FoundationOne Liquid CDx (F1 Liquid CDx)
P190032S010 · Foundation Medicine, Inc. · PQP · Jun 8, 2023 · Pathology
Device Facts
| Record ID | P190032S010 |
| Device Name | FoundationOne Liquid CDx (F1 Liquid CDx) |
| Applicant | Foundation Medicine, Inc. |
| Product Code | PQP · Pathology |
| Decision Date | Jun 8, 2023 |
| Decision | APPR |
| Device Class | Class 3 |
| Attributes | Software as a Medical Device, Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|
| P190032S010 · Jun 8, 2023 | FoundationOne Liquid CDx (F1 Liquid CDx) | Foundation Medicine, Inc. | BEACON (ARRAY-818-302) clinical trial plasma samples; Commercial clinical archives (residual cfDNA samples) | Retrospective clinical bridging study to demonstrate concordance between the F1LCDx assay and the clinical trial assay (CTA) and to establish clinical validity for identifying mCRC patients eligible for BRAFTOVI (encorafenib) + cetuximab. | Retrospective bridging study; Clinical concordance; Clinical validity; mCRC; BRAF V600E |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|
| BEACON (ARRAY-818-302) Clinical Bridging Study; Retrospective clinical bridging study; Follow-up/Duration: Not applicable (retrospective testing) | Patients with BRAF V600E mutation mCRC who had progressed after 1 or 2 prior regimens; Sample Size: 523 samples (402 from clinical trial, 121 from commercial sources); Number of Sites: Multicenter (BEACON trial sites) | Clinical Trial Assay (CTA) / Tissue-based testing | Overall Survival (OS) and Confirmed Overall Response Rate (ORR) |
Indications for Use
FoundationOne Liquid CDx is a qualitative next generation sequencing based in vitro diagnostic test that uses targeted high throughput hybridization-based capture technology to detect and report substitutions, insertions and deletions (indels) in 311 genes, rearrangements in eight (8) genes, and copy number alterations in three (3) genes. FoundationOne Liquid CDx utilizes circulating cell-free DNA (cfDNA) isolated from plasma derived from anti-coagulated peripheral whole blood of cancer patients collected in FoundationOne Liquid CDx cfDNA blood collection tubes included in the FoundationOne Liquid CDx Blood Sample Collection Kit. The test is intended to be used as a companion diagnostic to identify patients who may benefit from treatment with the targeted therapies listed in Table 1 in accordance with the approved therapeutic product labeling. Additionally, FoundationOne Liquid CDx is intended to provide tumor mutation profiling to be used by qualified health care professionals in accordance with professional guidelines in oncology for patients with solid malignant neoplasms. A negative result from a plasma specimen does not mean that the patient's tumor is negative for genomic findings. Patients who are negative for the mutations listed in Table 1 (see Note for NTRK1/2/3 and ROS1 fusions) should be reflexed to routine biopsy and their tumor mutation status confirmed using an FDA-approved tumor tissue test, if feasible.
Device Story
F1LCDx is a NGS-based IVD test for cfDNA isolated from plasma; uses hybridization-based capture to target 324 cancer-related genes; sequenced on Illumina NovaSeq 6000. Custom analysis pipeline detects substitutions, indels, copy number variants, and rearrangements. Performed as a centralized laboratory service by Foundation Medicine. Output provides genomic alteration reports used by oncologists to identify patients eligible for specific targeted therapies. Benefits include non-invasive tumor profiling for patients where tissue biopsy is unavailable or infeasible; supports clinical decision-making for personalized cancer treatment.
Clinical Evidence
Clinical bridging study using 523 samples (402 CTA+ clinical trial samples, 121 CTA- samples). Concordance between F1LCDx and enrolling CTA showed PPA 87.2% (95% CI: 83.1%, 90.4%) and NPA 97.1% (95% CI: 91.9%, 99.0%). Clinical efficacy (ORR, OS) in BRAF V600E positive mCRC patients treated with encorafenib + cetuximab was bridged from the BEACON trial. Sensitivity analysis confirmed robustness of efficacy estimates. Analytical validation included accuracy, precision, and LoD studies.
Technological Characteristics
NGS-based assay; hybridization-based capture of 324 genes; Illumina NovaSeq 6000 sequencing; cfDNA input ≥20ng; automated library construction and hybrid capture; proprietary analysis pipeline (BWA, SAMtools, Picard); single-site laboratory service.
Indications for Use
Indicated for patients with solid malignant neoplasms to provide tumor mutation profiling and as a companion diagnostic to identify patients eligible for specific targeted therapies (NSCLC, prostate, breast, colorectal, and solid tumors) based on detection of specific genomic alterations (e.g., ALK, EGFR, BRCA1/2, ATM, PIK3CA, MET, ROS1, NTRK1/2/3, BRAF V600E).
Regulatory Classification
Identification
A next generation sequencing (NGS) oncology panel is a device used for the qualitative detection of germline or somatic variants in one or more cancer-related genes. The device is intended to be used on DNA or RNA isolated from human clinical specimens.
Reference Devices
- therascreen BRAF V600E RGQ PCR Kit (P190026)
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
## I. GENERAL INFORMATION
Device Generic Name: Next generation sequencing oncology panel, somatic or germline variant detection systemDevice Trade Name: FoundationOne® Liquid CDx (F1 Liquid CDx)Device Procode: PQPApplicant's Name and Address: Foundation Medicine, Inc.
150 Second Street
Cambridge, MA 02141
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P190032/S010
Date of FDA Notice of Approval: June 8, 2023
The original Premarket Approval (PMA) (190032) for FoundationOne® Liquid CDx (F1LCDx) was approved on August 26, 2020 as a companion diagnostic for *BRCA1* and *BRCA2* alterations in metastatic castration-resistant prostate cancer (mCRPC) patients who may benefit from treatment with RUBRACA® (rucaparib) and *EGFR* activating mutations (Exon 19 deletions and L858R substitution mutation) in patients with advanced and metastatic non-small cell lung cancer (NSCLC) who may benefit from treatment with IRESSA® (gefitinib), TAGRISSO® (osimertinib), and TARCEVA® (erlotinib). Subsequently, additional PMA supplements were approved for expanding the indications for use of F1LCDx since its original approval. See Section VII for more details.
The current supplement was submitted to expand the indication for the FoundationOne Liquid CDx test as a companion diagnostic for the indication listed in the table below.
New Indication Being Sought in this PMA supplement submission.
| Tumor Type | Biomarker(s) Detected | Therapy |
| --- | --- | --- |
| Colorectal cancer (CRC) | *BRAF* V600E alteration | BRAFTOVI® (encorafenib) in combination with cetuximab |
## II. INDICATIONS FOR USE
FoundationOne Liquid CDx is a qualitative next generation sequencing based *in vitro* diagnostic test that uses targeted high throughput hybridization-based capture technology to detect and report substitutions, insertions and deletions (indels) in 311 genes, rearrangements in eight (8) genes, and copy number alterations in three (3) genes. FoundationOne Liquid CDx utilizes circulating cell-free DNA (cfDNA) isolated from
---PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
1 of 33
{1}
plasma derived from anti-coagulated peripheral whole blood of cancer patients collected in FoundationOne Liquid CDx cfDNA blood collection tubes included in the FoundationOne Liquid CDx Blood Sample Collection Kit. The test is intended to be used as a companion diagnostic to identify patients who may benefit from treatment with the targeted therapies listed in Table 1 in accordance with the approved therapeutic product labeling.
Table 1: Companion diagnostic indications
| Tumor Type | Biomarker(s) Detected | Therapy |
| --- | --- | --- |
| Non-small cell lung cancer (NSCLC) | ALK Rearrangements | ALECENSA® (alectinib) |
| | EGFR Exon 19 deletions and EGFR Exon 21 L858R substitution | EGFR tyrosine kinase inhibitors approved by FDA* |
| | EGFR Exon 20 insertions | EXKIVITY® (mobocertinib) |
| | MET single nucleotide variants (SNVs) and indels that lead to MET exon 14 skipping | TABRECTA® (capmatinib) |
| | ROS1 fusions** | ROZLYTREK® (entrectinib) |
| Prostate cancer | BRCA1, BRCA2, and ATM alterations | LYNPARZA® (olaparib) |
| | BRCA1, BRCA2 alterations | RUBRACA® (rucaparib) |
| Breast Cancer | PIK3CA mutations C420R, E542K, E545A, E545D [1635G>T only], E545G, E545K, Q546E, Q546R, H1047L, H1047R, and H1047Y | PIQRAY® (alpelisib) |
| Solid Tumors | NTRK1/2/3 fusions** | ROZLYTREK® (entrectinib) |
| Colorectal cancer (CRC) | BRAF V600E alteration | BRAFTOVI® (encorafenib) in combination with cetuximab |
*For the most current information about the therapeutic products in this group, go to: https://www.fda.gov/medical-devices/in-vitro-diagnostics/list-cleared-or-approved-companion-diagnostic-devices-in-vitro-and-imaging-tools#Group_Labeling
Additionally, FoundationOne Liquid CDx is intended to provide tumor mutation profiling to be used by qualified health care professionals in accordance with professional guidelines in oncology for patients with solid malignant neoplasms.
A negative result from a plasma specimen does not mean that the patient's tumor is negative for genomic findings. Patients who are negative for the mutations listed in Table 1 (see **Note for NTRK1/2/3 and ROS1 fusions) should be reflexed to routine biopsy and
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
2 of 33
{2}
their tumor mutation status confirmed using an FDA-approved tumor tissue test, if feasible.
**Note: when considering eligibility for ROZLYTREK® based on the detection of NTRK1/2/3 and ROS1 fusions, testing using plasma specimens is only appropriate for patients for whom tumor tissue is not available for testing.
Genomic findings other than those listed in Table 1 of the intended use statement are not prescriptive or conclusive for labeled use of any specific therapeutic product.
FoundationOne Liquid CDx is a single-site assay performed at Foundation Medicine, Inc. in Cambridge, MA.
### III. CONTRAINDICATIONS
There are no known contraindications.
### IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the FoundationOne Liquid CDx labeling.
### V. DEVICE DESCRIPTION
The FoundationOne Liquid CDx (F1LCDx) assay is performed exclusively as a laboratory service using circulating cell-free DNA (cfDNA) isolated from plasma derived from anti-coagulated peripheral whole blood from patients with solid malignant neoplasms. The assay employs a single DNA extraction method to obtain cfDNA from plasma from whole blood. Extracted cfDNA undergoes whole-genome shotgun library construction and hybridization-based capture of 324 cancer-related genes. All coding exons of 309 genes are targeted; select intronic or non-coding regions are targeted in three genes (refer to Table 2 for the complete list of genes reported by F1LCDx).
