnosis of adults with acute or chronic hepatitis B virus (HBV) infection and in the determination of the clinical status of HBV-infected individuals in conjunction with other HBV serological markers, for the laboratory diagnosis of HBV disease associated with HBV infection. This assay can also be used as an aid in the differential diagnosis in individuals displaying signs and symptoms of hepatitis in whom etiology is unknown. This assay is not intended for screening donors of blood or blood products or human cells, tissues, and cellular and tissue-based products (HCT/Ps). ## Atellica IM® HBc Total 2 Quality Control (HBcT2 QC) The Atellica IM HBc Total 2 (HBcT2) Quality Control material is for in vitro diagnostic use for monitoring the performance of the Atellica IM HBc Total 2 (HBcT2) assay using the Atellica IM systems. The performance of the Atellica IM HBcT2 Quality Control material has not been with any other anti-HBc Total assay. ## III. CONTRAINDICATIONS There are no known contraindications. ## IV. WARNINGS AND PRECAUTIONS The warnings and precautions can be found in the device labeling. ## V. DEVICE DESCRIPTION ### Assay Principle and Format The ADVIA Centaur/Atellica Anti-HBcT2 assay is a 2-wash antigen sandwich immunoassay in which antigens are bridged by antibody present in the patient sample. The Solid Phase contains a preformed complex of streptavidin-coated microparticles and biotinylated recombinant HBc antigen and is used to capture anti-HBc in the patient sample. The Lite Reagent contains recombinant HBc antigen labeled with acridinium ester and anti-human IgG Fab monoclonal antibody labeled with acridinium ester and is used to detect anti-HBc in the sample. The Ancillary Reagent, Solid Phase, and Ancillary Well Reagent are added to the sample, PMA P210019: FDA Summary of Safety and Effectiveness Data 2 of 27 {2} followed by Lite Reagent. Antibody-antigen complexes will form if anti-HBc antibodies (IgM and IgG) are present in the sample. A direct relationship exists between the amount of anti-HBc antibodies present in the patient sample and the amount of relative light units (RLUs) detected by the system. A result of reactive or nonreactive is determined according to the Index Value established with the calibrators. Assay procedure: The ADVIA Centaur system automatically performs the following actions: 1. Dispenses 50 µL of sample into a cuvette. 2. Dispenses 100 µL of Ancillary Reagent and incubates the mixture for 6 minutes at 37°C. 3. Dispenses 100 µL of Ancillary Well Reagent and 125 µL of Solid Phase, and incubates the mixture for 18 minutes at 37°C. 4. Washes the cuvette with ADVIA Centaur Wash 1. 5. Resuspends with 250 µL of ADVIA Centaur Wash 1 and incubates the mixture for 6 minutes at 37°C. 6. Dispenses 100 µL of Lite Reagent and incubates the mixture for 18 minutes at 37°C. 7. Washes the cuvette with ADVIA Centaur Wash 1. 8. Dispenses 300 µL of ADVIA Centaur Acid Reagent and 300 µL of ADVIA Centaur Base Reagent to initiate the chemiluminescent reaction. 9. Reports results. The Atellica IM automatically performs the following actions: 1. Dispenses 50 µL of sample into a cuvette. 2. Dispenses 100 µL of Ancillary Reagent into a cuvette, then incubates for 6 minutes at 37°C. 3. Dispenses 100 µL of Ancillary Well Reagent and 125 µL of Solid Phase, then incubates for 18 minutes at 37°C. 4. Performs a wash sequence using Atellica IM Wash. 5. Resuspends the particles in 250 µL of Atellica IM Wash. 6. Dispenses 100 µL of Lite Reagent, then incubates for 18 minutes at 37°C. 7. Performs a wash sequence using Atellica IM Wash. 8. Dispenses 300 µL each of Atellica IM Acid and Atellica IM Base to initiate the chemiluminescent reaction. 9. Reports results. Calibration The ADVIA Centaur/Atellica Anti-HBcT2 assay utilizes two-point calibration (Low Calibrator, High Calibrator). The assay utilizes a factory-set Master Curve. The Master Curve values are contained on the Master Curve card provided with each kit. The Master Curve and calibration are lot specific. The barcode reader or keyboard is used to enter the Master Curve values on the system. The two calibrators in the kit are run when the lot is first used or after expiration of the calibration interval. If the calibration run is valid as determined by prearranged parameters, the values are stored and used to "normalize" test values to the Master Curve. PMA P210019: FDA Summary of Safety and Effectiveness Data 3 of 27 {3} The Index value of the sample or control is read off the Master Curve. Individuals whose samples read at or above an Index of 1.0 are considered to be reactive for HBcT2. ## Controls The ADVIA Centaur HBcT2 QC and Atellica IM HBcT2 QC set contains Negative control (2 vials with 7 mL) and Positive control (2 vials with 7 mL). The performance of the ADVIA Centaur and Atellica IM HBcT2 assay is monitored by the use of ADVIA Centaur HBcT2 or Atellica IM HBcT2 QC Quality Controls at least once during each day when samples are analyzed or after a successful calibration. ## Interpretation of Results The system reports HBcT2 assay results in Index Values and as Nonreactive or Reactive: - Nonreactive: < 1.0 Index. These samples are considered negative. - Reactive: ≥ 1.0 Index. These samples are considered positive. ## VI. ALTERNATIVE PRACTICES AND PROCEDURES There are several other alternatives for the determination of HBV infection and its disease stage. Detection of anti-HBcT in patients who may be infected with the hepatitis B virus may also be accomplished with any commercially available FDA approved serological tests. This assay is one of several hepatitis marker assays that are often used together and in conjunction with clinical assessment and other laboratory test results in the diagnosis of the HBV infection. ## VII. MARKETING HISTORY ADVIA Centaur HBc Total 2 (HBcT2) assay and ADVIA Centaur HBc Total 2 Quality Control (HBcT2 QC) are marketed globally in several countries. The device has not been withdrawn to date from the market in any country for reasons relating to safety and effectiveness of the device. | Austria | Belgium | Switzerland | Czech Republic | | --- | --- | --- | --- | | Germany | Denmark | Estonia | Spain | | Finland | France | United Kingdom | Greece | | Croatia | Hungary | Ireland | Iceland | | Italy | Liechtenstein | Lithuania | Luxembourg | | Latvia | North Macedonia | Malta | Netherlands | | Norway | Poland | Romania | Serbia | | Sweden | Slovenia | Slovakia | Afghanistan | | Ukraine | United Arab Emirates | Cyprus | Hong Kong | | India | Iraq | Turkey | Saudi Arabia | | Singapore | Malaysia | Canada | Chile | | Chile | Peru | South Africa | Zimbabwe | | Madagascar | Bahrain | Pakistan | | PMA P210019: FDA Summary of Safety and Effectiveness Data 4 of 27 {4} PMA P210019: FDA Summary of Safety and Effectiveness Data 5 of 27 # VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH Below is a list of the potential adverse effects associated with the use of the device. Failure of the product to perform as indicated, or human