P170022S003 · Arj Medical, Inc. · OZA · Jan 11, 2024 · Microbiology
Device Facts
Record ID
P170022S003
Device Name
PyloPlus® UBT System
Applicant
Arj Medical, Inc.
Product Code
OZA · Microbiology
Decision Date
Jan 11, 2024
Decision
APPR
Regulation
21 CFR 866.3110
Device Class
Class 3
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
P170022S003 · Jan 11, 2024
PyloPlus® UBT System
Arj Medical, Inc.
Published clinical literature (meta-analysis and individual clinical studies)
The sponsor performed a literature search to identify existing clinical data (including a meta-analysis by Leal et al., 2011, and other studies from 1997-2009) to support the extrapolation of adult efficacy data to the pediatric population (ages 3-17).
Pediatric extrapolation; Literature review; Meta-analysis
The PyloPlus UBT system is intended for use in the qualitative detection of urease associated with H. pylori in the human stomach and is indicated as an aid in the initial diagnosis and post treatment monitoring of H. pylori infection in adults and pediatric patients ages 3-17 years old. The PyloPlus UBT system consists of the PyloPlus UBT Kit and the PyloPlus UBT analyzer. The analyzer is an infrared Spectrometer used for the measurement of the ratio of 13CO2 to 12CO2 in breath samples. The PyloPlus UBT system is for use by trained health care professionals as prescribed by a physician.
Device Story
PyloPlus UBT System is a diagnostic combination product consisting of a 13C-urea/citric acid drug component and an infrared spectrophotometer analyzer. Procedure: patient ingests reconstituted 13C-urea/citric acid; urease from gastric H. pylori decomposes urea into 13CO2 and NH4+; 13CO2 is absorbed into blood and exhaled. Analyzer measures 13CO2/12CO2 ratio in baseline and post-dose breath samples. Output: Delta Over Baseline (DOB) value. Used by trained healthcare professionals in clinical settings. DOB >= 3.0 indicates positive H. pylori infection; DOB < 3.0 indicates negative. Results aid physicians in diagnosing infection and monitoring post-treatment status, potentially avoiding invasive procedures like endoscopy.
Clinical Evidence
Clinical study evaluated 137 subjects (80 adults, 57 pediatric). Efficacy analysis included 116 evaluable subjects (76 adults, 40 pediatric) compared to FDA-cleared stool antigen test. Adult post-therapy cohort: 76% sensitivity, 98% specificity. Pediatric initial diagnosis cohort: 100% sensitivity, 89.9% specificity. Safety: no adverse events reported during clinical study. Pediatric efficacy supported by leveraging adult clinical data and literature meta-analysis (Leal et al., 2011) demonstrating high sensitivity/specificity in children.
Technological Characteristics
Combination product: 13C-urea/citric acid drug component and infrared spectrophotometer analyzer. Analyzer is a standalone unit manufactured by FAN. Measures 13C/12C isotope ratio in breath samples. DOB calculation based on baseline vs. post-dose breath samples. 13C-urea stability: 24 months at room temperature.
Indications for Use
Indicated for qualitative detection of urease associated with H. pylori in the human stomach; aid in initial diagnosis and post-treatment monitoring of H. pylori infection in adults and pediatric patients ages 3-17 years.
Regulatory Classification
Identification
Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
## I. GENERAL INFORMATION
Device Generic Name: Urea Breath Test (UBT)
Device Trade Name: PyloPlus® UBT System
Device Procode: OZA
Applicant's Name and Address: ARJ Medical, Inc.
209 State Street East
Oldsmar, FL 34677
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P170022/S003
Date of FDA Notice of Approval: January 11, 2024
The original PMA (P170022) was approved on February 18, 2020, and is indicated for use in the qualitative detection of urease associated with *H. pylori* in the human stomach and is indicated as an aid in the initial diagnosis of *H. pylori* infection in adults 18 years old and older. The SSED to support the indication is available on the CDRH website and is incorporated by reference here. The current supplement was submitted to expand the indication for the PyloPlus® UBT System
## II. INDICATIONS FOR USE
The PyloPlus UBT system is intended for use in the qualitative detection of urease associated with *H. pylori* in the human stomach and is indicated as an aid in the initial diagnosis and post treatment monitoring of *H. pylori* infection in adults and pediatric patients ages 3-17 years old. The PyloPlus UBT system consists of the PyloPlus UBT Kit and the PyloPlus UBT analyzer. The analyzer is an infrared Spectrometer used for the measurement of the ratio of $^{13}\text{CO}_2$ to $^{12}\text{CO}_2$ in breath samples. The PyloPlus UBT system is for use by trained health care professionals as prescribed by a physician.