Hybrid-capture selected libraries are sequenced with deep coverage using the NovaSeq 6000 platform. Sequence data are processed using a custom analysis pipeline designed to detect genomic alterations, including base substitutions and indels in 311 genes, copy number variants in three genes, and genomic rearrangements in eight genes. A subset of targeted regions in 75 genes is baited for increased sensitivity.
Table 2: Genomic Regions in which Variants are Reported by F1LCDx¹
| ABL1 [Exons 4-9] | ACVR1B | AKT1 [Exon 3] | AKT2 | AKT3 | ALK [Exons 20-29, Introns 18,19] | ALOX12B | AMER1 (FAM123B) | APC | AR |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| ARAF [Exons 4, 5, 7, 11, 13, 15, 16] | ARFRP1 | ARID1A | ASXL1 | ATM | ATR | ATRX | AURKA | AURKB | AXIN1 |
| AXL | BAP1 | BARD1 | BCL2 | BCL2L1 | BCL2L2 | BCL6 | BCOR | BCORL1 | BCR* [Introns 8, 13, |
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
3 of 33
{3}
| | | | | | | | | | 14] |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| BRAF [Exons 11-18, Introns 7-10] | BRCA1 [Introns 2, 7, 8, 12, 16, 19, 20] | BRCA2 [Intron 2] | BRD4 | BRIP1 | BTG1 | BTG2 | BTK [Exons 2, 15] | C11orf30 (EMSY) | C17orf39 (GID4) |
| CALR | CARD11 | CASP8 | CBFB | CBL | CCND1 | CCND2 | CCND3 | CCNE1 | CD22 |
| CD70 | CD74* [Introns 6-8] | CD79A | CD79B | CD274 (PD-L1) | CDC73 | CDH1 | CDK12 | CDK4 | CDK6 |
| CDK8 | CDKN1A | CDKN1B | CDKN2A | CDKN2B | CDKN2C | CEBPA | CHEK1 | CHEK2 | CIC |
| CREBBP | CRKL | CSF1R | CSF3R | CTCF | CTNNA1 | CTNNB1 [Exon 3] | CUL3 | CUL4A | CXCR4 |
| CYP17A1 | DAXX | DDR1 | DDR2 [Exons 5, 17, 18] | DIS3 | DNMT3A | DOT1L | EED | EGFR [Introns 7, 15, 24-27] | EP300 |
| EPHA3 | EPHB1 | EPHB4 | ERBB2 | ERBB3 [Exons 3, 6, 7, 8, 10, 12, 20, 21, 23, 24, 25] | ERBB4 | ERCC4 | ERG | ERRF11 | ESR1 [Exons 4-8] |
| ETV4* [Intron 8] | ETV5* [Introns 6, 7] | ETV6* [Introns 5, 6] | EWSR1* [Introns 7-13] | EZH2 [Exons 4, 16, 17, 18] | EZR* [Introns 9-11] | FAM46C | FANCA | FANCC | FANCG |
| FANCL | FAS | FBXW7 | FGF10 | FGF12 | FGF14 | FGF19 | FGF23 | FGF3 | FGF4 |
| FGF6 | FGFR1 [Introns 1, 5, Intron 17] | FGFR2 [Intron 1, Intron 17] | FGFR3 [Exons 7, 9 (alternative designation exon 10), 14, 18, Intron 17] | FGFR4 | FH | FLCN | FLT1 | FLT3 [Exons 14, 15, 20] | FOXL2 |
| FUBP1 | GABRA6 | GATA3 | GATA4 | GATA6 | GNA11 [Exons 4, 5] | GNA13 | GNAQ [Exons 4, 5] | GNAS [Exons 1, 8] | GRM3 |
| GSK3B | H3F3A | HDAC1 | HGF | HNF1A | HRAS [Exons 2, 3] | HSD3B1 | ID3 | IDH1 [Exon 4] | IDH2 [Exon 4] |
| IGF1R | IKBKE | IKZF1 | INPP4B | IRF2 | IRF4 | IRS2 | JAK1 | JAK2 [Exon 14] | JAK3 [Exons 5, 11, 12, 13, 15, 16] |
| JUN | KDM5A | KDM5C | KDM6A | KDR | KEAP1 | KEL | KIT [Exons 8, 9, 11, 12, 13, 17, Intron 16] | KLHL6 | KMT2A (MLL) [Introns 6, 8-11, Intron 7] |
| KMT2D (MLL2) | KRAS | LTK | LYN | MAF | MAP2K1 (MEK1) [Exons 2, 3] | MAP2K2 (MEK2) [Exons 2-4, 6, 7] | MAP2K4 | MAP3K1 | MAP3K13 |
| MAPK1 | MCL1 | MDM2 | MDM4 | MED12 | MEF2B | MEN1 | MERTK | MET | MITF |
| MKNK1 | MLH1 | MPL [Exon 10] | MRE11A | MSH2 [Intron 5] | MSH3 | MSH6 | MST1R | MTAP | MTOR [Exons 19, 30, 39, 40, 43-45, 47, 48, 53, 56] |
| MUTYH | MYB* [Intron 14] | MYC [Intron 1] | MYCL (MYCL1) | MYCN | MYD88 [Exon 4] | NBN | NF1 | NF2 | NFE2L2 |
| NFKBIA | NKX2-1 (TTF-1) | NOTCH1 | NOTCH2 [Intron 26] | NOTCH3 | NPM1 [Exons 4-6, 8, 10] | NRAS [Exons 2, 3] | NSD3 (WHSC1L1) | NT5C2 | NTRK1 [Exons 14, 15, Introns 8-11] |
| NTRK2 [Intron 12] | NTRK3 [Exons 16, Intron 1] | NUTM1* [Intron 1] | P2RY8 | PALB2 | PARK2 | PARP1 | PARP2 | PARP3 | PAX5 |
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
4 of 33
{4}
| | 17] | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| PBRM1 | PDCD1 (PD-1) | PDCD1L G2 (PD-L2) | PDGFRA [Exons 12, 18, Introns 7, 9, 11] | PDGFRB [Exons 12-21, 23] | PDK1 | PIK3C2B | PIK3C2G | PIK3CA [Exons 2, 3, 5-8, 10, 14, 19, 21 (Coding Exons 1, 2, 4-7, 9, 13, 18, 20)] | PIK3CB |
| PIK3R1 | PIM1 | PMS2 | POLD1 | POLE | PPARG | PPP2R1A | PPP2R2A | PRDM1 | PRKAR1A |
| PRKCI | PTCH1 | PTEN | PTPN11 | PTPRO | QKI | RAC1 | RAD21 | RAD51 | RAD51B |
| RAD51C | RAD51D | RAD52 | RAD54L | RAF1 [Exons 3, 4, 6, 7, 10, 14, 15, 17, Introns 4-8] | RARA [Intron 2] | RB1 | RBM10 | REL | RET [Introns 7, 8, Exons 11, 13-16, Introns 9-11] |
| RICTOR | RNF43 | ROS1 [Exons 31, 36-38, 40, Introns 31-35] | RPTOR | RSPO2* [Intron 1] | SDC4* [Intron 2] | SDHA | SDHB | SDHC | SDHD |
| SETD2 | SF3B1 | SGK1 | SLC34A2* [Intron 4] | SMAD2 | SMAD4 | SMARCA4 | SMARCB1 | SMO | SNCAIP |
| SOCS1 | SOX2 | SOX9 | SPEN | SPOP | SRC | STAG2 | STAT3 | STK11 (LKB1) | SUFU |
| SYK | TBX3 | TEK | TERC* [ncRNA] | TERT* [Promoter] | TET2 | TGFBR2 | TIPARP | TMPRSS2* [Introns 1-3] | TNFAIP3 |
| TNFRSF14 | TP53 | TSC1 | TSC2 | TYRO3 | U2AF1 | VEGFA | VHL | WHSC1 | WTI |
| XPO1 | XRCC2 | ZNF217 | ZNF703 | | | | | | |
$^{1}$As part of its FDA-approved intended use, the F1LCDx assay interrogates 324 genes, including 309 genes with complete exonic (coding) coverage and 15 genes with only select non-coding coverage (indicated with an *).
The reporting of rearrangements and copy number alterations are restricted to those genes included in Table 3, below.
**Table 3: Genes for which copy number alterations and rearrangements are reported for tumor profiling by F1LCDx**
| Alteration Type | Genes |
| --- | --- |
| Copy Number Alterations | *BRCA1, BRCA2, ERBB2* |
| Rearrangements | *ALK, BRCA1, BRCA2, NTRK1, NTRK2, NTRK3* |
The test report includes variants reported in the following levels:
### Level 1: Companion Diagnostics (CDx)
Clinical evidence should be presented from a prospectively designed clinical trial. Results can also be presented from a retrospective clinical bridging study demonstrating that the clinical endpoints are preserved using plasma samples in trials where enrollment was based on tissue test results. For follow-on markers, a clinical concordance study demonstrating non-inferiority to the original FDA-approved cfDNA-based companion diagnostic device (refer to Li, Meijuan. Statistical Methods for Clinical Validation of Follow-On Companion Diagnostic Devices via an External Concordance Study. Statistics in Biopharmaceutical Research. 8: 35-363, 2016) is required. In addition to the clinical
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
5 of 33
{5}
validation, analytical validation for each specific Level 1 CDx biomarker should be presented.
**Level 2: cfDNA Biomarkers with Strong Evidence of Clinical Significance in cfDNA**
For a Level 2 claim of cfDNA biomarkers with strong evidence of clinical significance, clinical validation needs to be from evidence presented with FDA-approved liquid biopsy companion diagnostic biomarkers for the specific tumor type at the biomarker or variant level. Such claims should also be supported by analytical performance for each biomarker from at least limit of detection (LoD), precision/ reproducibility, and accuracy studies.
**Level 3A: Biomarkers with Evidence of Clinical Significance in Tissue Supported by Strong Analytical Validation Using cfDNA and Concordance Between cfDNA and Tissue.**
Clinical evidence can be provided from tissue-based companion diagnostics. This should also be supported by analytical validation (LoD, precision, analytical accuracy, and concordance study to a tissue-based test) for the specific tumor type at the biomarker or variant level, using a representative approach for SNVs and indels. Evidence evaluating concordance between cfDNA- and tissue-samples for FDA-approved tissue markers should be demonstrated using an FDA-approved tissue test or a validated tissue test.
