Single Vision Spherical (SVS) Vision Correction The BAUSCH + LOMB ULTRA (samfilcon A) Contact Lens is indicated for extended wear for up to 7 days between removals for cleaning and disinfection or disposal of the lens, as recommended by the eye care practitioner. The lens is indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. Presbyopia Vision Correction The BAUSCH + LOMB ULTRA (samfilcon A) Contact Lens For Presbyopia is indicated for extended wear for up to 7 days between removals for cleaning and disinfection or disposal of the lens, as recommended by the eye care practitioner. The lens is indicated for the correction of refractive ametropia (myopia, hyperopia and astigmatism) and presbyopia in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens may be prescribed for add powers ranging from +0.75D to +5.00D. Astigmatism Vision Correction The BAUSCH + LOMB ULTRA (samfilcon A) Contact Lens For Astigmatism is indicated for extended wear for up to 7 days between removals for cleaning and disinfection or disposal of the lens, as recommended by the eye care practitioner. The lens is indicated for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism up to 5.00 diopters.
Device Story
Hydrophilic soft contact lens; samfilcon A material (siloxane methacrylate copolymer); 46% water content; MoistureSeal technology. Placed on cornea by patient; acts as refracting medium to focus light on retina. Corrects refractive ametropia (myopia, hyperopia, astigmatism) and presbyopia. Prescribed by eye care practitioner for extended wear (up to 7 days). Benefits include vision correction and convenience of overnight wear. Clinical decision-making based on practitioner assessment of refractive needs and ocular health.
Clinical Evidence
Prospective, multi-center, randomized, double-blinded 12-month clinical study (IDE #G130282) with 816 enrolled subjects (406 Ultra, 410 PureVision). Primary safety: rate of serious/significant non-serious AEs (3.0% Ultra vs 2.4% PureVision; non-inferiority met). Primary effectiveness: high contrast distance logMAR VA at 12 months (-0.063 Ultra vs -0.048 PureVision; non-inferiority met). No serious AEs reported. Secondary endpoints included wear time (6.7 days/week) and patient-reported symptoms.
Technological Characteristics
Material: samfilcon A (hydrophilic copolymer of siloxane methacrylate, siloxane cross-linker, N-vinyl pyrrolidone). Tint: Reactive Blue Dye 246 (21 CFR 73.3106). Water content: 46%. Oxygen permeability: 114 x 10⁻¹¹ (cm³O₂/sec/cm²/mmHg). Refractive index: 1.411. Sterilization: Moist heat (autoclave) to SAL 10⁻⁶. Packaging: Polypropylene blister, borate buffered saline with 0.5% poloxamine.
Indications for Use
Indicated for aphakic/not-aphakic persons (age 18+) with non-diseased eyes requiring correction of myopia, hyperopia, astigmatism (up to 5.00D), and/or presbyopia. Contraindicated for those with ocular inflammation, infection, corneal/conjunctival/eyelid disease, severe dry eyes, corneal hypoesthesia, systemic disease affecting the eye, or allergies to lens/solution ingredients.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Bausch + Lomb PureVision (balafilcon A) Contact Lenses
Submission Summary (Full Text)
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
I. GENERAL INFORMATION
Device Generic Name: samfilcon A soft (hydrophilic) contact lens
Device Trade Name: Bausch + Lomb ULTRA (samfilcon A) Contact Lenses
Device Procode: LPM
Applicant's Name and Address: Bausch + Lomb, Inc.
1400 Goodman Street
Rochester, NY 14609
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P170035
Date of FDA Notice of Approval: April 30, 2018
II. INDICATIONS FOR USE
Single Vision Spherical (SVS) Vision Correction
The BAUSCH + LOMB ULTRA (samfilcon A) Contact Lens is indicated for extended wear for up to 7 days between removals for cleaning and disinfection or disposal of the lens, as recommended by the eye care practitioner. The lens is indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity.
Presbyopia Vision Correction
The BAUSCH + LOMB ULTRA (samfilcon A) Contact Lens For Presbyopia is indicated for extended wear for up to 7 days between removals for cleaning and disinfection or disposal of the lens, as recommended by the eye care practitioner. The lens is indicated for the correction of refractive ametropia (myopia, hyperopia and astigmatism) and presbyopia in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism of 2.00 diopters or less, that does not interfere with visual acuity. The lens may be prescribed for add powers ranging from +0.75D to +5.00D.
Astigmatism Vision Correction
The BAUSCH + LOMB ULTRA (samfilcon A) Contact Lens For Astigmatism is indicated for extended wear for up to 7 days between removals for cleaning and disinfection or disposal of the lens, as recommended by the eye care practitioner. The lens is indicated for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not-aphakic persons with non-diseased eyes, exhibiting astigmatism up to 5.00 diopters.
PMA P170035: FDA Summary of Safety and Effectiveness Data
{1}
PMA P170035: FDA Summary of Safety and Effectiveness Data
Page 2
# III. CONTRAINDICATIONS
Do not use the Bausch + Lomb ULTRA (samfilcon A) Contact Lenses when any of the following conditions exist:
- Acute and subacute inflammation or infection of the anterior chamber of the eye
- Any eye disease, injury, or abnormality that affects the cornea, conjunctiva, or eyelids
- Severe insufficiency of lacrimal secretion (dry eyes)
- Corneal hypoesthesia (reduced corneal sensitivity)
- Any systemic disease that may affect the eye or be exaggerated by wearing contact lenses
- Allergic reactions of ocular surfaces or adnexa (surrounding tissue) that may be induced or exaggerated by wearing contact lenses or use of contact lens solutions
- Allergy to any ingredient, such as mercury or Thimerosal, in a solution which is to be used to care for the Bausch + Lomb Ultra (samfilcon A) Soft (hydrophilic) Contact Lens
- Any active corneal infection (bacterial, fungal, or viral)
- If eyes become red or irritated
# IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the Bausch + Lomb ULTRA (samfilcon A) Contact Lenses labeling.
# V. DEVICE DESCRIPTION
Bausch + Lomb ULTRA (samfilcon A) Contact Lenses are is 46% water and 54% samfilcon A material. The samfilcon A material is a hydrophilic copolymer of siloxane methacrylate, a siloxane cross-linker, and N-vinyl pyrrolidone. It is tinted blue for visibility with Reactive Blue Dye 246, a color additive that conforms to 21 CFR Part 73.3106. The Bausch + Lomb ULTRA (samfilcon A) Contact Lenses utilizes MoistureSeal® technology. In its hydrated state, the Bausch + Lomb ULTRA (samfilcon A) Contact Lens, when placed on the cornea acts as a refracting medium to focus light rays on the retina.
# LENS PARAMETERS
The lens designs include spherical, toric, and multifocal lenses in the following parameter ranges:
Diameter 13.5mm to 15.0mm
Center Thickness 0.05mm to 0.75mm (varies with power)
Base Curve 7.8mm to 9.5mm
Power Range +20.00D to -20.00D
Cylinder Power (Toric) -0.75D to -5.00D
{2}
Cylinder Axis (Toric) 10° to 180° in 10° increments
Add Power (Multi-Focal) +0.75D to +5.00D
The Ultra (samfilcon A) Contact Lens has the following physical properties:
Refractive Index: 1.411
Light Transmittance: C.I.E. value – at least 95%
Water Content: 46%
Specific Gravity: 1.048
Oxygen Permeability: 114 x 10⁻¹¹[cm³O₂(STP) x cm]/(sec x cm² x mmHg)@35°C (polarographic method)
VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are several other alternatives for the correction of refractive ametropia (myopia, hyperopia, and astigmatism) and presbyopia in aphakic and/or not-aphakic persons with non-diseased eyes. The currently available alternate practices and procedures for vision correction are other commercially available contact lenses approved for extended wear, spectacles, refractive keratoplasty, laser-assisted in-situ keratomileusis (LASIK), and corneal implants. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
VII. MARKETING HISTORY
The Bausch + Lomb ULTRA (samfilcon A) Contact Lenses were cleared for daily wear use under 510(k) K131208 on September 11, 2013, Summary information is available on the FDA website at:
https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfPMN/pmn.cfm?ID=K131208. The cleared lenses under K131208 are commercially available for daily wear in the United States, as well as in Canada, Mexico, European Union, South America, Australia, New Zealand, India, Russia, South Africa, United Arab Emirates, Asia Pacific, and many other countries worldwide. These lenses are not currently marketed for extended wear outside the United States. These lenses have not been withdrawn from the market for any reason related to the safety or effectiveness of the device.
