Guardant360 CDx
P200010S010 · Guardant Health, Inc. · PQP · Jan 27, 2023 · Pathology
Device Facts
| Record ID | P200010S010 |
| Device Name | Guardant360 CDx |
| Applicant | Guardant Health, Inc. |
| Product Code | PQP · Pathology |
| Decision Date | Jan 27, 2023 |
| Decision | APPR |
| Device Class | Class 3 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|
| P200010S010 · Jan 27, 2023 | Guardant360 CDx | Guardant Health, Inc. | Remnant plasma specimens from the EMERALD (NCT03778931) clinical study; Guardant Health clinical laboratory biobank | Retrospective analysis of remnant clinical trial samples was used to establish analytical accuracy, concordance, and stability of the Guardant360 CDx assay for the detection of ESR1 missense mutations. | Analytical accuracy; Concordance; Stability study; Remnant clinical samples |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|
| Accuracy Study (using EMERALD samples); Retrospective concordance study | Breast cancer patients enrolled in the EMERALD (NCT03778931) clinical study; Sample Size: 254 samples; Number of Sites: Multicenter (EMERALD study sites) | Externally validated mass spectrometry-based comparator method | Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) |
| Whole Blood, Plasma, and cfDNA Stability Studies; Retrospective stability analysis; Follow-up/Duration: Up to 7 days (whole blood); 45-46 days (plasma/cfDNA) | Patients with metastatic breast cancer (mBC); Sample Size: Varies by study (e.g., 44 plasma samples, 56 cfDNA samples) | Reference condition (Time zero/fresh processing) | Variant call concordance, sample-level molecule recovery |
Indications for Use
Guardant360® CDx is a qualitative next generation sequencing-based in vitro diagnostic device that uses targeted high throughput hybridization-based capture technology for detection of single nucleotide variants (SNVs), insertions and deletions (indels) in 55 genes, copy number amplifications (CNAs) in two (2) genes, and fusions in four (4) genes. Guardant360 CDx utilizes circulating cell-free DNA (cfDNA) from plasma of peripheral whole blood collected in Streck Cell-Free DNA Blood Collection Tubes (BCTs). The test is intended to be used as a companion diagnostic to identify patients who may benefit from treatment with the therapies listed in Table 1 in accordance with the approved therapeutic product labeling. A negative result from a plasma specimen does not assure that the patient's tumor is negative for genomic findings. Patients who are negative for the biomarkers listed in Table 1 should be reflexed to tissue biopsy testing for Table 1 biomarkers using an FDA-approved tumor tissue test, if feasible. Additionally, the test is intended to provide tumor mutation profiling to be used by qualified health care professionals in accordance with professional guidelines in oncology for cancer patients with any solid malignant neoplasms. The test is for use with patients previously diagnosed with cancer and in conjunction with other laboratory and clinical findings. Genomic findings other than those listed in Table 1 are not prescriptive or conclusive for labeled use of any specific therapeutic product. Guardant360 CDx is a single-site assay performed at Guardant Health, Inc.
Device Story
Guardant360 CDx is a single-site NGS-based liquid biopsy assay; inputs are 5-30 ng cfDNA from peripheral whole blood collected in Streck BCTs; processes cfDNA via library preparation, hybrid capture enrichment, and Illumina NextSeq 550 sequencing; bioinformatics pipeline reconstructs digitized signals using molecular barcodes to call SNVs, indels, CNAs, and fusions; output is a qualitative report identifying genomic alterations; used by oncologists to guide therapy selection (e.g., ORSERDU, TAGRISSO, RYBREVANT, ENHERTU, LUMAKRAS); benefits patients by identifying actionable mutations from plasma when tissue biopsy is not feasible or as a reflex test; results inform clinical decision-making regarding targeted therapy eligibility.
Clinical Evidence
Clinical validity established using pre-treatment plasma from the EMERALD (NCT03778931) Phase 3 study. Primary cohort included 228 subjects with ESR1 missense mutations (codons 310-547) detected by Guardant360 CDx. Efficacy endpoint was progression-free survival (PFS) by RECIST 1.1. Results showed HR 0.55 (95% CI 0.39-0.77, p=0.0005) for elacestrant vs. SOC in ESR1-mutated patients. Analytical validation included accuracy/concordance (PPA 97.6%, NPA 84.6% vs. mass spectrometry comparator), LoD, precision, and stability studies (whole blood, plasma, cfDNA).
Technological Characteristics
Targeted high-throughput hybridization-based capture NGS; Illumina NextSeq 550 sequencer; 55-gene panel; cfDNA input 5-30 ng; bioinformatics pipeline for variant calling; single-site laboratory assay; Streck BCTs for collection; reagents include biotinylated RNA probes; QC includes Variant Control and NTC; software-based analysis of SNVs, indels, CNAs, and fusions.
Indications for Use
Indicated for patients with non-small cell lung cancer (NSCLC) or breast cancer to identify eligibility for specific FDA-approved therapies based on detection of SNVs, indels, CNAs, and fusions in cfDNA. Also indicated for tumor mutation profiling in patients with any solid malignant neoplasm. No contraindications.
Regulatory Classification
Identification
A next generation sequencing (NGS) oncology panel is a device used for the qualitative detection of germline or somatic variants in one or more cancer-related genes. The device is intended to be used on DNA or RNA isolated from human clinical specimens.
Reference Devices
- cobas® EGFR Mutation Test v2
- FoundationOne® Liquid CDx
- QIAGEN therascreen® KRAS RGQ PCR Kit (P110027/S012)
- OncomineDx Target test (P160045/S027, P160045/S035)
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
## I. GENERAL INFORMATION
| Device Generic Name: | Next Generation Sequencing Oncology Panel, Somatic or Germline Variant Detection System |
| --- | --- |
| Device Trade Name: | Guardant360^{®} CDx |
| Device Procode: | PQP |
| Applicant’s Name and Address: | Guardant Health, Inc. 505 Penobscot Drive Redwood City, CA 94306 USA |
| Date(s) of Panel Recommendation: | None |
| Premarket Approval Application (PMA) Number: | P200010/S010 |
| Date of FDA Notice of Approval: | January 27, 2023 |
The original PMA (P200010) for Guardant360$^{®}$ CDx was approved on August 7, 2020 for the detection of genetic alterations in circulating cell-free DNA (cfDNA) from plasma of peripheral whole blood from patients who may benefit from one of the FDA-approved therapies for non-small cell lung cancer (NSCLC). Subsequently, additional PMA supplements were approved for expanding the indications for use of Guardant360$^{®}$ CDx for detecting *EGFR* exon 20 insertions, *KRAS* G12C mutations and *ERBB2/HER2* activating mutations (SNVs and exon 20 insertions) in NSCLC patients. The SSEDs to support the previously approved indications are available on the CDRH website.
The current panel-track supplement was submitted to expand the intended use and indications for use of Guardant360$^{®}$ CDx to include a companion diagnostic indication for the detection of *ESR1* missense mutations between codons 310 and 547 in breast cancer patients who may benefit from treatment with ORSERDU$^{TM}$ (elacestrant).
## II. INDICATIONS FOR USE
Guardant360$^{®}$ CDx is a qualitative next generation sequencing-based *in vitro* diagnostic device that uses targeted high throughput hybridization-based capture technology for detection of single nucleotide variants (SNVs), insertions and deletions (indels) in 55 genes, copy number amplifications (CNAs) in two (2) genes, and fusions in four (4) genes. Guardant360 CDx utilizes circulating cell-free DNA (cfDNA) from plasma of peripheral whole blood collected in Streck Cell-Free DNA Blood Collection Tubes (BCTs). The test is intended to be used as a companion diagnostic to identify patients
---PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
1 of 33
{1}
who may benefit from treatment with the therapies listed in Table 1 in accordance with the approved therapeutic product labeling.
Table 1. Companion Diagnostic Indications
| Indication | Biomarker | Therapy |
| --- | --- | --- |
| Non-small cell lung cancer (NSCLC) | EGFR exon 19 deletions, L858R, and T790M* | TAGRISSO® (osimertinib) |
| | EGFR exon 20 insertions | RYBREVANT® (amivantamab-vmjw) |
| | ERBB2/HER2 activating mutations (SNVs and exon 20 insertions) | ENHERTU® (fam-trastuzumab deruxtecan-nxki) |
| | KRAS G12C | LUMAKRAS™ (sotorasib) |
| Breast cancer | ESR1 missense mutations between codons 310 and 547 | ORSERDU™ (elacestrant) |
A negative result from a plasma specimen does not assure that the patient's tumor is negative for genomic findings. Patients who are negative for the biomarkers listed in Table 1 should be reflexed to tissue biopsy testing for Table 1 biomarkers using an FDA-approved tumor tissue test, if feasible.
*The efficacy of TAGRISSO® (osimertinib) has not been established in the EGFR T790M plasma-positive, tissue-negative or unknown population and clinical data for T790M plasma-positive patients are limited; therefore, testing using plasma specimens is most appropriate for consideration in patients from whom a tumor biopsy cannot be obtained.
Additionally, the test is intended to provide tumor mutation profiling to be used by qualified health care professionals in accordance with professional guidelines in oncology for cancer patients with any solid malignant neoplasms. The test is for use with patients previously diagnosed with cancer and in conjunction with other laboratory and clinical findings.
Genomic findings other than those listed in Table 1 are not prescriptive or conclusive for labeled use of any specific therapeutic product.
Guardant360 CDx is a single-site assay performed at Guardant Health, Inc.
### III. CONTRAINDICATIONS
There are no known contraindications.
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
2 of 33
{2}
# IV. WARNINGS AND PRECAUTIONS
Warnings and precautions are listed below:
- Alterations reported may include somatic (not inherited) or germline (inherited) alterations. The assay filters germline variants from reporting except for pathogenic BRCA1, BRCA2, ATM, and CDK12 alterations. However, if a reported alteration is suspected to be germline, confirmatory testing should be considered in the appropriate clinical context.
- The test is not intended to replace germline testing or to provide information about cancer predisposition.
- Somatic alterations in ATM and CDK12 are not reported by the test as they are excluded from the test's reportable range.
- Genomic findings from cfDNA may originate from circulating tumor DNA (ctDNA) fragments, germline alterations, or non-tumor somatic alterations, such as clonal hematopoiesis of indeterminate potential (CHIP).
- Allow the tube to fill completely until blood stops flowing into the tube. Underfilling of tubes with less than 5 mL of blood (bottom of the label indicates 5 mL fill when tube is held vertically) may lead to incorrect analytical results or poor product performance. This tube has been designed to fill with 10 mL of blood.
# V. DEVICE DESCRIPTION
Guardant360 CDx is a single-site test performed at Guardant Health, Inc. The test includes reagents, software, and procedures for testing cfDNA from whole blood samples. The test uses 5-30 ng of cfDNA for library construction and next generation sequencing (NGS). Sequencing data is processed using a customized bioinformatics pipeline designed to detect several classes of genomic alterations, including nucleotide substitutions, indels, copy number amplifications, and genomic fusions / rearrangements. The device is designed to sequence 74 genes, but only report pre-defined and de novo alterations within the 55 genes outlined in Table 2. The test's reportable range for SNVs and indels covers approximately 46,000 bases.
