EDWARDS SAPIEN 3 and SAPIEN 3 Ultra Transcatheter Heart Valve System

P140031S112 · Edwards Lifesciences, LLC · NPT · Sep 9, 2020 · Cardiovascular

Device Facts

Record IDP140031S112
Device NameEDWARDS SAPIEN 3 and SAPIEN 3 Ultra Transcatheter Heart Valve System
ApplicantEdwards Lifesciences, LLC
Product CodeNPT · Cardiovascular
Decision DateSep 9, 2020
DecisionAPPR
Device ClassClass 3
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
P140031S112 · Sep 9, 2020EDWARDS SAPIEN 3 and SAPIEN 3 Ultra Transcatheter Heart Valve SystemEdwards Lifesciences, LLCSociety of Thoracic Surgeons (STS)/American College of Cardiology (ACC) Transcatheter Valve Therapy (TVT) RegistryThe applicant used a retrospective database extract from the TVT Registry to evaluate the safety and effectiveness of the SAPIEN 3 THV in patients with failing transcatheter bioprosthetic aortic valves, serving as the primary clinical evidence for the PMA supplement.TVT Registry; Retrospective cohort; THV-in-THV; Real-world clinical practice

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
TVT Registry Analysis; Retrospective analysis of registry data; Follow-up/Duration: 1 year; Study Period: August 4, 2015 to June 9, 2018Patients with symptomatic heart disease due to failing transcatheter bioprosthetic aortic valve at high or greater risk for SAVR; Sample Size: 263; Number of Sites: 138Not applicable for this studyAll-cause mortality, stroke/TIA, valve reinterventions, valve performance (echocardiographic), NYHA classification, KCCQ score

Indications for Use

The Edwards SAPIEN 3 and SAPIEN 3 Ultra THV System is indicated for patients with symptomatic heart disease due to failing (stenosed, insufficient, or combined) of a surgical or transcatheter bioprosthetic aortic valve or a surgical bioprosthetic mitral valve who are judged by a heart team, including a cardiac surgeon, to be at high or greater risk for open surgical therapy (i.e., predicted risk of surgical mortality ≥ 8% at 30 days, based on the Society of Thoracic Surgeons (STS) risk score and other clinical co-morbidities unmeasured by the STS risk calculator).

Device Story

Transcatheter heart valve (THV) system for percutaneous replacement of failing bioprosthetic valves; balloon-expandable cobalt-chromium frame with bovine pericardial tissue leaflets; PET internal/external skirts for paravalvular leak reduction. Delivered via Commander or Certitude delivery systems; requires vascular access (e.g., transfemoral). Operated by heart team in clinical setting. Provides hemodynamic improvement for high-risk patients; reduces need for open-heart surgery. Output visualized via echocardiography; clinical decision-making guided by valve performance metrics and patient functional status (NYHA class, KCCQ scores).

Clinical Evidence

Retrospective analysis of 263 patients from the STS/ACC TVT Registry (THV-in-THV cohort). Primary endpoints: all-cause mortality (5.8% at 30 days, 18.2% at 1 year), stroke, valve reintervention, and hemodynamic performance. Mean aortic gradient decreased from 29.4 mmHg to 13.3 mmHg at 1 year. NYHA I/II status improved from 12.3% to 88.4% at 1 year. KCCQ scores increased from 38.7 to 73.7. Data supports safety and effectiveness for the expanded indication.

Technological Characteristics

Balloon-expandable cobalt-chromium (MP35N) frame; trileaflet bovine pericardial tissue valve; PET internal/external skirts. Leaflets treated with Carpentier-Edwards ThermaFix process. Delivery systems include flex-wheel articulation, tapered tips, and radiopaque markers. Compatible with eSheath or Certitude introducer sheaths. Non-patient contacting crimping accessories (polyester polyurethane foam/polyether urethane).

Indications for Use

Indicated for patients with symptomatic heart disease due to failing surgical or transcatheter bioprosthetic aortic valves or surgical bioprosthetic mitral valves, deemed at high or greater risk for open surgical therapy (STS risk score ≥ 8%). Contraindicated in patients unable to tolerate anticoagulation/antiplatelet regimens or those with active infections/endocarditis.

