JUVEDERM VOLBELLA XC

P110033S018 · Allergan · LMH · May 31, 2016 · General, Plastic Surgery

Device Facts

Record IDP110033S018
Device NameJUVEDERM VOLBELLA XC
ApplicantAllergan
Product CodeLMH · General, Plastic Surgery
Decision DateMay 31, 2016
DecisionAPPR
Device ClassClass 3
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
P110033S018 · May 31, 2016JUVEDERM VOLBELLA XCAllerganPostmarket surveillance data (global)Postmarket surveillance data from outside the United States were used to identify and report adverse events associated with the device in routine clinical use, including inflammatory reactions, loss of correction, and rare reports of blurry vision.Postmarket surveillance; Global clinical use; Adverse event reporting

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Global Postmarket Surveillance; Postmarket surveillance / observational; Study Period: Post-approval (outside US)Patients treated with JUVÉDERM VOLBELLA® XC outside the United States; Number of Sites: Multiple (Global)Not applicable for this studySafety and adverse event reporting

Indications for Use

JUVÉDERM VOLBELLA® XC injectable gel is indicated for injection into the lips for lip augmentation and for correction of perioral rhytids in adults over the age of 21.

Device Story

JUVÉDERM VOLBELLA® XC is a sterile, biodegradable, viscoelastic, clear, colorless, homogeneous gel implant consisting of crosslinked hyaluronic acid (15 mg/mL) and 0.3% lidocaine in a physiologic buffer. It is injected into the lips and perioral area by a physician to provide volume augmentation and correct rhytids. The device acts as a soft tissue filler. Clinical benefits include improved lip fullness and reduced perioral lines, with effects lasting up to one year. Patients benefit from aesthetic improvement and high satisfaction scores as measured by FACE-Q modules. The device is administered in a clinical setting.

Clinical Evidence

Prospective, double-blind, randomized, multicenter study (IDE #G130054) with 224 subjects (168 treatment, 56 control). Primary endpoint: non-inferiority to Restylane-L in mean change from baseline to month 3 in lip fullness (LFS2 scale). Results: Mean change 1.1 (VOLBELLA) vs 1.0 (control); 80.3% of VOLBELLA subjects achieved ≥1-point improvement. Secondary endpoints included POLSS (perioral lines) and FACE-Q (satisfaction). Safety data collected via 30-day diaries and investigator follow-ups through 12 months. No treatment-related serious adverse events reported.

Technological Characteristics

Crosslinked hyaluronic acid (15 mg/mL) produced by Streptococcus species; 0.3% w/w lidocaine; physiologic buffer. Sterile, nonpyrogenic, viscoelastic gel. Sterilized via moist heat process (autoclave) per ISO 17665-1 (SAL 10⁻⁶). Biocompatibility per ISO 10993.

Indications for Use

Indicated for lip augmentation and correction of perioral rhytids in adults aged 21 and older. Contraindicated in patients with severe allergies (history of anaphylaxis or multiple severe allergies), history of allergies to Gram-positive bacterial proteins, or history of allergies to lidocaine.

Reference Devices

Submission Summary (Full Text)