## III. CONTRAINDICATIONS
There are no known contraindications.
## IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the PyloPlus UBT System labeling.
---PMA P170022/S003: FDA Summary of Safety and Effectiveness Data
1 of 12
{1}
## **V. DEVICE DESCRIPTION**
The PyloPlus® UBT System (Urea Breath Test (UBT)) is intended for use in the qualitative detection of urease associated with *Helicobacter pylori* (*H. pylori*) in the human stomach. The PyloPlus UBT is a combination product that includes a diagnostic drug component, $^{13}$C urea/citric acid. The USP (U.S. Pharmacopeia) monograph for $^{13}$C-urea was provided and supports a stability claim of 24 months at room temperature under stable humidity. In the PyloPlus UBT, the pouched $^{13}$C-urea and pouched flavoring packet are reconstituted with water and then ingested by the subject. The $^{13}$C-urea is decomposed by urease associated with gastric *H. pylori* forming $^{13}$CO$_2$ and NH$_4^+$. The $^{13}$CO$_2$ is absorbed into the blood, and then exhaled in the breath. The result of the PyloPlus UBT is provided as the Delta Over Baseline (DOB) which is the difference between the ratio of $^{13}$CO$_2$/$^{12}$CO$_2$ in the post-dose sample and the corresponding ratio in the baseline sample. Analysis of the breath samples is performed by the PyloPlus UBT Analyzer. The analyzer is an infrared spectrophotometer that can distinguish the difference between $^{13}$C isotope and $^{12}$C that is most predominant in air. The Spectrophotometer is manufactured by FAN and is a standalone unit that measures the amount of $^{13}$C in each breath bag and then calculates the change in $^{13}$C between the baseline breath sample and post-dose breath sample. From the change in $^{13}$C levels, a ratio or DOB can be calculated. A DOB result $\geq 3.0$ is interpreted as positive for *H. pylori* infection, and a result $< 3.0$ is interpreted as negative for *H. pylori* infection in adult subjects.
## **VI. ALTERNATIVE PRACTICES AND PROCEDURES**
There are several other alternative methods for the detection of *H. pylori* in human specimens. One of the diagnostic methods entails an endoscopy procedure to obtain gastric biopsy and perform histology, immunohistochemical stains, culture, and rapid urease tests. Additional methods include serological assays to detect immunoglobulins (A, G, or M antibodies) to *H. pylori*, stool antigen test and other urea breath tests. Each alternative has its own advantages and disadvantages. A subject should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
## **VII. MARKETING HISTORY**
The PyloPlus UBT has been marketed in the US since its original approval on February 18, 2020.
---PMA P170022/S003: FDA Summary of Safety and Effectiveness Data
2 of 12
{2}
### VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Below is a list of the potential adverse effects (e.g., complications) associated with the use of the device.
Historical postmarket experience:
During post-approval use of UBTs, the following adverse events have been identified: anaphylactic reaction, hypersensitivity, rash, burning sensation in stomach, tingling in the skin, vomiting, and diarrhea. Because these reactions are reported voluntarily from a population of uncertain size, it is not always possible to establish a causal relationship to drug exposure.
There have been no adverse effects associated with the use of the device in the clinical studies, please see Section X below.
### IX. SUMMARY OF NON-CLINICAL STUDIES
#### A. Laboratory Studies
No new laboratory studies were conducted for the added target population claims.
#### B. Animal Studies
N/A
#### C. Additional Studies
N/A
### X. SUMMARY OF PRIMARY CLINICAL STUDY(IES)
The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of detection of *H. pylori* with the PyloPlus UBT System for the qualitative detection of urease associated with *H. pylori* in the human stomach and for the indication as an aid in the initial diagnosis and post treatment monitoring of *H. pylori* infection in adults and children 3-17 years old. Data from this clinical study were the basis for the Panel track PMA approval decision. A summary of the clinical study is presented below.
