P170018 · Physio-Control, Inc. · MKJ · Dec 21, 2018 · Cardiovascular
Device Facts
Record ID
P170018
Device Name
LIFEPAK® CR2 Defibrillator
Applicant
Physio-Control, Inc.
Product Code
MKJ · Cardiovascular
Decision Date
Dec 21, 2018
Decision
APPR
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
P170018 · Dec 21, 2018
LIFEPAK® CR2 Defibrillator
Physio-Control, Inc.
Commercial distribution data (OUS); Post-market surveillance data; Device return and malfunction reports
The FDA utilized evidence from ongoing commercial distribution in 33 countries to support the benefit-risk profile. Post-approval conditions require the sponsor to collect real-world performance data, including device returns, failure analysis, and a formal OUS post-market evaluation study of the cprINSIGHT algorithm in BLS settings.
OUS Post Market Evaluation Study; Post-market observational evaluation; Study Period: Ongoing
Basic Life Support (BLS) use environment patients
Not applicable for this study
Performance of cprINSIGHT Analysis Technology, identification of unanticipated use issues, and general device usability.
Indications for Use
The LIFEPAK® CR2 Defibrillator is indicated for use on patients 1 year of age and older in cardiopulmonary arrest. The patient must be unresponsive (unconscious), not breathing normally, and showing no signs of circulation (for example, no pulse, no coughing, or no movement). The cprCOACH™ Feedback Technology in the LIFEPAK CR2 defibrillator is indicated for use on cardiopulmonary arrest patients and provides CPR guidance in accordance with the AHA Guidelines for patients 1 year of age and older. The LIFEPAK® CR2 Defibrillator is intended for use by personnel who have been trained in its operation. Users should have received training in basic life support/AED, advanced life support, or a physician-authorized emergency medical response training program. The LIFEPAK® CR2 Defibrillator is indicated to be used with the QUIK-STEP Pacing/ECG Defibrillation Electrodes and the LIFEPAK CR2 Lithium Battery.
Device Story
LIFEPAK CR2 is a battery-powered automated external defibrillator (AED) for treating sudden cardiac arrest. It inputs ECG signals via QUIK-STEP electrodes; processes rhythm using Shock Advisory System (SAS) and cprINSIGHT Analysis Technology; and outputs voice/visual prompts and biphasic truncated exponential (BTE) electrical shocks. Used in public access settings by trained responders; device guides users through CPR and shock delivery. Semiautomatic models require manual shock button activation; fully automatic models deliver shocks autonomously. Output helps restore normal heart rhythm, potentially saving lives. Features include child mode, cprCOACH feedback, and wireless data transmission.
Clinical Evidence
No new prospective clinical trial conducted. Evidence based on literature and commercial use. Studies compared BTE vs. monophasic waveforms (Higgins et al., 2000; Van Alem et al., 2003), demonstrating superior first-shock efficacy for BTE. Animal study (Killingsworth et al.) in piglets supported pediatric safety/efficacy. Human factors testing (n=94) confirmed safe/effective use by intended users. SAS and cprINSIGHT performance validated against AHA recommendations using large datasets of adult and pediatric ECG segments, showing high sensitivity and specificity.
Indicated for cardiopulmonary arrest patients 1 year of age and older who are unresponsive, not breathing normally, and show no signs of circulation. Contraindicated for patients who are conscious or responsive.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
I. GENERAL INFORMATION
| Device Generic Name: | Automated External Defibrillator |
| --- | --- |
| Device Trade Name: | LIFEPAK® CR2 Defibrillator |
| Device Product Code: | MKJ |
| Applicant’s Name and Address: | Physio-Control, Inc.
11811 Willows Road NE
Redmond, WA 98052 USA |
| Date(s) of Panel Recommendation: | N/A |
| Premarketing Approval Application (PMA) Number: | P170018 |
| Date of FDA Notice of Approval: | December 21, 2018 |
II. INDICATIONS FOR USE
The LIFEPAK® CR2 Defibrillator is indicated for use on patients 1 year of age and older in cardiopulmonary arrest. The patient must be unresponsive (unconscious), not breathing normally, and showing no signs of circulation (for example, no pulse, no coughing, or no movement).
The cprCOACH™ Feedback Technology in the LIFEPAK CR2 defibrillator is indicated for use on cardiopulmonary arrest patients and provides CPR guidance in accordance with the AHA Guidelines for patients 1 year of age and older.
The LIFEPAK® CR2 Defibrillator is intended for use by personnel who have been trained in its operation. Users should have received training in basic life support/AED, advanced life support, or a physician-authorized emergency medical response training program.