{0} # SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED) ## I. GENERAL INFORMATION Device Generic Name: Injectable Dermal Filler Device Trade Name: JUVÉDERM VOLBELLA® XC Device Procode: LMH Applicant's Name and Address: Allergan 2525 Dupont Drive Irvine, CA 92612 Date(s) of Panel Recommendation: None Premarket Approval Application (PMA) Number: P110033/S018 Date of FDA Notice of Approval: 5/31/2016 Priority Review: No Expedited Access Pathway (EAP): No The original PMA for JUVÉDERM VOLUMA XC (P110033) was approved on 10/22/2013 and is indicated for deep (subcutaneous and/or supraperiosteal) injection for cheek augmentation to correct age-related volume deficit in the midface in adults over the age of 21. The SSED to support the indication is available on the CDRH website and is incorporated by reference here. JUVÉDERM VOLBELLA® XC is being submitted as a panel-track supplement (P110033/S018) to the JUVÉDERM VOLUMA XC PMA, P110033, to request changes in design or performance of the device, and a new indication for use of the device. The current supplement was submitted for JUVÉDERM VOLBELLA® XC for injection into the lips for lip augmentation and for correction of perioral rhytids in adults over the age of 21. ## II. INDICATIONS FOR USE JUVÉDERM VOLBELLA® XC injectable gel is indicated for injection into the lips for lip augmentation and for correction of perioral rhytids in adults over the age of 21. ## III. CONTRAINDICATIONS - JUVÉDERM VOLBELLA® XC is contraindicated for patients with severe allergies manifested by a history of anaphylaxis or history or presence of multiple severe allergies - JUVÉDERM VOLBELLA® XC contains trace amounts of Gram-positive bacterial proteins and is contraindicated for patients with a history of allergies to such material PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 1 {1} - JUVÉDERM VOLBELLA® XC contains lidocaine and is contraindicated for patients with a history of allergies to such material # IV. WARNINGS AND PRECAUTIONS The warnings and precautions can be found in the JUVÉDERM VOLBELLA® XC labeling. # V. DEVICE DESCRIPTION JUVÉDERM VOLBELLA® XC injectable gel is a sterile, biodegradable, nonpyrogenic, viscoelastic, clear, colorless, homogeneous gel implant. It consists of crosslinked hyaluronic acid (HA) produced by Streptococcus species of bacteria, formulated to a concentration of 15 mg/mL and 0.3% w/w lidocaine in a physiologic buffer. # VI. ALTERNATIVE PRACTICES AND PROCEDURES There are several other alternatives for lip augmentation and for correction of perioral rhytids. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle. Alternative therapies include autologous fat grafting, surgical facelift, and other soft tissue fillers approved by FDA for lip augmentation. Fat grafting is similar in result and usually requires multiple sessions. Fat grafting requires an invasive procedure to remove fat from the body (such as lipoplasty). Risks of fat grafting include donor site morbidity, graft resorption, fat necrosis, oil cyst, and uneven result. Surgical facelift is not directly comparable, but can decrease an aged look without providing additional volume. Facelift is a surgical procedure and carries the risks typically associated with a surgical procedure requiring general anesthesia and a prolonged recovery with scarring. # VII. MARKETING HISTORY JUVÉDERM VOLBELLA® XC received the CE Mark on October 7, 2011, under the name JUVÉDERM VOLBELLA with Lidocaine. In addition to being marketed throughout Europe, JUVÉDERM VOLBELLA with Lidocaine is currently marketed globally, including Australia, Canada, Brazil, Russia, Ukraine, Mexico, Korea, Taiwan, and Singapore. # VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH Below is a list of the potential adverse effects (e.g., complications) associated with the use of the device. PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 2 {2} Potential adverse effects (e.g., complications) associated with the use of the device, as well as for other devices in the same category, as reported in the clinical study, include tenderness, swelling, firmness (induration), lumps/bumps (mass), bruising, pain, redness, discoloration, and itching. The following reported adverse events were received from postmarket surveillance of the use of JUVÉDERM VOLBELLA® XC for lip augmentation outside the United States and were not observed in the clinical study. These adverse events, with a frequency of 5 or more events, are listed in order of prevalence: inflammatory reaction, loss/lack of correction, hematoma, allergic reaction, infection, paresthesia, herpes, migration, angioedema, and necrosis. In addition, two reports of blurry vision after injection into the highly vascularized periorbital area were received from postmarket surveillance for JUVÉDERM VOLBELLA® XC used outside of the United States. Reported treatment included anti-inflammatories. The outcomes for these two reports were either ongoing or unknown at time of last contact. For the specific adverse events that occurred in the clinical study, please see Section X below. ## IX. SUMMARY OF PRECLINICAL STUDIES ### A. Laboratory Studies #### Physical and Chemical Characterization JUVÉDERM VOLBELLA® XC has been extensively tested and characterized through physical and chemical analyses (Table 1). Degradation assays were also performed to ensure that JUVÉDERM VOLBELLA® XC naturally degrades in the body during its clinical lifespan. **Table 1: Summary of Key Bench Testing on JUVÉDERM VOLBELLA® XC** | Test | Purpose | Results | | --- | --- | --- | | NaHA Concentration (mg/g) | Ensures HA concentration meets specification | Passed | | Lidocaine Concentration (%w/w) | Ensures lidocaine concentration meets specification | Passed | | pH | Ensures pH meets specification | Passed | | Osmolarity (mOsmol/Kg) | Ensures osmolarity meets specification | Passed | | Extrusion Force (N) | Ensures extrusion force meets specification | Passed | | Residual Crosslinker (ppm) | Ensures residual crosslinker meets specifications | Passed | | Endotoxin (EU/Syringe) | Ensures endotoxin meets specification | Passed | PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 3 {3} Filled syringes are sterilized using a validated moist heat process in a pressurized autoclave. The sterilization cycle is validated according to ISO 17665-1 sterilization standard. The validated sterilization cycle provides a minimum Sterility Assurance Level (SAL) of 10⁻⁶. Stability data have been collected through 36 months at 25°C/60% relative humidity, through 12 months at 30°C/65% relative humidity, and through 6 months at 40°C/75% relative humidity. At each time point, product was evaluated for conformance with microbiological, physical, chemical, lidocaine HCl potency, and lidocaine-related degradants. Conformance with all specifications was confirmed. # **Biocompatibility Testing** JUVÉDERM VOLBELLA® XC was evaluated with in vitro and in vivo biocompatibility studies appropriate for devices in contact with tissue for greater than 30 days. The results of the tests are summarized in Table 2 below. The biocompatibility studies were performed in accordance with the Federal Good Laboratory Practices Regulations (21 CFR § 58), ISO 10993 and FDA's Blue Book memorandum G95-1 "Use of ISO-10993 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." Table 2: Summary of Biocompatibility Testing on JUVÉDERM VOLBELLA® XC | Test | Method | ISO standard | Results | | --- | --- | --- | --- | | Cytotoxicity | Agar overlay | 10993-5 | Non-cytotoxic | | Sensitization | Guinea pig maximization test | 10993-10 | Non-sensitizing | | Intracutaneous reactivity | 72 hours in rabbit | 10993-10 | Non-irritant | | Intracutaneous reactivity | 14 days in rabbits | 10993-10 | No acceptance criteria; slight irritant microscopically | | Acute Systemic Toxicity | Intraperitoneal injection in mice | 10993-11 | Not systemically toxic | | Subchronic Toxicity (13 Weeks) | Intradermal injection in rats | 10993-11 | Non-toxic | | Muscle Implantation (4 and 12 Weeks) | In rabbits | 10993-6 | Non-irritant | | Subcutaneous Implantation (4 and 12 weeks) | In rats | 10993-6 | Not causing local skin reaction macroscopically; slight irritant microscopically | | Pyrogenicity | Rabbit pyrogen study | USP <151> | Non-pyrogenic | | Genotoxicity | Bacterial Reverse Mutation, Micronucleus, and Chromosomal Aberration | 10993-3 | Non-genotoxic Non-mutagenic | PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 4 {4} Carcinogenicity risks: The excess cancer risks for JUVÉDERM VOLBELLA® XC range from 6.1 x 10⁻⁵ to 1.6 x 10⁻⁸ from lifetime exposure to residual BDDE based on a linear extrapolation method and a dose-response model. The excess cancer risks for JUVÉDERM VOLBELLA® XC are in the same range of acceptable cancer risks as other previously approved dermal filler products. # X. SUMMARY OF PRIMARY CLINICAL STUDIES The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness for JUVÉDERM VOLBELLA® XC for injection into the lips for lip augmentation and for correction of perioral rhytids in adults over the age of 21 in the US under IDE # G130054. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below. # A. Study Design Subjects were treated between November 25, 2013 and April 23, 2015. The database for this Panel Track Supplement reflected data collected through July 21, 2015 and included 224 subjects who were randomized and underwent treatment with either JUVÉDERM VOLBELLA® XC (N = 168) or control (N = 56) at the outset of the study. The control treatment was Restylane-L which is a legally marketed alternative with similar indications for use. There were 13 investigational sites. The study was a prospective, double-blind, randomized, controlled, multicenter clinical study of subjects seeking lip augmentation and correction of perioral rhytids. Subjects meeting inclusion/exclusion criteria were randomized 3:1 to treatment with either JUVÉDERM VOLBELLA® XC or control. # 1. Key Clinical Inclusion and Exclusion Criteria Inclusion criteria: - 22 years of age or older - Had a score of Minimal, Mild, or Moderate on the LFS2, as agreed upon by the Treating and Evaluating Investigators, and desired at least 1-point improvement in overall LFS2 score OR had Fitzpatrick skin phototype V or VI and had an overall lip fullness score of Marked or Very Marked on the LFS2, as agreed upon by the Treating and Evaluating Investigators, and desired treatment to the vermilion body of 1 or both lips - For treatment of perioral lines, had a POLSS score of Severe or Moderate as agreed upon by the Treating and Evaluating Investigators (subjects who met all other criteria were eligible for treatment in the vermilion body, oral commissures, vermilion border, Cupid's bow, and philtral columns) Exclusion criteria: PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 5 {5} - • Had dentures or any device covering all or part of the upper palate, and/or severe malocclusion or dentofacial or maxillofacial deformities as judged by the Treating Investigator - • Had received permanent facial implants (e.g., polymethylmethacrylate, silicone, polytetrafluoroethylene) anywhere in the face or neck, or was planning to be implanted with any of these products during the study - • Had undergone semipermanent dermal filler treatment (e.g., calcium hydroxylapatite, poly-L-lactic acid) in the lower face (below the orbital rim) within 24 months before enrollment or was planning to undergo such treatment during the study - • Had undergone temporary dermal filler treatment (e.g., hyaluronic acid or collagen) in the lower face (below the orbital rim) within 12 months before enrollment or was planning to undergo such treatment during the study - • Had undergone facial tissue augmentation with fat injections, botulinum toxin injections in the lower face (below the orbital rim), mesotherapy, or cosmetic procedures in the face or neck (e.g., face-lift, laser, photo-modulation, intense pulsed light, radio frequency, dermabrasion, moderate or greater depth chemical peel, or other ablative procedures) within 6 months before enrollment or was planning to undergo any of these procedures during the study - • Had used any lip plumping products within 10 days before enrollment or was planning to use such products during the study (study treatment was able to be delayed as necessary to accommodate this 10-day washout period) - • Had begun using any over-the-counter or prescription, oral or topical, anti-wrinkle products for the lips or around the mouth within 90 days before enrollment or was planning to begin using such products during the study (subjects who had been on a regimen of such products for at least 90 days were eligible for the study if they intended to continue their regimen throughout the study) - • Was on an ongoing regimen of anti-coagulation therapy (e.g., warfarin) or nonsteroidal anti-inflammatory drugs (e.g., aspirin, ibuprofen) or other substances known to increase coagulation time (e.g., herbal supplements with garlic or gingko) within 10 days of undergoing study device injections (study treatment was able to be delayed as necessary to accommodate this 10-day washout period) - • Was on a concurrent regimen of lidocaine or structurally related local anesthetics (e.g., bupivacaine) - • Had a history of anaphylaxis, atopy, or allergy to lidocaine, hyaluronic acid products, or Streptococcal protein, or was planning to undergo desensitization therapy during the study - • Had an active inflammation, infection, cancerous or precancerous lesion, or unhealed wound in the mouth area - • Had porphyria - • Had impaired cardiac conduction, severely impaired hepatic function, or severe renal dysfunction PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 6 {6} - Had any uncontrolled disease - Had severe cardiovascular disease - Females who were pregnant, nursing, or planning a pregnancy # 2. Follow-up Schedule In the pivotal study, subjects were randomized and underwent treatment with either JUVÉDERM VOLBELLA® XC or control at the outset of the study. Treatment occurred at up to 2 visits (initial and optional touch-up) approximately 30 days apart. Subjects completed a safety diary for 30 days after each treatment and attended safety follow-up visits at 3 and 14 days after each treatment. The follow-up period consisted of safety and effectiveness follow-up visits at 1, 3, 6, 9, and 12 months after the last treatment. Subjects were then eligible for a repeat treatment with JUVÉDERM VOLBELLA® XC, with post-treatment follow-up for 1 month after repeat treatment, at which time all subjects completed the study. The Treating Investigator determined the appropriate volume of VOLBELLA® XC or control to inject at initial, touch-up, and repeat treatments based on his/her clinical experience. Subjects received VOLBELLA® XC or control injections based on the randomization assignment. The maximum volume allowed for each lip was 1.5 mL at each treatment. Additional product was permitted to be injected into the perioral area as long as the total injection volume for an individual subject for the initial and touch-up treatments combined did not exceed 6.0 mL, and the volume of VOLBELLA® XC for the repeat treatment did not exceed 6.0 mL. # 3. Clinical Endpoints With regards to safety: the safety of JUVÉDERM VOLBELLA® XC in the lips and perioral area was evaluated by the presence, location, frequency, severity, and duration of ISRs after each treatment (initial, touch-up, and repeat) and any AEs throughout the study. ISRs were assessed by a subject safety diary for 30 days after each treatment. ISRs lasting beyond the 30-day diaries were considered adverse events (AEs). AEs were also reported by the Evaluating Investigator at follow-up visits. With regards to effectiveness: the primary effectiveness endpoint was the analysis of non-inferiority of JUVÉDERM VOLBELLA® XC relative to control in terms of change from baseline to month 3 in mean lip fullness based on Evaluating Investigator assessments using the validated 5-Point Lip Fullness Scale 2 (Table 3, Figure 1). Secondary endpoints included an analysis of the effectiveness of JUVÉDERM VOLBELLA® XC in the perioral region at month 3 based on the number and percent of responders on the POLSS (defined as a subjects demonstrating ≥ 1-point improvement from baseline) and the mean and 95% confidence interval for the change from baseline to month 3 in scores for the Satisfaction with Lips module of the FACE-Q questionnaire. PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 7 {7} With regard to success/failure criteria, the effectiveness of JUVÉDERM VOLBELLA® XC was demonstrated as noninferior to control if the lower limit of the 95% confidence interval for the difference in mean change in LFS2 from baseline to month 3 (JUVÉDERM VOLBELLA® XC minus control) was above the pre-specified noninferiority margin of -0.5. Table 3: 5-Point Lip Fullness Scale 2 (LFS2) | Score | Grade | Description | | --- | --- | --- | | 4 | Very Marked | Very significant red lip show, lower lip pout, and upper lip pout | | 3 | Marked | Significant red lip show and lower lip pout | | 2 | Moderate | Moderate red lip show with slight lower lip pout | | 1 | Mild | Some red lip show; no lower lip pout | | 0 | Minimal | Flat or nearly flat contour, minimal red lip show | Figure 1: Allergan Lip Fullness Scale 2 ![img-0.jpeg](img-0.jpeg) ### B. Accountability of PMA Cohort The subject disposition is depicted as in Figure 2. A total of 268 subjects enrolled in the PMA study. After enrollment, 42 subjects (15.7%) were screen failures, and 1 subject (0.4%) withdrew consent prior to randomization resulting in 225 subjects (84.0%) randomized to treatment. Of the 225 randomized subjects, 169 were randomized to treatment with JUVÉDERM VOLBELLA® XC and 56 were randomized to treatment with control. One subject in the JUVÉDERM VOLBELLA® XC group (1/169, 0.6%) was randomized but discontinued before treatment. Thus, the mITT and safety populations included 224 subjects: 168 subjects in the JUVÉDERM VOLBELLA® XC group and 56 subjects in the control group. Of the 224 mITT subjects, 201 completed PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 8 {8} follow-up through 12 months: 150 (88.8%) and 51 (91.1%) subjects in the JUVÉDERM VOLBELLA® XC and control groups, respectively. Repeat treatment with JUVÉDERM VOLBELLA® XC was administered to 168 subjects in total: 124 (73.4%) and 44 (78.6%) subjects in the JUVÉDERM VOLBELLA® XC and control groups (based on original randomization), respectively. Of the 224 mITT subjects, 167 (74.6%) completed the study: 123 (73.2%) and 44 (78.6%) subjects in the JUVÉDERM VOLBELLA® XC and control groups, respectively. The reasons for early discontinuation in the JUVÉDERM VOLBELLA® XC group were refusal of repeat treatment (11.9%, 20/168), loss to follow-up (7.1%, 12/168), withdrew consent (4.8%, 8/168), other (1.8%, 3/168), AE (0.6%, 1/168), and protocol violations (0.6%, 1/168). The most common reason given for refusal of repeat treatment in the JUVÉDERM VOLBELLA® XC group was satisfaction with current lip fullness. In the control group, the reasons for early discontinuation were refusal of repeat treatment (7.1%, 4/56), loss to follow-up (7.1%, 4/56), withdrew consent (1.8%, 1/56), other (3.6%, 2/56), and AE (1.8%, 1/56). PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 9 {9} Figure 2: Disposition of subjects ![img-1.jpeg](img-1.jpeg) ᵃ Enrolled population consists of all enrolled subjects screened for enrollment and recorded with screening log data. A subject is considered enrolled when s/he has signed the consent form. ᵇ The safety population consists of all subjects who received at least one study treatment. The modified intent-to-treat (mITT) population consists of all subjects who were randomized and received at least one study treatment. ᶜ The per-protocol (PP) population consists of all mITT subjects who do not have any significant protocol deviations that affect the primary effectiveness endpoint. PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 10 {10} ### C. Study Population Demographics and Baseline Parameters The demographics of the study population are typical for a pivotal study performed in the US. Subject demographics and pretreatment characteristics of the JUVÉDERM VOLBELLA® XC and control groups are presented in Table 4. Table 4: Subject Demographics and Pretreatment Characteristics (N = 224) | | **JUVÉDERM VOLBELLA® XC** (N = 168) % (n/N) | **Control** (N = 56) % (n/N) | | --- | --- | --- | | **Gender** | | | | Female | 97.6% (164/168) | 94.6% (53/56) | | Male | 2.4% (4/168) | 5.4% (3/56) | | **Age** | | | | Median | 53 | 55 | | Range | 22-78 | 23-75 | | **Race** | | | | Caucasian | 85.7% (144/168) | 87.5% (49/56) | | African-American | 8.9% (15/168) | 10.7% (6/56) | | Asian | 1.8% (3/168) | 1.8% (1/56) | | American Indian or Alaska Native | 1.2% (2/168) | 0% (0/56) | | Other | 2.4% (4/168) | 0% (0/56) | | **Ethnicity** | | | | Not Hispanic or Latino | 92.9% (156/168) | 91.1% (51/56) | | Hispanic or Latino | 7.1% (12/168) | 8.9% (5/56) | | **Fitzpatrick Skin Type** | | | | I | 10.7% (18/168) | 10.7% (6/56) | | II | 29.8% (50/168) | 26.8% (15/56) | | III | 31.5% (53/168) | 33.9% (19/56) | | IV | 15.5% (26/168) | 16.1% (9/56) | | V | 6.5% (11/168) | 5.4% (3/56) | | VI | 6.0% (10/168) | 7.1% (4/56) | | **Baseline Overall Lip Fullness (LFS2) Score** | | | | Very Marked | 0% (0/168) | 0% (0/56) | | Marked | 1.8% (3/168) | 5.4% (3/56) | | Moderate | 35.1% (59/168) | 26.8% (15/56) | | Mild | 43.5% (73/168) | 48.2% (27/56) | | Minimal | 19.6% (33/168) | 19.6% (11/56) | The median volume for initial and touch-up treatment combined was 2.6 mL (range 0.5 to 6.0 mL). The median volume for repeat treatment was 1.6 mL (range 0.3 to 4.0 mL). For each treatment, similar total volumes were used in each group. PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 11 {11} ## **D. Safety and Effectiveness Results** ### **1. Safety Results** The analysis of safety was based on the cohort of 224 patients, etc. available for the 13 month evaluation. The key safety outcomes for this study are presented below in Tables 5 to 7. Adverse effects are reported in Tables 8 to 12. #### Injection Site Responses Injection site responses (ISRs) were assessed by a subject safety diary for 30 days after initial treatment, touch-up treatment (if performed), and repeat treatment. The severity and duration of all ISRs reported by > 5% of subjects who completed post-treatment diary forms after initial treatment are summarized in Table 5 and Table 6, respectively. Table 7 shows the severity and duration of all ISRs after repeat treatment with Juvéderm VOLBELLA® XC reported by > 5% of subjects. Nearly all subjects (97.4%, 150/154 after initial treatment and 90.2%, 111/123 after repeat treatment) treated with Juvéderm VOLBELLA® XC reported at least 1 ISR. The most frequently reported ISRs after initial and repeat treatment were swelling (92.9%, 143/154, and 87.8%, 108/123, respectively), tenderness (89.6%, 138/154, and 83.7%, 103/123, respectively), firmness (89.0%, 137/154, and 80.5%, 99/123, respectively), and bruising (89.0%, 137/154, and 77.2%, 95/123, respectively). Other common ISRs were lumps/bumps (87.7%, 135/154), redness (83.1%, 128/154), and pain (80.5%, 124/154). Subjects reported the severity of their ISR after initial treatment as mild (14.7%, 22/150), moderate (45.3%, 68/150), or severe (40.0%, 60/150). Most ISRs lasted less than 2 weeks after initial (59.3%, 89/150) and repeat treatment (73.9%, 82/111), but 40.7% of the ISR lasted between 15-30 days of duration. The most common ISRs that lasted for 15 to 30 days after initial treatment were lumps/bumps (33.3%, 45/135), Dryness (25.0%, 2/8) and firmness (15.3%, 21/137). Similarly after repeat treatment, the most common ISRs lasting 15 to 30 days were lumps/bumps (22.4%, 22/98), firmness (13.1%, 13/99), and tenderness (3.9%, 4/103). After repeat treatment with JUVÉDERM VOLBELLA® XC, Similar types of ISRs were reported after repeat treatment for subjects in the JUVÉDERM VOLBELLA® XC (90.2%, 111/123) and control group (90.9% [40/44]). Subjects in JUVÉDERM VOLBELLA® XC and control group reported the severity of their ISR after repeat treatment as mild (18.9%, 21/111, and 12.5%, 5/40, respectively), moderate (44.1%, 49/111, and 55.0%, 22/40, respectively), or severe (36.9%, 41/111, 32.5%, 13/40, respectively). There were no significant differences in ISRs reported between JUVÉDERM VOLBELLA® XC and control. The incidence, severity, and duration of ISRs reported after the touch-up and repeat treatments were generally lower than those reported after initial treatment. PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 12 {12} **Table 5: Severity of Injection Site Responses after Initial Treatment$^{a}$ Occurring in > 5% of Treated Subjects** | Injection Site Response | JUVÉDERM VOLBELLA^{®} XC | | | | Control | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Incidence (n/N^{b}) | Severity^{c} | | | Incidence (n/N^{b}) | Severity^{c} | | | | | | Mild | Moderate | Severe | | Mild | Moderate | Severe | | Any ISR | 97.4% (150/154) | 14.7% (22/150) | 45.3% (68/150) | 40.0% (60/150) | 98.0% (50/51) | 8.0% (4/50) | 44.0% (22/50) | 48.0% (24/50) | | Swelling | 92.9% (143/154) | 23.1% (33/143) | 49.7% (71/143) | 27.3% (39/143) | 98.0% (50/51) | 16.0% (8/50) | 46.0% (23/50) | 38.0% (19/50) | | Tenderness | 89.6% (138/154) | 53.6% (74/138) | 32.6% (45/138) | 13.8% (19/138) | 92.2% (47/51) | 23.4% (11/47) | 66.0% (31/47) | 10.6% (5/47) | | Firmness | 89.0% (137/154) | 32.8% (45/137) | 48.2% (66/137) | 19.0% (26/137) | 92.2% (47/51) | 25.5% (12/47) | 59.6% (28/47) | 14.9% (7/47) | | Bruising | 89.0% (137/154) | 35.0% (48/137) | 40.9% (56/137) | 24.1% (33/137) | 90.2% (46/51) | 30.4% (14/46) | 47.8% (22/46) | 21.7% (10/46) | | Lumps/Bumps | 87.7% (135/154) | 43.0% (58/135) | 42.2% (57/135) | 14.8% (20/135) | 90.2% (46/51) | 30.4% (14/46) | 52.2% (24/46) | 17.4% (8/46) | | Redness | 83.1% (128/154) | 47.7% (61/128) | 39.1% (50/128) | 13.3% (17/128) | 88.2% (45/51) | 40.0% (18/45) | 44.4% (20/45) | 15.6% (7/45) | | Pain | 80.5% (124/154) | 58.9% (73/124) | 30.6% (38/124) | 10.5% (13/124) | 92.2% (47/51) | 42.6% (20/47) | 46.8% (22/47) | 10.6% (5/47) | | Discoloration | 41.6% (64/154) | 54.7% (35/64) | 34.4% (22/64) | 10.9% (7/64) | 49.0% (25/51) | 40.0% (10/25) | 36.0% (9/25) | 24.0% (6/25) | | Itching | 30.5% (47/154) | 76.6% (36/47) | 17.0% (8/47) | 6.4% (3/47) | 37.3% (19/51) | 63.2% (12/19) | 36.8% (7/19) | 0% (0/19) | | Dryness | 5.2% (8/154) | 37.5% (3/8) | 37.5% (3/8) | 25.0% (2/8) | 3.9% (2/51) | 0% (0/2) | 50% (1/2) | 50% (1/2) | $^{a}$ Does not include data after touch-up treatment $^{b}$ N denotes the number of subjects who recorded in the diaries after initial treatment $^{c}$ Maximum severity reported in the diary. Denominator for percentages by severity is the number of subjects with corresponding ISR PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 13 {13} **Table 6: Duration of Injection Site Responses after Initial Treatment$^{a}$ Occurring in > 5% of Treated Subjects** | Injection Site Response | JUVÉDERM VOLBELLA^{®} XC | | | | | Control | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Incidence (n/N^{b}) | Duration^{c} | | | | Incidence (n/N^{b}) | Duration^{c} | | | | | | | 1-3 Days | 4-7 Days | 8-14 Days | 15-30 Days | | 1-3 Days | 4-7 Days | 8-14 Days | 15-30 Days | | Any ISR | 97.4% (150/154) | 9.3% (14/150) | 30.0% (45/150) | 20.0% (30/150) | 40.7% (61/150) | 98.0% (50/51) | 6.0% (3/50) | 44.0% (22/50) | 8.0% (4/50) | 42.0% (21/50) | | Swelling | 92.9% (143/154) | 46.9% (67/143) | 34.3% (49/143) | 13.3% (19/143) | 5.6% (8/143) | 98.0% (50/51) | 42.0% (21/50) | 36.0% (18/50) | 12.0% (6/50) | 10.0% (5/50) | | Tenderness | 89.6% (138/154) | 47.8% (66/138) | 29.0% (40/138) | 15.2% (21/138) | 8.0% (11/138) | 92.2% (47/51) | 31.9% (15/47) | 36.2% (17/47) | 25.5% (12/47) | 6.4% (3/47) | | Firmness | 89.0% (137/154) | 39.4% (54/137) | 27.0% (37/137) | 18.2% (25/137) | 15.3% (21/137) | 92.2% (47/51) | 29.8% (14/47) | 40.4% (19/47) | 10.6% (5/47) | 19.1% (9/47) | | Bruising | 89.0% (137/154) | 32.8% (45/137) | 49.6% (68/137) | 13.9% (19/137) | 3.6% (5/137) | 90.2% (46/51) | 26.1% (12/46) | 65.2% (30/46) | 8.7% (4/46) | 0% (0/46) | | Lumps/Bumps | 87.7% (135/154) | 24.4% (33/135) | 25.2% (34/135) | 17.0% (23/135) | 33.3% (45/135) | 90.2% (46/51) | 32.6% (15/46) | 23.9% (11/46) | 4.3% (2/46) | 39.1% (18/46) | | Redness | 83.1% (128/154) | 65.6% (84/128) | 28.1% (36/128) | 5.5% (7/128) | 0.8% (1/128) | 88.2% (45/51) | 57.8% (26/45) | 35.6% (16/45) | 6.7% (3/45) | 0% (0/45) | | Pain | 80.5% (124/154) | 75.8% (94/124) | 18.5% (23/124) | 4.8% (6/124) | 0.8% (1/124) | 92.2% (47/51) | 61.7% (29/47) | 31.9% (15/47) | 6.4% (3/47) | 0% (0/47) | | Discoloration | 41.6% (64/154) | 64.1% (41/64) | 26.6% (17/64) | 6.3% (4/64) | 