**Level 3B: Biomarkers with Evidence of Clinical Significance in Tissue Supported by Analytical Validation Using cfDNA**
Clinical evidence can be provided from tissue-based companion diagnostics, with analytical validation supported by a representative approach for SNVs and indels from key analytical studies (such as LoD, accuracy, and precision).
**Level 4: Other Biomarkers with Potential Clinical Significance**
Biomarkers not categorized into Levels 1, 2, or 3 can be included under Level 4 for informational purposes or to be used to direct patients toward clinical trials for which they may be eligible. Such claims can be supported by clinical rationale for inclusion in the panel. Such rationale could also include peer-reviewed publications for genes/ variants in tissue, variant information from well curated public databases, or *in vitro* pre-clinical models. Analytical validation should be supported by a representative approach for SNVs and indels from key analytical studies (such as LoD, accuracy, and precision).
**FoundationOne® Liquid CDx cfDNA Blood Specimen Collection Kit Contents**
The test includes a blood specimen collection kit, which is sent to ordering laboratories. The shipping kit contains the following components:
- Specimen preparation and shipping instructions
- Two FoundationOne® Liquid CDx cfDNA Blood Collection Tubes (8.5 mL nominal fill volume per tube)
- Return shipping label
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
6 of 33
{6}
# Instruments
The F1LCDx assay is intended to be performed with the serial number-controlled instruments indicated in Table 4, below. All instruments are qualified by Foundation Medicine, Inc. (Foundation Medicine or FMI) under Foundation Medicine's Quality System.
Table 4: Instruments for use with the F1LCDx assay
| Instrument |
| --- |
| Illumina NovaSeq 6000 |
| Thermo Scientific Kingfisher Flex DW 96 |
| Hamilton STARlet STAR Liquid Handling Workstation |
# Test Process
All assay reagents including blood collection tubes included in the F1LCDx assay process are qualified by Foundation Medicine and are compliant with the medical device Quality System Regulation (QSR).
# A. Specimen Collection and Preparation
Whole blood specimens are collected in F1LCDx cfDNA Blood Collection Tubes (BCT) provided as a component of the F1LCDx specimen collection kit. Prior to cfDNA isolation, the plasma is collected from whole blood by centrifugation, which separates the plasma from the buffy coat (white blood cells) and red blood cells. The plasma layer is removed from the buffy coat to avoid contamination of cellular DNA into the plasma sample. A residual volume of plasma remains in the tube to avoid disturbing the buffy coat. A second spin of the separated plasma at high-speed further pellets cell debris and protein.
# B. DNA Extraction
Following the separation of plasma from whole blood, cfDNA is isolated from plasma using the KingFisher Flex Magnetic Particle Processor, which uses an efficient and automated method to purify cfDNA. The KingFisher Instrument uses magnetic rods to move nucleic acid through purification phases of binding, washing, and elution to yield high purity cfDNA. After isolating cfDNA, the Agilent 4200 TapeStation is used to quantify cfDNA.
# C. Library Construction
Library Construction (LC) begins with the normalization of cfDNA. The samples are purified, using AMPure XP Beads (Agencourt). Solid-phase reversible immobilization (SPRI) purification is used subsequent to library construction with the NEBNext kits (NEB), including mixes for end repair with blunt-end and 5'- phosphorylate the cfDNA fragments using T4 Polynucleotide Kinase and T4 DNA Polymerase. This step prepares the 3'-end for dA-addition while also preparing the 5'-end of the DNA fragment for ligation. Second, dA-addition will incorporate a single dAMP to the 3'-end of the End-Repaired material. After dA-addition, a universal Y-adaptor is ligated
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
7 of 33
{7}
onto each end of the DNA fragment using a DNA ligase. These steps are performed in 96-well plates (Eppendorf) on a liquid handling workstation (Hamilton STAR) using the “with-bead” protocol to maximize reproducibility and library yield. Dual-indexed (Foundation Medicine customized six base pair barcodes) sequencing libraries are PCR amplified with a high-fidelity DNA polymerase (HiFi™, Kapa) for ten cycles, SPRI purified and quantified by PicoGreen fluorescence assay (Invitrogen). Process matched control (PMC) is prepared and added to the plate with other cfDNA samples at the beginning of LC.
On May 25, 2022, F1LCDx was approved for a change in the design of the primers used for LC. Specifically, a change was made to the regions of the primer that hybridize to the universal Y-adaptors during the PCR amplification process, which resulted in increased amplification efficiency and reduced the minimum recommended cfDNA input level to 20ng.
### D. Hybrid Capture
Hybrid Capture (HC) begins with the normalization of each library from 500 ng to 2000 ng. Solution hybridization is performed using a >50-fold molar excess of a pool of individually synthesized 5'-biotinylated DNA 120 base pair oligonucleotides (Integrated DNA Technology) for baits. The baits target regions from 324 cancer-related genes including all coding exons of 309 genes and only select introns or non-coding regions in 15 genes. Baits were designed by appointing overlapping 120 bp DNA sequence intervals covering target exons (60 bp overlap) and introns (20 bp overlap), with a minimum of three baits per target; single nucleotide polymorphism (SNP) targets were allocated one bait each. Intronic baits were filtered for repetitive elements as defined by the University of California at Santa Cruz (UCSC) Genome Repeat Masker track. Hybrid selection of targets demonstrating reproducibly low coverage was boosted by increasing the number of baits for these targets.
Upon completion of the pre-capture normalization, blocking DNA (adaptor block, Cot, Salmon Sperm DNA) is added to the sequencing library and the mixture is lyophilized in a 96-well plate. The library is then re-suspended in nuclease-free water, heat denatured at 95°C for 5 minutes, temperature ramps from 95°C to 68°C to anneal blocking DNA, and then the samples are incubated at 68°C for a minimum of 5 minutes before the addition of the bait set reagent. After a 20-24-hour incubation, the library-bait duplexes are captured on paramagnetic MyOne™ streptavidin beads (Invitrogen) and off-target library is removed by washing one time with Saline Sodium Citrate (SSC) at 25°C and four times with SSC at 55°C. The PCR master mix is added to directly amplify the captured library from the washed beads. After amplification, the samples are SPRI purified and quantified by PicoGreen.
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
8 of 33
{8}
## E. Sequencing
Sequencing on the Illumina NovaSeq 6000 platform employs on-board cluster generation (OBCG) using patterned flow cell (FC) technology to generate monoclonal clusters via ExAmp from a single DNA template. The clusters are then sequenced using sequencing by synthesis (SBS) chemistry. The NovaSeq system is capable of sequencing up to two flow cells at a time. During OBCG, a single DNA template is introduced into each of the primer substrate layered nanowells of the flow cell, where the template is immediately and rapidly amplified by ExAmp. This rapid amplification prevents other DNA templates from binding, ensuring a monoclonal cluster is formed in each nanowell. The procedure allows for fixed size and spacing of the clusters which results in improved and more accurate resolution.
A growing nucleotide chain is created on the flow cell by incorporating fluorescently labeled, 3'-blocked deoxynucleoside triphosphates (dNTPs). After excitation by a laser, the camera captures the emission color of the incorporated, fluorescently labeled nucleotide. The 3'-block is then removed, reverting the nucleotide to its natural form, which allows the polymerase to add another base to the growing double strand of DNA. With each successive SBS cycle, a new fluorescently labeled 3'- blocked dNTP is added. SBS allows for millions of discrete clusters of clonal copies of DNA to be sequenced in parallel.
## F. Sequence Analysis
Sequence data are analyzed using mainly proprietary software developed by Foundation Medicine. External tools used include: 1) BWA (Burrows-Wheeler Aligner) v0.7.17, for aligning sequence reads to the genomic reference, 2) SAMtools v1.6 for utility operations, 3) Picard tools v1.56 for metrics calculations, and 4) Biopython for the pairwise2 sequence alignment module.
Reads from each Illumina flow cell are demultiplexed (sorted into sets of reads deriving from distinct samples), and their fragment barcodes (FBCs) are extracted and encoded into the read names. For each sample, read pairs with matching, valid FBCs are aligned and processed together to: 1) identify clusters of reads originating from the same original fragment; 2) merge overlapping read pairs into single reads, where possible; and 3) generate consensus reads representing all information in the set of reads for each cluster, encoding positions with mismatches (errors) with base quality 20. The consensus reads are then aligned to the reference genome to generate the 'consensus' binary alignment map (BAM).
For the detection of short variants (e.g., substitutions and small indels) in each target region of interest, a de novo assembly is performed. This is done using proprietary software to generate a de Bruijn graph including all k-mers in reads mapping to a particular locus. The graph is parsed to identify paths that
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
9 of 33
{9}
originate and terminate in reference nodes from the locus. Increased k-mer sizes may be used to account for ambiguities, cycles, and other problematic regions within the graph. The result of the graph traversal is a set of candidate variants. For each variant, there is a set of k-mers supporting the variant and a set of k-mers that would support the reference or another variant at the location.
Each candidate variant is then scanned against reads in the locus to identify which reads support either the candidate variant or a different variant or reference at the location. The cluster membership of the supporting reads is then assessed to determine which clusters show unambiguous support for the variant and which have conflicting assignments, indicating that the variant may have arisen as an error in sequencing or library preparation. The final variant calls are made based on a model that takes into account the coverage at the location, the number of supporting read clusters and their redundancy level, and the number of error-containing clusters.
### G. Report Generation
Approved results are annotated by automated software with CDx relevant information and are merged with patient demographic information and any additional information provided by Foundation Medicine as a professional service prior to approval and release by the laboratory director or designee.
### H. Internal Process Controls
#### Process Control
Each assay run includes a control sample run in duplicate. The control sample contains a pool of eleven HapMap cell lines and is used as a positive mutation detection control. Hundred different germline SNPs present across the entire targeted region are required to be detected by the analysis pipeline.
#### Sensitivity Control
The HapMap control pool used as the positive control is prepared to contain variants at 0.1%, 10% mutant allele frequency (MAF) which must be detected by the analysis pipeline to ensure expected sensitivity for each run.
#### Negative Control
Samples are barcoded molecularly at the library construction (LC) stage. Only reads with a perfect molecular barcode sequence are incorporated into the analysis. The Analysis Pipeline includes an algorithm that analyzes the SNP profile of each specimen to identify potential contamination that may have occurred prior to molecular barcoding.