Table 2. Genes Containing Alterations Detected by the Guardant360® CDx
| Alteration Type | Genes |
| --- | --- |
| Single Nucleotide Variants (SNVs) | AKT1, ALK, APC, AR, ARAF, ATM*, BRAF, BRCA1**, BRCA2**, CCND1, CDH1, CDK4, CDK6, CDK12*, CDKN2A, CTNNB1, EGFR, ERBB2, ESR1, FGFR1, FGFR2, FGFR3, GATA3, GNA11, GNAQ, HRAS, IDH1, IDH2, KIT, KRAS, MAP2K1, MAP2K2, MET, MLH1, MTOR, MYC, NF1, NFE2L2, NRAS, NTRK1, NTRK3, PDGFRA, PIK3CA, PTEN, RAF1, RET, RHEB, ROS1, SMAD4, SMO, STK11, TERT, TSC1, VHL |
| Indels | AKT1, ALK, APC, ATM*, BRAF, BRCA1**, BRCA2**, CDH1, CDK12*, CDKN2A, EGFR, ERBB2, ESR1, FGFR2, GATA3, HNF1A, HRAS, KIT, KRAS, MET, MLH1, NF1, PDGFRA, PIK3CA, PTEN, RET, ROS1, STK11, TSC1, VHL |
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
3 of 33
{3}
| Alteration Type | Genes |
| --- | --- |
| Copy Number Amplifications (CNAs) | ERBB2, MET |
| Fusions / Rearrangements | ALK, NTRK1, RET, ROS1 |
*Reporting is enabled for pathogenic germline alterations only. Somatic alterations will not be reported.
**Reporting is enabled for both germline and somatic alterations.
### Test Output
The test report includes variants reported in the following categories (Table 3):
Table 3. Category Definitions
| Category | Guardant360 CDx | | | Comments |
| --- | --- | --- | --- | --- |
| | Prescriptive use for a Therapeutic Product | Clinical Performance | Analytical Performance | |
| Category 1: Companion Diagnostic (CDx) | Yes | Yes | Yes | ctDNA biomarkers linked to the safe and effective use of the corresponding therapeutic product, for which Guardant360 CDx has demonstrated clinical performance shown to support therapeutic efficacy and strong analytical performance for the biomarker. |
| Category 2: ctDNA Biomarkers with Strong Evidence of Clinical Significance in ctDNA | No | No | Yes | ctDNA biomarkers with strong evidence of clinical significance presented by other FDA-approved liquid biopsy companion diagnostics for which Guardant360 CDx has demonstrated analytical reliability but not clinical performance. |
| Category 3A: Biomarkers with Evidence of Clinical Significance in tissue supported by: strong analytical validation using ctDNA | No | No | Yes | ctDNA biomarkers with evidence of clinical significance presented by tissue-based FDA-approved companion diagnostics or professional guidelines for which Guardant360 CDx has demonstrated analytical performance including analytical accuracy, and concordance of blood-based testing to tissue-based testing for the biomarker. |
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
4 of 33
{4}
| Category | Guardant360 CDx | | | Comments |
| --- | --- | --- | --- | --- |
| | Prescriptive use for a Therapeutic Product | Clinical Performance | Analytical Performance | |
| Category 3B: Biomarkers with Evidence of Clinical Significance in tissue supported by: analytical validation using ctDNA | No | No | Yes | ctDNA biomarkers with evidence of clinical significance presented by tissue-based FDA-approved companion diagnostics or professional guidelines for which Guardant360 CDx has demonstrated minimum analytical performance including analytical accuracy. |
| Category 4: Other Biomarkers with Potential Clinical Significance | No | No | Yes | ctDNA biomarkers with emergent evidence based on peer-reviewed publications for genes/variants in tissue, variant information from well-curated public databases, or in-vitro pre-clinical models, for which Guardant360 CDx has demonstrated minimum analytical performance. |
### Biomarker Rules for ERBB2 Activating Mutations Reported by Guardant360 CDx
The following ERBB2 activating mutations will be reported in Category 1 as a companion diagnostic (CDx) for patients with NSCLC for ENHERTU® (famtrastuzumab deruxtecan-nxki):
A775_G776insYVMA, Y772_A775dup, P780_Y781insGSP, G778_P780dup, G776delinsVC, G776_V777delinsCVCG, G776delinsLC, V777_S779dup, G776_V777insL, V777_G778insG, G778_S779insLPS, V777_G778insCG, A775_G776insV, A775_G776insTVMA, G776_V777insVGC, G778dup, G778_S779insCPG, L755S, V777L, G776S, S310F, G776V, V777M, S310Y, R678Q, T733I, L755M, L755P, L755W, D769N, D769H, D769Y, L755A, I767M, V842I, T862I, L869R, R896C, R896H, G776C, G776_V777insVC, S779_P780insVGS, I767F, T798I
### Biomarker Rules for ESR1 Missense Mutations Reported by Guardant360 CDx
Missense mutations between codons 310 and 547 of ESR1 will be reported in Category 1 as a companion diagnostic (CDx) for patients with breast cancer for ORSERDU (elacestrant).
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
5 of 33
{5}
# Test Kit Contents
The test includes the Guardant360 CDx Blood Collection Kit (BCK), which is sent to ordering laboratories. Each BCK contains two blood collection tubes. The BCK also contains supporting packaging materials, instructions for use and a return shipping label.
The BCK contains the following components:
- Streck blood collection tubes for specimen collection, stabilization, and transport of cfDNA; 2 per kit.
- Cushioning materials to prevent breakage of the blood collection tubes; 2 per kit
- Foam tray for protection of collection tubes during transport
- Absorbent sheet to be used during specimen shipping
- Biohazard specimen bag for protection during specimen transport
- Return shipping label for return of specimen to Guardant Health
- Barcodes for specimen identification and shipping instructions
- Instructions for Use for blood draw
- Patient welcome brochure which contains an overview of the test
- Test requisition form to complete to order Guardant360 CDx for a patient.
The test also includes the Guardant360 CDx Sample Preparation Kit (SPK), which is used in the Guardant Health Clinical Laboratory. The SPK contains reagents for library preparation, library enhancement, and cfDNA quantification/qualification. The kit is assembled into six (6) different boxes (referred to as box 1, 2, 3, 4a, 4b, and 4c) based on the usage of the reagents. The division of reagents amongst the boxes reflects different storage conditions and/or locations (e.g., different laboratory spaces).
# Instruments
Guardant360 CDx is intended to be performed with serial number-controlled instruments as indicated in Table 4. All instruments are qualified by Guardant Health, Inc. under the Guardant Health Quality System.
Table 4. Serial Number Controlled Instruments for Use with the Guardant360 CDx Assay
| Instrument |
| --- |
| Agilent Technologies 4200 TapeStation Instrument |
| Hamilton Company Microlab STAR |
| Hamilton Company Microlab STARlet |
| Illumina NextSeq 550 Sequencer |
| Veriti 96-Well Thermal Cycler |
# Test Process
# Whole Blood Collection and Shipping
The Guardant360 CDx Blood Collection Kit is used by ordering laboratories / physicians to collect whole blood specimens and ship them to the Guardant Health Clinical
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
6 of 33
{6}
Laboratory. A minimum of 5 mL whole blood must be received in order to achieve optimal performance for the Guardant360 CDx assay. Underfilling of tubes with less than 5 mL of blood may lead to incorrect analytical results or poor product performance.
## Plasma Isolation and cfDNA Extraction
Whole blood specimens are processed in the Guardant Health Clinical Laboratory within 7 days of blood collection. Plasma is isolated from both tubes of whole blood via centrifugation. One tube of plasma is stored, while the second tube is used for cfDNA extraction using the QIAGEN QIAsymphony SP Instrument and reagent system. The resulting cfDNA is quantified using the 4200 TapeStation. Input amounts ranging from 5 to 30 ng of cfDNA are further processed for each sample.
## Library Preparation and Enrichment
Reagents from the Guardant360 CDx Sample Preparation Kit are used during library preparation, enrichment, enrichment wash, and quantitation steps using the Veriti 96-Well Thermal Cycler, Microlab STAR and STARlet, and 4200 TapeStation Instruments. During library preparation, cfDNA fragment ends are repaired and library adapters containing inline barcodes are attached using blunt-end ligation. The resulting DNA is amplified by PCR to create libraries suitable for enrichment.
Amplified libraries are enriched for genes of interest using hybrid target capture with custom biotinylated RNA probes. Each enriched library is amplified by PCR using a unique index primer that also contains a sequencing flow cell attachment sequence. Amplified enriched libraries are pooled in equimolar amounts, denatured, and diluted to appropriate concentration for sequencing.
## DNA Sequencing
Paired-end sequencing by synthesis is performed with the Illumina NextSeq 550 Sequencing system. The amplified cfDNA is analyzed by parallel sequencing of amplified target genes to an average depth of coverage of greater than 2,700 unique molecules.
## Data Analysis and Reporting
The Guardant360 CDx Software uses a custom-developed analysis bioinformatics pipeline (BIP) software module. The BIP software module uses the raw data (output) from the targeted sequencing, partitions the data based on the sample index sequence (barcode) of each read to separate reads originating from individual samples, and executes a proprietary algorithmic reconstruction of the digitized sequencing signals based on molecular barcodes for high-fidelity molecule-based alteration calling downstream. The sequence data then undergoes an alignment process where it is mapped to the human genome (hg19) and an analysis of sequence alteration data is performed.
Alteration detection is conducted according to alteration calling metrics derived from clinical sample analysis. All alterations must pass alteration calling metrics as described in Table 5.
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
7 of 33
{7}
The SNV and indel cut-offs are defined in terms of mutant allele fraction (MAF) estimate, number and type of molecules supporting the alteration, pseudo-gene assessment, and likelihood ratio (LLR) score. The MAF estimate describes the calculated allelic fraction of an SNV or indel. The number of molecules describes the observed number of molecules meeting requirements for a particular alteration call. The LLR score is a calculated number that reflects how much observed support for the mutation exceeds expectations based on PCR and sequencing induced artifacts.
Table 5. Alteration Analytical Calling Threshold/Cut-Off Metrics
| SNV Calling Property | Metric |
| --- | --- |
| DNA Molecule Support | ≥ 2 |
| MAF Estimate | ≥ 0.001% |
| Log Likelihood Ratio | ≥ 0 |
| Indel Calling Property | Metric |
| DNA Molecule Support | ≥ 2 |
| MAF Estimate | ≥ 0.01% |
| Log Likelihood Ratio | ≥ 10 |
| CNA Calling Property | Metric |
| ERBB2 copy number | ≥ 2.18 |
| ERBB2 Z-score | ≥ 10 |
| ERBB2 amplification is not associated with chromosome-arm aneuploidy | TRUE |
| MET copy number | ≥ 2.16 |
| MET Z-score | ≥ 10 |
| MET amplification is not associated with chromosome-arm aneuploidy | TRUE |
| Fusion Calling Property | Metric |
| MAPQ score of supporting molecule to fusion sequence | > 30 |
| Number of unique fusion molecules | ≥ 2 |
| Number of unique fusion reads | > 2 |
The laboratory and physician receive a qualitative alteration-level result. A sample will receive an overall “Failed” result when any quality control (QC) metric is failed. Samples
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
8 of 33
{8}
failing any QC metric are automatically held and not released. The laboratory may attempt to rerun a patient sample that has failed a QC metric by using stored plasma or intermediate products.
