MagtraceTM and Sentimag(R) Magnetic Locatization System
P160053S002 · Endomagnetics Ltd., · PUV · Dec 6, 2022 · General, Plastic Surgery
Device Facts
| Record ID | P160053S002 |
| Device Name | MagtraceTM and Sentimag(R) Magnetic Locatization System |
| Applicant | Endomagnetics Ltd., |
| Product Code | PUV · General, Plastic Surgery |
| Decision Date | Dec 6, 2022 |
| Decision | APPR |
| Device Class | Class 3 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|
| P160053S002 · Dec 6, 2022 | MagtraceTM and Sentimag(R) Magnetic Locatization System | Endomagnetics Ltd., | Published European clinical studies; PostMag Registry (ISRCTN85167182); Institutional clinical data/reports | Real-world data from multiple European clinical studies and registries were used to provide supplemental evidence on the safety and effectiveness of the Magtrace/Sentimag system, specifically regarding detection rates, concordance with standard of care, and the incidence of long-term side effects like skin staining and MRI artifacts. | Retrospective cohort; Registry data; Published clinical literature; Post-market surveillance; Skin staining; MRI artifact |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|
| Summary of Published European Studies (Meta-analysis); Multi-center clinical studies/observational cohorts; Follow-up/Duration: Variable | Breast cancer patients undergoing sentinel lymph node biopsy; Sample Size: Total across 7 studies: ~1,000+ patients; Number of Sites: Multiple European centers | Radioisotope and/or Blue Dye | Sentinel lymph node detection rate, concordance |
| PostMag Registry (ISRCTN85167182); Registry / Observational; Follow-up/Duration: Up to 5 months | Breast cancer patients; Sample Size: 11 | Not applicable for this study | MRI artifact occurrence and resolution |
Indications for Use
The Magtrace® and Sentimag® Magnetic Localization System is indicated to assist in localizing lymph nodes draining a tumor site, as part of a sentinel lymph node biopsy procedure, in patients with breast cancer undergoing a mastectomy or lumpectomy. For patients undergoing lumpectomy, nipple sparing, nipple areolar sparing or skin sparing procedures, Magtrace is indicated to be injected only peritumorally.
Device Story
System consists of Magtrace (superparamagnetic iron oxide tracer) and Sentimag (hand-held magnetic probe/base unit). Surgeon injects Magtrace interstitially into breast; particles migrate to sentinel lymph nodes. Surgeon uses Sentimag probe to detect magnetic signal from nodes; probe generates electrical current from magnetic response; base unit converts signal to audible/visual feedback. Used in OR; operated by surgeon. Output guides node excision; replaces/complements radioisotope/blue dye. Benefits: no ionizing radiation; flexible injection timing; no special nuclear medicine handling; avoids anaphylaxis/necrosis risks of blue dye. Risks: MRI artifacts; skin staining (tattooing); rare anaphylaxis/bradycardia.
Clinical Evidence
Pivotal prospective, open-label, multi-center paired comparison (n=160, 147 evaluable). Primary endpoint: lymph node detection rate. Magtrace detection rate 94.3% vs. 93.5% for control (95% CI lower bound -2.1%, meeting non-inferiority margin of 5%). Concordance 98.0%. Safety profile similar to standard of care; adverse events included breast discoloration (16.3%) and ecchymosis (6.8%).
Technological Characteristics
Magtrace: aqueous suspension of carboxydextran-coated superparamagnetic iron oxide particles (3.5-10 nm core, 45-65 nm total diameter). Sentimag: susceptometer with hand-held probe (pick-up coils) and base unit. Energy: alternating magnetic field. Connectivity: standalone. Sterilization: aseptic filling (Magtrace). Software: signal processing for audible/visual feedback.
Indications for Use
Indicated for patients with breast cancer undergoing mastectomy or lumpectomy requiring sentinel lymph node biopsy. Contraindicated in patients with hypersensitivity to iron oxide or dextran, iron overload disease, metal implants in the axilla or chest, or those requiring post-lumpectomy breast MRI or SLNB before neoadjuvant chemotherapy where MRI is the primary monitoring tool.
Regulatory Classification
Identification
Locating lymph nodes draining a tumor as part of the sentinel lymph node biopsy procedure
Reference Devices
- Sienna+ (earlier variant of Magtrace)
- Technetium 99m sulfur colloid (standard of care)
- Isosulfan blue dye (standard of care)
- Lymphoseek (technetium Tc 99m tilmanocept)
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
# I. GENERAL INFORMATION
Device Generic Name: Magnetic Sentinel Node Detection System
Device Trade Name: Magtrace® and Sentimag® Magnetic Localization System
Device Procode: PUV
Applicant's Name and Address: Endomagnetics Ltd.
330 Cambridge Science Park
Milton Road, Cambridge, CB4 0WN, UK
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P160053/S002
Date of FDA Notice of Approval: December/06, 2022
The original PMA P160053 was approved on July 24, 2018 and is indicated to assist in localizing lymph nodes draining a tumor site, as part of a sentinel lymph node biopsy procedure, in patients with breast cancer undergoing a mastectomy. The SSED to support the indication is available on the CDRH website and is incorporated by reference here. The current supplement was submitted to expand the indication for the device to assist in localizing lymph nodes draining a tumor site, as part of a sentinel lymph node biopsy procedure, in patients with breast cancer.
# II. INDICATIONS FOR USE
The Magtrace® and Sentimag® Magnetic Localization System is indicated to assist in localizing lymph nodes draining a tumor site, as part of a sentinel lymph node biopsy procedure, in patients with breast cancer undergoing a mastectomy or lumpectomy. For patients undergoing lumpectomy, nipple sparing, nipple areolar sparing or skin sparing procedures, Magtrace is indicated to be injected only peritumorally.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 1
{1}
Magtrace® is intended and calibrated for use ONLY with the Sentimag® system.
### III. CONTRAINDICATIONS
- Known hypersensitivity to iron oxide or dextran compounds.
- Iron overload disease
- A metal implant in the axilla or in the chest.
- SLNB before neoadjuvant chemotherapy (NAC) where magnetic resonance imaging (MRI) will be the primary imaging used for monitoring the progress of NAC.
- Patients identified in advance to require post-lumpectomy imaging with breast MRI.
### IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the Magtrace® and Sentimag® Magnetic Localization System labeling.
### V. DEVICE DESCRIPTION
The Magtrace® and Sentimag® Magnetic Localization System is indicated to assist in localizing lymph nodes draining a tumor site, as part of a sentinel lymph node biopsy procedure, in patients with breast cancer undergoing a mastectomy and consists of:
- The Magtrace®
- The Sentimag®
Magtrace®
Magtrace® is a combination device/drug product consisting of a blackish-brown sterile aqueous suspension of carboxydextran-coated superparamagnetic iron oxide particles in Water for Injection (WFI) containing 0.3% w/v sodium chloride. Magtrace® is supplied as sterile (aseptically filled) in single-use glass vials containing a minimum of 2.2 ml to allow for a consistent 2.0 ml injection volume.
Each milliliter of Magtrace® contains approximately 28 milligrams of iron in the form of iron oxide. The recommended quantity of Magtrace® administered for use in patients is 2 ml with the equivalent iron content of 55 mg +/- 4 mg per injection.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 2
{2}
Magtrace® key characteristics include:
- Magnetic iron oxide core of 3.5-10 nm in diameter provides detectability by the Sentimag®
- Carboxydextran coating which brings the overall particle diameter to 45-65 nm, keeps the particles in solution, and prevents iron oxide aggregation
- 0.3% saline provides tonicity and allows uptake of the particles into the lymphatic system.
An schematic of the Magtrace® particles is shown below in Figure 1.
Figure 1: Magtrace® Particle Schematic

Sentimag®
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 3
{3}
Sentimag® is a susceptometer designed to deliver a small alternating magnetic field via a hand-held probe, and to electronically detect the presence of magnetic material in the vicinity of the probe head.
Two generations of this device are now available the original Sentimag (Sentimag Gen 2) and the new generation of the device Sentimag Gen 3.
In both cases, detection is performed via pick-up coils in the probe head, which generate electrical current from the magnetic materials response. This response current is passed through the probe cables and connectors to the Sentimag® base unit, where it is transformed into both audible and visual feedback for the surgeon. The base unit also contains the controls for operating the Sentimag® system that are located on the front of the unit, with a power switch on the back.
The main features of the original Sentimag® are:
- Portable base unit that can sit on a flat surface
- Audible and visual indications of magnetic material proximity:
- Magnetic signals indicated by variable pitch (audible), that increases as the probe is brought near Magtrace magnetic tracer material, and yellow LCD digits (visual);
- Extraneous or background signals indicated by low and constant pitch (audible) and red LCD digits (visual);
- Liquid crystal display (LCD) for numerical indication of signal strength and general unit information (e.g., volume, sensitivity setting).
- Choice of three (3) sensitivity settings, controlled by a knob mounted on the base unit
- Volume control knob on the base unit
- Push button mounted on the base unit that activates instrument balancing function that readies the system for measurement
- Detachable air-operated footswitch allowing remote operation of the balance function
- The detachable applied part is the probe assembly comprising a hand-held probe, a flexible cable of just under three (3) meters length, and colour-coded (black and white) connectors to plug the probe into the base unit.
- Applied Probe assembly is to be used in conjunction with a standard single-use sterile sheath (sold separately by OEM suppliers). Sheaths should be latex-free and at least 1 inch wide and 72 inches long.
The features of Sentimag Gen 3 are:
- Portable base unit that can sit on any flat surface
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 4
{4}
- Audible and visual indications of marker proximity
- Audible signal with a variable pitch (frequency) that increases as the probe is brought near the magnetic marker
- Volume control knob on the base unit
- Liquid crystal display (LCD) for numerical indication of signal strength in Counts mode.