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
10 of 33
{10}
# I. Classification Criteria for CDx Biomarkers Detected by F1LCDx
1. BRCA1 and BRCA2 alterations to identify patients eligible for rucaparib in prostate cancer:
The CDx classification criteria and the list of BRCA1/BRCA2 missense mutations for rucaparib, based on the trial prespecifications are described in Table 5 and Table 6; however, not all the missense mutations listed below were observed in the TRITON2, and PROfound clinical studies.
Table 5: Classification Criteria for Deleterious Tumor BRCA Variants
| Qualification Criteria | Sequence Classification | Methodology |
| --- | --- | --- |
| A BRCA1 or BRCA2 alteration that includes any of the sequence classifications | Protein truncating mutations | Sequence analysis identifies premature stop codons anywhere in the gene coding region, except, 3' of and including BRCA2 K3326* |
| | Splice site mutations | Sequence analysis identifies variant splice sequences at intron/exon junctions -/+ 2bp of exon starts/ends |
| | Homozygous deletions | Sequence analysis identifies deletions in both gene alleles of ≥1 exon in size |
| | Large protein truncating rearrangements | Sequence analysis identifies protein truncating rearrangements |
| | Deleterious missense mutations | Curated list |
Table 6: Deleterious BRCA Missense Alterations in rucaparib
| BRCA1 Alterations (Protein Change) | | | | | BRCA2 Alterations (Protein Change) | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| M1V | C44Y | R71T | R1699W | G1770V | M1V | R2336P | T2722R |
| M1T | C44F | R71M | R1699Q | M1775K | M1T | R2336L | D2723H |
| M1R | C47S | S770L | G1706R | M1775R | M1R | R2336H | D2723G |
| M1I | C47Y | R1495T | G1706E | C1787S | M1I | T2412I | G2724W |
| M18T | C47F | R1495M | A1708E | G1788V | D23N | R2602T | G2748D |
| L22S | C61S | R1495K | S1715R | P1812A | D23Y | W2626C | A2911E |
| I26N | C61G | E1559K | S1722F | A1823T | S142N | I2627F | E3002K |
| T37K | C61Y | E1559Q | V1736A | V1833M | S142I | R2659T | R3052W |
| C39R | C64R | T1685A | G1738R | W1837R | V159M | R2659K | D3095G |
| C39G | C64G | T1685I | G1738E | V1838E | V211I | E2663V | D3095E |
| C39Y | C64Y | D1692N | K1759N | | V211L | S2670L | N3124I |
| C39W | C64W | M1689R | L1764P | | Y600C | I2675V | N3187K |
| H41R | R71G | D1692H | I1766N | | K1530N | T2722K | |
| C44S | R71K | D1692Y | I1766S | | | | |
2. ATM, BRCA1 and BRCA2 alterations to identify patients eligible for olaparib in mCRPC:
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
11 of 33
{11}
**Table 7: Rules Applied to the Aforementioned Genes:**
| Qualification Criteria | Sequence Classification | Methodology | Comments |
| --- | --- | --- | --- |
| A gene alteration that includes any of the sequence classifications | Protein truncating mutations | Sequence analysis identifies premature stop codons anywhere in the gene coding region, except 3' of and including *BRCA2* K3326* | Does not include VUS. Includes mutations on the canonical transcript only for genes *ATM*, *BRCA1*, and *BRCA2*. |
| | Splice site mutations | Sequence analysis identifies variant splice sequences at intron/exon junctions -/+ 2bp of exon starts/ends | Does not include VUS. Includes indels that extend through ±2bp from the intron/exon junction. Includes mutations on the canonical transcript only for genes *ATM*, *BRCA1*, and *BRCA2*. |
| | Homozygous deletions | Sequence analysis identifies deletions in both gene alleles of ≥1 exon in size | Does not include VUS Only reported for *BRCA1* and *BRCA2*. Not reported for *ATM*. |
| | Large protein truncating rearrangements | Sequence analysis identifies protein truncating rearrangements | Does not include VUS |
| | Deleterious missense mutations | Curated list | Protein effects from list of missense mutations on the canonical transcript only for genes *ATM*, *BRCA1*, and *BRCA2*. |
Alterations reported are limited to those within the alteration-calling capabilities of FMI as of March 2, 2020. ATM missense mutations were identified from the ClinVar database. Should the calling capabilities expand, additional alterations that meet the above criteria may also be reported, per FDA approval.
**Table 8. List of Deleterious Missense Mutations by Protein Effect, Implemented on the Respective Canonical Transcript.**
| *BRCA1* | | *BRCA2* | | *ATM* | |
| --- | --- | --- | --- | --- | --- |
| Protein Effect (PE) | FMI Annotated PE | Protein Effect (PE) | FMI Annotated PE | Protein Effect (PE) | FMI Annotated PE |
| MIV | MIV | MIR | MIR | MIT | MIT |
| MII | MII | MII | MII | R2032K | R2032K |
| C6IG | C6IG | VI59M | VI59M | R2227C | R2227C |
| C64Y | C64Y | V211L | V211L | R2547 S2549del | R2547 S2549del |
| R7IG | R7IG | V211I | V211I | G2765S | G2765S |
| R7IK | R7IK | R2336P | R2336P | R2832C | R2832C |
| RI495M | RI495M | R2336H | R2336H | S2855 | S2855 |
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
12 of 33
{12}
| *BRCA1* | | *BRCA2* | | *ATM* | |
| --- | --- | --- | --- | --- | --- |
| Protein Effect (PE) | FMI Annotated PE | Protein Effect (PE) | FMI Annotated PE | Protein Effect (PE) | FMI Annotated PE |
| | | | | V2856delinsR1 | V2856delinsR1 S2855 V2856>R1 |
| EI559K | EI559K | | | R3008C | R3008C |
| DI692N | DI692N | | | R3008H | R3008H |
| DI692H | DI692H | | | [VUS from Jan 2016 HRR* List to be Excluded] | |
| RI699W | RI699W | | | V2424G | V2424G |
| AI708E | AI708E | | | [Excluded from Jan 2016 HRR List] | |
| GI788V | GI788V | | | K750K | splice site 2250G>A |
HRR = Homologous Recombination Repair genes
### Intronic Variants
| Gene | Chr | Position | Ref | Alt | dbSNP | FMI Protein Effect |
| --- | --- | --- | --- | --- | --- | --- |
| *ATM* | *chr11* | *108128198* | *T* | *G* | *rs730881346* | **[Variant Not Called by FMI]** |
| *ATM* | *chr11* | *108214102* | AGTGA | *A* | *rs730881295* | splice site 8418+5_8418+8delGTGA or splice site 8418+1_8418+4delGTGA |
3. *CDx classification criteria for EGFR alterations:*
- Base substitutions resulting in *EGFR* L858R
- In-frame deletions occurring within *EGFR* Exon 19
4. *ALK rearrangements to identify patients eligible for treatment with ALECENSA® (alectinib):*
CDx positivity for an *ALK* rearrangement is based on the following variant classification criteria:
- The *ALK* rearrangement must have pathogenic driver status (FMI driver status of "known" or "likely")
- AND the disease type must be NSCLC
- AND one of the following two conditions must hold:
1. The partner gene is *EML4*, or
2. The *ALK* breakpoint occurs within *ALK* intron 19
5. *SNVs and indels that lead to MET exon 14 skipping to identify patients eligible for treatment with TABRECTA® (capmatinib):*
A SNV or indel in *MET* shall be considered to result in skipping of exon 14 if one or more of the following criteria are met:
1. Deletions greater than or equal to 5 bp that affect positions -3 to -30 in the intronic region immediately adjacent to the splice acceptor site at the 5' boundary of *MET* exon 14.
2. Indels affecting positions -1 or -2 at the splice acceptor site of the 5' boundary of *MET* exon 14.
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
13 of 33
{13}
3. Base substitutions and indels affecting positions 0, +1, +2, or +3 at the splice donor site of the 3' boundary of MET exon 14.
6. Biomarker Rules for Rearrangements that Lead to NTRK1, NTRK2, or NTRK3 Fusions:
Rearrangements in NTRK1, NTRK2, or NTRK3 shall be considered CDx biomarker positive, that is, to lead to a NTRK1, NTRK2, or NTRK3 RNA fusion, if the following criterion is met:
- In-strand rearrangement events that may lead to an NTRK1, NTRK2 or NTRK3 RNA fusion with a previously reported or novel partner gene in which the kinase domain is not disrupted. This also includes rearrangement events that result in reciprocal fusions (NTRK may be on either the 5' or the 3' end of the detected fusion).
In this regard out-of-strand events are considered as non-fusion rearrangements and are classified as CDx biomarker negative. Intragenic fusions in which genomic rearrangement events are wholly internal to the NTRK1, NTRK2, or NTRK3 genes (i.e., NTRK1-NTRK1, NTRK2-NTRK2, NTRK3-NTRK3 events) are also considered biomarker negative. Unidentified partners (encoded as N/A) or LINC non-coding partners are also considered CDx biomarker negative.
7. Biomarker Rules for Rearrangements that Lead to ROS1 Fusions:
Rearrangements in ROS1 shall be considered CDx biomarker positive, i.e., to lead to ROS1 RNA fusion, if the following condition is met:
- In-strand rearrangement events that may lead to a ROS1 RNA fusion with another protein coding gene in which the ROS1 kinase domain is not disrupted. ROS1 must be on the 3' end of the detected fusion.
In this regard, out-of-strand events are considered as non-fusion rearrangements and are classified as CDx biomarker negative. Intragenic fusions in which genomic rearrangement events are wholly internal to the ROS1 (i.e., ROS1-ROS1 events) are also considered biomarker negative. Unidentified partners (encoded as N/A) or LINC non-coding partners are also considered CDx biomarker negative. ROS1 fusions with novel partners are required to be in frame.
8. EGFR exon 20 insertions to identify NSCLC patients eligible for treatment with EXKIVITY:
CDx positivity for EGFR exon 20 insertions is determined if the following criteria were met:
- Any in-frame insertions affecting amino acids 762 – 775 (inclusive) in EGFR exon 20
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
14 of 33
{14}
9. *BRAF V600E* alteration to identify metastatic CRC patients eligible for treatment with *BRAFTOVI* (encorafenib) in combination with cetuximab:
*CDx classification criteria for BRAF alterations:*
• Base alterations resulting in *BRAF* V600E
## VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are no FDA-approved CDx alternatives using cfDNA isolated from plasma for the detection of *BRAF* V600E alterations to identify patients with mCRC eligible for treatment with *BRAFTOVI* (encorafenib) in combination with cetuximab.
