TAXUS LIBERTE ATOM PACLITAXEL-ELUTING CORONARY STENT SYSTEM
P060008S008 · Boston Scientific Corp · NIQ · May 21, 2009 · Cardiovascular
Device Facts
| Record ID | P060008S008 |
| Device Name | TAXUS LIBERTE ATOM PACLITAXEL-ELUTING CORONARY STENT SYSTEM |
| Applicant | Boston Scientific Corp |
| Product Code | NIQ · Cardiovascular |
| Decision Date | May 21, 2009 |
| Decision | APPR |
| Device Class | Class 3 |
| Attributes | Therapeutic |
Indications for Use
The TAXUS Liberté Paclitaxel-Eluting Coronary Stent System (Monorail and Over-the-Wire Systems) is indicated for improving luminal diameter for the treatment of de novo lesions in native coronary arteries ≥ 2.25 to ≤ 4.00 mm in diameter in lesions ≤ 28 mm in length.
Device Story
Drug-eluting coronary stent system; consists of 316L stainless steel Liberté stent coated with paclitaxel in poly(styrene-b-isobutylene-b-styrene) (SIBS) polymer. Delivered via Monorail or Over-the-Wire catheter systems. Used by interventional cardiologists in cath labs to treat coronary artery disease. Stent provides mechanical scaffolding to maintain luminal diameter; polymer coating provides controlled release of paclitaxel to inhibit restenosis. Output is a permanently implanted stent. Benefits include reduced restenosis and target vessel revascularization compared to bare metal stents.
Clinical Evidence
Prospective, multicenter, single-arm trial (TAXUS ATLAS SV, N=261). Primary endpoint: 9-month in-segment percent diameter stenosis (%DS) via QCA. Results: 2.25 mm TAXUS Liberté was non-inferior to TAXUS Express DES (adjusted difference -7.3%, p<0.0001) and superior to bare metal Express (adjusted difference -13.4%, p<0.0001). MACE rates at 9 months were 12.8% for TAXUS Liberté vs 20.5% for DES control and 21.6% for BMS control.
Technological Characteristics
316L surgical grade stainless steel stent; SIBS polymer carrier; paclitaxel drug (1μg/mm²). Monorail (140 cm) or Over-the-Wire (135 cm) delivery systems. Balloon inflation nominal 9 ATM, rated burst 18 ATM. Ethylene oxide sterilization (AAMI/ISO 1135:1994).
Indications for Use
Indicated for patients ≥ 18 years old with stable or unstable angina or silent ischemia requiring PCI for de novo lesions in native coronary arteries (2.25-4.00 mm diameter, ≤ 28 mm length). Contraindicated in patients with hypersensitivity to 316L stainless steel, paclitaxel, or polymer components; patients unable to receive anti-platelet/anticoagulant therapy; or lesions preventing complete balloon inflation/stent placement.
Regulatory Classification
Identification
Stent, coronary, drug-eluting -- a metal scaffold with a drug coating placed via a delivery catheter into the coronary artery or saphenous vein graft to maintain the lumen. The drug coating is intended to inhibit restenosis.
Predicate Devices
Reference Devices
Submission Summary (Full Text)
{0}
# Summary of Safety and Effectiveness Data
# I. GENERAL INFORMATION
Device Generic Name: Drug-Eluting Coronary Stent System (NIQ)
Device Trade Name: TAXUS® Liberté® Atom™ Paclitaxel-Eluting Coronary Stent System (Monorail and Over-the-Wire Delivery Systems), 2.25 x 8-32 mm
Applicant's Name and Address: Boston Scientific Corporation
One Boston Scientific Place
Natick, MA 01760-1537
USA
Premarket Approval (PMA)
Application Number: P060008/S008
Date of Panel Recommendation: None
Date of FDA Notice of Approval: May 21, 2009
The original PMA (P060008) was approved on October 10, 2008 and is indicated for improving luminal diameter for the treatment of de novo lesions ≤ 28 mm in length in native coronary arteries ≥ 2.5 to ≤ 4.00 mm in diameter. The SSED to support the indication is available on the CDRH website and is incorporated by reference here: http://www.fda.gov/cdrh/pdf6/p060008b.pdf. This PMA supplement P060008/S008 was submitted to support expansion of the indications of the product, specifically, for improving luminal diameter for the treatment of de novo lesions in native coronary arteries ≥ 2.25 to ≤ 4.00 mm in diameter in lesions ≤ 28mm in length.
# II. INDICATIONS FOR USE
The TAXUS Liberté Paclitaxel-Eluting Coronary Stent System (Monorail and Over-the-Wire Systems) is indicated for improving luminal diameter for the treatment of de novo lesions in native coronary arteries ≥ 2.25 to ≤ 4.00 mm in diameter in lesions ≤ 28 mm in length.
page 1 8
{1}
### III. CONTRAINDICATIONS
Use of the TAXUS Liberté Paclitaxel-Eluting Coronary Stent System is contraindicated in patients with:
- Known hypersensitivity to 316L stainless steel
- Known hypersensitivity to paclitaxel or structurally-related compounds.
- Known hypersensitivity to the polymer or its individual components.
Coronary Artery Stenting is contraindicated for use in:
- Patients who can not receive recommended anti-platelet and/or anticoagulant therapy.
- Patients judged to have a lesion that prevents complete inflation of an angioplasty balloon or proper placement of the stent or delivery device.
### IV. WARNING AND PRECAUTIONS
The warnings and precautions can be found in the TAXUS Liberté Paclitaxel-Eluting Coronary Stent System Directions for Use (DFU).