The LIFEPAK® CR2 Defibrillator is indicated to be used with the QUIK-STEP Pacing/ECG Defibrillation Electrodes and the LIFEPAK CR2 Lithium Battery.
III. CONTRAINDICATIONS
The LIFEPAK® CR2 Defibrillator is not indicated for patients who are conscious or responsive.
IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the LIFEPAK CR2 labeling.
V. DEVICE DESCRIPTION
The LIFEPAK CR2 Defibrillator is a small, lightweight, battery powered automated external defibrillator (AED) intended for use by minimally trained responders to treat
PMA P170018: FDA Summary of Safety and Effectiveness Data
{1}
victims of sudden cardiac arrest. The LIFEPAK CR2 Defibrillator incorporates an easy to use interface for the responder to activate the device, apply the defibrillation electrodes, deliver a shock (semi-automatic model) or the device will automatically deliver a shock (automatic model), and perform CPR. Using voice prompts, graphics, audible tones and buttons, the responder is guided through device use.
Major accessories used with the LIFEPAK CR2 Defibrillator are the following:
- The LIFEPAK CR2 Defibrillator is shipped with disposable, single use QUIK-STEP Pacing/ECG Defibrillation Electrodes, which are pre-connected electrodes that can be used to treat both adults and children.
- The LIFEPAK CR2 Defibrillator is powered by a non-rechargeable lithium manganese dioxide battery with a typical capacity of 166 shocks at 200 joules or 800 minutes of operation when powered on.
- **Principle of Operation:**
Sudden cardiac arrest is usually caused by a malfunction in the heart's electrical system, called ventricular fibrillation (VF). This critical condition prevents the heart from pumping blood throughout the body. VF can cause death within seconds.
Defibrillation is a relatively simple procedure that involves placing defibrillation electrodes on a victim's exposed chest and delivering an electrical shock to the heart. The externally-delivered shock often restores the heart's electrical system to a normal rhythm. Combined with cardiopulmonary resuscitation (CPR), defibrillation provides the most effective care for victims in cardiac arrest.
In the LIFEPAK CR2 there are two (2) systems which are used for electrocardiogram (ECG) analysis. These systems recommend whether or not a patient should be given a defibrillation shock. These systems make it possible for individuals not trained to interpret ECG rhythms to provide potentially-lifesaving therapy to victims of ventricular fibrillation or pulseless ventricular tachycardia (VT).
- **Shock Advisory System™ (SAS):** The SAS algorithm is the same algorithm used in other PMA approved LIFEPAK defibrillators (P160012 and P160026). SAS is used to analyze the ECG rhythm during the first rhythm analysis after the defibrillation electrodes have been placed on the patient when CPR is not being performed. It is also used during subsequent rhythm analyses when the user has been instructed to stop CPR.
- **cprINSIGHT™ Analysis Technology:** The cprINSIGHT Analysis Technology is an additional analysis system that enables the defibrillator to analyze the patient's heart rhythm while CPR is being performed. This reduces pauses in chest compressions which are normally required for rhythm analysis and thus the ongoing CPR helps maintain circulation of the blood. When the ECG rhythm is determined to be nonshockable, the pause for analysis can be eliminated altogether, allowing for continuous CPR. When the rhythm is determined to be shockable, the necessary pause time in chest compressions is shortened to the
PMA P170018: FDA Summary of Safety and Effectiveness Data
Page 2
{2}
time needed for the rescuer to stand clear and deliver the shock. The cprINSIGHT Analysis Technology can be set to be on or off. As noted previously, the LIFEPAK CR2 Defibrillator always uses the SAS algorithm for the first rhythm analysis in each patient.
Further descriptive detail regarding the Shock Advisory System and cprINSIGHT Analysis Technology can be found within the LIFEPAK CR2 Defibrillator Operating Instructions.
- Device Features: A summary of the main features of the LIFEPAK CR2 Defibrillator are listed below. Details regarding each of these features can be found in the LIFEPAK CR2 Defibrillator Operating Instructions (labeling).
- Readiness for use indicator flashes every 6 seconds to indicate device status
- Pre-connected QUIK-STEP electrode tray
- cprCOACH Feedback Technology/compression detection provides CPR coaching (feedback) to user regarding compressions, when appropriate
- ClearVoice Technology - for delivery of voice prompts to the device user
- cprINSIGHT™ Analysis Technology
- Child Mode delivers lower energy levels that are appropriate for young children
- Dual Language capability
- Semiautomatic and fully automatic device models
- Biphasic truncated exponential (BTE) defibrillation waveform
- Shock Advisory System (SAS)
- Multiple configurable defibrillation shock energy levels from 150 to 360 joules for adult mode operation
- Multiple configurable defibrillation shock energy levels from 35 to 90 joules for child mode operation
- Simple User Interface
- Automatic Self-Tests - to assess device readiness for use
- Data Management - for capture of ECG/event data
- Wireless data transmission
For visual reference, diagrams of the LIFEPAK CR2 defibrillator are provided below in Figure 1 (exterior views) and Figure 2 (interior views). Additional diagrams and pictures of the device can be found in the LIFEPAK CR2 Operating Instructions.