---PMA P170022/S003: FDA Summary of Safety and Effectiveness Data
3 of 12
{3}
## **A. Study Design**
Adult subjects were treated between March 29, 2023, and May 16, 2023. The database for this Panel Track Supplement reflected data collected through May 2023 and included 80 subjects. There were 2 investigational sites.
The pediatric subjects were tested between November 14, 2022, to June 6, 2023. The database for this Panel Track Supplement reflected data collected through June 2023 and included 57 subjects. There were 5 investigational sites.
The studies were separate, multi-center, non-randomized, open label, clinical studies. Subjects were enrolled on a walk-in basis. Enrollment sites utilized a PyloPlus UBT Analyzer to document the results, but the results remained blinded to the investigator and treating physicians and no treatment decisions were made based on the investigational device results. Pediatric subjects were treatment naïve and adult subjects had been treated for *H. pylori* within the past 6 months at the time of testing.
The urea breath test uses a breath sample, pre-drug consumption as the baseline. Post drug consumption, a breath sample is collected and analyzed for an elevation in C13 which is indicative of detection of a urease metabolizing bacteria in the stomach. All subjects in both cohorts are compared to an FDA cleared stool antigen test using a 2x2 table and calculation the positive and negative percent agreement with a 95% two-sided confidence interval to determine clinical performance. Each subject acts as their own control by collecting a base line breath sample pre-drug treatment.
### **1. Clinical Inclusion and Exclusion Criteria for Adults**
Enrollment in the adult post-therapy study was limited to subjects who met the following inclusion criteria:
- Inclusion criteria:
a. Male or female age > 18 years of age
b. subjects who have been diagnosed with *H. pylori* and have been treated within the past 6 months
c. Naïve to *H. pylori* treatment in the past 4 weeks (including Proton Pump Inhibitors)
Patients were not permitted to enroll in the adult post-therapy study if they met any of the following exclusion criteria:
- Exclusion criteria:
a. Pregnant and/or lactating women.
b. Presence of a condition or abnormality that in the opinion of the Investigator would compromise the safety of the patient or the quality of the data.
c. Participation in other interventional trials.
d. Allergy to test substrates.
PMA P170022/S003: FDA Summary of Safety and Effectiveness Data
4 of 12
{4}
e. Antibiotics taken within 4 weeks of the testing.
f. Study subjects shall not consume the following items at least 1 hour prior to the PyloPlus UBT:
o Mouthwash
o Chewing Gum
o Carbonated Beverages
o Cigarette Smoke
o Acetone (to simulate the effect of ketone production that may result from some diets)
o Alcohol
o Food
# 2. Clinical Inclusion and Exclusion Criteria for Pediatric
Enrollment in the pediatric initial diagnosis study was limited to subjects who met the following inclusion criteria:
- Inclusion criteria:
a. Male or female age 3-17 at the time of visit.
b. Subject/Legal guardian (and subject when relevant) is willing to sign the Informed Consent/Assent Form
c. Naïve to H. pylori treatment in the past 4 weeks
Patients were not permitted to enroll in the pediatric study if they met any of the following exclusion criteria:
- Exclusion criteria:
a. Pregnant and/or lactating women.
b. Presence of a condition or abnormality that in the opinion of the Investigator would compromise the safety of the patient or the quality of the data.
c. Participation in other interventional trials.
d. Allergy to test substrates.
e. Antibiotics taken within 4 weeks of the testing.
f. Study subjects shall not consume the following items at least 1 hour prior to the PPUBT test:
o Mouthwash
o Chewing Gum
o Carbonated Beverages
o Cigarette Smoke
o Acetone (to simulate the effect of ketone production that may result from some diets)
o Alcohol
o Food
PMA P170022/S003: FDA Summary of Safety and Effectiveness Data
5 of 12
{5}
g. Children 12 years and older – to be excluded after a written notification from the sponsor is received at the site that the limit of approximately 1/3 of the sample size was achieved for this group.
