XIENCE NANO EVEROLIMUS ELUTING CORONARY STENT SYSTEM
P070015S054 · Abbott Vascular, Inc. · NIQ · May 24, 2011 · Cardiovascular
Device Facts
| Record ID | P070015S054 |
| Device Name | XIENCE NANO EVEROLIMUS ELUTING CORONARY STENT SYSTEM |
| Applicant | Abbott Vascular, Inc. |
| Product Code | NIQ · Cardiovascular |
| Decision Date | May 24, 2011 |
| Decision | APPR |
| Device Class | Class 3 |
| Attributes | Therapeutic |
Indications for Use
The XIENCE V Everolimus Eluting Coronary Stent System is indicated for improving coronary luminal diameter in patients with symptomatic heart disease due to de novo native coronary artery lesions (length ≤ 28 mm) with reference vessel diameters of 2.25 mm to 4.25 mm.
Device Story
Drug-eluting coronary stent system; consists of balloon-expandable cobalt-chromium stent (MULTI-LINK MINI VISION) coated with everolimus in non-erodible polymer (PBMA and PVDF-HFP). Delivered via Rapid-Exchange (RX) catheter system. Used by interventional cardiologists in cath labs to treat de novo native coronary artery lesions. Stent provides mechanical scaffolding; everolimus inhibits neointimal growth via mTOR pathway inhibition. Output is physical vessel support and localized drug delivery. Clinical benefit includes improved luminal diameter and reduced restenosis. Device is a line extension of XIENCE V system.
Clinical Evidence
Prospective, single-arm, multicenter registry (SPIRIT SV) of 150 subjects. Primary endpoint: Target Lesion Failure (TLF) at 1 year. Results: 1-year TLF rate 8.1% (upper 95% CL 13.03%), meeting performance goal of 20.4% (p<0.0001). Secondary endpoints included acute success (95.21% clinical device success) and stent thrombosis (1.5% ARC definite/probable at 1 year). Angiographic cohort (n=69) showed 240-day in-stent late loss of 0.20 ± 0.40 mm.
Technological Characteristics
Stent: L-605 Cobalt Chromium alloy. Coating: Everolimus (100 μg/cm²) in PBMA and PVDF-HFP non-erodible polymer. Delivery: RX platform, 143 cm working length, compatible with ≤ 0.014" guide wires. Sterilization: Ethylene Oxide (EtO) per ISO 11135-1. MR Conditional (1.5T/3T).
Indications for Use
Indicated for patients ≥ 18 years with symptomatic heart disease due to de novo native coronary artery lesions (length ≤ 28 mm) with reference vessel diameters 2.25 mm to 4.25 mm. Contraindicated in patients unable to receive anti-platelet/anti-coagulant therapy, with lesions preventing proper stent placement, or with hypersensitivity to everolimus, cobalt, chromium, nickel, tungsten, acrylic, or fluoropolymers.
Regulatory Classification
Identification
Stent, coronary, drug-eluting -- a metal scaffold with a drug coating placed via a delivery catheter into the coronary artery or saphenous vein graft to maintain the lumen. The drug coating is intended to inhibit restenosis.
Predicate Devices
- XIENCE V Everolimus Eluting Coronary Stent System (P070015)
- MULTI-LINK MINI VISION Coronary Stent System (P020047/S003)
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
## I. GENERAL INFORMATION
| Device Generic Name: | Drug Eluting Coronary Stent System (NIQ) |
| --- | --- |
| Device Trade Name: | XIENCE nano™ Everolimus Eluting Coronary Stent System (2.25 mm) |
| Device also distributed as: | PROMUS Everolimus Eluting Coronary Stent System |
| Applicant's Name and Address: | Abbott Vascular 3200 Lakeside Drive Santa Clara, CA 95054 |
| Date of Panel Recommendation: | none |
| Premarket Approval Application (PMA) Number: | P070015/S054 |
| Date of FDA Notice of Approval: | May 24, 2011 |
| Expedited: | Not Applicable |
The original PMA for the XIENCE V® Everolimus Eluting Coronary Stent System (XIENCE V Stent System) was approved on July 2, 2008 (P070015). The XIENCE V Stent System is indicated for improving coronary luminal diameter in patients with symptomatic heart disease due to de novo native coronary artery lesions (length ≤ 28 mm) with reference vessel diameters of 2.5 mm to 4.25 mm. The SSED to support the initial approval is available on the CDRH website and is incorporated by reference here: http://www.accessdata.fda.gov/cdrh_docs/pdf7/P070015b.pdf.
This PMA supplement, P070015/S054, was submitted to support approval of the 2.25 mm diameter XIENCE V stent and expansion of the indications of the product, specifically, for improving coronary luminal diameter in patients with symptomatic heart disease due to de novo native coronary artery lesions (length ≤ 28 mm) with reference vessel diameters of ≥ 2.25 mm to <2.50 mm. The 2.25 mm diameter XIENCE V Stent System will be referred to as the XIENCE nano™ Everolimus Eluting Coronary Stent System (XIENCE nano Stent System).
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 1 of 28
5
{1}
# II. INDICATIONS FOR USE
The XIENCE V Everolimus Eluting Coronary Stent System is indicated for improving coronary luminal diameter in patients with symptomatic heart disease due to de novo native coronary artery lesions (length ≤ 28 mm) with reference vessel diameters of 2.25 mm to 4.25 mm.
# III. CONTRAINDICATIONS
The XIENCE V stent is contraindicated for use in patients:
Who cannot receive anti-platelet and/or anti-coagulant therapy
- With lesions that prevent complete angioplasty balloon inflation or proper placement of the stent or stent delivery system
- With known hypersensitivity or contraindication to everolimus or structurally-related compounds, cobalt, chromium, nickel, tungsten, acrylic and fluoropolymers.
# IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the XIENCE V Stent System labeling.
# V. DEVICE DESCRIPTION
The components and characteristics of the XIENCE nano Stent System are identical to the XIENCE V Stent System approved in P070015. Please refer to the description provided in the original SSED for additional details. The characteristics of the XIENCE V Stent System are described in Table 1 below.
The XIENCE V Stent System (2.5mm – 4.0 mm) and XIENCE nano (2.25mm) (hereinafter referred to as XIENCE V stent or XIENCE V Stent System) is a device/drug combination product comprised of two regulated components:
- A device (MULTI-LINK VISION® Coronary Stent System or MULTI-LINK MINI VISION® Coronary Stent System)
- A drug coating (formulation of everolimus in a polymer coating)
Table 1 XIENCE V Stent System Product Description
| | XIENCE V Rapid-Exchange (RX) EECSS | XIENCE V Over-the-Wire (OTW) EECSS |
| --- | --- | --- |
| Available Stent Lengths (mm) | 8, 12, 15, 18, 23, 28 | 8, 12, 15, 18, 23, 28 |
| Available Stent Diameters (mm) | 2.25*, 2.5, 2.75, 3.0, 3.5, 4.0 | 2.5, 2.75, 3.0, 3.5, 4.0 |
| Stent Material | A medical grade L-605 Cobalt Chromium (CoCr) alloy MULTI-LINK VISION or | |
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 2 of 28
6
{2}
| | MULTI-LINK MINI VISION stent | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Drug Component | A conformal coating of a non-erodible polymer loaded with 100 μg/cm² of everolimus with a maximum nominal drug content of 181 μg on the largest stent (4.0 x 28 mm) | | | | | |
| Delivery System Working Length | 143 cm | | 143 cm | | | |
| Delivery System Design | Single access port to inflation lumen. Guide wire exit notch is located 30 cm from tip. Designed for guide wires ≤ 0.014". | | Sidearm adaptor provides access to balloon inflation/deflation lumen and guide wire lumen. Designed for guide wires ≤ 0.014". | | | |
| Stent Delivery System Balloon | A compliant, tapered balloon with two radiopaque markers to designate the stent placement on the balloon. | | | | | |
| Balloon Inflation Pressure | Nominal inflation pressure: 8 atm for the 2.25, 2.5 and 2.75 mm diameters; 9 atm for the 3.0, 3.5, and 4.0 mm diameters Rated Burst Pressure (RBP): 16 atm (1621 kPa) for all sizes | | | | | |
| Guiding Catheter Inner Diameter | ≥ 5F (0.056") | | | | | |
| Catheter Shaft Outer Diameter (nominal) | 2.25-3.0 mm 3.5-4.0 mm | | | 2.5 mm | 2.75 x 8 - 3.5 x 18 | 3.5 x 23 - 4.0 x 28 |
| | Distal: 0.032" | 0.035" | Distal: 0.032" | 0.034" | 0.036" | |
| | Proximal: 0.026" | 0.026" | Proximal: 0.042" | 0.042" | 0.042" | |
*The 2.25 mm diameter XIENCE V EECSS is only available on the RX platform.