- Audible and visual discrimination between magnetic signals and extraneous or background signals
- Magnetic signals indicated by variable pitch (audible) and yellow LCD digits (visual); extraneous or background signals indicated by low and constant pitch (audible) and red LCD digits (visual)
- Instrument-balancing function that readies the system for measurement
- Push button mounted on the base unit activates the balance function
- Two probes are compatible with the Sentimag base unit: a Standard Sentimag probe (18.5mm in diameter) and a Thinner (13.5mm diameter) Sentimag Probe
- Detachable electronic footswitch allows remote operation of the balance function
- The detachable applied part is the probe assembly comprising a hand-held probe, a flexible cable of 9 feet in length, and colour-coded (black and white) connector to plug the probe into the base unit
- Applied probe assembly is to be used in conjunction with a standard single-use sterile sheath (sold separately by OEM suppliers)
Sentimag Gen 3 operates in two distinct modes, Count mode and the Measure mode. In Count mode, the Sentimag Gen 3 detector works in exactly the same way as Sentimag Gen in that it can detect and locate Magtrace® for the tissue marker to be surgically removed with the sentinel node following detection. Measure mode is not available to be used with Magtrace and should not be used to detect Magtrace.
In Sentimag Gen 3 the display of the count has been expanded to 5 digits instead of 4 in the Gen 2 and the scale multiplier has been fixed to the equivalent of sensitivity/scale setting 2 in the Gen 2. The combination of these two features means that the Sensitivity/scale control knob is no longer required in Sentimag Gen 3.
The Base Unit, Probe, and Footswitch for both Sentimag Gen 2 and Gen 3 are shown in Figure 2 below.
Figure 2: Sentimag® Probe and Base Unit
Sentimag (Gen 2)
Sentimag (Gen 3)
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 5
{5}

## **VI. ALTERNATIVE PRACTICES AND PROCEDURES**
There are other alternatives to assist in localizing lymph nodes draining a tumor site, as part of a sentinel lymph node biopsy procedure. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle. The alternatives are Technetium radioisotope-labelled tracers and blue dye tracers.
## **VII. MARKETING HISTORY**
Sentimag®/Sienna+ (an earlier variant of the Magtrace®) has been commercially available in the European Union (EU) since 2013 and was available in the following countries: United Kingdom, Germany, France, Italy, Spain, Portugal, Netherlands, Austria, Czech Republic, Denmark, Croatia, Sweden, Slovakia, Turkey, Poland, Switzerland, Hong Kong, Singapore, New Zealand, and Australia. In 2017 Sienna+ was replaced with Magtrace which is currently available in the following countries: USA, Canada, United Kingdom, Germany, France, Italy, Spain, Portugal, Netherlands, Austria, Czech Republic, Denmark, Croatia, Sweden, Slovakia, Cyprus, Turkey, Israel, Poland, Switzerland, Hong Kong, South Africa, Serbia, Bosnia, Saudi Arabia, Bahrain, Morocco, New Zealand, and Australia. The Sentimag®/Sienna+ product was not withdrawn from any foreign market for any reason relating to the safety and effectiveness of the device. It was superseded by Magtrace which is a more user friendly version of the same device.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 6
{6}
Regarding the differences between Magtrace® and Sienna+, the Sienna+ was designed to be pre-mixed with saline immediately prior to administration, whereas Magtrace® has been formulated to contain 0.3% w/v sodium chloride and does not require premixing with saline. Apart from this addition of sodium chloride, Magtrace® is identical to Sienna+ and is considered acceptable a market history comparison.
### VIII. PROBABLE ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Below is a list of the probable adverse effects (e.g., complications) associated with Magtrace® and Sentimag® Magnetic Localization System include:
Magtrace® is intended for injection into the breast ONLY (interstitial injection).
When similar material to that used in Magtrace® has been injected directly into the bloodstream (intravenously), the following undesirable effects have been reported:
- Common (<2%) – pain at the injection site, vasodilation, paresthesia
- Uncommon (≥0.1% to <1%) – asthenia, back pain, injection site reactions, chest pain, nausea, vomiting, headache, taste changes, itching, rash, inflammatory response (localized redness and swelling) with intradermal injection.
- Rare (≥0.01% to <0.1%) – Hypersensitivity and anaphylaxis, hypertension, phlebitis, hyperesthesia, anxiety, dizziness, convulsion, parosmia, dyspnea, increased cough, rhinitis, eczema, urticaria.
For the specific adverse events that occurred in the clinical studies, please see Section X below.
### IX. SUMMARY OF NONCLINICAL STUDIES
#### A. Laboratory Studies
Table 1- Summary of Laboratory Studies
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 7
{7}
| Test | Purpose | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Biocompatibility- Sienna+/Magtrace® | | | |
| Cytotoxicity EN ISO 10993-5 | In Vitro Cytotoxicity test on Sienna+ | No toxicological or biologically critical cell damage. Note: Based on the close formulation similarity of Magtrace™ to Sienna+, further cytotoxicity testing was considered unnecessary and was not repeated for the new formulation. | PASS |
| Sensitization EN ISO 10993-10 | In Vitro maximization – Allergenicity test on Magtrace® equivalent | To show no allergenic potential in guinea pig. | PASS |
| Irritation & Intracutaneous Reactivity DIN ISO 10993-10 | In Vitro Irritation & Intracutaneous Reactivity Test on Sienna+ | Polar and apolar extracts not to cause any intracutaneous reactivity in rabbits within an observation period of 72 hours. Based on the close formulation similarity of Magtrace™ to Sienna+, further cytotoxicity testing was considered unnecessary | PASS |
| Systemic Toxicity EN ISO 10993-11 | Acute Toxicity: on Magtrace® equivalent | Single dose toxicity studies in rats, mice and dogs to show No toxicity or a Low acute toxicity with doses in the 12.5-20 mmol iron per kilogram of body weight range. | PASS |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 8
{8}
| Test | Purpose | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| **Biocompatibility- Sienna+/Magtrace®** | | | |
| Systemic Toxicity EN ISO 10993-11 | Subacute & Sub-chronic Toxicity on Magtrace® equivalent | In dogs an increase in serum iron and decrease in iron binding capacity was dose dependent. In 4-week studies in rats, an increase in serum iron and increase in liver and spleen weights was observed as dose-dependent at the end of the dosing period. A transient decrease in platelet counts was also observed this was shown to be due to the iron moiety and only observed in animal models only and not observed in human tests. Chronic toxicity (6-12 month repeated dose) is not deemed necessary as Sienna+/Magtrace® is given in single dose only. In conclusion, Sienna+ with a single dose of 1mmol per patient can be considered safe with regards to subacute toxicity. | PASS |
| Genotoxicity EN ISO 10993-3 | In-vitro (Ames test) and in-vivo tests (mice, micronucleus) tests on Magtrace® equivalent to detect mutagenic potential | To show no mutagenic potential. | PASS |
| **Cleaning – Sentimag®** | | | |
| Cleaning Validation Study | Validate the cleaning instructions for the Sentimag® system | Acceptance criteria in accordance with Guidance for Industry and FDA Staff – Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labelling and AAMI TIR30 (2011). | PASS |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 9
{9}
| Test | Purpose | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| **Biocompatibility- Sienna+/Magtrace®** | | | |
| **Lifetime Evaluation – Sentimag®** | | | |
| Lifetime Evaluation of the Sentimag® System | Estimate the maximum lifetime of the Sentimag® system | Lifetime of the device is estimated to last 5 years. | Pass |
| **Performance Testing – Pre-Clinical Bench Top Testing:** | | | |
| Investigation of performance for Sienna+ with Sentimag® system | Measure the detection distance of Sienna+ with the Sentimag® probe at the three sensitivity settings available on the device. | Maximum sensing distance of the 28μg Fe magnetic tracer sample was 8mm, 11mm and 14mm for Sensitivity setting 1, 2, and 3 respectively. For the 140μg sample, maximum sensing was achieved at 13mm, 18mm and 19mm for sensitivity setting 1, 2, and 3 current settings respectively. | PASS |
| Comparison of Generation 1 Sentimag® system, Generation 2 Sentimag® system and Generation 3 Sentimag System | The second generation probe was developed for: •Reduction in probe diameter to be similar to a Gamma probe •Increase in sensitivity for transcutaneous and small node detection •Greater resistance to thermal drift. | **Diameter** reduced from 24mm (Gen 1) to 18.5mm (Gen 2). Additionally, Gen 3 has a 13.5mm probe **Sensitivity** of Gen 2 system increased approximately 3.5x over that of the Gen 1 System. The sensitivity of Gen 3 for both standard diameter probe (18.5mm) and thinner probe (13.5mm) is exactly the same as Gen 2. | Pass |
### B. Electrical Safety Testing – Sentimag®
The purpose of these tests is to demonstrate the electrical safety of the Sentimag® System in accordance with the FDA Recognized Consensus Standard: AAMI/ANSI ES60601-1:2005 / IEC 60601-1:2005 + Corrigenda 2006 and 2007 Medical
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 10
{10}
electrical equipment — Part 1: General requirements for basic safety and essential performance, and CAN/CSA-C22.2 No. 60601-1:08 - Medical electrical equipment - Part 1: General requirements for safety and essential performance.
**Table 2: Electrical Safety Testing**
| Test Name | Acceptance Criteria | Results |
| --- | --- | --- |
| **Electrical Safety Testing – Sentimag®** | | |
| Marking Durability and Legibility Test | As defined in the standard | PASS |
| Power Input Test | As defined in the standard | PASS |
| Limitation of Voltage and/or Energy (Capacitance Discharge Test) | As defined in the standard | PASS |
| Enclosures and Protective covers (Access to live parts) | As defined in the standard | PASS |
| Grounding Impedance Test | As defined in the standard | PASS |
| Leakage Current Test | As defined in the standard | PASS |
| Dielectric Voltage Withstand Test | As defined in the standard | PASS |
| Mechanical Tests and Stability | As defined in the standard | PASS |
| Temperature Test | As defined in the standard | PASS |
| Spillage, Cleaning/Disinfection and Humidity Preconditioning | As defined in the standard | PASS |
| Abnormal Operation Tests | As defined in the standard | PASS |
| Creepage Distance and Air Clearance measurements | As defined in the standard | PASS |
| Insulation – Ball pressure test | As defined in the standard | PASS |
| Acoustic Energy Test | As defined in the standard | PASS |
| Actuating Parts Test | As defined in the standard | PASS |
The Sentimag® system has been tested, examined, and found to comply with the applicable requirements of UL 60601-1 Medical Electrical Equipment, Part 1 Requirements for Safety April 25, 2003, US National standard ANSI/AAMI ES60601-1: 2005 / A2:2010 – Medical Electrical Equipment, Part 1: General Requirements for Safety and Essential
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 11
{11}
Performance, and CAN/CSA-C22.2 No. 60601-1:08 - Medical electrical equipment - Part 1: General requirements for safety and essential performance.