VIII. PROBABLE ADVERSE EFFECTS OF THE DEVICE ON HEALTH
The specific probable adverse reactions that may occur include the following:
- Eyes stinging, burning, itching (irritation), or other eye pain
- Comfort is less than when lens was first placed on eye
- Abnormal feeling of something in the eye (foreign body, scratched area)
- Excessive watering (tearing) of the eyes
- Unusual eye secretions
- Redness of the eyes
- Reduced sharpness of vision (poor visual acuity)
PMA P170035: FDA Summary of Safety and Effectiveness Data
{3}
- Blurred vision, rainbows, or halos around objects
- Sensitivity to light (photophobia)
- Dry eyes
For the specific adverse events (AEs) that occurred during the Bausch + Lomb Ultra (samfilcon A) Contact Lens clinical study, please see Section X below.
## IX. SUMMARY OF NONCLINICAL STUDIES
### A. Laboratory Studies
#### 1. Biocompatibility
Non-clinical biocompatibility testing was conducted in accordance with FDA’s Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses, May 1994 and GLP regulation (21 CFR part 58) and applicable standards, which represents the basic level of biocompatibility testing for all contact lenses. These tests include: cytotoxicity, acute systemic toxicity, ocular irritation (<24 hours) and sensitization testing. To address biocompatibility issues related to the extended wear indication of greater than 24 hours, an ocular irritation study was conducted to support duration of exposure.
Non-clinical testing performed includes the following referenced standards:
- Biocompatibility testing per ISO 10993-5, ISO 10993-10, ISO 10993-11, and ISO 9394
#### All test results met the pre-established acceptance criteria.
The testing performed on the Ultra (samfilcon A) extended wear soft contact lens demonstrates the lens is safe. Non-clinical testing included conformance to predetermined specifications.
Table 1: Biocompatibility – Ultra (samfilcon A) extended wear soft contact lens
| Test Method | Acceptance | Results |
| --- | --- | --- |
| Cytotoxicity – Direct Contact ISO 10993-5, clause 8.3 | ≤ grade 2 (mild reactivity) | The test article is considered non-cytotoxic; there was no evidence of causing cell lysis or cytotoxicity to L-929 cells (Grade 0 – no reactivity). |
| Cytotoxicity – MEM Elution ISO 10993-5: clause 8.2 | ≤ grade 2 (mild reactivity) | The test article extracts showed no evidence of causing cell lysis or cytotoxicity (Grade 0 – no reactivity). |
PMA P170035: FDA Summary of Safety and Effectiveness Data
{4}
| Acute Systemic Toxicity ISO 10993-11, clause 5 | Success was defined as follows: None of the Treated mice exhibited a significantly greater reaction than the control mice. <2 Treated mice died or exhibited abnormal behavior and <3 Treated mice exhibited weight loss of >2 grams | There was no mortality or evidence of systemic toxicity from the extracts injected into mice. Each test article extract met the requirements of the study. |
| --- | --- | --- |
| Irritation – Ocular ISO 10993-10: Annex B.2 | No significant irritation over the reagent control | There was no evidence of significant irritation in the test eye or control eye of any animal. The sodium chloride and sesame oil test article extracts are not considered irritants to the ocular tissue of the rabbit. |
| Sensitization - Guinea Pig Maximization ISO 10993-10: clause 7.5 | Grade 0 reaction | The test article extracts showed no evidence of causing delayed dermal contact sensitization in the guinea pig. The test article is not considered a sensitizer in the guinea pig maximization test. |
| 22 Day Ocular Irritation ISO 9394 | Macroscopic ocular reaction grades (Draize scoring), minimal conjunctivitis (e.g. scores of 1) was not considered clinically significant. Biomicroscopic slit lamp data was evaluated in accordance with McDonald-Shadduck criteria. Lactic acid data was compared between groups to determine biological relevance. | No ocular irritation trends. Microscopic evaluation of the ocular tissue revealed that the test and control eyes were similar to each other and were essentially normal. The test and control values obtained in the corneal lactic acid analysis indicated that the samfilcon A contact lenses had no clinical effect on oxygen transport or corneal metabolism. |
PMA P170035: FDA Summary of Safety and Effectiveness Data
Page 5
{5}
The Bausch + Lomb ULTRA (samfilcon A) Contact Lenses are packaged in a polypropylene blister pack. The materials were subjected to cytotoxicity, systemic toxicity and ocular irritation testing. A summary of the testing is provided below:
Table 2: Biocompatibility – Bausch + Lomb ULTRA (samfilcon A) Contact Lenses packaging
| Test Method | Acceptance criteria | Results |
| --- | --- | --- |
| Cytotoxicity – MEM Elution ISO 10993-5, clause 8.2 | ≤ grade 2 (mild reactivity) | The test article extracts showed no evidence of causing cell lysis or cytotoxicity (Grade 0 – no reactivity). |
| Acute Systemic Toxicity ISO 10993-11, clause 5 | Success was defined as follows: None of the Treated mice exhibited a significantly greater reaction than the control mice. <2 Treated mice died or exhibited abnormal behavior and <3 Treated mice exhibited weight loss of >2 grams | There was no mortality or evidence of systemic toxicity from the extracts injected into mice. Each test article extract met the requirements of the study. |
| Irritation – Ocular ISO 10993-10, Annex B.2 | No significant irritation over the reagent control | No evidence of significant irritation in the test or control eye of any rabbit. |
2. Physicochemical Tests
Physicochemical tests were performed to demonstrate long term safety and stability of the properties of the material used to manufacture the Ultra lens. See the following table for a summary of results.
| Test | Purpose | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Preservative Uptake and Release | To determine the preservative uptake and release of the contact lens material | N/A | No significant uptake of either Polyaminopropyl Biguanide (PAPB) or Polyquaternium-1. |
| Compatibility | To determine compatibility | After 30- | Pass |
PMA P170035: FDA Summary of Safety and Effectiveness Data
{6}
| with Lens Care Products | of contact lens care products with contact lenses | cycles, contact lens parameters should be within the finished product specifications | |
| --- | --- | --- | --- |
| Extractables – Leachability | To determine if any monomer, initiator, or tint leached out during extraction with water | N/A | No detectable levels of monomers, initiator, or tint were found in any of the purified water leachate. |
| Extractables - Soxhlet Extraction | The quantity of extractables from the Soxhlet extraction | N/A | The extractables ranged from 0.52% - 1.01% for water and 1.31% - 1.59% for n-hexane |
3. Sterilization, Bioburden, and Shelf Life
The Bausch + Lomb ULTRA (samfilcon A) Contact Lenses are provided in a polypropylene blister and sealed with aluminum lid stock. The packaging solution is a sterile borate buffered saline with 0.5% poloxamine solution. The packaged lenses are steam sterilized in a validated autoclave. Each lens is labeled with the lens parameters, lot number and expiration date and placed in boxes with multiple packaging configurations and appropriate labeling.
Routine bioburden testing is performed prior to sterilization. This testing provides an assessment of the cleanliness of the devices being manufactured and the facility in general. The bioburden test method was validated in accordance with ISO Standard 11737-1:2006, "Sterilization of health care products – Microbiological methods – Part 1: Determination of the population of microorganisms on product."
The Bausch + Lomb ULTRA (samfilcon A) Contact Lenses are terminally sterilized by subjecting the finished device to moist heat sterilization. The moist heat sterilization cycle was validated using the overkill method (partial cycle approach) in accordance with Annex D of ISO Standard 17665-1:2006, "Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation and routine control of sterilization process for medical devices." The sterilization process for the device was validated to achieve a Sterility Assurance Level (SAL) of 10⁻⁶.
Shelf life studies have been conducted to verify that the packaging for the Bausch + Lomb ULTRA (samfilcon A) Contact Lenses maintains a sterile barrier and adequately protects the device through the expiration date on the package label,
PMA P170035: FDA Summary of Safety and Effectiveness Data
{7}
which is 3 years from the date of sterilization. Shelf life testing has also been conducted to verify that device physical and optical properties satisfy the requirements of the engineering drawings and product specification document through the 3 year labeled expiration date. All test samples satisfied all acceptance criteria (see Table 4).
Table 4: Sterility, Bioburden, and Shelf Life
| Test | Purpose | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Bioburden testing | Evaluate the cleanliness of the manufacturing process and facility | Vegetative growth:
Alert: ≥500
CFU/device
Action: ≥1,000
CFU/device
Spores:
Alert: ≥10
CFU/device
Action: ≥20
CFU/device | Pass |
| Moist Heat Sterilization Validation | Evaluate sterility | No positive biological indicators | Pass |
| Package Evaluation – Dye Penetration Testing | Evaluate whole package integrity | No evidence of dye across seal by a defined channel | Pass |
| Sterility testing | Evaluate sterility | Negative for growth | Pass |
| Shelf-life | To establish the expiration date | Finished Product Specifications | 3 years |
X. SUMMARY OF PRIMARY CLINICAL STUDY
The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of the Bausch + Lomb Ultra (samfilcon A) Contact Lenses when worn for up to 7 days extended wear for the optical correction of refractive ametropia and presbyopia in persons with non-diseased eyes in the United States under IDE # G130282. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below.