# A. Device Component Description
The XIENCE nano Stent System is a line extension to the currently approved XIENCE V Stent System, which was approved on July 2, 2008 (PMA P070015). The 2.25 mm diameter XIENCE V Stent System (XIENCE nano Stent System) is only available in the RX platform.
The device component is comprised of the balloon-expandable MULTI-LINK MINI VISION coronary stent pre-mounted onto the MULTI-LINK MINI VISION delivery system consisting of the Rapid Exchange (RX) platform. The MULTI-LINK MINI-VISION RX delivery systems were approved for deployment of the bare metal MULTI-LINK MINI-VISION stent in P020047/S003 (approved September 10, 2004).
The small XIENCE V stent design (2.25, 2.5, 2.75, and 3.0 mm diameters) is identical to the MULTI-LINK MINI VISION stent for the 2.0, 2.25, 2.5 mm diameters, and the MULTI-LINK VISION stent for the 2.75 mm and 3.0 mm diameter. All stent diameters will be available in 8-28 mm lengths.
# B. Drug Component Description
Identical to the XIENCE V stent, the XIENCE nano stent is coated with everolimus (active ingredient), embedded in a non-erodible polymer (inactive ingredient).
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 3 of 28
7
{3}
# B1. Everolimus
Everolimus is the active pharmaceutical ingredient in the XIENCE nano stent. It is a novel semi-synthetic macrolide immunosuppressant, synthesized by chemical modification of rapamycin (INN: sirolimus). The everolimus chemical name is 40-O-(2-hydroxyethyl)-rapamycin and the chemical structure is shown in Figure 1 below.

Figure 1 Chemical Structure of Everolimus
# B2. Inactive Ingredients
The XIENCE nano stent contains inactive ingredients including poly n-butyl methacrylate (PBMA), a polymer that adheres to the stent and drug coating, and PVDF-HFP which is comprised of vinylidene fluoride and hexafluoropropylene monomers as the drug matrix layer containing everolimus. PBMA is a homopolymer with a molecular weight of 264,000 to 376,000 dalton. PVDF-HFP is a non-erodible semi-crystalline random copolymer with a molecular weight of 254,000 to 293,000 dalton. The drug matrix copolymer is mixed with everolimus (83% /17% w/w polymer / everolimus ratio) and applied to the entire PBMA coated stent surface. The drug load is 100 µg/cm² for all product sizes. No topcoat layer is used. The chemical structure of the polymer components are shown in Figures 2a and 2b below.
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 4 of 28
8
{4}

Figure 2a Chemical Structure of Poly (n-butyl methacrylate) (PBMA)

Figure 2b Formula for Poly(Vinylidene Fluoride-Co-Hexafluoropropylene) (PVDF-HFP)
The product matrix, including nominal dosages of everolimus in each XIENCE nano stent is described in Table 2. The nominal everolimus content is based on stent design and length.
Table 2 XIENCE nano Stent System Product Matrix and Everolimus Content
| Model Number (RX) | Stent Diameter (mm) | Stent Length (mm) | Nominal Everolimus Content (μg) |
| --- | --- | --- | --- |
| 1009544-08 | 2.25 | 8 | 37 |
| 1009544-12 | 2.25 | 12 | 56 |
| 1009544-15 | 2.25 | 15 | 75 |
| 1009544-18 | 2.25 | 18 | 88 |
| 1009544-23 | 2.25 | 23 | 113 |
| 1009544-28 | 2.25 | 28 | 132 |
# C. Mechanism of Action
The mechanism by which the XIENCE nano stent inhibits neointimal growth as seen in pre-clinical and clinical studies has not been established. At the cellular level, everolimus inhibits growth factor-stimulated cell
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 5 of 28
9
{5}
proliferation. At the molecular level, everolimus forms a complex with the cytoplasmic protein FKBP-12 (FK 506 Binding Protein). This complex binds to and interferes with FRAP (FKBP-12 Rapamycin Associated Protein), also known as mTOR (mammalian Target Of Rapamycin), leading to inhibition of cell metabolism, growth and proliferation by arresting the cell cycle at the late G1 stage.
# VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are several other alternatives for the treatment of patients with coronary artery disease including exercise, diet, drug therapy, percutaneous coronary interventions (i.e., balloon angioplasty, atherectomy, bare metal stents, coated stents, and other drug-eluting stents), and coronary artery bypass grafting (CABG) surgery. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 6 of 28
10
{6}
# VII. MARKETING HISTORY
The XIENCE nano Stent System is commercially available in the following countries:
| Afghanistan | France | Mauritius | Suriname |
| --- | --- | --- | --- |
| Albania | French Polynesia | Morocco | Sweden |
| Algeria | French Guyana | Myanmar | Switzerland |
| Argentina | Georgia | Netherlands | Syria |
| Aruba | Germany | New Caledonia | Taiwan |
| Australia | Greece | New Zealand | Thailand |
| Austria | Guadeloupe | Nicaragua | Trinidad, Tobago |
| Bahamas | Guatemala | Niederl. Antill. | Tunisia |
| Bahrain | Guyana | Nigeria | Turkey |
| Bangladesh | Honduras | Norway | Uganda |
| Barbados | Hong Kong | Oman | Ukraine |
| Belarus | Hungary | Pakistan | United Arab Emirates |
| Belgium | Iceland | Panama | United Kingdom |
| Belize | India | Paraguay | Uruguay |
| Bermuda | Indonesia | Peru | Uzbekistan |
| Bolivia | Iran | Philippines | Venezuela |
| Brazil | Iraq | Poland | Vietnam |
| British Virgin Islands | Ireland | Portugal | Zimbabwe |
| Brunei | Israel | Qatar | |
| Bulgaria | Italy | Republic of Armenia | |
| Cambodia | Jamaica | Republic of Yemen | |
| Canada^{1} | Jordan | Reunion | |
| Cayman Islands | Kenya | Romania | |
| Chile | Kosovo | Russian Fed. | |
| China | Kuwait | San Marino | |
| Colombia | Latvia | Saudi Arabia | |
| Costa Rica | Lebanon | Serbia | |
| Cyprus | Libya | Singapore | |
| Czech Republic | Liechtenstein | Slovakia | |
| Denmark | Lithuania | Slovenia | |
| Dominican Republic | Luxembourg | South Africa^{2} | |
| Egypt | Macedonia | South Korea | |
| El Salvador | Malaysia | South Yemen | |
| Estonia | Malta | Spain | |
| Finland | Martinique | Sri Lanka | |
$^{1}$ XIENCE V 2.25mm stent is available under Special Access scheme in Canada
$^{2}$ XIENCE V (including 2.25 mm) is available under Section 21 approval
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 7 of 28
11
{7}
As of August 31, 2010, 79,131 XIENCE nano (2.25 mm) Stent Systems have been distributed outside the United States. The XIENCE nano Stent System has not been withdrawn from marketing in any country for any reason.