### C. Electromagnetic Compatibility Testing
The purpose of these tests is to demonstrate the electromagnetic compatibility of the Sentimag® system in accordance with these standards:
- FDA Recognized Consensus Standard: AAMI/ANSI IEC 60601-1-2:2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 3).
**Table 3:** Electromagnetic compatibility testing
| Test Name | Acceptance Criteria | Results |
| --- | --- | --- |
| **EMC Testing – Sentimag®** | | |
| EMC Testing | As defined in the standard | PASS |
| Verification EMC Testing | As defined in the standard | PASS |
| FCC EMC Testing | As defined in the FCC Rules - FCC Rules CFR47: 2008 Part 15.107 and 15.109 Class B | PASS |
| Conducted RF Immunity | As defined in the standard RF Immunity (EN61000-4-6) | PASS |
| Radiated Immunity Test | As defined in the standard Radiated Immunity (EN61000-4-3) | PASS |
### D. Animal Studies
**Table 4:** Animal Studies
| Test Name | Purpose | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| **Systemic Transport – Sienna+** | | | |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 12
{12}
| Sienna+ Transport Mechanism (Murine) | To determine Transport time and mechanism of transport of Sienna+ into Sentinel Lymph Node (SLN). Transport of Sienna+ into SLN was monitored in contrast to transport of immunologically marked Tetramethylrhodamine (TRITC)-positive Leukocytes. Time points: 10min, 30min, 1hr, 2hr, 24hr | Sienna+ appeared in SLN after 10 minutes, whereas TRITC-positive leukocytes were only detected at 24 hours. Rapid transport (minutes) - non interactive transport into lymphatic system. Slow transport (hours) - phagocytosis. Results demonstrate that the transport of the Sienna+ particles was mechanical and did not depend on cells or chemical means to transport the particles into the lymphatic system. | PASS |
| --- | --- | --- | --- |
| **Formulation – Sienna+** | | | |
| Sienna+ Formulation in Porcine model | Optimization of formulation for uptake into the lymphatic system in presence of different formulation components. Time points: 5min, 10min, 15 min, 30min, 1hr, 2hr, 24hr, 72hr | Presence of ion pair is essential for uptake into the lymphatic system. Presence of 0.3% w/w NaCl is the optimal concentration of excipient to enhance the uptake of Sienna+ into the lymphatic system | PASS |
## **X. SUMMARY OF PRIMARY CLINICAL STUDIES**
The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of Magtrace® and Sentimag® Magnetic Localization System for localizing lymph nodes draining a tumor site in patients with breast cancer, as part of a sentinel lymph node biopsy procedure in the U.S. (IDE #G140208, NCT02336737).
An additional supporting study was conducted in France (NCT01790399, See section XI), which was an open-label, multicenter, paired comparison of Sentimag® and Sienna+ and radioisotope with or without blue dye for sentinel lymph node detection in patients with breast cancer scheduled for sentinel node biopsy. Sienna+ is an earlier formulation of Magnecarbodex requiring dilution with saline prior to injection. Note: The French Sentimag Feasibility Trial is discussed in
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 13
{13}
the Summary of Supplemental Clinical Information section (Section XI) because it was only supporting clinical data.
Table 6: Clinical Studies
| Study | Products used | Study design | Location | Number of subjects (sites) |
| --- | --- | --- | --- | --- |
| U.S. SentimagIC trial G140208, NCT02336737 | Magtrace®, Sentimag® | Multi-center paired comparison with Radioisotope + Blue dye | US | 160 (6) |
| French Sentimag® Feasibility Trial, NCT01790399 | Sienna+, Sentimag® | Multi-center paired comparison with Radioisotope ± Blue dye | France | 115 (4) |
### A. Study Design
Patients were treated between January 9, 2015 and December 16, 2015. The database for this PMA reflected data collected through December 16, 2015 and included 160 patients. There were six (6) investigational sites in the United States.
The study was a pivotal, prospective, open label, multi-center, paired comparison study of the Magtrace®/Sentimag® system with the standard of care (Tc-99m radioisotope with blue dye) for the detection of lymph nodes in patients with breast cancer undergoing a sentinel lymph node biopsy (G140208). The trial was designed to provide powered evidence that the lymph node detection rate of the Magtrace®/Sentimag® system is non-inferior to the standard of care in patients with breast cancer and to summarize measures of product safety and performance.
The active control was Technetium 99 labeled sulfur colloid radioisotope in combination with isosulfan blue dye. The control was administered according to the standard of care at each site. All subjects underwent simultaneous lymph node mapping using Magtrace®, and with radioisotope with blue dye.
The trial sought to reject a null hypothesis that the true per lymph node detection rate for Magtrace® was worse than or equal to the true lymph node detection rate for standard of care by more than the non-inferiority margin δ, and support the alternative
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 14
{14}
hypothesis that the true lymph node detection rate of Magtrace® was no worse than the true lymph node detection rate for standard of care less the non-inferiority margin δ. That is:
$$\mathrm{H}_{0}: \mathrm{P}_{\mathrm{T}} - \mathrm{P}_{\mathrm{C}} \leq -\delta \text{ (inferior)}$$
$$\mathrm{H}_{\mathrm{a}}: \mathrm{P}_{\mathrm{T}} - \mathrm{P}_{\mathrm{C}} > -\delta \text{ (non-inferior)},$$
where P_T and P_C are the lymph node detection rates for Magtrace™ and standard of care Control, respectively, and δ is the non-inferiority margin.
The sample size calculation for the primary endpoint was performed using PASS 2008 and was based on a non-inferiority (one-sided) test of correlated proportions and the method of Nam with the following assumptions:
- Expected Sentimag®/Magtrace® (test) rate = 95%
- Expected standard of care (Control) rate = 95%
- Non-inferiority margin δ = 5%
- Assumed discordance rate = 8%
- Test significance level (α = 0.05 (1 -sided))
- Power (1-β) ≈ 0.85
A minimum of 265 nodes were required for each method. Given that ~ two (2) lymph nodes were expected per subject, it was anticipated that a total of 140 subjects would be required.
The expected per node detection rate for the standard of care combined technique was 94.6% based on the NSABP B-32 trial (Krag et al.¹).
### 1. Clinical Inclusion and Exclusion Criteria
Enrollment in the Sentimag® study (G140208) was limited to patients who met the following inclusion criteria:
- Subjects with a diagnosis of primary breast cancer or subjects with pure ductal carcinoma in situ (DCIS)
- Subjects scheduled for surgical intervention, with a sentinel lymph node biopsy procedure being a part of the surgical plan
- Subjects aged 18 years or more at the time of consent
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 15
{15}
- Subjects with an Eastern Cooperative Oncology Group (ECOG) performance status of Grade 0-2
- Subject has a clinical negative node status (i.e., T0-3, N0, M0)
Patients were not permitted to enroll in the Sentimag® study if they met any of the following exclusion criteria:
- The subject is pregnant or lactating
- The subject has clinical or radiological evidence of metastatic cancer including palpably abnormal or enlarged lymph nodes
- The subject has a known hypersensitivity to Isosulfan blue dye
- The subject has participated in another investigational drug study within 30 days of scheduled surgery
- Subject has had either a) previous axilla surgery, b) reduction mammoplasty, or c) lymphatic function that is impaired in the surgeon's judgment
- Subject has had preoperative radiation therapy to the affected breast or axilla
- Subject has received a Feraheme® (ferumoxytol) Injection within the past 6 months
- Subject has intolerance or hypersensitivity to iron or dextran compounds or to Magtrace™
- Subject has an iron overload disease
- Subject has pacemaker or another implantable device in the chest wall
## 2. Study Procedure and Follow-up Schedule
The study procedure flow is depicted in Figure 3 below.
Figure 3: Sentinel Node Biopsy Procedure Flow
Start of Procedure
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 16
{16}

Each SLN identified by Sentimag® and/or gamma probe or stained blue or black was excised and additional counts, with the excised node on the end of the probe, were taken with each detection system (Sentimag® and gamma probe) and recorded. In addition, nodes that were deemed highly clinically suspicious nodes (e.g., very hard and firm, or white colored consistent with gross tumor in the lymph node) were excised as sentinel nodes. Sentinel lymph node biopsy
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 17
{17}
(SLNB) was stopped when the residual count/signal in the axilla was less than 10% of the largest ex-vivo reading from an already excised node using that detection method.
All patients were scheduled to return for follow-up examinations at between 6 and 22 days post-procedure for a safety assessment postoperatively.
The study visits and assessments are summarized in Table 7.
Table 7: Study Visits and Data Collection Overview
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 18
{18}
| Procedure/ Assessment | Screening / Enrollment | Visit 1 Baseline / Medical History | Visit 2 Sentinel Node Biopsy Procedure | Visit 3 Post-procedure Evaluation (14 days +/- 8 days) | Unscheduled Visit |
| --- | --- | --- | --- | --- | --- |
| Inclusion / Exclusion Criteria | X | | | | |
| Informed Consent | X | | | | |
| Demographics, Medical / Surgical History | | | | | |
| Pregnancy test | | | X | | |
| Lymph node mapping and sentinel node biopsy procedure | | | X | | |
| Excised nodes sent for histological analysis & pathology evaluation | X | | X | | |
| SLN Biopsy results | | | | X | |
| Adverse Event Assessment | | X | X | X | X |
| Medications | | X | X | X | X |
| Device Deficiency Assessment | | | X | | |
| Study Completion | | | | X | |
### 3. Clinical Endpoints
#### Primary Safety Endpoint:
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 19
{19}
To provide evidence of the safety of Magtrace®/Sentimag® as indicated by adverse events and serious adverse events and their relatedness to the detection method or procedure.