A. Study Design
Subjects were treated between April 30, 2015 and August 3, 2016. The database for this PMA reflected data collected through August 3, 2016 and included 816 enrolled subjects. There were 406 subjects between the ages of 18 and 40 (812 eyes) in the
PMA P170035: FDA Summary of Safety and Effectiveness Data
Page 8
{8}
Bausch + Lomb Ultra investigational group and 410 subjects (820 eyes) in the Bausch + Lomb PureVision (balafilcon A) control group. There were 34 investigational sites. The study was a prospective, multi-center, two-arm cohort study, randomized, double-blinded, 12-month clinical study. Subjects were 1:1 randomized to one of the two study groups. Randomization was stratified by site.
For the primary safety analysis, the null hypothesis was that the proportion of subjects with serious or significant non-serious adverse events (AEs) in the Ultra lens group minus the proportion of subjects with serious or significant non-serious AEs in the PureVision lens group was 0.05 (5%) or more. The alternative hypothesis was that the difference in proportions was less than 0.05. If the null hypothesis was rejected, then the test lens would be statistically successful in demonstrating non-inferiority to the control lens.
For the primary effectiveness analysis, high contrast distance lens visual acuity (VA) was obtained at the Screening/Dispensing Visit, and at each follow-up visit. Scores were recorded as the number of letters correctly identified in the eye examination. VA was converted to the Log10 of the Minimum Angle of Resolution (logMAR) by using the score (number of letters), and the 12-Month visit VA was compared between treatment groups.
Continuous data were summarized using descriptive statistics: n, mean, median, standard deviation (SD), minimum, and maximum. Categorical data were presented by the total counts for each category and corresponding percentages.
Regarding sample size justification, when the sample size in each group is 326, a two-group large-sample normal approximation test of proportions with a one-sided 0.050 significance level will have 90% power to reject the null hypothesis that Ultra and PureVision are not equivalent in favor of the alternative hypothesis that the proportions in the two groups are equivalent, assuming that the expected difference in proportions is 0.000 and the proportion in the control group is 0.050. Accounting for a total discontinuation rate of up to 20%, at least 326 / (1.0 - 0.2) = 408 subjects were to be enrolled in each group for a total of approximately 816 subjects to be enrolled in the study.
Bausch + Lomb representatives monitored all study site locations to assess the data, quality, and study integrity in a manner consistent with applicable health authority regulations and the procedures adopted by Bausch + Lomb.
Monitoring visits and telephone consultations occurred as necessary, or per the monitoring plan, to verify that the rights and well-being of subjects were protected, the conduct of the investigation was in compliance with the currently approved protocol, the integrity of the data was maintained, the facilities remained acceptable, and test article accountability was ensured.
PMA P170035: FDA Summary of Safety and Effectiveness Data
Page 9
{9}
The control group was an active control; the control treatment, Bausch + Lomb PureVision (balafilcon A) Contact Lenses, is a legally marketed alternative with similar indications for use.
1. Clinical Inclusion and Exclusion Criteria
Enrollment in the Study to Evaluate the Safety and Effectiveness of a Silicone Hydrogel Soft Contact Lens When Worn on a 7-Day Extended Wear Basis was limited to subjects who met the following inclusion criteria. Study enrollment limitations included that subjects were required to be of legal age (at least 18) and no older than 40 on the date the Informed Consent Form was signed, subjects were required to be myopic and require lens correction from $-0.50$ diopters (D) to $-6.00\mathrm{D}$ in each eye, and subjects were required to have clear central corneas and be free of any anterior segment disorders.
Subjects were not permitted to enroll in the Study to Evaluate the Safety and Effectiveness of a Silicone Hydrogel Soft Contact Lens When Worn on a 7-Day Extended Wear Basis study if they met any of the following exclusion criteria: Subjects were participating in any drug or device clinical investigation within two weeks prior to entry into this study and/or during the period of study participation, subjects who had an active ocular disease or who were using any ocular medication, subjects who were not correctable to 42 letters (0.1 logMAR) in each eye with soft spherical contact lenses, subjects who were using any systemic medications that would, in the Investigator's opinion, affect ocular physiology or lens performance, subjects with any Grade 2 or greater finding during the slit lamp examination, and subjects with corneal infiltrates, of any grade.
2. Follow-up Schedule
All subjects were evaluated according to the follow-up schedule below (Table 5):
Table 5
Follow-up Visit Schedule
| Visit | Exam | Target | Visit Window |
| --- | --- | --- | --- |
| 1 | Screening/Dispensing | Day 1 | |
| 2 | 1-Day Follow-up | Day 2 | |
| 3 | 1-Week Follow-up | Day 7 | Day 5-9 |
| 4 | 1-Month Follow-up | Day 30 | Day 26-34 |
| 5 | 3-Month Follow-up | Day 90 | Day 83-97 |
| 6 | 6-Month Follow-up | Day 180 | Day 166-194 |
| 7 | 9-Month Follow-up | Day 270 | Day 256-284 |
| 8 | 12-Month Follow-up | Day 365 | Day 351-379 |
At the screening/dispensing visit, a slit lamp biomicroscope examination was conducted, a spherocylindrical refraction (and best spectacle corrected visual acuity (BSCVA)) was measured, and keratometry was performed. Following the contact lens fitting, symptoms/complaints were assessed, distance high contrast
PMA P170035: FDA Summary of Safety and Effectiveness Data
Page 10
{10}
visual acuity was measured, contact lens over-refraction and visual acuity were measured, and lens wettability, deposits, centration and movement were assessed.
During the follow-up visits, the slit lamp examination was repeated, symptoms/complaints were assessed, distance high contrast visual acuity was measured, change in visual acuity from baseline was determined, contact lens over-refraction and visual acuity were measured, and lens wettability, deposits, centration and movement were assessed. Contact lenses were replaced monthly. Adverse events and complications were recorded at all visits.
The key timepoints are shown below in the tables summarizing safety and effectiveness.
## 3. Clinical Endpoints
With regards to safety,
- The primary safety endpoint was the rate of serious or significant non-serious adverse events during the 12-month follow-up.
Serious adverse events are those events that result in, or have potential to cause, either permanent impairment of an ocular function or damage to an ocular structure, and may necessitate medical or surgical intervention.
Serious adverse events may include any hazardous, sight-threatening conditions occurring after exposure to the test article, including the following:
- A presumed infectious ulcer (defined as a progressive erosion of the corneal tissue);
- Any central or paracentral (within 6mm of cornea) corneal event that results in permanent opacification (such as vascularization);
- Any serious adverse ophthalmic events including hypopyon and hyphema;
- Any neovascularization within the central 6 mm of the cornea;
- Permanent loss of two or more lines of BSCVA;
- All cases of iritis.
Significant Non-Serious Adverse Events include:
- Peripheral non-progressive non-infectious corneal ulcers;
- All symptomatic corneal infiltrative events;
- All cases of corneal staining greater than or equal to Grade 3;
- A temporary loss of two or more lines of BSCVA (for greater than or equal to 2 weeks);
- Neovascularization cases Grade 2 or greater;
- Any ocular event that necessitates temporary lens discontinuation of greater than or equal to 2 weeks.
Secondary safety endpoints included the proportion of eyes with slit lamp findings Grade 2 or greater; AEs and medically treated events; ungraded slit lamp findings; best-spectacle corrected VA line change from baseline; mean subjective
PMA P170035: FDA Summary of Safety and Effectiveness Data
Page 11
{11}
responses to comfort-related symptoms and complaints; lens wettability; lens deposits; centration and movement assessments; changes in keratometry; and changes in refraction.
With regards to effectiveness,
- The primary effectiveness endpoint was high contrast, distance visual acuity (VA) with dispensed lenses at the 12-Month Follow-up Visit.
- The secondary effectiveness endpoint was lens wear time reported as average extended lens wear time (days/week) since last visit.
The null hypothesis for each primary endpoint was that the Ultra lens is inferior to the PureVision lens using the predetermined threshold, or non-inferiority bound, for each endpoint. Rejecting the null hypothesis would provide evidence that the Test lens is at least as good as, or non-inferior to, the Control lens.
With regard to success/failure criteria, the applicant did not define any specific study success criteria beyond the safety and effectiveness endpoints.
The primary and secondary study endpoints were analyzed in two population groups: Dispensed eyes (all eyes dispensed contact lenses) and Eligible Dispensed eyes. Safety endpoints were analyzed using the Dispensed population, and effectiveness endpoints were analyzed using the Eligible Dispensed population.
## B. Accountability of PMA Cohort
At the time of database lock, of 816 subjects enrolled in the PMA study, 100% (816) subjects were available for analysis at the completion of the study, the 12-month final visit evaluated for safety and effectiveness as the basis for the PMA submission.