- Device User Interface: The LIFEPAK CR2 is designed for public access ease of use by reducing the number of steps traditionally required for defibrillation. As such, the device has a simple user interface. For example: The LIFEPAK CR2 is shipped ready to use with preprogrammed biphasic escalating energy and operating settings. The defibrillator includes an integral lid that protects the electrodes and the same button that opens the lid, turns the device power on (see Figure 1). The defibrillator voice prompts and audible tones guide the device user, including when to apply electrodes, and provides instructions for standing clear or pressing the flashing shock
PMA P170018: FDA Summary of Safety and Effectiveness Data
Page 3
{3}
button, depending upon the version of the device being used (fully automatic or semiautomatic).
- Device Versions: The LIFEPAK CR2 is offered in two (2) models: fully automatic or semiautomatic.
- Fully automatic AED: The fully automatic model of the LIFEPAK CR2 only requires the responder to assess the patient, open the defibrillator, apply the electrodes to the victim's chest and follow the voice prompts. The defibrillator will automatically analyze the victim's heart rhythm, charge, and if a shockable rhythm is detected, the defibrillator will alert the responder that a defibrillation pulse (shock) will be delivered and to stand clear of the patient, and then deliver the shock without any responder assistance. The defibrillator delivers shocks through the defibrillation electrodes on the victim's chest.
- Semiautomatic AED: The semiautomatic model of the LIFEPAK CR2 evaluates the heart rhythm and if a shockable rhythm is detected, the voice prompt instructs the responder to press the flashing "shock" button on the defibrillator to deliver the shock through the defibrillation electrodes on the victim's chest.

Figure 1: LIFEPAK CR2 defibrillator - top and bottom exterior views

| Items in Figure 1 | Feature | Description |
| --- | --- | --- |
| 1 | Readiness indicator | The green LED flashes every 6 seconds if the defibrillator is ready for use. The indicator is steady on if the defibrillator is turned on, and steady off if the defibrillator needs attention. |
| 2 | Latch | To open the defibrillator, insert a finger in the recessed area and pull up. |
| 3 | USB port | The USB connection is used to connect the defibrillator to a computer, to establish communication with LIFELINK™ Central AED Program Manager or LIFENET® System. |
| 4 | Battery compartment | Insert the LIFEPAK CR2 lithium battery into the battery compartment until it clicks into place. |
| 5 | Warning symbol | See General Dangers and Warnings in the LIFEPAK CR2 Operating Instructions. |
PMA P170018: FDA Summary of Safety and Effectiveness Data
{4}

Figure 2: LIFEPAK CR2 defibrillator - interior view
| Item in Figure 2 | Feature | Description |
| --- | --- | --- |
| 1 | Language button | If the defibrillator has two (2) languages installed, press the LANGUAGE button to switch between languages. |
| 2 | On/Off button | The defibrillator turns on automatically when the lid is opened. Press and hold the ON/OFF button for 3 seconds to turn the defibrillator off. Press again to turn the defibrillator back on. The green LED is illuminated when the defibrillator is on. |
| 3 | Child Mode button | Press the CHILD MODE button to switch between Adult mode and Child mode. When Child mode is selected, the green LED above the CHILD MODE button is illuminated and a voice prompt announces "Child mode." |
| 4 | Shock button/indicator | On fully automatic defibrillators, the SHOCK indicator flashes when the defibrillator is preparing to deliver a shock. On semiautomatic defibrillators, press the flashing SHOCK button to deliver a shock to the patient. |
| 5 | Red handle | During a cardiac arrest event, the red handle is pulled to reveal the electrodes. |
| 6 | Speaker | Provides audio voice prompts and tones. |
| 7 | Noise sensor | The noise sensor monitors the noise level of the surrounding area while the defibrillator is in use. |
# VI. ALTERNATIVE PRACTICES AND PROCEDURES
Defibrillation is the only currently available treatment for termination of VF or pulseless VT. Public access defibrillation is designed to provide potentially lifesaving treatment prior to the arrival of emergency personnel.