### V. DEVICE DESCRIPTION
The TAXUS Liberté Paclitaxel-Eluting Coronary Stent System is a device / drug combination product comprised of two regulated components: a device (Liberté Coronary Stent System) and a drug product (a formulation of paclitaxel contained in a polymer coating). The components and characteristics of the TAXUS Liberté Atom Paclitaxel-Eluting Coronary Stent System are identical to the small vessel model TAXUS Liberté Paclitaxel-Eluting Coronary Stent System approved in P060008. Please refer to the device description provided in the original SSED for additional details. The characteristics of the TAXUS Liberté Atom stent system are described in Table 1.
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 2
9
{2}
Table 1: TAXUS Liberté Stent System Product Description
| | TAXUS Liberté Monorail® Stent Delivery System | TAXUS Liberté Over-the-Wire Stent Delivery System |
| --- | --- | --- |
| Available Stent Lengths (mm) | 8, 12, 16, 20, 24, 28, 32 | |
| Available Stent Diameters (mm) | 2.25 | |
| Stent Material | A 316L surgical grade stainless steel Liberté® stent | |
| Drug Product | A conformal coating of a polymer carrier loaded with 1μg/mm² paclitaxel in a slow release (SR)* formulation applied to the stent with a maximum nominal drug content of 155 μg on the largest stent (2.25 x 32 mm). | |
| Delivery System | | |
| Working Length | 140 cm | 135 cm |
| Delivery System Y- Adapter Ports | Single access port to inflation lumen. Guidewire exit port is located approximately 25 cm from tip. Designed for guidewire ≤ 0.014" | Y-Connector (Side arm for access to balloon inflation/deflation lumen. Straight arm is continuous with shaft inner lumen). Designed for guidewire ≤ 0.014" |
| Stent Delivery | A balloon, nominally 0.4 mm longer than the stent, with two radiopaque markers. | |
| Balloon Inflation Pressure | Nominal Inflation Pressure: 9 ATM Rated Burst Inflation Pressure: 18 ATM | |
| Guide Catheter Inner Diameter | ≥ 0.058" | ≥ 0.066" |
| Catheter Shaft Outer Diameter | 1.8F proximally and 2.7F distally | 3.2F proximally, 2.7F distally |
*Release rate is a function of weight/weight ratio of polymer and drug. (SR) is the formulation that was studied clinically and is used in the marketed product
# VI. ALTERNATIVE PRACTICES OR PROCEDURES
Treatment of patients with coronary artery disease may include exercise, diet, drug therapy, percutaneous coronary interventions (such as angioplasty, and placement of bare metal stents, coated stents, and other drug eluting stents), and coronary artery bypass surgery (CABG).
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 3
10
{3}
# VII. MARKETING HISTORY
The TAXUS Liberté Paclitaxel-Eluting Coronary Stent System 2.25 mm size is commercially available in the following countries:
| • Albania | • Dominican Rep | • Kuwait | • Portugal |
| --- | --- | --- | --- |
| • Algeria | • Dutch Antilles | • Latvia | • Qatar |
| • Antigua/Barbuda | • Ecuador | • Lebanon | • Romania |
| • Argentina | • Egypt | • Libya | • Russia |
| • Armenia | • El Salvador | • Liechtenstein | • Saudi Arabia |
| • Aruba | • Estonia | • Lithuania | • Scotland |
| • Australia | • Finland | • Luxembourg | • Serbia/Montenegro |
| • Austria | • France | • Macau | • Singapore |
| • Bahamas | • Georgia | • Macedonia | • Slovakia |
| • Bahrain | • Germany | • Malaysia | • Slovenia |
| • Bangladesh | • Great Britain | • Malta | • South Africa |
| • Barbados | • Greece | • Martinique | • Spain |
| • Belarus | • Guatemala | • Mauritania | • Sri Lanka |
| • Belgium | • Guyana | • Mauritius | • Sudan |
| • Belize | • Haiti | • Mexico | • Suriname |
| • Bermuda | • Honduras | • Moldavia | • Sweden |
| • Bolivia | • Hong Kong | • Morocco | • Switzerland |
| • Bosnia | • Hungary | • Myanmar | • Syria |
| • Brazil | • Iceland | • Nepal | • Taiwan |
| • Brunei | • India | • Netherlands | • Thailand |
| • Bulgaria | • Indonesia | • New Zealand | • Trinidad/Tobago |
| • Chile | • Iran | • Nicaragua | • Tunisia |
| • China | • Iraq | • Norway | • Turkey |
| • Colombia | • Ireland | • Oman | • United Arab Emirates |
| • Costa Rica | • Israel | • Pakistan | • Ukraine |
| • Croatia | • Italy | • Panama | • Uruguay |
| • Cyprus | • Jamaica | • Paraguay | • Venezuela |
| • Czech Republic | • Jordan | • Peru | • Vietnam |
| • Denmark | • Kenya | • Philippines | • West Bank Gaza Strip |
| • Djibouti | • Korea | • Poland | • Yemen |
As of April 2009, over 94,000 TAXUS Liberté 2.25 mm stents have been distributed outside the U.S. No products have been withdrawn from the market in any country for any reason.