Results from samples passing all QC metrics are formatted onto an IVD results report with CDx relevant information (Category 1) and all other biomarkers (Categories 2-4) within the LIMS system. The IVD results report will be populated with patient-specific information and may be merged with additional information provided by Guardant Health as a professional service prior to approval and release by the laboratory director or designee.
### **Quality Control Measures**
The Guardant360 CDx Sample Preparation Kit includes the Variant Control, which is engineered to contain known positive and negative alterations and is treated as a sample. Additionally, a no template negative control (NTC) is run in parallel with patient samples.
The Variant Control consists of a mixture of cfDNA from multiple human cancer cell lines containing all four alteration types, SNVs, indels, CNAs and fusions. The control is treated as a sample and processed starting from 15 ng cfDNA input through sequencing where it is analyzed for the presence and absence of the specific alterations.
Although the Variant Control does not contain all the alterations that the test is capable of detecting, concordant detection of alterations targeted in the Variant Control indicates that assay is performing as expected across the panel.
In addition to assessing Variant Control performance within a batch, the test is assessing multiple per-sample in-process and post-sequencing analytical metrics for each of the patient samples tested. These metrics provide in depth analytical QC information that complements Variant Control performance data and is specific and informative to that sample performance.
The NTC samples are absent of a DNA template, so cfDNA extraction, library preparation, and enrichment steps are expected to result in background level metrics.
## **VI. ALTERNATIVE PRACTICES AND PROCEDURES**
There are FDA approved companion diagnostic (CDx) alternatives for the detection of some of the genetic alterations using cfDNA isolated from plasma samples to those that are listed in Table 1 of the Guardant360 CDx intended use statement. These approved alternative CDx tests are listed in Table 6 below. Each alternative has its own advantages and disadvantages. A patient should fully discuss any alternative with their physician to select the most appropriate method that best meets expectations and lifestyle. For additional details see list of FDA Cleared or Approved Companion Diagnostic Devices at https://www.fda.gov/medical-devices/in-vitro-diagnostics/list-cleared-or-approved-companion-diagnostic-devices-in-vitro-and-imaging-tools
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
9 of 33
{9}
**Table 6: List of FDA-Approved CDx Assays for Genes Targeted by the Guardant360 CDx**
| Gene and Variant | Indication | Therapy | Device (PMA #) | Company | Technology |
| --- | --- | --- | --- | --- | --- |
| *EGFR* T790M | NSCLC | TAGRISSO® (osimertinib) | cobas® EGFR Mutation Test v2 | Roche Molecular Systems, Inc. | Polymerase Chain Reaction (PCR) |
| *EGFR* L858R and exon 19 deletions | | | FoundationOne® Liquid CDx | Foundation Medicine, Inc. | NGS |
Note: There are no FDA-approved CDx alternatives using cfDNA isolated from plasma for the detection of *KRAS* G12C mutations for the identification of patients with NSCLC eligible for treatment with LUMAKRAS™ (sotorasib). However, there is one FDA approved CDx alternative for the detection of *KRAS* G12C mutation in patients with NSCLC using tissue specimens for treatment with LUMAKRAS™ (sotorasib): QIAGEN *therascreen*® KRAS RGQ PCR Kit (See SSED for P110027/S012).
Similarly, there are no FDA-approved CDx alternatives using cfDNA isolated from plasma for the detection of *EGFR* exon 20 insertions for the identification of patients with NSCLC eligible for treatment with RYBREVANT® (amivantamab-vmjw). There are no FDA-approved CDx alternatives using cfDNA isolated from plasma for the detection of *ERBB2* activating mutations (SNVs and exon 20 insertions) for the identification of patients with NSCLC eligible for treatment with ENHERTU® (fam-trastuzumab deruxtecan-nxki). Also, there are no FDA-approved CDx alternatives using cfDNA isolated from plasma for the detection of *ESR1* missense mutations (subject of this PMA Supplement) in patients with breast cancer (BC) eligible for treatment with ORSERDU (elacestrant).
However, there is an FDA approved CDx (Life Technologies Corporation’s OncomineDx Target test) alternative for the detection of *EGFR* exon 20 insertions in NSCLC patients using tissue specimens for treatment with RYBREVANT® (amivantamab-vmjw) (See labeling for P160045/S027). There is also an FDA approved (Life Technologies Corporation’s OncomineDx Target test) CDx alternative for the detection of *ERBB2* activating mutations (SNVs and exon 20 insertions) in patients with NSCLC using tissue specimens for treatment with ENHERTU® (fam-trastuzumab deruxtecan-nxki) (See SSED for P160045/S035).
## VII. MARKETING HISTORY
Guardant Health, Inc. initially designed and developed the Guardant360 laboratory developed test (Guardant360 LDT), and the first commercial sample was tested in 2012 to detect the presence of genomic alterations in plasma isolated from whole blood. The Guardant360 LDT is not FDA-cleared or -approved.
The Guardant360 CDx Premarket Approval (PMA) application was FDA-approved on August 7, 2020, and subsequently commercialized in the USA. The approved PMA and its supplements that affected the Intended Use are listed in Table 7.
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
10 of 33
{10}
**Table 7. Marketing History**
| Submission No. | Date of Approval | Biomarker | Patient Population | Drug |
| --- | --- | --- | --- | --- |
| P200010 | August 7, 2020 | *EGFR* exon 19 deletions, L858R, and T790M | NSCLC | TAGRISSO® (osimertinib) |
| P200010/S001 | May 21, 2021 | *EGFR* exon 20 insertions | NSCLC | RYBREVANT® (amivantamab-vmjw) |
| P200010/S002 | May 28, 2021 | *KRAS* G12C | NSCLC | LUMAKRAS™ (sotorasib) |
| P200010/S008 | August 11, 2022 | *ERBB2/HER2* activating mutations (SNVs and exon 20 insertions) | NSCLC | ENHERTU® (fam-trastuzumab deruxtecan-nxki) |
### **VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH**
Failure of the device to perform as expected or failure to correctly interpret test results may lead to incorrect test results, and subsequently, may lead to inappropriate patient management decisions. Patients with false positive results may undergo treatment with the therapy listed in the intended use statement without clinical benefit and may experience adverse reactions associated with the therapy. Patients with false negative results may not be considered for treatment with the indicated therapy. There is also a risk of delayed results, which may lead to delay of treatment with indicated therapy.
### **IX. SUMMARY OF NONCLINICAL STUDIES**
#### **A. Laboratory Studies**
The primary evidence for supporting the performance of Guardant360 CDx in detecting *ESR1* missense mutations between codons 310-547 (hereafter referred to as *ESR1* missense mutations) was from the data presented using intended use specimens across all validation studies. In addition to the existing platform-level validation data (P200010, P200010/S001 and P200010/S002), analytical accuracy/concordance, limit of detection (LoD), precision at the LoD and sample stability (whole blood, plasma, cfDNA) studies, were conducted to support the indication for the *ESR1* missense mutation biomarker.
For Guardant360 CDx platform-level validation (P200010), performance characteristics were established using plasma-derived cfDNA samples from a wide range of cancer types. Each study included CDx variants, as well as a broad range of representative alteration types (SNVs, indels, CNAs, and fusions/rearrangements) in various genomic contexts across several genes. The platform validation studies included samples with *ESR1* missense mutations in plasma specimens from patients with BC and without BC. These results from the platform-level validation (P200010, P200010/S001, and P200010/S002) that have been leveraged to support Guardant360 CDx detection of *ESR1* missense mutations are referenced below. Additional *ESR1*-specific analytical validation studies are described below.
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
11 of 33
{11}
## 1. Analytical Accuracy/Concordance
Accuracy for the detection of *ESR1* missense mutations was established relative to an externally validated mass spectrometry-based comparator method by comparing the results from 259 samples from the EMERALD (NCT03778931) clinical study. The analysis included 258 samples that passed QC on both assays. Among the 258 samples, two pairs of samples were identified to be duplicate as each sample pair was drawn from the same subject on different days. Therefore, these two pairs of samples (n=4) were excluded from the analysis and 254 samples were included in the Accuracy Study.
A summary of positive percent agreement (PPA), negative percent agreement (NPA), and the corresponding 95% two-sided exact confidence intervals (CIs) is provided in Table 8. Out of the 254 remaining samples, 121 samples (47.6%) were positive for an eligible *ESR1* variant by both Guardant360 CDx and the comparator method [Guardant360 CDx (+) / Comparator (+)], while 110 samples (43.3%) were negative by both platforms [Guardant360 CDx (-) / Comparator (-)]. Three samples (1.2%) were negative by Guardant360 CDx but positive by the comparator method [Guardant360 CDx (-) / Comparator (+)], and 20 samples (7.9%) were positive by Guardant360 CDx but negative by the comparator method [Guardant360 CDx (+) / Comparator (-)].
**Table 8. Summary of Concordance Between Guardant360 CDx and the Comparator for *ESR1* Missense Mutations**
| Alteration Type | Guardant360 CDx (+), Comparator (+) | Guardant360 CDx (+), Comparator (-) | Guardant360 CDx (-), Comparator (+) | Guardant360 CDx (-), Comparator (-) | Patients (n) | PPA (95% CI)* | NPA (95% CI)* |
| --- | --- | --- | --- | --- | --- | --- | --- |
| *ESR1* missense mutations | 121 | 20 | 3 | 110 | 254 | 97.6% (93.1% - 99.5%) | 84.6% (77.2% - 90.3%) |
\*The Clopper-Pearson Exact Method was used for the confidence interval (CI) analysis.
Among the 254 samples, 191 samples were enrolled in the EMERALD (NCT03778931) clinical study and have valid drug efficacy data, i.e., progression free survival (PFS) data. To understand if the low NPA of the Accuracy Study was due to the low sensitivity of the comparator method or the erroneous detection of the Guardant360 CDx, drug response data for the patients with discordant samples [Guardant360 CDx (+) / Comparator (-)], who were enrolled in the clinical study based on the CDx result, were analyzed.
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
12 of 33
{12}
Patients with discordant [Guardant360 CDx (+) / Comparator (-)] results in the elacestrant arm had a median PFS of 4.58 months, which is similar to the median PFS (3.78 months) observed in the patients assigned to the elacestrant arm in the clinical trial. Moreover, the discordant samples [Guardant360 CDx (+) / Comparator (-)] in the standard of care (SOC) arm had a median PFS of 1.87 months, which is the same median PFS observed in the total subjects assigned to the SOC arm. Refer to Section X for further details regarding the EMERALD (NCT03778931) clinical study.