Regulatory Classification

Identification

To replace a patient's aortic heart valve. They are different from the classified device (heart valves) in that they are placed percutaneously and do not require open chest surgery or a cardiotomy for placement.

Reference Devices

Submission Summary (Full Text)

{0} # SUMMARY OF SAFETY AND EFFECTIVENESS DATA ## I. GENERAL INFORMATION | Device Generic Name: | Aortic valve, prosthesis, percutaneously delivered | | --- | --- | | Device Trade Name: | Edwards SAPIEN 3 and SAPIEN 3 Ultra Transcatheter Heart Valve System | | Device Procode: | NPT | | Applicant Name and Address: | Edwards Lifesciences LLC One Edwards Way Irvine, CA 92614 | | Date of Panel Recommendation: | None | | Premarket Approval Application (PMA) Number: | P140031/S112 | | Date of FDA Notice of Approval: | September 9, 2020 | The original PMA of the Edwards SAPIEN 3 Transcatheter Heart Valve (THV) System, P140031, was approved on June 17, 2015, with an indication for patients deemed to be at high or greater risk for surgical aortic valve replacement (SAVR). The indication was later expanded in PMA Supplement P140031/S010 on August 18, 2016, to include patients deemed to be at intermediate risk for SAVR. The indication was further expanded in PMA Supplement P140031/S028 on June 5, 2017 to include patients with a failing (stenosed, insufficient, or combined) surgical bioprosthetic aortic or mitral valve who are deemed to be at high or greater risk for redo SAVR or surgical mitral valve replacement. More recently, in PMA Supplement P140031/S085, which was approved on August 16, 2019, the indication of the SAPIEN 3 and SAPIEN 3 Ultra THV System was expanded to include patients deemed to be at low risk for SAVR. The SSEDs to support the above indications are available on the following FDA websites and are incorporated by reference herein: https://www.accessdata.fda.gov/cdrh\\_docs/pdf14/P140031b.pdf https://www.accessdata.fda.gov/cdrh\\_docs/pdf14/P140031S010b.pdf https://www.accessdata.fda.gov/cdrh\\_docs/pdf14/P140031S028b.pdf https://www.accessdata.fda.gov/cdrh\\_docs/pdf14/P140031S085B.pdf The current Panel Track PMA Supplement expands the indication of the Edwards SAPIEN 3 and SAPIEN 3 Ultra THV System to include patients with a failing transcatheter PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 1 {1} bioprosthetic aortic valve (i.e., THV-in-THV) who are deemed to be at high or greater risk for SAVR. ## II. INDICATIONS FOR USE The Edwards SAPIEN 3 and SAPIEN 3 Ultra THV System is indicated for patients with symptomatic heart disease due to failing (stenosed, insufficient, or combined) of a surgical or transcatheter bioprosthetic aortic valve or a surgical bioprosthetic mitral valve who are judged by a heart team, including a cardiac surgeon, to be at high or greater risk for open surgical therapy (i.e., predicted risk of surgical mortality ≥ 8% at 30 days, based on the Society of Thoracic Surgeons (STS) risk score and other clinical co-morbidities unmeasured by the STS risk calculator). ### III. CONTRAINDICATIONS The Edwards SAPIEN 3 and SAPIEN 3 Ultra THV System are contraindicated in patients who cannot tolerate an anticoagulation/antiplatelet regimen or who have active bacterial endocarditis or other active infections. ### IV. WARNINGS AND PRECAUTIONS The warnings and precautions can be found in the Edwards SAPIEN 3 and SAPIEN 3 Ultra THV System labeling. ### V. DEVICE DESCRIPTION The Edwards SAPIEN 3 THV (model 9600TFX, 20, 23, 26, and 29 mm), as shown in Figure 1, is comprised of a balloon-expandable, radiopaque, cobalt-chromium (MP35N) frame, a trileaflet bovine pericardial tissue valve, a polyethylene terephthalate (PET) internal fabric skirt, and a PET external sealing skirt for reduction of paravalvular regurgitation. The leaflets are treated according to the Carpentier-Edwards ThermaFix process. Figure 1: SAPIEN 3 Transcatheter Heart Valve ![img-0.jpeg](img-0.jpeg) The Edwards SAPIEN 3 Ultra THV (model 9750TFX, 20, 23, and 26 mm), as shown in Figure 2, is a design iteration of the SAPIEN 3 THV, with a knitted outer skirt featuring a velour texture