RHA® Dynamic Volume, RHA®4
P170002S048 · Teoxane S.A. · LMH · Sep 18, 2025 · General, Plastic Surgery
Device Facts
| Record ID | P170002S048 |
| Device Name | RHA® Dynamic Volume, RHA®4 |
| Applicant | Teoxane S.A. |
| Product Code | LMH · General, Plastic Surgery |
| Decision Date | Sep 18, 2025 |
| Decision | APPR |
| Device Class | Class 3 |
| Attributes | Therapeutic |
Indications for Use
RHA4® is indicated for injection into the deep dermis to superficial subcutaneous tissue for the correction of moderate to severe dynamic facial wrinkles and folds, such as nasolabial folds (NLF), in adults aged 22 years or older. RHA®4 is indicated for injection into the subcutaneous to supraperiosteal layers for cheeks augmentation and/or correction of age-related midface contour deficiencies, in adults aged 22 years or older. RHA® Dynamic Volume is indicated for injection into the deep dermis to superficial subcutaneous tissue for the correction of moderate to severe dynamic facial wrinkles and folds, such as nasolabial folds (NLF), in adults aged 22 years or older. RHA® Dynamic Volume is indicated for injection into the subcutaneous to supraperiosteal layers for cheeks augmentation and/or correction of age-related midface contour deficiencies, in adults aged 22 years or older.
Device Story
RHA®4 and RHA® Dynamic Volume are sterile, biodegradable, viscoelastic sodium hyaluronate (NaHA) gel implants crosslinked with BDDE. RHA®4 contains lidocaine; RHA® Dynamic Volume contains mepivacaine. Injected by physicians into deep dermis, subcutaneous, or supraperiosteal layers using needles or cannulas. Used for midface volume restoration and wrinkle correction. Provides immediate structural support and volume; results sustained over 12 months. Clinical benefit measured via Teoxane Midface Volume Deficiency Scale (TMVDS) and patient-reported outcomes (FACE-Q).
Clinical Evidence
Prospective, randomized, blinded-evaluator, multicenter, non-inferiority study (N=201). Primary endpoint: TMVDS change from baseline at 8 weeks post-treatment. RHA®4 mean change -1.3 vs. Control -1.3 (95% CI difference -0.22 to 0.25). Responder rate (≥1-grade improvement) 87.0% (RHA®4) vs 90.5% (Control). Long-term safety/effectiveness evaluated over 52 weeks. No device-related serious adverse events reported.
Technological Characteristics
Viscoelastic NaHA gel (23 mg/g) crosslinked with BDDE; pH 7.3. RHA®4 contains 0.3% lidocaine; RHA® Dynamic Volume contains 0.3% mepivacaine. Supplied in 1.2 ml syringes with 27G needles. Biocompatibility per ISO 10993 (cytotoxicity, sensitization, genotoxicity, systemic toxicity). Shelf life 24 months at ≤25°C.
Indications for Use
Indicated for adults aged 22+ for correction of moderate to severe dynamic facial wrinkles/folds (e.g., nasolabial folds) and for cheek augmentation/correction of age-related midface contour deficiencies. Contraindicated in patients with severe allergies (anaphylaxis history), known hypersensitivity to gram-positive bacterial proteins or amide-type local anesthetics (lidocaine/mepivacaine), and bleeding disorders.
Predicate Devices
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
## I. GENERAL INFORMATION
Device Generic Name: Injectable Dermal Filler
Device Trade Name: RHA®4 and RHA® Dynamic Volume
Device Procode: LMH (Implant, Dermal, For Aesthetic Use)
Applicant's Name and Address: TEOXANE S.A.
Les Charmilles
Rue de Lyon, 105
CH - 1203 Geneva, Switzerland
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P170002/S048
Date of FDA Notice of Approval: September 18, 2025
RHA®4 was approved on October 19, 2017 under the original PMA P170002 and RHA® Dynamic Volume (formerly RHA®4 Mepi) was approved on October 13, 2023 under panel-track supplement P170002/S026. Both were approved for injection into the deep dermis to superficial subcutaneous tissue for the correction of moderate to severe dynamic facial wrinkles and folds, such as nasolabial folds (NLFs), in adults aged 22 years or older. The SSEDs to support the indication of RHA®4 and RHA®4 Dynamic Volume for the correction of moderate to severe dynamic facial wrinkles and folds, such as nasolabial folds (NLFs) are available on the CDRH website and are incorporated by reference here.
The current supplement was submitted to expand the indication of RHA®4 and RHA® Dynamic Volume for injection into the subcutaneous to supraperiosteal layers for cheeks augmentation and/or correction of age-related midface contour deficiencies, in adults aged 22 years or older.
## II. INDICATIONS FOR USE
RHA4® is indicated for injection into the deep dermis to superficial subcutaneous tissue for the correction of moderate to severe dynamic facial wrinkles and folds, such as nasolabial folds (NLF), in adults aged 22 years or older.
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 1
{1}
RHA®4 is indicated for injection into the subcutaneous to supraperiosteal layers for cheeks augmentation and/or correction of age-related midface contour deficiencies, in adults aged 22 years or older.
RHA® Dynamic Volume is indicated for injection into the deep dermis to superficial subcutaneous tissue for the correction of moderate to severe dynamic facial wrinkles and folds, such as nasolabial folds (NLF), in adults aged 22 years or older.
RHA® Dynamic Volume is indicated for injection into the subcutaneous to supraperiosteal layers for cheeks augmentation and/or correction of age-related midface contour deficiencies, in adults aged 22 years or older.
### III. CONTRAINDICATIONS
- RHA®4 and RHA® Dynamic Volume are contraindicated for patients with severe allergies manifested by a history of anaphylaxis or history or presence of multiple severe allergies.
- RHA®4 and RHA® Dynamic Volume contain trace amounts of gram-positive bacterial proteins and is contraindicated for patients with a history of allergies to such material.
- RHA®4 and RHA® Dynamic Volume should not be used in patients with previous hypersensitivity to local anesthetics of the amide type, such as lidocaine or mepivacaine.
- RHA®4 and RHA® Dynamic Volume should not be used in patients with bleeding disorders.
### IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the RHA®4 and RHA® Dynamic Volume labeling.
### V. DEVICE DESCRIPTION
RHA®4 and RHA® Dynamic Volume are viscoelastic, sterile, non-pyrogenic, clear, colorless, and biodegradable gel implants. They are produced with sodium hyaluronate (NaHA) with a concentration of 23 mg/g obtained from bacterial fermentation using a Streptococcus equi bacterial strain, crosslinked with 1,4-butanediol diglycidyl ether (BDDE) and reconstituted in a physiological buffer (pH 7.3).
RHA®4 contains 0.3% lidocaine hydrochloride monohydrate to reduce pain on injection.
RHA® Dynamic Volume contains 0.3% mepivacaine hydrochloride to reduce pain on injection.
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 2
{2}
RHA®4 and RHA® Dynamic Volume are supplied in individual blisters containing a 1.2 ml treatment syringe with two 27G ½ inch hypodermic needles.
## VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are several other alternatives for cheek augmentation and correction of age-related midface contour deficiencies: surgical implants, autologous fat injections, face-lift surgery, approved soft tissue filler for this indication. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
## VII. MARKETING HISTORY
RHA®4 has been sold in the United States since 2020 and RHA®4 Mepi for the approved indication when injected into the dynamic facial wrinkles and folds will be launched late 2025. RHA®4 has been commercially available in the European Union and registered in more than 45 countries around the world since 2015. It has been approved for a wide range of indications including for midface volume deficiency. Neither RHA®4 or RHA®4 Mepi has been removed from the marketplace for any reasons related to safety or effectiveness.
## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Common treatment responses which can occur with the use of RHA®4, RHA® Dynamic Volume and other dermal fillers, include bruising, discoloration, firmness (induration), itching, lumps/bumps (injection site mass), pain, redness, swelling and tenderness. All these common treatment responses were seen in the clinical studies.
In addition to the common treatment responses noted above, the following adverse events were reported from use in the nasolabial folds and other locations of the face, as part of the post-marketing surveillance on the use of RHA®4 in and outside the United States. The following adverse events were reported on the use of RHA®4 worldwide with a prevalence equal or superior to one occurrence for 100,000 syringes: skin edema, skin swelling, injection site mass (inflammatory or non-inflammatory), skin induration, injection site inflammation, pain, erythema, granuloma, vascular complications, ecchymosis, skin infection and tenderness.
Additionally, other less frequent adverse reactions have also been reported, including implant migration, allergic reaction, skin discoloration/Tyndall effect, abscess, overcorrection, pruritus, anaphylactic reaction, skin necrosis, urticaria, blister, scab, angioedema, chapped lips, dermatitis, dry skin, fibrosis, herpes breakout, numbness, pustules, telangiectasia and visual impairment.
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 3
{3}
Delayed-onset inflammation near the site of dermal filler injections is one of the known adverse events associated with dermal fillers. Cases of delayed-onset inflammation have been reported to occur at the dermal filler treatment site following viral or bacterial illnesses or infections, vaccinations, or dental procedures. Typically, the reported inflammation was responsive to treatment or resolved on its own.
Additionally, the following rare but serious adverse events that are associated with intravascular injection of soft tissue filler material in the face have been reported in the literature: vision impairment (acute or permanent), blindness, cerebral ischemia or cerebral hemorrhage leading to stroke, and skin necrosis
In many cases the symptoms resolved without any treatment. Reported treatments included the use of (in alphabetical order): analgesics, antibiotics, antihistamines, anti-inflammatories, anti-viral, drainage, excision, implant dissolution (hyaluronidase), incision, massage and vasodilators. Outcomes for these reported events ranged from resolved to ongoing at the time of last contact.
For the specific adverse events that occurred in the clinical study, please see Section X below.