The results show that Guardant360 CDx effectively detects ESR1 missense mutations in patients with BC. The discordant samples [Guardant360 CDx (+) / Comparator (-)] had MAF values ranging from 0.04% to 0.56%, with the majority of samples (17 of 20 samples) harboring MAF values below the LoD (Refer to Section IX.A.2.b for further details regarding the LoD study). The discordance between the comparator method and Guardant360 CDx may be due to the stochastic detection of mutations below LoD of both assays, along with sensitivity differences between Guardant360 CDx and the comparator method.
## 2. Analytical Sensitivity
### a. Limit of Blank (LoB)
Please refer to the Summary of Safety and Effectiveness Data P200010 (Section IX.A.3.a) for Guardant360 CDx platform-level analytical sensitivity data for LoB. There were no false positives for ESR1 missense mutations among 240 replicates tested across three unique reagent lots.
### b. Limit of Detection (LoD)
The LoDs for ESR1 E380Q, Y537S and D538G were established using patient sample pools derived from patients with BC at 5 ng and 30 ng cfDNA input. E380Q is located in exon 7, and Y537S and D538G are located in exon 10 of the ESR1 gene. All of these mutations are representative of prevalent ESR1 missense mutations. Trial pools targeting 2.16% MAF at 5 ng input (pool 1) and 0.4% MAF at 30 ng input (pool 2) were characterized with Guardant360 CDx. For cfDNA at 5 ng input, the LoD was estimated using 5 MAF levels ranging from 0.10% to 2.16% (0.10%, 0.65%, 0.97%, 1.45% and 2.16%) with 24 replicates per level. For cfDNA at 30 ng input, the LoD was estimated using 5 MAF levels ranging from 0.017% to 0.40% (0.017%, 0.12%, 0.18%, 0.27% and 0.40%) with 14 replicates per level. Since fewer than three observed MAF levels had detection rates between 10% and 90%, an empirical estimate of the LoD was used, which was defined as the lowest MAF observed for which all MAF levels at or above this MAF had a hit rate ≥ 95%. The LoDs for ESR1 E380Q, Y537S and D538G are summarized in Table 9.
The established LoDs were confirmed for these prevalent ESR1 mutations by testing clinical sample pools from patients with BC near the established LoDs (refer to Table 9) across 24 replicates at 5 ng input using a combined LoD Confirmation and Precision Study. Refer to Section IX.A.4 for further details regarding the combined LoD Confirmation and Precision Study.
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
13 of 33
{13}
Table 9. Summary of Established LoDs for Prevalent ESR1 Missense Mutations in BC Clinical Samples
| ESR1 Missense Mutation | LoD, 5ng input MAF (%) | LoD, 30 ng input MAF (%) |
| --- | --- | --- |
| E380Q | 1.0 | 0.3 |
| Y537S | 1.0 | 0.3 |
| D538G | 1.1 | 0.2 |
In addition to confirming the LoDs for the prevalent mutations, a combined LoD Confirmation and Precision Study was also performed using sample pools from patients with BC at 5 ng cfDNA input with a 2-3% targeted MAF (2-3x LoD for the prevalent mutations) across 24 replicates for three rare ESR1 mutations (H356D, G442A and S463P). The ESR1 mutations H356D, G442A and S463P are located in exons 6, 8 and 9, respectively. The LoDs for ESR1 G442A and S463P were confirmed; however, the study result of the LoD confirmation and precision study for ESR1 H356D has a PPA of 83.3% or 91.7%, which did not achieve agreement of ≥ 95% across all conditions for H356D at 5 ng of cfDNA input (Table 10).
Table 10. Summary of Confirmed LoDs for ESR1 Missense Mutations in Clinical Samples from Patients with BC
| ESR1 Missense Mutation | LoD, 5ng input (MAF%) | Number Positive/ Number Expected | PPA (95% CI)** |
| --- | --- | --- | --- |
| H356D | 2.1^ | 20/24 | 83.3% (62.6% - 95.3%) |
| H356D | 3.1^ | 22/24 | 91.7% (73.0% - 99.0%) |
| G442A | 2.3* | 24/24 | 100% (85.8%-100%) |
| S463P | 2.8* | 24/24 | 100% (85.8%-100%) |
^LoD values were not confirmed for H356D as the PPA did not achieve agreement of ≥95%.
*LoD values were confirmed at 2.1x (G422A) and 2.6x (S463P) of the prevalent ESR1 mutation LoD values.
**The Clopper-Pearson Exact Method was used for the CI analysis.
Coverage analysis showed that the mean coverage at the H356D locus (independent of variant detection) was lower than the coverage for the other two ESR1 missense mutations (G442A and S463P). The mean Non-Singleton Coverage (NSC) of H356D is 522.52, while that of G442A and S463P are 583.00 and 567.83, respectively. The reason for the study did not demonstrate
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
14 of 33
{14}
agreement of ≥ 95% H356D may be due to stochastic, near-threshold effects at a reduced coverage. Given these reasons and the rarity of H356D mutations (0.88% of the patients enrolled in the EMERALD (NCT03778931) clinical study were observed to harbor a H356D mutation), the study result was deemed acceptable. The LoD confirmation study results (Table 10) suggest that the LoDs for theses variants are higher (~2x-3x higher) than the established LoDs for the variants in Table 9 at 5 ng cfDNA input.
### 3. Analytical Specificity
Please refer to the Summary of Safety and Effectiveness Data P200010 and P200010/S002 (Section IX.A.4) for Guardant360 CDx platform validation of analytical specificity, including endogenous and microbial interfering substances and in silico specificity. The effect of potential exogenous interfering substances that may carry over from cfDNA extraction on assay performance was evaluated in the PMA supplement (P200010/S002).
### 4. Precision
The purpose of the precision study was to demonstrate the repeatability and within-site reproducibility of Guardant360 CDx for detecting representative ESR1 missense mutations (H356D, E380Q, G442A, S463P, Y537S and D538G) through closeness of agreement between measured qualitative detection in replicates using different combinations of reagent lots, instruments, operators, and days. The study was conducted with sample pools from patients with BC harboring three prevalent missense mutations (E380Q, Y537S and D538G) or three rare mutations (H356D, G442A and S463P).
For prevalent mutations (E380Q, Y537S and D538G), precision was evaluated using 5 ng input with MAF levels targeted at 1x LoD. Each variant was tested across six unique reagent lot-instrument-operator combinations with four replicates per combination (6x4=24 samples). This study successfully verified the precision of Guardant360 CDx for detecting the three representative ESR1 prevalent mutations (E380Q, Y537S and D538G) within and between different reagent lots, instrument sets, and operator groups with samples near LoD processed on different runs and days in the Guardant Health Clinical Laboratory (Table 11).
For rare mutations (H356D, G442A and S463P), the precision study was performed using 5 ng input at analyte levels targeted to 2.0-3.0% MAF (1.8x-2.9x of the established LoD (i.e., 1.1% MAF) for the prevalent ESR1 missense mutations). Each variant was tested across six unique reagent lot-instrument-operator combinations, with four replicates per combination (6x4=24 samples). The PPA of ESR1 SNVs G442A and S463P were observed to be 100%; however, the ESR1 H356D mutation showed a PPA of 83.3% or 91.7% at 2.1% MAF and 3.1% MAF, respectively, which did not demonstrate agreement of ≥ 95% (Table 11). By evaluating 25 additional replicates from the same sample pool for H356D, PPA of 95.9% was achieved.
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
15 of 33
{15}
**Table 11. Summary of Precision Results for Representative *ESR1* Missense Mutations**
| *ESR1* Missense Mutation | Observed MAF% | Relative LoD Level* | Number Positive/ Number Expected | PPA (95% CI)*** |
| --- | --- | --- | --- | --- |
| E380Q | 1.0 | 1.0x | 24/24 | 100% (85.8%-100%) |
| Y537S | 1.0 | 1.0x | 23/24 | 95.8% (78.9% - 99.9%) |
| D538G | 1.1 | 1.0x | 23/24 | 95.8% (78.9% - 99.9%) |
| H356D | 2.1** | 2.0x | 20/24 | 83.3% (62.6% - 95.3%) |
| H356D | 3.1** | 2.9x | 22/24 | 91.7% (73.0% - 99.0%) |
| G442A | 2.3 | 2.1x | 24/24 | 100% (85.8%-100%) |
| S463P | 2.8 | 2.6x | 24/24 | 100% (85.8%-100%) |
*Compared to the established LoD for the prevalent *ESR1* missense mutations.
**Note that the observed MAF is the average variant MAF from all samples with a reported variant (i.e., excluding dropouts).
***The Clopper-Pearson Exact Method was used for the CI analysis.
The original PMA (P200010) comprised negative precision data from 72 self-declared cancer-free age-matched healthy donors. 240 replicates were tested at 30 ng inputs across three precision combinations of operator group, instrument combination, reagent lots and start dates. No *ESR1* false positive mutations were detected (NPA 100%, 240/240). These data are leveraged to support this PMA supplement. Please refer to the Summary of Safety and Effectiveness Data for P200010 (Section IX.A.5) for details on precision for mutation-negative samples, precision starting from plasma extraction, and precision studies in support of the Guardant360 CDx platform validation.
##### **5. Carryover/Cross-Contamination**
Please refer to the Summary of Safety and Effectiveness Data of P200010 (Section IX.A.6) for platform-level carryover/cross-contamination data for Guardant360 CDx.
##### **6. Reagent Lot Interchangeability**
Please refer to the Summary of Safety and Effectiveness Data P200010 (Section IX.A.7) for Guardant360 CDx platform validation of reagent lot interchangeability.
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
16 of 33
{16}
## 7. Stability
Please see the Summary of Safety and Effectiveness Data P200010 (Section IX.A.8) for Guardant360 CDx platform level reagent and sample stability, including whole blood stability, plasma stability, cfDNA stability, and intermediate sample stability. Additional stability studies were performed to verify the platform level stability for whole blood, plasma and cfDNA samples, which evaluated samples from patients with BC harboring *ESR1* missense mutations.
### a. Whole Blood Stability
The purpose of this study was to demonstrate the stability of whole blood specimens used for Guardant360 CDx collected in the Guardant360 BCK, across the expected range of sample transport and storage conditions for up to 7 days after blood collection prior to plasma isolation. While whole blood stability from patients with multiple different cancer types has been examined previously (P200010), this study focuses specifically on specimens derived from patients with metastatic breast cancer (mBC).
A total of four BCTs were drawn from each of 11 patients with mBC and subjected to the conditions described in Table 12 below. For each patient, a single tube of blood was processed to plasma one day after the blood draw (stored at room temperature). Plasma was then shipped on dry ice to Guardant Health. This constituted the reference condition. In addition to the reference tube, three more blood tubes per donor were shipped as whole blood to Guardant Health and subjected to Condition 1 (Summer profile), Condition 2 (Winter profile) or Condition 3 (Room temperature), as described in Table 12. After conditioning, plasma was isolated on the 8th day after blood collection and run using the Guardant360 CDx assay.