on one side. PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 2 {2} **Figure 2: SAPIEN 3 Ultra Transcatheter Heart Valve** ![img-1.jpeg](img-1.jpeg) The Edwards Commander Delivery System (models 9600LDS20, 9600LDS23, 9600LDS26, and 9600LDS29), as shown in Figure 3, includes a handle that provides a flex wheel for articulation of the flex catheter, a tapered tip at the distal end of the delivery system, a balloon catheter for deployment of the THV, and radiopaque markers. It is used when a long access route is planned. The Commander Delivery System uses the Edwards eSheath Introducer Set (models 914ES and 916ES), which is an off-the-shelf device cleared in 510(k) K200258, to establish vascular access. **Figure 3: Edwards Commander Delivery System** ![img-2.jpeg](img-2.jpeg) The Edwards Certitude Delivery System (models 9630TA20, 9600SDS20, 9630TA23, 9600SDS23, 9630TA26, 9600SDS26, 9630TA29, and 9600SDS29), as shown in Figure 4, includes a handle with a flex wheel for articulation of the balloon catheter and extension tubing. It is used when a short access route is planned. **Figure 4: Edwards Certitude Delivery System** ![img-3.jpeg](img-3.jpeg) The Edwards Certitude Introducer Sheath (models 9600IS18 and 9600IS21), as shown in Figure 5, is intended to be used with the Certitude Delivery System. It has a radiopaque marker for visualization of the sheath tip and non-radiopaque depth markings on the distal end. The proximal end of the introducer sheath includes a flush tube and three hemostasis valves. PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 3 {3} Figure 5: Edwards Certitude Introducer Sheath ![img-4.jpeg](img-4.jpeg) The Qualcrimp crimping accessory, as shown in Figure 6, is a non-patient contacting device that is placed around the THV to protect the leaflets during the crimping process. It is manufactured of tubular polyester polyurethane foam and laminated cylindrically on both the inner and outer surfaces with a polyether urethane material. Figure 6: Qualcrimp Crimping Accessory ![img-5.jpeg](img-5.jpeg) The Edwards Crimper (model 9600CR), as shown in Figure 7, is comprised of various molded plastic components which compress the valve to a controlled aperture. The aperture is created by rotating the handle until it abuts the crimp stopper. The Edwards Crimper is used with a Crimp Stopper to correctly crimp the THV. Figure 7: Edwards Crimper ![img-6.jpeg](img-6.jpeg) ## VI. ALTERNATIVE PRACTICES AND PROCEDURES There are other alternatives for patients with a failing transcatheter bioprosthetic aortic valve, including percutaneous balloon valvuloplasty (BAV) for temporary relief of stenosis, surgical replacement of the degenerated device, and palliative medical therapy without an obstruction-relieving procedure. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle. PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 4 {4} ## **VII. MARKETING HISTORY** The Edwards SAPIEN 3 THV is currently approved in the following foreign countries for the aortic THV-in-THV indication: - Austria - Algeria - Argentina - Bahrain - Belgium - Bosnia and Herzegovina - Brazil - Bulgaria - Canada - Chile - Colombia - Costa Rica - Croatia - Cyprus - Czechia - Denmark - Ecuador - Estonia - Finland - France - Germany - Greece - Hong Kong - Hungary - Iran - Ireland - Israel - Italy - Kuwait - Latvia - Lebanon - Lithuania - Luxembourg - Malta - Netherlands - Oman - Pakistan - Paraguay - Peru - Poland - Portugal - Qatar - Romania - South Africa - Slovakia - Slovenia - South Korea - Spain - Sweden - Taiwan - Thailand - Tunisia - United Arab Emirates - Uruguay - Vietnam It has not been withdrawn from marketing for any reason related to its safety or effectiveness. The SAPIEN 3 Ultra THV has not been marketed in the United States or any foreign country for the aortic THV-in-THV indication. ## **VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH** Below is a list of potential adverse effects (e.g., complications) associated with the use of the device. - Death - Stroke/transient ischemic attack, clusters or neurological deficit - Paralysis - Permanent disability - Respiratory