### IX. SUMMARY OF NON-CLINICAL STUDIES
#### A. Laboratory Studies
Table 1: Physical and Chemical Testing – Requirements for RHA®4 and RHA® Dynamic Volume
| Test | Purpose | Results |
| --- | --- | --- |
| NaHA content | To confirm the NaHA concentration meets specifications | Passed |
| Sterility | To ensure the product is sterile | Passed |
| Bacterial Endotoxins | To confirm the endotoxins count in the device meets specifications | Passed |
| pH | To confirm the pH of the gel meet specifications | Passed |
| Residual crosslinker content | To confirm the residual crosslinker content of the gel meets specifications | Passed |
| Anesthetic content Lidocaine for RHA®4 Mepivacaine for RHA® Dynamic Volume | To confirm the anesthetic concentration of the gel meets specifications | Passed |
| Impurities deriving from anesthetic | To confirm impurities in the gel meet specifications | Passed |
| Extrusion force | To confirm the extrusion force meets specifications | Passed |
| Rheology: mechanical properties of the gel | To confirm phase angle δ of the gel meets specifications | Passed |
| Appearance of the device | To control visually the absence of irregularities and defects in the device | Passed |
| Gel content (RHA® Dynamic Volume) | To ensure gel meets specifications | Passed |
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 4
{4}
## B. Biocompatibility Studies
Table 2: Summary of biocompatibility studies for RHA®4 and RHA® Dynamic Volume
| Test | Method | ISO Standard or USP | Results |
| --- | --- | --- | --- |
| Cytotoxicity | In vitro mammalian cell culture test | ISO 10993-5 | Same cytotoxic potential as control*. |
| Sensitization | Guinea pig maximization study | ISO 10993-10 | No delayed sensitization. |
| Intracutaneous reactivity | Intradermal injection in rabbits. | ISO 10993-10** | Level of reactivity similar or slightly less than its control*. RHA®4 was non-irritant at 26 days. RHA® Dynamic Volume was non-irritant at 20 days. |
| Pyrogenicity | Rabbit pyrogen study | USP 36 NF 31 | Non-pyrogenic. |
| Genotoxicity | Ames test (bacterial reverse mutation study) | ISO 10993-3 | Non-mutagenic. |
| Genotoxicity | Mouse lymphoma assay | ISO 10993-3 | Non- mutagenic. |
| Genotoxicity | Mouse peripheral blood micronucleus test (for RHA®4 only) | ISO 10993-3 | Non-genotoxic. |
| Acute systemic toxicity | Mice intraperitoneal study | ISO 10993-11 | No evidence of acute systemic toxicity. |
| Sub-acute and subchronic systemic toxicity | Intradermal injection in Sprague-Dawley rats (for RHA®4 only) | ISO 10993-11 | No toxicological concern. |
| Intradermal implantation | Intradermal implantation in rats | ISO 10993-6 | The test article was classified as non-irritant. No adverse response microscopically or macroscopically. After 52 weeks, degradation had started. |
(*) Note: The control device was an FDA approved Hyaluronic Acid soft tissue filler, with similar characteristics. The control product is legally marketed with similar indications for use
(**) Note: This standard has been updated in 2021 and now covers only sensitization tests, ISO 10993-23 is now covering irritation tests. However, in vivo intracutaneous tests were carried out based on a methodology which has not been modified. Therefore, results remain valid accordingly for irritation assays performed on RHA® and RHA® Dynamic Volume products.
## C. Stability studies
Stability data have been collected on three batches per product (RHA®4 and RHA® Dynamic Volume) through a 24-month stability study conducted at 25°C ± 2°C and 60% ± 5% relative humidity (ICH long term storage conditions). As part of stability testing, products were characterized via microbiological, physical, and chemical specifications parameters, including lidocaine hydrochloride monohydrate content for RHA®4 and mepivacaine hydrochloride content for RHA® Dynamic Volume, and lidocaine-related or mepivacaine-related degradation products. Conformance of all
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 5
{5}
batches with all specifications was confirmed up to the end of the studies. Therefore, RHA®4 and RHA® Dynamic Volume dermal fillers have a 24-month shelf life with recommended storage conditions up to 25°C (77°F).
## X. SUMMARY OF PRIMARY CLINICAL STUDY
The applicant performed one clinical study to establish a reasonable assurance of safety and effectiveness when injecting RHA®4 into the subcutaneous to supraperiosteal layers for cheek augmentation and/or correction of age-related midface contour deficiencies in adults aged 22 years or over in the U.S. under IDE # G210086. Data from this clinical study was the basis for the PMA approval decision. Those data were leveraged for RHA® Dynamic Volume as they are identical products with a different anesthetic agent. Clinical study IDE G190118 had demonstrated the non-inferiority of mepivacaine to lidocaine in the device. A summary of the clinical study with RHA®4 is presented below.
Please note: RHA® Dynamic Volume is strictly identical to RHA®4 except for the small amount of anesthetic medicine: RHA® Dynamic Volume contains mepivacaine and RHA®4 contains lidocaine. Both anesthetics agents are of the same family with the same mechanisms of effect. Details about the supporting clinical data of the equivalency of both products were documented under SSED of panel-track supplement P170002/S026. RHA® Dynamic Volume and RHA®4 have the same indication. The long-term safety and effectiveness of RHA® Dynamic Volume were evaluated in a clinical study (presented hereafter) using RHA®4 over 52 weeks after initial or touch-up treatments and for 12 weeks following retreatment.
### A. Study Design
Subjects were treated between October 29, 2021 and January 29, 2024. The database for this PMA reflected data collected through March 19, 2024 and included 201 treated subjects with 152 subjects who received injection with RHA®4 and 49 subjects with the control. There were 8 U.S. investigational sites.
The study was a controlled, randomized, blinded evaluator, between subject, multicenter, prospective non-inferiority clinical study to evaluate the safety and effectiveness of RHA®4 for cheek augmentation and/or midface contour deficiencies against a control. Subjects meeting inclusion/exclusion criteria were randomized 3:1 ratio into the treatment or control group (Juvederm® Voluma™ XC [Allergan]). The control was an FDA approved hyaluronic acid soft tissue filler with similar characteristics to RHA®4 and marketed for midface correction. The study duration was nearly 16 months and included repeat treatment.
#### 1. Clinical Inclusion and Exclusion Criteria
Enrollment in the pivotal study was limited to subjects who met the following inclusion criteria
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 6
{6}
- Outpatient, male or female of any race, 22 years of age or older. Female patients of childbearing potential must have had a negative urine pregnancy test (UPT) at Visit 1 and practice 1 a reliable method of contraception throughout the study:
- Had midface volume loss of grades 2 to 3 on the Teoxane Midface Volume Deficiency Scale (TMVDS) (ranging from 0 = absent to 4 = very severe). The blinded live evaluation (BLE) and treating investigator (TI) must independently assess and agree that this criterion was met; however, concordance of fullness was not required. If the assessments of the TI and the BLE were the same or differ by exactly 1 point on the scale, this difference was considered acceptable.
- Had cheeks of the same TMVDS grade on the left and right side of the face (i.e., approximate bilateral symmetry).
- Was willing to abstain from facial aesthetic procedures/therapies that could interfere with study evaluations (e.g., other fillers, botulinum toxin injections, laser, or chemical resurfacing, facial plastic surgery) for the duration of the study
Patients were not permitted to enroll in the study if they met any of the following exclusion criteria:
- Had either one of the following assessments during the vision tests: Snellen acuity test worse than 20/40 (with corrections, if applicable); abnormal confrontational visual field test; or abnormal ocular motility test
- Had known hypersensitivity or previous allergic reaction to any component of the study or the control device
- Had known sensitivity to local anesthetics of the amide type, including known or suspected lidocaine hypersensitivity, history of multiple severe allergies, or history of anaphylactic shock
- Had known susceptibility to keloid formation, hypertrophic scarring, or clinically significant skin pigmentation disorders as judged by the TI
- Had a history of severe chronic systemic diseases, including but not limited to, poorly controlled diabetes mellitus (all types), congestive heart failure, severe liver disease, severe kidney disease, and others that are likely to interfere with the measured parameters or to put the subjects to an undue risk as judged by the TI
- Had a history of connective tissue disease (rheumatoid arthritis, scleroderma, systemic lupus erythematosus)
- Had a need for clinically significant, as judged by the TI, and continuous medical treatment within 2 weeks prior to Visit 1
- Had an active inflammation, infection, cancerous or precancerous lesion, or unhealed wound on the midface, or has undergone radiation treatment in the area to be treated
- Had any significant concomitant skin disease present within 6 months prior to Visit 1 (such as moderate to severe rosacea, acne vulgaris, actinic keratosis, or seborrheic keratosis) that
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 7
{7}
may interfere with study assessments and interpretation of study results or would require medical or surgical intervention on the cheek during the study
- Exhibited any physical attribute that may prevent assessment or treatment of the midface volume deficiency, such as excessive facial hair (men with a beard are required to shave before each study visit), traumatic or noticeable facial scars, tattoos, and/or excessive hyperpigmentation in the treatment areas, as judged by the TI
- Had midface volume deficit due to congenital defect, trauma, abnormalities in adipose tissue related to immune-mediated diseases such as generalized lipodystrophy (e.g., juvenile dermatomyositis), partial lipodystrophy (e.g., Barraquer-Simons syndrome), and inherited disease
- Had known prolonged bleeding times because of disease or anticoagulant medication (including warfarin or low-dose aspirin). If food supplements with potential anticoagulant or antiplatelet effect (e.g., fish oil, omega-3, vitamin E, herbal supplements with garlic, or gingko biloba) were taken, the subject must have stopped them and waited 1 week or until bleeding times returned to normal before injecting. Supplements should not have been used for 1 week after the injection
- Had previously undergone or was planning to undergo prohibited treatment/procedures during certain time periods
- Had received any investigational product within 90 days prior to Visit 1 or planned to participate in another investigation during the course of the study
- Had medical or psychiatric conditions that may increase the risk associated with study participation or may interfere with the interpretation of study results or compliance of the subject and would make the subject inappropriate for entry into this study, as judged by the TI
- Had clinically significant alcohol or drug abuse, or history of poor cooperation or unreliability, as judged by the TI
- Was study staff or a close relative to study staff (e.g., parents, children, siblings, or spouse).
- Had unstable weight, as judged by the TI
- Had no teeth or misses significant portion of dentition that changes the face appearance; subjects with dentures sufficiently replacing missing natural teeth are allowed
- Currently received or had received oral or injectable corticosteroids, immunosuppressants or chemotherapy within 3 months before Visit 1. Inhaled and topical corticotherapy not including head or neck was permitted
# 2. Follow-up Schedule
At Visit 1, subjects were randomized to RHA®4 or control treatment group. They received injection in their midface region as defined by the protocol. The Treating Investigator (TI) determined the injection technique and depth of injection (subcutaneous and/or supraperiosteal). RHA®4 may be injected with a 27G ½” needle or 25G 2” cannula using a multilayering technique. The control
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 8
{8}
device was injected as per its IFU). The maximum volume of administration of RHA®4 was 3 mL per treatment session on each side, with a maximum of 6 mL at each treatment session. After 4 weeks, subjects may receive additional treatment with the same device as originally injected (RHA®4 or control) to achieve optimal correction if deemed necessary by the TI.
Following any injection (initial, touch up, retreatment), subjects were given a 30-Day diary to daily record common treatment responses (CTR) and any other adverse observations. They were instructed to record the severity of each CTR as mild, moderate, or severe.