**Table 12. Description of Whole Blood Storage Conditions**
| Condition | BCT per Patient | Storage Condition / Processing |
| --- | --- | --- |
| Reference (Ref) | 1 | Reference condition: Plasma processing (day 1 after blood collection). |
| Condition 1 (C1) | 1 | Summer Profile Storage: 4h at 22°C, 6h at 37°C and 56h at 22°C, 6h at 37°C) plus remaining time at room temperature. |
| Condition 2 (C2) | 1 | Winter Profile Storage: 4h at 18°C, 6h at 0°C, 56h at 10°C, and 6h at 0°C plus remaining time at room temperature. |
| Condition 3 (C3) | 1 | Room Temperature Storage: Storage at room temperature 18-25°C. |
Among the 44 samples, one sample failed the 'On Target Rate' post-sequencing QC metric, and four samples from one patient were excluded from the analysis because the reference condition sample had less than 5ng input. The remaining 39 samples resulted in 10 evaluable patient groups for the
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
17 of 33
{17}
Winter Profile (C2) and Room Temperature (C3) conditions and 9 evaluable patient groups for the Summer Profile condition (C1). All storage conditions demonstrated acceptable performance (Table 13). All samples in each group demonstrated acceptable sample-level molecule recovery, as assessed by depth of NSC coverage across the panel. The fold change of median NSC in the test condition over the reference condition or time zero ranged from 0.87 to 1.00.
Table 13. Whole Blood Stability Summary Results
| | | | | Room Temp |
| --- | --- | --- | --- | --- |
| | | | | |
| Sample-Level Molecule Recovery | Median of NSC fold (condition / reference) | 0.87 | 0.94 | 1.00 |
| Exon-Level Molecule Recovery | 90% LCL of Fraction of Exons with Expected Coverage | 98.26% | 98.71% | 98.55% |
| Variant Call Concordance* | PPA (n/N) | 83.33% (10/12) | 75.00% (9/12) | 75.00% (9/12) |
| | PPA (≥ 1X LoD) (n/N) | 100.00% (6/6) | 100.00% (6/6) | 100.00% (6/6) |
| | NPA (n/N) | >99.99% (25,315/ 25,316) | 100.00% (27,849/ 27,849) | >99.99% (27,848/ 27,849) |
| Variant Call Concordance (ESR1)** | PPA (n/N) | 85.71% (6/7) | 85.71% (6/7) | 71.43% (5/7) |
| | PPA (≥ 1X LoD) (n/N) | 100.00% (3/3) | 100.00% (3/3) | 100.00% (3/3) |
| | NPA (n/N) | 99.51% (204/205) | 100.00% (226/226) | 99.56% (225/226) |
LCL, Lower Confidence Limit.
*The variant call concordance analysis is based on the results from all detectable variants.
**The variant call concordance analysis is limited to the ESR1 variants.
Exon-level coverage was also acceptable for all conditions evaluated. The fraction of exons with relative exon level coverage difference between each condition and the reference (Time zero) within 2σ (2 × 0.204) was 98.56%-98.95%, which demonstrate that there was no preferential drop-out of relative exon-level coverage exceeding expected levels due to random variation. The
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
18 of 33
{18}
lower bound of the 90% exact binomial CI for the fraction of genomic targeted exonic regions with relative exon-level NSC is shown in Table 13.
PPAs were also calculated for the SNVs, insertion-deletion mutations (indels) and copy number alterations (CNAs) in the reportable range: 11 SNVs and 1 indel. All conditions showed variant call concordance PPA of 75.00% - 83.33%. The low PPA for variant call concordance is probably due to the sub-LoD sample level (0.14x - 0.90x LoD) for some samples. PPA above LoD was 100.00% for all conditions (Table 13). For the NPA analysis, a total of 2,533 genomic positions were considered for each sample pair. The ratio of genomic positions in agreement and the total number of assessed positions are shown in parentheses.
To demonstrate biomarker-specific variant call concordance, the PPA and NPA calculations were repeated for ESR1 missense mutations within the biomarker definition (i.e., codons 310-547). The PPA was observed to be 85.71% for both the Summer and Winter Profiles and 71.43% for the Room Temperature Profile. One variant - Y537N (c.1609T>A), which had a sample level of 0.25x LoD, was not detected in all storage conditions; the other variant - L536Q (c.1607_1608delTCinsAG), which had a sample level of 0.90x LoD, was not detected in C3 (Room Temperature). Detection below or close to the LoD threshold is expected to be stochastic. For variants ≥ 1x LoD in the reference condition, the PPA was observed to be 100% for all storage conditions. The NPA calculations were limited to only consider the 21 genomic locations that overlapped the biomarker definition that were found in either this study or in P200010. The NPA was observed to be 99.50% in the summer profile, 100.00% in the Winter profile, and 99.56% in the Room Temperature profile (Table 13).
The results confirm the observations from P200010 for BC samples that whole blood storage in Guardant360 BCKs preserves cfDNA quantity and quality for up to seven days and in extreme weather shipping conditions.
### b. Plasma Stability
The purpose of this study was to demonstrate the stability of frozen plasma samples at -80°C ± 10°C, which is representative of the expected range of sample storage conditions for the specimen type. While plasma stability has been examined previously (P200010) for all variant types across multiple cancer indications, this study focuses specifically on specimens derived from patients with advanced breast cancer.
The study samples were selected in reverse chronological order from remnant BC plasma specimens in the Guardant Health biobank that were sequenced using Guardant360 CDx at least 46 days before the study start date. Two BCTs from 22 cancer patients, 44 samples in total, were collected and run on Guardant360 CDx, with plasma stored at the specified storage conditions (Table 14).
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
19 of 33
{19}
Table 14. Description of Plasma Storage Conditions
| Condition | BCT per Patient | Storage Condition / Processing |
| --- | --- | --- |
| Reference | 1 | Reference condition: cfDNA extraction after plasma isolation within 48 hours of delivery. |
| Condition | 1 | Storage condition: Storage of plasma at -80 ± 10°C for ≥ 45 days before cfDNA extraction. |
All 44 samples passed their respective in-process and post-sequencing QC metrics. In the tested storage condition, samples demonstrated acceptable sample-level molecule recovery, relative exon-level coverage, variant call concordance for all variants, including variant call concordance for ESR1 (Table 15).
Sample level molecule recovery showed a NSC fold change of 0.94. The lower bound of the 90% CI for the relative exon coverage metric for the storage condition was 98.05%. The PPA was also calculated for a total of 43 variants (37 SNVs and 6 Indels). The storage condition showed variant call concordant PPA of 88.37%. The PPA above LoD was 96.97%. The NPA across the reportable range was 99.99% (Table 15).
Table 15. Plasma Stability Summary Results
| Study Endpoint | Metric | Study Result |
| --- | --- | --- |
| Sample-Level Molecule Recovery | Median of NSC fold (condition/reference) | 0.94 |
| Exon-Level Molecule Recovery | 90% LCL of Fraction of Exons with Expected Coverage | 98.05% |
| Variant Call Concordance* | PPA (n/N) | 88.37% (38/43) |
| | PPA (≥ 1X LoD) (n/N) | 96.97% (32/33) |
| | NPA (n/N) | 99.99% (55,958/55,961) |
| Variant Call Concordance (ESR1)** | PPA (n/N) | 73.68% (14/19) |
| | PPA (≥ 1X LoD) (n/N) | 88.89% (8/9) |
| | NPA (n/N) | 100.00% (510/510) |
* The variant call concordance analysis is based on the results from all detectable variants.
**The variant call concordance analysis is limited to the ESR1 variants.
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
20 of 33
{20}
The PPA and NPA calculations were also performed for ESR1 missense mutations within the biomarker definition. The PPA was observed to be 73.68%. Among the five discordant variants, four variants were detected below or near the LoD threshold (0.10x – 0.87x LoD). Detection below or close to the LoD threshold is expected to be stochastic. For variants ≥ 1x LoD in the reference condition, the PPA was observed to be 88.89% for the storage condition, and one variant – D351N (c.1051G>A), which had a sample level of 1.77x LoD was not detected in the storage condition. The NPA was observed to be 100% in the storage condition (Table 15).
Based on these study results, plasma may be stored at -80 ± 10°C for ≤ 45 days before cfDNA extraction.
# c. cfDNA Stability
The purpose of this study was to evaluate the stability of cfDNA samples stored at -20°C ± 5°C for 45 days (Table 16). While cfDNA stability has been examined previously (P200010) for all variant types across multiple cancer indications, this study focused specifically on specimens derived from patients with advanced breast cancer. In total, 56 samples were collected from 28 patients. One sample failed the post-enrichment quantification step due to high molarity and was excluded from the analysis. The remaining 55 samples passed all QC metrics with a 98.2% pass rate, resulting in 27 evaluable sample pairs.
Table 16. Description of cfDNA Storage Conditions
| Condition | BCT per Patient | Storage Condition / Processing |
| --- | --- | --- |
| Reference | 1 | Reference condition: Quantitation, dilution and library preparation on the same day as cfDNA extraction. |
| Condition | 1 | Quantitation and dilution on the same day as cfDNA extraction, followed by storage of cfDNA at -20°C ± 5°C for 46 days plus 1 freeze/thaw cycle before library preparation. |
In the tested storage condition, samples demonstrated acceptable sample-level molecule recovery, relative exon-level coverage, and variant call concordance (Table 17).
Sample level molecule recovery showed a NSC fold change of 1.05. The lower bound of the 95% CI for the relative exon coverage metric for the storage condition was 90.30%. The PPA was also calculated for a total of 106 variants (96 SNVs and 10 Indels). The storage condition showed a variant call concordance PPA of 89.62%, and the PPA above LoD was 96.05%. The NPA across the reportable range was 99.97% (Table 17).
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
21 of 33
{21}
In total, 74 ESR1 SNVs were detected in the reference condition. The PPA was observed to be 89.19%. The variant call discordance is probably due to the sub-LoD detection in the reference condition (variants at 0.13x - 0.90x LoD sample level). For variants ≥ 1x LoD in the reference condition, the PPA was observed to be 96.00% for the storage condition. The NPA was observed to be 99.36% in the storage condition (Table 17).
Table 17. cfDNA Stability Summary Results
| Study Endpoint | Metric | Study Result |
| --- | --- | --- |
| Sample-Level Molecule Recovery | Median of NSC fold (condition/reference) | 1.05 |
| Exon-Level Molecule Recovery | 95% LCL of Fraction of Exons with Expected Coverage | 90.30% |
| Variant Call Concordance* | PPA (n/N) | 89.62% (95/106) |
| | PPA (≥ 1X LoD) (n/N) | 96.05% (73/76) |
| | NPA (n/N) | 99.97% (69,018/69,040) |
| Variant Call Concordance (ESR1)** | PPA (n/N) | 89.19% (66/74) |
| | PPA (≥ 1X LoD) (n/N) | 96.00% (48/50) |
| | NPA (n/N) | 99.36% (1,243/1,251) |
*The variant call concordance analysis is based on the results from all detectable variants.