Of the 816 subjects enrolled, 10 (20 eyes) were ineligible at baseline (2 [4 eyes] in the Ultra group and 8 [16 eyes] in the PureVision group). All of the ineligible subjects were discontinued. Of the 816 enrolled subjects, 669 (82.0%) eligible subjects (1,338 eyes) completed the study, 136 (16.7%) eligible subjects (272 eyes) were discontinued, and one (0.1%) eligible subject (2 eyes) was non-dispensed.
Subject accountability is presented in Table 6.
PMA P170035: FDA Summary of Safety and Effectiveness Data
Page 12
{12}
Table 6
Ultra Clinical Study Subject Accountability
| | Ultra | | PureVision | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | n | % | n | % | n | % |
| Enrolled (N) | 406 | | 410 | | 816 | |
| Eligible at Baseline | 404 | 99.5 | 402 | 98.0 | 806 | 98.8 |
| --Completed study visit schedule and exited | 340 | 83.7 | 329 | 80.2 | 669 | 82.0 |
| --Discontinued | 63 | 15.5 | 73 | 17.8 | 136 | 16.7 |
| --Non-Dispensed | 1 | 0.2 | 0 | | 1 | 0.1 |
| Ineligible at Baseline | 2 | 0.5 | 8 | 2.0 | 10 | 1.2 |
| --Completed study visit schedule and exited | 0 | | 0 | | 0 | |
| --Discontinued | 2 | 0.5 | 8 | 2.0 | 10 | 1.2 |
| --Non-Dispensed | 0 | | 0 | | 0 | |
Percentages (n/N x 100) are based on the number of enrolled subjects within each treatment group.
Regarding reasons for discontinuation, 63 of the 406 Ultra subjects (15.5%), and 73 of the 410 PureVision subjects (17.8%) were discontinued. The major reasons for discontinuation included voluntary withdrawal (24 subjects (3%) for Ultra and 56 subjects (6.8% for PureVision) and lost to follow-up (26 subjects (3.2%) for Ultra and 32 subjects (3.9% for PureVision). Outcomes were similar for adverse events (7 subjects (0.9%) for Ultra and 4 subjects (0.5% for PureVision) and for study-related symptoms/complaints (15 subjects (0.1%) for Ultra and 15 subjects (0.2% for PureVision). Reasons for discontinuation by eye are presented in Table 7.
Table 7
Reasons for Discontinuation by Eye (Dispensed Eyes)
| Reason for Discontinuation | Ultra N = 810 n (%) | PureVision N = 820 n (%) | Total N = 1630 n (%) |
| --- | --- | --- | --- |
| Adverse event | 7 (0.9) | 4 (0.5) | 11 (0.7) |
| Positive slit lamp findings | 0 | 2 (0.2) | 2 (0.1) |
| Study related symptoms/complaints | 15 (1.9) | 15 (1.8) | 30 (1.8) |
| Unacceptable distance lens visual acuity (VA) | 1 (0.1) | 2 (0.2) | 3 (0.2) |
| Unacceptable lens centration | 0 | 0 | 0 |
| Unacceptable lens movement | 3 (0.4) | 0 | 3 (0.2) |
| Ineligible at baseline | 4 (0.5) | 15 (1.8) | 19 (1.2) |
| Inability to maintain recommended wearing schedule | 10 (1.2) | 4 (0.5) | 14 (0.9) |
| Lack of motivation | 0 | 4 (0.5) | 4 (0.2) |
PMA P170035: FDA Summary of Safety and Effectiveness Data
{13}
| Reason for Discontinuation | Ultra N = 810 n (%) | PureVision N = 820 n (%) | Total N = 1630 n (%) |
| --- | --- | --- | --- |
| Failure to follow subject instructions | 2 (0.2) | 2 (0.2) | 4 (0.2) |
| Lost to follow-up | 26 (3.2) | 32 (3.9) | 58 (3.6) |
| Instillation of a non-medically indicated solution not specified in the protocol | 0 | 0 | 0 |
| Pregnant | 12 (1.5) | 14 (1.7) | 26 (1.6) |
| Other eye discontinued | 10 (1.2) | 5 (0.6) | 15 (0.9) |
| Death | 0 | 0 | 0 |
| Voluntary withdrawal | 24 (3.0) | 56 (6.8) | 80 (4.9) |
| Missed >2 follow-up visits | 0 | 0 | 0 |
| Investigator decision not in the best medical interest of the subject to continue the study | 4 (0.5) | 0 | 4 (0.2) |
| Other | 12 (1.5) | 7 (0.9) | 19 (1.2) |
| Total | 130 (16.0) | 162 (19.8) | 292 (17.9) |
Percentages (n/N x 100) are based on the number of dispensed eyes in each treatment group.
## C. Study Population Demographics and Baseline Parameters
The demographics of the study population are typical for a randomized, prospective, multi-center clinical study performed in the United States.
Table 8 presents the demographic data for the Ultra test group vs. the PureVision control group. Subject demographics were similar between the groups. The mean age was approximately 29 years for both lens groups, females represented more than half of both lens groups, and most subjects in both lens groups were white.
Table 8 Demographics (Eligible, Dispensed Subjects)
| | Ultra N=403 | PureVision N=402 |
| --- | --- | --- |
| Age (years) | | |
| N | 403 | 402 |
| Mean (SD) | 29.8 (6.2) | 29.2 (6.1) |
| Median | 30.0 | 29.0 |
| Min, Max | 18, 40 | 18, 40 |
| Gender: n (%) | | |
| Female | 263 (65.3) | 265 (65.9) |
| Male | 140 (34.7) | 137 (34.1) |
PMA P170035: FDA Summary of Safety and Effectiveness Data
{14}
| Race: n (%) | | |
| --- | --- | --- |
| White | 326 (80.9) | 328 (81.6) |
| American Indian/ Alaska Native | 2 (0.5) | 1 (0.2) |
| Native Hawaiian/ Pacific Islander | 0 | 0 |
| Black/ African American | 33 (8.2) | 30 (7.5) |
| Asian | 15 (3.7) | 19 (4.7) |
| Other | 23 (5.7) | 17 (4.2) |
| Refuses to Answer | 4 (1.0) | 7 (1.7) |
Demographics are reported for all eligible, dispensed subjects. Percentages (n/N x 100) are based on the number of subjects with responses.
## D. Safety and Effectiveness Results
### 1. Safety Results
The analysis of safety was based on the dispensed cohort of 669 subjects available for the 12 month evaluation: 340 Ultra subjects and 329 PureVision subjects. The key safety outcomes for this study are presented below in Tables 9 to 24. Adverse effects are reported in Tables 9 and 10.
### Adverse effects that occurred in the PMA clinical study:
There were no serious adverse events reported for either group.
Overall, 3.0% (24/810) of the Ultra eyes experienced significant non-serious adverse events during the study, compared to 2.4% (20/820) of the PureVision eyes; non-inferiority was met using a predetermined threshold of 5.0%. There were 3 incidents each (0.4%) of peripheral non-progressive non-infectious corneal ulcer associated with the Ultra test and PureVision control lens groups. The adverse events also included 17 (2.1%) Ultra and 8 (1.0%) PureVision incidents of symptomatic corneal infiltrative events, and 6 PureVision control lens ocular events that necessitated ≥2 weeks temporary lens discontinuation. The incidence rates of eyes with significant non-serious adverse events are presented in Table 9.
Table 9 Eyes with Significant Non-Serious Adverse Events
| | Ultra | PureVision |
| --- | --- | --- |
| | n (%) | n (%) |
| Dispensed Eyes (N) | 810 | 820 |
| Total Events | 24 | 20 |
| Eyes with an Event | 24 (3.0) | 20 (2.4) |
| Peripheral Non-Progressive Non-Infectious Corneal Ulcer | 3 (0.4) | 3 (0.4) |
| Symptomatic Corneal Infiltrative Event | 17 (2.1) | 8 (1.0) |
| Corneal Staining Greater than or equal to Grade 3 | 4 (0.5) | 3 (0.4) |
| Temporary Loss of 2 or more lines of BSCVA | 0 | 0 |
| Neovascularizarion of Grade 2 or greater for ≥ 2 weeks | 0 | 0 |
| Ocular Event that necessitates temporary lens discontinuation | 0 | 6 (0.7) |
PMA P170035: FDA Summary of Safety and Effectiveness Data
{15}
for $\geq 2$ weeks
Percentages (n/N x 100) are based on the number of dispensed eyes in each treatment group.
Other adverse events were grouped by graded findings $\geq$ Grade 2 and ungraded findings. There were 50 (6/2%) reports of upper lid tarsal conjunctival abnormalities associated with Ultra test lens compared to 73 (9.6%) with PureVision, and 12 (1.5%) Ultra test lens reports of corneal infiltrates compared to 6 (0.7%) with PureVision. Table 10 summarizes the outcomes for Other Adverse Events.