PMA P170018: FDA Summary of Safety and Effectiveness Data
{5}
PMA P170018: FDA Summary of Safety and Effectiveness Data
Page 6
# VII. MARKETING HISTORY
The LIFEPAK CR2, QUIK-STEP Pacing/ECG Defibrillation Electrodes, and the LIFEPAK CR2 Lithium Battery are new products and have not been marketed in the US.
The LIFEPAK CR2 and the accessories have been in commercial distribution in non-US countries since October 2016. As of October 2018, the LIFEPAK CR2 is in commercial distribution in the following 33 non-US countries:
- Albania
- Australia
- Austria
- Bahrain
- Canada
- Denmark
- Finland
- France
- Germany
- Hungary
- Iran
- Ireland
- Israel
- Italy
- Japan
- Kuwait
- Liechtenstein
- Luxembourg
- Malaysia
- Malta
- Netherlands
- New Zealand
- Norway
- Panama
- Peru
- Poland
- Singapore
- Spain
- Sweden
- Switzerland
- Turkey
- United Arab Emirates
- UK
The LIFEPAK CR2 and the accessories have not been withdrawn from the market in the United States or any foreign country.
# VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Below is a list of the potential adverse effects (e.g., complications) associated with the use of the LIFEPAK CR2.
- Failure to identify shockable arrhythmia;
- Failure to deliver a defibrillation shock in the presence of VF or pulseless VT, which may result in death or permanent injury;
- Inappropriate energy delivery which could cause failed defibrillation or post-shock dysfunction;
- Myocardial damage;
- Fire hazard in the presence of high oxygen concentration or flammable anesthetic agents;
- Electromagnetic interference (EMI) from the defibrillator impacting other devices especially during charge and energy transfers;
- Incorrectly shocking a pulse sustaining rhythm and inducing VF or cardiac arrest;
- Bystander shock from patient contact during defibrillation shock;
{6}
- Interaction with pacemakers;
- Skin burns around the electrode placement area;
- Allergic dermatitis due to sensitivity to materials used in electrode construction; and
- Minor skin rash.
## IX. SUMMARY OF PRECLINICAL STUDIES
The LIFEPAK CR2 underwent bench and animal testing as well as software verification and validation appropriate for major level of concern devices. This non-clinical information included:
## A. Bench Testing
Table 1 summarizes the major bench testing conducted to demonstrate performance of the LIFEPAK CR2, including conformance with applicable performance standards.
Table 1. Major Bench Testing
| Test Title | Results |
| --- | --- |
| Electrical Safety (IEC 60601-1:2005 Edition 3 and IEC 60601-2-4:2010 Edition 3) | Pass |
| Electromagnetic Compatibility (IEC 60601-1-2:2014 Edition 4) with Home Use environment requirements | Pass |
| Software Verification/Validation | Pass |
| Environmental Testing | Pass |
| Shock Advisory System Performance Testing | Pass |
| cprINSIGHT Analysis Technology Performance Testing | Pass |
| Motion Detection Testing | Pass |
| System Design Verification | Pass |
| Hardware Design Verification | Pass |
| Functional Testing | Pass |
| Battery Testing | Pass |
| Energy Accuracy and Waveform Testing | Pass |
| Time to Shock Ready Testing | Pass |
| Mechanical (Vibration, Drop, etc.) Testing | Pass |
| Cleaning Tests | Pass |
| Packaging Tests | Pass |
## B. Shelf Life Testing
Shelf life testing was conducted to demonstrate a 4-year shelf life for the QUIK-STEP Electrodes and 4-year standby life for the Lithium manganese dioxide battery used with the LIFEPAK CR2.
## C. Software
The software for the LIFEPAK CR2 was verified/validated and documented as a Major Level of Concern device according to the FDA guidance document "Guidance
PMA P170018: FDA Summary of Safety and Effectiveness Data
Page 7
{7}
for the Content of Premarket Submissions for Software Contained in Medical Devices." The documentation included level of concern, software description, device hazard analysis, software requirements specification, software architecture diagrams, software design specifications, requirements traceability matrix, software development environment description, verification and validation documentation, revision level history, report of unresolved anomalies, discussion of tools to detect run-time errors, and cybersecurity documentation. Unit, integration, and system-level testing was documented and demonstrated that the software for the LIFEPAK CR2 performs as intended.
# D. Defibrillation Waveform
The LIFEPAK CR2 delivers a Biphasic Truncated Exponential waveform, with voltage and duration compensation for patient impedance.