# **Primary Effectiveness Endpoint:**
The primary effectiveness endpoint was the lymph node detection rate, which is defined as the number of lymph nodes identified by a specific method (Magtrace™/Sentimag® or Control) divided by the total number of lymph nodes detected.
# **Success/Failure Criteria:**
The study was considered a success if Magtrace™/Sentimag® demonstrated a statistically significantly non-inferior lymph node detection rate compared to the Control, with a 5% non-inferiority margin. If the lower bound of the one-sided 95% confidence interval for the difference between detection rates at the nodal level was greater than -5%, then the study was considered a success.
# **B. Accountability of PMA Cohort**
At the time of database lock, of 160 patients enrolled in the PMA study, 147 patients (91.9%) completed the study and are available for analysis. Patient accountability is shown in Figure 4.
Thirteen (13) patients withdrew from the study prior to sentinel lymph node biopsy procedure as follows: five (5) patients withdrew themselves, and eight (8) patients were withdrawn by investigators for the following reasons:
- Two (2) received the incorrect isotope injection (Lymphoseek (technetium Tc 99m tilmanocept) instead of Tc-99m sulfur colloid)
- Two (2) were found not to meet the inclusion/exclusion criteria
- One (1) was withdrawn due to concerns regarding her history of thalassemia
- One (1) was found to have axillary metastasis on a PET scan
- One (1) was withdrawn as there was no study coordinator on site to record the study data
- One (1) patient opted for chemotherapy prior to surgery
The primary analysis set was the modified intent to treat (mITT) cohort comprising all subjects who completed the study procedures (n=147).
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 20
{20}
Figure 4: SentimagIC trial patient accountability tree

### C. Study Population Demographics and Baseline Clinicopathological Characteristics
Patient demographic characteristics are shown in Table 8 with the patient baseline clinicopathological characteristics given in Table 9.
Table 8: Study Population Demographics
| | Overall (N=147) |
| --- | --- |
| Race (not mutually exclusive, %) | |
| American Indian or Alaska Native | 0.0% |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 21
{21}
| | Overall (N=147) |
| --- | --- |
| Asian | 4.8 % |
| Black or African American | 7.5% |
| Pacific Islander | 0.0% |
| White | 82.3% |
| Other | 6.1% |
| Ethnicity (n/N (%)) | |
| Hispanic or Latino | 11.6% |
| Not Hispanic or Latino | 88.4% |
| Mean Age (SD) | 61.1 (12.3) |
| Mean Weight in lbs (SD) | 167.1 (38.5) |
| Mean Height in inches (SD) | 63.7 (2.6) |
| Mean Body Mass Index (BMI Kg/m² (SD)) | 29.0 (6.9) |
| Menopausal status | |
| Premenopausal | 19.0% |
| Perimenopausal | 3.4% |
| Postmenopausal | 77.6% |
Table 9: Baseline Patient Clinicopathological Characteristics
| Type of surgery* | |
| --- | --- |
| Wide local excision/Lumpectomy | 103/147 (70.1) |
| Mastectomy | 43/147 (29.3) |
| Tumor location | |
| Upper Outer Quadrant (UOQ) | 74/147 (50.3) |
| Upper Inner Quadrant (UIQ) | 28/147 (19) |
| Lower Inner Quadrant (LIQ) | 10/147 (6.8) |
| Lower Outer Quadrant (LOQ) | 26/147 (17.7) |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 22
{22}
| Central/Areolar | 9/147 (6.1) |
| --- | --- |
| Pathological tumor size | |
| pTis | 13/135 (9.6) |
| pT1a | 19/135 (14.1) |
| pT1b | 30/135 (22.2) |
| pT1c | 33/135 (24.4) |
| pT2 | 33/135 (24.4) |
| pT3 | 7/135 (5.2) |
| Tumor grade | |
| I | 45/135 (33.3) |
| II | 51/135 (37.8) |
| III | 37/135 (27.4) |
| IV | 0/135 (0.0) |
| Not assessable | 2/135 (1.5) |
| Estrogen Receptor (ER) Status (n/N (%)) | |
| Positive | 113/135 (83.7) |
| Negative | 13/135 (9.6) |
| Not performed | 9/135 (6.7) |
| Progesterone Receptor (PR) Status (n/N (%)) | |
| Positive | 87/135 (64.4) |
| Negative | 39/135 (28.9) |
| Not performed | 9/135 (6.7) |
| Human Epidermal Growth Factor Receptor (HER2) Status (n/N (%)) | |
| Positive | 13/135 (9.6) |
| Negative | 105/135 (77.8) |
| Not performed | 17/135 (12.6) |
* One patient had SLNB only
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 23
{23}
## D. Safety and Effectiveness Results
# 1. Safety Results
The analysis of safety was based on the cohort of 147 evaluable patients. The key safety outcomes for this study are presented below. Adverse effects are reported in Tables 10 and 11.
Adverse events that occurred in the PMA clinical study:
A total of 69 adverse events were reported in 56/147 (38.1%) subjects, and of these adverse events, 9 (13.0%) were considered serious adverse events (SAE).
The most common adverse events were breast discoloration/hyperpigmentation, which occurred in 16.3% (24/147) of subjects and ecchymosis/bruising, which occurred in 6.8% (10/147) of subjects. Breast discoloration was not observed in patients that underwent mastectomy (43/147 or 29.3%) at follow up visit between 6-22 days post-surgery.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 24
{24}
Table 10: Adverse events by type
| Adverse Event Type | Events (N) | Subjects n (%) |
| --- | --- | --- |
| Total Adverse Events | 69 | 56 (38.1) |
| Breast Discoloration/Hyperpigmentation | 24 | 24 (16.3) |
| Ecchymosis / Bruising | 10 | 10 (6.8) |
| Pain | 5 | 5 (3.4) |
| Other | 5 | 5 (3.4) |
| Gastrointestinal Disorder | 3 | 3 (2.0) |
| Cellulitis | 3 | 3 (2.0) |
| Skin Ischemia | 3 | 3 (2.0) |
| Cardiac Disorder | 3 | 3 (2.0) |
| Rash | 2 | 2 (1.4) |
| Erythema | 2 | 2 (1.4) |
| Respiratory Disorder | 1 | 1 (0.7) |
| Hypertension | 1 | 1 (0.7) |
| Hypotension | 1 | 1 (0.7) |
| Pulmonary Embolism | 1 | 1 (0.7) |
| Musculoskeletal Disorder | 1 | 1 (0.7) |
| Psychological Disorder | 1 | 1 (0.7) |
| Allergic Reaction | 1 | 1 (0.7) |
| Pleural Effusion | 1 | 1 (0.7) |
| Inflammation | 1 | 1 (0.7) |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 25
{25}
Table 11 shows Magtrace®-related adverse events. If an adverse event was assessed as having an “undetermined” relationship, it was conservatively considered “related.”
Twenty (20) events occurring in 20 subjects (13.6%) were related to Magtrace®, and six (6) events occurring in six (6) subjects (4.1%) were assessed as having an undetermined relatedness in relation to Magtrace®. There were nine (9) serious adverse events in the study. After data analysis, seven (7) out of the nine (9) SAEs were unrelated to the Magtrace®, and two (2) of the nine (9) SAEs were found to be undetermined (Bradycardia and Anaphylaxis).
Table 11: Magtrace®-Related Adverse Events
| | Magtrace™-Related Adverse Events | |
| --- | --- | --- |
| Adverse Event Type | Events N | Subjects n (%) |
| Total Adverse Events | 26 | 25 (16.3) |
| Breast Discoloration/Hyperpigmentation^{1} | 23 | 23 (15.6) |
| Erythema | 1 | 1 (0.7) |
| Anaphlaxis^{2} | 1 | 1 (0.7) |
| Cardiac Disorder^{3} | 1 | 1 (0.7) |
$^{1}$Breast Discoloration: The degree and duration of skin staining is unknown. Skin staining was not observed in patients that underwent mastectomy (43/147) at follow-up visit between 6-22 days post-surgery.
$^{2}$Anaphlaxis: During the procedure the patient developed tongue swelling, hypotension, and tachycardia and was treated with epinephrine and steroids and the event resolved that day.
$^{3}$Cardiac Disorder: Thirty (30) minutes after injection bradycardia followed by pulselessness treated with atropine, CPR with intubation and the event resolved.
## 2. Effectiveness Results
The analysis of effectiveness was based on the 147 evaluable patients who completed the study. Key effectiveness outcomes are presented in Tables 12 to Table 17.
### Primary Endpoint Analysis
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 26
{26}
The primary endpoint was the lymph node detection rate, which is defined as the number of lymph nodes identified by a specific method (Magtrace® or Control) divided by the total number of lymph nodes detected (n=369). The Magtrace®/Sentimag® had a detection rate 94.3% and the control detected 93.5% of the total nodes detected. The difference in detection rates between the methods (Magtrace® - Control) was 0.8% with a 95% one-sided lower confidence bound of -2.1%.
Table 12: Summary of Overall mITT Study Results
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 27
{27}
| | G140208 Pivotal Study Breast Cancer | |
| --- | --- | --- |
| | Magtrace® n = 147 | Radioisotope with blue dye n = 147 |
| Nodes detected (n) | 348 | 345 |
| Per node lymph node detection rate % (95% CI) | 94.3% (91.9%, 96.7%) | 93.5% (91.0%, 96.0%) |
| Per patient lymph node detection rate % (95% CI) | 98.6% (95.2%, 99.8%) | 98.0% (94.2%, 99.6%) |
| Overall per patient concordance % (95% CI) | 98.0% (94.2%, 99.6%) | |
| Patients with at least one positive (metastatic) node (n) | 22 | |
| Detection rate for patients with at least one metastatic node % (95% CI) | 95.5% (86.8%, 100.0%) | 95.5% (86.8%, 100.0%) |
Table 13: The nodal detection rates
| | Magtrace™ | | |
| --- | --- | --- | --- |
| Control (Radioisotope and Blue Dye) | Detected | Not Detected | Total |
| Detected | 326 (88.3%) | 19 (5.2%) | 345 (93.5%) |
| Not Detected | 22 (6.0%) | 2 (0.5%) | -- |
| Total | 348 (94.3%) | -- | 369^{1} (100.0%) |
$^{1}$Four sentinel lymph nodes are excluded due to missing data for Magnetic (Magtrace®) count, Radioisotope count and/or Blue Dye.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 28
{28}
There were 41 discordant nodes in 29 subjects; 19 were found by control only and 22 were found by Sentimag® only.