All subjects were scheduled to return for follow-up examinations at 4, 8, 16, 24, 36, and 52 weeks after last treatment (initial or touch-up). The primary effectiveness endpoint (TMVDS) was evaluated at 8 weeks after last treatment. Subjects were followed for 52 weeks to evaluate the long-term safety, safety of retreatment and other secondary endpoints of RHA®4. Subjects were offered retreatment 52 weeks after last treatment. Retreatment was with RHA®4 irrespective of the assigned device at initial and touch-up treatment. Adverse events and complications were recorded at all visits.
The key timepoints are shown below in the tables summarizing safety and effectiveness.
**Table 3: Treatment schedule and assessments**
| | | Treatment (Tx) groups: RHA®4/ Control | Safety Assessments | Effectiveness Assessments |
| --- | --- | --- | --- | --- |
| Initial treatment | V1 | 0 weeks Initial treatment | Visual assessments CTR diary AEs Pain VAS | BLE: TMVDS, NLF-WSRS, TIOHS TI: TMVDS Subject: Face-Q, TMNSQ |
| | V1a | 3 days - Phone call | CTR diary AEs | N/A |
| | V1b | 2 weeks | Visual assessments CTR diary AEs | N/A |
| Optional touch-up | V2 | 4 weeks Optional TU | Visual assessments CTR diary AEs Pain VAS if TU tx | TI: TMVDS, GAIS Subject: Face-Q, GAIS, Subject satisfaction, TMNSQ |
| If touch-up | V2a | 3 days after TU – Phone call | CTR diary AEs | N/A |
| | V2b | 2 weeks after TU | Visual assessments | N/A |
| | V2c | 4 week after TU | CTR diary AEs | TI: TMVDS, GAIS Subject: Face-Q, GAIS, Subject satisfaction, TMNSQ |
| | V3 | 8 weeks after last tx **Primary endpoint assessment** | Visual assessments AEs | BLE: TMVDS, NLF-WSRS, TIOHS, GAIS TI: TMVDS, GAIS Subject: Face-Q, GAIS, Subject satisfaction, TMNSQ |
| | V4 | 16 weeks after last tx | | |
| | V5 | 24 weeks after last tx | | |
| | V6 | 36 weeks after last tx | | |
| Optional retreatment | V7 | 52 weeks after last tx – Exit or | Visual assessments CTR diary AEs | BLE: TMVDS, NLF-WSRS, TIOHS, GAIS TI: TMVDS, GAIS |
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 9
{9}
| | | Treatment (Tx) groups: RHA®4/ Control | Safety Assessments | Effectiveness Assessments |
| --- | --- | --- | --- | --- |
| | | Optional retreatment with RHA® Dynamic Volume for both treatment groups | Pain VAS if retreatment | Subject: Face-Q, GAIS, Subject satisfaction, TMNSQ |
| If retreatment | V7a | +3 days after retreatment | CTR diary AEs | N/A |
| | V7b | 2 weeks after retreatment | Visual assessments | N/A |
| | V7c | 4 weeks after retreatment | CTR diary AEs | BLE: TMVDS, NLF-WSRS, TIOHS, GAIS TI: TMVDS, GAIS Subject: Face-Q, GAIS, Subject satisfaction, TMNSQ |
| | V7d | 12 weeks after retreatment - Exit | Visual assessments AEs | |
Abbreviations: AE: Adverse Event; BLE: Blinded Live Evaluator; CTR: Common Treatment Response; FACE-Q: Satisfaction with cheeks questionnaire; GAIS: Global Aesthetic Improvement Scale; NLF-WSRS: Nasolabial Folds Wrinkle Severity Scale; TI: Treatment Investigator; TIOHS: Teoxane Infraorbital Hollows Scale TMNSQ: Teoxane Midface Naturalness Subject Questionnaire; TMVDS: Teoxane Midface Volume Deficit Scale; TU: Touch-up; Tx: Treatment; VAS: Visual Analog Scale
### 3. Clinical Endpoints
Safety was evaluated through a 30-Day patient Common Treatment Response (CTR) diary (after each injection), measures of injection site pain, visual assessments and AE assessments at each visit.
Subjects recorded the presence, duration, and severity of CTRs that may occur following the injection of a dermal filler, for the first 30 days after each treatment (initial, touch-up, and retreatment) in a patient diary: redness, pain, tenderness, firmness, swelling, lumps/bumps, bruising, itching, discoloration, and other. Subjects could fill out the description of the symptoms in the “other” category and it was automatically categorized as an AE. In addition, the CTR diary captured the occurrence of events associated with visual disturbances and possible symptoms of intravascular injection. Such events were automatically categorized as AEs and the subject was required to seek immediate medical assistance.
CTRs were not considered AEs unless the duration and/or severity were in excess of that typically observed following injection of a dermal filler and were clinically significant as determined by the TI. Additionally, CTRs that were present on the last day of diary entry, regardless of severity, were automatically recorded as AEs.
The TI assessed all AEs and recorded details of seriousness, severity, duration, and action taken with study device, as well as relationship to the study device. For statistical analysis, the maximal severity reported for the AE was used, even if the AE presented as being less severe at some point during the event.
Safety was also evaluated at each visit with visual assessment tests (before and 30 min post-injections, and at each study visit after the last treatment).
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 10
{10}
Pain at the injection site(s) was self-assessed by the subject using a 100 mm Visual Analog Scale (VAS), with the left end representing “no pain” and the right end representing “worst pain” during injection and 5, 15 and 30 minutes after injection.
Effectiveness was measured by assessing midface volume improvement based on the validated Teoxane Midface Volume Deficiency Scale (TMVDS$^{1}$) (Table 4) from pre-injection midface volume treated with the RHA$^{®}$4 compared to the improvement from pre-injection midface volume treated with the control device, as assessed by the BLE at 8 weeks after the last injection (initial or touch-up).
**Table 4: Teoxane Midface Volume Deficiency Scale (TMVDS)**
| Category | Grade | Description |
| --- | --- | --- |
| Absent | 0 | - **Full or round face.** - No visible signs of volume deficit noted in the overall midface. |
| Mild | 1 | - **Presence of slight flattening of the anteromedial cheek or very minor contour deficit in the lateral malar area.** - Presence of mild infra-orbital hollows and/or mild nasolabial folds. |
| Moderate | 2 | - **Presence of moderate flattening of the anteromedial cheek and/or moderate hollowing in the lateral malar area.** - Descent of the facial fat pads creating moderate nasolabial folds and mild to moderate infra orbital hollows. - Mild pre-jowl sulcus and/or mild to moderate naso-jugal folds extending beyond the mid pupil. |
| Severe | 3 | - **Significant flattening of the anteromedial cheek and/or severe submalar hollowing.** - Obvious descent of the facial fat pads creating bony prominence of the zygoma. - Moderate to severe infra-orbital hollowing and nasolabial folds. - May have moderate presence of pre-jowl sulcus and jowls. |
| Very Severe | 4 | - **Very severe facial wasting.** - Obvious descent of the facial fat pads creating very deep, heavy / nasolabial folds and jowls. - Very severe infra orbital hollowing and deep naso-jugal folds. - Prominence of bony landmarks, visibility of facial musculature, and/or moderate to severe pre-jowl sulcus. |
The primary effectiveness endpoint was the difference in the Teoxane Midface Volume Deficiency Scale (TMVDS) change from Baseline to 8 weeks after last injection (initial or touch-up) between subjects treated with RHA$^{®}$4 and those treated with the control. Assessment of the subject’s midface volume was based on the validated Teoxane Midface Volume Deficiency Scale (TMVDS) as rated by the Blinded Live Evaluator. The proportion of responders with a $\geq 1$-grade point improvement on the TMVDS at 8 weeks after last injection (initial or touch-up) when compared to pretreatment (Baseline) was calculated. A change in the TMVDS $\geq 1$ grade compared to pretreatment would be considered clinically meaningful.
Secondary effectiveness endpoints throughout the course of the study included: change from Baseline and responder rate of TMVDS score as rated by the BLE and TI (responder is a change of $\geq 1$-grade
$^{1}$ The development and validation of the TMVDS has been published in the *Journal Dermatologic Surgery*, by Bury et al., in July 2025 (Creation and validation of a photonumeric scale for assessment of midface volume deficit).
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 11
{11}
on the TMVDS); Global Aesthetic Improvement (GAI) as assessed by the BLE, TI and subject; Change from Baseline and responder rate of TMVDS score as rated by the Independent Photographic Reviewer (IPR) at 8 weeks after last treatment; Change from Baseline and responder rate of the severity of the nasolabial folds as assessed by the BLE using the validated Teoxane Nasolabial Fold Wrinkle Severity Rating Scale (NLF-WSRS); Change from Baseline and responder rate of the severity of the infraorbital hollows as assessed by the BLE using the validated Teoxane Infra-Orbital Hollows Scale (TIOHS); Impact and effectiveness of study treatment from the subjects' perspective, as assessed by the *Satisfaction with cheeks* domain of the validated FACE-Q® patient-reported outcome measurement at rest and when smiling, by the subject satisfaction questionnaire and the naturalness of the cheek subject questionnaire;
With regard to success/failure criteria, the effectiveness of RHA®4 would be demonstrated if the TMVDS change from Baseline for subjects treated with RHA®4 was statistically non-inferior to the change from Baseline for subjects treated with the control with a non-inferiority margin of $\geq 0.5$, and the proportion of responders among subjects treated with the control is $\geq 70\%$.
To demonstrate the TMVDS (BLE) change from Baseline at Visit 3 (week 8) after last treatment of RHA®4 was non-inferior to that of Control, the following hypotheses for the difference in means was tested:
$$H_0: d_{RHA4} - d_{Control} \geq 0.5$$
$$H_1: d_{RHA4} - d_{Control} < 0.5$$
If the upper bound of the two-sided 95% confidence interval (CI) of the difference of RHA®4 Minus Control was $< 0.5$, then the non-inferiority was demonstrated (with 0.5 as the non-inferiority margin).
## **B. Accountability of PMA Cohort**
At the time of database lock, a total of 257 subjects were screened. 56 subjects did not meet the inclusion/exclusion criteria resulting in 201 enrolled subjects. Out of these enrolled subjects:
- • 201 subjects (100%) were assigned to the Safety Population
- • 201 subjects (100%) to the ITT population
- • 173 subjects (86.1%) to the PP population (21 subjects (13.8%) in RHA®4 group and 7 subjects (14.3%) in control group: 13 subjects missed visit 2 (to allow for touch-up) or visit 3 (primary endpoint); 12 subjects had deviation of effectiveness assessment impacting the primary endpoint; 8 subjects had their primary endpoint visit out-of-window, 2 subjects had a deviation related to the administration of the device and 1 subject was related to an exclusion criteria).