error in use of the product may lead to a false result. Repeatedly erroneous false positive or false negative anti-HBc results could lead to inappropriate initiation or cessation of antiviral therapy. The risk of incorrect test results is inherent with all in vitro diagnostic products. Therefore, the above potential risks are not unusual in the laboratory setting and should be evaluated in conjunction with other clinical indicators. When used according to the instructions in the package insert, there are no known direct adverse effects of this device on the health of the user. Standard good laboratory practices are considered sufficient to minimize risks to the end user. # IX. SUMMARY OF NONCLINICAL STUDIES ## Cut-off Determination The ADVIA Centaur/Atellica Anti-HBcT2 assay is a qualitative assay to detect human IgG and IgM antibodies against the core antigen of the Hepatitis B virus. The ADVIA Centaur/Atellica Anti-HBcT2 assay results are reported as Nonreactive or Reactive with 'Index' units. An Index value of $\geq 1.00$ indicates the presence of anti-HBc antibodies in a sample, whereas, an Index value $< 1.00$ indicates that the sample is negative for anti-HBc antibodies. Internal master curve standards were made with increasing concentrations of anti-HBc antibodies in the feasibility phase of development. Each of the standard levels was assigned an Index value such that the resulting dose – response curve (RLU vs. Index) was used to calculate the concentration (Index units) of anti-HBc antibodies in samples. The assignment of the master curve standards was done such that a preliminary assay cut off was established at 1.00 Index while maintaining a separation between anti-HBc antibody positive and negative samples. This was done by evaluating known positive and negative samples, as determined by the HBcT2 index value, during the development of this assay via a qualitative method comparison to the HBcT2 assay which was used as the initial predicate device for research purposes. During the development and verification phase of this assay development project, population studies were conducted with the reagent lots to further optimize the cutoff. The HBcT2 interpretation of samples that were discordant based on vendor Certificates of Analysis (CoA) was compared to an FDA-approved Anti-HBc II assay. Based on this data, the assignment of the master curve standards and calibrators were linearly adjusted and validated. {5} # Precision The ADVIA Centaur/Atellica Anti-HBcT2 assay precision in various sample matrices was examined in a 20-day precision protocol using 3 lots of reagent. Two controls and 5 specimens in 2 matrices (serum and EDTA plasma) were used to measure the precision of the assay at different Index levels. The specimens were assayed in duplicate with 2 runs per day for 20 days ( $N = 80$ for each sample). The results in the following table are from one lot. Calculations for within-run, between-run, between-day, and total precision were performed according to Clinical and Laboratory Standards Institute (CLSI) Document EP05-A3. Table 1: ADVIA Centaur Precision | Specimen Type | Na | Mean (Index | Repeatability | | Within-Laboratory Precision | | | --- | --- | --- | --- | --- | --- | --- | | | | | SDb (Index | CVc % | SD (Index | CV % | | Plasma 1 | 80 | 0.39 | 0.02 | N/Ad | 0.09 | N/A | | Plasma 2 | 80 | 0.78 | 0.03 | N/A | 0.06 | N/A | | Plasma 3 | 80 | 1.56 | 0.07 | 4.6 | 0.10 | 6.7 | | Plasma 4 | 80 | 2.38 | 0.11 | 4.5 | 0.15 | 6.3 | | Plasma 5 | 80 | 6.71 | 0.33 | 4.9 | 0.64 | 9.6 | | Serum 1 | 80 | 0.30 | 0.02 | N/A | 0.09 | N/A | | Serum 2 | 80 | 0.63 | 0.04 | N/A | 0.06 | N/A | | Serum 3 | 80 | 1.53 | 0.09 | 5.8 | 0.10 | 6.4 | | Serum 4 | 80 | 2.30 | 0.09 | 4.0 | 0.16 | 7.1 | | Serum 5 | 80 | 6.13 | 0.34 | 5.6 | 0.57 | 9.3 | | Control 1 (negative) | 80 | 0.31 | 0.02 | N/A | 0.08 | N/A | | Control 2 positive | 80 | 3.49 | 0.18 | 5.2 | 0.25 | 7.2 | aNumber of measurements bStandard deviation Coefficient of variation ${}^{\mathrm{d}}$ Not applicable. Results remained nonreactive throughout the study Table 2: Atellica IM Precision | Specimen Type | Na | Mean (Index | Repeatability | | Within-Laboratory Precision | | | --- | --- | --- | --- | --- | --- | --- | | | | | SDb (Index | CVc % | SD (Index | CV % | | Plasma 1 | 75d | 0.13 | 0.03 | N/Ae | 0.03 | N/A | | Plasma 2 | 80 | 0.62 | 0.04 | N/A | 0.04 | N/A | | Plasma 3 | 80 | 1.38 | 0.06 | 4.3 | 0.06 | 4.3 | PMA P210019: FDA Summary of Safety and Effectiveness Data {6} | Plasma 4 | 80 | 2.25 | 0.06 | 2.7 | 0.06 | 2.7 | | --- | --- | --- | --- | --- | --- | --- | | Plasma 5 | | | | | | 3.7 | | Serum 1 | 77 | 0.10 | 0.03 | N/A | 0.05 | N/A | | Serum 2 | 80 | 0.46 | 0.04 | N/A | 0.07 | N/A | | Serum 3 | 80 | 1.41 | 0.07 | 4.8 | 0.12 | 8.3 | | Serum 4 | 80 | 2.17 | 0.10 | 4.4 | 0.16 | 7.2 | | Serum 5 | 80 | 5.73 | 0.27 | 4.7 | 0.39 | 6.8 | | Control 1 (negative) | 80 | 0.24 | 0.02 | N/A | 0.04 | N/A | | Control 2 positive | 80 | 3.52 | 0.10 | 2.9 | 0.14 | 3.9 | aNumber of measurements bStandard deviation Coefficient of variation Samples recovering below 0.70 Index are not included in analysis Not applicable. Results remained nonreactive throughout the study # System Reproducibility The system reproducibility of the ADVIA Centaur/Atellica Anti-HBcT2 assay was evaluated on ADVIA Centaur XP, XPT and the Atellica IM systems at two external US sites and one internal sponsor site within the US. A six-member panel, QC and calibrators were assayed in triplicate for five days, two runs per day, at the three sites. The end of the first run on a testing day was separated from the start of the second run by approximately two hours. Each site ran three reagent lots with their respective calibrators and controls. The system reproducibility was determined in accordance with CLSI Document EP05-A3 The results obtained with the ADVIA Centaur XP, XPT, and Atellica IM are summarized below in the following tables. PMA P210019: FDA Summary of Safety and Effectiveness Data {7} Table 3: ADVIA Centaur XP Reproducibility Across Sites and Reagent Lots | | | | Repeatability | | Run | | Day | | Lot | | Within-Lab | | Site | | Reproducibility | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Sample | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | High Calibrator | 270 | 2.01 | 0.08 | 3.8% | 0.05 | 2.5% | 0.00 | 0.0% | 0.06 | 2.9% | 0.09 | 4.6% | 0.15 | 7.6% | 0.19 | 9.4% | | Low Calibrator | 270 | 0.61 | 0.03 | 5.6% | 0.02 | 4.0% | 0.01 | 1.5% | 0.00 | 0.0% | 0.04 | 7.0% | 0.06 | 9.6% | 0.07 | 11.9% | | Negative Control | 270 | 0.25 | 0.02 | 6.9% | 0.03 | 12.7% | 0.00 | 0.0% | 0.02 | 8.3% | 0.04 | 14.4% | 0.03 | 13.5% | 0.05 | 21.4% | | Positive Control | 270 | 3.28 | 0.14 | 4.3% | 0.09 | 2.8% | 0.00 | 0.0% | 0.18 | 5.4% | 0.17 | 5.1% | 0.25 | 7.7% | 0.35 | 10.7% | | Serum 1 | 270 | 0.50 | 0.03 | 5.6% | 0.03 | 6.3% | 0.00 | 0.0% | 0.01 | 2.9% | 0.04 | 8.4% | 0.03 | 5.5% | 0.05 | 10.4% | | Serum 2 | 270 | 0.83 | 0.03 | 4.2% | 0.03 | 3.9% | 0.00 | 0.0% | 0.03 | 4.2% | 0.05 | 5.7% | 