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 4 11
{4}
# VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Potential adverse events (in alphabetical order) which may be associated with the use of a coronary stent in native coronary arteries include but are not limited to:
- Abrupt stent closure
- Acute myocardial infarction
- Allergic reaction to anti-coagulant and/or antiplatelet therapy, contrast medium, or stent materials
Angina
- Arrhythmias, including ventricular fibrillation (VF) and ventricular tachycardia (VT)
- Arteriovenous fistula
Cardiac tamponade
Cardiogenic shock/Pulmonary edema
Coronary aneurysm
Death
Dissection
- Emboli, distal (air, tissue or thrombotic material or material from devices(s) used in the procedure)
Heart failure
Hematoma
- Hemorrhage, required transfusion
Hypotension/Hypertension
- Infection, local or systemic
Ischemia, myocardial
- Pain, at the access site
Perforation or Rupture of coronary artery
Pericardial effusion
Pseudoaneurysm, femoral
- Renal Failure
Respiratory Failure
- Restenosis of stented segment
- Stent embolization or migration
- Stent thrombosis/occlusion
Stroke/cerebrovascular accident /TIA
Total occlusion of coronary artery
Vessel spasm
Vessel trauma requiring surgical repair or reintervention
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 5
12
{5}
Potential adverse events not captured above, that may be unique to the paclitaxel drug coating:
- Allergic/immunologic reaction to drug (paclitaxel or structurally-related compounds) or the polymer stent coating (or its individual components)
- Alopecia
- Anemia
- Blood product transfusion
- Gastrointestinal symptoms
- Hematologic dyscrasia (including leukopenia, neutropenia, thrombocytopenia)
- Hepatic enzyme changes
- Histologic changes in vessel wall, including inflammation, cellular damage or necrosis
- Myalgia/Arthralgia
- Peripheral neuropathy
For the specific events that occurred in the clinical study, please see Section X.
# IX. SUMMARY OF PRECLINICAL STUDIES
A series of non-clinical laboratory studies were performed – those related to the stent and the stent delivery system [i.e. the stent on either the Monorail (MR) or Over-The-Wire (OTW) stent delivery system (SDS)], the polymer substance [i.e., poly(styrene-b-isobutylene-b-styrene) (SIBS)], the drug substance (i.e., paclitaxel) and the finished combination product (i.e., TAXUS Liberté Paclitaxel-Eluting Coronary Stent).
# A. Biocompatibility Studies
The biocompatibility testing information included in the original PMA submission (P060008, approved October 10, 2008) is directly applicable to this PMA supplement. Please refer to the original SSED for details regarding the biocompatibility studies.
Additional testing was conducted on SIBS coated Liberté stents to evaluate the stent and stent delivery system separately. These biocompatibility studies are summarized in Table 2.
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 6
13
{6}
Table 2: Biocompatibility Test Summary
| Test Name | Test Description | Test Article and Results |
| --- | --- | --- |
| Systemic Toxicity | ISO 10993-11: Systemic Toxicity (Acute) | Polymer-only coated stent: Pass (non-toxic) |
| | ISO 10993-11: 90-Day Chronic Toxicity (Rat, Subcutaneous Implant) | Polymer-only coated stent: Pass (non-toxic) |
| Pyrogenicity | ISO 10993-11: Systemic Toxicity (Material-Mediated Rabbit) | Polymer-only coated stent: Pass (non-pyrogenic) |
| Hemocompatibility | ISO 10993-4: Complement Activation C3a and SC5b-9 Assay | Polymer-only coated stent: Pass (Negative for C3a and SC5b-9 assays) |
# B. In Vivo Pharmacokinetics
The non-clinical chemistry information included in the original PMA submission (P050008, approved October 10, 2008) are directly applicable to this PMA Supplement; therefore, the information is not repeated herein.
# C. In Vitro Engineering Testing
The in vitro engineering testing included in the original PMA submission (P050008, approved October 10, 2008) is directly applicable to this PMA Supplement. Therefore, only the testing information specific to the 2.25 mm diameter stent are summarized below in Table 3. "Pass" denotes that the test results met product specifications and/or the recommendation in the guidance documents.
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 7 14
{7}
Table 3: Stent and Delivery Catheter Engineering Testing
| Test | Description of Test | Conclusion |
| --- | --- | --- |
| **Stent Dimensional and Functional Attributes** | | |
| Radial Stiffness and Radial Strength | Testing was conducted to determine the ability of the TAXUS Liberté stent to resist collapse under short term and long term loads. | Pass |
| **Stent Delivery System Dimensional and Functional Attributes** | | |
| Delivery, Deployment and Retraction | The delivery, deployment and retraction of the TAXUS Liberté Stent System was assessed by testing guidewire movement at rated burst pressure. Testing demonstrated that the TAXUS Liberté stent system could be delivered to the target location, deployed, and retracted, thus met required product specifications. | Pass |
| Balloon Fatigue | TAXUS Liberté stent systems across the range of stent/balloon lengths and diameters were required to complete 10 pressurization cycles to Rated Burst Pressure (RBP). The results show statistically that, with 95% confidence, 90% of the catheters will not experience balloon, shaft, or proximal/distal seal loss of integrity at or below the maximum recommended rated balloon burst pressure. | Pass |
| Stent Securement for Unsheathed Stents | Testing was conducted to assess the forces required to displace a crimped TAXUS Liberté stent from the delivery systems (1) after tracking through a simulated tortuous artery model and then through a simulated lesion and (2) during non-coaxial withdrawal. All stent systems met specification. | Pass |
| Balloon Catheter Withdrawal Resistance | TAXUS Liberté is to have statistically lower withdrawal forces as compared to TAXUS Express^{2} when analyzed as a pooled population with 95% confidence. | Pass |
| **Stent, System and Coating Durability Testing** | | |
| Accelerated Durability Testing | Overlapping stents in a curve were evaluated in an accelerated in vitro test of approximately 10 year equivalent (400 million cycles) real time and met visual requirements for coating integrity and strut damage. | Pass |
| Coating Durability | The coating durability of the TAXUS Liberté stent coating was assessed pre-expansion and post deployment beyond rated burst pressure. The test results demonstrate that the paclitaxel/SIBS coating displays good durability and coating integrity that will be maintained throughout the lifetime of the coated stent implant. | Pass |
### D. Coating Characterization Testing
The coating characterization information included in the original PMA submission (P060008, approved October 10, 2008) is directly applicable to this PMA supplement. Therefore, only the testing information specific to the 2.25 mm diameter stent is summarized below in Table 4.