Table 14: Findings of Discordant Lymph Nodes
| Overall discordant Nodes | Rate | Number of Nodes Detected by Test but not Control | Number of Nodes Detected by Control but not Test |
| --- | --- | --- | --- |
| 41/369 | 11.1% | 22 (in 16/29 patients) | 19 (in 13/29 patients) |
All of the discordant nodes had no clinical impact as:
- All malignant SLNs were concordant
- All discordant SLNs were benign. (See Table 20 malignant nodes table)
Table 15: Sentinel Node per-Node Detection Rates by Radioisotope Alone
| | Magtrace™ | | |
| --- | --- | --- | --- |
| Radioisotope | Detected | Not Detected | Total |
| Detected | 319 (86.4%) | 19 (5.1%) | 338 (91.6%) |
| Not Detected | 29 (7.9%) | 2 (0.5%) | -- |
| Total | 348 (94.3%) | -- | 369^{1} (100.0%) |
$^{1}$Four sentinel lymph nodes are excluded due to missing data for Magnetic (Magtrace®) count, Radioisotope count and/or Blue Dye.
Table 16: Sentinel Node per-Node Detection Rates by Blue Dye Alone
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 29
{29}
| | Magtrace® | | |
| --- | --- | --- | --- |
| Blue Dye | Detected | Not Detected | Total |
| Detected | 175 (47.4%) | 5 (1.4%) | 180 (48.8%) |
| Not Detected | 173 (46.9%) | 16 (4.3%) | -- |
| Total | 348 (94.3%) | -- | 369¹ (100.0%) |
¹Four sentinel lymph nodes are excluded due to missing data for Magnetic (Magtrace®) count, Radioisotope count and/or Blue Dye.
Table 17: Sentinel Node per-Subject Detection Rates by Method
| | Magtrace® | | |
| --- | --- | --- | --- |
| Control (Radioisotope and Blue Dye) | At Least 1 Node Detected | No Nodes Detected | Total |
| At Least 1 Node Detected | 144/147 (98.0%) | 0 (0.0%) | 144 (98.0%) |
| No Nodes Detected | 1/147 (0.7%) | 1/147 (0.7%) | -- |
| Total | 145/147 (98.6%) | -- | 147 (100.0%) |
### Other Endpoint Analysis
Table 18: Results of Other Per Node Endpoints
| Per Node Endpoints |
| --- |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 30
{30}
| | **n/N Rate (95% CI)** |
| --- | --- |
| **Overall Nodal Concordance** Number of nodes identified by both test and Control out of all nodes identified | 326/369 (88.3%) CI (85.1%, 91.6%) |
| **Overall Nodal Discordance** Number of nodes identified by either test or Control (but not by both) out of all nodes identified | 41/369 (11.1%) CI (7.9%, 14.3%) |
| **Nodal concordance** Number of nodes identified by both test and Control out of nodes identified by Control | 326/345 (94.5%) CI (92.1%, 96.9%) |
| **Reverse nodal concordance** Number of nodes identified by both test and Control out of nodes identified by test | 326/348 (93.7%) CI (91.1%, 96.2%) |
Table 19: Number of Lymph Nodes Detected per Subject Assessed for Each Method.
| | Mean (S.D) | Median | Range |
| --- | --- | --- | --- |
| Magtrace® | 2.4 (1.19) | 2 | 0-6 |
| Control | 2.4 (1.34) | 2 | 0-6 |
| Radioisotope | 2.3 (1.38) | 2 | 0-6 |
| Blue Dye | 1.2 (0.93) | 1 | 0-4 |
### 3. Subgroup Analysis
#### Per node endpoints for cancer positive (malignant) nodes
The nodal status was reported as the percentage of histologically malignant nodes detected by a specific detection method (magnetic; combined radioisotope and blue dye; radioisotope alone; blue dye alone) on a per node and a per subject basis.
Of the 25 confirmed analyzable positive (malignant) nodes in the mITT analysis set, 96.0% (24/25) with a 95% CI of (88.3%, 100.0%) were identified by both the Control radioisotope or blue dye, and Magtrace®. One (1) node was not
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 31
{31}
identified by either Control or Magtrace®, but was considered 'highly clinically suspicious' in the judgment of the investigator. All the nodes identified by either Magtrace® or Control were identified by both Magtrace® and Control. Blue dye detected 60.0% (15/25).
Of the 24 malignant nodes identified by both Magtrace® and Control, 19 contained macrometastasis, and five (5) contained micrometastasis. The one node that was not identified by either Control or Magtrace® but was considered clinically suspicious contained a macrometastasis.
Table 20: Sentinel lymph node detection of malignant nodes - per node
| | Magtrace™ | | |
| --- | --- | --- | --- |
| Control (Radioisotope or Blue Dye) | Cancer Positive Detected | Cancer Positive Not Detected | Total |
| Cancer Positive Detected | 24 (96.0%) | 0 (0.0%) | 24 (96.0%) |
| Cancer Positive Not Detected | 0 (0.0%) | 1 (4.0%) | -- |
| Total | 24 (96.0%) | -- | 25^{1} (100.0%) |
One additional positive node (and the one subject with this node) is excluded from analyses discussed above since it did not meet any of the criteria for a sentinel lymph node. This node, subject 06-018, Node 4, was one of two (2) nodes excised in a single piece of tissue: subject 06-018, Nodes 3 and 4. Node 3 had a Magtrace® and radioisotope signal and was recorded as a sentinel lymph node. Node 4 did not meet any of the pre-determined criteria for a sentinel lymph node and was therefore recorded as a non-sentinel lymph node. Upon histopathological analysis Node 4 was found to be malignant.
# 4. Pediatric Extrapolation
In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 32
{32}
### **E. Financial Disclosure**
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 13 investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data.
## **XI. SUMMARY OF SUPPLEMENTAL CLINICAL INFORMATION**
The supplemental clinical information includes:
1. French Clinical Study NCT01790399
2. Subgroup Analyses of Mastectomy Cohort
3. Protocol Deviations in the Pivotal (Sentimag®) study
4. Device Failures in the Pivotal (Sentimag®) Study
5. Magnetic Resonance Imaging (MRI) Artifact
6. Published literature studies
### **1. French Feasibility Study Summary:**
A feasibility study was conducted in France (NCT No: NCT01790399). This was an investigator-led multi-center paired comparison of Sienna+ and Sentimag® with radioisotope ± Blue dye. Sienna+ is a previous formulation of the same iron oxide particles, which required dilution with saline prior to injection.
A. French Study Title: Detection of Sentinel Node using Sentimag®/Sienna+ for breast cancer: A feasibility study.
B. Overview of Feasibility Trial
Patients were treated between January 30, 2013, and January 22, 2014.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 33
{33}
# C. Patient Disposition
Number enrolled: n=115
Number of evaluable patients: n=108
Withdrew: n=7, 1 withdrew consent, 1 did not receive study drug, the remainder had missing data due to data entry fault at the time of surgery
Number of participating Centers: n=4
# D. Study Objectives
- Primary: To evaluate the feasibility of the sentinel lymph node identification technique using the Sentimag® device (manual magnetometer)/Sienna+ (superparamagnetic iron-oxide tracer).
- Secondary: To evaluate the reliability of the technique compared with benchmark methods (isotopic and/or colorimetric).
# E. Clinical Endpoints
Safety Endpoint:
- Rates of adverse events and serious adverse events were recorded.
Primary Endpoint:
- The primary endpoint of this trial was the proportion of successful procedures for SLN identification (identification rate per patient) by the magnetic method compared with the standard method (isotopes with or without patent blue).
Other Endpoints:
- The secondary endpoint evaluated the concordance of sentinel nodes detected with magnetic and standard method. The concordance is reported by patient and by node.
- Concordance per subject is defined as the number of subjects in whom the magnetic technique agrees with the standard technique (i.e., subjects in whom either both identified a node, or neither identified a node) divided by the total number of evaluable subjects.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 34
{34}
- Concordance per node is defined as the number of nodes in whom the magnetic technique agrees with the standard technique (i.e., nodes detected by either both techniques or neither technique) divided by the total number of evaluable nodes.
Success/Failure Criteria:
- A successful procedure was defined as the detection of at least one magnetic sentinel node for the magnetic method; and at least one node radioactive and/or blue (if blue dye was used) for the standard method.
# F. Study Design
# Methodology:
- The investigated devices were the Sentimag® probe system and Sienna+ magnetic tracer. Sienna+ was diluted with 3ml of 0.9% saline prior to injection.
The control products used were: Nanocis® or Nanocoll albumin colloids radiolabeled with Technetium 99m isotope; with or without patent blue dye.
- Patients received the radioisotope injection first; either the day before or day of surgery, per the usual custom of the center. After induction of anesthesia, the Sienna+ was administered followed by blue dye.
- Sentinel Node Detection was first performed with Sentimag® followed by gamma probe and blue dye. All nodes identified by any method were removed.
Radioisotope (Technetium albumin colloid) was injected according to the standard of care protocol at each site. Forty-five (45) of 108 patients (45/108, 42%) also received a blue dye injection shortly prior to surgery at sites where blue dye was standard protocol. Sienna+ was injected at least 20 minutes prior to initiating sentinel lymph node mapping.
Lymph node detection was performed intraoperatively using the Sentimag® probe to identify magnetic nodes, followed by the use of a handheld gamma probe to identify radioactive ('hot') nodes. Any blue or black/brown stained nodes, and any nodes judged to be highly clinically suspicious by the surgeon were also excised. The excised nodes were evaluated using histopathology.
The percentage of lymph nodes identified by each technique was presented with a 95% confidence interval. The comparison of discordant pairs (identified or non-identified SLN) was conducted using the McNemar test per patient and
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 35
{35}
per lymph node. To detect a 5% discrepancy percentage between the two (2) techniques with a 95% confidence interval of 0.04, 115 evaluable patients needed to be enrolled.