3.1% (2/64) | 49.0% (25/51) | 68.0% (17/25) | 20.0% (5/25) | 4.0% (1/45) | 8.0% (2/25) | | Itching | 30.5% (47/154) | 72.3% (34/47) | 17.0% (8/47) | 8.5% (4/47) | 2.1% (1/47) | 37.3% (19/51) | 78.9% (15/19) | 21.1% (4/19) | 0% (0/19) | 0% (0/19) | | Dryness | 5.2% (8/154) | 37.5% (3/8) | 0% (0/8) | 37.5% (3/8) | 25.0% (2/8) | 3.9% (2/51) | 0% (0/2) | 50.0% (1/2) | 0% (0/2) | 50.0% (1/2) | $^{a}$ Does not include data after touch-up treatment $^{b}$ N denotes the number of subjects who recorded in the diaries after initial treatment $^{c}$ Maximum reported successive occurrence of an ISR. Denominator for percentages by duration is the number of subjects with corresponding ISR PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 14 {14} **Table 7: Severity and Duration of Injection Site Responses after Repeat Treatment with JUVÉDERM VOLBELLA® XC Occurring in > 5% of Treated Subjects** | Injection Site Response | Incidence (n/N^{a}) | Severity^{b} | | | Duration^{c} | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | Mild | Moderate | Severe | 1-3 Days | 4-7 Days | 8-14 Days | 15-30 Days | | Any ISR | 90.2% (111/123) | 18.9% (21/111) | 44.1% (49/111) | 36.9% (41/111) | 18.0% (20/111) | 30.6% (34/111) | 25.2% (28/111) | 26.1% (29/111) | | Swelling | 87.8% (108/123) | 36.1% (39/108) | 41.7% (45/108) | 22.2% (24/108) | 50.9% (55/108) | 33.3% (36/108) | 13.9% (15/108) | 1.9% (2/108) | | Tenderness | 83.7% (103/123) | 47.6% (49/103) | 35.9% (37/103) | 16.5% (17/103) | 52.4% (54/103) | 27.2% (28/103) | 16.5% (17/103) | 3.9% (4/103) | | Firmness | 80.5% (99/123) | 39.4% (39/99) | 39.4% (39/99) | 21.2% (21/99) | 39.4% (39/99) | 22.2% (22/99) | 25.3% (25/99) | 13.1% (13/99) | | Lumps/Bumps | 79.7% (98/123) | 41.8% (41/98) | 39.8% (39/98) | 18.4% (18/98) | 39.8% (39/98) | 22.4% (22/98) | 15.3% (15/98) | 22.4% (22/98) | | Bruising | 77.2% (95/123) | 36.8% (35/95) | 43.2% (41/95) | 20.0% (19/95) | 40.0% (38/95) | 43.2% (41/95) | 16.8% (16/95) | 0.00% (0/95) | | Pain | 72.4% (89/123) | 44.9% (40/89) | 47.2% (42/89) | 7.9% (7/89) | 68.5% (61/89) | 15.7% (14/89) | 12.4% (11/89) | 3.4% (3/89) | | Redness | 69.9% (86/123) | 48.8% (42/86) | 37.2% (32/86) | 14.0% (12/86) | 62.8% (54/86) | 29.1% (25/85) | 7.0% (6/86) | 1.2% (1/86) | | Discoloration | 30.9% (38/123) | 60.5% (23/38) | 31.6% (12/38) | 7.9% (3/38) | 76.3% (29/38) | 21.1% (8/38) | 2.6% (1/38) | 0.00% (0/38) | | Itching | 26.0% (32/123) | 50.0% (16/32) | 46.9% (15/32) | 3.1% (1/32) | 71.9% (23/32) | 18.8% (6/32) | 6.3% (2/32) | 3.1% (1/32) | $^{a}$ N denotes the number of subjects who recorded in the diaries after initial and repeat treatment with JUVÉDERM VOLBELLA® XC $^{b}$ Maximum severity reported in the diary. $^{c}$ Maximum reported successive occurrence of an ISR. ### Adverse Events Adverse events (AEs) were reported by Evaluating Investigators throughout the study: 104/168 (61.9%) of JUVÉDERM VOLBELLA® XC and 37/56 (66.1%) of control subjects experienced at least one treatment emergent adverse events, TEAEs (related and unrelated to the treatment). ISRs that lasted beyond the 30-day diaries were also considered AEs. After initial treatment (or touch-up treatment, if performed), treatment-related AEs were reported in 50.0% (84/168) of subjects treated with JUVÉDERM VOLBELLA® XC and 51.8% (29/56) of subjects treated with control. All the treatment emergent adverse events with an incidence rate > 5% were treatment related. A summary of treatment-related AEs after initial/touch-up treatment is provided in Table 8. The severity and duration of treatment-related AEs reported by > 5% of subjects after initial treatment (or touch-up treatment) are summarized in Table 9 and Table 10. PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 15 {15} Most subjects treated with JUVÉDERM VOLBELLA® XC experienced mild (42.3%, 71/168) or moderate (17.3%, 29/168) treatment-related AEs, only 1.2% (2/168) of subjects experienced severe treatment-related AE. Similar results were observed after treatment with control (41.1% [23/56], 23.2% [13/56], and 7.1% [4/56] of subjects experienced mild, moderate and severe AEs, respectively). Regardless of treatment group, the treatment-related AEs generally required no action to be taken and resolved without sequelae (49.4% [83/168] and 51.8% [29/56] of subjects in VOLBELLA® XC group and in control group, respectively). Treatment-related TEAEs resolved within 7 days of onset for 22.0% (37/168) of subjects, 8-14 days for 11.3% (19/168) of subjects, 15-30 days for 12.5% (21/168) of subjects, and longer than 30 days for 30.4% (51/168) of subjects treated with JUVÉDERM VOLBELLA® XC. Treatment-related TEAEs with a duration of more than 180 days include: injection site mass (3.0%, 5/168 subjects); injection site erythema (1.2%, 2/168); injection site pain, injection site swelling, and wound (puncture marks on right and left sides of upper lip per investigator term) (each 0.6%, 1/168). PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 16 {16} **Table 3: Summary of Treatment-Related Adverse Events after Initial/Touch-up Treatment** | | JUVÉDERM VOLBELLA^{®} XC | | Control | | | --- | --- | --- | --- | --- | | | Subjects | Events | Subjects | Events | | | (N = 168) n (%) | (N = 345) n (%) | (N = 56) n (%) | (N = 111) n (%) | | Overall | 84 (50.0) | 345 (100) | 29 (51.8) | 111 (100) | | At Injection Site | | | | | | Upper Lip | 67 (39.9) | 140 (40.6) | 27 (48.2) | 50 (45.0) | | Lower Lip | 60 (35.7) | 133 (38.6) | 22 (39.3) | 47 (42.3) | | Perioral Lines | 21 (12.5) | 27 (7.8) | 5 (8.9) | 7 (6.3) | | Oral Commissures | 22 (13.1) | 40 (11.6) | 5 (8.9) | 7 (6.3) | | Not at Injection Site | 5 (3.0) | 5 (1.4) | 0 (0.0) | 0 (0.0) | | Duration | | | | | | ≤ 7 Days | 37 (22.0) | 89 (25.8) | 15 (26.8) | 51 (45.9) | | 8-14 Days | 19 (11.3) | 47 (13.6) | 7 (12.5) | 11 (9.9) | | 15-30 Days | 21 (12.5) | 39 (11.3) | 4 (7.1) | 9 (8.1) | | > 30 Days | 51 (30.4) | 167 (48.4) | 15 (26.8) | 36 (32.4) | | Not yet Resolved^{a} | 3 (1.8) | 3 (0.9) | 1 (1.8) | 4 (3.6) | | Severity | | | | | | Mild | 71 (42.3) | 269 (78.0) | 23 (41.1) | 63 (56.8) | | Moderate | 29 (17.3) | 71 (20.6) | 13 (23.2) | 34 (30.6) | | Severe | 2 (1.2) | 5 (1.4) | 4 (7.1) | 14 (12.6) | | Causality | | | | | | Device | 58 (34.5) | 205 (59.4) | 19 (33.9) | 42 (37.8) | | Procedure | 44 (26.2) | 139 (40.3) | 19 (33.9) | 69 (62.2) | | Unknown | 1 (0.6) | 1 (0.3) | 0 (0.0) | 0 (0.0) | | Outcome | | | | | | Resolved without sequelae | 83 (49.4) | 342 (99.1) | 29 (51.8) | 107 (96.4) | | Ongoing^{a} | 3 (1.8) | 3 (0.9) | 1 (1.8) | 4 (3.6) | | Death | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | | Treatment Required | | | | | | No | 78 (46.4) | 308 (89.3) | 25 (44.6) | 91 (82.0) | | Medication | 18 (10.7) | 36 (10.4) | 9 (16.1) | 18 (16.2) | | Procedure | 1 (0.6) | 1 (0.3) | 1 (1.8) | 2 (1.8) | $^{a}$ The ongoing treatment-related AEs were ongoing at study exit. No additional information is available on the resolution of these AEs. PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 17 {17} **Table 4: Severity of Treatment-Related Adverse Events after Initial/Touch-Up Treatment Occurring in > 5% of Treated Subjects** | Adverse Event | JUVÉDERM VOLBELLA® XC (N = 168) | | | | Control (N = 56) | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Subjects^{a} | Mild^{a} | Moderate^{a} | Severe^{a} | Subjects^{a} | Mild^{a} | Moderate^{a} | Severe^{a} | | Injection Site Mass | 32.1% (54/168) | 26.8% | 4.2% | 1.2% | 26.8% (15/56) | 17.9% | 5.4% | 3.6% | | Injection Site Bruising | 17.9% (30/168) | 10.7% | 7.1% | 0% | 19.6% (11/56) | 10.7% | 8.9% | 0% | | Injection Site Pain | 11.9% (20/168) | 7.7% | 4.2% | 0% | 21.4% (12/56) | 8.9% | 8.9% | 3.6% | | Injection Site Induration | 8.9% (15/168) | 7.7% | 0.6% | 0.6% | 7.1% (4/56) | 3.6% | 3.6% | 0% | | Injection Site Swelling | 8.3% (14/168) | 4.8% | 3.6% | 0% | 12.5% (7/56) | 3.6% | 3.6% | 5.4% | | Injection Site Dryness | 4.2% (7/168) | 3.0% | 1.2% | 0% | 5.4% (3/56) | 1.8% | 3.6% | 0% | $^{a}$ The percentages are based on the number of subjects who received treatment with the corresponding product. **Table 5: Duration of Treatment-Related Adverse Events after Initial/Touch-Up Treatment Occurring in > 5% of Treated Subjects** | Adverse Event | JUVÉDERM VOLBELLA® XC | | | | | | Control | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Events % (n/N^{a}) | ≤ 7 Days^{b} % | 8-14 Days^{b} % | 15-30 Days^{b} % | > 30 Days^{b} % | Not yet Resolved % | Events % (n/N^{a}) | ≤ 7 Days^{b} % | 8-14 Days^{b} % | 15-30 Days^{b} % | > 30 Days^{b} % | Not yet Resolved % | | Injection Site Mass | 29.0% (100/345) | 5.0% | 1.0% | 15.0% | 76.0% | 3.0% | 26.1% (29/111) | 10.3% | 3.4% | 3.4% | 69.0% | 13.8% | | Injection Site Bruising | 15.9% (55/345) | 52.7% | 2.9% | 14.5% | 3.6% | 0% | 18.0% (20/111) | 60.0% | 30.0% | 10.0% | 0% | 0% | | Injection Site Pain | 12.2% (42/345) | 50.0% | 9.5% | 4.8% | 35.7% | 0% | 18.9% (21/111) | 85.7% | 9.5% | 0% | 4.8% | 0% | | Induration Site Induration | 12.5% (43/345) | 0% | 4.7% | 0% | 95.3% | 0% | 6.3% (7/111) | 0% | 0% | 0% | 100% | 0% | | Injection Site Swelling | 9.3% (32/345) | 43.8% | 18.8% | 6.3% | 31.3% | 0% | 15.3% (17/111) | 64.7% | 0% | 11.8% | 23.5% | 0% | | Injection Site Dryness | 3.8% (13/345) | 0% | 46.2% | 0% | 53.8% | 0% | 4.5% (5/111) | 60.0% | 0% | 40.0% | 0% | 0% | $^{a}$ N denotes the number of subjects who received treatment with the corresponding product. $^{b}$ The percentages by duration are based on the number of subjects with the corresponding treatment-related adverse event. Treatment-related AEs after initial/touch-up treatment occurring in ≤ 5% of subjects included chapped