insufficiency or respiratory failure - Hemorrhage requiring transfusion or intervention - Cardiovascular injury including perforation or dissection of vessels, ventricle, atrium, septum, myocardium or valvular structures that may require intervention - Pericardial effusion or cardiac tamponade PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 5 {5} - Thoracic bleeding - Embolization including air, calcific valve material or thrombus - Infection including septicemia and endocarditis - Heart failure - Myocardial infarction - Renal insufficiency or renal failure - Conduction system defect which may require a permanent pacemaker - Arrhythmia - Retroperitoneal bleed - Arteriovenous (AV) fistula or pseudoaneurysm - Reoperation - Ischemia or nerve injury or brachial plexus injury - Restenosis - Pulmonary edema - Pleural effusion - Bleeding - Anemia - Abnormal lab values (including electrolyte imbalance) - Hypertension or hypotension - Allergic reaction to anesthesia, contrast media, or device materials - Hematoma - Syncope - Pain or changes at the access site - Exercise intolerance or weakness - Inflammation - Angina - Heart murmur - Fever - Cardiac arrest - Cardiogenic shock - Emergency cardiac surgery - Cardiac failure or low cardiac output - Coronary flow obstruction/transvalvular flow disturbance - Device thrombosis requiring intervention - Valve thrombosis - Device embolization - Device migration or malposition requiring intervention - Left ventricular outflow tract obstruction - Valve deployment in unintended location - Valve stenosis - Structural valve deterioration (wear, fracture, calcification, leaflet tear/tearing from the stent posts, leaflet retraction, suture line disruption of components of a prosthetic valve, thickening, stenosis) - Device degeneration - Paravalvular or transvalvular leak - Valve regurgitation PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 6 {6} - Hemolysis - Device explants - Nonstructural dysfunction - Mechanical failure of delivery system, and/or accessories - Non-emergent reoperation For the specific adverse events that occurred in the real-world clinical practice, please see Section X. # IX. SUMMARY OF PRECLINICAL STUDIES A summary of previously reported preclinical studies can be found in the SSEDs for the original PMA P140031 and PMA Supplement P140031/S028. # X. SUMMARY OF PRIMARY CLINICAL DATA The applicant performed an analysis of the real-world off-label use data captured in the Society of Thoracic Surgeons (STS)/American College of Cardiology (ACC) Transcatheter Valve Therapy (TVT) Registry to establish a reasonable assurance of the safety and effectiveness of the Edwards SAPIEN 3 THV System in patients receiving THV-in-THV treatment. The data from the TVT Registry were the basis of the PMA approval decision. A summary of the clinical data is presented below. Note that the clinical data set did not include the SAPIEN 3 Ultra THV System. However, the results obtained on the Edwards SAPIEN 3 THV System are considered applicable to the Edwards SAPIEN 3 Ultra THV System based on prior demonstration of device comparability in PMA Supplement P140031/S074. # A. Study Design A database extract was performed on August 9, 2019. Patients were excluded if their previous transcatheter aortic valve replacement (TAVR) procedure was performed prior to April 23, 2007 (first implant in the US under the PARTNER trial). The database extract yielded 404 patients that had undergone a THV-in-THV procedure with an Edwards SAPIEN 3 (N=402) or Edwards SAPIEN 3 Ultra (N=2). These patients were treated between August 4, 2015 and July 11, 2019 at 188 participating hospitals. To obtain more complete 1-year follow-up data, a treatment cutoff date of June 9, 2018 was then applied to the data set obtained above, which yielded 263 patients (SAPIEN 3 THV only) treated at 138 participating hospitals. The cutoff date was 14 months before the database extract date, which included a +60-day window for the 1-year visit. These 263 patients constituted the clinical data set used to support this application. # 1. Clinical Inclusion and Exclusion Criteria The