# 2. Follow-up Schedule
All subjects were first identified as candidates for enrollment into the study. Results between the two test procedures were blinded to the operators. Adult subjects were enrolled who had a positive test for *H. pylori* win the past 6 months and had completed treatment at least 4 weeks prior to entering the post-therapy cohort. Pediatric subjects were enrolled for initial diagnosis of *H. pylori* infections without prior knowledge of disease status. The PyloPlus UBT System test was performed once, and the results were compared to the stool antigen results to determine performance.
# 3. Clinical Endpoints
With regards to safety, the consumption of the urea/citric acid mix, is considered low risk. The collection of breath prior to drug consumption and post drug consumption is considered non-invasive with minimal risk.
With regards to effectiveness, the non-invasive clinical test demonstrated sensitivity and specificity values of 76%, and 98%, respectively for the adult post-therapy cohort and sensitivity and specificity values of 100%, and 89.9%, respectively for the pediatric initial diagnosis cohort.
With regard to success/failure criteria, of the 80 adult subjects enrolled, 76 subjects had a stool antigen result from which an evaluable result was established and corresponding UBT result that could be used for analysis. Of the 57 pediatric subjects enrolled, 40 subjects had a stool antigen result from which an evaluable result was established and corresponding UBT result that could be used for analysis. The primary analysis of the pediatric cohort was safety with efficacy being a secondary analysis.
# **B. Accountability of PMA Cohort**
At the time of study completion, June 2023, a total of one hundred thirty-seven (137) subjects enrolled in the PMA study were available for safety analysis, with results from 84.7% (N=116) of subjects available for efficacy analysis. Table 1. Below is the study breakdown for the number of subjects in each study cohort.
One hundred sixteen (116) subjects (76 adults and 40 pediatric) had stool antigen results with a corresponding UBT result available and were determined to be evaluable for efficacy.
---PMA P170022/S003: FDA Summary of Safety and Effectiveness Data
6 of 12
{6}
**Table 1. Breakdown of adult and pediatric analysis**
| Study analysis cohorts. | Safety analysis | Efficacy analysis is a subset of the safety analysis cohort | Total |
| --- | --- | --- | --- |
| Adult safety | 80 | 76 | 80 |
| Pediatric safety | 57 | 40 | 57 |
| Total | | | 137 |
### **C. Study Population Demographics and Baseline Parameters**
The demographics of the study population are typical for an all-comers study performed in the US.
Of the one hundred thirty-seven (137) subjects tested, 80 were adults and 57 were children. All adult and pediatric subjects were evaluable for safety analysis. As noted below, the analysis of effectiveness was based on 116 evaluable subjects (76 adults and 40 pediatric subjects).
There were 80 evaluable adult subjects included in the safety analysis. There were 76 evaluable adult subjects in the efficacy analysis who completed both the UBT and stool antigen test. Of the evaluable 80 adult subjects, 67.5% (54/80) were female and 32.5% (26/80) were male. Subjects were aged 21 to 93 with 29% (23/80) over aged 65. The collected demographic information combined race and ethnicity data. Therefore only 1 subject was listed as Black/Hispanic with the remaining 79 subjects listed as Caucasian/Hispanic.
There were 57 evaluable pediatric subjects included in the safety analysis. There were 40 evaluable pediatric subjects in the efficacy study who completed both the UBT and the stool antigen test. Of the evaluable 57 subjects for safety, 42% (24/57) were male and 58% (33/57) were female. Subjects were aged from 3-17 years, with 68.4% (39/57) under the age of 12. Hispanic subjects comprised 47% (27/57) of the study population, Caucasian subjects 26.3% (15/57), African American subjects 7% (4/57) and Asian subjects 1.8% (1/57). 10 subjects (17.5%) were listed as Caucasian/Hispanic, other or unknown.
### **D. Safety and Effectiveness Results**
#### **1. Safety Results**
The analysis of safety was based on the 137 prospectively enrolled over a 7-month time span.
PMA P170022/S003: FDA Summary of Safety and Effectiveness Data
7 of 12
{7}
# **Adverse effects that occurred in the PMA clinical study:**
No adverse events were reported for the ingestion of $^{13}\text{C}$ urea/citric acid during the clinical study.
# 2. Effectiveness Results
The analysis of effectiveness was based on the 116 evaluable subjects, 76 adults and 40 pediatric subjects tested over the 7-month enrollment period. Key effectiveness outcomes are presented in Table 2 for adults and Table 3 for the pediatric cohort.