lips, dizziness, dry lip, general physical condition abnormal, headache, lip disorder (lumps), lip injury, oral herpes, presyncope, wound, and injection site discoloration, edema, erythema, exfoliation, pruritus, reaction, discomfort, hyperaesthesia, nodule, papule, hypoaesthesia, laceration, and paraesthesia. PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 18 {18} After initial/touch-up treatment, some AEs occurred weeks to months after the injection procedure. Treatment-related AEs with onset after 30 days were reported for 7 subjects in the JUVÉDERM VOLBELLA® XC group as follows: - In 1 subject, mild injection site mass occurred between 30 to 60 days after initial/touch-up treatment. This AE required no treatment and resolved without sequelae. - In 3 subjects, mild injection site nodules, edema, mass, and swelling occurred between 60 to 90 days after initial/touch-up treatment. The nodules, edema, and mass required no treatment and resolved without sequelae. The swelling was treated with acetaminophen and also resolved without sequelae. - In 2 subjects, injection site swelling and lip disorder (lumps) occurred between 90 to 180 days after initial/touch-up treatment. The swelling was treated with doxycycline and resolved without sequelae. The lip disorder required no treatment and also resolved without sequelae. - In 2 subjects, injection site mass occurred after 180 days. These AEs required no treatment and resolved without sequelae. After repeat treatment with JUVÉDERM VOLBELLA® XC, treatment-related AEs were reported in 13.7% (17/124) and 15.9% (7/44) of subjects in the JUVÉDERM VOLBELLA® XC and control groups, respectively. A summary of treatment-related AEs after repeat treatment with JUVÉDERM VOLBELLA® XC by randomization group is provided in Table 11. The treatment-related AEs were similar between the two groups after repeat treatment. The severity and duration of AEs reported by > 5% of subjects after repeat treatment are summarized in Table 12. Similar types of treatment-related AEs were reported after repeat treatment, but with a lower incidence rate, severity, and duration compared to initial/touch-up treatment. Treatment-related AEs after repeat treatment occurring in ≤ 5% of subjects included chapped lips, injection site bruising, edema, induration, and pain. There were no treatment-related serious adverse events reported in the study. PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 19 {19} **Table 6: Summary of Treatment-Related Adverse Events after Repeat Treatment** | | JUVÉDERM VOLBELLA® XC | | Control* | | | --- | --- | --- | --- | --- | | | Subjects | Events | Subjects | Events | | | (N = 124) n (%) | (N = 46) n (%) | (N = 44) n (%) | (N = 10) n (%) | | Overall | 17 (13.7) | 46 (100) | 7 (15.9) | 10 (100) | | At Injection Site | | | | | | Upper Lip | 12 (9.7) | 16 (34.8) | 6 (13.6) | 6 (60.0) | | Lower Lip | 12 (9.7) | 17 (37.0) | 3 (6.8) | 3 (30.0) | | Perioral Lines | 7 (5.6) | 7 (15.2) | 0 (0.0) | 0 (0.0) | | Oral Commissures | 3 (2.4) | 4 (8.7) | 1 (2.3) | 1 (10.0) | | Not at Injection Site | 1 (0.8) | 2 (4.3) | 0 (0.0) | 0 (0.0) | | Duration | | | | | | ≤ 7 Days | 7 (5.6) | 12 (26.1) | 0 (0.0) | 0 (0.0) | | 8-14 Days | 2 (1.6) | 3 (6.5) | 1 (2.3) | 1 (10.0) | | 15-30 Days | 3 (2.4) | 4 (8.7) | 1 (2.3) | 1 (10.0) | | > 30 Days | 9 (7.3) | 26 (56.5) | 2 (4.5) | 2 (20.0) | | Not yet Resolved | 1 (0.8) | 1 (2.2) | 4 (9.1) | 6 (60.0) | | Severity | | | | | | Mild | 10 (8.1) | 27 (58.7) | 3 (6.8) | 4 (40.0) | | Moderate | 9 (7.3) | 18 (39.1) | 3 (6.8) | 5 (50.0) | | Severe | 1 (0.8) | 1 (2.2) | 1 (2.3) | 1 (10.0) | | Causality | | | | | | Device | 8 (6.5) | 18 (39.1) | 3 (6.8) | 5 (50.0) | | Procedure | 10 (8.1) | 26 (56.5) | 2 (4.5) | 2 (20.0) | | Unknown | 1 (0.8) | 2 (4.3) | 2 (4.5) | 3 (30.0) | | Outcome | | | | | | Resolved without sequelae | 16 (12.9) | 45 (97.8) | 4 (9.1) | 4 (40.0) | | Ongoing | 1 (0.8) | 1 (2.2) | 4 (9.1) | 6 (60.0) | | Death | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | | Treatment Required | | | | | | No | 15 (12.1) | 43 (93.5) | 7 (15.9) | 10 (100) | | Medication | 2 (1.6) | 3 (6.5) | 0 (0.0) | 0 (0.0) | \*denotes subjects who received initial treatment in control but repeat treatment with JUVÉDERM VOLBELLA® XC PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 20 {20} Table 7: Severity and Duration of Treatment-Related Adverse Events after Repeat Treatment Occurring in > 5% of Treated Subjects | Adverse Event | JUVÉDERM VOLBELLA® XC | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Subjects % (n/N) | Severity | | | Events % (n/N) | Duration | | | | | | | | Mild^{a} | Moderate^{a} | Severe^{a} | | ≤ 7 Days^{b} | 8-14 Days^{b} | 15-30 Days^{b} | > 30 Days^{b} | Not yet Resolved^{b} | | Injection Site Mass | 7.3% (9/124) | 5.6 % | 0.8% | 0.8% | 32.6% (15/46) | 0% | 0% | 13.3% | 80.0% | 6.7% | $^{a}$ The percentages by severity are based on the number of subjects who received repeat treatment. $^{b}$ The percentages by duration are based on the number of events for the corresponding treatment-related adverse event. ### Subgroup analyses Subgroup analyses of ISRs and device-related AEs were performed by Fitzpatrick skin phototype (grouped as I/II, III/IV, and V/VI), age group (< 30, 30 to 50, 51 to 65, and > 65 years), volume injected (< or ≥ the median injected volume of 2.6 mL), and investigational site (Table 13). For ISRs after initial/touch-up treatment, there were no significant differences between the treatment and control groups for any subgroup in terms of overall ISR incidence or individual incidences. ISRs after repeat treatment were similar within each subgroup category. For device-related AEs after initial/touch-up treatment, no significant differences between the treatment and control groups were observed for any subgroup in terms of overall incidence of device-related AEs. Device-related AEs after repeat treatment were similar within each subgroup category. Table 13: Incidence of Device-Related Adverse Events after Initial/Touch-up Treatment by Subgroup | | JUVÉDERM VOLBELLA® XC n/N (%) | Control n/N (%) | | --- | --- | --- | | Fitzpatrick Skin Phototype | | | | I/II | 42.6% (29/68) | 42.9% (9/21) | | III/IV | 43.0% (34/79) | 57.1% (16/28) | | V/VI | 14.3% (3/21) | 14.3% (1/7) | | Age | | | | <30 years | 28.6% (2/7) | 0% (0/1) | | 30-50 years | 35.1% (20/57) | 41.2% (7/17) | | 51-65 years | 40.9% (36/88) | 51.5% (17/33) | | >65 years | 50.0% (8/16) | 40.0% (2/5) | | Median Volume | | | | <Median Volume (2.6 mL) | 31.7% (26/82) | 34.6% (9/26) | | ≥Median Volume (2.6 mL) | 46.5% (40/86) | 56.7% (17/30) | PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 21 {21} | | JUVÉDERM VOLBELLA® XC n/N (%) | Control n/N (%) | | --- | --- | --- | | **Investigational Site** | | | | Site 10001 | 26.7% (4/15) | 66.7% (4/6) | | Site 10002 | 22.2% (2/9) | 25.0% (1/4) | | Site 10003 | 42.9% (3/7) | 50.0% (1/2) | | Site 10004 | 35.3% (6/17) | 40.0% (2/5) | | Site 10005 | 30.8% (4/13) | 0.0% (0/2) | | Site 10006 | 9.1% (1/11) | 0.0% (0/1) | | Site 10007 | 70.6% (12/17) | 83.3% (5/6) | | Site 10008 | 46.7% (7/15) | 50.0% (2/4) | | Site 10009 | 54.5% (6/11) | 71.4% (5/7) | | Site 10010 | 78.6% (11/14) | 50.0% (2/4) | | Site 10012 | 54.5% (6/11) | 25.0% (1/4) | | Site 10013 | 4.5% (1/22) | 20.0% (2/10) | | Site 10014 | 50.0% (3/6) | 100.0% (1/1) | ## 2. Effectiveness Results The analysis of effectiveness was based on the 224 evaluable patients at the 3-month time point. Key effectiveness outcomes are presented in Tables 14 to 15. ### Primary Endpoint The primary analysis was to evaluate the non-inferiority (non-inferiority margin = 0.5 points) of Juvéderm VOLBELLA® XC to control. The primary effectiveness endpoint was the mean change in overall lip fullness from baseline to month 3 based on the Evaluating Investigator’s assessments of overall lip fullness using the validated Allergan Lip Fullness Scale 2 (LFS2). The primary endpoint of the study was met with non-inferiority of Juvéderm VOLBELLA® XC to control. The mean change from baseline to month 3 on the LFS2 score was 1.1 for subjects treated with JUVÉDERM VOLBELLA® XC and 1.0 for subjects treated with control, with 80.3% (122/152) of subjects treated with JUVÉDERM VOLBELLA® XC and 70.8% (34/48) of subjects treated with control showing a ≥ 1 point improvement in overall lip fullness (Table 14). PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 22 {22} **Table 14: Primary Effectiveness Endpoint: Change from Baseline to Month 3 in Overall Lip Fullness (LFS2) Score (Modified Intent-to-treat Population)** | | JUVÉDERM VOLBELLA® XC (N = 165) | Control (N = 53) | P-value | Difference (Lower Limit of 95% Confidence Interval) | | --- | --- | --- | --- | --- | | N | 152 | 48 | | | | Mean | 1.1 | 1.0 | 0.609 | 0.07 | | SD | 0.75 | 0.85 | | (-0.15) | | Median | 1.0 | 1.0 | | | | Min, Max | 0, 4 | 0, 3 | | | | 95% CI | (0.9, 1.2) | (0.8, 1.2) | | | | P-value | < 0.001 | < 0.001 | | | Throughout the follow-up period, JUVÉDERM VOLBELLA® XC continued to provide a clinically significant improvement in lip fullness ($\geq 1$-point mean improvement on the LFS2 scale), with a majority of subjects treated with JUVÉDERM VOLBELLA® XC demonstrating improvement through 1 year (Table 15). **Table 15: Effectiveness Results Beyond Primary Endpoint** | | JUVÉDERM VOLBELLA® XC | | --- | --- | | | % (n/N) | | **6 Months** | 71.1% (106/149) | | **9 Months** | 65.1% (95/146) | | **1 Year** | 61.8% (76/123) | #### Secondary Endpoints Secondary measures included Evaluating Investigators' assessment of subjects' perioral lines using the validated perioral lines severity scale (POLSS) and subjects' satisfaction