# - Shock Advisory Algorithms
The results of performance testing of the SAS algorithm and the cprINSIGHT algorithm are shown in Tables 2-5. This data has been reported per the American Heart Association (AHA) Scientific Statement: "Automatic External Defibrillators for Public Access Defibrillation: Recommendations for Specifying and Reporting Arrhythmia Algorithm Performance, Incorporating New Waveforms, and Enhancing Safety," Circulation 1997; 95:1677-1682. $^{1}$
Table 2. AHA Recommendations and SAS Performance for Adult Patients
| Rhythm Category | Sample Size (AHA Recommended) | Sample Size (Tested) | Performance Goal (AHA Recommended) | Test Result | One-Sided 90% Lower Confidence Limit |
| --- | --- | --- | --- | --- | --- |
| | Shockable | | | | |
| Coarse VF | 200 | 206 | >90% sensitivity | 99.0% sensitivity | 97.4% |
| Rapid VT, pulseless | 50 | 65 | >75% sensitivity | 84.6% sensitivity | 77.3% |
| | Nonshockable | | | | |
| NSR | 100 | 509 | >99% specificity | 100.0% specificity | 99.5% |
| Other QRS | 30 | 749 | >95% specificity | 98.9% specificity | 98.3% |
| Asystole | 100 | 124 | >95% specificity | 100.0% specificity | 98.2% |
| | Intermediate | | | | |
| Fine VF | 25 | 32 | Report only | 53.1% shocked | 40.4% |
| Other VT | 25 | 27 | Report only | 29.6% shocked | 18.1% |
PMA P170018: FDA Summary of Safety and Effectiveness Data
{8}
Table 3. AHA Recommendations and SAS Performance for Pediatric Patients
| Rhythm Category | Sample Size (Tested) | Performance Goal (AHA Recommended) | Test Result | One-Sided 90% Lower Confidence Limit |
| --- | --- | --- | --- | --- |
| Shockable | | | | |
| Coarse VF | 63 | >90% sensitivity | 98.4% sensitivity | 94.0% |
| Nonshockable | | | | |
| NSR | 69 | >99% specificity | 100.0% specificity | 96.7% |
| Other QRS | 507 | >95% specificity | 100.0% specificity | 99.5% |
| Asystole | 60 | >95% specificity | 100.0% specificity | 96.2% |
| Intermediate | | | | |
| Fine VF | 1 | Report only | 100.0% shocked | 10.0% |
Table 4. AHA Recommendations and cprINSIGHT Analysis Performance for Adult Patients, results of testing on rhythm segments during CPR periods
| Rhythm category | Sample Size | Decided Without a Pause | Performance Goal (AHA Recommended) | Test Result | One-sided 90% Lower Confidence Limit |
| --- | --- | --- | --- | --- | --- |
| Shockable | | | | | |
| Coarse VF | 1282 | 80.6% | >90% sensitivity | 96.4% sensitivity | 95.6% |
| Rapid VT, pulseless | 52 | 61.5% | >75% sensitivity | 65.6% sensitivity | 52.8% |
| Nonshockable | | | | | |
| NSR | 72 | 93.1% | >99% specificity | 100.0% specificity | 96.6% |
| Other QRS | 2527 | 79.3% | >95% specificity | 97.8% specificity | 97.3% |
| Asystole | 1790 | 59.9% | >95% specificity | 97.0% specificity | 96.2% |
| Intermediate | | | | | |
| Fine VF | 113 | 39.8% | Report only | 48.9% shocked | |
| Other VT | 49 | 67.3% | Report only | 33.3% shocked | |
Table 5. AHA Recommendations and cprINSIGHT Analysis Performance for Pediatric Patients, results of testing on rhythm segments during CPR periods
| Rhythm category | Sample Size | Decided Without a Pause | Performance Goal (AHA Recommended) | Test Result | One-sided 90% Lower Confidence Limit |
| --- | --- | --- | --- | --- | --- |
| Shockable | | | | | |
| Coarse VF | 38 | 84.2% | >90% sensitivity | 100.0% sensitivity | 93.1% |
| Rapid VT, pulseless | 0 | -- | -- | -- | -- |
| Nonshockable | | | | | |
| NSR | 13 | 100% | >99% specificity | 100.0% specificity | 83.8% |
PMA P170018: FDA Summary of Safety and Effectiveness Data
{9}
| Rhythm category | Sample Size | Decided Without a Pause | Performance Goal (AHA Recommended) | Test Result | One-sided 90% Lower Confidence Limit |
| --- | --- | --- | --- | --- | --- |
| Other QRS | 472 | 72.0% | >95% specificity | 99.7% specificity | 98.9% |
| Asystole | 176 | 72.2% | >95% specificity | 97.6% specificity | 94.8% |
| Intermediate | | | | | |
| Fine VF | 0 | -- | -- | -- | -- |
| Other VT | 1 | 100% | Report only | 0 shocked | |
# Animal Study
An experimental study by C. Killingsworth titled "Defibrillation threshold and cardiac responses using an external biphasic defibrillator with pediatric and adult adhesive patches in pediatric-sized piglets" provides evidence on safety and effectiveness of the therapy used in child mode.[2]
The LIFEPAK CR2 delivers the same waveform when treating adults and children, although the amplitude of the pulses reflects the lower energy settings. The Killingsworth article describes testing with shocks generated by the LIFEPAK 12 defibrillator/monitor which delivers shocks with a waveform similar to the shocks generated by the LIFEPAK CR2. In the study, defibrillation thresholds were measured in 10 pigs weighing from 3.8 to $20.1\mathrm{kg}$ (see Figure 4 below). For adult-sized electrode pads, the defibrillation threshold was $2.1 \pm 0.65$ $\mathrm{J / kg}$ ; for pediatric-sized electrode pads, it was $2.4 \pm 0.81$ $\mathrm{J / kg}$ ; the difference was not statistically significant.