0.06 | 7.2% | 0.08 | 10.1% | | Serum 3 | 270 | 1.47 | 0.06 | 4.0% | 0.02 | 1.7% | 0.02 | 1.5% | 0.12 | 7.9% | 0.07 | 4.6% | 0.11 | 7.2% | 0.17 | 11.7% | | Serum 4 | 270 | 2.56 | 0.11 | 4.4% | 0.04 | 1.7% | 0.03 | 1.2% | 0.21 | 8.3% | 0.12 | 4.9% | 0.19 | 7.5% | 0.31 | 12.2% | | Serum 5 | 270 | 5.48 | 0.26 | 4.7% | 0.06 | 1.2% | 0.07 | 1.3% | 0.48 | 8.8% | 0.28 | 5.1% | 0.40 | 7.3% | 0.68 | 12.5% | | Serum 6 | 270 | 8.71 | 0.38 | 4.4% | 0.16 | 1.8% | 0.00 | 0.0% | 0.66 | 7.6% | 0.41 | 4.7% | 0.61 | 7.0% | 0.98 | 11.3% | Table 4: ADVIA Centaur XPT Reproducibility Across Sites and Reagent Lots | | | | Repeatability | | Run | | Day | | Lot | | Within-Lab | | Site | | Reproducibility | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Sample | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | High Calibrator | 270 | 1.96 | 0.09 | 4.4% | 0.05 | 2.6% | 0.00 | 0.0% | 0.16 | 8.2% | 0.10 | 5.1% | 0.10 | 5.2% | 0.21 | 11.0% | | Low Calibrator | 270 | 0.57 | 0.03 | 5.4% | 0.03 | 4.8% | 0.00 | 0.0% | 0.05 | 9.4% | 0.04 | 7.3% | 0.04 | 6.5% | 0.08 | 13.5% | | Negative Control | 270 | 0.20 | 0.02 | 11.6% | 0.02 | 8.3% | 0.01 | 4.5% | 0.06 | 29.5% | 0.03 | 14.9% | 0.02 | 12.1% | 0.07 | 35.2% | | Positive Control | 270 | 3.21 | 0.14 | 4.2% | 0.08 | 2.6% | 0.00 | 0.0% | 0.19 | 5.8% | 0.16 | 5.0% | 0.17 | 5.2% | 0.30 | 9.2% | | Serum 1 | 270 | 0.45 | 0.03 | 6.4% | 0.01 | 3.0% | 0.00 | 0.0% | 0.07 | 16.4% | 0.03 | 7.1% | 0.02 | 3.4% | 0.08 | 18.1% | | Serum 2 | 270 | 0.79 | 0.03 | 3.8% | 0.02 | 3.2% | 0.00 | 0.0% | 0.10 | 12.7% | 0.04 | 5.0% | 0.03 | 4.4% | 0.11 | 14.3% | | Serum 3 | 270 | 1.42 | 0.06 | 4.1% | 0.03 | 1.9% | 0.00 | 0.0% | 0.18 | 12.6% | 0.07 | 4.6% | 0.06 | 4.4% | 0.20 | 14.1% | | Serum 4 | 270 | 2.50 | 0.10 | 3.9% | 0.04 | 1.8% | 0.00 | 0.0% | 0.29 | 11.5% | 0.11 | 4.3% | 0.12 | 4.7% | 0.33 | 13.1% | | Serum 5 | 270 | 5.36 | 0.25 | 4.7% | 0.08 | 1.5% | 0.00 | 0.0% | 0.63 | 11.8% | 0.26 | 4.9% | 0.26 | 4.8% | 0.73 | 13.7% | | Serum 6 | 270 | 8.48 | 0.37 | 4.3% | 0.05 | 0.6% | 0.06 | 0.7% | 0.81 | 9.6% | 0.38 | 4.4% | 0.48 | 5.6% | 1.01 | 11.9% | PMA P210019: FDA Summary of Safety and Effectiveness Data {8} Table 5: Atellica IM Reproducibility Across Sites and Reagent Lots | | | | Repeatability | | Run | | Day | | Lot | | Within-Lab | | Site | | Reproducibility | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Sample | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | High Calibrator | 270 | 2.04 | 0.04 | 1.9% | 0.03 | 1.5% | 0.02 | 0.7% | 0.02 | 1.2% | 0.05 | 2.5% | 0.02 | 1.1% | 0.06 | 3.0% | | Low Calibrator | 270 | 0.63 | 0.02 | 2.9% | 0.01 | 2.0% | 0.01 | 2.3% | 0.02 | 3.6% | 0.03 | 4.2% | 0.00 | 0.0% | 0.03 | 5.5% | | Negative Control | 270 | 0.21 | 0.02 | 7.6% | 0.01 | 5.3% | 0.02 | 8.1% | 0.06 | 30.1% | 0.03 | 12.3% | 0.01 | 6.0% | 0.07 | 33.1% | | Positive Control | 270 | 3.29 | 0.07 | 2.1% | 0.05 | 1.5% | 0.03 | 0.8% | 0.23 | 7.0% | 0.09 | 2.7% | 0.04 | 1.1% | 0.25 | 7.6% | | Serum 1 | 270 | 0.41 | 0.02 | 4.6% | 0.02 | 3.7% | 0.00 | 0.0% | 0.09 | 21.2% | 0.02 | 5.8% | 0.00 | 0.0% | 0.09 | 22.0% | | Serum 2 | 270 | 0.76 | 0.02 | 2.7% | 0.01 | 1.5% | 0.01 | 1.7% | 0.07 | 9.0% | 0.03 | 3.4% | 0.01 | 0.9% | 0.07 | 9.7% | | Serum 3 | 270 | 1.35 | 0.04 | 2.6% | 0.01 | 0.6% | 0.02 | 1.2% | 0.13 | 9.5% | 0.04 | 3.0% | 0.00 | 0.0% | 0.13 | 10.0% | | Serum 4 | 270 | 2.40 | 0.06 | 2.4% | 0.03 | 1.4% | 0.02 | 1.0% | 0.20 | 8.4% | 0.07 | 3.0% | 0.00 | 0.2% | 0.21 | 8.9% | | Serum 5 | 270 | 5.09 | 0.17 | 3.4% | 0.12 | 2.4% | 0.06 | 1.2% | 0.41 | 8.1% | 0.22 | 4.3% | 0.00 | 0.0% | 0.47 | 9.1% | | Serum 6 | 270 | 8.38 | 0.21 | 2.6% | 0.00 | 0.0% | 0.09 | 1.1% | 0.43 | 5.2% | 0.23 | 2.8% | 0.07 | 0.8% | 0.50 | 5.9% | PMA P210019: FDA Summary of Safety and Effectiveness Data {9} # Sensitivity Study with HBV WHO First International Standard The analytical sensitivity of the ADVIA Centaur/Atellica Anti-HBcT2 assay was evaluated using the World Health Organization (WHO) First International Standard for anti-Hepatitis B core antigen (anti-HBc). The WHO Standard was reconstituted and diluted according to manufacturer's instructions. A stock solution of the standard was used to prepare a dilution series with negative human serum. The first level in the dilution series (A) was prepared by mixing 1 part of the NIBSC 95/522 50 IU/mL stock solution with 1 part of anti-HBc negative serum. The serial dilutions were then prepared by adding anti-HBc negative serum. Each diluted sample was tested in triplicate with the ADVIA Centaur/Atellica Anti-HBcT2 assay using an ADVIA Centaur XP, XPT, and Atellica IM instrument. The observed dose was calculated using 2-point calibration. The quantitative value of WHO material detected at cutoff (Index = 1.00) was calculated using linear regression analysis. Analytical sensitivity evaluated with three different reagent lots of ADVIA Centaur/Atellica Anti-HBcT2 assay on ADVIA Centaur XP, XPT, and Atellica IM instruments yield only minor differences from the WHO International Standard in all cases, as shown below in Table 6. The WHO 95/522 International Unit per milliliter (IU/mL) concentration at the assay cutoff was determined to be 0.28 IU/mL. Table 6: Value of WHO material (IU/mL) detected at HBcT2 cutoff (Index = 1.00) and Correlation Coefficient | Reagent Lot | Platform | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | ADIVA Centaur XP | | ADVIA Centaur XPT | | Atellica IM | | | | IU/mL | R² | IU/mL | R² | IU/mL | R² | | 136991 | 0.225 | 0.998 | 0.267 | 0.996 | 0.218 | 0.980 | | 136990 | 0.234 | 0.997 | 0.231 | 0.995 | 0.211 | 0.995 | | 136987 | 0.254 | 0.997 | 0.231 | 0.995 | 0.251 | 0.992 | # Seroconversion Sensitivity Commercially available HBV patient seroconversion panels were tested using the ADVIA Centaur/Atellica Anti-HBcT2 assay to determine the seroconversion sensitivity of the assay. The performance of the ADVIA Centaur/Atellica Anti-HBcT2 assay on the seroconversion panels matched the performance of the reference assay. The following results were obtained with the ADVIA Centaur XP instrument: Table 7: Seroconversion Sensitivity | Panel ID | Reference Anti-HBc Total Assay - ReactiveFrom Initial Draw Date | | ADVIA Centaur HBcT2 Assay versusReference Anti-HBc Total Assay | | --- | --- | --- | --- | | | ADVIA Centaur HBcT2 Assay (Days) | Reference (Days) | Difference in Bleed Numbersa | | HBV6278 | | | 0 | | HBV6281 | 41 | 41 | 0 | | | | | 0 | PMA P210019: FDA Summary of Safety and Effectiveness Data {10} PMA P210019: FDA Summary of Safety and Effectiveness Data 11 of 27 | HBV9099 | 74 | 74 | 0 | | --- | --- | --- | --- | | | | | 0 | | SCPHBV1 | 29 | 29 | 0 | | | | | +1 | a The difference in bleed numbers is relative to the reference assay. For example, a "+1" means that the reference assay required 1 additional bleed before reactivity was determined as compared to the time point when the ADVIA Centaur assay confirmed reactive. # Cross-Reactivity The ADVIA Centaur/Atellica Anti-HBcT2 assay was evaluated for potential cross-reactivity with viral antibodies and disease state