# VIII. SUMMARY OF PRECLINICAL STUDIES
A series of non-clinical laboratory studies related to the XIENCE nano product were performed. Studies included those performed on the bare metal stent system (MULTI-LINK VISION or MULTI-LINK MINI VISION stent mounted on the stent delivery system), the coated stent alone, the polymer-only coated stent alone (the MULTI-LINK VISION or MULTI-LINK MINI VISION with the PBMA primer layer and PVDF-HFP polymer layer), or the finished combination product (XIENCE nano Stent System).
# A1. Biocompatibility Testing
The biocompatibility testing information included in the original PMA submission (P070015, approved July 2, 2008) is directly applicable to this PMA supplement. Please refer to the original SSED for details regarding the biocompatibility studies.
The complement activation study is being leveraged from the testing conducted on a stent system with identical stent material, drug coating formulation, drug dose density, sterilization cycle and packaging components. This additional biocompatibility study for complement activation was performed to augment the hemocompatibility studies of XIENCE V Stent System provided in PMA P070015. The outcomes of the biocompatibility tests provide passing results.
# A2. In Vitro Engineering Testing
The in vitro engineering testing included in the original PMA submission (P070015, approved July 2, 2008) is directly applicable to this PMA supplement. Based on the similarities in design, test results were leveraged from the MULTI-LINK VISION and MINI VISION Coronary Stent System (CSS), or XIENCE V Stent System testing provided in PMA P070015. The in vitro bench testing plan to support commercial approval of the XIENCE nano Stent System was re-evaluated in light of the release of US FDA April 18, 2010 "Guidance for Industry and FDA Staff - Non-Clinical Engineering Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems," which superseded the previous guidance document. Therefore, only the testing information specific to the XIENCE nano (2.25 mm diameter) stent is summarized below in Table 3. "Pass" denotes that the test results met product specifications and/or the recommendation in the guidance documents.
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 8 of 28
12
{8}
Table 3 In Vitro Engineering Studies
| Test | Test Description | Results |
| --- | --- | --- |
| **Material Characterization Testing** | | |
| Corrosion Testing | The initial pitting corrosion testing conducted on the MULTI-LINK VISION stents (P020047) is leveraged to support the approval of the XIENCE nano Stent System. | PASS |
| Fretting Corrosion | Evaluates the presence of fretting corrosion on the stent after ten year accelerated fatigue testing (400 million cycles) in an overlapped configuration on a 15 mm static bend. The test conducted for the small XIENCE V stent is leveraged for the XIENCE nano stent due to the identical material, design, and manufacturing process. No evidence of fretting corrosion was found. The results met all acceptance criteria. | PASS |
| **Stent Dimensional and Functional Attributes** | | |
| Stent Percent Surface Area | Determines the metal to artery ratio by dividing the abluminal stent surface area by the theoretical cylindrical stented vessel area. It is a theoretical calculation based upon a 3D computational solid model of the stent design. The calculated value for the XIENCE nano at the nominal 2.25 mm diameter is 15%. | Descriptive Only |
| Stent Uniformity of Expansion Test | Determines the uniformity of expansion along the stent length. The XIENCE nano units were inflated to either nominal or post-dilated inner diameters, deflated, and diameter measurements were taken at various points along the stent length. Measurements were averaged and all XIENCE nano stents met product specifications. | PASS |
| Stent Percent Length Change (Foreshortening) Test | Determines the difference in stent length pre-and post-expansion to either nominal or post-dilated inner diameters. All XIENCE nano stents met product specifications. | PASS |
| Stent Percent Recoil Test | Quantifies the amount of recoil of the stent after balloon expansion. The system was inflated to either nominal or post-dilated diameters and measurements were taken of the stent diameter at various locations along the stent length. The system was then deflated and the same measurements taken. The percent recoil is calculated by subtracting the average stent inner diameter (ID) without the balloon from the average stent ID with the balloon, dividing by the average stent ID with the balloon and multiplying by 100. All XIENCE nano stents met product specifications. | PASS |
| Radial Stiffness | Radial stiffness was evaluated on the XIENCE nano stent as a characterization test. | For characterization only |
| Stent Radial (Hoop) Strength Test | Testing was conducted to determine the radial strength of the stent under compression force. Stents were expanded to either nominal or post-dilated diameters, placed in an Instron tester, and subjected to incrementally increasing compression forces. The pressure at which deformation is no longer completely reversible was recorded. All XIENCE nano stents met product specifications. | PASS |
| Finite Element Analysis (FEA) | The XIENCE nano stent has identical stent designs, material properties, maximum expansion diameters, and fatigue loading steps, fatigue safety factors as that of the small XIENCE V stents and hence the in-depth FEA analysis conducted on the XIENCE V stent (P070015) can be leveraged in support of the XIENCE nano stent. The testing indicates adequate margins of safety for the XIENCE nano stent. A supplemental FEA was also performed at nominal expansion diameter of 2.25 mm to verify the worst-case expansion diameter for stress/strain and fatigue safety for XIENCE nano. The analysis | PASS |
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 9 of 28
13
{9}
| Test | Test Description | Results |
| --- | --- | --- |
| | simulated overlapped stent deployment and incorporated a bending radius of 15 mm and verified that the worst case for stress/strain as well as the lowest fatigue safety factor occurs at maximum post-dilation (3.5 mm) and not at nominal expansion of 2.25 mm. The supplemental FEA, therefore, confirmed that the FEA testing submitted for the XIENCE V stent can be appropriately leveraged to verify the short and long term structural behavior of the XIENCE nano stent under worst-case deployment conditions. Results of the FEA were used to determine the worst-case test size to be evaluated in accelerated durability testing and chronic coating particulates. Based upon these results testing can be leveraged from the XIENCE V stent for both accelerated durability testing and chronic coating particulates. | |
| Magnetic Resonance Imaging (MRI) | The initial MRI test results for the XIENCE V Stent System were leveraged to support the approval of the XIENCE nano Stent System. In addition to the leveraged testing, supplemental information and calculations support MR Conditional labeling for the XIENCE nano and XIENCE V Stent Systems under MR imaging at 1.5 T and 3 T and at lengths up to 68 mm for a whole body averaged SAR of 2 W/kg (normal operating conditions) at spatial gradients up to 2500 G/cm. The conservative maximum temperature rise cited in the labeling for these conditions will be 3°C, with no cooling effects from perfusion or blood flow. | PASS |
| **Delivery System Dimensional and Functional Attributes** | | |
| Catheter Dimensional Measurements | The following characteristics were tested to conform to the applicable specifications: Tip Length, Tip Seal Length, Tip Unsealed Length, Proximal Unsealed Balloon Shaft, Total Catheter Length & Distal Catheter Length, Guide Wire Lumen Dimensions (Tip Inner Diameter (ID) & Distal Shaft Junction Notch ID), Stent Placement, Balloon Shoulder to Marker Alignment, Balloon Working Length, Proximal Shaft Marker Locations (Femoral Marker & Brachial Marker), Delivery System Outer Diameters (Distal Shaft OD, Mid Shaft OD, Proximal Shaft OD, Tip Entry OD, Guide Wire Notch OD). All XIENCE nano Stent Systems met product specifications. | PASS |