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 12
{12}
**Table 5: Subject Accountability and Disposition**
| | RHA^{®}4 | Control |
| --- | --- | --- |
| Screened | 257 | |
| Enrolled | 201 | |
| Safety population | 152 | 49 |
| ITT population | 152 | 49 |
| PP population | 131 | 42 |
| Number of subjects who completed the study | 126 (82.9%) | 43 (87.8%) |
| Number of subjects at primary endpoint visit (V3) | 137 (90.1%) | 45 (91.8%) |
| Number of subjects at V7 | 128 (84.2%) | 44 (89.8%) |
| Number of subjects who received repeat treatment at V7 | 77 (50.7%) | 29 (59.2%) |
| Number of subjects withdrawn from study | 26 (17.1%) | 6 (12.2%) |
| Reason for discontinuation | | |
| *A subject or legal representative withdrawal* | 18 (11.8%) | 1 (2.0%) |
| *Lost to follow-up* | 6 (3.9%) | 3 (6.1%) |
| *Sponsor decision* | 1 (0.7%) | 1 (2.0%) |
| *Other* | 1 (0.7%) | 1 (2.0%) |
All percentages are based on the number of subjects by group in the safety population.
### **C. Study Population Demographics and Baseline Parameters**
The demographics of the study population are typical for a pivotal study performed in the United States.
The study population included female and male subjects who were $\geq 22$ years old. The study ensured that subjects meeting the inclusion criteria were representative of gender and ethnicity of the U.S. population who may use RHA$^{®}$4 implant for midface volume deficiency. Subjects had a mean age of approximately 55 years, and most subjects were female (approximately 89.6% (180/201) of study subjects).
The majority of subjects were white (87.0% - 174/201), with 12.5% (25/201) of subjects identifying as Black or African American. 19.9% (40/201) of subjects were of Hispanic or Latino ethnicity. Fitzpatrick Skin Types were appropriately represented with all predefined minimal sample size thresholds being attained for subjects with skin types IV to VI, with approximately 71.1% (143/201) and 28.9% (58/201) of subjects had skin types I-III and IV-VI, respectively (and 14% (28/201) of subjects having skin types V/VI) (with 9.0% (18/201) of type V and 5.0% (10/201) of type VI).
**Table 6: Demographic Characteristics at Baseline**
| Demographic Variable | RHA^{®}4 N=152 | Control N=49 | Total N=201 |
| --- | --- | --- | --- |
| **Age (years)** | | | |
| N (missing) | 152 (0) | 49 (0) | 201 (0) |
| Mean $\pm$ SD | 55.4 $\pm$ 9.97 | 55.6 $\pm$ 7.85 | 55.5 $\pm$ 9.48 |
| Min, Max | 24, 79 | 34, 68 | 24, 79 |
| **Gender** | | | |
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 13
{13}
| Demographic Variable | RHA^{®}4 N=152 | Control N=49 | Total N=201 |
| --- | --- | --- | --- |
| N (missing) | 152 (0) | 49 (0) | 201 (0) |
| Male | 18 (11.8%) | 3 (6.1%) | 21 (10.4%) |
| Female | 134 (88.2%) | 46 (93.9%) | 180 (89.6%) |
| **Race^{a}** | | | |
| N (missing) | 152 (0) | 48 (1) | 200 (1) |
| American Indian or Alaska Native | 0 (0.0%) | 1 (2.1%) | 1 (0.5%) |
| Asian | 3 (2.0%) | 0 (0.0%) | 3 (1.5%) |
| Black or African American | 19 (12.5%) | 6 (12.5%) | 25 (12.5%) |
| White | 132 (86.8%) | 42 (87.5%) | 174 (87.0%) |
| **Ethnicity** | | | |
| N (missing) | 152 (0) | 49 (0) | 201 (0) |
| Hispanic or Latino | 31 (20.4%) | 9 (18.4%) | 40 (19.9%) |
| Not-Hispanic or Latino | 120 (78.9%) | 39 (79.6%) | 159 (79.1%) |
| Not available | 1 (0.7%) | 1 (2.0%) | 2 (1.0%) |
| **Fitzpatrick Skin Type** | | | |
| N | 152 (0) | 49 (0) | 201 (0) |
| **I-III** | **107 (70.4%)** | **36 (73.5%)** | **143 (71.1%)** |
| I | 5 (3.3%) | 0 (0.0%) | 5 (2.5%) |
| II | 40 (26.3%) | 15 (30.6%) | 55 (27.4%) |
| III | 62 (40.8%) | 21 (42.9%) | 83 (41.3%) |
| **IV-VI** | **45 (29.6%)** | **13 (26.5%)** | **58 (28.9%)** |
| IV | 23 (15.1%) | 7 (14.3%) | 30 (14.9%) |
| V | 14 (9.2%) | 4 (8.2%) | 18 (9.0%) |
| VI | 8 (5.3%) | 2 (4.1%) | 10 (5.0%) |
Abbreviations: N = number of subjects; SD = standard deviation
All percentages are based on the number of subjects by group with non-missing information for each variable.
a. A subject could be counted in several categories.
### Extent of exposure
The number of treatment sessions is described in Table 7 below. All subjects who received retreatment were injected with RHA$^{®}$4 only irrespective of their initial group assignment.
**Table 7: Number of treatment session – Safety population**
| Treatment | RHA^{®}4 (N=152) | Control (N=49) | Total (N=202) |
| --- | --- | --- | --- |
| Initial treatment | 152 (100%) | 49 (100%) | 201 (100%) |
| Touch-up treatment | 96 (63.2%) | 41 (83.7%) | 137 (68.2%) |
| Retreatment | 77 (50.7%) | 29 (59.2%) | 106 (52.7%) |
| Number of subjects receiving 1 treatment (i.e., only initial treatment) | 36 (23.7%) | 4 (8.2%) | 40 (19.9%) |
| Number of subjects receiving all 3 injections (initial, touch-up and retreatment) | 57 (37.5%) | 25 (51.0%) | 82 (40.6%) |
*All percentages are based on the number of subjects by group in the population.*
A smaller proportion of subjects (63.2%) in the RHA$^{®}$4 group required a touch-up when compared to the subjects in the Control group (83.7%). Similarly, a smaller proportion of subjects received 3 injections (initial treatment, touch-up and retreatment) in the RHA$^{®}$4 group than in the Control group (37.5% vs 51.0%).
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 14
{14}
The maximum volume allowed was 6.0 mL per injection session, with a maximum of 3.0 mL per cheek per session.
The average volume injected for initial treatment was similar between treatment groups with volumes of 2.87±1.13 mL and 2.50±0.98 mL in the RHA®4 and Control groups, respectively. Similar volumes were injected at initial treatment for left and right cheeks, in both treatment groups. Injection volumes used for touch-up treatment in both cheeks were similar in both treatment groups, with 1.33±0.38 mL and 1.21±0.69 mL in the RHA®4 and Control groups, respectively.
Volumes injected to achieve optimal cosmetic results were comparable in both treatment groups: a total volume of 3.70±1.56 mL and 3.51±1.37 mL was used in the RHA®4 and Control groups, respectively. Volume injected at retreatment (with RHA®4 only) at the end of the study was slightly higher in subjects initially assigned to receive the control device, with 2.10±0.81 mL compared to 1.74±0.98 mL in subjects initially assigned to receive RHA®4. No meaningful difference was observed between volumes injected in the left and right cheek.
The majority of treatments used the cannula only or combined with the needle at initial, touch-up and retreatment. Cannula only was more often used with RHA®4 than with the control: 53.9% (82/152) and 34.7% (17/49) at initial treatment respectively. A similar difference was also observed at touch-up treatment (RHA®4 only was used at retreatment). The rate of using only the needle was highest for touch-up treatment: 37.5% (36/152) and 41.5% (17/49) for RHA®4 and the control respectively as shown in Table 8. RHA®4 and the control were mainly injected using the multilayering technique (i.e. combination of subcutaneous and supraperiosteal depths of injection) at initial treatment: 75.0% (114/152) and 73.5% (36/459) respectively.
**Table 8: Method of injection and injection depth – Safety population**
| Treatment | RHA®4 (N=152) | Control (N=49) | Total (N=202) |
| --- | --- | --- | --- |
| **Initial treatment** | | | |
| Needle only | 19 (12.5%) | 7 (14.3%) | 26 (12.9%) |
| Cannula only | 82 (53.9%) | 17 (34.7%) | 99 (49.3%) |
| Needle + Cannula | 51 (33.6%) | 25 (51.0%) | 76 (37.8%) |
| Subcutaneous | 32 (21.1%) | 1 (2.0%) | 33 (16.4%) |
| Supraperiosteal | 6 (3.9%) | 12 (24.5%) | 18 (9.0%) |
| Both | 114 (75.0%) | 36 (73.5%) | 150 (74.6%) |
| **Touch-up treatment** | | | |
| Needle only | 36 (37.5%) | 17 (41.5%) | 53 (38.7%) |
| Cannula only | 39 (40.6%) | 10 (24.4%) | 49 (35.8%) |
| Needle + Cannula | 21 (21.9%) | 14 (34.1%) | 35 (25.5%) |
| Subcutaneous | 15 (15.6%) | 7 (17.1%) | 22 (16.1%) |
| Supraperiosteal | 19 (19.8%) | 14 (34.1%) | 33 (24.1%) |
| Both | 62 (64.6%) | 20 (48.8%) | 82 (59.9%) |
| **Retreatment** | | | |
| Needle only | 15 (19.5%) | 8 (27.6%) | 23 (21.7%) |
| Cannula only | 48 (62.3%) | 15 (51.7%) | 63 (59.4%) |
| Needle + Cannula | 14 (18.2%) | 6 (20.7%) | 20 (18.9%) |
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 15
{15}
| Treatment | RHA^{®}4 (N=152) | Control (N=49) | Total (N=202) |
| --- | --- | --- | --- |
| Subcutaneous | 28 (36.4%) | 11 (37.9%) | 39 (36.8%) |
| Supraperiosteal | 14 (18.2%) | 4 (13.8%) | 18 (17.0%) |
| Both | 35 (45.5%) | 14 (48.3%) | 49 (46.2%) |
## **D. Safety and Effectiveness Results**
### **1. Safety Results**
The analysis of safety was based on the cohort of 201 subjects available for up to 52 weeks evaluation after initial or touch-up treatment. The common treatment responses for this study are presented below in Table 9 to Table 11. Adverse device effects are reported in Table 12 to Table 16.
Safety of the RHA$^{®}$4 implant when injected into the midface was evaluated through a 30-Day patient Common Treatment Response (CTR) diary which was completed after each injection, AE assessments at each visit, visual assessment at each visit and measurement of injection site pain.