Table 10 Other Adverse Events (Dispensed Eyes Over All Follow-up Visits)
| | Ultra N=808 n (%) | PureVision N=816 n (%) |
| --- | --- | --- |
| Graded Findings >= Grade 2 | | |
| Asymptomatic corneal infiltrates | 2 (0.2%) | 1 (0.1%) |
| Upper lid tarsal conjunctival abnormalities | 50 (6.2%) | 73 (9.6%) |
| Corneal Infiltrates | 12 (1.5%) | 6 (0.7%) |
| | | |
| Ungraded Findings | | |
| Conjunctivitis | 28 (3.5%) | 30 (3.7%) |
Percentages (n/N x 100) are based on the number of dispensed eyes in each treatment group.
Biomicroscopy findings for each dispensed eye, assessed at the Screening/ Dispensing Visit (baseline) and at each follow-up visit, were graded for severity on a scale from 0 (No Finding) to 4 (Severe Finding). Over All Follow-up Visits, $\geq$ Grade 2 findings were noted in 114 $(14.1\%)$ eyes in the Ultra group and 152 $(18.6\%)$ eyes in the PureVision group. Table 11 provides comparisons between groups for eyes with slit lamp findings $\geq$ Grade 2.
Table 11 Graded Slit Lamp Findings Over All Follow-up Visits, Eyes with Findings $\geq$ Grade 2 (Dispensed Eyes)
| Finding | Ultra N=808 n (%) | | PureVision N=816 n (%) | |
| --- | --- | --- | --- | --- |
| | Absent | Present | Absent | Present |
| Any Finding | 694 (85.9) | 114 (14.1) | 664 (81.4) | 152 (18.6) |
| Epithelial Edema | 804 (99.5) | 4 (0.5) | 812 (99.5) | 4 (0.5) |
| Epithelial Microcysts | 808 (100.0) | 0 | 815 (99.9) | 1 (0.1) |
| Corneal Staining | 775 (96.2) | 31 (3.8) | 739 (91.5) | 69 (8.5) |
| Limbal Injection | 783 (96.9) | 25 (3.1) | 799 (97.9) | 17 (2.1) |
PMA P170035: FDA Summary of Safety and Effectiveness Data
{16}
| Finding | Ultra N=808 n (%) | | PureVision N=816 n (%) | |
| --- | --- | --- | --- | --- |
| | Absent | Present | Absent | Present |
| | | | | |
| Bulbar Injection | 768 (95.0) | 40 (5.0) | 784 (96.1) | 32 (3.9) |
| Corneal Infiltrates | 796 (98.5) | 12 (1.5) | 810 (99.3) | 6 (0.7) |
| Corneal Neovascularization | 808 (100.0) | 0 | 816 (100.0) | 0 |
| Upper Lid Tarsal Conjunctival Abnormalities | 758 (93.8) | 50 (6.2) | 740 (90.7) | 76 (9.3) |
Percentages (n/N x 100) are based on the number of dispensed eyes with non-missing scores.
Table 12 presents Eyes with Device-Related Medically Treated Adverse Events. There were 14 (1.7%) Ultra test lens related events compared to 8 (1.0%) PureVision control lens related events.
Table 12 Eyes with Device-Related Medically Treated Adverse Events (Dispensed Eyes)
| | Ultra N=810 n (%) | PureVision N=820 n (%) |
| --- | --- | --- |
| | | |
| Eyes with medically treated lens-related significant non-serious AEs | | |
| Total Non-Missing | 810 | 820 |
| Yes | 14 (1.7%) | 8 (1.0%) |
| No | 796 (98.3%) | 812 (99.0%) |
| Missing | 0 | 0 |
Percentages (n/N x 100) are based on the number of dispensed eyes with non-missing scores.
Ungraded slit lamp findings were marked as either present or absent for each eye. Over All Follow-up Visits, for Any Finding, there were 146 (18.1%) eyes in the Ultra group and 101 (12.4%) eyes in the PureVision group with ungraded slit lamp findings. Regarding anterior segment abnormalities, which included but were not limited to circumlimbal conjunctival staining, Grade 1-2 inferior papillae, pinguecula, and pterygium, there were 109 (13.5%) reports with the Ultra test lenses, compared to 60 (7.4%) with the PureVision control lenses. Also included were 9 (1.1%) Ultra test lens reported new corneal scars, compared to 3 (0.4%) PureVision control lens new corneal scars. In addition, regarding symptoms associated with corneal infiltrates, there were 19 (2.4%) Ultra test lens reported incidents, compared to 9 (1.1%) PureVision control lens incidents. Table
PMA P170035: FDA Summary of Safety and Effectiveness Data
{17}
13 presents the ungraded slit lamp findings for the Ultra and PureVision groups, based upon all dispensed eyes.
Table 13 Ungraded Slit Lamp Findings (Dispensed Eyes)
| | Eyes With Findings | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Ultra
N=810 | | | | PureVision
N=820 | | | |
| | Absent | | Present | | Absent | | Present | |
| | n | % | n | % | n | % | n | % |
| Any Finding | | | | | | | | |
| Screening/Dispensing | 773 | 95.4 | 37 | 4.6 | 785 | 95.7 | 35 | 4.3 |
| Over All Follow-up Visits | 662 | 81.9 | 146 | 18.1 | 715 | 87.6 | 101 | 12.4 |
| 1-Day Follow-up Visit | 772 | 96.3 | 30 | 3.7 | 796 | 98.5 | 12 | 1.5 |
| 1-Week Follow-up Visit | 743 | 92.2 | 63 | 7.8 | 785 | 97.4 | 21 | 2.6 |
| 1-Month Follow-up Visit | 743 | 94.1 | 47 | 5.9 | 756 | 95.7 | 34 | 4.3 |
| 3-Month Follow-up Visit | 705 | 93.0 | 53 | 7.0 | 727 | 96.7 | 25 | 3.3 |
| 6-Month Follow-up Visit | 677 | 94.0 | 43 | 6.0 | 705 | 96.3 | 27 | 3.7 |
| 9-Month Follow-up Visit | 647 | 93.0 | 49 | 7.0 | 663 | 96.4 | 25 | 3.6 |
| 12-Month Follow-up Visit | 645 | 94.6 | 37 | 5.4 | 638 | 97.0 | 20 | 3.0 |
Percentages (n/N x 100) are based on the number of eyes with non-missing scores in each treatment group.
High contrast distance BSCVA was obtained at the Screening/Dispensing Visit (baseline) and again at the Exit Visit. VA was calculated as the $\mathrm{Log_{10}}$ of the Minimum Angle of Resolution (logMAR) by converting the number of letters correctly identified to the logMAR equivalent. An eye was considered to have worsened 2 or more Snellen lines if the difference in logMAR VA $\geq 0.2$ (exit BSCVA - baseline BSCVA). Over All Follow-up Visits, there were no eyes in either group with a 2 or more line decrease in BSCVA from baseline to exit (refer to Table 14).
Table 14 High Contrast Distance VA Line Change: Baseline BSCVA Versus Exit BSCVA (Dispensed Eyes) Eyes With a 2 or More Line Decrease in VA
| | Ultra
N=810 | | PureVision
N=820 | |
| --- | --- | --- | --- | --- |
| | Absent
n (%) | Present
n (%) | Absent
n (%) | Present
n (%) |
| Exit BSCVA | 776 (100.0) | 0 | 772 (100) | 0 |
Percentages (n/N x 100) are based on the number of dispensed eyes with non-missing scores in each treatment group.
PMA P170035: FDA Summary of Safety and Effectiveness Data
{18}
Symptoms/complaints were collected for each eye for each subject at the Screening/ Dispensing Visit after the study lens was inserted and at each scheduled visit based on the subject's experience with the study lenses. Subject symptoms and complaints were assessed on a scale from 0 to 100, with 0 denoting the most unfavorable symptoms/complaints; the rating scale included descriptors at 20-point intervals to help standardize subject responses. The outcomes indicated there were generally more reports of unfavorable symptoms/complaints associated with the PureVision lens group. Table 15 presents the proportion of eyes with unfavorable symptoms/complaints (scores < 50) compared between the Ultra and PureVision groups, based upon all dispensed eyes.
Table 15 Summary of Unfavorable Symptoms/Complaints: Over All Follow-up Visits (Dispensed Eyes)
| Symptoms/Complaints | Ultra N=810 n (%) | PureVision N=820 n (%) |
| --- | --- | --- |
| Burning/Stinging Upon Insertion | 28 (3.5) | 48 (6.0) |
| Comfort Upon Insertion | 29 (3.6) | 46 (5.7) |
| Vision Upon Insertion | 15 (1.9) | 18 (2.2) |
| Overall Comfort | 72 (8.9) | 130 (15.9) |
| Comfort at End of Day | 86 (10.6) | 150 (18.4) |
| Dryness | 106 (13.1) | 208 (25.5) |
| Ease of Handling/Insertion | 24 (3.0) | 22 (2.7) |
| Ease of Handling/Removal | 20 (2.5) | 30 (3.7) |
| Itchiness | 82 (10.1) | 96 (11.8) |
| Lens Cleanness Upon Removal | 42 (5.2) | 86 (10.7) |
| Lens Handling | 11 (1.4) | 20 (2.5) |
| Overall Impression | 45 (5.6) | 93 (11.4) |
| Redness | 86 (10.6) | 65 (8.0) |
| Vision | 35 (4.3) | 64 (7.8) |
| Vision at End of Day | 44 (5.4) | 76 (9.3) |
| Vision in Low Light | 32 (4.0) | 67 (8.2) |
Percentages (n/N x 100) are based on the number of eyes with non-missing scores in each treatment group. Scores are captured on a scale from 0 to 100, with 100 being the most favorable score. Unfavorable symptoms/complaints are defined as scores < 50.