Table 4: Coating Characterization Testing
| Test | Description of Test |
| --- | --- |
| **Material Characterization** | |
| Particulates | Particulate levels were evaluated for the TAXUS Liberté stent system post deployment, over-expansion and simulated use. |
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 8
15
{8}
### E. Chemistry Manufacturing and Controls (CMC) Testing
The CMC information included in the original PMA submission (P060008, approved October 10, 2008) is directly applicable to this PMA Supplement; therefore, the information is not repeated herein.
### F. Stability
Stability studies were conducted to establish a shelf life/expiration date for the TAXUS Liberté Atom Paclitaxel-Eluting Coronary Stent System. Based on these studies, a shelf life of 18 months is appropriate.
### G. Sterilization
The TAXUS Liberté Atom Paclitaxel-Eluting Coronary Stent System (Monorail and Over-The-Wire) is sterilized using ethylene oxide sterilization and has been validated per AAMI/ISO 1135:1994 “Medical Devices – Validation and Routine Control of Ethylene Oxide Sterilization.”
Results obtained from the sterilization studies show that the product satisfies a minimum Sterility Assurance Level (SAL) of 10⁻⁶.
The amount of bacterial endotoxin was verified to be within the specification limit for TAXUS Liberté Atom Paclitaxel-Eluting Coronary Stent System.
### H. Animal Studies
The non-clinical studies included in the original PMA submission (P060008, approved October 10, 2008) are directly applicable to this PMA Supplement; therefore, the information is not repeated herein.
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 9
16
{9}
X.
# SUMMARY OF PRIMARY CLINICAL STUDY
# TAXUS ATLAS Small Vessel Trial
The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of coronary drug-eluting stent implantation with the 2.25 mm TAXUS Liberté Coronary Stent System for improving luminal diameter for the treatment of de novo lesions in native coronary arteries 2.25 mm in diameter and ≤ 28mm in length in the US under IDE #G040100. Data from this clinical study were the basis for this PMA approval decision. A summary of the clinical study is presented below.
# Study Design
The TAXUS ATLAS Small Vessel (SV) trial is a prospective, multicenter, single-arm, trial to evaluate the safety and efficacy of the TAXUS Liberté 2.25 mm stent in the treatment of small (RVD ≤ 2.5mm) de novo coronary artery lesions. The primary endpoint of the study is percent diameter stenosis (%DS) of the analysis segment at 9 month follow-up as determined by QCA. Patients were enrolled after successful predilation and received a TAXUS Liberté paclitaxel-eluting stent with a diameter of 2.25 mm. All patients were to have angiographic follow-up at 9 months. A total of 261 patients were enrolled in TAXUS ATLAS SV study from 23 centers in 3 countries (New Zealand, Singapore, and the United States).
The outcomes of TAXUS ATLAS SV study were compared to 2 historical controls derived from the TAXUS V de novo clinical trial.
The first objective of the TAXUS ATLAS Small Vessel study was to demonstrate non-inferiority for the angiographic outcome of %DS for the 2.25 mm TAXUS Liberté stent when compared to the TAXUS Express Stent. Therefore, the treatment group is compared to a case-matched Control group derived from TAXUS V de novo. In order to case-match this first Control group, all TAXUS V patients randomized to the TAXUS Express group with (1) a reference vessel diameter (RVD) by visual estimate ≤ 2.5 mm, (2) a lesion length by visual estimate ≥ 10 mm and ≤ 28 mm, and (3) receiving 1 planned 2.25 mm study stent were included. This resulted in inclusion of 75 out of 577 patients randomized into the TAXUS Express treatment arm of TAXUS V de novo.
The second objective of the TAXUS ATLAS Small Vessel study was to demonstrate superiority for the angiographic outcome of %DS for the 2.25 mm TAXUS Liberté stent when compared to the bare metal Express stent. Therefore, the treatment group is compared to a case-matched Control group derived from TAXUS V de novo. In order to case-match this second Control group, all TAXUS V patients randomized to the bare metal Express group with (1) a reference vessel diameter (RVD) by visual estimate ≤ 2.5 mm, (2) a lesion length by visual estimate ≥ 10 mm and ≤ 28 mm, and (3) receiving 1 planned 2.25 mm or 2.5mm study stent were included. This resulted in
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 10
17
{10}
inclusion of 155 out of 579 patients randomized into the bare metal Express treatment arm of TAXUS V de novo.
Patients were consented, and then were to be followed-up at 1, 4, and 9 months post-index procedure and 1, 2, 3, 4, and 5 years thereafter. The primary endpoint was reached after all enrolled patients had completed the 9 month follow-up.
Patients were to receive aspirin and clopidogrel or ticlopidine pre-procedure. Following the procedure patients were prescribed to mandatory clopidogrel (or ticlopidine) use for 6 months. In addition, patients were prescribed mandatory aspirin use for 9 months, with indefinite use of aspirin highly recommended.
The trial utilized a clinical event committee (CEC) to adjudicate potential Major Adverse Clinical Events (MACE), including all reported death and stent thromboses. During the course of the trial, a data monitoring committee (DMC) reviewed aggregate safety data to monitor for incidence of MACE and other trends that may warrant modification of the trial. A core lab was used for central assessment of angiography.