However, *therascreen* *BRAF* V600E RGQ PCR Kit (QIAGEN GmbH) is an FDA-approved CDx for the detection of *BRAF* V600E mutation in mCRC to identify patients eligible for *BRAFTOVI* (encorafenib) in combination with cetuximab with formalin-fixed, paraffin-embedded (FFPE) tissue specimens (P190026).
There are FDA-approved alternatives for the detection of select CDx and tumor profiling genetic alterations using either cfDNA isolated from plasma samples or FFPE tissue specimens. For additional details see FDA List of Cleared or Approved Companion Diagnostic Devices at: https://www.fda.gov/media/119249/download. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
## VII. MARKETING HISTORY
The F1LCDx assay was FDA-approved on August 26, 2020, and subsequently commercialized in the United States. The F1LCDx assay has been marketed in the United States, the European Union, and in several other foreign countries since the approval. On September 21, 2022, the companion diagnostic indication for F1LCDx to identify patients with ovarian cancer harboring *BRCA1* or *BRCA2* alterations for treatment with RUBRACA® (rucaparib) was removed. The approved PMA supplements that affected the intended use are listed in Table 9.
**Table 9. Marketing History**
| Submission No. | Date of Approval | Biomarker/Update | Indication | Drug |
| --- | --- | --- | --- | --- |
| P200006 | October 26, 2020 | *ALK* Rearrangements | NSCLC | ALECENSA® (alectinib) |
| | | *PIK3CA* alterations | Breast Cancer | PIQRAY® (alpelisib) |
| P200016 | November 6, 2020 | *BRCA1*, *BRCA2*, and *ATM* alterations | Prostate Cancer | LYNPARZA® (olaparib) |
| P190032/S001 | July 15, 2021 | *MET* single nucleotide variants (SNVs) and indels that lead to *MET* exon 14 skipping | NSCLC | TABRECTA® (capmatinib) |
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
15 of 33
{15}
| Submission No. | Date of Approval | Biomarker/Update | Indication | Drug |
| --- | --- | --- | --- | --- |
| P190032/S004 | December 22, 2022 | *NTRK1/2/3* fusions | Solid Tumors | ROZLYTREK® (entrectinib) |
| | | *ROS1* fusions | NSCLC | |
| P190032/S008 | December 19, 2022 | *EGFR* Exon 19 deletions and *EGFR* Exon 21 L858R alteration | NSCLC | *EGFR* tyrosine kinase inhibitors approved by FDA |
| P190032/S005 | May 3, 2023 | *EGFR* Exon 20 insertions | NSCLC | EXKIVITY® (mobocertinib) |
## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Failure of the device to perform as expected or failure to correctly interpret test results may lead to incorrect F1LCDx assay results, and subsequently, inappropriate patient management decisions. Patients with false positive CDx biomarker results may undergo treatment with one of the therapies listed in the intended use statement without clinical benefit and may experience adverse reactions associated with the therapy. Patients with false negative results may not be considered for treatment with the indicated targeted therapy. There is also a risk of delayed results, which may lead to delay of treatment with the indicated therapy. For the specific adverse events related to the approved therapeutics, please see approved drug product labels.
For the specific adverse events that occurred in the clinical study, please see the FDA approved package insert for BRAFTOVI (encorafenib) in combination with cetuximab which is available at Drugs@FDA.
## IX. SUMMARY OF NONCLINICAL STUDIES
### A. Laboratory Studies
The evidence in support of the analytical performance of F1LCDx in detecting *BRAF* V600E alteration is presented in this section. Analytical accuracy/concordance and precision near the LoD studies were conducted to support the indication for *BRAF* V600E alteration using clinical samples.
For F1LCDx platform-level validation (P190032), due to the challenges with obtaining sufficient volume of clinical specimens, analytical performance characteristics were established for some of the studies using contrived samples, which consisted of enzymatically sheared cell line DNA spiked into human plasma from healthy donors, extracted according to the assay’s standard procedure, and diluted with cfDNA isolated from healthy donor plasma. A contrived sample functional characterization (CSFC) study was conducted to demonstrate comparable performance of sheared cell line DNA samples spiked into plasma as compared to cfDNA isolated from plasma specimens obtained from cancer positive intended use patient specimens. These contrived samples were used to establish the LoD for short variants evaluated in the platform-level LoD establishment study. For information regarding the platform-level validation, refer to Section IX.A. in P190032 Summary of Safety and Effectiveness Data.
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
16 of 33
{16}
### 1. Analytical Accuracy/Concordance
An analytical accuracy/concordance study was performed to demonstrate the concordance between F1LCDx and an externally validated NGS assay (evNGS) for the detection of *BRAF* V600E alteration. For this study, 50 frozen plasma samples were identified from clinical trial assay-positive (CTA+) patients enrolled in the ARRAY-818-302 (BEACON) trial and 153 residual cfDNA samples including 40 biomarker positives and 113 biomarker negatives were sourced from FMI's clinical archives. Of the 153 residual cfDNA samples, seven (7) were excluded due to LC/HC QC failure or evNGS ctDNA QC failure. Of the 50 frozen plasma samples from CTA+ patients, seven (7) were excluded from the analysis due to diluted DNA concentration being out of acceptable range or evNGS ctDNA QC failure.
Analytical concordance of F1LCDx for detecting *BRAF* V600E alteration was determined with 189 samples tested by F1LCDx assay (Table 10). Since archived specimens were selected based on a previous version of the device and confirmed by the evNGS assay, calculation of percent agreement (PPA) and negative percent agreement (NPA) is presented adjusted for the enrichment of *BRAF* V600E positives in the concordance evaluation sample cohort. Adjusted PPA has a point estimate of 92.31% with a 95% two-sided CI of (78.45%, 100.00%). Adjusted NPA has a point estimate of 100.00% with a 95% two-sided CI of (96.53%, 100.00%). For informational purposes, unadjusted PPA, NPA, PPV (positive predictive value) and NPV (negative predictive value) are also displayed.
**Table 10. Concordance summary for *BRAF* V600E alteration by F1LCDx and the evNGS**
| | evNGS | | | PPV/NPV (Unadjusted) (95% CI^{1}) | |
| --- | --- | --- | --- | --- | --- |
| | | *BRAF* V600E alteration positive | *BRAF* V600E alteration negative | | Total |
| F1LCDx | ***BRAF* V600E alteration positive** | 81 | 0 | 81 | PPV: 100% (95.47%, 100%) |
| | ***BRAF* V600E alteration negative** | 1* | 107 | 108 | NPV: 99.07% (94.94%, 99.84%) |
| | **Total** | 82 | 107 | 189 | |
| | **PPA/NPA (Unadjusted) (95% CI^{1})** | PPA: 98.78% (93.41%, 99.78%) | NPA: 100% (96.53%, 100%) | | |
$^{1}$Calculated with Wilson 2-sided 95% CI
*This sample has very low supporting reads (4) and variant allele fraction (0.000544) in F1LCDx, which did not pass F1LCDx positive calling threshold.
The one (1) sample that was *BRAF* V600E alteration negative by F1LCDx and *BRAF* V600E alteration positive by the evNGS was discordant due to low supporting reads (4) and variant allele fraction (VAF), 0.0544% VAF in F1LCDx, which did not pass F1LCDx positive calling threshold.
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
17 of 33
{17}
The lowest VAF F1LCDx test result observed in the accuracy study was 1.02% VAF; therefore, accuracy of F1LCDx was not demonstrated for samples with variants below this VAF level. A limitation is included in the device labeling to address the uncertainty of the accuracy of F1LCDx for detecting BRAF V600E alteration with VAFs below those evaluated in the analytical concordance study (refer to Section XII.C. below). Although accuracy is not supported by direct comparison to orthogonal method, the clinical trial included specimens below the LoD and efficacy was demonstrated in specimens below the LoD supporting the conclusion that positives below the LoD are true positives for BRAF.
## 2. Analytical Sensitivity
a. Limit of Blank (LoB)
The LoB of F1LCDx was evaluated in the platform LoB study for PMA P190032 (refer to Section IX.A.3.a. in the Summary of Safety and Effectiveness Data for P190032).
A supplemental LoB study was performed for F1LCDx to support the updated LC input range (20-60ng) by collecting whole blood samples from 44 healthy donors and preparing two plasma cfDNA replicates per donor for a total of 88 cfDNA sample replicates. Additionally, one matched gDNA replicate per donor was isolated from buffy coat and mechanically fragmented for F1LCDx testing to obtain non-tumor variant (e.g., germline) information to support the LoB analysis. One cfDNA replicate was excluded from the analysis due to failure at the DNA extraction step. BRAF V600E alterations were not observed in any of the replicates, and therefore, the false positive rate was 0% and the LoB for BRAF V600E alteration is zero.
b. Limit of Detection (LoD)
The LoD of BRAF V600E alteration was not established as part of this PMA supplement. However, the LoD for single nucleotide variants (SNVs) in contrived samples was established as part of the LoD study for PMA P190032 (refer to Section IX.A.3.b in the Summary of Safety and Effectiveness Data for P190032). In this study, the median LoD for short variants was determined to be 0.40% VAF in the enhanced sensitivity region of the bait set, which includes BRAF, and the LoD for BRAF V600E alteration was determined to be 0.33% VAF using the empirical hit rate approach. The LoD was confirmed with clinical samples in the precision study (refer to Section IX.A.3, below). The confirmed LoD was 0.70%.
## 3. Precision and Reproducibility
a. Within-Laboratory (Intermediate) Precision
A precision study was conducted using two (2) clinical CRC samples harboring BRAF V600E alteration, however just one sample was close to the LoD for BRAF V600E (~2x, the second sample was ~16x LoD).
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
18 of 33
{18}
Repeatability including intra-run performance (run on the same plate under the same conditions) and reproducibility including inter-run performance (run on different plates under different conditions) were assessed and compared in duplicates across two (2) different sequencers and three (3) different reagent lots, across multiple days of performance by multiple operators.
The results for the precision study for the clinical sample are summarized in Tables 11 and 12, below.