with their lips using the validated Satisfaction with Lips module of the FACE-Q questionnaire. At 3 months, 65.4% (53/81) of subjects treated with JUVÉDERM VOLBELLA® XC showed a $\geq 1$ point improvement in POLSS scores from baseline at rest. At 1 year, 66.2% (45/68) of subjects treated with JUVÉDERM VOLBELLA® XC maintained improvement in perioral lines severity at rest. At 3 months, 96.1% (147/153) of subjects treated with JUVÉDERM VOLBELLA® XC reported improvement in satisfaction with their lips, based on the Satisfaction with Lips module of the FACE-Q questionnaire, with the mean score increasing from 38.5 at baseline to 76.5. At 1 year, 79.7% PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 23 {23} (98/123) of subjects reported improved satisfaction with their lips over baseline, with a mean score of 59.6. #### Other Effectiveness Assessments Additional effectiveness measures included Evaluating Investigators' assessment of subjects' upper and lower lip fullness using the LFS2, oral commissures severity using the oral commissures severity scale (OCSS), global aesthetic improvement using the global aesthetic improvement scale (GAIS), and Satisfaction with Lip Lines module of the FACE-Q. Improvements in upper and lower lip fullness were similar to the improvements seen in overall lip fullness through 1 year. Subjects treated with JUVÉDERM VOLBELLA® XC in the perioral lines and oral commissures also saw improvement in perioral lines severity at maximal contraction and oral commissures severity through 1 year. On the GAIS at 3 months, 92.9% (143/154) of subjects in the JUVÉDERM VOLBELLA® XC group were scored as 'improved' or 'much improved' in appearance. At 1 year, the percentage of subjects 'improved' or 'much improved' was 58.5% (72/123) in the JUVÉDERM VOLBELLA® XC group. At 3 months, subjects treated with JUVÉDERM VOLBELLA® XC reported a mean score improved to 72.9 from 37.5 at baseline on the Lip Lines module of the FACE-Q. The mean score was 56.3 in the JUVÉDERM VOLBELLA® XC group at 1 year. Within the JUVÉDERM VOLBELLA® XC randomization group, repeat treatment with JUVÉDERM VOLBELLA® XC was administered to 124 subjects. The effectiveness profile after repeat treatment was similar to that after initial treatment, with 94.3% (115/122) of subjects showing at least a 1-point improvement in lip fullness, based on the Evaluating Investigator assessment at 1 month after repeat treatment. ### 3. Subgroup Analyses Subgroup analyses of the primary effectiveness endpoint were performed by gender, age group, baseline overall LFS2 score, Fitzpatrick skin phototype (I/II, III/IV, and V/VI), volume injected (< or ≥ the median total injected volume of 2.6 mL), and investigational site. Similar results were observed in the JUVÉDERM VOLBELLA® XC and control groups within each subgroup (Table 16). Some subgroups (e.g., investigational sites, males, subjects < 30 years of age) had a small number of subjects, such that the analysis cannot support conclusions for those subgroups. PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 24 {24} **Table 16: Subgroup Effectiveness Analyses – Mean (± Standard Deviation) Change from Baseline in Overall Lip Fullness at Month 3 (mITT Population)** | Subgroup | JUVÉDERM VOLBELLA^{®} XC (N = 168) | | Control (N = 56) | | | --- | --- | --- | --- | --- | | | Subgroup N | Result | Subgroup N | Result | | **Baseline overall LFS2 score** | | | | | | Minimal | 32 | 1.7 (0.97) | 10 | 1.8 (0.79) | | Mild | 65 | 1.0 (0.61) | 25 | 1.0 (0.73) | | Moderate | 55 | 0.8 (0.54) | 13 | 0.3 (0.48) | | **Volume injected at initial and touch-up treatment combined** | | | | | | < median (2.6 mL^{a}) | 72 | 0.8 (0.60) | 20 | 0.7 (0.81) | | ≥ median (2.6 mL^{a}) | 80 | 1.3 (0.80) | 28 | 1.3 (0.80) | | **Fitzpatrick skin phototype** | | | | | | I/II | 67 | 1.1 (0.88) | 18 | 1.0 (0.91) | | III/IV | 71 | 1.1 (0.66) | 26 | 1.0 (0.87) | | V/VI | 14 | 0.8 (0.43) | 4 | 0.8 (0.50) | | **Gender** | | | | | | Male | 2 | 0.5 (0.71) | 2 | 0.5 (0.71) | | Female | 150 | 1.1 (0.75) | 46 | 1.0 (0.86) | | **Age Group** | | | | | | < 30 years | 4 | 1.3 (0.50) | 0 | N/A | | 30 to 50 years | 53 | 0.9 (0.68) | 13 | 0.7 (0.75) | | 51 to 65 years | 81 | 1.1 (0.74) | 30 | 1.2 (0.91) | | > 65 years | 14 | 1.4 (1.02) | 5 | 0.8 (0.45) | | **Investigational site** | | | | | | 10001 | 15 | 1.5 (0.52) | 4 | 1.8 (0.50) | | 10002 | 8 | 0.6 (0.52) | 3 | 1.0 (0.00) | | 10003 | 7 | 0.9 (0.38) | 2 | 0.0 (0.00) | | 10004 | 15 | 0.3 (0.46) | 5 | 0.4 (0.55) | | 10005 | 13 | 0.9 (0.28) | 2 | 0.0 (0.00) | | 10006 | 8 | 1.0 (0.76) | 1 | 1.0 (N/A) | | 10007 | 15 | 0.8 (0.41) | 5 | 0.8 (0.45) | | 10008 | 15 | 0.7 (0.59) | 4 | 0.5 (0.58) | | 10009 | 11 | 1.6 (0.81) | 7 | 2.0 (0.82) | | 10010 | 14 | 2.1 (0.83) | 3 | 1.7 (1.15) | | 10012 | 11 | 1.0 (0.45) | 4 | 0.5 (0.58) | | 10013 | 14 | 1.3 (0.91) | 7 | 0.9 (0.69) | | 10014 | 6 | 1.0 (0.00) | 1 | 2.0 (N/A) | $^{a}$ The median total volume injected across all injection sites at the initial and touch-up treatments combined PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 25 {25} ### **E. Financial Disclosure** The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 26 investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data. ## **XI. SUMMARY OF SUPPLEMENTAL CLINICAL INFORMATION** ### **European Clinical Study** In this prospective, randomized, multicenter study, 280 subjects desiring lip volume enhancement were treated with JUVÉDERM VOLBELLA® XC or RESTYLANE-L®. The subjects returned to the investigational sites at quarterly intervals for follow-up evaluations with the Treating Investigator. Subjects could receive repeat treatment with JUVÉDERM VOLBELLA® XC at months 6, 9, or 12 if the investigator determined that the subject's LFS2 score had returned to baseline, or at month 12 if the subject's LFS2 score was lower than the treatment goal as assessed by the investigator. At 1 month after repeat treatment, subjects returned for a final follow-up visit and were then exited from the study. Common and expected ISRs were collected via 30-day diaries after each treatment. The incidence, severity, and duration of ISRs for subjects treated with JUVÉDERM VOLBELLA® XC after initial treatment are shown in Table 17. Most subjects reported an ISR after treatment, with the most common being swelling, tenderness, and firmness. The majority of ISRs were mild to moderate in severity and resolved within 14 days. The incidence of ISRs after touch-up treatment was lower than that after initial treatment. The ISRs after repeat treatment were similar to those after initial treatment. PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 26 {26} **Table 17: Injection Site Responses by Severity and Duration Occurring in > 5% of Treated Subjects after Initial Treatment$^{a}$ with JUVÉDERM VOLBELLA$^{®}$ XC** | ISR | Incidence (n/N^{b}) | Severity^{c} | | | Duration^{d} | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | Mild | Moderate | Severe | 1-3 Days | 4-7 Days | 8-14 Days | 15-30 Days | | Any ISR | 95.5% (126/132) | 17.5% | 50.0% | 32.5% | 23.8% | 28.6% | 26.2% | 21.4% | | Swelling | 90.9% (120/132) | 31.7% | 46.7% | 21.7% | 52.5% | 40.8% | 5.8% | 0.8% | | Tenderness | 87.1% (115/132) | 49.6% | 42.6% | 7.8% | 57.4% | 27.8% | 10.4% | 4.3% | | Firmness | 82.6% (109/132) | 46.8% | 38.5% | 14.7% | 45.9% | 24.8% | 18.3% | 11.0% | | Bruising | 78.0% (103/132) | 35.9% | 41.7% | 22.3% | 35.0% | 48.5% | 15.5% | 1.0% | | Lumps/Bumps | 76.5% (101/132) | 37.6% | 49.5% | 12.9% | 35.6% | 25.7% | 17.8% | 20.8% | | Redness | 76.5% (101/132) | 54.5% | 38.6% | 6.9% | 73.3% | 19.8% | 6.9% | 0% | | Pain | 68.9% (91/132) | 50.5% | 44.0% | 5.5% | 80.2% | 16.5% | 3.3% | 0% | | Itching | 21.2% (28/132) | 64.3% | 25.0% | 10.7% | 82.1% | 14.3% | 3.6% | 0% | | Discoloration | 17.4% (23/132) | 65.2% | 30.4% | 4.3% | 73.9% | 21.7% | 0% | 4.3% | $^{a}$ Does not include data after touch-up treatment $^{b}$ N denotes the number of subjects who recorded in the diaries after initial treatment $^{c}$ Maximum severity reported in the diary. Denominator for percentages by severity is the number of subjects with corresponding ISR $^{d}$ Maximum reported successive occurrence of an ISR. Denominator for percentages by duration is the number of subjects with corresponding ISR ISRs lasting beyond the 30-day diaries were considered AEs. AEs were also reported by the Evaluating Investigator at follow-up visits. Among the 139 subjects treated with JUVÉDERM VOLBELLA$^{®}$ XC at the initial treatment, 14 (10.1%) experienced 22 treatment-related AEs before repeat treatment. The most common treatment-related AE was injection site mass (lumps/bumps). Subjects treated with JUVÉDERM VOLBELLA$^{®}$ XC experienced mild (7.9%, 11/139) or moderate (2.9%, 4/139) treatment-related AEs. In general, the treatment-related AEs required no action and resolved without sequelae. Among the 139 subjects who were treated with JUVÉDERM VOLBELLA$^{®}$ XC at initial treatment and received repeat treatment with JUVÉDERM VOLBELLA$^{®}$ XC, 3 experienced 4 treatment-related AEs after repeat treatment. These AEs include 3 reports of injection site mass and 1 report of oral herpes. Of the 4 AEs, 2 were mild (1 injection site mass [lumps/bumps] and 1 oral herpes) and 2 were severe (both events were injection site mass [lumps/bumps] in 1 subject). The 2 mild AEs resolved without sequelae. At the end of the study, the 2 severe AEs of injection site mass (lumps/bumps) occurring in 1 subject had not yet resolved. PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 27 {27} Among the 141 subjects who were treated with RESTYLANE-L® at initial treatment and received repeat treatment with JUVÉDERM VOLBELLA® XC, 4 experienced 8 treatment-related AEs after repeat treatment. These AEs include 6 reports of injection site induration (firmness), 1 report of injection site mass (lumps/bumps), and 1 report of oral paresthesia. Of these 8 AEs, 6 were mild (4 reports of injection site induration [firmness], injection site mass [lumps/bumps], and oral paresthesia), 1 was moderate (injection site induration [firmness]), and 1 was severe (injection site induration [firmness]). All of these treatment-related AEs required no action and resolved without sequelae. # **XII. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION** In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the General and Plastic Surgery Devices Advisory Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel. # **XIII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES** # **A. Effectiveness Conclusions** Assessment of product effectiveness is based on the results of the U.S. pivotal study. These submitted data provided a reasonable assurance that the device is effective for injection into the lips for lip augmentation and for correction of perioral rhytids in adults over the age of 21. The specific conclusions are: - The primary analysis was to evaluate the non-inferiority (non-inferiority margin = 0.5 points) of Juvéderm VOLBELLA® XC to control. The primary effectiveness endpoint was the mean change in overall lip fullness from baseline to month 3 based on the Evaluating Investigator's assessments of overall lip fullness using the validated Allergan Lip Fullness Scale 2 (LFS2). The primary endpoint of the study was met with non-inferiority of Juvéderm VOLBELLA® XC to control. The mean change from baseline to month 3 on the LFS2 score was 1.1 for subjects treated with JUVÉDERM VOLBELLA® XC and 1.0 for subjects treated with control, with 80.3% (122/152) of subjects treated with JUVÉDERM VOLBELLA® XC and 70.8% (34/48) of subjects treated with control showing a ≥ 1 point improvement in overall lip fullness. - Throughout the follow-up period, JUVÉDERM VOLBELLA® XC continued to provide a clinically significant improvement in lip fullness (≥ 1-point mean improvement on the LFS2 scale), with 61.8% subjects treated with JUVÉDERM VOLBELLA® XC demonstrating improvement through 1 year. - At 3 months, 65.4% (53/81) of subjects treated with JUVÉDERM VOLBELLA® XC showed a ≥ 1 point improvement in POLSS scores from baseline at rest. At 1 PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 28 {28} year, 66.2% (45/68) of subjects treated with JUVÉDERM VOLBELLA® XC maintained improvement in perioral lines severity at rest. - At 3 months, subjects treated with JUVÉDERM VOLBELLA® XC reported a mean score improved to 76.5 points from the baseline score of 38.5 points on the Satisfaction with Lips module of the FACE-Q. The mean score at 1 year was 59.6 points in the JUVÉDERM VOLBELLA® XC group. - The effectiveness profile after repeat treatment was similar to that after initial treatment, with 94.3% (115/122) of subjects showing at least a 1-point improvement in lip fullness, based on the Evaluating Investigator assessment at 1 month after repeat treatment. # **B. Safety Conclusions** The risks of the device are based on nonclinical laboratory and/or animal studies as well as data collected in the clinical study conducted to support PMA approval as described above. The submitted data provide a reasonable assurance of safety of JUVÉDERM VOLBELLA® XC for lip augmentation and for correction of perioral rhytids in adults over the age of 21. The specific conclusions are: - Almost all subjects (97.4%, 150/154 after initial treatment and 90.2%, 111/123 after repeat treatment) treated with Juvéderm VOLBELLA® XC reported at least 1 ISR. The most frequently reported ISRs after initial and repeat treatment were swelling, tenderness, firmness, bruising, lumps/bumps, redness and pain. Subjects reported the severity of their ISR after initial treatment as mild (14.7%, 22/150), moderate (45.3%, 68/150), or severe (40.0%, 60/150). - Most ISRs lasted less than 2 weeks after initial (59.3%, 89/150) and repeat treatment (73.9%, 82/111), but 40.7% of the ISR lasted between 15-30 days of duration. The most common ISRs that lasted for 15 to 30 days after initial treatment were lumps/bumps (33.3%, 45/135), Dryness (25.0%, 2/8) and firmness (15.3%, 21/137). Similarly after repeat treatment, the most common ISRs lasting 15 to 30 days were lumps/bumps (22.4%, 22/98), firmness (13.1%, 13/99), and tenderness (3.9%, 4/103). - In the time between initial treatment and before repeat treatment, 104 subjects (61.9%, 104/168) in the treated mITT population with VOLBELLA® XC treatment experienced 426 adverse events (treatment emergent adverse event, TEAE). - After initial treatment, treatment-related adverse events were reported in 50.0% (84/168) of subjects treated with Juvéderm VOLBELLA® XC. Most subjects treated with Juvéderm VOLBELLA® XC experienced mild (42.3%, 71/168) or moderate (17.3%, 29/168) treatment-related AEs, only 1.2% (2/168) of subjects experienced severe treatment-related AEs. The treatment-related AEs generally required no action to be taken and resolved without sequelae (98.8%, 83/84). PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 29 {29} - Similar types of treatment-related AEs were reported after repeat treatment, but with a lower incidence rate (13.7%, [17/124]), severity, and duration compared to initial/touch-up treatment. - There were no deaths or treatment-related serious adverse events (SAEs) reported in the study. - Similar proportions of ISRs and AEs were observed between VOLBELLA® XC group and the control groups for the following subgroup analysis: volume injected (< or ≥ the median injected volume of 2.6 mL), Fitzpatrick skin phototype (grouped as I/II, III/IV, and V/VI), age group (< 30, 30 to 50, 51 to 65, and > 65 years), and investigational site. ### C. Benefit-Risk Determination The probable benefits of the device are also based on data collected in a clinical study conducted to support PMA approval as described above. The study was a prospective, controlled study using a validated scale and blinded, live evaluations. The study investigated the safety and effectiveness of JUVÉDERM VOLBELLA® XC for lip augmentation and correction of perioral rhytids. The primary effectiveness endpoint was the analysis of non-inferiority of JUVÉDERM VOLBELLA® XC relative to control treatment in terms of change from baseline to month 3 in mean lip fullness based on Evaluating Investigator assessments using the 5-Point Lip Fullness Scale 2 (LFS2). The data are considered to be as robust as possible for an aesthetic endpoint. The primary endpoint of the study was met with non-inferiority of Juvéderm VOLBELLA® XC to control. Summary of effectiveness conclusions is provided above. Additional factors to be considered in determining probable risks and benefits of JUVÉDERM VOLBELLA® XC injection included: Nearly all subjects (97.4%, 150/154 after initial treatment and 90.2%, 111/123 after repeat treatment) experienced an injection site response. Summary of safety conclusions is provided above. The probable benefits outweigh the probable risks, as determined by the robustness of the effectiveness results, the lack of any long term sequelae, and the improvement of subject satisfaction on the FACE-Q questionnaires. The risks of short term adverse outcomes seen after injection and rare adverse events are sufficiently well understood for patients to make informed decisions about device use. ### Patient Perspectives Patient perspectives considered during the review included: - Despite the frequency of ISRs, patients are willing to accept the probable risk of these harmful events as shown through patient-reported outcomes. - At 3 months, 96.1% (147/153) of subjects treated with JUVÉDERM VOLBELLA® XC reported improvement in satisfaction with their lips, based on the Satisfaction PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 30 {30} with Lips module of the FACE-Q questionnaire, with the mean score increasing from 38.5 at baseline to 76.5. At 1 year, 79.7% (98/123) of subjects reported improved satisfaction with their lips over baseline, with a mean score of 59.6. - At 3 months, subjects treated with JUVÉDERM VOLBELLA® XC reported a mean score improved to 72.9 from 37.5 at baseline on the Lip Lines module of the FACE-Q. The mean score was 56.3 in the JUVÉDERM VOLBELLA® XC group at 1 year. - In the JUVÉDERM VOLBELLA® XC group, 124 subjects received repeat treatment and 20 subjects refused repeat treatment at 12 months. The most common reason given for refusal of repeat treatment was satisfaction with current lip fullness. No subjects refused repeat treatment because of the occurrence of AEs or ISRs. In conclusion, given the available information above, the data support the use of JUVÉDERM VOLBELLA® XC for injection into the lips for lip augmentation and for correction of perioral rhytids in adults over the age of 21, and the probable benefits outweigh the probable risks. # **D. Overall Conclusions** The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the indications for use. # **XIV. CDRH DECISION** CDRH issued an approval order on 5/31/2016. The applicant's manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820). # **XV. APPROVAL SPECIFICATIONS** Directions for use: See device labeling. Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling. Post-approval Requirements and Restrictions: None PMA P110033/S018: FDA Summary of Safety and Effectiveness Data Page 31
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...