| Delivery, Deployment, and Retraction | The delivery, deployment and retraction evaluated for system deliverability, balloon marker placement, stent radiopacity, system pullback out of deployed stent and into the guide catheter, and compatibility with accessory devices. Testing demonstrated that the stent system could be delivered to the target location, deployed, and retracted, thus meeting required product specifications. | PASS |
| Balloon Rated Burst Pressure | Statistically demonstrates with 95% confidence, at least 99.9% of the XIENCE nano Stent Systems will not rupture below the rated burst pressure (RBP) and to demonstrate that at a 95% confidence level, at least 99% of the XIENCE nano Stent Systems will not rupture below the maximum labeled compliance (MLC) pressure. All systems met product specifications and confidence/reliability limits. | PASS |
| Unconstrained Balloon Fatigue | Statistically demonstrates with 95% confidence, at least 90% of the XIENCE nano Stent Systems will sustain 10 repeated inflations to the rated burst pressure inside the stent. All XIENCE nano Stent Systems met product specifications. | PASS |
| Stent Diameter vs. Balloon Pressure | Determines how the diameter of a deployed balloon varies with applied balloon pressures. All XIENCE nano Stent Systems met | PASS |
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 10 of 28
14
{10}
| Test | Test Description | Results |
| --- | --- | --- |
| (Compliance) | product specifications. | |
| Soft Tip Tensile | Determines the tensile strength of the soft tip. All XIENCE nano Stent Systems met product specifications. | PASS |
| Distal Delivery System Tensile | Determines the tensile strength of the distal portion of the delivery system. All XIENCE nano Stent Systems met product specifications. | PASS |
| Proximal Adaptation Tensile Strength | Determines the tensile strength of the proximal adaptation of the delivery system. All XIENCE nano Stent Systems met product specifications. | PASS |
| Delivery System Crossing Profile – Crimped Stent Outer Diameter | Determines the crimped stent outer diameter. Measurements were taken at various locations along the length of the stent and averaged to calculate the mean outer diameter. All XIENCE nano Stent Systems met product specifications. | PASS |
| Delivery System Balloon Inflation/Deflation Times | Determines the amount of time required to inflate or deflate the delivery catheter balloon. Inflation times were tested for information only. All XIENCE nano Stent Systems met product specifications for deflation times. | PASS |
| Stent Dislodgement | Determines the amount of force required to displace a stent in both the distal and proximal direction from its original, crimped position on the delivery system balloon after a pre-conditioning step where the system is tracked through a tortuous artery model. All XIENCE nano Stent Systems met product specifications. | PASS |
| Delivery System Guiding Catheter Pullback | Statistically demonstrates that with 95% confidence, at least 99% of the XIENCE nano Stent Systems can be successfully retracted back into a 5F guiding catheter after tracking through a simulated tortuous model prior to the deployment of the stent. All systems met product specifications and confidence/reliability limits. | PASS |
| Delivery System Preparation | Evaluates the ease of preparing the XIENCE nano Stent System using the aspiration method. All XIENCE nano Stent Systems met product specifications. | PASS |
| Delivery System Inner Member Collapse | Verifies that irreversible collapse of the inner member does not occur at or below 300 psi. All XIENCE nano Stent Systems met product specifications. | PASS |
| Catheter Kink and Flexibility Test | Determines the radius of curvature at which the delivery system kinks. All XIENCE nano Stent Systems met product specifications. | PASS |
| Catheter Torque Test - Turns to Failure | Determines the minimum number of rotations to break joints and/or materials or to lose functional integrity of the delivery system. All XIENCE nano Stent Systems met product specifications. | PASS |
### A3. Coating Characterization Testing
The coating characterization testing included in the original PMA submission (P070015, approved July 2, 2008) is directly applicable to this PMA supplement. Therefore, only the testing information specific to the XIENCE nano (2.25 mm diameter) stent is summarized below in Table 4. "Pass" denotes that the test results met product specifications and/or the recommendation in the guidance documents.
Table 4 Coating Characterization Testing
| Test | Test Description | Results |
| --- | --- | --- |
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 11 of 28
15
{11}
| Test | Test Description | Results |
| --- | --- | --- |
| Stent Coating Durability | | |
| Coating Surface Integrity | Determines the stent coating surface integrity of the XIENCE nano stent after tracking through a tortuosity fixture, expansion, and post-dilated to RBP. Defect quantities and sizes were recorded. The compromised coating area was calculated as a percentage of entire coated stent surface. All stents met product specifications. | PASS |
| Particulate - Beaker Method (Over-expansion) | Determines the particulate matter generated during deployment and over expansion of the XIENCE nano stent in a beaker of water. The distal end (balloon and stent) was inserted into glassware filled with clean water. The stents were deployed and post-dilated to the maximum stent diameter. After agitation, aliquots of the water were withdrawn and the particles quantities and sizes were counted and recorded. All stents met product specifications. | PASS |
| Particulate - Tracking Method (Simulated Use) | Determines the particulate matter after navigating simulated, challenging vasculature followed by deployment. The XIENCE nano system was tracked through a simulated tortuous artery model and the stent was deployed unconstrained to RBP inside the simulated vasculature. Water was drawn through the vasculature and the particle quantities and sizes were counted and recorded. All stents met product specifications. | PASS |
### A4. Chemistry, Manufacturing, & Controls (CMC) Testing
The CMC information included in the original PMA submission (P070015, approved July 2, 2008) is directly applicable to this PMA Supplement; therefore the information is not repeated herein.
### A5. Stability/ Shelf Life
A formal stability study was conducted to establish a shelf life / expiration date for the XIENCE nano Stent System. Testing included appearance, total content, drug release, degradation products, oxygen content, molecular weight and polydispersity, microbial challenge (packaging integrity) and endotoxin (pyrogen), particulates and butylated hydroxytoluene (BHT) content. Testing to establish container closure integrity was conducted to ensure sterility was maintained during the shelf life of the product. Functional testing of the stent system was conducted on aged product. The data generated to-date support a shelf life of 1 year.
### A6. Sterilization
The XIENCE nano Stent System is sterilized using ethylene oxide (EtO) sterilization. The sterilization parameters used to sterilize XIENCE nano are identical to the parameters used to sterilize XIENCE V. The cycle is validated per the ISO 11135-1: 2007 Medical Devices – Validation and Routine Control of Ethylene Oxide Sterilization.
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 12 of 28
16
{12}
Results obtained from the sterilization studies show that the product satisfies a minimum Sterility Assurance Level (SAL) of 10-6. In addition, the amount of bacterial endotoxins was verified to be within the specification limits.
# A7. Animal Studies
The in vivo animal studies included in the original PMA submission (P070015, approved July 2, 2008) are directly applicable to this PMA Supplement; therefore, the information is not repeated herein.
# IX. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Below is a list of the potential adverse effects (e.g., complications) associated with the use of the XIENCE V stent.