initial database extract included all patients who received a commercially available PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 7 {7} Edwards SAPIEN 3 or Edwards SAPIEN 3 Ultra THV in an aortic THV-in-THV procedure. The final data set was a subset of the initial database extract as described above. # 2. Follow-up Schedule All patients were followed post-implantation according to their local standards of care. The TVT Registry collects follow-up data at 30 days and 1 year. # 3. Clinical Endpoints Data entered into the TVT Registry were collected through standardized data collection forms. The endpoints analyzed in this application included: death, stroke/transient ischemic attack (TIA), valve reinterventions, key site reported adverse events, valve performance based on echocardiographic data, New York Heart Association (NYHA) classification, and the Kansas City Cardiomyopathy Questionnaire (KCCQ) score. The analyses in the application focused on 30-day and 1-year time points. # **B. Accountability of PMA Cohorts** At the time of database extract, 242 of the 263 patients were eligible for the 30-day visit and 216 (89.3%) completed the visit within the 30-day follow-up window, defined as the period between 21 days post-procedure and 75 days post-procedure. At 1 year, 200 patients were eligible for the 1-year visit and 136 (68.0%) completed the visit within the follow-up window, defined as the period between 305 days post-procedure and 425 days post-procedure. A detailed summary of the patient accountability at 30 days and 1 year is shown in Table 3. **Table 3: Patient Visit Accountability (AI Population)** | | 30-day Visit | 1-year Visit | | --- | --- | --- | | Total patients | 263 | 263 | | Non-eligible | 21 | 63 | | Death | 19 | 47 | | Withdrawal | 1 | 3 | | Lost to follow-up | 1 | 13 | | Eligible | 242 | 200 | | Follow-up visit completed | 89.3% (216) | 68.0% (136) | | Missed visit | 10.7% (26) | 32.0% (64) | The “Attempted Implant” population consisted of all patients for whom the first vascular access was attempted. The “Valve Implant” population consisted of those patients for whom the valve implant procedure has started and a “No” was indicated for both ‘procedure aborted’ and “conversion to open heart surgery”. The number of patients in the analysis population is shown in Table 4. PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 8 {8} **Table 4: Analysis Populations** | Analysis Population | Number of Patients | | --- | --- | | Attempted Implant Population | 263 | | Valve Implant Population | 261 | ### **C. Study Population Demographics and Baseline Characteristics** The demographics and baseline characteristics of the patients, as shown in Table 5, represent an elderly, multimorbid cohort of patients, consistent with the high operative risk of the population. **Table 5: Patient Demographics and Baseline Characteristics (AI Population)** | Demographics and Baseline Characteristics | Summary Statistics* (N = 263) | | --- | --- | | Age - years | 78.9 ± 10.5 | | Male sex | 55.1% (145/263) | | Society of Thoracic Surgeons (STS) score | 10.2 ± 8.6 (242) | | New York Heart Association (NYHA) class | | | I/II | 12.3% (32/261) | | III/IV | 87.7% (229/261) | | Previous myocardial infarction | 26.6% (70/263) | | Previous intervention | | | Coronary artery bypass grafting (CABG) | 28.1% (74/263) | | Percutaneous coronary intervention (PCI) | 34.0% (89/262) | | Prior aortic valvuloplasty | 13.7% (36/263) | | Stroke or Cerebrovascular accident (CVA) | 18.3% (48/263) | | Peripheral vascular disease (PVD) | 32.1% (84/262) | | Atrial fibrillation/flutter | 48.7% (128/263) | | Permanent pacemaker | 32.2% (84/261) | | Porcelain aorta | 8.0% (21/262) | | Hostile chest | 8.7% (23/263) | | Echocardiographic findings (Valve Implant Population) | | | Valve area (cm^{2}) | 1.0 ± 0.5 (115) | | Mean gradient (mmHg) | 29.4 ± 19.0 (135) | | Mean left ventricular ejection fraction (LVEF), % | 49.3 ± 15.1 (257) | | Moderate or severe aortic regurgitation | 79.3% (207/261) | | Moderate or severe mitral regurgitation | 42.1% (98/233) | \*Continuous measures - Mean ± SD (Total