**The variant call concordance analysis is limited to the ESR1 variants.
Based on these study results, extracted cfDNA may be stored at -20°C for ≤ 45 days for Guardant360 CDx analysis.
### 8. Guard Banding/Robustness
Please refer to the Summary of Safety and Effectiveness Data P200010/S002 (Section IX.A.6) for Guardant360 CDx platform level Guard Banding/Robustness study.
### 9. General Lab Equipment and Reagent Evaluation
Please see the Summary of Safety and Effectiveness Data P200010 (Section IX.A.9) for Guardant360 CDx platform validation of general lab equipment and reagents, including cfDNA extraction as well as other instruments and reagents.
### B. Animal Studies
No animal studies were conducted using Guardant360 CDx.
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
22 of 33
{22}
# X. SUMMARY OF PRIMARY CLINICAL STUDY
The safety and effectiveness of the Guardant360 CDx assay for identifying patients with mBC with ESR1 missense mutations between codons 310 and 547 (hereafter referred to as ESR1 missense mutations) who may benefit from treatment with ORSERDU (elacestrant) was demonstrated through testing of cfDNA in pre-treatment plasma specimens from patients enrolled into the EMERALD (NCT03778931) clinical study, which was conducted to support the efficacy of ORSERDU (elacestrant). In the EMERALD clinical study, patients were prospectively stratified based on Guardant360 CDx detection of ESR1 missense mutations. A summary of the Guardant360 CDx clinical validation study is presented below.
# A. EMERALD (NCT03778931) Clinical Study Design
The EMERALD clinical study was an international, multicenter, randomized, open-label, active-controlled, event-driven, Phase 3 clinical study comparing the safety and efficacy of ORSERDU (elacestrant) to the standard of care (SOC) treatment options of fulvestrant or an aromatase inhibitor (AI) in post-menopausal women and men with ER+/HER2- mBC. Eligible patients were randomized in a 1:1 ratio to either ORSERDU (elacestrant) or SOC and stratified by ESR1 mutation status using Guardant360 CDx and other criteria described in the clinical study protocol.
Patients from the primary EMERALD registration population positive for ESR1 mutations by Guardant360 CDx were included in the diagnostic study primary clinical efficacy cohort to assess the clinical validity of Guardant360 CDx to identify BC patients with ESR1 missense mutations to aid in the selection of patients for treatment with ORSERDU (elacestrant) therapy.
# 1. Clinical Study Inclusion and Exclusion Criteria
Patients in the primary EMERALD registration population were included in the diagnostic study efficacy cohort if the selection criteria were met. The key inclusion criteria are summarized below:
- Male or postmenopausal female
- Histologically- or cytologically-proven adenocarcinoma of the breast with evidence of either locally advanced disease not amenable to resection or radiation therapy with curative intent or metastatic disease not amenable to curative therapy
- Must be appropriate candidates for endocrine monotherapy
- Must have ER+ and HER2- tumor status confirmed
- Must have previously received at least one and no more than two lines of endocrine therapy, either a monotherapy or as a combination therapy with another agent; prior treatment with a CDK4/6 inhibitor in combination with either fulvestrant or an AI; and no more than one line of cytotoxic chemotherapy in the mBC setting
- Measurable disease or non-measurable bone-only disease.
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
23 of 33
{23}
## 2. Follow-up Schedule
The Guardant360 CDx diagnostic study involved only testing and analysis of plasma samples; as such, no additional patient subject follow-up was conducted in regard to the diagnostic study.
## 3. Clinical Endpoints
The clinical endpoint used to assess ORSERDU (elacestrant) efficacy in the EMERALD (NCT03778931) clinical study primary objective was progression free survival (PFS) by RECIST version 1.1 as assessed by independent central review (ICR) or death from any cause.
## 4. Diagnostic Objective and Endpoints
The primary objective of the diagnostic clinical study was to demonstrate the safety and effectiveness of Guardant360 CDx as a companion diagnostic to aid in the selection of breast cancer patients with ESR1 missense mutations for ORSERDU (elacestrant) therapy. This was a co-development study utilizing plasma samples and the clinical outcome data from the EMERALD (NCT03778931) clinical study. Guardant360 CDx was used to test all available samples.
The primary objective was assessed by comparing the efficacy of ORSERDU (elacestrant) to that of subjects on SOC therapy in patients harboring ESR1 mutations detected by Guardant360 CDx. The primary endpoint is the same as that used for the clinical study, PFS by RECIST 1.1 as assessed by ICR.
## B. Accountability of PMA Cohort for the Guardant360 CDx Clinical Study for ESR1 Missense Mutations
The EMERALD (NCT03778931) clinical study registration population (Full Analysis Set, FAS) included 478 subjects, of which 228 subjects had ESR1 missense mutations detected by Guardant360 CDx, while 249 subjects did not have an ESR1 missense mutation detected by Guardant360 CDx (Figure 1). Of note, one patient was excluded from the diagnostic study efficacy analysis due to QC failure with the final Guardant360 CDx bioinformatics software.
Figure 1. Guardant360 CDx ESR1 Missense Mutations Efficacy Analysis Patient Accountability and Analysis Set Definitions

PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
24 of 33
{24}
### C. Study Population Demographics and Baseline Parameters for the Guardant360 CDx Clinical Study for ESR1 Missense Mutations
Demographics and baseline clinical characteristics of patients enrolled in the EMERALD (NCT03778931) clinical study was categorized relative to the diagnostic study populations, as defined by Guardant360 CDx results. As shown in Tables 18 and 19, the diagnostic study primary efficacy population ESR1-mutated (ESR1-mut) and ESR1-mutation-not-detected (ESR1-mut-nd) were well balanced. The ORSERDU (elacestrant) and SOC treatment arms were also well balanced.
Table 18. Clinical Effectiveness Analysis Subgroup Demographics
| | Elacestrant | | SOC | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | ESR1-mut | ESR1-mut-nd | ESR1-mut | ESR1-mut-nd | ESR1-mut | ESR1-mut-nd |
| Analysis set: | 115 | 124 | 113 | 125 | 228 | 249 |
| Age, years | | | | | | |
| N (missing) | 115 (0) | 124 (0) | 113 (0) | 125 (0) | 228 (0) | 249 (0) |
| Mean | 62.7 | 62.4 | 62.0 | 64.4 | 62.4 | 63.4 |
| SD | 12.25 | 11.91 | 11.74 | 10.03 | 11.98 | 11.03 |
| Median | 64.0 | 63.0 | 63.0 | 64.0 | 63.0 | 64.0 |
| Min | 28 | 24 | 32 | 41 | 28 | 24 |
| Max | 89 | 84 | 83 | 82 | 89 | 84 |
| | | | | | | |
| Age Group (years) n (%) n (missing) | 115 (0) | 124 (0) | 113 (0) | 125 (0) | 228 (0) | 249 (0) |
| >=18 - <50 | 15 (13.0) | 18 (14.5) | 19 (16.8) | 10 (8.0) | 34 (14.9) | 28 (11.2) |
| >=50 - <65 | 47 (40.9) | 55 (44.4) | 43 (38.1) | 55 (44.0) | 90 (39.5) | 110 (44.2) |
| >=65 - <75 | 36 (31.3) | 28 (22.6) | 34 (30.1) | 31 (24.8) | 70 (30.7) | 59 (23.7) |
| >=75 | 17 (14.8) | 23 (18.5) | 17 (15.0) | 29 (23.2) | 34 (14.9) | 52 (20.9) |
| <65 | 62 (53.9) | 73 (58.9) | 62 (54.9) | 65 (52.0) | 124 (54.4) | 138 (55.4) |
| >=65 | 53 (46.1) | 51 (41.1) | 51 (45.1) | 60 (48.0) | 104 (45.6) | 111 (44.6) |
| | | | | | | |
| Race n (%) [1] | | | | | | |
| N (missing) | 94 (21) | 96 (28) | 92 (21) | 102 (23) | 186 (42) | 198 (51) |
| American Indian or Alaska Native | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| Asian | 5 (5.3) | 11 (11.5) | 8 (8.7) | 8 (7.8) | 13 (7.0) | 19 (9.6) |
| Black or African American | 4 (4.3) | 1 (1.0) | 4 (4.3) | 3 (2.9) | 8 (4.3) | 4 (2.0) |
| Native Hawaiian or Other Pacific Islander | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| White/Caucasian | 84 (89.4) | 84 (87.5) | 80 (87.0) | 90 (88.2) | 164 (88.2) | 174 (87.9) |
| Other | 1 (1.1) | 0 (0.0) | 0 (0.0) | 1 (1.0) | 1 (0.5) | 1 (0.5) |
| | | | | | | |
| Gender n (%) n (missing) | 115(0) | 124 (0) | 113 (0) | 125 (0) | 228 (0) | 249 (0) |
| Male | 0 (0.0) | 6 (4.8) | 0 (0) | 1 (0.8) | 0 (0.0) | 7 (2.8) |
| Female | 115 (100.0) | 118 (95.2) | 113 (100.0) | 124 (99.2) | 228 (100.0) | 242 (97.2) |
| | | | | | | |
| Ethnicity n (%) n (missing) | 115 (0) | 124 (0) | 113 (0) | 125 (0) | 228 (0) | 249 (0) |
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
25 of 33
{25}
| | Elacestrant | | SOC | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | ESR1-mut | ESR1-mut-nd | ESR1-mut | ESR1-mut-nd | ESR1-mut | ESR1-mut-nd |
| Hispanic or Latino | 10 (8.7) | 9 (7.3) | 10 (8.8) | 8 (6.4) | 20 (8.8) | 17 (6.8) |
| Non-Hispanic or Latino | 92 (80.0) | 102 (82.3) | 88 (77.9) | 102 (81.6) | 180 (78.9) | 204 (81.9) |
| Unknown | 13 (11.3) | 13 (10.5) | 15 (13.3) | 15 (12.0) | 28 (12.3) | 28 (11.2) |
| | | | | | | |
| Height (cm) n (missing) | 113 (2) | 123 (1) | 112 (1) | 124 (1) | 225 (3) | 247 (2) |
| Mean | 161.98 | 162.62 | 160.65 | 161.24 | 161.27 | 161.93 |
| SD | 7.454 | 8.230 | 6.482 | 7.743 | 6.998 | 8.003 |
| Median | 160.00 | 161.00 | 160.40 | 162.00 | 160.30 | 162.00 |
| Min | 143.0 | 144.8 | 145.0 | 142.0 | 143.0 | 142.0 |
| Max | 183.0 | 190.0 | 173.0 | 183.0 | 183.0 | 190.0 |
| | | | | | | |
| Weight (kg) n (missing) | 115 (0) | 124 (0) | 113 (0) | 125 (0) | 228 (0) | 249 (0) |
| Mean | 73.41 | 72.04 | 71.87 | 72.83 | 72.65 | 72.43 |
| SD | 17.145 | 15.092 | 16.455 | 16.443 | 16.787 | 15.758 |
| Median | 69.00 | 70.00 | 69.10 | 72.00 | 69.05 | 70.45 |
| Min | 42.0 | 44.0 | 44.0 | 42.0 | 42.0 | 42.0 |
| Max | 135.0 | 125.7 | 124.0 | 132.3 | 135.0 | 132.3 |
| | | | | | | |
| BMI (kg/m2) n (missing) | 113 (2) | 123 (1) | 112 (1) | 124 (1) | 225 (3) | 247 (2) |
| Mean | 28.07 | 27.13 | 27.88 | 27.95 | 27.97 | 27.55 |
| SD | 6.058 | 4.901 | 6.012 | 5.752 | 6.023 | 5.350 |
| Median | 26.30 | 27.03 | 27.41 | 26.75 | 26.48 | 26.85 |
| Min | 17.5 | 18.2 | 16.9 | 16.5 | 16.9 | 16.5 |
| Max | 52.7 | 40.9 | 45.1 | 47.8 | 52.7 | 47.8 |
| | | | | | | |
| ECOG Performance Status n (%) n (missing) | 115 (0) | 124 (0) | 113 (0) | 125 (0) | 228 (0) | 249 (0) |
| 0 | 67 (58.3) | 76 (61.3) | 62 (54.9) | 73 (58.4) | 129 (56.6) | 149 (59.8) |
| 1 | 48 (41.7) | 48 (38.7) | 51 (45.1) | 51 (40.8) | 99 (43.4) | 99 (39.8) |
| >1 | 0 (0.0) | 0 (0.0) | 0 (0.0) | 1 (0.8) | 0 (0.0) | 1 (0.4) |
ESR1-mut = ESR1-mutated, ESR1-mut-nd = ESR1-mutation-not-detected, SD = Standard Deviation, Min = Minimum, Max = Maximum, BMI = Body Mass Index, ECOG = Eastern Cooperative Oncology Group,
[1] Subjects may select more than 1 race.