Lens wettability was assessed at the Screening/Dispensing Visit on the newly dispensed lens and at each follow-up visit, using a scale of Optimal, Slight, Mild, Moderate and Severe. Over All Follow-up Visits, 294 (36.3%) eyes in the Ultra group exhibited suboptimal (< 100% of the anterior lens surface rated wettable)
PMA P170035: FDA Summary of Safety and Effectiveness Data
Page 19
{19}
lens wettability, compared to 308 (37.9%) eyes in the PureVision group. Table 16 presents lens wettability over all follow-up visits.
Table 16 Lens Wettability Over All Follow-up Visits (Dispensed Eyes)
| | Ultra
N=810
n (%) | PureVision
N=820
n (%) |
| --- | --- | --- |
| Optimal:
100% of anterior lens
surface is wettable | 516 (63.7) | 504 (62.1) |
| Slight:
Presence of small (<0.1
mm), individual,
discrete, non-wetting
areas | 196 (24.2) | 205 (25.2) |
| Mild:
Presence of single area
of non-wetting between
0.1 mm and 0.5 mm in
size | 52 (6.4) | 47 (5.8) |
| Moderate:
Presence of several
areas of non-wetting,
each between 0.1 mm
and 0.5 mm in size | 45 (5.6) | 47 (5.8) |
| Severe:
Presence of one or more
non-wetting areas
greater than
0.5 mm in size | 1 (0.1) | 9 (1.1) |
Percentages (n/N x 100) are based on the number of eligible, dispensed eyes with non-missing scores in each treatment group. Events with multiple visits in a visit category are counted once for the most severe lack of wettability. Over All Follow-up Visits summarizes the worst case over all follow-up visits.
Three qualities of lens deposits were assessed at the Screening/ Dispensing Visit (for the dispensed lenses) and at each follow-up visit. Type of deposits were captured as none, crystalline, crust-like, film, or spots; percent coverage of deposits were assessed as none, 1-25%, 26-50%, 51-75% or 76-100%; and degree of deposits were assessed as none, light, medium, or heavy. Over All Follow-up Visits, 60 (7.4%) eyes in the Ultra group and 83 (10.2%) eyes in the PureVision group had suboptimal lens deposits (a deposit degree rating of medium or heavy, regardless of the type or percent coverage). Tables 17 and 18 present summaries of the proportions of eyes with suboptimal lens deposits.
PMA P170035: FDA Summary of Safety and Effectiveness Data
Page 20
{20}
Table 17
Eyes with Suboptimal Lens Deposits (Dispensed Eyes)
| Visit | Ultra N=810 | | PureVision N=812 | |
| --- | --- | --- | --- | --- |
| | Absent n (%) | Present n (%) | Absent n (%) | Present n (%) |
| Over All Follow-up Visits | 750 (92.6) | 60 (7.4) | 729 (89.8) | 83 (10.2) |
Percentages (n/N x 100) are based on the number of dispensed eyes with non-missing scores in each treatment group. The total scores are used for analysis. Suboptimal lens deposits are defined as a degree of deposits of medium or heavy.
Table 18
Lens Deposits: Over All Follow-Up Visits by Deposit Category (Dispensed Eyes)
| | Ultra n (%) | PureVision n (%) |
| --- | --- | --- |
| Total Non-missing (N) | 810 | 812 |
| None | 334 (41.2) | 313 (38.5) |
| Light | 416 (51.4) | 416 (51.2) |
| Medium | 60 (7.4) | 79 (9.7) |
| Heavy | 0 | 4 (0.5) |
Percentages (n/N x 100) are based on the number of dispensed eyes with non-missing scores in each treatment group. Over All Follow-up Visits summarizes the worst case over all follow-up visits.
Contact lens centration was assessed at the Screening/Dispensing Visit on the newly dispensed lens and at each follow-up visit, using the following scale:
- Excellent (fully centered);
- Good (slight decentration, no corneal exposure);
- Fair (decentration, intermittent corneal exposure);
- Poor (incomplete corneal coverage and/or edge lift).
Suboptimal lens centration was defined as any rating other than excellent. Over All Follow-up Visits, lens centration was reported as excellent in 715 (88.3%) eyes in the Ultra group and 737 (90.8%) eyes in the PureVision group (Table 19). Suboptimal lens centration is depicted in Table 20.
PMA P170035: FDA Summary of Safety and Effectiveness Data
{21}
Table 19 Lens Centration Over All Follow-up Visits (Dispensed Eyes)
| | Ultra n (%) | PureVision n (%) |
| --- | --- | --- |
| Total Non-Missing (N) | 810 | 812 |
| Excellent (fully centered | 715 (88.3) | 737 (90.8) |
| Good (slight decentration, no corneal exposure) | 89 (11.0) | 71 (8.7) |
| Fair (decentration, intermittent corneal exposure) | 2 (0.2) | 3 (0.4) |
| Poor (incomplete corneal coverage and/or edge lift) | 4 (0.5) | 1 (0.1) |
Percentages (n/N x 100) are based on the number of dispensed eyes with non-missing scores in each treatment group. Over All Follow-up Visits summarizes the worst case over all follow-up visits.
Table 20 Eyes with Suboptimal Lens Centration (Dispensed Eyes)
| Visit | Ultra N=810 | | PureVision N=820 | |
| --- | --- | --- | --- | --- |
| | Absent n (%) | Present n (%) | Absent n (%) | Present n (%) |
| Screening/Dispensing | 772 (95.5) | 36 (4.5) | 806 (98.3) | 14 (1.7) |
| Over All Follow-up Visits | 715 (88.3) | 95 (11.7) | 737 (90.8) | 75 (9.2) |
Percentages (n/N x 100) are based on the number of dispensed eyes with non-missing scores in each treatment group.
Contact lens movement was assessed at the Screening/Dispensing Visit on the newly dispensed lens and at each follow-up visit using the following scale:
Adequate
- Excessive $(>0.6\mathrm{mm})$
Insufficient $(< 0.2\mathrm{mm})$
- Adherence
Over All Follow-up Visits, lens movement was reported as adequate for 730 $(90.1\%)$ eyes in the Ultra group and 771 $(95.0\%)$ eyes in the PureVision group (Table 21). Suboptimal lens movement is defined as a rating other than adequate. There were 80 $(9.9\%)$ eyes in the Ultra group and 41 $(5.0\%)$ eyes in the PureVision group. Over All Follow-up Visits with suboptimal lens movement. The majority of the Ultra suboptimal lens movement was due to insufficient movement; see Table 22. In addition, over All Follow-up Visits, there were 6 $(0.7\%)$ eyes in the Ultra group and 4 $(0.5\%)$ eyes in the PureVision groups with lens adherence.
PMA P170035: FDA Summary of Safety and Effectiveness Data
{22}
Table 21 Suboptimal Lens Movement (Dispensed Eyes)
| Visit | Ultra N=810 | | PureVision N=820 | |
| --- | --- | --- | --- | --- |
| | Absent n (%) | Present n (%) | Absent n (%) | Present n (%) |
| Over All Follow-up Visits | 730 (90.1) | 80 (9.9) | 771 (95) | 41 (5.0) |
Percentages (n/N x 100) are based on the number of dispensed eyes with non-missing scores in each treatment group. The Total scores are used for analysis. Suboptimal lens movement is defined as a score other than Adequate.
Table 22 Lens Movement Over All Follow-up Visits (Dispensed Eyes)
| | Ultra n (%) | PureVision n (%) |
| --- | --- | --- |
| Total Non-Missing (N) | 810 | 812 |
| Adequate | 730 (90.1) | 771 (95.0) |
| Excessive (> 0.6 mm) | 4 (0.5) | 13 (1.6) |
| Insufficient (< 0.2 mm) | 70 (8.6) | 24 (3.0) |
| Adherence | 6 (0.7) | 4 (0.5) |
Percentages (n/N x 100) are based on the number of dispensed eyes with non-missing scores in each treatment group. Over All Follow-up Visits summarizes the worst case over all follow-up visits where Adherence is considered most severe, followed by Insufficient and Excessive.