#### **Common Treatment Responses after initial treatment**
CTR data for initial treatment are presented in Table 9. CTRs for touch-up and repeat treatment were of the same proportions.
Out of 201 subjects, 187 subjects returned their diaries, with 141/152 in the RHA$^{®}$4 group and 46/49 in the control group. Out of those subjects, 92.0% (172/187) experienced at least 1 CTR: 90.8% (128/141) of subjects in the RHA$^{®}$4 group and 95.7% (44/46) of subjects in the control group (each subject may have reported more than 1 CTR).
After initial treatment, CTRs incidence rate was similar between treatment groups. The most common CTRs were tenderness (experienced by 81.3% (152/187) of subjects overall), firmness (75.9% (142/187) of subjects overall), swelling (67.4% (126/187) of subjects overall), and lumps/bumps (53.5% (100/187) of subjects overall).
For the RHA$^{®}$4 group, within the diaries having at least 1 CTR, 94.5% (121/128) of the diaries contained only CTRs of mild or moderate severity, while 5.5% (7/128) of the diaries contained at least 1 CTR of severe severity. This proportion was similar in both treatment groups: 5.5% (7/128) in RHA$^{®}$4 group and 6.8% (3/44) in the control group.
The most frequent severe CTR reported was firmness (4 in RHA$^{®}$4 and 2 in control group) and lumps/bumps (4 in RHA$^{®}$4 and 1 in control group). Those are consistent and anticipated from previous similar studies following an injection into the cheeks.
The duration of CTR incidence was similar between study groups. In the RHA$^{®}$4 group, 59.6% (84/128) of subjects experienced a CTR up to 7 days and 29.1% (41/128) up to 30 days. In the
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 16
{16}
control group, 54.3% (25/44) of subjects experienced a CTR up to 7 days and 30.4% (14/44) up to 30 days.
By the end of the diary, nearly all CTRs had resolved. On the last day of the diary, 12.8% (18/128) of subjects in the RHA$^{®}$4 group were experiencing a CTR with 8.7% (4/44) of subjects in the control group. The most frequent CTRs were lumps/bumps and firmness.
**Table 9: CTRs incidence rate – initial treatment – Safety population**
| CTR | RHA^{®}4 | Control | Total |
| --- | --- | --- | --- |
| Number of subjects | 152 | 49 | 201 |
| Number of CTR diaries returned | 141 | 46 | 187 |
| ≥1 CTR | 128 (90.8%) | 44 (95.7%) | 172 (92.0%) |
| Redness | 61 (43.3%) | 17 (37.0%) | 78 (41.7%) |
| Pain | 70 (49.6%) | 20 (43.5%) | 90 (48.1%) |
| Tenderness | 113 (80.1%) | 39 (84.8%) | 152 (81.3%) |
| Firmness | 108 (76.6%) | 34 (73.9%) | 142 (75.9%) |
| Swelling | 98 (69.5%) | 28 (60.9%) | 126 (67.4%) |
| Lumps/Bumps | 77 (54.6%) | 23 (50.0%) | 100 (53.5%) |
| Bruising | 74 (52.5%) | 16 (34.8%) | 90 (48.1%) |
| Itching | 17 (12.1%) | 2 (4.3%) | 19 (10.2%) |
| Discoloration | 32 (22.7%) | 6 (13.0%) | 38 (20.3%) |
Abbreviation: CTR = Common Treatment Response. All percentages are based on the number of CTR diaries returned by injection by subgroup in the population.
**Table 10: CTRs by maximum severity – initial treatment – Safety population**
| CTR | Severity | RHA^{®}4 | Control | Total |
| --- | --- | --- | --- | --- |
| Number of subjects | | 152 | 49 | 201 |
| Number of CTR diaries returned | | 141 | 46 | 187 |
| ≥1 CTR | | 128 (90.8%) | 44 (95.7%) | 172 (92.0%) |
| At least 1 CTR | Mild | 69 (53.9%) | 26 (59.1%) | 95 (55.2%) |
| | Moderate | 52 (40.6%) | 15 (34.1%) | 67 (39.0%) |
| | Severe | 7 (5.5%) | 3 (6.8%) | 10 (5.8%) |
| Redness | Mild | 51 (83.6%) | 13 (76.5%) | 64 (82.1%) |
| | Moderate | 10 (16.4%) | 3 (17.6%) | 13 (16.7%) |
| | Severe | 0 | 1 (5.9%) | 1 (1.3%) |
| Pain | Mild | 55 (78.6%) | 18 (90.0%) | 73 (81.1%) |
| | Moderate | 15 (21.4%) | 2 (10.0%) | 17 (18.9%) |
| | Severe | 0 | 0 | 0 |
| Tenderness | Mild | 86 (76.1%) | 33 (84.6%) | 119 (78.3%) |
| | Moderate | 26 (23.0%) | 5 (12.8%) | 31 (20.4%) |
| | Severe | 1 (0.9%) | 1 (2.6%) | 2 (1.3%) |
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 17
{17}
| CTR | Severity | RHA*4 | Control | Total |
| --- | --- | --- | --- | --- |
| Firmness | Mild | 72 (66.7%) | 21 (61.8%) | 93 (65.5%) |
| | Moderate | 32 (29.6%) | 11 (32.4%) | 43 (30.3%) |
| | Severe | 4 (3.7%) | 2 (5.9%) | 6 (4.2%) |
| Swelling | Mild | 68 (69.4%) | 24 (85.7%) | 92 (73.0%) |
| | Moderate | 29 (29.6%) | 4 (14.3%) | 33 (26.2%) |
| | Severe | 1 (1.0%) | 0 | 1 (0.8%) |
| Lumps/Bumps | Mild | 43 (55.8%) | 14 (60.9%) | 57 (57.0%) |
| | Moderate | 30 (39.0%) | 8 (34.8%) | 38 (38.0%) |
| | Severe | 4 (5.2%) | 1 (4.3%) | 5 (5.0%) |
| Bruising | Mild | 55 (74.3%) | 15 (93.8%) | 70 (77.8%) |
| | Moderate | 17 (23.0%) | 1 (6.3%) | 18 (20.0%) |
| | Severe | 2 (2.7%) | 0 | 2 (2.2%) |
| Itching | Mild | 15 (88.2%) | 2 (100%) | 17 (89.5%) |
| | Moderate | 2 (11.8%) | 0 | 2 (10.5%) |
| | Severe | 0 | 0 | 0 |
| Discoloration | Mild | 24 (75.0%) | 6 (100%) | 30 (78.9%) |
| | Moderate | 8 (25.0%) | 0 | 8 (21.1%) |
| | Severe | 0 | 0 | 0 |
Abbreviation: CTR = Common Treatment Response. All percentages are based on the number of subjects with the specific CTR by injection by subgroup in the population.
Maximum severity (Severe>Moderate>Mild) reported by the subject for the particular CTR in the diary for at least one day.
**Table 11: CTRs by total duration – initial treatment – Safety population**
| Total Duration (days) | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| CTR after initial injection | Group | Subjects Experiencing the CTR | 1-3 | 4-7 | 8-14 | 15-30 | Last Day Diary |
| Any CTR | RHA4 | 128 (90.8%) | 104 (73.8%) | 84 (59.6%) | 55 (39.0%) | 41 (29.1%) | 18 (12.8%) |
| | Control | 44 (95.7%) | 33 (71.7%) | 25 (54.3%) | 20 (43.5%) | 14 (30.4%) | 4 (8.7%) |
| | Total | 172 (92.0%) | 137 (73.3%) | 109 (58.3%) | 75 (40.1%) | 55 (29.4%) | 22 (11.8%) |
| Redness | RHA4 | 61 (43.3%) | 41 (29.1%) | 11 (7.8%) | 4 (2.8%) | 5 (3.5%) | 2 (1.4%) |
| | Control | 17 (37.0%) | 11 (23.9%) | 5 (10.9%) | 0 | 1 (2.2%) | 0 |
| | Total | 78 (41.7%) | 52 (27.8%) | 16 (8.6%) | 4 (2.1%) | 6 (3.2%) | 2 (1.1%) |
| Pain | RHA4 | 70 (49.6%) | 39 (27.7%) | 17 (12.1%) | 8 (5.7%) | 6 (4.3%) | 1 (0.7%) |
| | Control | 20 (43.5%) | 9 (19.6%) | 5 (10.9%) | 5 (10.9%) | 1 (2.2%) | 0 |
| | Total | 90 (48.1%) | 48 (25.7%) | 22 (11.8%) | 13 (7.0%) | 7 (3.7%) | 1 (0.5%) |
| Tenderness | RHA4 | 113 (80.1%) | 34 (24.1%) | 39 (27.7%) | 24 (17.0%) | 16 (11.3%) | 2 (1.4%) |
| | Control | 39 (84.8%) | 10 (21.7%) | 15 (32.6%) | 9 (19.6%) | 5 (10.9%) | 1 (2.2%) |
| | Total | 152 (81.3%) | 44 (23.5%) | 54 (28.9%) | 33 (17.6%) | 21 (11.2%) | 3 (1.6%) |
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 18
{18}
| **Total Duration (days)** | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| **CTR after initial injection** | **Group** | **Subjects Experiencing the CTR** | **1-3** | **4-7** | **8-14** | **15-30** | **Last Day Diary** |
| Firmness | RHA4 | 108 (76.6%) | 34 (24.1%) | 35 (24.8%) | 19 (13.5%) | 20 (14.2%) | 6 (4.3%) |
| | Control | 34 (73.9%) | 10 (21.7%) | 9 (19.6%) | 5 (10.9%) | 10 (21.7%) | 2 (4.3%) |
| | Total | 142 (75.9%) | 44 (23.5%) | 44 (23.5%) | 24 (12.8%) | 30 (16.0%) | 8 (4.3%) |
| Swelling | RHA4 | 98 (69.5%) | 47 (33.3%) | 32 (22.7%) | 13 (9.2%) | 6 (4.3%) | 1 (0.7%) |
| | Control | 28 (60.9%) | 16 (34.8%) | 8 (17.4%) | 1 (2.2%) | 3 (6.5%) | 0 |
| | Total | 126 (67.4%) | 63 (33.7%) | 40 (21.4%) | 14 (7.5%) | 9 (4.8%) | 1 (0.5%) |
| Lumps/ Bumps | RHA4 | 77 (54.6%) | 30 (21.3%) | 13 (9.2%) | 12 (8.5%) | 22 (15.6%) | 15 (10.6%) |
| | Control | 23 (50.0%) | 7 (15.2%) | 5 (10.9%) | 4 (8.7%) | 7 (15.2%) | 1 (2.2%) |
| | Total | 100 (53.5%) | 37 (19.8%) | 18 (9.6%) | 16 (8.6%) | 29 (15.5%) | 16 (8.6%) |
| Bruising | RHA4 | 74 (52.5%) | 19 (13.5%) | 28 (19.9%) | 17 (12.1%) | 10 (7.1%) | 3 (2.1%) |
| | Control | 16 (34.8%) | 5 (10.9%) | 6 (13.0%) | 5 (10.9%) | 0 | 0 |
| | Total | 90 (48.1%) | 24 (12.8%) | 34 (18.2%) | 22 (11.8%) | 10 (5.3%) | 3 (1.6%) |
| Itching | RHA4 | 17 (12.1%) | 7 (5.0%) | 3 (2.1%) | 5 (3.5%) | 2 (1.4%) | 1 (0.7%) |
| | Control | 2 (4.3%) | 2 (4.3%) | 0 | 0 | 0 | 0 |
| | Total | 19 (10.2%) | 9 (4.8%) | 3 (1.6%) | 5 (2.7%) | 2 (1.1%) | 1 (0.5%) |
| Discoloration | RHA4 | 32 (22.7%) | 17 (12.1%) | 5 (3.5%) | 6 (4.3%) | 4 (2.8%) | 4 (2.8%) |
| | Control | 6 (13.0%) | 3 (6.5%) | 3 (6.5%) | 0 | 0 | 0 |
| | Total | 38 (20.3%) | 20 (10.7%) | 8 (4.3%) | 6 (3.2%) | 4 (2.1%) | 4 (2.1%) |
Abbreviation: CTR = Common Treatment Response.