# G. Clinical Inclusion and Exclusion Criteria
Enrolment in the French Study was limited to patients who met the following inclusion criteria:
- Female patients with invasive or micro-invasive breast cancer proven by histology or cytology regardless of the histology type
- cT0/cT1/cT2 (up to 5 cm) cN0 clinic and/or echographic previously untreated (chemotherapy or neo-adjuvant hormonotherapy)
- Aged 18 years or over
- Scheduled for breast surgery and axillary staging by sentinel lymph node
- Female patient using effective contraception (BHCG negative)
- Patient affiliated to a health insurance system
- Informed consent signed by the patient
Patients were not permitted to enroll in the French Study if they met any of the following exclusion criteria:
- T3 or T4 tumor (> 5 cm, cutaneous or muscular infiltration, or inflammatory cancer)
- Existence of an axillary adenopathy suspected clinically or in imaging
- Bifocal or multi-focal tumors known before surgery
- History of mammary of axillary surgery
- Metastatic patient
- Patient with a contra-indication to anesthesia and/or surgery
- Intolerance or hypersensitivity:
- to iron or dextran or superparamagnetic iron oxide particles
- to the patent blue dye in centers where it is currently used
- Patient unable to receive a radioactive isotope for excision of the sentinel lymph node
- Allergy to radioactive product
- Iron excess disease
- Cardiac stimulator or any other device implantable in the thoracic wall
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 36
{36}
- Unable to be medically monitored in the study for geographic, social or mental reasons
- Patient deprived of their freedom or under guardianship
- Pregnant or breast-feeding
# H. Patient accountability
One hundred fifteen (115) subjects were enrolled at four (4) investigational sites in France and 108 subjects completed the Sentinel Lymph Node Biopsy (SLNB) procedure. Seven (7) subjects were not evaluable: one (1) did not receive the Sienna+ injection; one (1) subject withdrew consent prior to the SLNB procedure; and five (5) had missing data for the Sentimag® technique due to a data entry fault in the operating room.
The patient accountability tree is shown in Figure 5.
Figure 5: Study 2 patient accountability tree

PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 37
{37}
# I. Study population demographics
The median age was 58 years (range 29-79). Histopathological analysis showed that 89% of tumors were invasive carcinoma. Baseline clinicopathologic characteristics for the French Study population are shown in Table 21.
Table 21. Demographic and Clinicopathologic Baseline Characteristics for the French Study Population
| | **N = 108** | **%** |
| --- | --- | --- |
| **Age** | | |
| ≤ 50 | 29 | 27 |
| 51-69 | 62 | 57 |
| ≥ 70 | 17 | 16 |
| **BMI** | | |
| Thin | 3 | 3 |
| Normal | 44 | 41 |
| Overweight | 40 | 37 |
| Obese | 18 | 17 |
| Morbidly obese | 2 | 2 |
| Missing | 1 | |
| **Hormonal status** | | |
| Active | 26 | 24 |
| Pre-menopausal | 5 | 5 |
| Menopausal | 77 | 71 |
| **Location of the lesion** | | |
| Upper inner quadrant | 26 | 24 |
| Upper outer quadrant | 62 | 57 |
| Lower-inner quadrant | 5 | 5 |
| Lower-outer quadrant | 9 | 8 |
| Retro-areolar | 5 | 1 |
| **Histology type** | | |
| Invasive root carcinoma | 96 | 89 |
| Invasive lobular | 9 | 8 |
| Other | 3 | 3 |
| **SBR Grade** | 37 | 34 |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 38
{38}
| | **N = 108** | **%** |
| --- | --- | --- |
| II | 58 | 54 |
| III | 13 | 12 |
| **Hormonal receptors** | | |
| Estrogen receptors | | |
| Negative | 9 | 8 |
| Positive | 99 | 92 |
| Progesterone receptors | | |
| Negative | 28 | 26 |
| Positive | 80 | 74 |
| **HER status (in IHC)** | | |
| 0 | 60 | 57 |
| + | 29 | 27 |
| ++ | 8 | 8 |
| +++ | 9 | 8 |
| Missing (#5, #6) | 2 | |
| **KI67** | | |
| ≤ 15 | 70 | 67 |
| >15 | 35 | 33 |
| Median (range) | 10 | (0- |
| Missing (#6, #8, #99) | 3 | 90) |
# J. Safety & Effectiveness Results
# Safety results:
Seventy (70) subjects had post-operative complications. The most common adverse events were breast discoloration/hyperpigmentation, which occurred in 22 subjects and seroma (noted as "punctured lymphocele") which occurred in 14 subjects.
Three (3) serious adverse events were recorded in two (2) subjects: one subject was hospitalized for a bacterial infection and one subject had two (2) separate hematoma events not related to the study. No serious adverse events related to the device were reported.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 39
{39}
Effectiveness results:
Table 22: Primary Endpoint Analysis
| | Sienna + (Magnecarbodex) n=108 | Radioisotope with/without Blue Dye n= 108 |
| --- | --- | --- |
| Nodes Detected (n) | 208 | 193 |
| Per Patient Lymph Node Detection Rate % (95% CI) | 97.2% (92.1%, 99.4%) | 95.4% (89.5%, 98.5%) |
| Overall per Patient Concordance % (95% CI) | 96.3% (90.8%, 99.0%) | |
Table 23: Detection Concordance for Cancer Positive Nodes
| Per Patient | Sienna cancer + | Sienna Cancer - |
| --- | --- | --- |
| Control cancer + | 43 | 1 |
| Control cancer - | 2 | 0 |
### Primary endpoint analysis
The primary endpoint of this trial was the proportion of successful procedures for Sentinel Lymph Node (SLN) identification (identification rate per patient) by the magnetic method compared with the standard method (isotopes with or without patent blue).
In total, 220 SLNs were collected from 106 patients. The identification of at least one SLN with standard method was achieved in 95.4% of patients (103/108, 95%CI: 89.5–98.5) and with Sienna+ in 97.2% of patients (105/108, 95%CI: 92.1–99.4).
The concordance rate per subject of the two (2) mapping methods (magnetic and isotopic ± patent blue) was 96.3%, 95%CI: 90.8–99.0). The discordance rate of both methods per subject was 3.7% (4/108, CI: 1.0–9.2%). The p-value for the Exact McNemar test was p = 0.6250, which means that there is insufficient statistical evidence that the two methods are discordant.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 40
{40}
# **Per node endpoints**
Among the 220 SLNs removed, 214 were subjected to statistical analysis (six (6) nodes had intraoperative tracer values missing). A mean [SD] of 2.08 [0.943] SLNs per subject were identified. The mean number of magnetic nodes identified was 2.01 [0.976] per subject and the mean of standard nodes identified was 1.94 [0.968]. The nodal concordance rate was 88.3% (95%CI: 83.2–92.3).
# **Endpoints for subjects with positive nodes**
Forty-six patients (46, 43.4%) had nodal involvement with 21 (45.7%) presenting micrometastasis and 25 (54.3%) presenting macrometastasis. The per subject malignancy detection rate was 95.7% (44/46, 95%CI: 85.2–99.5) for the standard method and 97.8% (45/46, 95%CI: 88.4–99.9) for the magnetic technique.
Among these node-positive patients, the concordance rate was 93.5% (43/46, 95% CI: 82.1%; 98.6%). For the 61 involved SLNs included in the calculation, the concordance rate was 86.9% (53/61, 95% CI: 75.8%; 94.2%).
Table 24 summarizes the per-patient and per-node endpoints.
Table 24: Per node and per patient lymph node detection rates for Sienna+ and Radioisotope in NCT01790399
| | French NCT01790399Study | |
| --- | --- | --- |
| | Sienna+ n = 108 | Radioisotope with or without blue dye n = 108 |
| Nodes detected (n) | 208 | 193 |
| Per node lymph node detection rate % (95% CI) | 97.2% | 90.2% |
| Per patient lymph node detection rate % (95% CI) | 97.2% (92.1%, 99.4%) | 95.4% (89.5%,98.5%) |
| Overall per patient concordance % (95% CI) | 96.3% (90.8%, 99.0%) | |
| Patients with at least one positive node n | 46 | |
| Detection rate for patients with at least one metastatic node % (95% CI) | 97.8% (88.4, 99.9) | 95.7% (85.2, 99.5) |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 41
{41}
# K. Protocol Deviations
A total of 36 protocol deviations was reported in 29.6% (34) of subjects. The most common protocol deviation was incorrect βHCG pregnancy testing or testing out of the specified timeframe. This deviation occurred 13 times and at all four (4) sites. The deviations that occurred did not negatively impact the scientific soundness or the data integrity of the clinical study.
# L. Financial Disclosure
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The supplemental clinical study included 14 investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data.
# M. NCT01790399 Feasibility Safety & Effectiveness Conclusions
The study success criterion was met showing no significant discrepancy between the per subject detection rates for the two (2) techniques. The investigational device produced a similar risk profile to Control with no unanticipated adverse device effects. The analysis of this study provides valid scientific evidence to support the safety and effectiveness of Sentimag®/Sienna+ to assist in detecting and localizing lymph nodes draining a tumor site in breast cancer, as part of a SLNB procedure.
# 1. Subgroup Analyses of Mastectomy Cohort
Forty-three (43) of the 160 patients in the pivotal trial underwent mastectomy with SLNB. The demographics of this cohort are shown in Table 25 below.