Adverse events (in alphabetical order) which may be associated with coronary stent use in native coronary arteries include, but are not limited to:
- Abrupt closure
- Access site pain, hematoma or hemorrhage
- Acute myocardial infarction
- Allergic reaction or hypersensitivity to contrast agent or cobalt, chromium, nickel, tungsten, acrylic and fluoropolymers; and drug reactions to antiplatelet drugs or contrast agent
- Aneurysm
- Arterial perforation and injury to the coronary artery
- Arterial rupture
- Arteriovenous fistula
- Arrhythmias, atrial and ventricular
- Bleeding complications, which may require transfusion
- Cardiac tamponade
- Coronary artery spasm
- Coronary or stent embolism
- Coronary or stent thrombosis
- Death
- Dissection of the coronary artery
- Distal emboli (air, tissue or thrombotic)
- Emergent or non-emergent coronary artery bypass graft surgery
- Fever
- Hypotension and/or hypertension
- Infection and pain at insertion site
- Injury to the coronary artery
- Ischemia (myocardial)
- Myocardial infarction (MI)
- Nausea and vomiting
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 13 of 28
17
{13}
Palpitations
Peripheral ischemia (due to vascular injury)
Pseudoaneurysm
Renal failure
Restenosis of the stented segment of the artery
Shock/pulmonary edema
Stroke/cerebrovascular accident (CVA)
Total occlusion of coronary artery
Unstable or stable angina pectoris
Vascular complications including at the entry site which may require vessel repair
Vessel dissection
Zortress®, the oral formulation of everolimus developed by Novartis Pharmaceuticals Corporation, has been evaluated in clinical trials and is approved in the United States for the prophylaxis of organ rejection in adult kidney transplant recipients at low-moderate immunologic risk, at the dose of 1.5 mg/day. Outside the United States, Zortress® is sold under the brand name Certican® in more than 70 countries. Everolimus is also approved in the United States under the name of Afinitor® for the treatment of patients with advanced renal cell carcinoma (cancer) after failure of treatment with sunitinib or sorafenib, at doses of 5 to 20 mg/day when taken by mouth. The following list includes the known risks of everolimus at the oral doses indicated above:
- Abdominal pain (including upper abdominal pain)
- Anemia
- Angioedema
- Anorexia
- Asthenia
- Constipation
- Cough
- Delayed wound healing/fluid accumulation
- Diarrhea
- Dyslipidemia (including hyperlipidemia and hypercholesterolemia)
- Dyspnea
- Dysgeusia
- Dyspepsia
- Dysuria
- Dry skin
- Edema (peripheral)
- Epistaxis
- Fatigue
- Headache
- Hematuria
- Hyperglycemia (may include new onset of diabetes)
- Hyperlipidemia
- Hyperkalemia
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 14 of 28
18
{14}
- Hypertension
- Hypokalemia
- Hypomagnesemia
- Hypophosphatemia
- Increased serum creatinine
- Infections and serious infections: bacterial, viral, fungal, and protozoal infections (may include herpes virus infection, polyoma virus infection which may be associated with BK virus associated nephropathy, and/or other opportunistic infections)
- Insomnia
- Interaction with strong inhibitors and inducers of CY3PA4
- Leukopenia
- Lymphoma and other malignancies (including skin cancer)
- Male infertility (azospermia and/or oligospermia)
- Mucosal inflammation (including oral ulceration and oral mucositis)
- Nausea
- Neutropenia
- Non-infectious pneumonitis
- Pain: extremity, incision site and procedural, and back
- Proteinuria
- Pruritus
- Pyrexia
- Rash
- Stomatitis
- Thrombocytopenia
- Thrombotic microangiopathy (TMA)/Thrombotic thrombocytopenic purpura (TTP)/ Hemolytic uremic syndrome (HUS)
- Tremor
- Urinary tract infection
- Upper respiratory tract infection
- Vomiting
Live vaccines should be avoided and close contact with those that have had live vaccines should be avoided. Fetal harm can occur when administered to a pregnant woman. There may be other potential adverse events that are unforeseen at this time.
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 15 of 28
19
{15}
# X. SUMMARY OF PRIMARY CLINICAL STUDIES
The XIENCE nano safety and effectiveness information is derived from the SPIRIT Small Vessel Registry (SPIRIT SV trial). The SPIRIT SV trial evaluated the safety and efficacy of the XIENCE nano Stent System in improving coronary luminal diameter in subjects with symptomatic heart disease due to a maximum of two de novo native coronary artery lesions in small vessels (≥ 2.25 mm to < 2.50 mm), each in a different epicardial vessel. Follow-up through 1 year is currently available, and yearly follow-up for clinical parameters through 5 years is ongoing.
# A. Study Design
The SPIRIT SV trial is a prospective, single-arm, open-label, US multicenter registry study using XIENCE nano Stent System with stent diameter of 2.25 mm. The trial enrolled a total of 150 subjects at 33 sites. The first 69 enrolled subjects who received the XIENCE nano stent were assigned to the angiographic cohort. The clinical trial design for SPIRIT SV is summarized in Table 5 below.
The SPIRIT SV trial follow-up included office visits at 30 ± 7 days, 240 ± 28 days, and at 1 year ± 28 days. An office visit or telephone follow-up is required annually for years 2 through 5. An ECG was performed for all subjects at the 30-day visit, 240-day visit, and 1-year visit. Angiographic follow-up was also required at the 240-day visit for the angiographic cohort. The primary endpoint of the SPIRIT SV trial was target lesion failure (TLF) at 1 year. TLF is a composite endpoint of cardiac death, target vessel MI and clinically indicated Target Lesion Revascularization (CI-TLR).
Subjects enrolled in the trial had: 1) one target lesion (treated with one XIENCE nano stent), 2) two target lesions (treated with two XIENCE nano stents), or 3) one target lesion (treated with one XIENCE nano stent) and one non-target lesion (treated with a commercial size of the XIENCE V Stent System). Planned overlap was allowed for both the target and non-target lesions, but only with a commercial size of the XIENCE V Stent System. Bailout was allowed with a commercial XIENCE V stent or the XIENCE nano stent. A total of 69 subjects were enrolled in the angiographic cohort.
The primary analysis of the SPIRIT SV data was performed on the Full Analysis Set (FAS) population which was defined as the subjects who received the XIENCE nano stent. The Intent to Treat (ITT) population was defined as the subjects enrolled into the study, regardless of the treatment actually received; this population excludes de-registered subjects.
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 16 of 28
20
{16}
Table 5 SPIRIT Small Vessel Trial Design
| | SPIRIT Small Vessel |
| --- | --- |
| Study Type/Design | • Multi-center • Non-randomized • Open-label • Non-blinded • Single-arm |
| Number of Subjects Enrolled | Total : 150 |
| Treatment | Up to two de novo lesions in different epicardial vessels |
| Lesion Size | RVD: ≥ 2.25 < 2.50 mm Length: ≤ 28 mm |
| Stent Sizes (XIENCE nano) | Diameter: 2.25 mm Length: 8, 18, 28 mm |
| Post-procedure Antiplatelet Therapy | Clopidogrel 12 months minimum (or ticlopidine per site standard), aspirin indefinitely |
| Primary Endpoint | Target lesion failure (TLF) at 1-year (composite of cardiac death, target vessel MI and clinically-indicated TLR) |
| Co-Primary Endpoint | None |
| Major Secondary Endpoint | None |
| Clinical Follow-up | 30, 240, days, 1 to 5 years |
| Angiographic Follow-up | 240 days (n = 69) |
| Status | One year reported |
# 1. Clinical Inclusion and Exclusion Criteria
Enrollment in the SPIRIT SV trial was limited to subjects who met the eligibility criteria and who provided a signed informed consent form prior to enrollment. Subjects had to be at least 18 years old, with evidence of myocardial ischemia based on the presence of angina, silent ischemia, a positive functional study or reversible ECG changes consistent with ischemia. Female subjects with childbearing potential had to have a negative pregnancy test within 7 days of the index procedure.
The key angiographic inclusion criteria applied to the target and non-target lesion are listed below:
- De novo lesions in native coronary arteries (Target lesion: RVD ≥ 2.25 mm to < 2.50 mm; Non-target lesion: RVD ≥ 2.5 mm to ≤ 4.25 mm; lesion length ≤ 28 mm) with one lesion in one vessel or two lesions in two separate epicardial vessels.
- If two target lesions or one target and one non-target lesion, both lesions must have satisfied the angiographic eligibility criteria of %DS of ≥ 50% and < 100% with a TIMI flow of ≥ 1.