All percentages are based on the number of CTR diaries returned by injection by subgroup in the population.
152 subjects were treated with initial injection in RHA®4 group and 141 CTR diaries were returned. 49 subjects were treated with initial injection in Control group and 46 CTR diaries were returned.
Total duration is defined as the number of days +1 between the first and last occurrence of the CTR in the diary, no matter if there was a gap between the several occurrences.
The proportion of diaries having at least 1 CTR were similar after touch-up: 79.8% (75/94) in RHA®4 group against 77.5% (31/40) in the control group. Similarly to initial treatment, 98.8% (83/84) of the diaries contained only CTRs of mild or moderate severity for the RHA®4 group against 90.3% (28/31) in the control group.
CTRs were stratified by Fitzpatrick Skin Type (FST). All FST were appropriately represented with 71% and 29% of subjects with FST I-III and IV-VI, respectively. In the RHA®4 group, 96.0% (97/101) and 77.5% (31/40) of subjects with FST I-III and FST IV-VI, respectively experienced at least 1 CTR. In the control group, 100% (33/33) and 76.9% (10/13) of subjects with FST I-III and FST IV-VI, respectively experienced at least 1 CTR. Tenderness and firmness were the most common CTRs in both FST subgroups in the RHA®4 and control groups.
At initial injection, the incidence rate of CTRs was also stratified by volume. 65.8% (100/152) of subjects in RHA®4 group and 83.7% (41/49) in the control group received ≤3 mL. 34.2% (52/152) of subjects in RHA®4 group and 16.3% (8/49) in the control group received >3 mL. Based on the
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 19
{19}
number of returned diaries, 85.9% (79/92) of subjects in RHA®4 who received ≤3 mL experienced at least 1 CTR against 100% (49/49) for those who received >3 mL. In the control group, 94.7% (36/38) of subjects experienced at least 1 CTR if they received ≤3 mL and 100% (8/8) if they received >3 mL.
When injecting using the multilayering technique (i.e. injection at both depths: subcutaneous and supraperiosteal), the incidence rate of subjects experiencing at least 1 CTR was 94.3% (99/105) in RHA®4 group and 100% (33/33) in the control group. When compared to injections performed at a unique depth (subcutaneous only or supraperiosteal only), no conclusion could be made on the impact of injection depth on CTR occurrence.
After initial injection, in the RHA®4 group, 100% (19/19) of subjects reported a CTR if the administration used needle only, 82.2% (60/73) reported a CTR after cannula injection only, and 100% (49/49) reported a CTR after needle + cannula injection. Rates were similar in the control group with, 100% (7/7) of subjects reported a CTR if the administration used needle only, 93.8% (16/17) reported a CTR after cannula injection only, and 95.7% (23/25) reported a CTR after needle + cannula injection. Overall, the cannula only administration of RHA®4 had the lowest incidence of CTRs out of all administration methods and groups for all 3 injections; initial injection (82.2%, 60/73), touch-up (60.5%, 23/38) and retreatment (74.6%, 47/63).
The TI reviewed all CTRs to ensure they were elevated as appropriate to the status of an AE. CTRs were not considered AEs unless the duration and/or severity were in excess of that typically observed following injection of a dermal filler and were clinically significant as determined by the TI. However, CTRs that were noted on the last day of the CTR diary were recorded automatically as AEs regardless of their severity (30-day rule). Overall, for CTRs that were automatically elevated to the level of an AE after 30 days, the TI determined that all the AEs were of “mild” intensity.
### **Adverse Device Effects that occurred in the PMA clinical study**
All Adverse Device Effects (Table 12) were the types and frequency of adverse events that are typically experienced following the injection of a dermal filler into the cheeks.
From Visit 1 to Visit 7 (i.e., not including retreatment injections), 84 ADEs were reported (Table 12):
- 23.7% (36/152) subjects (67 events) in the RHA®4 group
- 16.3% (8/49) subjects (17 events) in the control group
Overall, most ADEs were obtained from the diary. In the RHA®4 group (from V1 to V7):
- 62.7% (42/67) of the ADEs were CTRs automatically elevated to AEs because they were present on the last day of the diary.
- 16.4% (11/67) of the ADEs were automatically elevated to AEs because they were reported as “Other” in the diary.
- 12.0% (8/67) of the ADEs were reported from a pre-identified list of AEs in the subject diary but were not a CTR. Out of these 8 ADEs, none were related to visual disturbances.
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 20
{20}
• 9.0% (6/67) of the ADEs were identified by the TI during a visit.
**Table 12: Adverse Events profile overview (initial, touch-up) Visit 1 to Visit 7 – Safety population**
| From V1 to V7 | RHA^{®}4 (N=152) | | Control (N=49) | | Total (N=201) | |
| --- | --- | --- | --- | --- | --- | --- |
| | Number of events | Number of subjects (%) | Number of events | Number of subjects (%) | Number of events | Number of subjects (%) |
| **Overall** | | | | | | |
| Number of subjects | | 152 | | 49 | | 201 |
| All AEs | 136 | 68 (44.7%) | 43 | 19 (38.8%) | 179 | 87 (43.3%) |
| AESIs | 10 | 5 (3.3%) | 0 | 0 | 10 | 5 (2.5%) |
| ADEs | 67 | 36 (23.7%) | 17 | 8 (16.3%) | 84 | 44 (21.9%) |
| UADEs | 0 | 0 | 0 | 0 | 0 | 0 |
| Serious AEs | 1 | 1 (0.7%) | 2 | 2 (4.1%) | 3 | 3 (1.5%) |
| Device-related Serious AEs | 0 | 0 | 0 | 0 | 0 | 0 |
Abbreviations: ADE = adverse device effect; AE = adverse event; AESI = adverse event of special interest; N = number of subjects; SAE = serious adverse event; SAFT = safety (population); UADE = unanticipated adverse device effect
Notes: Number (%) of subjects with at least 1 AE in the category. All percentages are based on the number of subjects by group in the population. An AESI is any new vision disturbances including, but not limited to, any loss of vision, blurriness, double vision, pain in or around the eye, blindness, blind spot or shadow in the visual field, trouble moving eyes, ocular hypotonia, ptosis, etc. that occurs after any administration of the filler. A related AE is an AE that was judged by the investigator to be Possibly Related or Probably Related or Causal relationship to study device or study procedure. AEs with a missing relationship were counted as Related. A device-related SAE is a related AE also considered as serious.
The type of ADEs were also comparable between RHA$^{®}$4 and the control groups as shown by Table 13.
Similar safety profiles were observed between treatment groups: a total of 23.7% (36/152) and 16.3% (8/49) subjects in RHA$^{®}$4 and control groups respectively, experienced at least 1 ADE between Visit 1 and Visit 7.
The most common events were:
- Injection site mass, experienced in 11.2% (17/152) and 6.1% (3/49) subjects in the RHA$^{®}$4 and control groups, respectively.
- Injection site induration, experienced in 5.3% (8/152) and 10.2% (5/49) subjects in the RHA$^{®}$4 and control groups, respectively.
- Injection site pain, experienced in 3.3% (5/152) and 2.0% (2/49) subjects in the RHA$^{®}$4 and control groups, respectively.
There were no noticeable differences of the frequencies of ADEs between the RHA$^{®}$4 and control groups.