specimens. The anti-HBcT serostatus of each specimen was verified using an FDA approved reference assay. The following results were obtained using the ADVIA Centaur XP instrument: Table 8: Cross-Reactivity | Substance | Number Tested | Number of Reactive Anti-HBc Total Results | | | --- | --- | --- | --- | | | | ADVIA Centaur HBcT2 Assay | Reference Assay | | Anti-nuclear antibody (ANA) | | | 2 | | Cytomegalovirus (CMV) IgG | 15 | 0 | 0 | | Cytomegalovirus (CMV) IgM | | | 0 | | Epstein-Barr virus (EBV) IgG | 15 | 0 | 0 | | Epstein-Barr virus (EBV) IgM | | | 0 | | Flu vaccine recipient | 15 | 0 | 0 | | Human anti-mouse antibody (HAMA) | | | 2 | | Hepatitis A infection (HAV) | 20 | 4 | 4 | | Hepatitis C infection (HCV) | | | 7 | | Herpes simplex virus (HSV) IgG | 15 | 0 | 0 | | Herpes simplex virus (HSV) IgM | | | 0 | | Human immunodeficiency virus (HIV 1/2) | 15 | 6 | 6 | | Multiparity | | | 1 | | Non-viral liver disease | 15 | 1 | 0 | | Rheumatoid arthritis | | | 1 | | Rubella IgG | 15 | 0 | 0 | | Syphilis IgG | 15 | 3 | 3 | | Systemic lupus erythematosus (SLE) | 20 | 1 | 1 | | Toxoplasma IgG | 21 | 0 | 0 | {11} # Interference with Endongeneous Substances The sensitivity of the ADVIA Centaur/Atellica Anti-HBcT2 assay to interference by endogenous substances was evaluated at the concentrations indicated in Table 9. Samples from 4 matrixes (serum, K2 EDTA, Li heparin, Na heparin) were tested on the ADVIA Centaur XP. Samples were made from pools of patient units with anti-HBc positive pool spiked or without (negatives). Two spiked samples were low positives (target index of 1.5-2.5) and intermediate positives (target index of $>3.0$ ), respectively. None of the evaluated endogenous substances resulted a change of the test result. Table 9: Summary of Endogenous Substances used in the Interference Study | Substance | Substance Test Concentration Units | | --- | --- | | Hemolyzed | 500 mg/dL | | Icteric | 60 mg/dL | | Icteric | 40 mg/dL | | Lipemic | 1000 mg/dL | | Hyperproteinemic | 12.0 g/dL | | Hypoproteinemic | 3.5 g/dL | | Hyper IgG | 60 mg/mL | | Biotin | 3500 ng/mL | | Cholesterol | 500 mg/dL | # Specimen Matrix Equivalency The study was performed on the ADVIA Centaur XP instrument to assess the influence of different matrices on the results of the ADVIA Centaur/Atellica Anti-HBcT2 assay. A total of 50 nonreactive matched sets were collected in five different blood collection tubes (Serum Glass, SST, K2 EDTA, Li Hep, and Na Hep). 10 matched sets were used to compare nonreactive blood collective tubes versus serum glass while 40 matched sets were separated and spiked with 1 unique reactive positive serum sample per matched set to 4 different target levels in order to span the assay Index range of $0.07 - 10.00$ Index. The target ranges were 0.75 Index (High Negative), 1.50 Index (Low Positive), 2.50 Index (Moderate Positive), and $3.00 - 10.00$ Index (High Positive) respectively. Samples were tested on one ADVIA Centaur XP instrument using one reagent lot. Data were analyzed by Deming regression and results for each tube type were compared to the results of serum glass. The results for all matrices were plotted on a XY graph compared to the values obtained from serum samples and weighted Deming regression fit was used to evaluate the variability. The correlation coefficient was calculated using Pearson correlation. Matrix equivalency was determined in accordance with CLSI Document EP09-A3, Measurement Procedure PMA P210019: FDA Summary of Safety and Effectiveness Data {12} Comparison and Bias Estimation Using Patient Samples; Approved Guideline—Third Edition. All samples demonstrated acceptable assay equivalency between matrices throughout the measurement range (Table 10). Table 10: Regression Statistics for ADVIA Centaur HBcT2 Matrix Comparisons | Tubes Types | Slope | Intercept | r | | --- | --- | --- | --- | | SST vs Serum Glass | 0.960 | 0.043 | 98.3% | | K2 EDTA vs Serum Glass | 0.979 | 0.025 | 98.1% | | Li Heparin vs Serum Glass | 1.000 | -0.053 | 96.8% | | Na Heparin vs Serum Glass | 1.042 | -0.093 | 97.1% | # Stability Studies # a. Specimen Stability Studies were performed to determine the stability of patient samples under different storage conditions. Specimens were collected in the following tube types: serum glass tubes, serum separator tubes (SST), K2EDTA plasma tubes, sodium heparin plasma tubes, and lithium heparin plasma tubes. The effect of multiple freeze/thaw cycles, time to centrifugation, and onboard storage on the stability of the samples was also evaluated. A minimum of ten samples were used for each sample handling with samples spanning the assay range including: <1.00 Index (Non-reactive samples) 1.00-3.00 Index (Low Reactive) >3.00 Index Nonreactive anti-HBc samples were spiked with anti-HBc Reactive samples in order to achieve the desired sample distribution. Samples were aliquoted and placed in the appropriate storage/stress conditions. A baseline Index value, t0, was established for each sample. For reactive samples bias $\pm 15\%$ of t0 was determined by trend analysis. Non-reactive samples remained non-reactive. Additionally, samples stored frozen were subjected to 6 freeze-thaw cycles. Bias lower than $\pm 15\%$ versus t0 were observed. The results of the sample handling and storage temperature studies, freeze/thaw, and frozen $(-20^{\circ}\mathrm{C})$ studies support the claims for stability for all samples and matrices (serum or SST, EDTA plasma and Lithium Heparin plasma) (Table 11). Table 11: Summary of the Results for the Sample Stability Study Types | Storage Condition | Claimed Stability | | --- | --- | | Time to Centrifugation | 24 hours | | Primary Tube Storage | 7 Days | | Onboard Storage | 24 hours | | Room Temp Storage (25°C) | 3 Days | PMA P210019: FDA Summary of Safety and Effectiveness Data {13} | Refrigerated Storage (2-8°C) | 7 Days | | --- | --- | | Frozen Storage (-20°C) | 12 Months | | Freeze/Thaw Cycles | 5 | b. Reagent Stability Reagent Real Time Stability Study (Shelf-Life) Real time stability testing was performed on three lots of ADVIA Centaur Anti-HBcT2 on the ADVIA Centaur XP instrument and three lots of Atellica IM HBcT2 on the Atellica IM instrument to determine reagent shelf-life when stored at 2–8°C. Each lot was stored at 2–8°C and was tested at pre-determined checkpoints from T = 0 to 108 weeks (ADVIA Centaur Anti-HBe2) or 124-138 weeks (Atellica Anti-HBe2). The results support the shelf-life claim of 24 months for ADVIA Centaur and Atellica IM HBcT2 assay stored at 2–8°C. Stability of the Calibration Interval The stability of the working calibration curve was evaluated on the ADVIA Centaur XP, XPT and Atellica IM instruments with two lots of ADVIA Centaur HBe2 and two lots of Atellica HBcT2. Reagents were placed on onboard ADVIA Centaur XP and Atellica IM for the duration of the study. Testing was performed with calibrators (Low, High) and controls (Control 1, Control 2) run as samples and patient pools spiked with anti-HBc. All samples were tested in replicates