# 1. Clinical Inclusion and Exclusion Criteria
Enrollment in the TAXUS ATLAS Small Vessel study was limited to patients who met the following inclusion criteria:
1. Patient is \(\geq 18\) years old.
2. Eligible for percutaneous coronary intervention (PCI)
3. Documented stable angina pectoris (Canadian Cardiovascular Society [CCS] Classification 1, 2, 3, or 4) or unstable angina pectoris with documented ischemia (Braunwald Class IB-C, IIB-C, or IIIB-C), or documented silent ischemia
4. Left ventricular ejection fraction (LVEF) of \(\geq 25\%\)
5. Acceptable candidate for coronary artery bypass grafting (CABG)
6. Patient or legal guardian understands the study requirements and the treatment procedures and provides written Informed Consent before any study-specific tests or procedures are performed
7. Willing to comply with all specified follow-up evaluations
Patients were not permitted to enroll in the TAXUS ATLAS Small Vessel study if they met any of the following exclusion criteria:
1. Known sensitivity to paclitaxel
2. Any previous, concurrent or planned treatment with a non-study anti-restenotic drug-coated or DES
3. Planned use of both the study stent and a non-study stent, (i.e., commercial stent) in the treatment of the target vessel
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 11
18
{11}
4. Previous or planned treatment with intravascular brachytherapy in the target vessel
5. Planned CABG \(\leq 9\) -months post-index procedure
6. MI within 72 hours prior to the index procedure and/or creatine kinase (CK) \(>2x\) the local laboratory's ULN (refers to a measured value twice the upper limits of normal on the day of the index procedure) unless CK-MB is \(<2x\) ULN. A CK-MB laboratory result must be available on the day of the procedure prior to treatment for patients with recent MI (72 hours to 30 days) or unstable angina if the CK is \(2x\) ULN. Patients with stable angina may be enrolled without lab results being received; however, baseline CK must be drawn immediately pre-procedure
7. Cerebrovascular Accident (CVA) within the past 6 months
8. Cardiogenic Shock characterized by systolic pressure \(< 80 \mathrm{~mmHg}\) and/or central filling pressure \(>20 \mathrm{~mmHg}\), or cardiac index \(< 1.8\) liters/minute/m² or intra-aortic balloon pump or intravenous inotropes were needed to maintain a systolic pressure \(>80 \mathrm{~mmHg}\) and a cardiac index \(>1.8\) liters/minute/m²
9. Acute or chronic renal dysfunction (creatinine \(>2.0 \mathrm{mg} / \mathrm{dL}\) or \(177 \mu \mathrm{mol} / \mathrm{L}\))
10. Contraindication to ASA, or to both clopidogrel and ticlopidine
11. Leukopenia (leukocyte count \(< 3.5 \times 10^{9} / \text{liter}\))
12. Thrombocytopenia (platelet count \(< 100,000 / \mathrm{mm}^3\)) or thrombocytosis \((>750,000 / \mathrm{mm}^3)\)
13. Active peptic ulcer or active gastrointestinal (GI) bleeding
14. Known allergy to stainless steel
15. Any prior true anaphylactic reaction to contrast agents; defined as known anaphylactoid or other non-anaphylactic allergic reactions to contrast agents that cannot be adequately pre-medicated prior to the index procedure
16. Patient was currently, or had been treated with paclitaxel or other chemotherapeutic agents within 12-months of the index procedure
17. Anticipated treatment with paclitaxel or oral rapamycin during any period in the 9-months after the index procedure
18. Male or female with known intention to procreate within 3 months after the index procedure
19. Female of childbearing potential with a positive pregnancy test within 7 days before the index procedure, or lactating
20. Life expectancy of less than 24 months due to other medical conditions
21. Co-morbid condition(s) that could limit the patient's ability to participate in the study, compliance with follow-up requirements or impact the scientific integrity of the study
22. Currently participating in another investigational drug or device study that has not completed the primary endpoint or that clinically interferes with the endpoints of this study
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 12
19
{12}
## 2. Follow-up Schedule
Follow-up included clinical assessments at 1, 4 and 9 months. All patients were to undergo angiographic follow-up at 9 months. In addition, patients agreed to annual telephone follow-up for clinical parameters through 5 years post-procedure.
## 3. Clinical Endpoints
The primary endpoint for the study was the percent diameter stenosis (%DS) of the analysis segment at 9 months, as determined by QCA. Secondary endpoints included clinical procedural and technical success, utilization parameters (e.g., procedure and fluoroscopic time), MACE rates at each follow-up time point, stent thrombosis rate, target vessel failure, target vessel revascularization, and QCA measurements (binary restenosis, in-stent %DS, MLD and late loss).
### A. Accountability of PMA Cohort
Figure 1: TAXUS ATLAS Small Vessel Group versus DES Control

PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 13
20
{13}
Figure 2: TAXUS ATLAS Small Vessel Group versus BMS Control

# B. Study Population Demographics and Baseline Parameters
TAXUS ATLAS Small Vessel versus DES Control
Baseline characteristics were comparable between the 2 groups with no statistically significant differences in demographics, cardiac history and cardiac risk factors. In terms of comorbidities, more patients in the DES Control Group had a known history of peripheral vascular disease (PVD) (13.3%) as compared with patients in the TAXUS ATLAS Small Vessel Group (5.7%, P=0.0274).
Most of the baseline angiographic lesion characteristics assessed by the Core Lab were comparable between the 2 groups. However, patients in the TAXUS ATLAS Small Vessel Group had lesions that were longer (14.53±6.89 mm) than those observed in the DES Control Group (11.84±5.69 mm, P=0.0025), and a greater proportion had ACC/AHA Category C lesions (34.5% versus 16.2%, respectively, P=0.0026). Lesion location demonstrated a significant difference between the groups, with more target lesions in the TAXUS ATLAS Small Vessel group being in the mid portion of the target vessels, and more lesions in the DES Control group being in the proximal portion of the target vessel. Eccentric lesions were observed more frequently in the DES Control Group (59.5% versus 28.7% for the TAXUS ATLAS Small Vessel Group, P<0.0001). Notably, differences in baseline characteristics were not expected to affect the primary endpoint due to propensity-score adjustment.