**Table 11. Reproducibility results for *EGFR* exon 20 insertions**
| *BRAF* V600E alteration | Previously Established LoD VAF^{1} | Mean VAF Calculated | Concordant /Total (n/N) | Reproducibility (%) 95% CI^{2} | Fold LoD |
| --- | --- | --- | --- | --- | --- |
| 1799T>A | 0.33% | 0.7% | 24/24 | 100% (86.2, 100) | 2.12x |
$^{1}$0.33% VAF represents the VAF evaluated in the platform LoD establishment study using contrived samples (See Section IX.A.2.b)
$^{2}$Calculated with Wilson 2-sided 95% CI
**Table 12. Repeatability results for *EGFR* exon 20 insertions**
| *BRAF* V600E alteration | Previously Established LoD VAF^{1} | Mean VAF Calculated | Concordant /Total (n/N) | Repeatability (%) 95%CI^{2} | Fold LoD |
| --- | --- | --- | --- | --- | --- |
| 1799T>A | 0.33% | 0.7% | 12/12 | 100% (75.75, 100) | 2.12x |
$^{1}$0.33% VAF represents the VAF evaluated in the platform LoD establishment study using contrived samples (See Section IX.A.2.b)
$^{2}$Calculated with Wilson 2-sided 95% CI
## **B. Animal Studies**
No animal studies were conducted using the F1LCDx assay.
## **C. Additional Studies**
None
## **X. SUMMARY OF PRIMARY CLINICAL STUDIES**
The reasonable assurance of safety and effectiveness for F1LCDx for detection of *BRAF* V600E alteration in patients with mCRC who may benefit from treatment with BRAFTOVI (encorafenib) in combination with cetuximab was established through a clinical bridging study using clinical plasma specimens from patients enrolled in the BEACON (ARRAY-818-302) trial, as well as *BRAF* V600E alteration-negative tissue-matched plasma samples from commercial sources. Data from this clinical study were the basis for the PMA supplement approval decision.
A summary of the clinical study is presented below.
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
19 of 33
{19}
# A. Study Design
# 1. BEACON (ARRAY-818-302) Study Design
The BEACON trial is a Phase 3 multicenter, randomized, open-label, 3-arm study designed to evaluate the efficacy and safety of encorafenib + cetuximab + binimetinib (referred to as the Triplet arm) and encorafenib + cetuximab (referred to as the Doublet arm) versus Investigator's choice of either irinotecan/cetuximab or FOLFIRI/cetuximab (Control arm) in patients with BRAF V600E mutation mCRC whose disease had progressed after 1 or 2 prior regimens in the metastatic setting. The primary endpoints of the BEACON study were overall survival (OS) (Triplet arm vs. Control arm) and confirmed overall response rate (ORR) by Blinded independent central review (BICR) per Response Evaluation Criteria in Solid Tumors (RECIST) v1.1 (Triplet arm vs. Control arm). The secondary endpoints of the BEACON study were OS and ORR by BICR per RECIST v1.1 of the Doublet Arm compared against the Control Arm. This SSED presents analysis from the Doublet Arm against the Control Arm, in order to align with NDA labelling.
Patients were enrolled into the BEACON trial based on tumor tissue samples using either a local laboratory developed test (LDT), followed by confirmation of the local LDT result via a central clinical trial assay (CTA) test or using the CTA for enrollment directly.
# 2. Clinical Bridging Study Design
A clinical bridging study was conducted to evaluate: 1) the concordance between the F1LCDx assay and the CTA for the detection of BRAF V600E alteration for patient samples from the Doublet and Control Arms, and 2) the clinical validity of F1LCDx in identifying mCRC patients with BRAF V600E alteration who may be eligible for treatment with BRAFTOVI (encorafenib) in combination with cetuximab.
Plasma samples from the BEACON trial patients were collected by the therapeutic investigational sites per the study protocol and study documents and shipped to the central testing laboratories. All patients with available plasma samples from the Doublet and Control Arms (i.e., patient population that supported the BRAFTOVI (encorafenib) in combination with cetuximab approval) from the BEACON trial were tested by F1LCDx as part of this clinical bridging study. Additionally, tissue and plasma-matched CTA-negative samples procured from commercial sources were also included in the clinical bridging study.
# 3. Clinical Inclusion and Exclusion Criteria
The inclusion/exclusion criteria for the retrospective testing of plasma samples in the clinical bridging study were:
Sample inclusion criteria:
- Samples from enrolled patients from Doublet and Control Arms with informed consent provided in the BEACON trial
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
20 of 33
{20}
- Samples from commercial sources must have a valid CTA negative result from tissue
- Specimens in frozen plasma
- Samples must meet F1LCDx operational testing requirements
Sample exclusion criteria:
- Tissue, other liquid samples
- Samples that do not meet F1LCDx operational testing requirements
Specimens included in the clinical bridging study were tested according to the standard testing protocol for the F1LCDx assay test with a minimum recommended cfDNA input of ≥ 20 ng for the library construction step.
# 4. Follow-up Schedule
The F1LCDx clinical bridging study involved only retrospective testing of plasma samples; as such, no additional patient follow-up was conducted.
# 5. Clinical Endpoints
This SSED presents analysis from the Doublet Arm against the Control Arm (which was the secondary endpoint), in conjunction with the corresponding NDA (210496) (See Section D. Safety and Effectiveness Results).
The Endpoints were the following:
- OS, defined as the time from randomization to death due to any cause, of Doublet Arm vs. Control Arm
- Confirmed ORR (by BICR) per RECIST, v1.1 of Doublet Arm vs. Control Arm.
# B. Accountability of PMA Cohort
A total of 561 samples, including 441 plasma samples from the BEACON clinical trial and 120 commercial BRAF V600E negative plasma samples, were identified for the clinical bridging study. Out of the 561 samples, 37 CTA unevaluable samples (31 from the clinical trial and 6 procured) and one CTA positive procured sample were excluded from the clinical bridging analysis, resulting a total of 523 samples included in the bridging analysis.
Among the 441 samples from the Doublet and Control Arms from the BEACON trial, 8 samples were CTA-negative and were removed from the positive population in the bridging analysis; 31 samples were CTA unevaluable and therefore were excluded from the clinical bridging analysis. Out of the remaining 402 CTA+ clinical trial samples, 42 CTA+ samples did not have a plasma sample available for F1LCDx testing, 32 failed the F1LCDx QC metrics, resulting in a total of 328 BRAF V600E-positive samples that had both CTA and F1LCDx evaluable results.
Among the 120 tissue and plasma-matched procured samples, 1 sample was CTA+, and 6 samples were CTA unevaluable, 14 failed the F1LCDx QC metrics, resulting in 99
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
21 of 33
{21}
procured BRAF V600E-negative samples that had F1LCDx evaluable results.
Additionally, 6 out of 8 CTA- samples from the clinical trial that had F1LCDx evaluable results were included in the negative population in the analysis.
For the clinical bridging analysis, the CTA+ set contains 402 CTA+ (328 F1LCDx-evaluable; F1LCDx-evaluable refers to samples with valid F1LCDx testing results) clinical trial samples, and the CTA-negative set consists of 121 CTA- samples, where 113 CTA- samples (99 F1LCDx-evaluable) are from procured samples and 8 CTA- (6 F1LCDx-evaluable) samples are from the clinical trial, resulting in a total of 105 (99+6) CTA- and F1LCDx-evaluable samples.
The accountability for all samples reported in this study are detailed in Figure 1 below.
Figure 1. Sample Processing Flow Chart

### C. Study Population Demographics and Baseline Parameters
The demographics and baseline disease characteristics were similar between F1LCDx-evaluable and F1LCDx-unevaluable, and CTA-positive patients (Table 13).
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
22 of 33
{22}
Table 13. Comparison of baseline demographic and clinical characteristics between the CDx-evaluable patients and the CDx-unevaluable patients
| Baseline characteristic | CTA-positive | F1LCDx-evaluable | F1LCDx-unevaluable | p-value comparing the two subsets* |
| --- | --- | --- | --- | --- |
| AGE | | | | 0.17 |
| Mean | 59.43 | 59.79 | 57.82 | |
| Min | 27.00 | 27.00 | 29.00 | |
| Q1 | 53.00 | 53.00 | 50.25 | |
| Median | 61.00 | 61.00 | 59.50 | |
| Q3 | 68.00 | 68.25 | 65.75 | |
| Max | 91.00 | 91.00 | 80.00 | |
| SD | | 12.00 | 11.80 | |
| SEX | | | | 1 |
| Female | 212 (52.74%) | 173 (52.74%) | 39 (52.7%) | |
| Male | 190 (47.26%) | 155 (47.26%) | 35 (47.3%) | |
| ECOG | | | | 0.13 |
| 0 | 197 (49.00%) | 154 (46.95%) | 43 (58.11%) | |
| 1 | 201 (50.00%) | 171 (52.13%) | 30 (40.54%) | |
| 2 | 4 (1.00%) | 3 (0.91%) | 1 (1.35%) | |
| LOC_TUMOR (Location of primary tumor) | | | | 0.38 |
| Both Sides | 28 (6.97%) | 24 (7.32%) | 4 (5.41%) | |
| Left Colon | 137 (34.08%) | 109 (33.23%) | 28 (37.84%) | |
| Right Colon | 213 (52.99%) | 178 (54.27%) | 35 (47.3%) | |
| Unknown Colon | 24 (5.97%) | 17 (5.18%) | 7 (9.46%) | |
| INVOL_ORGAN (Involvement of ≥3 organs) | | | | 0.12 |
| No | 217 (53.98%) | 171 (52.13%) | 46 (62.16%) | |
| Yes | 185 (46.02%) | 157 (47.87%) | 28 (37.84%) | |
| LIVER_METAS (Presence of liver metastases) | | | | 0.24 |
| No | 170 (42.29%) | 134 (40.85%) | 36 (48.65%) | |
| Yes | 232 (57.71%) | 194 (59.15%) | 38 (51.35%) | |
| PRI_TUMOR (Primary tumor removed) | | | | 0.36 |
| Completely Resected | 228 (56.72%) | 190 (57.93%) | 38 (51.35%) | |
| Partially Resected or Unresected | 174 (43.28%) | 138 (42.07%) | 36 (48.65%) | |
| PRE_THERAPY (Previous lines of therapy) | | | | 0.91 |
| 1 | 264 (65.67%) | 216 (65.85%) | 48 (64.86%) | |
| 2 | 137 (34.08%) | 111 (33.84%) | 26 (35.14%) | |
| 3 | 1 (0.25%) | 1 (0.3%) | 0 (0%) | |
| HIGH_MICRO (High microsatellite instability) | | | | 0.23 |
| N/A | 76 | 40 | 36 | |
| No | 295 (90.49%) | 263 (91.32%) | 32 (84.21%) | |
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
23 of 33
{23}
| Baseline characteristic | CTA-positive | F1LCDx-evaluable | F1LCDx-unevaluable | p-value comparing the two subsets* |
| --- | --- | --- | --- | --- |
| Yes | 31 (9.51%) | 25 (8.68%) | 6 (15.79%) | |
| BASE_CARCI (Baseline carcinoembryonic antigen level >5 μg/liter) | | | | 0.00 |
| N/A | 1 | 0 | 1 | |
| No | 100 (24.94%) | 72 (21.95%) | 28 (38.36%) | |
| Yes | 301 (75.06%) | 256 (78.05%) | 45 (61.64%) | |
| BASE_CPRO (Baseline C-reactive protein level >10 mg/liter) | | | | 0.69 |
| N/A | 6 | 5 | 1 | |
| No | 244 (61.62%) | 197 (60.99%) | 47 (64.38%) | |
| Yes | 152 (38.38%) | 126 (39.01%) | 26 (35.62%) | |
| IRINOTECAN_STA TUS (Irinotecan status) | | | | 0.37 |
| No | 196 (48.76%) | 156 (47.56%) | 40 (54.05%) | |
| Yes | 206 (51.24%) | 172 (52.44%) | 34 (45.95%) | |
*p-value was from nonparametric Mann-Whitney Test for continuous measures, and Fisher-Freeman-Halton Test for categorical measures between the CDx-evaluable and CDx-unevaluable sets
### D. Safety and Effectiveness Results
#### 1. Safety Results
The safety with respect to treatment with BRAFTOVI (encorafenib) in combination with cetuximab was addressed during the review of the supplemental NDA and is not addressed in detail in this Summary of Safety and Effectiveness Data. The evaluation of safety was based on the analysis of adverse events (AEs), clinical laboratory evaluations, physical examinations, and vital signs. No adverse events were reported in connection with the bridging study used to support this PMA supplement, as the study was performed retrospectively using banked samples.