of 4 (ADVIA Centaur) or replicates of 3 (Atellica IM). In addition to testing open onboard reagents, fresh static packs were tested at each timepoint. The results support the calibration interval claim of 21 days on ADVIA Centaur XP and XPT instruments and 42 days on the Atellica IM. Calibrator Open Vial Stability The study was performed with one lot of the ADVIA Centaur Anti-HBcT2 on the ADVIA Centaur XP instrument to establish the open vial stability at 2–8°C of calibrators versus the unopened vials. At each of the scheduled timepoints, a set of calibrators were assayed. Calibration for this study used a stored 2-point curve, based upon the calibration of the zero timepoint. Data was calculated from the 2-point calibration using calibrators held -40°C. After the use of calibrator and controls, vials were returned to refrigerated storage until testing at the next time point. Materials were tested at multiple timepoints and were compared to Time 0. The results support the open vial stability claim of 70 days for ADVIA Centaur Anti-HBe2 Calibrators. A similar study was conducted to determine how long an open vial of the Atellica IM HBcT2 Calibrators can remain on the Sample Handler of the Atellica IM Analyzer and provide acceptable results. The results of that study support the open vial stability claim up to 25 hours on the sample handler. Calibrator On-Board Stability ADVIA Centaur HBcT2 calibrators were tested on ADVIA Centaur XP over the course of 9 hours. The same study was conducted with Atellica HBcT2 calibrators tested on an Atellica IM instrument. At each of the scheduled timepoints, a set of calibrators were assayed. Calibration for this study used a stored 2-point curve, based upon the calibration of the zero PMA P210019: FDA Summary of Safety and Effectiveness Data {14} timepoint. Data was calculated from the 2-point calibration using calibrators held at $-40^{\circ}\mathrm{C}$ . When Calibrators were run as samples, all results met the acceptance criteria up to 9 hours. The results support the on-board stability claim of 9 hours for the ADVIA Centaur Anti-HBcT2 Calibrators on the ADVIA Centaur XP and XPT instruments and Atellica Anti-HBcT2 Calibrators on the Atellica IM instrument. # X. SUMMARY OF PRIMARY CLINICAL STUDIES The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of the ADVIA Centaur/Atellica Anti-HBcT2 assay. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below. # A. Study Design A multisite study was conducted to evaluate the performance of the ADVIA Centaur/Atellica Anti-HBcT2 assay. The study consists of a Qualitative Method Comparison Agreement with a reference assay (FDA-approved Anti-HBcT assay). Subjects were assessed for hepatitis markers using commercially available, FDA-approved reference assays. The serological assessment included the following 6 HBV markers: hepatitis B virus surface antigen (HBsAg), hepatitis B e antigen (HBeAg), IgM antibody to hepatitis B core antigen (anti-HBc IgM), total antibody to hepatitis B virus core antigen (anti-HBc Total), hepatitis B e antibody (anti-HBe), and antibody to hepatitis B virus surface antigen (anti-HBs). Testing was conducted at three US sites with three reagent lots. # B. Study Population Demographics and Baseline Parameters A total of 1595 prospective subjects were enrolled for this study. The study population was considered either at risk for hepatitis B (due to lifestyle, behavior, occupation, or known exposure events) or had signs-and-symptoms of hepatitis B infection. Table 12 summarizes the patient population demographics: Table 12: Subject Demographics | Subject Demographics | | | | | --- | --- | --- | --- | | | | N | (%) | | Characteristic | Population | 1595 | 100.0% | | Category | Dialysis Patient | 105 | 6.6% | | | High Risk Patient | 619 | 38.8% | | | Hepatitis B | 628 | 39.4% | | | Pregnant Patient | 193 | 12.1% | | | Transplant Recipient | 50 | 3.1% | | Gender | Female | 837 | 52.5% | | | Male | 758 | 47.5% | PMA P210019: FDA Summary of Safety and Effectiveness Data {15} | Subject Demographics | | | | | --- | --- | --- | --- | | | | N | (%) | | | Unknown | 0 | 0.0% | | State of Residence | Arizona | 1 | 0.1 | | | California | 436 | 27.2 | | | Florida | 582 | 36.5 | | | Massachusetts | 1 | 0.1 | | | Michigan | 5 | 0.3 | | | Minnesota | 533 | 33.6 | | | North Dakota | 1 | 0.1 | | | Nevada | 1 | 0.1 | | | Texas | 2 | 0.1 | | | Virginia | 1 | 0.1 | | | Wisconsin | 1 | 0.1 | | | Other | 31 | 1.9 | | | Africa | 26 | 1.6 | | Continent of Origin | Asia | 36 | 2.2 | | | Caribbean | 116 | 7.3 | | | Central America | 21 | 1.3 | | | Europe | 7 | 0.4 | | | North America | 1361 | 85.4 | | | South America | 27 | 1.7 | | | South Pacific | 1 | 0.1 | | Ethnicity | Non-Hispanic/Non-Latino | 1345 | 84.3 | | | Hispanic/Latino | 250 | 15.7 | | Race | White | 483 | 30.3 | | | Black or African American | 973 | 61.0 | | | Asian | 39 | 2.4 | | | Native Am/Alaska Native | 54 | 3.4 | | | Native HI/Pacific Islander | 6 | 0.4 | | | Other | 9 | 0.6 | | | multi-race | 23 | 1.4 | | | unknown/declined | 8 | 0.5 | | Age (Years) | Mean | 47.13 | | | | Median | 50.0 | | | | SD | 13.997 | | | | Min | 21.0 | | | | Max | 85.0 | | # Percent Agreement: ADVIA Centaur/Atellica IM Anti-HBe2 assay was tested on all three platforms, ADVIA Centaur XP, XPT and Atellica IM systems and the agreement calculated against an FDA-approved aHBcT reference assay. An overview of the detailed results in the following PMA P210019: FDA Summary of Safety and Effectiveness Data {16} tables of the Percent Positive Agreement (PPA) and Negative Percentage of Agreement (NPA) versus the reference assay on the different instrument systems is below (Tables 13 - 15). Table 13: Agreement Table with ADVIA Centaur XP | ADVIA Centaur XP HBcT2 Assay | Reference Assay | | Total | | --- | --- | --- | --- | | | Reactive | Nonreactive | | | Reactive | 434 | 19 | 453 | | Nonreactive | 9 | 1133 | 1142 | | Total | 443 | 1152 | 1595 | % Positive Agreement = 98.0% (434/443) 95% Confidence Interval = 96.2%–98.9% % Negative Agreement = 98.4% (1133/1152) 95% Confidence Interval = 97.4%–98.9% Table 14: Agreement Table with ADVIA Centaur XPT | ADVIA Centaur XPT HBcT2 Assay | Reference Assay | | Total | | --- | --- | --- | --- | | | Reactive | Nonreactive | | | Reactive | 435 | 22 | 457 | | Nonreactive | 8 | 1130 | 1138 | | Total | 443 | 1152 | 1595 | % Positive Agreement = 98.2% (435/443) 95% Confidence Interval = 96.5%–99.1% % Negative Agreement = 98.1% (1130/1152) 95% Confidence Interval = 97.1%–98.7% Table 15: Agreement Table with Atellica IM | Atellica IM HBcT2 Assay | Reference Assay | | Total | | --- | --- | --- | --- | | | Reactive | Nonreactive | | | Reactive | 433 | 16 | 449 | | Nonreactive | 10 | 1136 | 1146 | | Total | 443 | 1152 | 1595 | % Positive Agreement = 97.7% (433/443) 95% Confidence Interval = 95.9%–98.8% % Negative Agreement = 98.6% (1136/1152) 95% Confidence Interval = 97.8%–99.1% ## HBV Disease Classification: Patients were assessed for hepatitis markers using commercially