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 14
21
{14}
# TAXUS ATLAS Small Vessel versus BMS Control
The groups were well matched with respect to most baseline patient characteristics. As compared with the BMS Control Group, a larger proportion of patients in the TAXUS ATLAS Small Vessel Group were with CCS Class 1 angina (4.5% versus 10.0%, respectively, P=0.0469), however, patients with clinically more significant angina class 2, 3 or 4 were similarly distributed between the 2 groups.
Several differences between study groups were noted. Although inclusion criteria for RVD were identical in the TAXUS ATLAS Small Vessel study and for the TAXUS V patients included in the BMS Control group, based on the angiographic analysis at the Core Lab, patients in the TAXUS ATLAS Small Vessel Group had lesions with a smaller RVD than those in the BMS Control Group (2.02±0.30 mm versus 2.20±0.34 mm, respectively, P<0.0001). Patients in the TAXUS ATLAS Small Vessel Group had a smaller %DS (67.26±10.91% versus 72.10±10.69%, respectively, P<0.0001). Patients with a BMS had more complex target lesions as compared with the TAXUS ATLAS Small Vessel Group (59.4% were eccentric lesions versus 28.7% for patients in the TAXUS ATLAS Small Vessel Group, P<0.0001, and 79.4% were classified as B2/C as compared with 69.0% in the TAXUS ATLAS Small Vessel Group, p=0.0213). These differences were not expected to affect outcomes variables, as propensity score adjustments were made.
For patients with only study stents implanted, maximum diameter implanted was greater in the BMS Control Group (2.39±0.17 mm) versus the TAXUS ATLAS Small Vessel Group (2.25±0.00 mm, P<0.0001), as expected, due to the differences in protocol design. The total length implanted in patients in the BMS Control Group (23.3±7.9 mm) was greater than that implanted in the TAXUS ATLAS Small Vessel Group (21.5±7.0 mm, P=0.0173).
# C. Safety and Effectiveness Results
Major Adverse Cardiac Events (MACE) reported in the TAXUS ATLAS Small Vessel Study are presented in Table 5 below. Please note that this study was not powered to assess clinical outcomes; however, the following safety outcomes were observed:
- MI rates (both Q-wave and non-Q-wave) were comparable between the two groups at 9 months (overall MI rate was \(2.7\%\) in the TAXUS ATLAS Small Vessel group versus \(4.1\%\) in the DES Control group).
- Cardiac death rates were also comparable between the two groups at 9 months (0.8% in the TAXUS ATLAS Small Vessel group versus 2.7% in the DES Control group).
- Stent thrombosis events were comparable between the two groups at 9 months (0.4% in the TAXUS ATLAS Small Vessel group versus 1.4% in the DES Control group).
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 15
22
{15}
Table 5: TAXUS ATLAS Small Vessel Major Adverse Cardiac Events (MACE) From Post-Procedure to Latest Follow-Up
| | TAXUS ATLAS Small Vessel 2.25 mm to 1 Year | | |
| --- | --- | --- | --- |
| | TAXUS ATLAS Small Vessel (N=261) | TAXUS Express DES Control (N=75) | Express BMS Control (N=155) |
| In-Hospital MACE | 1.9% (5/261) | 2.7% (2/75) | 1.9% (3/155) |
| 30-Day MACE, overall | 1.9% (5/261) | 4.1% (3/74) | 2.6% (4/154) |
| 9-Month MACE, overall | 12.8% (33/258) | 20.5% (15/73) | 21.6% (33/153) |
| Cardiac Death | 0.8% (2/258) | 2.7% (2/73) | 0.7% (1/153) |
| MI | 2.7% (7/258) | 4.1% (3/73) | 2.6% (4/153) |
| Q-Wave MI | 0.8% (2/258) | 1.4% (1/73) | 0.0% (0/153) |
| Non-Q-Wave MI | 1.9% (5/258) | 2.7% (2/73) | 2.6% (4/153) |
| TVR, Overall | 10.1% (26/258) | 17.8% (13/73) | 19.6% (30/153) |
| TLR, Overall | 5.8% (15/258) | 13.7% (10/73) | 17.6% (27/153) |
| Non-TLR, Overall | 6.6% (17/258) | 6.8% (5/73) | 5.9% (9/153) |
| 1-Year MACE | 13.4% (33/247) | 26.8% (19/71) | 28.4% (42/148) |
| Cardiac Death | 1.2% (3/247) | 4.2% (3/71) | 0.7% (1/148) |
| MI | 2.4% (6/247) | 4.2% (3/71) | 2.7% (4/148) |
| Q-Wave MI | 0.8% (2/247) | 1.4% (1/71) | 0.0% (0/148) |
| Non-Q-Wave MI | 1.6% (4/247) | 2.8% (2/71) | 2.7% (4/148) |
| TVR, Overall | 10.5% (26/247) | 22.5% (16/71) | 26.4% (39/148) |
| TLR, Overall | 6.1% (15/247) | 16.9% (12/71) | 22.3% (33/148) |
| Non-TLR, Overall | 6.9% (17/247) | 8.5% (6/71) | 8.8% (13/148) |
| 1-Year Stent Thrombosis | 0.4% (1/243) | 1.5% (1/67) | 1.4% (2/146) |
### TAXUS ATLAS Small Vessel versus DES Control
The pre-specified primary endpoint of non-inferiority of in-segment %DS in the 2.25 mm TAXUS Liberté stent versus the TAXUS Express stent was met with propensity score adjustment. The 9-month difference in the in-segment %DS was -7.3%. As the upper 1-sided 95% confidence bound for the difference was -0.8%, this was less than the pre-specified non-inferiority margin of 10.00% (P<0.0001). In addition, the primary endpoint was met without propensity score adjustment. Non-inferiority testing was also performed in the ITT population, which was met with and without adjustment for propensity score (Table 6).