The key angiographic exclusion criteria applied to the target and non-target lesion are listed below:
- Location in an arterial or vein graft
- Aorto-ostial location
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 17 of 28
21
{17}
- Left main location
- Lesion located within 2 mm of the origin of the LAD and LCX
- Extreme angulation (≥ 90°) or excessive tortuosity (≥ two 45° angles) proximal to or within the target or non-target lesion
- Heavy calcification proximal to or within the target or non-target lesion
- Target or non-target vessel contain thrombus
- Another clinically significant lesion located in the same epicardial vessel as the target or non-target lesion
- Target or non-target vessel has been previously treated with any type of PCI < 9 months prior to index procedure
- Vessel not intended to be treated with the XIENCE nano Stent System or commercial XIENCE V Stent System has been previously treated with any type of PCI < 90 days prior to index procedure.
- Clinically significant lesions in any vessel or side branch that may require PCI ≤ 90 days after the index procedure are also excluded.
# 2. Follow-up Schedule
All subjects will be followed for a five-year period. All subjects were required to have an office follow-up visit at 30 days, 240 days, and at 1 year. Annual office or telephone follow-up visits are required at years 2 through 5.
# 3. Stent Thrombosis Definitions
Stent Thrombosis was defined in the protocol as clinical presentation of acute coronary syndrome with angiographic evidence of stent thrombosis (thrombus within or adjacent to the treated target lesion), or in the absence of angiography, any unexplained death (at any time following the index procedure) or acute MI in the distribution of the target lesion within 30 days of the index procedure. Stent thrombosis was categorized as acute (≤ 1 day), subacute (> 1 day to ≤ 30 days), or late (> 30 days) relative to the index procedure.
Stent thrombosis was defined by ARC criteria:
# Timing:
- Acute ST: 0 to 24 hours post stent implantation
- Subacute ST: >24 hours - 30 days post stent implantation
- Late ST: 30 days to 1 year post stent implantation
- Very late ST: > 1 year post stent implantation
# Level of probability:
- Definite ST - considered to have occurred by either angiographic (with at least one of the following: acute onset of ischemic symptoms at rest; new ischemic changes suggestive of acute ischemia; or a typical rise and fall of cardiac biomarkers) or pathologic confirmation.
- Probable ST - considered to have occurred after intracoronary stenting in the following cases:
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 18 of 28
22
{18}
1. Any unexplained death within the first 30 days.
2. Irrespective of the time after the index procedure, any MI which is related to documented acute ischemia in the territory of the implanted stent without angiographic confirmation of ST and in the absence of any other obvious cause.
- Possible ST - considered to have occurred with any unexplained death following 30 days after the intracoronary stenting until the end of trial follow-up.³
# 4. Clinical Endpoints
The primary endpoint of the SPIRIT SV was target lesion failure (TLF) at 1 year. TLF was defined as the composite endpoint of Cardiac Death, Target Vessel Myocardial Infarction (Q-wave and non-Q-wave), and clinically-indicated Target Lesion Revascularization (CI-TLR) by Coronary Artery Bypass Graft (CABG) or Percutaneous Coronary Intervention (PCI).
Other key secondary endpoints to examine the safety and efficacy included the following:
- Acute Success: (Combined clinical/angiographic)
- Clinical Device Success⁴ (Per lesion basis, for target lesions treated by the XIENCE nano Stent with or without planned overlap)
- Clinical Procedural Success⁵ (Per subject basis, for all target and non-target lesions)
- Clinical Endpoint in hospital and at each follow-up point (30 days, 240 days, 1, 2, 3, 4 and 5 years) (for ALL lesions)
- Component Endpoints (each evaluated per protocol definition and ARC definition)
All Death (cardiac, vascular, non-cardiovascular)
Target Vessel MI - Q-wave and non Q-wave (defined as MI not clearly attributable to a non-target vessel)
Non Target Vessel MI (Q-wave, Non Q-wave)
CI-TLR
Clinically indicated Target Vessel Revascularization (TVR = TLR and non-TLR in TV)
³ All stent thrombosis data presented reflect definite + probable stent thrombosis only.
⁴ Clinical Device Success was defined as the successful delivery and deployment of the first study stent intended to be implanted at the intended target lesion (or in the overlapping stent setting, a successful delivery and deployment of the intended first and second investigational stents) and successful withdrawal of the stent delivery system with attainment of final residual stenosis of <50% of the target lesion by QCA (or by visual estimation if QCA unavailable). Bailout lesions were included as device success only if the above criteria for clinical device success were met for the bailout stent.
⁵ Clinical Procedural Success was defined as the achievement of a final in-stent diameter stenosis (DS) of <50% (by QCA) using the assigned device and with any adjunctive devices, without the occurrence of cardiac death, target vessel MI (per protocol definition), or repeat coronary revascularization of the target lesion during the hospital stay (up to 7 days if a subject is still in the hospital). If QCA %DS was not available, procedure success data were considered missing.
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 19 of 28
23
{19}
All TLR (CI and non-CI)
All TVR (CI and non-CI)
All Coronary Revascularization (TVR and non-TVR)
• Composite Endpoints
Cardiac Death/MI
Cardiac Death/All MI/CI-TLR
All Death/All MI/All Coronary Revascularization
• Stent Thrombosis (to be evaluated per protocol definition and ARC)
By Timing (Acute, Subacute, Late and Very late)
By Evidence (Definite, Probable and Possible)
• Angiographic Endpoints at 240 days (for target lesions of the angiographic cohort)
In-segment, in-stent, proximal and distal LL
In-segment, in-stent, proximal and distal % diameter stenosis (%DS)
In-segment, in-stent, proximal and distal angiographic binary restenosis
(ABR) rate
### B. Accountability of PMA Cohort
There were a total of 150 subjects enrolled in SPIRIT SV trial at 33 US sites which included a total of 69 investigators. The FAS population (primary analysis population) consisted of 144 subjects as 6 subjects did not receive the XIENCE nano stent. Subjects in the trial were terminated if a subject withdrew consent, the physician withdrew the subject, the subject was lost to follow-up or the subject died. In the FAS population, 3.5% (5/144) of subjects were considered terminated from the study on or prior to the 30-day follow-up, 4.9% (7/144) of subjects were terminated on or prior to the 240-day follow-up. In the angiographic cohort 5.8% (4/69) of subjects were terminated on or prior to the 240-day follow-up. In the FAS population, 6.9% (10/144) of subjects were terminated on or prior to the 1-year follow-up.
### C. Study Population Demographics and Baseline Parameters
In the FAS population, the mean age was 62.97 ± 10.59 years and 61.8% (89/144) were male. In subjects treated with the XIENCE nano Stent System, 22.9% (32/140) were tobacco users, 81.9% (118/144) were hypertensive requiring medication, 86.5% (122/141) had hypercholesterolemia requiring medication, and 39.2% (56/143) were diabetic. Additionally, 68.8% (99/144) of the subjects had stable angina and 27.1% (39/144) had unstable angina. A prior cardiac intervention at the target vessel had occurred in 20.6% (29/141) of subjects, while 27.5% (39/142) of subjects had a prior MI, and 54.2% (65/120) had a family history of premature coronary artery disease.
In the FAS population, LAD coronary artery treatment was performed in 40.7% (59/145) of lesions, LCX/ramus treatment in 31.0% (45/145), and RCA treatment in 28.3% (41/145). In the FAS population, the incidence of calcified or eccentric target lesions were 9.0% (13/144) and 10.4% (15/144), respectively. TIMI 3 flow was
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 20 of 28
24
{20}
present in 92.4% (134/145) of target lesions. In the FAS population, the frequency of type B1 or type B2 target lesions (according to ACC/AHA classification) was 47.6% (69/145) and 23.4% (34/145), respectively. In the FAS population the mean pre-procedure lesion length was 13.38 ± 5.31 mm, pre-procedure RVD 2.13 ± 0.23 mm, pre-procedure MLD 0.55 ± 0.20 mm, and pre-procedure %DS 72.84 ± 9.26.