Figure 4: Defibrillation Thresholds of 10 Pigs
Figure 4 is from Killingsworth article showing defibrillation threshold expressed in energy (Joules) vs. body weight with pediatric-size electrodes $(\triangle)$ and adult-size electrodes $(\blacksquare)$ . Also shown are regression lines for pediatric (dashed line) and adult (solid line) patches.
The same study also investigated the effect of much larger shocks, with the same waveform, on heart function. Delivering shocks all the way up to the maximum
PMA P170018: FDA Summary of Safety and Effectiveness Data
{10}
adult dose of 360 J (without using an attenuator) to these small pigs produced only brief-duration, transient ST-segment, rhythm, and hemodynamic changes. The authors concluded, “Despite the delivery of very large cumulative doses and multiple large individual shocks during the course of each experiment, myocardial function in these small animals was remarkably resilient.” This study demonstrates that, with the biphasic waveform employed by the LIFEPAK 12 defibrillator/monitor and the LIFEPAK CR2 defibrillator, there is a large safety factor between the dose needed to defibrillate, and the dose that would cause harm.
The animal study discussed above was leveraged to support the reasonable assurance of safety and effectiveness of the proposed device in the pediatric subpopulation of children 1 to 8 years old for the LIFEPAK CR2 defibrillator.
# X. SUMMARY OF CLINICAL STUDIES
A new, prospective clinical trial was not conducted using the LIFEPAK CR2 Defibrillator. However, publications of several prior clinical studies $^{3,4}$ , conducted by Physio-Control were considered relevant in demonstrating the safety and effectiveness of the LIFEPAK CR2 were provided to support premarket approval. In particular, the identical ECG analysis Shock Advisory System and biphasic truncated exponential (BTE) (ADAPTIV biphasic waveform) used in the LIFEPAK 500 defibrillators deployed in the following clinical studies, is also used in the LIFEPAK CR2.
The studies summarized in the PMA are listed in Table 7. Further study details can be found within the summaries of the publications noted below; full citations of the publications are provided in section XV of this SSED.
Table 6. List of Supporting Clinical Studies
| Clinical Study # | Clinical Study | Related Publication |
| --- | --- | --- |
| Study #1 | Comparison of monophasic vs. biphasic waveforms: in-hospital trial | Higgins, et al., 20003 |
| Study #2 | Comparison of monophasic vs. biphasic waveforms: out-of-hospital trial | Van Alem, et al., 20034 |
# A. Clinical Study #1 - Monophasic vs. biphasic waveforms: in-hospital trial
This prospective, double-blinded, randomized clinical trial compared the first shock efficacies of Physio-Control ADAPTIV 200J BTE waveform, 130 J BTE and 200 J monophasic damped sine shocks (MDS) shocks in the electrophysiology lab, delivered from modified LIFEPAK 7 defibrillators (an early defibrillator cleared under $510(\mathrm{k})$ K810154, but modified to have the same BTE waveform as the LIFEPAK CR Plus and the LIFEPAK EXPRESS).
Methods: VF was induced in 154 patients. After $19 \pm 10$ seconds of VF, a randomized transthoracic shock was administered. Mean first shock success rates for the three groups were compared.
PMA P170018: FDA Summary of Safety and Effectiveness Data
{11}
Results: First shock VF termination rates were 61/68 (90%) for the 200 J monophasic, 39/39 (100%) for the 200 J biphasic, and 39/47 (83%) for 130 J biphasic shocks.
Conclusion: The 200 J biphasic shocks were superior in first-shock effectiveness to both 200 J MDS shocks and 130 J BTE shocks. There were no significant differences in hemodynamic parameters between the three (3) groups after successful shocks. The 200 J biphasic shocks were more effective than monophasic and the 130 J BTE shocks and may allow earlier termination of VF in cardiac arrest patients.