available, FDA-approved reference assays. Each patient's HBV infection status was determined based on the reactive (+) / nonreactive (-) patterns of six HBV reference serological markers obtained from a single specimen (Table 16): hepatitis B virus surface antigen (HBsAg), hepatitis B virus e antigen (HBeAg), total antibody to hepatitis B virus core antigen (anti-HBc Total), IgM PMA P210019: FDA Summary of Safety and Effectiveness Data {17} antibody to hepatitis B core antigen (anti-HBc IgM), total antibody to HBeAg (anti-HBe), and total antibody to hepatitis B virus surface antigen (anti-HBs). There were 30 unique reference marker patterns observed using the FDA-approved assays (Table 16). Table 16: Interpretation of hepatitis B Serologic Test Results used for Classifications | HBV Classification (n) | HBsAg | HBeAg | Anti-HBc IgM | Anti-HBc Total | Anti-HBe | Anti-HBs (>10 mIU/mL) | | --- | --- | --- | --- | --- | --- | --- | | Acute (1) | + | + | + | + | + | - | | Acute (2) | + | + | + | + | - | - | | Chronic (3) | + | + | - | + | + | - | | Chronic (19) | + | + | - | + | - | - | | Chronic (49) | + | - | - | + | + | - | | Chronic (1) | + | - | - | + | - | + | | Chronic (5) | + | - | - | + | - | - | | Early Recovery (4) | - | - | + | + | + | + | | Early Recovery (2) | - | - | + | + | - | + | | Early Recovery (9) | - | - | - | + | + | - | | Recovery (102) | - | - | - | + | + | + | | Recovery (3) | - | - | - | - | + | + | | Recovered (156) | - | - | - | + | - | + | | Recovered (67) | - | - | - | + | - | - | | HBV Vaccine Response (539) | - | - | - | - | - | + | | Not Previously Infected (599) | - | - | - | - | - | - | | Unclassified (2) | + | - | - | - | - | + | | Unclassified (2) | + | - | - | - | - | - | | Unclassified (5) | - | + | - | - | - | + | | Unclassified (9) | - | + | - | - | - | - | | Unclassified (2) | - | - | + | - | - | - | | Unclassified (2) | - | - | - | - | + | - | | Unclassified (1) | + | + | - | - | - | + | | Unclassified (4) | + | + | - | - | - | - | PMA P210019: FDA Summary of Safety and Effectiveness Data 18 of 27 {18} | HBV Classification (n) | HBsAg | HBeAg | Anti-HBc IgM | Anti-HBc Total | Anti-HBe | Anti-HBs (>10 mIU/mL) | | --- | --- | --- | --- | --- | --- | --- | | Unclassified (1) | - | - | Equivocal | - | - | - | | Unclassified (1) | - | - | Equivocal | + | + | - | | Unclassified (2) | - | - | Equivocal | + | + | + | | Unclassified (1) | + | - | Equivocal | + | + | - | | Unclassified (1) | Conf Invalid | - | - | - | - | - | | Unclassified (1) | - | + | - | + | + | + | | + = reactive - = non-reactive | | | | | | | Samples from patients who fell into the following categories: acute, chronic, early recovery, recovery, immune natural infection (recovery), recovered, HBV vaccine response, not previously infected, and unknown serostatus categories, were included in the study and used for data analysis. Thirty-five samples returned a result of unknown serological status (2.17%). The performance of the device was evaluated with the samples included. # Expected Values All of the samples from the 1595 unique patients including pregnant women were tested with the ADVIA Centaur/Atellica Anti-HBcT2. - 619 patients $(38.4\%)$ were from the population considered at risk for hepatitis B (high risk) due to lifestyle, behavior, occupation, or known exposure events. - 629 patients $(39.0\%)$ were from the signs and symptoms population The ADVIA Centaur XP Anti-HBcT2 results for the prospective population for all sites combined by age group and gender are summarized below (Table 17). Table 17: Expected Values with ADVIA Centaur XP | Age Range (Years) | Gender | Reactive | | Nonreactive | | Total Number Tested | | --- | --- | --- | --- | --- | --- | --- | | | | Na | %b | N | % | | | 21-30 | Male | 3 | 4.2 | 69 | 95.8 | 72 | | | Female | 4 | 2.0 | 193 | 98.0 | 197 | | | Overall | 7 | 2.6 | 262 | 97.4 | 269 | | 31-40 | Male | 13 | 15.5 | 71 | 84.5 | 84 | | | Female | 31 | 16.1 | 161 | 83.9 | 192 | | | Overall | 44 | 15.9 | 232 | 84.1 | 276 | | | Male | 48 | 34.3 | 92 | 65.7 | 140 | PMA P210019: FDA Summary of Safety and Effectiveness Data {19} PMA P210019: FDA Summary of Safety and Effectiveness Data 20 of 27 | 41-50 | Female | 35 | 24.8 | 106 | 75.2 | 141 | | --- | --- | --- | --- | --- | --- | --- | | | Overall | 83 | 29.5 | 198 | 70.5 | 281 | | 51-60 | Male | 115 | 40.9 | 166 | 59.1 | 281 | | | Female | 78 | 40.0 | 117 | 60.0 | 195 | | | Overall | 193 | 40.5 | 283 | 59.5 | 476 | | 61-70 | Male | 72 | 47.4 | 80 | 52.6 | 152 | | | Female | 38 | 39.2 | 59 | 60.8 | 97 | | | Overall | 110 | 44.2 | 139 | 55.8 | 249 | | 71-92 | Male | 11 | 37.9 | 18 | 62.1 | 29 | | | Female | 5 | 33.3 | 10 | 66.7 | 15 | | | Overall | 16 | 36.4 | 28 | 63.6 | 44 | | Total | Male | 262 | 34.6 | 496 | 65.4 | 758 | | | Female | 191 | 22.8 | 646 | 77.2 | 837 | | | Overall | 453 | 28.4 | 1142 | 71.6 | 1595 | ## C. Safety and Effectiveness Results ### Safety Results The safety of this device is related to the efficacy described below as incorrect results may lead to patient mismanagement. ### Effectiveness Results #### Prospective Population The performance of the ADVIA CentaurAtellica Anti-HBcT2 assay was evaluated against the risk groups, the disease classification, and the subpopulations. The performance on the ADVIA Centaur XP, ADVIA Centaur CP, and Atellica IM instruments is presented separately and combined. #### By Risk Group The performance of the ADVIA Centaur/Atellica Anti-HBcT2 assay was evaluated in the signs and symptoms prospective populations on each assay system and compared to the reference assay (Tables 18 - 20). The percent agreement and confidence intervals for the prospective population by risk group is presented below (Tables 21 - 23). 1) Signs and Symptom prospective population: Table 18: ADVIA Centaur XP -Comparison of Results in the Signs and Symptoms Prospective Population | ADVIA Centaur XP HBcT2 Assay | Reference Assay | | Total | | --- | --- | --- | --- | | | Reactive | Nonreactive | | | Reactive | 189 | 12 | 201 | | Nonreactive | 5 | 570 | 575 | {20} | Total | 194 | 582 | 776 | | --- | --- | --- | --- | % Positive Agreement = 97.4% (189/194) 95% Confidence Interval = 94.1%–98.9% % Negative Agreement = 97.9% (571/583) 95% Confidence Interval = 96.4%–98.8% Table 19: ADVIA Centaur XPT -Comparison of Results in the Signs and Symptoms Prospective Population | ADVIA Centaur XPT HBcT2 Assay | Reference Assay | | Total | | --- | --- | --- | --- | | | Reactive | Nonreactive | | | Reactive | 189 | 14 | 203 | | Nonreactive | 5 | 568 | 573 | | Total | 194 | 582 | 776 | % Positive Agreement = 97.4% (189/194) 95% Confidence Interval = 94.1%–98.9% % Negative Agreement = 97.6% (568/582) 95% Confidence Interval = 96.0%–98.6% Table 20: Atellica IM - Comparison of Results in the Signs and Symptoms Prospective Population | Atellica IM HBcT2 Assay | Reference Assay | | Total | | --- | --- | --- | --- | | | Reactive | Nonreactive | | | Reactive | 189 | 11 | 200 | | Nonreactive | 5 | 571 | 576 | | Total | 194 | 582 | 776 | % Positive Agreement = 