**Table 2. Adult post-therapy efficacy compared to stool antigen.**
| PyloPus UBT system | Stool Antigen Test | | |
| --- | --- | --- | --- |
| | Positive | Negative | Total |
| Positive | 19 | 1^{2} | 20 |
| Negative | 7^{1} | 49 | 56 |
| Total | 26 | 50 | 76 |
$^{1}$Seven (7) of the false negative results were followed up with an FDA reviewed UBT and found to be negative by UBT.
$^{2}$The false positive result was followed up with an FDA reviewed UBT and found to be positive by UBT.
Positive Percent Agreement 73.1% [95% CI (54.0%-86.3%)]
Negative Percent Agreement 98% [95% CI (89.7%-99.7%)]
**Table 3. Pediatric initial diagnosis secondary analysis (efficacy) compared to stool antigen.**
| PyloPus UBT system | Stool Antigen Test | | |
| --- | --- | --- | --- |
| | Positive | Negative | Total |
| Positive | 1 | 4 | 5 |
| Negative | 0 | 35 | 35 |
| Total | 1 | 39 | 40 |
Positive Percent Agreement 100%
Negative Percent Agreement 90% [95% CI (76.4-86.0%)]
# 3. Subgroup Analyses
No other subgroup analysis was conducted.
# 4. Pediatric Extrapolation
The data in this premarket application as well as leveraging existing clinical data were used to support the reasonable assurance of safety and effectiveness of the proposed device in pediatric subjects.
a) The framework set forth by the FDA Guidance entitled “Leveraging Existing Clinical Data for Extrapolation to Pediatric Uses of Medical Devices” was utilized to determine the acceptability of the sponsor’s proposal. As stated in the guidance, “Leveraging relevant available clinical data, when appropriate, may lead to more devices being granted
PMA P170022/S003: FDA Summary of Safety and Effectiveness Data
8 of 12
{8}
marketing authorization for pediatric indications, which will increase the availability of medical devices with appropriate labeling to support safe and effective device use in pediatric” subjects. Valid scientific evidence in the guidance is defined as well-controlled investigations, partially controlled studies, case histories and reports of significant human experience from which FDA can be reasonably assured that device will be safe and effective when used as intended. The scientific evidence provided by the sponsor was considered sufficient to allow pediatric extrapolation from adult data for initial and post- treatment efficacy.
b) The Sponsor additionally performed a literature search and provided the following references for FDA review. The literature search identified Leal et al., 2011, a meta-analysis which included thirty-one articles and 135 studies which overall demonstrated a sensitivity of 95% and specificity of 93.5% in children. They also included four other studies published from 1997 to 2009. These included Rowland et al., 1997 and Bazzoli et al., 2000 which both demonstrated a high sensitivity with a DOB of 3.5. Elitsur et al, 2009 demonstrated high sensitivity with a DOB of 2.4 and Machado et al., 2004 which demonstrated no relationship between the DOB value and age including in young children down to 1 year of age.
c) The sponsor fully leveraged clinical data from their device and literature to extrapolate effectiveness for the pediatric post-therapy detection of urease activity.
d) The literature search and pediatric extrapolation from adult efficacy is appropriate because of the large margin of safety associated with the drug, the risk profile of the diagnosis and the over 10 years experience we have with this class of UBT using the same drug formulation with the pediatric population. Therefore, the provision of only safety data from the clinical study is acceptable. Based on the literature search, maintaining the same DOB of 3 in adult and pediatric populations is also acceptable.
# XI. FINANCIAL DISCLOSURE
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included five (5) investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data.
PMA P170022/S003: FDA Summary of Safety and Effectiveness Data
9 of 12
{9}
## **XII. SUMMARY OF SUPPLEMENTAL CLINICAL INFORMATION**
N/A
## **XIII. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION**
Device did not go to Panel.
## **XIV. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES**
### **A. Effectiveness Conclusions**
Performance characteristics of PyloPlus UBT System for *H. pylori* were similar across age groups, race, gender and study sites. The primary endpoint analysis was conducted to determine the sensitivity and specificity of the PyloPlus UBT kit for *H. pylori* to an FDA cleared stool antigen test for the adult post-therapy cohort with results for the 76 subjects who abstained from PPIs. When compared to the stool antigen test, the observed sensitivity and specificity values were 76%, and 98%, respectively. When subjects were compared to an FDA cleared breath test, the sensitivity and specificity were 100% and 100%, respectively.