## D. Safety and Effectiveness Results
The results are presented in Table 6 (Primary endpoint), Table 7 (Clinical Results) Table 8 (Stent Thrombosis Results), Table 9 (Angiographic Results), and Figure 3 (TLF Free Survival). These analyses are based on the FAS population.
In the FAS population, the 1-year TLF rate was 8.1% for which the upper limit of the one-sided 95% confidence interval was 13.03%, which met the pre-specified performance goal of 20.4% (p<0.0001).
Table 6 SPIRIT Small Vessel Primary Endpoint Results
| Primary Endpoint | XIENCE nano (N=144) | Upper 1-Sided 95% CL | P-Value^{1} |
| --- | --- | --- | --- |
| 1 Year TLF | 8.1% (11/136) | 13.03% | <0.0001 |
Notes:
- N is the total number of subjects.
- TLF includes cardiac death, target vessel MI (per protocol definition) and clinically-indicated TLR.
- Time Frame includes follow-up window (365 + 28 days).
$^{1}$ One-sided p-value by testing against the performance goal of 20.4% using exact test at 0.05 significance level.
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 21 of 28
25
{21}
Table 7 SPIRIT Small Vessel Registry Clinical Endpoint Results through 1 Year
| XIENCE nano | Per Protocol Definition |
| --- | --- |
| ACUTE SUCCESS | ITT* (N = 149) |
| Clinical Device Success | 95.21% (139/146) |
| Clinical Procedure Success | 97.93% (142/145) |
| CLINICAL ENDPOINTS | FAS (N = 144) |
| Component Endpoints | |
| All Death | 1.5% (2/136) |
| Cardiac Death | 1.5% (2/136) |
| Non-Cardiac Death | 0.0% (0/136) |
| Target Vessel MI | 1.5% (2/136) |
| Non Target Vessel MI | 0.0% (0/136) |
| Clinically-Indicated TLR (CI-TLR) | 5.1% (7/136) |
| Clinically-Indicated TVR (CI-TVR) | 8.8% (12/136) |
| All TLR | 6.6% (9/136) |
| All TVR | 10.3% (14/136) |
| All Revascularization | 14.7% (20/136) |
| Composite Endpoints | |
| Cardiac Death or MI | 2.9% (4/136) |
| Cardiac Death or All MI or CI-TLR | 8.1% (11/136) |
| All Death or All MI or All Revascularization | 16.9% (23/136) |
| Notes: - N is the total number of subjects; L is the number of lesions. - Per protocol MI definition was used for Target Vessel MI, Non Target Vessel MI and all composite endpoints. - Time Frame includes follow-up window (365 ± 28 days). - Non Target Vessel MI includes MI not attributed to the treated vessel. - All Revascularization includes TVR and non-TVR, and non-treated vessel revascularization. * The ITT population provides the most accurate estimate of successful XIENCE nano stent implantation because it includes all subjects regardless of whether the attempted implantation of XIENCE nano stent was successful. | |
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 22 of 28
26
{22}
Table 8 SPIRIT Small Vessel Stent Thrombosis Results through 1 Year
| XIENCE nano | FAS (N = 144) | |
| --- | --- | --- |
| Stent Thrombosis | Per Protocol Definition | Per ARC Definition (Definite + Probable) |
| Acute (≤ 1 day) | 0.0% (0/143) | 0.0% (0/143) |
| Subacute (>1 – 30 days) | 0.7% (1/140) | 0.7% (1/140) |
| Acute/Subacute (0 – 30 days) | 0.7% (1/140) | 0.7% (1/140) |
| Late (31 – 393 days) | 1.5% (2/135) | 0.7% (1/135) |
| Overall (0 – 393 days) | 2.2% (3/136) | 1.5% (2/136) |
Table 9 SPIRIT Small Vessel 240-Day Angiographic Results (Angiographic Cohort¹)
| XIENCE nano | FAS (N = 69) (L = 69) |
| --- | --- |
| 240-day Late Loss (mm) | |
| In-Stent | 0.20 ± 0.40 (52) |
| In-Segment | 0.16 ± 0.41 (52) |
| Proximal | 0.21 ± 0.35 (34) |
| Distal | 0.00 ± 0.28 (45) |
| 240-day % Diameter Stenosis | |
| In-Stent | 12.86 ± 19.58 (52) |
| In-Segment | 20.85 ± 22.53 (52) |
| Proximal | 14.31 ± 13.16 (37) |
| Distal | 10.40 ± 8.45 (46) |
| 240-day Angiographic Binary Restenosis | |
| In-Stent | 3.8% (2/52) |
| In-Segment | 9.6% (5/52) |
| Proximal | 2.7% (1/37) |
| Distal | 0.0% (0/46) |
Notes:
N is the total number of subjects. L is the total number of lesions.
¹ Per Protocol defined qualifying angiogram with follow-up window extended to 268 days.
240-day angiographic data is only available for 52 subjects.
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 23 of 28
27
{23}
TLF

| TLF | Event Free | Event Rate |
| --- | --- | --- |
| XIENCE nano | 92.1% | 7.9% |
Note:
- Time Frame includes follow-up window (365 ÷ 28 days).
Figure 3 SPIRIT Small Vessel Registry: Kaplan Meier Time-to-Event Curve for TLF through 1 Year
# Gender-Based Analysis
Cardiovascular disease is the leading cause of death for both women and men in the U.S. and coronary artery disease is a major cause of morbidity and mortality in women. It is estimated that the prevalence of coronary artery disease in the United States is 9.1% (9,200,000) in males and 7.0% (8,400,000) in females for adults at least 20 years old according to the American Heart Association 2010 Update.⁶ However, it is estimated that only 36% of annual PCIs are performed in women.⁷ In PCI clinical trials, women represent only 25-35% of the enrolled populations, and there are relatively little gender-specific data. The disproportionate enrollment distribution in these trials may be partly attributable to gender differences in symptoms and pathophysiology,⁸ which may lead to under-diagnosis and under-referral of female patients with CAD. Women tend to have
⁶ Lloyd-Jones D, Adams R, Carnethon M, De Simone G, et al. Heart disease and stroke statistics--2010 update: a report from the American Heart Association. Circulation 2010; 121:e46-215.
⁷ Lloyd-Jones D, Adams R, Carnethon M, De Simone G, Ferguson TB, Flegal K, et al. Heart disease and stroke statistics--2009 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Circulation 2009; 119(3):e21-181.
⁸ Shaw LJ, Bairey Merz CN, Pepine CJ, et al. Insights From the NHLBI-Sponsored Women's Ischemia Syndrome Evaluation (WISE) Study: Part I: Gender Differences in Traditional and Novel Risk Factors, Symptom Evaluation, and Gender-Optimized Diagnostic Strategies. J Am Coll Cardiol 2006 47: S4-20.
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 24 of 28
28
{24}
worse clinical outcomes compared to men, most likely due to a higher baseline risk profile and more complex angiographic characteristics.9,10,11
The applicant also performed a post-hoc multivariable gender analysis of the SPIRIT Small Vessel clinical trial. However, it should be noted that the SPIRIT SV trial was not powered to detect any differences between genders, and a subgroup analysis was not pre-specified in the Statistical Analysis Plan. Therefore, due to the limited sample size and the small number of events, the results from the analysis below should be considered exploratory without definitive conclusions.
The baseline SPIRIT SV trial subject characteristics stratified by gender are shown in Table 10. Compared to the men, women were more likely to be diabetics treated with insulin [27.3% (15/55) for females versus 8.0% (7/88) for males], and less likely to be smokers [13.0% (7/54) smokers for females versus 29.1% (25/86) smokers for males]. In terms of baseline angiographic characteristics, the differences in lesion length and pre-procedure RVD, MLD, or %DS were not found to differ significantly between females and males in this post-hoc analysis.