Table 19. Clinical Effectiveness Analysis Subgroup Baseline Clinical Characteristics
| | | | SOC | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | | | ESR1-mut-nd |
| Years Since Initial Diagnosis | | | | | | |
| N (missing) | 115 (0) | 124 (0) | 113 (0) | 125 (0) | 228 (0) | 249 (0) |
| Mean | 7.49 | 8.63 | 8.41 | 8.90 | 7.95 | 8.77 |
| SD | 6.527 | 6.372 | 6.985 | 7.742 | 6.759 | 7.080 |
| Median | 4.92 | 6.76 | 5.75 | 6.42 | 5.42 | 6.63 |
| Min | 0.2 | 0.7 | 0.9 | 0.5 | 0.2 | 0.5 |
| Max | 28.4 | 32.2 | 31.0 | 40.1 | 31.0 | 40.1 |
| Stage at Initial Diagnosis | | | | | | |
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
26 of 33
{26}
| | SOC | | | Total | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | | | *ESR1-mut-nd* |
| I | 15 (13.0) | 20 (16.1) | 11 (9.7) | 18 (14.4) | 26 (11.4) | 38 (15.3) |
| II | 27 (23.5) | 53 (42.7) | 39 (34.5) | 42 (33.6) | 66 (28.9) | 95 (38.2) |
| III | 1 (0.9) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 1 (0.4) | 0 (0.0) |
| IIIA | 5 (4.3) | 14 (11.3) | 6 (5.3) | 14 (11.2) | 11 (4.8) | 28 (11.2) |
| IIIB | 4 (3.5) | 3 (2.4) | 0 (0.0) | 3 (2.4) | 4 (1.8) | 6 (2.4) |
| IIIC | 4 (3.5) | 7 (5.6) | 6 (5.3) | 1 (0.8) | 10 (4.4) | 8 (3.2) |
| IV | 42 (36.5) | 20 (16.1) | 38 (33.6) | 38 (30.4) | 80 (35.1) | 58 (23.3) |
| Unknown | 17 (14.8) | 7 (5.6) | 12 (10.6) | 9 (7.2) | 29 (12.7) | 16 (6.4) |
| **T Stage at Initial Diagnosis** | | | | | | |
| T1 | 18 (15.7) | 29 (23.4) | 20 (17.7) | 23 (18.4) | 38 (16.7) | 52 (20.9) |
| T2 | 29 (25.2) | 48 (38.7) | 40 (35.4) | 49 (39.2) | 69 (30.3) | 97 (39.0) |
| T3 | 13 (11.3) | 18 (14.5) | 6 (5.3) | 11 (8.8) | 19 (8.3) | 29 (11.6) |
| T4 | 11 (9.6) | 7 (5.6) | 10 (8.8) | 13 (10.4) | 21 (9.2) | 20 (8.0) |
| Unknown | 3 (2.6) | 2 (1.6) | 5 (4.4) | 2 (1.6) | 8 (3.5) | 4 (1.6) |
| **N Stage at Initial Diagnosis** | | | | | | |
| N0 | 16 (13.9) | 35 (28.2) | 15 (13.3) | 38 (30.4) | 31 (13.6) | 73 (29.3) |
| N1 | 34 (29.6) | 45 (36.3) | 37 (32.7) | 31 (24.8) | 71 (31.1) | 76 (30.5) |
| N2 | 14 (12.2) | 14 (11.3) | 10 (8.8) | 14 (11.2) | 24 (10.5) | 28 (11.2) |
| N3 | 9 (7.8) | 7 (5.6) | 11 (9.7) | 11 (8.8) | 20 (8.8) | 18 (7.2) |
| Unknown | 1 (0.9) | 3 (2.4) | 8 (7.1) | 4 (3.2) | 9 (3.9) | 7 (2.8) |
| **M Stage at Initial Diagnosis** | | | | | | |
| M0 | 41 (35.7) | 82 (66.1) | 51 (45.1) | 67 (53.6) | 92 (40.4) | 149 (59.8) |
| M1 | 27 (23.5) | 15 (12.1) | 26 (23.0) | 27 (21.6) | 53 (23.2) | 42 (16.9) |
| Unknown | 6 (5.2) | 7 (5.6) | 4 (3.5) | 4 (3.2) | 10 (4.4) | 11 (4.4) |
| **Stage at Baseline** | | | | | | |
| IIA | 1 (0.9) | 0 (0.0) | 0 (0.0) | 1 (0.8) | 1 (0.4) | 1 (0.4) |
| IIIA | 0 (0.0) | 2 (1.6) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 2 (0.8) |
| IIIC | 0 (0.0) | 1 (0.8) | 1 (0.9) | 0 (0.0) | 1 (0.4) | 1 (0.4) |
| IV | 8 (7.0) | 4 (3.2) | 7 (6.2) | 11 (8.8) | 15 (6.6) | 15 (6.0) |
| IVA | 1 (0.9) | 2 (1.6) | 2 (1.8) | 1 (0.8) | 3 (1.3) | 3 (1.2) |
| IVB | 1 (0.9) | 2 (1.6) | 1 (0.9) | 2 (1.6) | 2 (0.9) | 4 (1.6) |
| IVC | 0 (0.0) | 1 (0.8) | 1 (0.9) | 0 (0.0) | 1 (0.4) | 1 (0.4) |
| Unknown | 91 (79.1) | 103 (83.1) | 88 (77.9) | 103 (82.4) | 179 (78.5) | 206 (82.7) |
| **T Stage at Baseline** | | | | | | |
| T1 | 2 (1.7) | 6 (4.8) | 2 (1.8) | 3 (2.4) | 4 (1.8) | 9 (3.6) |
| T2 | 6 (5.2) | 7 (5.6) | 8 (7.1) | 7 (5.6) | 14 (6.1) | 14 (5.6) |
| T3 | 3 (2.6) | 3 (2.4) | 0 (0.0) | 4 (3.2) | 3 (1.3) | 7 (2.8) |
| T4 | 8 (7.0) | 4 (3.2) | 4 (3.5) | 7 (5.6) | 12 (5.3) | 11 (4.4) |
| Unknown | 24 (20.9) | 30 (24.2) | 25 (22.1) | 29 (23.2) | 49 (21.5) | 59 (23.7) |
| **N Stage at Baseline** | | | | | | |
| N0 | 8 (7.0) | 6 (4.8) | 3 (2.7) | 9 (7.2) | 11 (4.8) | 15 (6.0) |
| N1 | 4 (3.5) | 10 (8.1) | 6 (5.3) | 7 (5.6) | 10 (4.4) | 17 (6.8) |
| N2 | 4 (3.5) | 3 (2.4) | 3 (2.7) | 4 (3.2) | 7 (3.1) | 7 (2.8) |
| N3 | 3 (2.6) | 3 (2.4) | 1 (0.9) | 4 (3.2) | 4 (1.8) | 7 (2.8) |
| Unknown | 24 (20.9) | 28 (22.6) | 27 (23.9) | 27 (21.6) | 51 (22.4) | 55 (22.1) |
| **M Stage at Baseline** | | | | | | |
| M0 | 3 (2.6) | 5 (4.0) | 0 (0.0) | 6 (4.8) | 3 (1.3) | 11 (4.4) |
| M1 | 27 (23.5) | 33 (26.6) | 25 (22.1) | 37 (29.6) | 52 (22.8) | 70 (28.1) |
| Unknown | 13 (11.3) | 13 (10.5) | 14 (12.4) | 10 (8.0) | 27 (11.8) | 23 (9.2) |
| **Sites of Disease** | | | | | | |
| Breast | 24 (20.9) | 15 (12.1) | 21 (18.6) | 28 (22.4) | 45 (19.7) | 43 (17.3) |
| Bone | 101 (87.8) | 91 (73.4) | 93 (82.3) | 91 (72.8) | 194 (85.1) | 182 (73.1) |
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
27 of 33
{27}
| | Elacestrant | | SOC | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | ESR1-mut | ESR1-mut-nd | ESR1-mut | ESR1-mut-nd | ESR1-mut | ESR1-mut-nd |
| Bone only | 14 (12.2) | 24 (19.4) | 14 (12.4) | 15 (12.0) | 28 (12.3) | 39 (15.7) |
| Lymph Nodes | 34 (29.6) | 34 (27.4) | 27 (23.9) | 41 (32.8) | 61 (26.8) | 75 (30.1) |
| Visceral [1] | 81 (70.4) | 82 (66.1) | 83 (73.5) | 85 (68.0) | 164 (71.9) | 167 (67.1) |
| Brain | 3 (2.6) | 1 (0.8) | 2 (1.8) | 1 (0.8) | 5 (2.2) | 2 (0.8) |
| Liver | 60 (52.2) | 62 (50.0) | 64 (56.6) | 49 (39.2) | 124 (54.4) | 111 (44.6) |
| Lung | 27 (23.5) | 29 (23.4) | 31 (27.4) | 37 (29.6) | 58 (25.4) | 66 (26.5) |
| Other Sites | 26 (22.6) | 21 (16.9) | 18 (15.9) | 30 (24.0) | 44 (19.3) | 51 (20.5) |
| Abdominal Cavity | 2 (1.7) | 3 (2.4) | 1 (0.9) | 4 (3.2) | 3 (1.3) | 7 (2.8) |
| Adrenal Gland | 5 (4.3) | 3 (2.4) | 5 (4.4) | 4 (3.2) | 10 (4.4) | 7 (2.8) |
| Cervix Uteri | 0 (0.0) | 0 (0.0) | 1 (0.9) | 0 (0.0) | 1 (0.4) | 0 (0.0) |
| Chest Wall | 0 (0.0) | 3 (2.4) | 3 (2.7) | 8 (6.4) | 3 (1.3) | 11 (4.4) |
| Esophagus | 0 (0.0) | 1 (0.8) | 0 (0.0) | 1 (0.8) | 0 (0.0) | 2 (0.8) |
| Head And Neck | 1 (0.9) | 1 (0.8) | 0 (0.0) | 0 (0.0) | 1 (0.4) | 1 (0.4) |
| Intestine | 0 (0.0) | 0 (0.0) | 0 (0.0) | 1 (0.8) | 0 (0.0) | 1 (0.4) |
| Kidney | 1 (0.9) | 0 (0.0) | 0 (0.0) | 2 (1.6) | 1 (0.4) | 2 (0.8) |
| Mediastinum | 6 (5.2) | 3 (2.4) | 1 (0.9) | 2 (1.6) | 7 (3.1) | 5 (2.0) |
| Other | 1 (0.9) | 4 (3.2) | 1 (0.9) | 4 (3.2) | 2 (0.9) | 8 (3.2) |
| Pancreas | 1 (0.9) | 2 (1.6) | 0 (0.0) | 1 (0.8) | 1 (0.4) | 3 (1.2) |
| Pericardium | 1 (0.9) | 0 (0.0) | 0 (0.0) | 1 (0.8) | 1 (0.4) | 1 (0.4) |
| Skin | 6 (5.2) | 3 (2.4) | 5 (4.4) | 4 (3.2) | 11 (4.8) | 7 (2.8) |
| Soft Tissue | 5 (4.3) | 1 (0.8) | 3 (2.7) | 2 (1.6) | 8 (3.5) | 3 (1.2) |
| Spleen | 1 (0.9) | 1 (0.8) | 0 (0.0) | 0 (0.0) | 1 (0.4) | 1 (0.4) |
| Stomach | 1 (0.9) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 1 (0.4) | 0 (0.0) |
| Thyroid Gland | 0 (0.0) | 0 (0.0) | 0 (0.0) | 1 (0.8) | 0 (0.0) | 1 (0.4) |
| Number of Metastatic Sites | | | | | | |
| 0 | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| 1 | 16 (13.9) | 35 (28.2) | 19 (16.8) | 27 (21.6) | 35 (15.4) | 62 (24.9) |
| 2 | 43 (37.4) | 31 (25.0) | 34 (30.1) | 38 (30.4) | 77 (33.8) | 69 (27.7) |
| >=3 | 44 (38.3) | 34 (27.4) | 41 (36.3) | 41 (32.8) | 85 (37.3) | 75 (30.1) |
[1] Includes lung, liver, brain, pleural, and peritoneal involvement
### D. Safety and Effectiveness Results
#### 1. Safety Results
Data regarding the safety of ORSERDU (elacestrant) therapy are presented in the original drug approval. Refer to the ORSERDU (elacestrant) label for more information. No adverse events were reported in the conduct of the diagnostic studies used to support this sPMA.
For the specific adverse events that occurred in the clinical studies, please see the ORSERDU (elacestrant) FDA approved package insert which is available at Drugs@FDA.
#### 2. Effectiveness Results
PFS in Patients Positive by Guardant360 CDx for ESR1 Missense Mutations
To demonstrate the clinical validity of Guardant360 CDx for the identification of patients with ER+/HER2- mBC with ESR1 missense mutations for treatment with
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
28 of 33
{28}