## B. Clinical Study #2 - Monophasic vs. biphasic waveforms: out-of-hospital trial
In this publication by Van Alem, et al., the authors noted "Evidence suggests that biphasic waveforms are more effective than monophasic waveforms for defibrillation in out-of-hospital cardiac arrest (OHCA), yet their performance has only been compared in un-blinded studies." The authors subsequently conducted and reported on a clinical trial comparing the effectiveness of the LIFEPAK 500 defibrillation waveform (monophasic versus biphasic). Specifically, the success of BTE and MDS shocks for defibrillation were compared in a prospective, randomized, double blind clinical trial of out-of-hospital (OOH) cardiac arrest patients.
Note: The identical ECG analysis Shock Advisory System and BTE (ADAPTIV biphasic waveform) used in the LIFEPAK 500 AED is also used in the LIFEPAK CR Plus and LIFEPAK EXPRESS AEDs.
Methods: First responders were equipped with either a Physio-Control LIFEPAK 500 MDS or BTE (ADAPTIV biphasic waveform) AED in a random fashion. Patients in VF received BTE or MDS first shocks of 200 J. The ECG was recorded for subsequent analysis continuously. The success of the first shock as a primary endpoint was removal of VF and required a return of an organized rhythm for at least two (2) QRS complexes, with an interval of <5 seconds, within 1 minute after the first shock. The secondary endpoint was termination of VF at 5 seconds.
Results: VF was the initial recorded rhythm in 120 patients in OHCA, 51 patients received BTE and 69 received MDS shocks. The median time from collapse to first shock was 9 minutes for the monophasic shock and 11 minutes for the BTE. The success rate of 200 J first shocks was significantly higher for BTE than for MDS shocks, 35/51 (69%) and 31/69 (45%), P=0.01. Termination of VF at 5 seconds after the first shock was 91% for the monophasic shock and 98% for BTE waveform. Return of spontaneous circulation was 61% for the Physio-Control defibrillation shock.
In a logistic regression model, the odds ratio of success for a BTE shock was 4.01 (95% CI 1.01-10.0), adjusted for baseline cardiopulmonary resuscitation, VF-amplitude, and time between collapse and first shock. No difference was found with respect to the secondary endpoint, termination of VF at 5 seconds (RR 1.07 95% CI: 0.99-1.11) and with respect to survival to hospital discharge (RR 0.73 95% CI: 0.31-1.70).
PMA P170018: FDA Summary of Safety and Effectiveness Data
Page 12
{12}
Conclusion: The authors concluded that BTE-waveform AEDs provide significantly higher rates of successful defibrillation with return of an organized rhythm in OHCA than MDS waveform AEDs.
## C. Human Factors
Human Factors testing was conducted to demonstrate that the LIFEPAK CR2 Defibrillator (semi-automatic and automatic mode) can be used safely and effectively for its indication for use by intended users in its intended use environment. Testing was carried out primarily through simulated use test by having representative users (n=94) perform tasks associated with primary operating functions of the device. The results of testing demonstrated that the LIFEPAK CR2 user interface supports safe and effective use of the device.
## D. Pediatric Extrapolation
In this premarket application, the applicant submitted the cprINSIGHT algorithm to reduce the pauses between chest compressions in adults and pediatric patients. Because the cprINSIGHT algorithm uses similar measures as the SAS algorithm, it is expected that the cprINSIGHT algorithm will also perform similarly well on both adult and pediatric ECGs. The SAS algorithm performs well for both adult and pediatric ECGs because the ECG measures employed avoid characteristics of the ECG that vary between adults and children.
## E. Financial Disclosure
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The clinical studies included six (6) investigators of which none were full-time or part-time employees of the sponsor and two (2) investigators had disclosable financial interests/arrangements as defined in 21 CFR 54.2(a), (b), (c) and (f) and described below:
- Compensation to the investigator for conducting the study where the value could be influenced by the outcome of the study: 1 investigator
- Significant payment of other sorts: 1 investigator
- Proprietary interest in the product tested held by the investigator: none
- Significant equity interest held by investigator in sponsor of covered study: none
The applicant has adequately disclosed the financial interest/arrangements with clinical investigators. Statistical analyses were conducted by FDA to determine whether the financial interests/arrangements had any impact on the clinical study outcome. The information provided does not raise any questions about the reliability of the data.
PMA P170018: FDA Summary of Safety and Effectiveness Data
Page 13
{13}
PMA P170018: FDA Summary of Safety and Effectiveness Data
Page 14
# XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION
In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Cardiovascular Device Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel on January 25, 2011 as part of the 515(i) process. The majority of the panel recommended that AEDs be regulated as Class III PMAs to have better oversight of device manufacturing and post-market performance.