97.4% (189/194) 95% Confidence Interval = 94.1%–98.9% % Negative Agreement = 98.1% (571/582) 95% Confidence Interval = 96.6%–98.9% 2) High-Risk Prospective Population: Table 21: ADVIA Centaur XP -Comparison of Results in the High-Risk Prospective Population | ADVIA Centaur XP HBcT2 Assay | Reference Assay | | Total | | --- | --- | --- | --- | | | Reactive | Nonreactive | | | Reactive | 245 | 7 | 252 | | Nonreactive | 4 | 563 | 567 | | Total | 249 | 570 | 819 | % Positive Agreement = 98.4% (245/249) 95% Confidence Interval = 95.9%–99.4% % Negative Agreement = 98.8% (563/570) 95% Confidence Interval = 97.5%–99.4% PMA P210019: FDA Summary of Safety and Effectiveness Data {21} Table 22: ADVIA Centaur XPT – Comparison of Results in the High-Risk Prospective Population | ADVIA Centaur XPT HBcT2 Assay | Reference Assay | | Total | | --- | --- | --- | --- | | | Reactive | Nonreactive | | | Reactive | 246 | 8 | 254 | | Nonreactive | 3 | 562 | 565 | | Total | 249 | 570 | 819 | % Positive Agreement = 98.8% (246/249) 95% Confidence Interval = 96.5%-99.6% % Negative Agreement = 98.6% (562/570) 95% Confidence Interval = 97.3%-99.3% Table 23: Atellica IM – Comparison of Results in the High-Risk Prospective Population | Atellica IM HBcT2 Assay | Reference Assay | | Total | | --- | --- | --- | --- | | | Reactive | Nonreactive | | | Reactive | 244 | 5 | 249 | | Nonreactive | 5 | 565 | 570 | | Total | 249 | 570 | 819 | % Positive Agreement = 98.0% (244/249) 95% Confidence Interval = 95.4%-99.1% % Negative Agreement = 99.1% (565/570) 95% Confidence Interval = 98.0%-99.6% a. By HBV Serological Classification A total of 1595 samples including the 35 unclassified serostatus samples were evaluated using the ADVIA Centaur/Atellica Anti-HBcT2 assay on the ADVIA Centaur XP, XPT, and Atellica IM and a reference aHBcT assay for each sample classification (Table 16). The agreement and 95% CIs between the ADVIA Centaur/Atellica IM Anti-HBcT2 assay and a reference aHBcT assay for each HBV classification are presented in the Tables 24 - 26. Table 24: Positive and Negative Percent Agreements Between ADVIA Centaur XP HBcT2 assay and reference aHBcT assay by HBV Classification | ADVIA Centaur XP Percent Agreement with Reference Assay | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | HBV Classification | Positive Percent Agreement | | | Negative Percent Agreement | | | | | (x/N) | % | 95% CI | (x/N) | % | 95% CI | | Acute, N=3 | (3/3) | 100% | 43.9-100% | NA | NA | NA | | Chronic, N=77 | (77/77) | 100% | 95.2-100% | NA | NA | NA | | Early Recovery, N=15 | (15/15) | 100% | 79.6-100% | NA | NA | NA | | Recovered, N=223 | (207/209) | 99.0% | 96.6-99.7% | (6/14) | 42.9% | 21.4-67.4% | | Recovery, N=105 | (101/101) | 100% | 96.3-100% | (3/4) | 75.0% | 30.1-95.4% | | HBV Vaccine Response, N=539 | (20/25) | 80.0% | 60.9-91.1% | (509/514) | 99.0% | 97.7-99.6% | | Not Previously Infected, N=599 | (3/5) | 60.0% | 23.1-88.2% | (591/594) | 99.5% | 98.5-99.8% | | Unclassified, N=34 | (8/8) | 100% | 67.6-100% | (24/26) | 92.3% | 75.9-97.9% | PMA P210019: FDA Summary of Safety and Effectiveness Data {22} Table 25: Positive and Negative Percent Agreements Between ADVIA Centaur XPT HBcT2 assay and reference aHBcT assay by HBV Classification | ADVIA Centaur XPT Percent Agreement with Reference Assay | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | HBV Classification | Positive Percent Agreement | | | Negative Percent Agreement | | | | | (x/N) | % | 95% CI | (x/N) | % | 95% CI | | Acute, N=3 | (3/3) | 100% | 43.9-100.0% | NA | NA | NA | | Chronic, N=77 | (77/77) | 100% | 95.2-100.0% | NA | NA | NA | | Early Recovery, N=15 | (15/15) | 100% | 79.6-100.0% | NA | NA | NA | | Recovered, N=223 | (207/209) | 99.0% | 96.6-99.7% | (6/14) | 42.9% | 21.4-67.4% | | Recovery, N=105 | (101/101) | 100% | 96.3-100.0% | (3/4) | 75.0% | 30.1-95.4% | | HBV Vaccine Response, N=539 | (20/25) | 80.0% | 60.9-91.1% | (509/514) | 99.0% | 97.7-99.6% | | Not Previously Infected, N=599 | (3/5) | 60.0% | 23.1-88.2% | (589/594) | 99.2% | 98.0-99.6% | | Unclassified, N=34 | (8/8) | 100% | 67.6-100.0% | (23/26) | 88.5% | 71.0-96.0% | | Total, N=1595 | (434/443) | 98.0% | 96.2-98.9% | (1133/1152) | 98.4% | 97.4-98.9% | Table 26: Positive and Negative Percent Agreements Between Atellica IM HBcT2 assay and reference aHBcT assay by HBV Classification | Atellica IM Analyzer Percent Agreement with Reference Assay | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | HBV Classification | Positive Percent Agreement | | | Negative Percent Agreement | | | | | (x/N) | % | 95% CI | (x/N) | % | 95% CI | | Acute, N=3 | (3/3) | 100% | 43.9-100.0% | NA | NA | NA | | Chronic, N=77 | (77/77) | 100% | 95.2-100.0% | NA | NA | NA | | Early Recovery, N=15 | (15/15) | 100% | 79.6-100.0% | NA | NA | NA | | Recovered, N=223 | (207/209) | 99.0% | 96.6-99.7% | (7/14) | 50.0% | 26.8-73.2% | | Recovery, N=105 | (101/101) | 100% | 96.3-100.0% | (3/4) | 75.0% | 30.1-95.4% | | HBV Vaccine Response, N=539 | (19/25) | 76.0% | 56.6-88.5% | (511/514) | 99.4% | 98.3-99.8% | | Not Previously Infected, N=599 | (3/5) | 60.0% | 23.1-88.2% | (591/594) | 99.5% | 98.5-99.8% | | Unclassified, N=34 | (8/8) | 100% | 67.6-100% | (24/26) | 92.3% | 75.9-97.9% | | Total, N=1595 | (433/443) | 97.7% | 95.9-98.8% | (1136/1152) | 98.6% | 97.8-99.1% | ## b. Pregnant Population Serum samples from United States were included in the study (N = 193). Samples were tested from pregnant women with either signs and symptoms of hepatitis B or with risk factors for HBV infection, who were in the first (62/193, 32.1%), second (61/193, 31.6.0%), or third trimester (70/193, 36.3%) of pregnancy. Results of the testing (reactive and nonreactive) were compared using the ADVIA Centaur HBcT2 assay and the reference HBcT assay for the prenatal population in their first, second, and third trimester, for all testing sites: Table 27: Percent Agreement and Confidence Intervals: Prenatal Populations | Trimester | Positive Agreement | | Negative Agreement | | | --- | --- | --- | --- | --- | | | % (x/n) | 95% CI | % (x/n) | 95% CI | | First | 100% (2/2) | 34.2%-100% | 100% (60/60) | 94.0%-100% | | Second | 100% (6/6) | 61.0%-100% | 98.2% (54/55) | 90.4%-99.7% | | Third | 100% (2/2) | 34.2%-100% | 100% (68/68) | 94.7%-100% | | Total | 100% (10/10) | 72.2%-100% | 99.5% (182/183) | 97.0%-99.9% | PMA P210019: FDA Summary of Safety and Effectiveness Data {23} # Pediatric Extrapolation In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population. ## D. Financial Disclosure The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 5 investigators of which 1 was a full-time or part-time employee of the sponsor and none had disclosable financial interests/arrangements as defined in 21 CFR 54.2(a), (b), (c) and (f) and described below: - Compensation to the investigator for conducting the study where the value could be influenced by the outcome of the study: 0 - Significant payment of other sorts: 0 - Proprietary interest in the product tested held by the investigator: 0 - Significant