Table 23 summarizes keratometry changes from baseline to Exit Visit for all dispensed eyes. There were 25 eyes (3.2%) in the Ultra group and 19 eyes (2.5%) in the PureVision group with a change in keratometry (absolute value of $\geq 1.00$ D).
Table 23 Keratometry Changes (Absolute Value) from Baseline to Exit Visit in the Meridian with the Largest Change (Dispensed Eyes)
| | Ultra N=776 n (%) | PureVision N=772 n (%) |
| --- | --- | --- |
| Diopters | | |
| 0.00-0.99 | 751 (96.8) | 753 (97.5) |
| 1.00-1.99 | 22 (2.8) | 13 (1.7) |
| 2.00-2.99 | 2 (0.3) | 2 (0.3) |
| 3.00-3.99 | 1 (0.1) | 4 (0.5) |
| 4.00-4.99 | 0 | 0 |
| ≥5.00 | 0 | 0 |
| Missing | 34 | 48 |
Percentages (n/N x 100) are based on the number of non-missing dispensed eyes with responses. Baseline measurements are obtained at the Screening/Dispensing Visit.
PMA P170035: FDA Summary of Safety and Effectiveness Data
{23}
Table 24 summarizes spherocylindrical refraction changes from baseline to Exit Visit for all dispensed eyes. There were 2 eyes (0.3%) in the Ultra group and 5 eyes (0.6%) in the PureVision group with a change in refraction of ≥ 1.00 D.
Table 24
Spherocylindrical Refractive Changes (Absolute Value) from Baseline to Exit Visit (Dispensed Eyes)
| | Ultra
N=776 | PureVision
N=772 |
| --- | --- | --- |
| Spherocylindrical Refractive Absolute Change (D) | | |
| Mean (SD) | 0.10 (0.16) | 0.11 (0.17) |
| Median | 0.00 | 0.00 |
| Min, Max | 0.00, 1.00 | 0.00, 1.00 |
| | | |
| Categorical Summary: n (%) | | |
| 0.00 - 0.99D | 774 (99.7%) | 767 (99.4%) |
| 1.00 - 1.99D | 2 (0.3%) | 5 (0.6%) |
| 2.00 - 2.99D | 0 | 0 |
| 3.00 - 3.99D | 0 | 0 |
| 4.00 - 4.99D | 0 | 0 |
| ≥5.00 | 0 | 0 |
Percentages (n/N x 100) are based on the number of non-missing dispensed eyes in each treatment group with responses. Spherocylindrical refraction, or Mean Refractive Spherical Equivalent (MRSE) is calculated as sphere measurement + (cylinder measurement/2).
2. Effectiveness Results
The analysis of effectiveness was based on the Eligible, Dispensed cohort of 658 subjects at the 12-month time point.
Key effectiveness outcomes are presented in Tables 25 to 29.
The primary effectiveness endpoint was high contrast, distance visual acuity (VA) with dispensed lenses at the 12-Month Follow-up Visit. All eyes dispensed study lenses for eligible subjects were included in the analysis under the treatment they actually received. High contrast distance lens VA was obtained at the Screening/Dispensing Visit, and at each follow-up visit. Scores were recorded as the numbers of letters correctly identified in the eye examination. VA was converted to the Log10 of the Minimum Angle of Resolution (logMAR) by using the score (number of letters). The Ultra lens was found to be non-inferior to the PureVision lens with regard to distance high contrast lens VA with a non-inferiority margin of 0.06. VA at the 12-Month visit was comparable between treatments (Table 25). Approximately 97% of subject eyes in the clinical study achieved at least 20/25 with the Bausch + Lomb ULTRA (samfilcon A) contact lens.
PMA P170035: FDA Summary of Safety and Effectiveness Data
Page 24
{24}
Table 25 High Contrast Distance logMAR Lens VA at the 12-Month Follow-up Visit (Eligible, Dispensed Eyes)
| Ultra | | PureVision | |
| --- | --- | --- | --- |
| n | LSMean (SE) | n | LSMean (SE) |
| 667 | -0.063 (0.002) | 649 | -0.048 (0.002) |
Analyses are based on the number of eligible, dispensed eyes with non-missing scores in each treatment group.
LSMean = Least Square Mean
The secondary effectiveness endpoint was lens wear time reported as average extended lens wear time (days/week) since last visit. For this parameter, the subjects in both groups reported mean extended wearing times of 6.7 days per week over all visits (see Table 26).
Table 26 Average Weekly Wear Time (Days per Week) (Eligible, Dispensed Subjects)
| Visit | Ultra | | PureVision | |
| --- | --- | --- | --- | --- |
| | n | LSMean (SE) | n | LSMean (SE) |
| Over All Follow-up Visits | 400 | 6.7 (0.031) | 396 | 6.7 (0.031) |
Over All Follow-up Visits is based on the average weekly wear time over all follow-up visits.
With regard to the line change in visual acuity from baseline over all follow-up visits, Over All Follow-up Visits, 25 (3.0%) eyes in the Ultra group and 30 (3.8%) eyes in the PureVision group experienced a worsening of $\geq 2$ lines ( $\geq 10$ letters) from Dispensing Lens VA at any time point. Table 27 presents the line change in visual acuity from baseline over all follow-up visits.
Table 27 High Contrast Distance Lens VA Line Change: Screening/Dispensed Lens VA versus Follow-up Lens VA Over All Follow-up Visits (Eligible, Dispensed Eyes)
| Line Change | Ultra N=806 n (%) | PureVision N=796 n (%) |
| --- | --- | --- |
| -9 | 1 (0.1) | 0 |
| -8 | 0 | 0 |
| -7 | 0 | 0 |
| -6 | 0 | 0 |
| -5 | 0 | 0 |
| -4 | 2 (0.2) | 2 (0.3) |
| -3 | 2 (0.2) | 3 (0.4) |
| -2 | 20 (2.5) | 25 (3.1) |
| -1 | 198 (24.6) | 261 (32.8) |
PMA P170035: FDA Summary of Safety and Effectiveness Data
{25}
Percentages (n/N x 100) are based on the number of eligible, dispensed eyes with non-missing scores in each treatment group. Positive line changes reflect improved VA. Over All Follow-up Visits summarizes the worst case over all the follow-up visits.
Contact Lens Performance was evaluated at all visits for a number of parameters, based upon a scale of 0 (no symptoms) to 4 (severe symptoms: the lens was very irritating or annoying, and could not be tolerated). For each performance parameter, a categorical parameter was defined as unfavorable if the evaluation was a 3 (moderate) or a 4 (severe), and favorable if the evaluation was 0-2 (no, slight, or mild symptoms).
Over All Follow-up Visits, there were differences between the Ultra group and PureVision group in the proportion of eyes with unfavorable lens performance ratings, as there was a higher proportion of eyes with unfavorable lens performance ratings for the PureVision group for Discomfort, Spectacle Blur, Variable Vision, Blurred Vision, and Lens Needs Cleaning. Over All Follow-up Visits, there were similar proportions of eyes between the two groups with unfavorable lens performance ratings for the remaining parameters which were assessed in Table 28.
Table 28 Eyes with Unfavorable Lens Performance Ratings Over All Scheduled Follow-up Visits (Eligible, Dispensed Eyes)
| Lens Performance Parameter | Ultra N=805 n (%) | PureVision N=800 n (%) |
| --- | --- | --- |
| Discomfort | 128 (15.9) | 181 (22.6) |
| Excessive Tearing | 24 (3.0) | 16 (2.0) |
| Photophobia | 21 (2.6) | 11 (1.4) |
| Halos | 9 (1.1) | 13 (1.6) |
| Itching/Burning | 67 (8.3) | 68 (8.5) |
| Spectacle Blur | 8 (1.0) | 24 (3.0) |
| Variable Vision | 22 (2.7) | 39 (4.9) |
| Blurred Vision | 43 (5.3) | 66 (8.3) |
| Lens Needs Cleaning | 48 (6.0) | 88 (11.0) |
| Handling | 12 (1.5) | 22 (2.8) |
Only data from scheduled visits are used. Percentages (n/N x 100) are based on the number of eyes with non-missing scores in each treatment group. An unfavorable performance is a score of 3 or 4.
PMA P170035: FDA Summary of Safety and Effectiveness Data
{26}
At the start of each follow-up visit, as part of the worn lens evaluation, the subject was asked whether they used rewetting drops. The frequency (count and percent) of subjects using rewetting drops is provided with percent based on the total non-missing responses. Over All Follow-up Visits, 722 (89.6%) eyes in the Ultra group used rewetting drops compared to 748 eyes (93.5%) in the PureVision group (Table 29).
Table 29
Rewetting Drop Usage Over All Follow-up Visits
(Eligible, Dispensed Eyes)
| Used Rewetting Drops | |
| --- | --- |
| Ultra
N=806
n (%) | PureVision
N=804
n (%) |
| Yes (%) | Yes (%) |
| 722 (89.6) | 748 (93.5) |
Summaries are based on usage reported at scheduled visits. The percentages (n/N x 100) are based on the number of total non-missing eyes.