ORSERDU (elacestrant), the primary diagnostic study objective was assessed by comparing the efficacy of ORSERDU (elacestrant) relative to SOC in patients positive for ESR1 missense mutations by Guardant360 CDx.
The PFS hazard ratio (HR) observed in the ESR1-mut population treated with ORSERDU (elacestrant) vs. SOC was 0.55 (95% CI 0.39 – 0.77, p=0.0005, Figure 2). Similar results were seen in the sensitivity analysis using an unstratified Cox Proportional Hazard model with an observed HR of 0.53 (95% CI 0.38 - 0.74, p=0.0002). Demonstration of clinical efficacy in the ESR1-mut population is further supported by clear separation of the treatment arms in the Kaplan-Meir plot of PFS (Figure 2). The median PFS in the ESR1-mut population treated with ORSERDU (elacestrant) was 3.78 months (95% CI 2.17 – 7.26) vs. 1.87 months for SOC (95% CI 1.87 – 2.14) (Table 20).
Figure 2. Progression-Free Survival for Elacestrant versus SOC in ESR1-mut Subjects

PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
29 of 33
{29}
**Table 20. Efficacy Results for EMERALD (Patients with *ESR1* Missense Mutations)**
| | ORSERDU (N = 115) | SOC (Fulvestrant or an Aromatase Inhibitor) (N=113) |
| --- | --- | --- |
| **Progression-free Survival (PFS)^{a}** | | |
| Number of PFS Events, n (%) | 62 (53.9) | 78 (69.0) |
| Median PFS months^{b} (95% CI) | 3.78 (2.17, 7.26) | 1.87 (1.87, 2.14) |
| Hazard ratio^{c} (95% CI) | 0.55 (0.39, 0.77) | |
| p-value^{d} (stratified log-rank) | 0.0005 | |
| **Overall Survival (OS)** | | |
| Number of OS Events, n (%) | 61 (53.0) | 60 (53.1) |
| Hazard ratio^{c} (95% CI) | 0.90 (0.63, 1.30) | |
| p-value^{d} (stratified log-rank) | NS^{e} | |
$^{a}$ PFS results based on blinded imaging review committee.
$^{b}$ Kaplan-Meier estimate; 95% CI based on the Brookmeyer-Crowley method using a linear transformation.
$^{c}$ Cox proportional hazards model stratified by prior treatment with fulvestrant (yes vs no) and visceral metastasis (yes vs no).
$^{d}$ Stratified log-rank test two-sided p-value.
$^{e}$ NS – Not statistically significant.
### 3. Pediatric Extrapolation
In this premarket approval application, existing clinical data was not leveraged to support approval of a pediatric patient population.
### E. Financial Disclosure
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included one investigator who was a full-time of the sponsor and had disclosable financial interests/arrangements as defined in 21 CFR 54.2(a), (b), (c) and (f) and described below:
- Compensation to the investigator for conducting the study where the value could be influenced by the outcome of the study: [0]
- Significant payment of other sorts: [0]
- Proprietary interest in the product tested held by the investigator: [0]
- Significant equity interest held by investigator in sponsor of covered study: [1]
The applicant has adequately disclosed the financial interest/arrangements with clinical investigators. Statistical analyses were conducted by FDA to determine whether the financial interests/arrangements had any impact on the clinical study outcome. The information provided does not raise any questions about the reliability of the data.
PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
30 of 33
{30}
## **XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION**
In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA supplement was not referred to the Molecular and Clinical Genetics Panel, an FDA advisory committee, for review and recommendation.
## **XII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES**
### **A. Effectiveness Conclusions**
For the intended use of identifying patients with BC with *ESR1* missense mutations for treatment with ORSERDU (elacestrant), the effectiveness of Guardant360 CDx was demonstrated through analytical studies using patient samples from the intended use population and a diagnostic study utilizing Guardant360 CDx results and outcome data from the EMERALD (NCT03778931) clinical study. The data from the analytical validation and clinical studies support the reasonable assurance of safety and effectiveness of Guardant360 CDx when used in accordance with the indications for use. Data from the EMERALD (NCT03778931) clinical study show that patients with BC who had qualifying *ESR1* missense mutations received benefit from treatment with ORSERDU (elacestrant) and support the addition of the CDx indication to Guardant360 CDx.
### **B. Safety Conclusions**
The risks of the device are based on data collected in the analytical studies conducted to support PMA approval as described above. Guardant360 CDx is an *in vitro* diagnostic test, which involves testing of cfDNA extracted from the plasma of whole blood routinely collected as part of the diagnosis and patient care.
Failure of the device to perform as expected or failure to correctly interpret test results may lead to incorrect test results, and subsequently, inappropriate patient management decisions in cancer treatment. Patients with false positive results may undergo treatment with one of the therapies listed in Table 1 of the intended use statement without clinical benefit and may experience adverse reactions associated with the therapy. Patients with false negative results may not be considered for treatment with the indicated therapy. There is also a risk of delayed results, which may lead to delay of treatment with the indicated therapy.
### **C. Benefit-Risk Determination**
The clinical benefit of Guardant360 CDx for the identification of breast cancer patients with *ESR1* missense mutation(s) for treatment with ORSERDU (elacestrant) was demonstrated in a prospective analysis of efficacy and safety data obtained from the EMERALD (NCT03778931) clinical study. The supporting clinical validation analysis demonstrates PFS HR (0.55, 95% CI 0.39 – 0.77) for subjects from the EMERALD clinical study positive by Guardant360 CDx for *ESR1* mutation(s) (*ESR1*-mut) treated with ORSERDU (elacestrant) relative to SOC which provides for a meaningful clinical benefit, given the patient population. These results support the
---PMA P200010/S010: FDA Summary of Safety and Effectiveness Data
31 of 33
{31}
use of Guardant360 CDx as an aid in identification of breast cancer patients with *ESR1* missense mutation(s) for ORSERDU (elacestrant) treatment.
There is a potential risk associated with the use of Guardant360 CDx, for identification of breast cancer patients with *ESR1* missense mutation(s) for treatment with ORSERDU (elacestrant), mainly due to 1) false positives, false negatives, and failure to provide a result and 2) incorrect interpretation of test results by the user. The risks of Guardant360 CDx are associated with the potential mismanagement of patients resulting from false results of the test. Patients who are determined to be false positive by the test may be exposed to a drug that is not beneficial which may lead to adverse events or may have delayed access to treatments that could be more beneficial. A false negative result may prevent a patient from accessing a potentially beneficial drug.
The likelihood of false results was assessed by analytical and clinical validation studies, which partially mitigate the probable risk of the Guardant360 CDx, discussed above.
The clinical and analytical performance of the device included in this submission demonstrate that the assay is expected to perform with acceptable accuracy, mitigating the potential for false results.
# 1. 1. Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
In conclusion, given the above information, the data support that for the indications of the Guardant360 CDx device discussed above, the probable benefits outweigh the probable risks.
# **D. Overall Conclusions**
The data in this application support the reason…