Table 10: Demographics, Risk Factors, and Baseline Angiographic Characteristics for All Female and All Male Subgroups in the SPIRIT SV Clinical Trial
| Subject Characteristics | Females (N=55) (M=55) | Males (N=89) (M=90) | P Value |
| --- | --- | --- | --- |
| **Baseline Demographics, Mean ± SD (n)** | | | |
| Age (year) | 64.74 ± 11.28 (55) | 61.88 ± 10.05 (89) | 0.1260 |
| **Baseline Risk Factors, % (No./total)** | | | |
| All Diabetes | 49.1% (27/55) | 33.0% (29/88) | 0.0778 |
| Diabetes Treated with Insulin | 27.3% (15/55) | 8.0% (7/88) | 0.0034 |
| Current Tobacco Use | 13.0% (7/54) | 29.1% (25/86) | 0.0378 |
| Hypertension Requiring Medication | 83.6% (46/55) | 80.9% (72/89) | 0.8243 |
| Hypercholesterolemia Requiring Medication | 85.2% (46/54) | 87.4% (76/87) | 0.8012 |
| Stable Angina | 72.7% (40/55) | 66.3% (59/89) | 0.4632 |
| Unstable Angina | 25.5% (14/55) | 28.1% (25/89) | 0.8474 |
| Prior MI | 20.4% (11/54) | 31.8% (28/88) | 0.1759 |
| **Target Vessel, % (No./total)** | | | |
| LAD | 45.5% (25/55) | 37.8% (34/90) | 0.3876 |
| Circumflex or Ramus | 30.9% (17/55) | 31.1% (28/90) | 1.0000 |
| RCA | 23.6% (13/55) | 31.1% (28/90) | 0.3502 |
| LMCA | 0.0% (0/55) | 0.0% (0/90) | N/A |
| **Pre-Procedure QCA Analysis, Mean ± SD (m)** | | | |
| Lesion Length (mm) | 13.62 ± 5.81 (55) | 13.23 ± 5.00 (89) | 0.6808 |
| Pre-Procedure RVD (mm) | 2.11 ± 0.24 (55) | 2.14 ± 0.22 (90) | 0.4615 |
9 Mahoney EM, Jurkovitz CT, Chu H, Becker ER, Culler S, Kosinski AS, et al. Cost and cost-effectiveness of an early invasive vs conservative strategy for the treatment of unstable angina and non-ST-segment elevation myocardial infarction. Jama 2002; 288(15):1851-8.
10 Akhter N, Milford-Beland S, Roe MT, Piana RN, Kao J, Shroff A. Gender differences among patients with acute coronary syndromes undergoing percutaneous coronary intervention in the American College of Cardiology-National Cardiovascular Data Registry (ACC-NCDR). Am Heart J 2009; 157(1):141-8.
11 Vaina S, Voudris V, Morice M-C, de Bruyne B, Colombo A, Macaya C, Richardt, G, Fajadet, J et al. Effect of gender differences on early and mid-term clinical outcome after percutaneous or surgical coronary revascularization in patients with multivessel coronary artery disease: Insights from ARTS I and ARTS II. EuroInterv. 2009; 4(4):492-501.
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 25 of 28
29
{25}
| Subject Characteristics | Females (N=55) (M=55) | Males (N=89) (M=90) | P Value |
| --- | --- | --- | --- |
| Pre-Procedure MLD (mm) | 0.55 ± 0.19 (55) | 0.55 ± 0.20 (90) | 0.9914 |
| Pre-Procedure Percent Diameter Stenosis (%DS) | 72.74 ± 8.31 (55) | 72.90 ± 9.83 (90) | 0.9150 |
Table 11 presents key clinical outcomes through 1 year in female and male subjects in the SPIRIT SV clinical trial. At 1 year, post-hoc analyses of the SPIRIT SV trial are limited with regards to conclusions that could be drawn regarding any differences in adverse event rates between female and male subjects due to the small sample size.
Table 11: Clinical Results for All Female and All Male Subgroups in the SPIRIT SV Clinical Trial through 1 year
| SPIRIT SV | Females (N=55) | Males (N=89) |
| --- | --- | --- |
| All Death | 0.0% (0/51) | 2.4% (2/85) |
| Cardiac Death | 0.0% (0/51) | 2.4% (2/85) |
| Non-Cardiac Death | 0.0% (0/51) | 0.0% (0/85) |
| Target Vessel MI | 0.0% (0/51) | 2.4% (2/85) |
| Cardiac Death or Target Vessel MI | 0.0% (0/51) | 4.7% (4/85) |
| Bleeding Complication | 5.9% (3/51) | 2.4% (2/84) |
| Stent Thrombosis | | |
| Protocol defined | 0.0% (0/51) | 3.5% (3/85) |
| ARC definite + probable | 0.0% (0/51) | 2.4% (2/85) |
| TLF | 11.8% (6/51) | 5.9% (5/85) |
| Ischemia-Driven TLR | 11.8% (6/51) | 1.2% (1/85) |
| Ischemia-Driven TVR, non TL | 7.8% (4/51) | 4.7% (4/85) |
# XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION
In accordance with the provisions of section 515(c)(2) of the Act as amended by the Safe Medical Devices Act of 1990, this PMA supplement was not referred to the Circulatory Systems Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA supplement substantially duplicates information previously reviewed by this panel.
# XII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES
The safety and effectiveness of the XIENCE nano Everolimus Eluting Coronary Stent System are based on the results obtained from: evaluation of biocompatibility; in vitro engineering testing; coating characterization; chemistry, manufacturing and controls
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 26 of 28
30
{26}
information; in vivo animal testing; sterilization information; stability testing; and clinical studies. These tests revealed the following:
# A. Safety Conclusions
The biocompatibility testing conducted on the XIENCE nano Stent System and the in vivo pharmacokinetics evaluation and in vivo animal testing conducted on the XIENCE V Stent System demonstrate that the acute and chronic in vivo performance characteristics of the product provide reasonable assurance of safety and acceptability for clinical use.
The in vitro engineering testing conducted on the stent and delivery systems demonstrated that the performance characteristics met the product specifications and the coating characterization testing adequately described the important attributes of the everolimus/polymer coating. The chemistry, manufacturing, and controls information ensures that product meeting specifications will be released.
The test results obtained from the sterilization testing demonstrated that the product can be adequately sterilized and is acceptable for clinical use. The stability testing demonstrated that the product can be labeled with a shelf life of 12 months.
# B. Effectiveness Conclusions
The results of the SPIRIT Small Vessel Registry showed that the primary composite endpoint of target vessel failure (defined as the composite of cardiac death, target vessel myocardial infarction and clinically-indicated target lesion revascularization) at one year was 8.1%, for which the upper limit of the one-sided 95% confidence interval was 13.03%, which met the prespecified performance goal of 20.4% (p<0.0001). This composite endpoint contains both safety and effectiveness components.
The clinical testing conducted demonstrated that the XIENCE nano Everolimus Eluting Coronary Stent System provides a reasonable assurance of safety and effectiveness when used in accordance with the instructions for use.
# XIII. CDRH DECISION
CDRH issued an approval order on May 24, 2011.
The applicant's manufacturing facilities were inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820).
# XIV. APPROVAL SPECIFICATIONS
Directions for use: See device labeling.
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 27 of 28
31
{27}
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling.
Post-approval Requirements and Restrictions: See approval order.
PMA P070015/S054: FDA Summary of Safety and Effectiveness Data
Page 28 of 28
32