equity interest held by investigator in sponsor of covered study: 0 The applicant has adequately disclosed the financial interest/arrangements with clinical investigators. Statistical analyses were conducted by FDA to determine whether the financial interests/arrangements had any impact on the clinical study outcome. The information provided does not raise any questions about the reliability of the data ## XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Microbiology Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel. ## XII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES ### A. Effectiveness Conclusions The effectiveness of the ADVIA Centaur/Atellica Anti-HBcT2 assay has been demonstrated by the sensitivity and specificity which has been comparable with the current commercially available FDA-approved anti-HBcT assays among all populations tested. The results from both the nonclinical and clinical studies indicate that the ADVIA Centaur/Atellica Anti-HBcT2 assay is safe and effective for the in vitro qualitative detection of total antibodies to the hepatitis B core antigen (Anti-HBc) in serum and plasma (EDTA, lithium heparin, and sodium heparin) collected from adults. PMA P210019: FDA Summary of Safety and Effectiveness Data {24} B. Safety Conclusions The risk of the device is based on data collected in the non-clinical and clinical studies conducted to support PMA approval as described above. Based on the results from both studies, the ADVIA Centaur/Atellica Anti-HBcT2 assay, when used according to the provided directions and in conjunction with all relevant clinical and laboratory findings, should be safe to use and poses minimal risk to the patient due to false test results. C. Benefit-Risk Determination The benefits of the assay are as an aid in the diagnosis of individuals with acute or chronic hepatitis B virus (HBV) infection, and in the determination of the clinical status of HBV-infected individuals in conjunction with other HBV serological markers, for the laboratory diagnosis of HBV disease associated with HBV infection, and as an aid in the differential diagnosis in individuals displaying signs and symptoms of hepatitis in whom etiology is unknown. Test results can facilitate initiation of appropriate monitoring, antiviral medications, and improved patient knowledge regarding the Hepatitis B infection. Treatment for appropriate patients can mitigate the sequelae of hepatitis B infection and may result in improved morbidity and mortality in these patients. Additionally, diagnosis and appropriate treatment can potentially decrease transmission and disease burden in the general population as well as in populations at high risk for hepatitis B infection. Accurate diagnosis of HBV infection also leads clinicians to evaluate and subsequently treat patients for human immunodeficiency virus (HIV) and hepatitis C virus (HCV) if indicated as these viruses share common risk factors and modes of transmission with HBV, and patients are often coinfected. The risks associated with the device, when used as intended, are those related to the risk of false test results, failure to correctly interpret the test results and failure to correctly operate the instrument. Risks of false positive tests includes improper patient management, including treatment for hepatitis B with antiviral medication. Antiviral medical has risks including toxicity and more rarely allergic reactions. Over time, viral resistance in patients who are co-infected but undiagnosed with other viruses using the same antiviral medication, such as HIV, can lead to viral resistance, however the chance of an undiagnosed co-infection in a patient tested for hepatitis B is exceedingly unlikely. These risks are somewhat mitigated by the fact that this test is generally sent as part of a panel, and incongruous test results in a hepatitis panel would lead a clinician to retest the patient before starting treatment. Risks of false negative tests include potentially missing the opportunity to treat a patient who has hepatitis B infection and whose clinical picture warrants antiviral treatment. A clinician may falsely believe that a patient is not acutely or chronically infected, but rather is currently susceptible or immune to the infection. False negative results may lead a clinician to vaccinate an infected patient. This risk is somewhat PMA P210019: FDA Summary of Safety and Effectiveness Data 25 of 27 {25} mitigated by the fact that this test is usually ordered as part of a panel of hepatitis B tests, and incongruous test results in a hepatitis panel would lead a clinician to retest the patient. 1. Patient Perspective Patient perspectives considered during the review included: This submission either did not include specific information on patient perspectives for this device. In conclusion, given the available information above, the data support that for the qualitative determination of total antibodies to the core antigen of the hepatitis B virus (HBV) in human adult serum and plasma, the probable benefits outweigh the probable risks. D. Overall Conclusions The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the indications for use. The data from the nonclinical studies demonstrated acceptable analytical sensitivity, precision, and analytical specificity of the ADVIA Centaur/Atellica Anti-HBcT2 assay when used according to the instructions for use as stated in the labeling, the warnings, and precautions, and limitations sections of the labeling. The clinical studies have shown that the ADVIA Centaur/Atellica Anti-HBcT2 assay, when compared to the FDA approved comparator, has a similar ability to detect the presence of anti-HBcT antibodies in specimens from individuals with chronic hepatitis B, or those recovered from HBV infection. The assay has also demonstrated that it has no cross-reactivity with viral antibodies or other cross-reactants in the specimens from individuals with medical conditions unrelated to the HBV infection. The probable clinical benefits outweigh the potential risks for the proposed assay considering the performance of the device in the clinical trial and the low risk and associated risk mitigations in clinical practice. The proposed assay labelling will facilitate accurate assay implementation and interpretation of results. The assay may provide substantial benefits to patients as an accurate and sensitive aid in determining HBV seroconversion in conjunction with other diagnostic information. XIII. CDRH DECISION CDRH issued an approval order on July 27, 2022. The applicant's manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820). PMA P210019: FDA Summary of Safety and Effectiveness Data 26 of 27 {26} PMA P210019: FDA Summary of Safety and Effectiveness Data 27 of 27 # XIV. APPROVAL SPECIFICATIONS Directions for use: See device labeling. Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling. Post-approval Requirements and Restrictions: See approval order.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...