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 21
{21}
**Table 13: ADEs sorted by SOC and PT (Visit 1 to Visit 7 – initial treatment to 52 weeks follow-up excluding retreatment) – Safety population**
| From V1 to V7 SOC PT | RHA^{®}4 (N=152) | | Control (N=49) | | Total (N=201) | |
| --- | --- | --- | --- | --- | --- | --- |
| | Number of events | Number of subjects | Number of events | Number of subjects | Number of events | Number of subjects |
| **Any ADE** | **67** | **36 (23.7%)** | **17** | **8 (16.3%)** | **84** | **44 (21.9%)** |
| **General Disorders and Admin. Site Conditions** | **55** | **32 (21.1%)** | **13** | **7 (14.3%)** | **68** | **39 (19.4%)** |
| Injection Site Mass | 18 | 17 (11.2%) | 4 | 3 (6.1%) | 22 | 20 (10.0%) |
| Injection Site Induration | 8 | 8 (5.3%) | 5 | 5 (10.2%) | 13 | 13 (6.5%) |
| Injection Site Pain | 6 | 5 (3.3%) | 1 | 1 (2.0%) | 7 | 6 (3.0%) |
| Injection Site Discoloration | 5 | 5 (3.3%) | 0 | 0 | 5 | 5 (2.5%) |
| Injection Site Bruising | 3 | 3 (2.0%) | 0 | 0 | 3 | 3 (1.5%) |
| Injection Site Swelling | 2 | 2 (1.3%) | 1 | 1 (2.0%) | 3 | 3 (1.5%) |
| Injection Site Warmth | 2 | 2 (1.3%) | 1 | 1 (2.0%) | 3 | 3 (1.5%) |
| Injection Site Hypoaesthesia | 1 | 1 (0.7%) | 1 | 1 (2.0%) | 2 | 2 (1.0%) |
| Mass | 2 | 2 (1.3%) | 0 | 0 | 2 | 2 (1.0%) |
| Administration Site Acne | 1 | 1 (0.7%) | 0 | 0 | 1 | 1 (0.5%) |
| Injection Site Deformation | 3 | 1 (0.7%) | 0 | 0 | 3 | 1 (0.5%) |
| Injection Site Discomfort | 1 | 1 (0.7%) | 0 | 0 | 1 | 1 (0.5%) |
| Injection Site Erythema | 1 | 1 (0.7%) | 0 | 0 | 1 | 1 (0.5%) |
| Injection Site Pruritus | 1 | 1 (0.7%) | 0 | 0 | 1 | 1 (0.5%) |
| Swelling Face | 1 | 1 (0.7%) | 0 | 0 | 1 | 1 (0.5%) |
| Injection Site Discolouration | 1 | 1 (0.7%) | 0 | 0 | 1 | 1 (0.5%) |
| Injection Site Haematoma | 1 | 1 (0.7%) | 0 | 0 | 1 | 1 (0.5%) |
| Injection Site Vesicles | 0 | 0 | 2 | 1 (2.0%) | 2 | 1 (0.5%) |
| Mass | 1 | 1 (0.7%) | 0 | 0 | 1 | 1 (0.5%) |
| Swelling Face | 0 | 0 | 1 | 1 (2.0%) | 1 | 1 (0.5%) |
| **Nervous System Disorders** | **6** | **5 (3.3%)** | **2** | **2 (4.1%)** | **8** | **7 (3.5%)** |
| Headache | 6 | 5 (3.3%) | 2 | 2 (4.1%) | 8 | 7 (3.5%) |
| **Eye Disorders** | **4** | **2 (1.3%)** | **0** | **0** | **4** | **2 (1.0%)** |
| Periorbital pain | 2 | 2 (1.3%) | 0 | 0 | 2 | 2 (1.0%) |
| Eye pain | 1 | 1 (0.7%) | 0 | 0 | 1 | 1 (0.5%) |
| Ocular Hyperaemia | 1 | 1 (0.7%) | 0 | 0 | 1 | 1 (0.5%) |
| **Skin and Subcutaneous Tissue Disorders** | **2** | **2 (1.3%)** | **0** | **0** | **2** | **2 (1.0%)** |
| Pruritus | 1 | 1 (0.7%) | 0 | 0 | 1 | 1 (0.5%) |
| Urticaria | 1 | 1 (0.7%) | 0 | 0 | 1 | 1 (0.5%) |
| **Musculoskeletal And Connective Tissue Disorders** | **0** | **0** | **2** | **1 (2.0%)** | **2** | **1 (0.5%)** |
| Masticatory pain | 0 | 0 | 2 | 1 (2.0%) | 2 | 1 (0.5%) |
Abbreviations: ADE = adverse device effect; MedDRA = Medical Dictionary for Regulatory Activities; N = number of subjects; PT = preferred term; SAFT = safety (population); SOC = system organ class; V = visit
All percentages were based on the number of subjects by group in the population. ADEs were coded using MedDRA version 24.0. AEs are displayed by descending subject frequency of SOC, then descending subject frequency of PT within SOC in the Total column, then alphabetically.
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 22
{22}
The majority 82.0% (55/67) of ADEs in the RHA®4 group were mild and the remaining were moderate 17.9% (12/67) (Table 14). In the control group, the severity was similar: 76.5% (13/17) were mild and 23.5% (4/17) were moderate. There were no severe ADEs reported in both treatment groups.
Of the 67 ADEs reported between Visit 1 and Visit 7 for the RHA®4 group (after initial and touch-up treatment), 46.3% (31/67) ADEs lasted 1 to 3 days, 82.1% (55/67) ADEs were resolved within 14 days. There was 1 event of injection site mass in the RHA®4 group which lasted more than 90 days. It was mild in severity and resolved without treatment and sequelae. Overall, most ADEs resolved within 30 days and the proportion of subjects with reported ADEs was similar across the 2 treatment groups.
After retreatment, most ADEs also resolved within 30 days. They were all resolved in less than 90 and all ADEs resolved at the time of study exit.
**Table 14: ADEs sorted by severity (Visit 1 to Visit 7 – initial treatment to 52 weeks follow-up excluding retreatment) – Safety population**
| From V1 to V7 Severity | RHA®4 (N=152) | | Control (N=49) | | Total (N=201) | |
| --- | --- | --- | --- | --- | --- | --- |
| | Number of events | Number of subjects | Number of events | Number of subjects | Number of events | Number of subjects |
| **Any ADE** | **67** | **36 (23.7%)** | **17** | **8 (16.3%)** | **84** | **44 (21.9%)** |
| Mild | 55 | 32 (21.1%) | 13 | 8 (16.3%) | 68 | 40 (19.9%) |
| Moderate | 12 | 9 (5.9%) | 4 | 3 (6.1%) | 16 | 12 (6.0%) |
| Severe | 0 | 0 | 0 | 0 | 0 | 0 |
Abbreviations: ADE = adverse device effect.
All percentages are based on the number of subjects by group in the Safety population.
ADEs with a missing severity were counted as Severe.
**Table 15: ADEs sorted by duration (Visit 1 to Visit 7 – initial treatment to 52 weeks follow-up excluding retreatment) – Safety population**
| From Visit 1 to Visit 7 | Duration of ADE Characteristics (days) | Duration of AE (days) | RHA®4 (N=152) | | Control (N=49) | | Total (N=201) | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Number of events | Number of subjects | Number of events | Number of subjects | Number of events | Number of Subjects |
| **Any ADE** | | | **67** | **36 (23.7%)** | **17** | **8 (16.3%)** | **84** | **44 (21.9%)** |
| | 1-3 | | 31 | 19 (12.5%) | 6 | 4 (8.2%) | 37 | 23 (11.4%) |
| | 4-7 | | 10 | 9 (5.9%) | 3 | 3 (6.1%) | 13 | 12 (6.0%) |
| | 8-14 | | 14 | 7 (4.6%) | 3 | 2 (4.1%) | 17 | 9 (4.5%) |
| | 15 - 30 | | 7 | 6 (3.9%) | 4 | 3 (6.1%) | 11 | 9 (4.5%) |
| | 31 - 90 | | 4 | 4 (2.6%) | 1 | 1 (2.0%) | 5 | 5 (2.5%) |
| | >90 | | 1 | 1 (0.7%) | 0 | 0 | 1 | 1 (0.5%) |
| N (missing) | | | 67 (0) | | 17 (0) | | 84 (0) | |
| Mean (SD) | | | 12.4 (28.76) | | 9.5 (9.06) | | 11.8 (25.98) | |
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 23
{23}
| From Visit 1 to Visit 7 | | RHA®4 (N= 152) | | Control (N= 49) | | Total (N= 201) | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Duration of ADE Characteristics (days) | Duration of AE (days) | Number of events | Number of subjects | Number of events | Number of subjects | Number of events | Number of Subjects |
| 95% CI | | 5.3-19.4 | | 4.9-14.2 | | 6.1-17.4 | |
| Mean | | | | | | | |
| Min to Max | | 1 to 227 | | 1 to 36 | | 1 to 227 | |
Abbreviations: ADE = adverse device effect. CI = Confidence Interval; Max = Maximum; Min = Minimum; SD = Standard Deviation.
All percentages are based on the number of subjects by group in the Safety population.
For ongoing ADEs, missing end date is assigned to date of last contact to derive current duration of the event.
At time of this analysis, 1 ADE was ongoing in the category 31 - 90 days.
All Fitzpatrick Skin Types (FST) were appropriately represented, with 71.1% (143/201) and 28.9% (58/201) of subjects being of skin types I-III and types IV-VI, respectively. Proportion of subjects being FST I-III vs IV-VI were similar within treatment groups:
- FST I-III: 70.4% (107/152) and 73.5% (36/49) of subjects in the RHA®4 and control groups, respectively
- FST IV-VI: 29.6% (45/152) and 26.5% (13/49) of subjects in the RHA®4 and control groups, respectively
In RHA®4 treatment group, 22.4% (24/107) and 26.7% (12/45) subjects with FST I-III and with FST IV-VI, respectively, experienced at least 1 ADE. In the control group, 11.1% (4/36) and 30.8% (4/13) subjects with FST I-III and with FST IV-VI, respectively, experienced at least 1 ADE. There were no observed differences in rates of ADEs across FST groups. The ADEs profiles were similar. RHA®4 and control treatment groups present a similar profile. Similar results were found following retreatment.
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 24
{24}
**Table 16: ADE profile overview by Fitzpatrick Skin Type (Visit 1 to Visit 7 – initial treatment to 52 weeks follow-up excluding retreatment) – Safety population**
| FST | From V1 to V7 | RHA^{®}4 (N= 152) | | Control (N= 49) | | Total (N= 201) | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Number of events | Number of subjects | Number of events | Number of subjects | Number of events | Number of subjects |
| **I-III** | Number of subjects | | 107 | | 36 | | 143 |
| | Any ADE | 45 | 24 (22.4%) | 7 | 4 (11.1%) | 52 | 28 (19.6%) |
| **IV-VI** | Number of subjects | | 45 | | 13 | | 58 |
| | Any ADE | 22 | 12 (26.7%) | 10 | 4 (30.8%) | 32 | 16 (27.6%) |
| | **p-value** | **0.6764** | | | | | |
Abbreviations: ADE = adverse device effect. N = number of subjects; SAFT = safety (population)
Notes: Number (%) of subjects with at least 1 ADE in the subgroup. All percentages were based on the number of subjects by subgroup in the population. P-values were based on a Fisher exact test, 2-sided at 5% to test independence or not of the subgroup. Only subgroups with at least 10 subjects from the RHA$^{®}$4 initial treatment group were analyzed.
All ethnicities were appropriately represented, with 19.9% (40/201) and 79.1% (159/201) of subjects being Hispanic and Non-Hispanic or Latino, respectively. There were no observed differences in rates of ADEs across ethnicity groups. RHA$^{®}$4 and control treatment groups present a similar profile. Similar results were found following retreatment.
Sex was appropriately represented, with 10.4% (21/201) and 89.6% (180/201) of male and female subjects, respectively. There were no observed differences in rates of ADEs across sex. Similar results were found following retreatment.
The median age was 56 years old with 52.2% (105/201) of subjects aged 22 to 56 years old and 47.8% (96/201) of subjects older than 56 years old. In the RHA$^{®}$4 group, 19.8% (16/81) of subjects 56 years old or younger experienced an ADE against 28.2% (20/71) of subjects older than 56 years old. Subjects in the control group had a similar incidence rate of ADEs as subjects of the same age in the RHA$^{®}$4 group. Similar results were found following retreatment.