no.); Categorical measures – % (no./Total no.) PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 9 {9} ## D. Safety and Effectiveness Results ### 1. Safety Endpoints The Kaplan-Meier estimates of site-reported adverse events through 1 year are presented in Table 6. The Kaplan-Meier curve for all-cause mortality is shown in Figure 8. The all-cause mortality rate was 5.8% at 30 days and 18.2% at 1 year, including a cardiovascular death rate of 2.7% at 30 days and 5.4% at 1 year. Other relatively more frequent adverse events included conduction/native pacer disturbance requiring pacer (8.2% at 30 days and 10.6% at 1 year) and valve-related readmission (4.3% at 30 days and 8.6% at 1 year). **Table 6: Site Reported Adverse Events (AI Population)** | Adverse Event | Kaplan-Meier Rate* | | | --- | --- | --- | | | 30 Days (N=263) | 1 Year (N=263) | | All-cause death | 5.8% (15, 15) | 18.2% (45, 45) | | Cardiovascular death | 2.7% (7, 7) | 5.4% (13, 13) | | All stroke | 2.3% (6, 6) | 2.8% (7, 7) | | Ischemic stroke | 1.9% (5, 5) | 2.4% (6, 6) | | Undetermined stroke | 0.4% (1, 1) | 0.4% (1, 1) | | Transient ischemic attack (TIA) | 0.8% (2, 2) | 1.9% (4, 4) | | Major vascular complication | 0.4% (1, 1) | 1.0% (2, 2) | | Major bleeding | 1.2% (3, 3) | 2.5% (7, 5) | | Myocardial infarction | 0.8% (3, 2) | 3.1% (6, 5) | | New requirement for dialysis | 0.4% (1, 1) | 1.6% (3, 3) | | Conduction/native pacer disturbance requiring pacer | 8.2% (21, 21) | 10.6% (25, 25) | | Conduction/native pacer disturbance requiring implantable cardioverter defibrillator (ICD) | 0.4% (1, 1) | 2.2% (4, 4) | | Aortic valve re-intervention | 0.4% (1, 1) | 1.2% (2, 2) | | Unplanned other cardiac surgery or intervention | 2.4% (6, 6) | 4.5% (9, 9) | | Unplanned vascular surgery or intervention | 1.5% (4, 4) | 2.6% (6, 6) | | Device thrombosis | 0.4% (1, 1) | 1.0% (2, 2) | | Valve-related readmission | 4.3% (11, 11) | 8.6% (22, 18) | \*Kaplan-Meier rate - (no. of events, no. of patients with the event). PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 10 {10} **Figure 8: All-Cause Mortality through 1 Year (AI Population)** ![img-7.jpeg](img-7.jpeg) **Note:** The 95% confidence intervals were calculated without multiplicity adjustment. The adjusted confidence intervals could be wider than presented here. As such, confidence intervals are provided to illustrate the variability only and should not be used to draw any statistical conclusion. ## 2. Effectiveness Endpoints ### Valve Performance The echocardiographic valve performance results are shown in Figures 9-11. The decrease in gradients were sustained through 1 year. The mean aortic gradient decreased from 29.4 mmHg at baseline to 14.4 mmHg at 30 days, which was maintained through 1 year (13.3 mmHg). Moderate or severe total aortic regurgitation was observed in 79.3% of the patients at baseline, which decreased to 4.6% at 30 days and 3.4% at 1 year. The proportion of patients with ≥ moderate paravalvular regurgitation was 4.5% at 30 days and 2.6% at 1 year. PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 11 {11} Figure 9: Mean Aortic Gradient (VI Population) ![img-8.jpeg](img-8.jpeg) Note: Line plot with mean and standard error. The total number of patients at each visit time point only counted the patients with valid values. Figure 10: Total Aortic Regurgitation (VI Population) ![img-9.jpeg](img-9.jpeg) PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 12 {12} **Figure 11: Paravalvular Regurgitation (VI Population)** ![img-10.jpeg](img-10.jpeg) ### NYHA Functional Class The NYHA functional class distributions by visit are presented in Figure 12. At baseline, 87.7% of patients were in NYHA III/IV. At 1 year, the majority (88.4%) of patients were in NYHA I/II. **Figure 12: NYHA Class by Visit (VI Population)** ![img-11.jpeg](img-11.jpeg) PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 13 {13} ## Length of Stay The mean index hospitalization stay was 4.9 days, which included an average of 1.7 days in the intensive care unit (ICU), as summarized in Table 7. **Table 7: Index Hospitalization (AI Population)** | | Length of Stay (days)* | | --- | --- | | Index hospitalization duration | 4.9 ± 0.3 (263) | | Intensive care stay | 1.7 ± 0.2 (255) | *Mean ± SE (Total no.) ## Quality of Life The results for the KCCQ overall summary score are presented in Figure 13. The mean score increased from 38.7 at baseline to 65.3 and 73.7 at 30 days and 1 year, respectively. **Figure 14: KCCQ Overall Summary Score (VI Population)** ![img-12.jpeg](img-12.jpeg) **Note:** Line plot with mean and standard error. The total number of patients at each visit time point only counted patients with valid values. ### 3. Other Study Observations #### Procedural Information The procedural information is summarized in Table 9. General anesthesia was used in the majority (70%) of patients. Conversion to open heart surgery occurred in two patients due to ventricular rupture and annulus rupture, respectively. PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 14 {14} **Table 9: Procedural Data Summary (AI Population)** | Procedural Data | Summary Statistics* | | --- | --- | | Operator reason for procedure | | | Inoperable/extreme risk | 20.8% (54/259) | | High risk | 69.1% (179/259) | | Intermediate risk | 9.7% (25/259) | | Low risk | 0.4% (1/259) | | Implant approach | | | Transfemoral | 95.8% (252/263) | | Transapical | 1.1% (3/263) | | Transaortic | 0.8% (2/263) | | Subclavian/axillary | 0.8% (2/263) | | Transseptal | 0.4% (1/263) | | Transcarotid | 1.1% (3/263) | | Valve size | | | 20 mm | 2.3% (6/263) | | 23 mm | 35.0% (92/263) | | 26 mm | 30.8% (81/263) | | 29 mm | 31.9% (84/263) | | Cardiopulmonary bypass | 0.4% (1/263) | | Cardiopulmonary bypass status | | | Emergent | 100.0% (1/1) | | Cardiopulmonary bypass time, minutes | 254.0 ± NA (1) | | Type of anesthesia | | | General anesthesia | 70.0% (184/263) | | Moderate sedation | 29.7% (78/263) | | Combination | 0.4% (1/263) | | Total procedure time, minutes | 108.5 ± 4.3 (263) | | Device implanted successfully | 98.9% (260/263) | | Procedure aborted | 0.0% (0/263) | | Conversion to open heart surgery | 0.8% (2/263) | | Ventricular rupture | 1 | | Annulus rupture | 1 | | Mechanical assist device in place at start of procedure | 0.4% (1/263) | PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 15 {15} | Procedural Data | Summary Statistics* | | --- | --- | | Catheter-based assist device | 100.0% (1/1) | *Continuous measures - mean ± SE (n); categorical measures - % (no./Total no.) #### 4. Pediatric Extrapolation In this premarket application, existing clinical data were not leveraged to support approval of a pediatric patient population. ### **XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION** In accordance with the provisions of section 515(c)(2) of the Act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Circulatory System Devices panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel. ### **XII. CONCLUSIONS DRAWN FROM THE PRECLINICAL AND CLINICAL STUDIES** #### **A. Effectiveness Conclusions** The patients who underwent aortic THV-in-THV procedures captured in the TVT Registry overall demonstrated clinically significant improvements in valve hemodynamics from baseline to 1 year. On average, the aortic valve pressure gradient decreased from 29.4 mmHg at baseline to 13.3 mmHg at 1 year. The proportion of patients with moderate or severe total aortic regurgitation decreased from 79.3% at baseline to 3.4% at 1 year. The improvements in clinical outcomes were demonstrated in patients' functional status and quality of life. The majority (88.4%) of patients were in NYHA I/II at 1 year as compared to 12.3% at baseline. Similarly, clinically significant improvement was observed in the KCCQ overall summary score, which increased from 38.7 at baseline to 73.7 at 1 year on average. #### **B. Safety Conclusions** The risks of the device are based on nonclinical laboratory studies and clinical data collected in the TVT Registry to support PMA approval as described above. The Kaplan-Meier estimate of the all-cause mortality rate was 5.8% at 30 days (compared to a mean STS score of 10.2) and 18.2% at 1 year. The Kaplan-Meier estimates of all stroke, conduction/native pacer