Please refer to Drugs@FDA for complete safety information on BRAFTOVI (encorafenib) in combination with cetuximab.
#### 2. Effectiveness Results
##### a. Concordance Analysis with enrollment CTA
Since patients in the BEACON trial were enrolled based on testing tumor tissue samples using either a local LDT followed by central CTA confirmation or using CTA for enrollment directly, a study was conducted to evaluate the concordance between the LDT and CTA results. Data supported that prescreened LDT and CTA results are highly concordant. Therefore, CTA+ population and CTA+LDT+ double positive population are consistent and the impact of the pre-screening with the LDT is negligible. Direct bridging from the CTA to the F1LCDx ignoring the pre-screening by LDT was performed in this clinical study
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
24 of 33
{24}
to support the intended use of the F1LCDx. As described above, a total of 523 samples (402 CTA+ samples from the clinical trial and 121 CTA- procured or clinical trial samples) were included in the clinical bridging study. Concordance of the F1LCDx assay with the enrolling CTA was demonstrated with the CDx- evaluable population (328 CTA+ and 105 CTA- samples) and summarized in Table 14, below.
Table 14. Concordance for BRAF V600E mutation between F1LCDx in plasma and the CTA in tissue
| | CTAs | | | |
| --- | --- | --- | --- | --- |
| | | Detected | Not Detected | Total |
| F1LCDx | Detected | 286 | 3 | 289 |
| | Not Detected | 42 | 102 | 144 |
| | Unevaluable | 74 | 16 | 90 |
| | Total | 402 | 121 | 523 |
| Agreement Statistics Excluding CDx- Unevaluable Results | | PPA: 87.2% (286/328) 95% CI¹: (83.1%, 90.4%) | NPA: 97.1% (102/105) 95% CI¹: (91.9%, 99.0%) | |
| Percent Unevaluable | | 18.4% (74/402) | 13.2% (16/121) | |
¹Calculated with Wilson 2-sided 95% CI
The PPA was 87.2% (286/328)) with 95% two-sided CI (83.1%, 90.4%) and NPA was 97.1% (102/105) with 95% CI (91.9%, 99.0%) after excluding CDx- unevaluable results. Since patients were enrolled and initially tested by LDT and/or central CTA, the PPV and NPV were calculated using the PPA and NPA, after adjusting for the prevalence of BRAF V600E alteration among the intention-to-treat (ITT) population. The prevalence estimate used in the adjusted agreement was 10% and 15%. In this analysis, F1LCDx demonstrated an adjusted PPV of 77.2% with 95% two-sided CI (59.4%, 100%) and NPV of 98.6% with 95% two-sided CI (98.2%, 98.9%) at 10% prevalence. At the estimated prevalence of 15%, F1LCDx demonstrated an adjusted PPV of 84.3% with 95% two-sided CI (69.9%, 100%) and NPV of 97.7% with 95% two-sided CI (97.1%, 98.3%).
The discordance between the CTA and F1LCDx among BRAF V600E positive patients in the primary analysis were evaluated. Of the 42 CTA- positive/F1LCDx-negative samples, 27 had no BRAF V600E alteration detected by the F1LCDx pipeline. Fifteen (15) samples had an BRAF V600E alteration detected by F1LCDx but were filtered out for failing the pipeline's quality threshold. All 42 CTA-positive/F1LCDx-negative samples were from patients in the Doublet and Control arms of the BEACON trial that had clinical outcome data (refer to Section X.D.2.b, below). Note that all 3 CTA- negative /F1LCDx- positive samples were from patients in the Control arm of the BEACON trial that had clinical outcome data.
Based on the PPA between F1LCDx and the enrolling CTAs, which were predominately tissue-based tests, as shown in Table 15, F1LCDx may miss a
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
25 of 33
{25}
proportion of patients with mCRC with BRAF V600E alteration who may derive benefit from BRAFTOVI (encorafenib) in combination with cetuximab.
Therefore, reflex testing using tissue specimens to an FDA approved tissue test will be required, if feasible, if the plasma test is negative.
# b. Bridging clinical outcome from CTA to F1LCDx
The clinical efficacy of BRAFTOVI (encorafenib) in combination with cetuximab in the clinical trial was measured by OS and ORR with either confirmed complete response (CR) or partial response (PR) based on BICR per RECIST v1.1 of the Doublet Arm compared against the Control Arm. Only clinical samples with clinical outcome data were used in this part of the study analysis.
The ORR for the Doublet Arm and Control Arm as well as the ORR difference are reported in Table 15 for the following subpopulations: CTA+, F1LCDx+|CTA+, F1LCDx-|CTA+, and F1LCDx unevaluable |CTA+. Table 15 also summarizes the median overall survival (OS) by the Kaplan-Meier method for each arm as well as the log(hazard ratio) [log(HR)] with 95% two-sided CI for each of the aforementioned subpopulations. The ORR differences between the Doublet and Control Arms were 18.41% (12.74%, 24.55%), 17.06% (10.51%, 24.22%), 17.39% (-2.39%, 37.14%) and 25.74% (9.73%, 43.10%) for CTA+, F1LCDx+|CTA+, F1LCDx-|CTA+, and F1LCDx unevaluable |CTA+ population, respectively.
Table 15. Efficacy by BRAF V600E alteration status in biomarker subgroups
| | CTA+ | F1LCDx+ | CTA+ | F1LCDx- | CTA+ | F1LCDx unevaluable | CTA+ |
| --- | --- | --- | --- | --- | --- | --- | --- |
| # Total | 402 | 286 | 42 | 74 |
| ORR for Doublet Arm | 19.90% | 18.49% | 17.39% | 28.13% |
| ORR for Control Arm | 1.49% | 1.43% | 0.00% | 2.38% |
| ORR Difference (95% two-sided CI)* | 18.41% [12.74%, 24.55%] | 17.06% [10.51%, 24.22%] | 17.39% [-2.39%, 37.14%] | 25.74% [9.73%, 43.10%] |
| Median OS (months) for Doublet Arm | 9.49 | 7.62 | NA§ | 18.89 |
| Median OS (months) for Control Arm | 5.88 | 5.38 | 12.16 | 7.16 |
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
26 of 33
{26}
| | CTA+ | F1LCDx+ | CTA+ | F1LCDx- | CTA+ | F1LCDx unevaluable | CTA+ |
| --- | --- | --- | --- | --- | --- | --- | --- |
| log(HR) (95% two-sided CI) | -0.51 [-0.76, -0.26] | -0.47 [-0.75, -0.19] | -2.72 [-4.71, -0.74] | -0.44 [-1.23, 0.34] |
*CI was calculated using the Newcombe method.
§The estimated median OS is NA due to the small number of events in this group (3 events).
The F1LCDx-positive intended use population consists of both F1LCDx-positive | CTA-positive and F1LCDx-positive | CTA-negative subjects and the final drug efficacy in F1LCDx-positive can be estimated as a weighted average of drug efficacy in F1LCDx-positive | CTA-positive and drug efficacy in F1LCDx-positive | CTA-negative with weight Pr(CTA+| F1LCDx +) and 1- Pr(CTA+| F1LCDx +) respectively. The bridging efficacy analysis to evaluate final drug efficacy in F1LCDx-positive intended use population was performed for log (HR) and ORR difference, separately (Table 16).
The ORR difference between the Doublet and Control Arms in F1LCDx-positive subjects were 13.18% (95% CI: 6.86% -19.50%), 14.34% (95% CI: 8.48% - 20.21%), 15.12% (95% CI: 9.48% - 20.76%), 15.90% (95% CI: 10.42% - 21.38%) and 17.06% (95% CI: 11.67% - 22.46%) when a range of ORR values for the F1LCDx-positive | CTA-negative subjects were assumed as 0%, 30%, 50%, 70%, and 100% of the observed ORR in the F1LCDx-positive | CTA-positive subjects, assuming that the prevalence of CTA-positive subjects in the intended use population is 10%.
The log (HR) between the Doublet Arm and Control Arm in F1LCDx-positive subjects ranges from -0.36 (95% CI: -0.61, -0.12) to -0.47 (95% CI: -0.70, -0.24) when the log(HR) values for the F1LCDx-positive | CTA-negative subjects were assumed to be between the worst-case scenario (no drug efficacy in F1LCDx-positive | CTA-negative subjects) and the best case scenario (full drug efficacy in F1LCDx-positive | CTA-negative subjects as that in F1LCDx-positive | CTA-positive subjects) and we also assume that the prevalence of CTA-positive subjects in the intended use population is 10%..
The efficacy for the F1LCDx+ population was also calculated assuming that the prevalence of CTA-positive subjects in the intended use population is 15% and the results are showed in Table 16.