Performance characteristics of PyloPlus UBT System for *H. pylori* were similar across age groups, race, gender, and study sites. The secondary endpoint analysis of efficacy for pediatric subjects 3-17 years of age was conducted to determine the sensitivity and specificity of the PyloPlus UBT kit for *H. pylori* to an FDA cleared stool antigen test. Results were available for 40 subjects; when compared to the stool antigen test, the observed sensitivity (n=1) and specificity (n=39) values were 100%, and 89.9%, respectively.
### **B. Safety Conclusions**
The adverse effects (AE) of the device are based on data collected in clinical studies to support PMA approval as described above. No adverse effects were reported during this clinical trial.
### **C. Benefit-Risk Determination**
The probable benefits of the device are also based on data collected in the clinical study conducted to support PMA approval, as described above. The benefits include providing a means of rapid, non-invasive detection of *H. pylori* in the stomach for adults with active peptic ulcer disease (PUD), certain types of gastric lymphoma or gastric cancer, or other conditions where suspicion of *H. pylori* infection warrants testing, and treatment if present. The device, along with others like it that utilize the
---PMA P170022/S003: FDA Summary of Safety and Effectiveness Data
10 of 12
{10}
same mechanism for detection of *H. pylori*, provide an alternative to other means of *H. pylori* diagnosis that may require an invasive procedure, or may be unable to detect active, current infection.
The probable risks of the device are also based on data collected in a clinical study conducted to support PMA approval, as described above. The risks associated with the device, when used for pre-treatment diagnosis of *H. pylori*, are those related to the risk of false test results, failure to correctly interpret the test results, and failure to correctly operate the instrument. A false positive *H. pylori* diagnostic test risks overtreatment with antimicrobials used in treatment regimens for *H. pylori*. A false negative result may result in delays in treatment of *H. pylori* infection, which could result in worsening of certain associated gastrointestinal illnesses, such as peptic ulcer disease. Adverse events related to use of the device were not observed during the clinical study and are expected to be rare.
No additional factors to be considered in determining probable risks and benefits for the PyloPlus UBT System are included.
#### 1. Patient Perspectives
This submission did not include specific information on patient perspectives for this device, or the information did not serve as part of the basis of the decision to approve or deny the PMA for this device.
In conclusion, given the available information above, the data supports the intended use of PyloPlus UBT system in the qualitative detection of urease associated with *H. pylori* in the human stomach and for use as an aid in the initial diagnosis of *H. pylori* infection in adults and children 3-17 years old aid in the post therapy detection of urease associated with *H. pylori* infection in adult subjects and children 17 years old. The probable benefits outweigh the probable risks for this device.
### **D. Overall Conclusions**
The data in this application support the reasonable assurance of safety and effectiveness of this device and drug ($^{13}$C urea/citric acid) when used in accordance with the indications for use. The submitted clinical studies have shown that the PyloPlus UBT System, when used for the detection of *H. pylori*, has a similar performance as a stool antigen test, to aid in the post-therapy detection of urease associated with *H. pylori* infection in adult subjects. Furthermore, the data and evidence provided indicate results do not differ in the pediatric population as the $^{13}$C-enriched urea dosage amount and DOB cut-off point would be the same. ARJ Medical, Inc. has effectively used the CDRH guidance titled *Leveraging Existing*
---PMA P170022/S003: FDA Summary of Safety and Effectiveness Data
11 of 12
{11}
*Clinical Data for Extrapolation to Pediatric Uses of Medical Devices* to support the claim that the PyloPlus UBT analyzer is safe, effective and provides results as intended when used in children 3-17 years of age.
# **XV. CDRH DECISION**
CDRH issued an approval order on January 11, 2024.
The applicant’s manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820).
# **XVI. APPROVAL SPECIFICATIONS**
Directions for use: See device labeling.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling.
Post-approval Requirements and Restrictions: N/A
# **XVII. REFERENCES**
N/A
---PMA P170022/S003: FDA Summary of Safety and Effectiveness Data
12 of 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.