# XII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES
A. Effectiveness Conclusions: The preclinical and clinical information (including commercial use and publications) provided in this PMA did not identify any effectiveness concerns associated with use of the LIFEPAK CR2 (including the battery and QUIK-STEP electrode accessories) for the defibrillators' indications for use.
B. Safety Conclusions: The preclinical and clinical information (including commercial use and publications) provided in this PMA did not identify any safety concerns associated with use of the LIFEPAK CR2 (including the battery and QUIK-STEP electrode accessories) for the defibrillators' indications for use.
C. Benefit-Risk Conclusions: The LIFEPAK CR2 and accessories (battery and QUIK-STEP electrodes), are new devices not previously cleared by FDA. However, they are in commercial distribution outside the US.
Overall, the information in this PMA demonstrates the safe and effective use of the LIFEPAK CR2 (including the battery and QUIK-STEP electrode accessories) in ongoing, commercial use, for the defibrillators' indications for use and suggests risks in the use of LIFEPAK CR2 defibrillator (and accessories) are outweighed by the benefit of potentially resuscitating cardiac arrest victims.
## Patient Perspectives
This application did not include specific information on patient perspectives for the device.
D. In conclusion, the information and data in this application provides reasonable assurance that the LIFEPAK CR2 AED and accessories (battery and QUIK-STEP electrodes) are safe and effective when used in accordance with the indications for use.
# XIII. CDRH DECISION
CDRH issued an approval order on December 21, 2018. FDA has developed unique conditions of approval to pursue real world information and in response to panel comments from the 515i Panel discussed in Section XI above. The final conditions of approval cited in the approval order are described below.
{14}
The applicant will provide the following non-clinical information as part of the annual report, which may be followed by a PMA supplement, where applicable:
1. The number of devices returned to the applicant for cause from domestic sources, with a breakdown into:
a. Those returned for normal end-of-life; and
b. Those returned with any alleged failures or malfunctions, including a summary of root causes and the frequency of occurrence for each identified root cause.
2. The number of replacement defibrillation pads and replacement batteries issued to customers domestically for all causes.
3. A summary of information available to you related to individual domestic uses of your device that may include, but is not limited to:
a. Defibrillation success and the number of shocks required for success; and
b. Identification of any error codes or malfunctions during use and their related MDR number.
4. A listing of any safety alerts, technical service bulletins, user communications, or recalls for devices under this PMA.
The applicant will continue to collect data from their OUS Post Market Evaluation Study. The purpose of this post market evaluation is to survey the performance of the LIFEPAK CR2 Defibrillator in the Basic Life Support (BLS) use environment by characterizing the performance of the cprINSIGHT Analysis Technology algorithm, identifying any unanticipated use issues or performance concerns directly attributable to the cprINSIGHT algorithm and collect general device usability data from end users.
The applicant's manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820).
## XIV. APPROVAL SPECIFICATIONS
Directions for use: See device labeling.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, and Cautions in the device labeling.
Post-approval Requirements and Restrictions: See approval order.
## XV. REFERENCES
The following publications are particularly relevant to the clinical safety and effectiveness of the LIFEPAK CR2.
1) American Heart Association (AHA) Scientific Statement: “Automatic External Defibrillators for Public Access Defibrillation: Recommendations for Specifying and
PMA P170018: FDA Summary of Safety and Effectiveness Data
Page 15
{15}
Reporting Arrhythmia Algorithm Performance, Incorporating New Waveforms, and Enhancing Safety," Circulation 1997; 95:1677-1682.
2) Killingsworth CR, Melnick SB, Chapman FW, Walker RG, Smith WM, Ideker RE, Walcott GP. Defibrillation threshold and cardiac responses using an external biphasic defibrillator with pediatric and adult adhesive patches in pediatric-sized piglets. Resuscitation 2002; 55:177-185.
3) Higgins SL, Herre JM, Epstein AE, Greer, SG, Freidman PL, Gleva ML, Porterfield JG, Chapman FW, Finkel ES, Schmitt PW, Nova RC, Greene HL. A Comparison of Biphasic and Monophasic Shocks for External Defibrillation. Prehospital Emergency Care 2000;4(4):305-313.
4) Van Alem AP, Chapman FW, Lank P, Hart AAM, Koster RW. A prospective, randomised and blinded comparison of first shock success of monophasic and biphasic waveforms in out-of-hospital cardiac arrest. Resuscitation 2003;58(1):17-24.
PMA P170018: FDA Summary of Safety and Effectiveness Data
Page 16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.