Table 25: Demographics of the Mastectomy Patient Cohort (Pivotal Study)
| Characteristic | n/N (%) or Mean (SD) |
| --- | --- |
| Race (not mutually exclusive, n/N (%)) | |
| American Indian or Alaska Native | 0/43 (0%) |
| Asian | 2/43 (4.7%) |
| Black or African American | 2/43 (4.7%) |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 42
{42}
| Native Hawaiian or Other Pacific Islander | 0/43 (0%) |
| --- | --- |
| White | 36/43 (83.7%) |
| Other | 3/43 (7.0%) |
| Ethnicity (n/N (%)) | |
| Hispanic or Latino | 5/43 (11.6%) |
| Not Hispanic or Latino | 38/43 (88.4%) |
| Age | 54.7 (11.7) |
| BMI | 26.8 (5.4) |
| Endpoint | |
| Magtrace^{TM} per node detection rate | 116/123 (94.3%) |
| Control per node detection rate | 115/123 (93.5%) |
| Magtrace^{TM} per subject detection rate | 43/43 (100%) |
| Control per subject detection rate | 43/43 (100%) |
| Node positive subjects: | 6/43 (14.0%) |
The baseline clinical pathological characteristics of the mastectomy cohort are shown in Table 26 below:
Table 26: Baseline Clinicopathologic Characteristics of the Mastectomy Patient Cohort
| Tumor location | |
| --- | --- |
| Upper Outer Quadrant (UOQ) | 23/43 (53.5%) |
| Upper Inner Quadrant (UIQ) | 10/43 (23.3%) |
| Lower Inner Quadrant (LIQ) | 1/43 (2.3%) |
| Lower Outer Quadrant (LOQ) | 5/43 (11.6%) |
| Central/Areolar | 4/43 (9.3%) |
| Pathological tumor size | |
| pTis | 6/38 (15.8%) |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 43
{43}
| pT1a | 2/38 (5.3%) |
| --- | --- |
| pT1b | 4/38 (10.5%) |
| pT1c | 9/38 (23.7%) |
| pT2 | 13/38 (34.2%) |
| pT3 | 4/38 (10.5%) |
| Tumor grade | |
| I | 6/38 (15.8%) |
| II | 19/38 (50%) |
| III | 11/38 (28.9%) |
| IV | 0/38 (0%) |
| Not assessable | 2/38 (5.3%) |
| Estrogen Receptor (ER) Status (n/N (%)) | |
| Positive | 34/43 (79.1%) |
| Negative | 4/43 (9.3%) |
| Not performed | 5/43 (11.6%) |
| Progesterone Receptor (PR) Status (n/N (%)) | |
| Positive | 25/43 (58.1%) |
| Negative | 13/43 (30.2%) |
| Not performed | 5/43 (11.6%) |
| Human Epidermal Growth Factor Receptor (HER2) Status (n/N (%)) | |
| Positive | 2/43 (4.7%) |
| Negative | 33/43 (76.7%) |
| Not performed | 8/43 (18.6%) |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 44
{44}
Table 27: Per node detection rates for mastectomy patients from the mITT group
| | Magtrace™ (mITT nodal analysis) | | |
| --- | --- | --- | --- |
| Control (Radioisotope and Blue Dye) | Detected | Not Detected | Total |
| Detected | 108/123 (87.8%) | 7/123 (5.7%) | 115/123 (93.5%) |
| Not Detected | 8/123 (6.5%) | 0/123 (0%) | |
| Total | 116/123 (94.3%) | | 123/123 (100%) |
Table 28: Malignant node per node detection rates for mastectomy patients from the mITT group
| | Magtrace™ (mITT nodal analysis of malignant nodes) | | |
| --- | --- | --- | --- |
| Control (Radioisotope and Blue Dye) | Malignant Detected | Malignant Not Detected | Total |
| Detected | 8/8 (100%) | 0/8 (0%) | 8/8 (100%) |
| Not Detected | 0/8 (0%) | 0/8 (0%) | |
| Total | 8/8 (100%) | | 8/8 (100%) |
## 2. Protocol Deviations in the Pivotal (Sentimag®) study
In total, 29 protocol deviations were reported in 17.5% (28) of subjects. The most common protocol deviation was the use of Lymphoseek (technetium Tc 99m tilmanocept) as the radioisotope Control versus the protocol-required radiolabeled sulfur colloid radioisotope. This deviation occurred 13 times at three (3) different sites. The deviations that occurred did not negatively impact the scientific soundness or the data integrity of the clinical study. However, subjects in whom Lymphoseek was used were excluded from the PP analysis as this met one of the pre-specified criteria for exclusion from the PP analysis set.
## 3. Device Failures in the Pivotal (Sentimag®) Study
Four (4) Sentimag® device failures were reported in four (4) subjects. No adverse effects occurred as a result of the device failures.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 45
{45}
# 4. Magnetic Resonance Imaging (MRI) Artifact
Magtrace® can cause image artifacts during magnetic resonance imaging (MRI) near injection and drainage site. These artifacts may be present long-term.
- Information from European sample cases and reports indicate that the artifact persists, often unchanged, for at least 25 months.
- The artifact from the device may make large parts of the images completely uninterpretable and nondiagnostic.
Magtrace® may also travel to regions away from the injection site such as liver, spleen, etc. if injected directly into the blood stream. In such cases the presence of Magtrace® may cause image artifacts during Magnetic Resonance Imaging (MRI) of those regions. Some manipulation of scan parameters may be required to compensate for the artifact.
Magtrace® residues have not been reported to produce artifacts affecting imaging in Contrast Enhanced Digital Mammography (CEDM), X-ray, PET, PET/CT, CT, or ultrasound studies.
Table 29 summarizes per patient or per breast occurrence of imaging artifacts in mastectomy patients.
In the study conducted by Krischer et al.², 24 subjects participated of which two (2) had bilateral mastectomy treatment making in total 26 breast cancer cases. Of these, 18 underwent Breast Conserving Surgery (BCS), and eight (8) underwent mastectomy. Of the BCS cases, the data from one subject (PID 15) was not interpretable due to breathing artifacts, leaving 17 interpretable BCS cases. There were two (2) bilateral surgeries, but no bilateral mastectomies. Subject PID 3 had a Right mastectomy and a left lumpectomy and subject PID 17 had bilateral lumpectomy. Therefore, in total, eight (8) patients underwent mastectomy, of whom one also had a lumpectomy in the contralateral breast. None of the cases show the occurrence of artifact.
In the SentimagIC pivotal study, 43/147 subjects had mastectomy. Of these, imaging was available for 2/43 plus a further subject 05-012 who received lumpectomy in the study and mastectomy after the study completed. None of the cases show the occurrence of artifact.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 46
{46}
Table 29: Per patient and per breast occurrence of artifact in post-mastectomy MRI
| Source | Number of post mastectomy images | Per patient (per breast) occurrence of artifact |
| --- | --- | --- |
| Krischer et al.^{2} (see reference: Krischer et al., Feasibility of breast MRI after sentinel procedure for breast cancer with superparamagnetic tracers, Eur J Surg Oncol. 2018 Jan;44(1):74-79.) | 24 subjects participated, of which two (2) had bilateral mastectomy, making 26 total breast cancer cases. • eight (8) mastectomies • 18 BCS (including one after chemotherapy) One subject (PID 15) not interpretable due to breathing artifacts and movement. Therefore 25 breast cancers eligible for analysis: • eight (8) mastectomies • 17 BCS Bilateral cases were: PID 3 right mastectomy, left lumpectomy; and PID 17 bilateral lumpectomy | 0/8 (0/8) |
| SentimagIC pivotal study | 43/147 subjects had mastectomy. Of these, imaging was available for 2/43 plus a further subject 05-012 who received lumpectomy in the study and mastectomy after the study completed. | 0/3 (0/5) (Only 3/5 breasts received Magtrace®) |
| Total | | 0/11 (0/15) |
Table 30 summarizes the type of mastectomy conducted after which the subject underwent MRI treatment. As noted above, there is no incidence of MRI artifacts observed in any of the cases outlined below.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 47
{47}
Table 30: Type of mastectomy before MRI
| Study | Type of mastectomy | Sienna injection technique | Incidence of MRI artifact |
| --- | --- | --- | --- |
| Krischer, 2018 paper | 8 subjects received mastectomy (non-skin or nipple sparing)** | Sub-areolar interstitial | 0/8 (None visible) |
| SentimagIC pivotal study Subject 05-012b,c* | Bilateral. Non-skin or nipple sparing | Sub-cutaneous, sub-areolar | None visible |
| SentimagIC pivotal study Subject 05-018 | Bilateral. Non-skin or nipple sparing | Sub-cutaneous, sub-areolar | None visible |
| SentimagIC pivotal study Subject 06-030 | Skin-sparing | Sub-cutaneous, sub-areolar | None visible |
*Subject 05-012 received lumpectomy surgery in the study, but subsequently bilateral mastectomy, after which these MRI scans were obtained.