3. Subgroup Analyses
In this premarket application, there were no preoperative characteristics that were evaluated for potential association with outcomes.
4. Pediatric Extrapolation
In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population.
E. Financial Disclosure
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 34 investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data.
XI. SUMMARY OF SUPPLEMENTAL CLINICAL INFORMATION
An Overnight Corneal Swelling Study was conducted to demonstrate the percentage change in overnight average central 2mm corneal thickness was not more than 2% greater with overnight wear of the Bausch + Lomb Ultra (samfilcon A) contact lens compared to no contact lens wear in the same eye of the same participants.
PMA P170035: FDA Summary of Safety and Effectiveness Data
{27}
The applicant also conducted two feasibility clinical studies to evaluate the comfort and fitting characteristics on the Ultra Contact Lens when treated with OPTI-FREE PureMoist Multi-Purpose Disinfecting Solution.
No issues regarding device safety or lack of effectiveness were raised by the results from these studies.
## XII. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION
In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Ophthalmic Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
## XIII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES
### A. Effectiveness Conclusions
The overall effectiveness of the Bausch + Lomb Ultra (samfilcon A) contact lens was demonstrated based upon the 12-month results of the IDE clinical investigation.
The primary effectiveness endpoint was high contrast, distance visual acuity (VA) with dispensed lenses at the 12-Month Follow-up Visit. At the 12-Month Follow-up Visit, mean LogMAR was -0.063 for the Ultra test lens and -0.048 for the PureVision control lens, indicating comparable outcomes. Approximately 97% of subject eyes in the clinical study achieved at least 20/25 with the Bausch + Lomb ULTRA (samfilcon A) contact lens.
The secondary effectiveness endpoint was lens wear time reported as average extended lens wear time (days/week) since last visit. The average wearing time since the last visit was 6.7 (±0.031) days for both the Ultra and the PureVision groups over all follow-up visits.
With regard to the line change in visual acuity from baseline, Over All Follow-up Visits, 25 (3.0%) eyes in the Ultra group and 30 (3.8%) eyes in the PureVision group experienced a worsening of ≥ 2 lines (≥ 10 letters) from Dispensing Lens VA.
Over All Follow-up Visits, there were significant differences between the Ultra group and PureVision group in the proportion of eyes with unfavorable lens performance ratings. There was a higher proportion of eyes with unfavorable lens performance ratings for the PureVision group for Discomfort, Spectacle Blur, Variable Vision, Blurred Vision, and Lens Needs Cleaning. Over All Follow-up Visits, the proportions of eyes with unfavorable lens performance ratings were similar between treatment groups or the remaining parameters which were assessed.
PMA P170035: FDA Summary of Safety and Effectiveness Data
Page 28
{28}
Regarding the usage of rewetting drops, the frequency of subjects using rewetting drops was determined with percent based on the total non-missing responses. Over All Follow-up Visits, 722 (89.6%) eyes in the Ultra group used rewetting drops compared to 748 eyes (93.5%) in the PureVision group.
The effectiveness results from the PMA clinical trial provided evidence that the study outcomes met the acceptance criteria.
## B. Safety Conclusions
The risks of the device are based on nonclinical laboratory studies as well as data collected in the pivotal clinical study conducted to support PMA approval as described above.
The primary safety endpoint was the rate of serious or significant non-serious adverse events during the 12-month follow-up. There were no serious adverse events reported for either lens group.
Overall, 3.0% (24/810) of the Ultra eyes experienced significant non-serious adverse events during the study, compared to 2.4% (20/820) of the PureVision eyes; non-inferiority was met using a predetermined threshold of 5.0%.
Biomicroscopy findings for each dispensed eye were graded for severity on a scale from 0 to 4. Over All Follow-up Visits, ≥ Grade 2 findings were noted in 114 (14.1%) eyes in the Ultra group and 152 (18.6%) eyes in the PureVision group.
Ungraded slit lamp findings were reported. Over All Follow-up Visits, for Any Finding there were 146 (18.1%) eyes in the Ultra group and 101 (12.4%) eyes in the PureVision group with ungraded slit lamp findings.
High contrast distance BSCVA was obtained at the Screening/Dispensing Visit (baseline) and at the Exit Visit. An eye was considered to have worsened 2 or more Snellen lines if the difference in logMAR VA ≥ 0.2 (exit BSCVA – baseline BSCVA). Over All Follow-up Visits, there were no eyes in either group with a 2 or more line decrease in BSCVA from baseline to exit.
Symptoms/complaints were collected for each eye for each subject at the Screening/Dispensing Visit following lens insertion and at follow-up. The outcomes indicated there were generally more reports of unfavorable symptoms/complaints associated with the PureVision lens group.
Lens wettability was assessed at the Screening/Dispensing Visit on the newly dispensed lens and at each follow-up visit, using a scale of Optimal, Slight, Mild, Moderate, and Severe. Over All Follow-up Visits, the proportion of eyes with suboptimal lens wettability, 294 (36.3%) eyes in the Ultra group, was comparable to the 308 (37.9%) eyes in the PureVision group.
PMA P170035: FDA Summary of Safety and Effectiveness Data
Page 29
{29}
Suboptimal degree of lens deposits was defined as a deposit degree rating of medium or heavy, regardless of the type or percent coverage. Over All Follow-up Visits, 60 (7.4%) eyes in the Ultra group and 83 (10.2%) eyes in the PureVision group had suboptimal lens deposits.
Contact lens centration was assessed at the Screening/Dispensing Visit on the newly dispensed lens and at each follow-up visit, using a scale of Excellent to Poor. Suboptimal lens centration was defined as any rating other than excellent. Over All Follow-up Visits, lens centration was reported as excellent in 715 (88.3%) eyes in the Ultra group and 737 (90.8%) eyes in the PureVision group.
Contact lens movement was assessed at the Screening/Dispensing Visit on the newly dispensed lens and at each follow-up visit. Over All Follow-up Visits, lens movement was reported as adequate for 730 (90.1%) eyes in the Ultra group and 771 (95.0%) eyes in the PureVision group. There were 80 (9.9%) eyes in the Ultra group and 41 (5.0%) eyes in the PureVision group. Over All Follow-up Visits with suboptimal lens movement. The majority of the Ultra suboptimal lens movement was due to insufficient movement.
Keratometry changes from baseline to Exit Visit were determined for all dispensed eyes. There were 25 eyes (3.2%) in the Ultra group and 19 eyes (2.5%) in the PureVision group with a change in keratometry (absolute value of ≥ 1.00 D).
Regarding spherocylindrical refraction changes from baseline to Exit Visit for all dispensed eyes, there were 2 eyes (0.3%) in the Ultra group and 5 eyes (0.6%) in the PureVision group with a change in refraction of ≥ 1.00 D.
The safety results from the PMA clinical trial provided evidence that the study outcomes met the acceptance criteria.
## C. Benefit-Risk Determination
The probable benefits of the device are based on data collected in a clinical study conducted to support PMA approval and other clinical studies as described above. The benefits of the subject device include vision correction, as well as the convenience of wearing contact lenses overnight, i.e., without removal, for up to 1 week.
Additional factors to be considered in determining probable risks and benefits for the Bausch + Lomb Ultra (samfilcon A) Soft (hydrophilic) Contact Lens include:
a. The results of the clinical study can be considered generalizable to the intended market or target patient population.
PMA P170035: FDA Summary of Safety and Effectiveness Data
Page 30
{30}
b. Clinical data were collected using a study design that included randomized treatment and masking of subjects and evaluators.
c. Medical adverse events and complications (e.g., risks of infection, inflammation, corneal edema, etc.) are similar to those associated with most other extended wear contact lenses.
1. Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
In conclusion, given the available information above, the data support that for the visual correction of refractive ametropia, the probable benefits of the Bausch + Lomb Ultra (samfilcon A) Contact Lens outweigh the probable risks.
D. Overall Conclusions
The data in this premarket application support the reasonable assurance of safety and effectiveness of the subject device when used in accordance with the Indications for Use. The effectiveness endpoints related to adequate high contrast, distance visual acuity and lens wearing time were met, demonstrating the ability of the Bausch + Lomb Ultra (samfilcon A) Contact Lenses to correct ametropia as well as to provide acceptable lens comfort. The incidence rates of eyes with significant non-serious adverse events were comparable between the two lens groups.
XIV. CDRH DECISION
CDRH issued an approval order on April 30, 2018.
The applicant's manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820).
XV. APPROVAL SPECIFICATIONS
Directions for use: See device labeling.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling.
Post-approval Requirements and Restrictions: See approval order.
PMA P170035: FDA Summary of Safety and Effectiveness Data
Page 31
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.