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 16
23
{16}
Table 6: TAXUS ATLAS Small Vessel Primary Endpoint with DES Control
| Per Protocol Population | TAXUS Liberté^{®} Atom^{TM} (N=254) | TAXUS^{®} Express^{®} DES Control (N=73) | Difference [Upper 1-Sided 95% CL] | P-Value | Δ |
| --- | --- | --- | --- | --- | --- |
| Follow-up In-Segment Percent Diameter Stenosis | | | | | |
| Adjusted for the propensity score | 32.2 | 39.6 | -7.3 [-0.8] | <0.0001 | 10.00% |
| Unadjusted | 31.70±18.23 (207) (4.07, 100.00) | 37.69±23.32 (54) (5.36, 100.00) | -5.99 [-1.10] | <0.0001* | 10.00% |
| | | | | | |
| Intent-to-Treat Population | TAXUS Liberté Atom (N=261) | TAXUS Express DES Control (N=75) | Difference [Upper 1-Sided 95% CL] | P-Value | Δ |
| Follow-up In-Segment Percent Diameter Stenosis | | | | | |
| Adjusted for the propensity score | 32.4 | 40.1 | -7.7 [-1.1] | <0.0001 | 10.00% |
| Unadjusted | 32.09±18.38 (211) (4.07, 100.00) | 38.36±23.64 (55) (5.36, 100.00) | -6.27 [-1.38] | <0.0001* | 10.00% |
*Variances unequal: Satterthwaite's approximate t statistic.
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 17
24
{17}
Figure 3: Cumulative Frequency Distribution of In-Stent Percent Diameter Stenosis by QCA, Intent-to-Treat, All Patients (N=336)

| | Post-Procedure | | Follow-Up | | Pre-Procedure | |
| --- | --- | --- | --- | --- | --- | --- |
| | TAXUS® Liberté® Atom™ (N=261) | TAXUS® Express® DES Control (N=75) | TAXUS Liberté Atom (N=261) | TAXUS Express DES Control (N=75) | TAXUS Liberté Atom (N=261) | TAXUS Express DES Control (N=75) |
| N | 255 | 72 | 207 | 54 | 261 | 74 |
| Median | 9.43 | 7.41 | 20.04 | 17.40 | 67.45 | 68.56 |
| Minimum | -29.41 | -21.17 | -16.12 | -18.31 | 36.49 | 41.29 |
| Maximum | 38.19 | 26.90 | 100.00 | 100.00 | 95.37 | 100.00 |
| Mean | 8.64 | 6.36 | 23.35 | 27.73 | 67.26 | 69.97 |
| SD | 11.10 | 9.60 | 20.89 | 28.12 | 10.91 | 10.59 |
| COV | 128.51% | 150.91% | 89.47% | 101.43% | 16.22% | 15.14% |
| Diff (95% CI) | 2.28 [-0.55, 5.10] | | -4.38 [-11.14, 2.38] | | -2.71 [-5.51, 0.08] | |
COV = coefficient of variation.
Post-Procedure and Follow-up measurements are in-stent measurements.
Clinical evaluations and survival analyses demonstrate that MACE, TVR, and TVF were similar for patients in the TAXUS ATLAS Small Vessel Group and the DES Control Group. The 9-month TLR rates were statistically significantly lower in TAXUS ATLAS Small Vessel Group (5.8%) as compared with the DES Control Group (13.7%, P=0.0244) (Table 7). After propensity score adjustment, the differences no longer remained statistically significant (6.5% versus 12.6%, P=0.1472). The safety profile of the 2.25 mm TAXUS Liberté stent is similar to that of the DES Control stent. The rates for MACE and MACE components (cardiac death, MI, and TVR) demonstrate that the TAXUS Liberté stent is as safe as the commercially-available TAXUS Express stent for use in small vessels.
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 18
25
{18}
Table 7: TAXUS ATLAS Small Vessel Clinical Results
| | 9 months (ITT population) | | | 1 year (per protocol population**) | | |
| --- | --- | --- | --- | --- | --- | --- |
| | TAXUS ATLAS Small Vessel (N=261) | TAXUS Express DES Control (N=75) | P-Value | TAXUS ATLAS Small Vessel (N=254) | TAXUS Express DES Control (N=73) | P-Value |
| **EFFICACY** | | | | | | |
| TVR, Overall | 10.1% (26/258) | 17.8% (13/73) | 0.0705 | 10.5% (26/247) | 22.5% (16/71) | 0.0084 |
| TLR, Overall | 5.8% (15/258) | 13.7% (10/73) | 0.0244 | 6.1% (15/247) | 16.9% (12/71) | 0.0039 |
| TLR, PCI | 5.8% (15/258) | 12.3% (9/73) | 0.0581 | 6.1% (15/247) | 15.5% (11/71) | 0.0107 |
| TLR, CABG | 0.0% (0/258) | 1.4% (1/73) | 0.2205* | 0.0% (0/247) | 1.4% (1/71) | 0.2233* |
| Non-TLR, Overall | 6.6% (17/258) | 6.8% (5/73) | 1.0000* | 6.9% (17/247) | 8.5% (6/71) | 0.6530 |
| Non-TLR, PCI | 6.6% (17/258) | 6.8% (5/73) | 1.0000* | 6.9% (17/247) | 8.5% (6/71) | 0.6530 |
| Non-TLR, CABG | 0.4% (1/258) | 0.0% (0/73) | 1.0000* | 0.4% (1/247) | 0.0% (0/71) | 1.0000* |
| **SAFETY** | | | | | | |
| Total Death | 1.2% (3/259) | 2.7% (2/73) | 0.3035* | 2.8% (7/249) | 4.3% (3/69) | 0.4569* |
| Cardiac Death or MI | 3.5% (9/258) | 5.5% (4/73) | 0.4938* | 3.6% (9/247) | 7.0% (5/71) | 0.3204* |
| Cardiac Death | 0.8% (2/258) | 2.7% (2/73) | 0.2123* | 1.2% (3/247) | 4.2% (3/71) | 0.1275* |
| MI | 2.7% (7/258) | 4.1% (3/73) | 0.4643* | 2.4% (6/247) | 4.2% (3/71) | 0.4234* |
| Q-wave MI | 0.8% (2/258) | 1.4% (1/73) | 0.5277* | 0.8% (2/247) | 1.4% (1/71) | 0.5327* |
| Non-Q-wave MI | 1.9% (5/258) | 2.7% (2/73) | 0.6519* | 1.6% (4/247) | 2.8% (2/71) | 0.6187* |
| Stent Thrombosis | 0.4% (1/256) | 1.4% (1/72) | 0.3914* | 0.4% (1/243) | 1.5% (1/67) | 0.3861* |
* P-values are two-sided from Fisher's exact test; p-values without * are two-sided from the Chi-square test.