**Data on the type of mastectomy obtained from the author via a personal communication.
The summary of studies on MRI Artifact is presented in Table 31 to Table 33.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 48
{48}
**Table 31 (All Injection Methods)**
| Study [Reference] | Amount of Magtrace Injected and Location (e.g., subareolar, periareolar, retroareolar, peritumoral, etc.) | Net Number of Subjects Participated/ Studied | Number (and %) of Subjects with Image Artifacts | Size of the Artifacts (mean, range) | Did Artifacts Resolve Completely (Yes (%), No (%)) | Time taken to resolve artifacts (mean, and range) | Radiology reported outcome (no artifacts, minimal artifacts, suitable for diagnosis/ non-diagnostic, significantly impaired) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Krischer 2018 | 2 ml Sienna+ 3ml Saline. 5 ml Retroareolar | n=17 evaluable | 13/17 (76.5%) | Not specified | Not specified | Not specified | 4/17 (23.5%) Grade 2 - without restriction, 10/17 (58.8%) Grade 1-with restriction, 3/17 (17.6%) Grade 0 – impossible |
| Aribal 2021 | 2 ml Sienna+ 3ml Saline. 5 ml Periareolar | n=25 | 21/25 (84%) | Not specified | Not specified | Not specified | No artifact 4/25(16%) Focal area of signal void artifact (less than 5mm) 10/25 (40%) Segmental area of signal void artifact 3/25(12%) Non diagnostic 8/25(32%) |
| PostMag Registry (Reference No. ISRCTN85167182) This study is not yet published. | 2ml Peritumoral | n=11 | 10 (91%) | Mean artifact Volume 10 cm^{3} Range: 0 cm^{3} - 28 cm^{3} | Yes (18%) No (82%) | Mean 3 months Range: 1 month to 5 months | No Artifacts 1/11 (9%) Minimal Artifacts suitable for diagnosis 6/11 (55%) Significantly impaired, but could be diagnostic 3/11 (25%) Non diagnostic, 1/11 (9%) |
| Unpublished study | 2ml Peritumoral | n=2 | 1(50%) | Mean artifact Volume 5 cm^{3} Range: 0 cm^{3} - 10 cm^{3} | Yes (50%) No (50%) | Not applicable, Magtrace intraoperative followed by MRI | No Artifacts 0/2 (0%) Minimal Artifacts suitable for diagnosis 2/2 (100%) Non diagnostic, significantly impaired 0/2 (0%) |
| Sentimag IC Study Reference: NCT02336737 IDE No. G140208 | 2ml Subareolar | 147/13 | 12 (92%) | Mean 68 cm^{3} Range 0 cm^{3}- 204 cm^{3} | Yes (8%) No (92%) | Mean: 10 months, Range: 1 month to 15 months | No Artifacts 1/13 (8%) Minimal Artifacts suitable for diagnosis 4/13 (31%) Significantly impaired but could be diagnostic 5/13 (39%) Non diagnostic, 3/13 (23%) |
| Shrotria, Stuart European Journal of Surgical Oncology 46 (2020) P136 Abstract 121. | 2ml Periareolar | 160/14 | Not confirmed, assumed 100% | Not known | Not confirmed, assumed No (100%) | Not known | No Artifacts 0/14 (0%) Minimal Artifacts suitable for diagnosis 11/14 (78%) Significantly impaired 0/14 (0%) Non diagnostic, 3/14 (21%) |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 49
{49}
**Table 32 (Peritumoral Injections)**
| Study [Reference] | Amount of Magtrace Injected | Net Number of Subjects Participated/ Studied | Number (and %) of Subjects with Image Artifacts | Size of the Artifacts (mean, range) | Did Artifacts Resolve Completely (Yes (%), No (%)) | Time taken to resolve artifacts (mean, and range) | Radiology reported outcome (no artifacts, minimal artifacts, suitable for diagnosis/ non-diagnostic, significantly impaired) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| PostMag Registry (Reference No. ISRCTN85167182) This study is not yet published. | 2ml Peritumoral | n=11 | 10 (91%) | Mean artifact Volume 10 cm^{3} Range: 0 cm^{3} -28 cm^{3} | Yes (18%) No (82%) | Mean 3 months Range: 1 month to 5 months | No Artifacts 1/11 (9%) Minimal Artifacts suitable for diagnosis 6/11 (55%) Significantly impaired, but could be diagnostic 3/11 (25%) Non diagnostic, 1/11 (9%) |
| Unpublished study | 2ml Peritumoral | n=2 | 1(50%) | Mean artifact Volume 5 cm^{3} Range: 0 cm^{3} -10 cm^{3} | Yes (50%) No (50%) | Not applicable, Magtrace intraoperative followed by MRI | No Artifacts 0/2 (0%) Minimal Artifacts suitable for diagnosis 2/2 (100%) Non diagnostic, significantly impaired 0/2 (0%) |
**Table 33 (Not Peritumoral Injections, All other methods of Injection)**
| Study [Reference] | Amount of Magtrace Injected | Net Number of Subjects Participated/ Studied | Number (and %) of Subjects with Image Artifacts | Size of the Artifacts (mean, range) | Did Artifacts Resolve Completely (Yes (%), No (%)) | Time taken to resolve artifacts (mean, and range) | Radiology reported outcome (no artifacts, minimal artifacts, suitable for diagnosis/ non-diagnostic, significantly impaired) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Krischer 2018 | 2 ml Sienna+ 3ml Saline. 5 ml Retroareolar | n=17 evaluable | 13/17 (76.5%) | Not specified | Not specified | Not specified | 3/17 (17.6%) grade 0 – impossible, 10/17 (58.8%) Grade 1-with restriction, 4/17 (23.5%) Grade 2 - without restriction |
| Aribal 2021 | 2 ml Sienna+ 3ml Saline. 5 ml Periareolar | n=25 | 21/25 (84%) | Not specified | Not specified | Not specified | No Artefacts 26/34 (81%) Minimal Artefacts suitable for diagnosis 6/34 (18%) Significantly impaired, but could be diagnostic 1/34 (3%) Non diagnostic, 1/34 (6%) |
| Sentimag IC Study Reference: NCT02336737 IDE No. G140208 | 2ml Subareolar | 147/13 | 12 (92%) | Mean 68 cm^{3} Range 0 cm^{3} - 204 cm^{3} | Yes (8%) No (92%) | Mean: 10 months, Range: 1 month to 15 months | No Artifacts 1/13 (8%) Minimal Artifacts suitable for diagnosis 4/13 (31%) |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 50
{50}
| Study [Reference] | Amount of Magtrace Injected | Net Number of Subjects Participated/ Studied | Number (and %) of Subjects with Image Artifacts | Size of the Artifacts (mean, range) | Did Artifacts Resolve Completely (Yes (%), No (%)) | Time taken to resolve artifacts (mean, and range) | Radiology reported outcome (no artifacts, minimal artifacts, suitable for diagnosis/ non-diagnostic, significantly impaired) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | | | | Significantly impaired but could be diagnostic 5/13 (39%) Non diagnostic, 3/13 (23%) |
| Shrotria, Stuart European Journal of Surgical Oncology 46 (2020) P136 Abstract 121. | 2ml Periareolar | 160/14 | Not confirmed, assumed 100% | Not known | Not confirmed, assumed No (100%) | Not known | No Artifacts 0/14 (0%) Minimal Artifacts suitable for diagnosis 11/14 (78%) Significantly impaired 0/14 (0%) Non diagnostic, 3/14 (21%) |
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 51
{51}
# 5. Published literature studies
Further studies: Seven (7) European studies have been carried out for which the data are published. These are summarized in the Table 34 along with the supporting publications (Note: The French NCT01790399 Study is the Houpeau study in Table 34).
Table 34: Summary of Published European Studies
| Author | Douek^{7} | Thill^{8} | Rubio^{10} | Ghilli^{11} | Houpeau^{12} | Pinero^{13} | Karakatsanis^{14} |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Centers | 7 | 4 | 1 | 3 | 4 | 9 | 7 |
| Locations | UK, Netherlands | Germany, Poland, Switzerland | Spain | Italy | France | Spain | Sweden, Denmark |
| Patients enrolled | 160 | 150 | 100 | 185 | 108 | 181 | 206 |
| Control technique | Isotope + Blue dye | Isotope | Isotope | Isotope | Isotope + Blue dye | Isotope | Isotope + Blue dye |
| Per patient detection rate (proportion of patients in whom at least one node is found) | | | | | | | |
| Test: | 94.4% | 98.0% | 96.0% | 98.4% | 97.2% | 97.8% | 97.6% |
| | 151/160 | 147/150 | 96/100 | 182/185 | 105/108 | 177/181 | 201/206 |
| Control: | 95.0% | 97.3% | 93.0% | 97.8% | 95.4% | 98.3% | 97.1% |
| | 152/160 | 146/150 | 93/100 | 181/185 | 103/108 | 178/181 | 200/206 |
| Per node detection rate: (Proportion of total nodes found) | | | | | | | |
| Test: | 80.0% | 97.3% | N/A | 95.0% | 97.2% | 91.0% | 93.3% |
| | 323/404 | 283/291 | | 342/360 | 208/214 | 292/321 | 376/403 |
| Control: | 73.5% | 91.8% | N/A | 94.2% | 90.2% | 86.3% | 91.3% |
| | 297/404 | 267/291 | | 339/360 | 193/214 | 277/321 | 368/403 |
| Mean nodes detected per patient: | | | | | | | |
| Test: | 2.0 | 1.9 | 2.2 | 1.8 | 1.9 | 1.6 | 1.8 |
| Control: | 1.9 | 1.8 | 1.77 | 1.8 | 1.8 | 1.5 | 1.8 |
Skin staining was observed only in the MONOS study (Karakatsanis study in Table 34, above) in which two (2) of a total of 57 mastectomy patients who had received Sienna+ (Magtrace®) showed signs of skin staining. Skin staining was resolved in both patients in 3 months post-surgery. The first subject had received subcutaneous, periareolar injection subsequent to which she had undergone skin sparing mastectomy. The position of the stain was towards upper outer quadrant and
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 52
{52}
the size of the stain 1 x 2 cm. The stain had disappeared after 3 months. The second subject also received subcutaneous, periareolar injection subsequent to which she underwent classic mastectomy. The position of the stain was also towards upper outer quadrant and the size of the stain 1 x 2 cm. In this case the stain also disappeared after 3 months.
In the SentimagIC pivotal study (G140208, NCT02336737) where 43 of 147 patients underwent mastectomy, no skin staining was observed at follow-up visit between 6-22 days post-surgery and the surgeons did not record the type of mastectomy. No subsequent follow up was recorded.
Tables 35 to 38 summarize skin staining in patients for a number of studies.
PMA P160053/S002: FDA Summary of Safety and Effectiveness Data
Page 53
{53}
**Table 35 Skin Staining with All Injection Methods**
| Study [Reference] | Amount of Magtrace Injected and Location (e.g., sub-areolar (SA), peri-areolar (PA), retro-areolar (RA), peri-tumoral (PT), etc.) 5ml = 2ml Sienna+ diluted with 3ml saline | Net Number of Subjects Participated/ Studied | Numbers of lumpectomy vs mastectomy procedures | Number of Subjects with Skin Staining | Did Skin Staining Resolve Completely (Yes (%), No (%)) As % of patients with staining | Time taken to resolve skin staining (mean, and range) | Patient reported outcome (no-problem/problem-not important/problem-important/unavailable) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Rubio 2015 | 5ml SA | 120 | n=103 lumpectomy n=17 mastectomy | 20% no timepoint specified | Not stated specifically 'The tattoo usually vanishes over time.' | Not stated specifically 'In our study patients still have some discoloration after 6 months of the injection. This tattoo is similar to the one produced by the blue dye in terms of numbers and time of stain' | Unavailable |
| Karakatsanis 2016 | 5ml SA | 206 Follow up for 186 | n=154…