Table 16. Estimated Efficacy for the F1LCDx+ Population
| | δ̄_{CDx+} with 95% CI [log(HR)] | δ̄_{CDx+} with 95% CI (ORR difference) |
| --- | --- | --- |
| | prev = 10% | |
| c*=0% | -0.36 [-0.61, -0.12] | 13.18 [6.86, 19.50] |
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
27 of 33
{27}
| | $$\hat{\delta}_{CDx+}$$ with 95% CI [log(HR)] | $$\hat{\delta}_{CDx+}$$ with 95% CI (ORR difference) |
| --- | --- | --- |
| c=30% | -0.40 [-0.63, -0.16] | 14.34 [8.48, 20.21] |
| c=50% | -0.42 [-0.65, -0.18] | 15.12 [9.48, 20.76] |
| c=70% | -0.44 [-0.67, -0.21] | 15.90 [10.42, 21.38] |
| c=100% | -0.47 [-0.70, -0.24] | 17.06 [11.67, 22.46] |
| | prev = 15% | |
| c=0% | -0.40 [-0.65, -0.15] | 14.39 [8.20, 20.58] |
| c=30% | -0.42 [-0.67, -0.17] | 15.19 [9.25, 21.14] |
| c=50% | -0.43 [-0.68, -0.19] | 15.73 [9.90, 21.55] |
| c=70% | -0.45 [-0.69, -0.21] | 16.26 [10.52, 22.01] |
| c=100% | -0.47 [-0.71, -0.23] | 17.06 [11.37, 22.76] |
*c is the ratio of efficacy between F1LCDx+|CTA- and F1LCDx+|CTA+ populations.
Samples that were positive by F1LCDx for BRAF V600E alteration in the clinical bridging study had VAFs as low as 0.11%. Since F1LCDx does not have a pre-specified VAF cut-off for positivity, but rather is based on quality threshold based on select metrics for each variant position, it is possible that patients with VAFs below the limit of detection (0.70%) and below that which was enrolled in the trial (e.g., 0.11%) can be reported as positive. The clinical effectiveness of F1LCDx has not been demonstrated for patients with mCRC that have BRAF V600E alteration with VAF below 0.11%. A limitation addressing the uncertainty of the clinical effectiveness of the device for BRAF V600E alteration with VAFs below those evaluated in the clinical study is included as a limitation to the device (also refer to Section XII. C. below).
# c. Sensitivity Analysis
Sensitivity analyses with regard to missing values were conducted to evaluate the robustness of the ORR estimates considering F1LCDx unevaluable patients enrolled in the BEACON trial. Samples were considered missing if the samples were not tested, if they were tested but returned an invalid result, or if they did not satisfy the F1LCDx cfDNA minimum input requirement (i.e., ≥ 20ng).
Amongst all CTA-positive NDA patients, 18.4% did not have a F1LCDx result (74/402). To evaluate the impact of the F1LCDx unevaluable population, the distribution of patients for baseline covariates and disease characteristics was compared among the CTA-positive population, the F1LCDx-evaluable/CTA-positive subpopulation, and F1LCDx-unevaluable/CTA-positive subpopulation. A multiple imputation method was utilized to account for patients with missing or non-evaluable F1LCDx results (n=74).
The imputed ORR difference between the Doublet and Control Arms by BICR was estimated to range from 14.54% (95% CI: 8.27%, 20.81%) to 18.91(95% CI: 12.49%, 25.34%) when a range of ORR values for the F1LCDx-
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
28 of 33
{28}
unevaluable/CTA-positive subjects were assumed as 0%, 30%, 50%, 70%, and 100% of the observed ORR in the F1LCDx-positive /CTA-positive subjects, which is similar to the ORR difference for the CTA-positive population based on the observed data [18.41% (95% CI: 12.74%, 24.55%)]. The imputed log (HR) was estimated to range from -0.41% (95% CI: -0.66%, -0.15%) to -0.53 (95% CI: -0.82%, -0.24%) when a range of log (HR) for the F1LCDx-unevaluable/CTA-positive subjects were assumed as 30%, 50%, 70%, and 100% of the observed log (HR) in the F1LCDx-positive /CTA-positive subjects, which is similar to the log (HR) for the CTA-positive population based on the observed data [-0.51 (95% CI: -0.76, -0.26)]. Thus, the sensitivity analysis demonstrated the robustness of the clinical efficacy estimate. The sensitivity analysis results demonstrate that the estimated drug efficacy in the F1LCDx-positive set of the primary bridging studies for ORR difference and log (HR) remain robust to missing F1LCDx results.
3. Pediatric Extrapolation
In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population.
E. Financial Disclosure
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included one (1) investigator which was a full-time employee of the sponsor and had disclosable financial interests/arrangements as defined in 21 CFR 54.2(a), (b), (c) and (f) and described below:
- Compensation to the investigator for conducting the study where the value could be influenced by the outcome of the study: [0]
- Significant payment of other sorts: [0]
- Proprietary interest in the product tested held by the investigator: [1]
- Significant equity interest held by investigator in sponsor of covered study: [0]
The applicant has adequately disclosed the financial interest/arrangements with clinical investigators. Statistical analyses were conducted by FDA to determine whether the financial interests/arrangements had any impact on the clinical study outcome. The information provided does not raise any questions about the reliability of the data.
XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION
In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Molecular and Clinical Genetics Panel, an FDA advisory committee, for review and recommendation because
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
29 of 33
{29}
the information in the PMA substantially duplicates information previously reviewed by this panel.
# XII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES
# A. Effectiveness Conclusions
The effectiveness of F1LCDx to identify BRAF V600E alteration in patients with mCRC who may benefit from treatment with BRAFTOVI (encorafenib) in combination with cetuximab was demonstrated through clinical bridging studies using specimens from patients enrolled into the BEACON trial. The data from the analytical validation and clinical bridging studies support the reasonable assurance of safety and effectiveness of the F1LCDx assay when used in accordance with the indications for use. Data from the BEACON trial show that patients with mCRC harboring BRAF V600E alteration received benefit from treatment with BRAFTOVI (encorafenib) in combination with cetuximab and support the addition of the CDx indication to F1LCDx.
# B. Safety Conclusions
The risks of the device are based on data collected in the analytical and clinical validation studies conducted to support sPMA approval, as described above. The F1LCDx assay is an in vitro diagnostic test, which involves testing of cfDNA extracted from blood or plasma.
Failure of the device to perform as expected or failure to correctly interpret test results may lead to incorrect test results, and subsequently, inappropriate patient management decisions in cancer treatment. Patients with false positive results may undergo treatment with one of the therapies listed in Table 1 of the intended use statement without clinical benefit and may experience adverse reactions associated with the therapy. Patients with false negative results may not be considered for treatment with the indicated therapy. There is also a risk of delayed results, which may lead to delay of treatment with indicated therapy.
# C. Benefit-Risk Determination
The probable benefit of the F1LCDx assay in identifying patients with mCRC with BRAF V600E alteration for treatment with BRAFTOVI (encorafenib) in combination with cetuximab was demonstrated through clinical bridging studies using specimens from patients enrolled into the BEACON (ARRAY-818-302) trial.
For patients with mCRC with BRAF V600E alteration positive status, the ORR difference for the CTA-positive patients was 18.41% (95% CI: 12.74%, 24.55%), and the log(HR) was -0.51(95% CI: -0.76, -0.26). Clinical outcome for patients with mCRC with BRAF V600E alteration positive status by the CTA and F1LCDx indicated an ORR difference of 17.06% (95% CI:10.51%, 24.22%) and log(HR) of -0.47 (95% CI: -0.75, -0.19), which were comparable to the ORR difference and log (HR) in the CTA-positive population and provides evidence of a meaningful clinical benefit in this population. The observed ORR difference and log(HR) for the F1LCDx BRAF V600E alteration positive patients supports probable benefit of
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
30 of 33
{30}
F1LCDx in selecting BRAF V600E alteration positive mCRC patients for treatment with BRAFTOVI (encorafenib) in combination with cetuximab.
There is potential risk associated with the use of this device, mainly due to 1) false positive, false negatives, or failure to provide a result, and 2) incorrect interpretation of test results by the user. The risks of the F1LCDx assay are associated with the potential mismanagement of patients resulting from false results of the test. Patients who are determined to be false positive by the test may be exposed to a drug that is not beneficial which may lead to adverse events or may have delayed access to treatments that could be more beneficial. A false negative result may prevent a patient from accessing a potentially beneficial drug.
The risks of false results are partially mitigated by the analytical and clinical performance of the device, as summarized above, including the analytical accuracy, and clinical concordance and bridging efficacy studies. In addition, the risks of false negative results are partially mitigated by a recommendation that those patients whose plasma generates a negative result for those alterations included in Table 1, including BRAF V600E alteration, should have their tumor mutation status verified by using an FDA-approved tumor tissue test, if feasible. Additional factors to consider in determining probable risks and benefits for F1LCDx included: the availability of alternative tests. Of note, there are no FDA-approved CDx alternatives using cfDNA isolated from plasma for the detection of BRAF V600E alteration to identify patients with mCRC for treatment with BRAFTOVI (encorafenib) in combination with cetuximab, and this device may meet an unmet clinical need; however, there is an FDA-approved tissue test available for the identification of mCRC patients with BRAF V600E mutation for treatment with BRAFTOVI (encorafenib) in combination with cetuximab. The advantage of F1LCDx over the FDA-approved tissue test, is that it offers a non-invasive method to obtain DNA from the cancer, rather than a tissue biopsy and may meet an unmet need for patients who cannot otherwise provide a biopsy. Another factor to consider in the benefit-risk profile is that this device would provide additional potential benefit to the patient by profiling 324 cancer related genes. However, given uncertainties that remain based on the analytical and clinical validation data, the following limitations are included in the device labeling:
- The analytical accuracy for the FoundationOne Liquid CDx assay for detection of BRAF V600E alteration has not been demonstrated for samples with <1.02% VAF
- BRAFTOVI (encorafenib) in combination with cetuximab efficacy has not been established in patients with BRAF V600E alteration <0.11 % VAF tested with FoundationOne Liquid CDx
The overall clinical and analytical validation data support that for F1LCDx, and the indications noted in the intended use statement, the probable benefits outweigh the probable risks.
# 1. Patient Perspective
PMA P190032/S010: FDA Summary of Safety and Effectiveness Data
31 of 33
{31}
This submission did not include specific information on patient perspectives for this device.
In conclusion, given the available information above, the data support that for the selection of patients with mCRC with BRAF V600E alteration for treatment with BRAFTOVI (encorafenib) in combination with cetuximab, the probable benefits outweigh the probable risks.
# D. Overall Conclusions
The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the indications for use. Data from the clinical bridging study support the performance of F1LCDx as an aid for the identification of BRAF V600E alteration in patient…