The number of subjects who experienced ADEs analyzed by subgroup of volume were divided into: ≤3 mL (141 subjects) and/or with >3 to ≤6 mL (115 subjects) and >6 mL (18 subjects). A subject may be counted more than once because a subject may be reported after experiencing an ADE following initial treatment and then again after touch-up (in another bracket of volume). Most subjects were injected with ≤3 mL (141 subjects) and/or with >3 to ≤6 mL (115 subjects). The incidence of ADEs between study groups was similar for subjects injected with these 2 volume ranges. Incidence rates for subjects injected with: ≤3 mL were 36 events in 20/100 subjects (20.0%) in RHA$^{®}$4 group and 12 events in 7/41 subjects (17.1%) in control group; >3 to ≤6 mL were 29 events in 17/90 subjects (18.9%) in RHA$^{®}$4 group and 5 events in 3/25 subjects (12.0%) in control group. Only 18 subjects received >6 mL to ≤9 mL and only 1 of them (in the RHA$^{®}$4 group) developed an ADE, however only 3 subjects in the control group received this volume.
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 25
{25}
Similarly, the depth of injection may be different after initial treatment and touch-up and a subject may experience an ADE after both and be counted twice in different injection depth sub-groups. The analyses with relatively small and disproportionate number of subjects treated per depth of injection (subcutaneous only and supraperiosteal only) does not allow any meaningful conclusions regarding the incidence rate of ADEs per injection depths. However, for subjects injected with the multilayering technique (both subcutaneously and supraperiosteally), the incidence of ADEs was similar between treatment groups, with 33 (27.0%, 33/122) subjects in the RHA$^{®}$4 group and 7 (17.5%, 7/40) subjects in the Control group.
Subjects were injected using several administration methods: needle only, cannula only or both. The administration method may have been different between initial and touch-up treatments. Because of the high number of combinations of administration methods, the analysis of the ADEs incidence rate by method of administration was no conclusive.
Similarly, the incidence rates of ADEs per injection technique did not provide any meaningful conclusion due to the disproportionate use of some injection techniques and very low numbers in others.
### **Adverse Events of Special Interests (AESI)**
Eleven events of visual disturbances (vision blurred, headache, blepharospasm, photopsia, periorbital pain, ocular hyperemia, eye pain, periorbital pain, blurred vision, pain around the eyes, intermittent floaters in the eye) in five subjects were reported as Adverse Events of Special Interests (AESI). All were in subjects randomly assigned to RHA$^{®}$4 group including one event after retreatment. All events were mild. None were related to device. Four events from two subjects were related to the injection procedure. None of the subjects required treatment or outside interventions, except for one subject who experienced blurred vision which was mild and unlikely related to the study device or procedure, who used lubricating eye daily drops for 3 days. All events resolved without sequelae or the TI considered that no additional follow-up was necessary.
There were no Serious Adverse Events (SAE) that were device related and no Unanticipated Adverse Device Effects (UADE). There were no deaths, and no subjects prematurely withdrew due to an ADE.
### **Pain at injection**
Injection pain during injection (initial, touch-up and retreatment) was measured using a 10cm (100mm) Visual Analog Scale (VAS), and is presented in millimeters (i.e., 100 points within the 10cm VAS). Injection pain was assessed immediately, 5, 15, and 30 minutes after filler injection.
The average level of pain noted during initial study device injections was similar between treatment groups, with $7.2 \pm 17.32$ and $5.5 \pm 14.96$ in the RHA$^{®}$4 and the Control groups, respectively. Injection site pain was reduced to $2.6 \pm 9.01$ mm and $1.9 \pm 7.40$ mm in the RHA$^{®}$4 and the Control groups, respectively by 5 minutes post-injection. Pain after 15 minutes was $1.3 \pm 5.33$ and $0.4 \pm 2.86$ in the RHA$^{®}$4 and the Control groups, respectively. Pain after 30 minutes was $1.1 \pm 5.87$ and $0.0 \pm 0.00$ in the RHA$^{®}$4 and the Control groups, respectively.
Similar findings were observed following touch-up and retreatment injections.
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 26
{26}
## 2. Effectiveness Results
The analysis of effectiveness was based on the per-protocol (PP) population of 131 subjects who received treatment with RHA®4 and 42 who received treatment with the control. A supportive analysis was also performed in the intent-to-treat (ITT) population of 152 subjects who received treatment with RHA®4 and 49 subjects who received treatment with the control.
The primary effectiveness endpoint was to assess the effectiveness (non-inferiority) of RHA®4 versus the control on adding volume in the midface region 8 weeks after the last treatment (initial or touch-up). For the primary endpoint, the change from Baseline for subjects treated with RHA®4 had to be statistically non-inferior to the change from Baseline for subjects treated with the control at 8 weeks after the last treatment as assessed by the Blinded Live Evaluator (BLE) using the TMVDS. The difference in the TMVDS change from Baseline to 8 weeks was used to establish non-inferiority.
As shown in Table 17, subjects treated with RHA®4 had a mean TMVDS change at week 8 from Baseline of -1.3, and subjects treated with Control had a similar TMVDS change of -1.3.
**Table 17: TMVDS Grade Change from Baseline at primary endpoint (week 8)**
| Category (PP Population) | RHA®4 (N=131) | Control (N=42) |
| --- | --- | --- |
| **TMVDS change from Baseline at Visit 3** | | |
| N | 131 | 42 |
| Mean (SD) | -1.3 (0.80) | -1.3 (0.65) |
| 95% CI Mean | -1.5 - -1.2 | -1.5 - -1.1 |
| Min to Max | -3 to 1 | -2 to 0 |
| Bootstrap method | | |
| Estimate of difference in means RHA4 – Control | 0.02 | |
| 95% CI for difference in means | -0.22 – 0.25 | |
| Category (ITT Population*) | RHA®4 (N=152) | Control (N=49) |
| **TMVDS change from Baseline at Visit 3** | | |
| N | 152 | 49 |
| Mean (SD) | -1.3 (0.75) | -1.3 (0.61) |
| 95% CI Mean | -1.4 - -1.2 | -1.5 - -1.1 |
| Min to Max | -3.0 - -1.0 | -2.0 - -1.0 |
| Bootstrap method | | |
| Estimate of difference in means RHA4 – Control | 0.01 | |
| 95% CI for difference in means | -0.20-0.22 | |
Abbreviations: ANCOVA = analysis of covariance; BLE = blinded live evaluator; CI = confidence interval; n = number of observations; N = number of subjects; PP = per population; SD = standard deviation; TMVDS = Teoxane Midface Volume Deficit Scale
Baseline is defined as the last non missing measurement preceding the initial treatment.
Difference in means was calculated by Bootstrap estimate using 1,000 samples. 95% CI was based on the percentile interval. A non-parametric model was used as all assumptions needed to use an ANCOVA model were not met.
*In the ITT population, missing values were imputed using mean imputation before bootstrapping. 19 subjects (15 RHA4, 4 Control) were missing primary endpoint data in the ITT population.
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 27
{27}
When repeated with the ITT population, the TMVDS change from Baseline assessed by BLE at 8 weeks was similar to that of PP population in RHA$^{®}$4 and Control groups, respectively, with -1.3 in both treatment groups.
Table 18 presents the TMVDS responder rate as assessed by BLE at week 8. In RHA$^{®}$4 group, 114 (87.0%) subjects were considered as responders at week 8, whereas 38 (90.5%) subjects responded to Control treatment. The co-primary endpoint would support non-inferiority if the Control device responder rate was $\geq 70\%$. The threshold was chosen at $\geq 70\%$ to make sure the Control device was an effective comparator, to determine the good effectiveness of the tested device (i.e., RHA$^{®}$4). This assures that RHA$^{®}$4 would be non-inferior to an effective control device at 8 weeks after last treatment. More than 90% of subjects treated with the Control device were responders, the co-primary was therefore met.
**Table 18: Responder rate (BLE) at primary endpoint (Week 8) – PP population**
| Responders | RHA^{®}4 (N=131) | Control (N=42) |
| --- | --- | --- |
| N | 131 | 42 |
| Responder^{a} [95% CI] | 114 (87.0%) [80.2 – 91.7%] | 38 (90.5%) [77.9 – 96.2%] |
| Not Responder | 17 (13.0%) | 4 (9.5%) |
| Difference in responder rate (RHA4 – Control) [95% CI] | -3.5% [-12.4 – 9.9%] | |
Abbreviations: BLE = blinded live evaluator; CI = confidence interval; PP = per population; N = number of subjects; TMVDS = Teoxane Midface Volume Deficit Scale
Baseline was defined as the last non-missing measurement preceding the initial treatment. To successfully achieve the co-primary endpoint, the proportion of responders with a $\geq 1$ grade point on the TMVDS at Visit 3 when compared to pretreatment for the control device should be $\geq 70\%$. The 95% CI for responder and for difference in responder rate were obtained using Wilson method.
All percentages are based upon the total number of subjects by group with non-missing data for the TMVDS at Week 8 (Visit 3).
a: Responder: Improvement from Baseline of $\geq 1$ grade TMVDS
When repeated with the ITT population, the responder rate of the Control was $\geq 70\%$ at 8 weeks after last treatment confirming the criteria of the co-primary endpoint.
At 8 weeks after last treatment, as the upper bound of the confidence intervals of the TMVDS of the Change from Baseline of RHA$^{®}$4 was $< 0.5$ and the lower bound of the responder rate of the Control was $\geq 70\%$ (PP and ITT), treatment with RHA$^{®}$4 is non-inferior to treatment with Control.
RHA$^{®}$4 demonstrated non-inferiority to the Control device based on TMVDS change from Baseline at 8 weeks after last treatment in both the PP and ITT analysis populations.
All secondary endpoints were performed using ITT population as observed (no imputation).
PMA P170002/S048: FDA Summary of Safety and Effectiveness Data
Page 28
{28}
**Figure 1: TMVDS rate of responders (≥1-grade difference from Baseline), as assessed by the Blinded Live Evaluator, thought the follow-up period – Observed ITT population**

*Descriptive only – The 95% confidence intervals presented have not been adjusted for multiple comparisons.*
The responder rate of RHA$^{®}$4 was similar to the responder rate of the Control throughout the course of the study (52 weeks after last treatment). Overall, we observed an expected decrease in the change from Baseline in TMVDS over time on both treatment groups. Results showed that subjects injected with RHA$^{®}$4 were able to sustain volume augmentation over time, as the volumizing effect was still visible in approximately 65% of subjects at 1-year post treatment. To achieve this effect, 75% (114/152) of subjects were injected (Table 8) with the multilayering technique (subcutaneous and supraperiosteal layers) for their initial treatment, and 64.6% (62/96) for their touch-up treatment.
The analysis grade (Table 19) change from Baseline showed th…