disturbance requiring pacer, and valve-related readmission were 2.8%, 10.6%, and 8.6%, respectively, at 1 year. #### **C. Benefit-Risk Determination** The probable benefits of the aortic THV-in-THV treatment with the SAPIEN 3 THV include PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 16 {16} improved valve hemodynamic performance, improved functional status as measured by the NYHA classification, and improved quality of life as measured by the KCCQ. The probable risks of the aortic THV-in-THV treatment with the SAPIEN 3 THV include procedural and late complications such as death, stroke, and conduction/native pacer disturbance requiring pacer. #### 1. Patient Perspectives This submission did not include specific information on patient perspectives for this device. However, since transcatheter valve replacement with an Edwards SAPIEN 3 THV provides a less invasive alternative to surgical valve replacement, FDA believes many patients and their physicians would prefer the transcatheter valve replacement therapy as an alternative. In conclusion, given the available information above, the data support that for patients with a failing (stenosed, insufficient, or combined) previously implanted transcatheter bioprosthetic aortic valve who are at high or greater risk for reoperative surgical aortic valve replacement, the probable benefits of implanting an Edwards SAPIEN 3 THV outweigh the probable risks. #### **D. Overall Conclusions** The data in this application support the reasonable assurance of safety and effectiveness of the SAPIEN 3 THV in treating patients with symptomatic heart disease due to failing of a transcatheter bioprosthetic aortic valve who are judged by a heart team, including a cardiac surgeon, to be at high or greater risk for open surgical therapy. FDA has determined this conclusion is also applicable to the SAPIEN 3 Ultra THV. ### **XIII. CDRH DECISION** CDRH issued an approval order on September 9, 2020. The final conditions of approval cited in the approval order are described below. The applicant must participate in and support continued surveillance of the Edwards SAPIEN 3 and SAPIEN 3 Ultra THV System used for the aortic THV-in-THV treatment: 1. **SAPIEN 3 and SAPIEN 3 Ultra Aortic THV-in-THV Surveillance:** The applicant has agreed to work with the TVT Registry to ensure that FDA surveillance occurs for commercial uses of the Edwards SAPIEN 3 and SAPIEN 3 Ultra THV System for the aortic THV-in-THV indication. The surveillance is to continue to monitor the performance of the Edwards SAPIEN 3 and SAPIEN 3 Ultra THV System in the real-world setting. It will involve all consecutive patients treated within the first 3 years after the PMA approval that are entered into the TVT Registry (enrollment period). The applicant has also agreed to link the data to the Centers for Medicare and Medicaid Services (CMS) claims database for long-term surveillance of these patients through 10 years post implantation (follow-up duration). This surveillance will monitor the PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 17 {17} following: (1) device success (intra-procedure); (2) all-cause mortality, all stroke, life-threatening/major bleeding, new requirement for dialysis, peri-procedural myocardial infarction, and repeat procedure for valve-related dysfunction (surgical or interventional therapy) at 30 days and 12 months; (3) neurological (non-stroke), vascular complications, and quality of life (KCCQ) outcomes at 30 days and 12 months; and (4) all-cause mortality, all stroke, and repeat procedure for valve-related dysfunction (surgical or interventional therapy) at 2-10 year post implantation. The applicant’s manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820). #### **XIV. APPROVAL SPECIFICATIONS** Directions for use: See final approved labeling (Instructions for Use). Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the final labeling (Instructions for Use). Post-approval Requirements and Restrictions: See Approval Order. PMA P140031/S112: FDA Summary of Safety and Effectiveness Data Page 18
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...