**After 9 months, the TAXUS ATLAS Small Vessel study population was reduced to a pre-specified cohort (per protocol population), which consists of all patients who received a study stent at baseline
The secondary endpoints included several angiographic parameters, including in-stent percent diameter stenosis, binary restenosis, MLD, and late loss. Nine-month MLD in the analysis segment was statistically significantly greater in the TAXUS ATLAS Small Vessel Group versus the DES Control Group (1.41±0.45 mm versus 1.26±0.51 mm, respectively, P=0.0323). Late loss (both in-stent and in the analysis segment) was statistically significantly lower for patients in the TAXUS ATLAS Small Vessel Group versus the DES Control Group (in-stent late loss: 0.28±0.45 mm for the TAXUS ATLAS Small Vessel Group versus 0.44±0.61 mm for the DES
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 19
26
{19}
Control Group, P=0.0297; analysis segment late loss 0.16±0.40 mm for the TAXUS ATLAS Small Vessel Group versus 0.33±0.52 mm for the DES Control Group, P=0.0085), although these p-values have not been adjusted for multiple comparisons. A summary of the angiographic results are provided in Table 8.
Table 8: TAXUS ATLAS Small Vessel 9-Month Angiographic Results
| Angiographic Outcomes* | TAXUS ATLAS Small Vessel (N=261) | TAXUS Express DES Control (N=75) | P-Value |
| --- | --- | --- | --- |
| MLD (mm), In-stent | | | |
| Post-Procedure | 1.83±0.26(255) | 1.93±0.26(72) | 0.1603 |
| 9-Month | 1.59±0.48(207) | 1.47±0.60(54) | 0.1365 |
| MLD (mm), Analysis Segment | | | |
| Post-Procedure | 1.53±0.33(261) | 1.59±0.34(74) | 0.8459 |
| 9-Month | 1.41±0.45(211) | 1.26±0.51(55) | 0.0323 |
| % DS, In-stent | | | |
| Post-Procedure | 8.64±11.10(255) | 6.36±9.60(72) | 0.1148 |
| 9-Month | 23.33±20.89(207) | 27.73±28.12(54) | 0.2051 |
| % DS, Analysis Segment | | | |
| Post-Procedure | 23.96±10.38(261) | 23.59±10.68(74) | 0.7918 |
| 9-Month | 32.09±18.38(211) | 38.36±23.64(55) | 0.0351 |
| Late Loss, In-stent (mm) | 0.23±0.45(207) | 0.44±0.61(54) | 0.0297 |
| Late Loss, Analysis Segment (mm) | 0.16±0.40(211) | 0.33±0.52(55) | 0.0085 |
| Binary Restenosis | | | |
| In-stent restenosis | 13.0% (27/207) | 25.9% (14/54) | 0.0205 |
| Analysis segment restenosis | 18.5% (39/211) | 32.7% (18/55) | 0.0219 |
*Includes all patients in the angiographic subset.
P-values are not adjusted for multiple comparisons.
This trial was not adequately powered to compare the rate of low frequency events to a bare metal control group, nor was it sized to determine the rate of low frequency events with a pre-specified precision.
### TAXUS ATLAS Small Vessel versus BMS Control
The pre-specified primary endpoint of superiority of in-segment %DS in the 2.25 mm TAXUS Liberté stent versus the bare metal Express stent with propensity score adjustment was met. The 9-month %DS in the TAXUS ATLAS Small Vessel Group was 31.9% versus 45.3% in the BMS Control Group (P<0.0001). In addition, the primary endpoint was met without propensity score adjustment (Table 9).
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 20
27
{20}
Table 9: TAXUS ATLAS Small Vessel Primary Endpoint with BMS Control
| Per Protocol Population | TAXUS Liberté^{®} Atom^{TM} (N=254) | Express^{®} BMS Control (N=152) | Difference [95% CI] | P-Value |
| --- | --- | --- | --- | --- |
| Follow-up In-Segment Percent Diameter Stenosis | | | | |
| Adjusted for the propensity score | 31.9 | 45.3 | -13.4 [-18.7,-8.0] | <0.0001 |
| Unadjusted | 31.70±18.23 (207) (4.07, 100.00) | 45.61±23.48 (105) (7.29, 100.00) | -13.91 [-18.64, -9.18] | <0.0001 |
| | | | | |
| Intent-to-Treat Population | TAXUS Liberté Atom (N=261) | Express BMS Control (N=155) | Difference [Upper 1-Sided 95% CL] | P-Value |
| Follow-up In-Segment Percent Diameter Stenosis | | | | |
| Adjusted for the propensity score | 31.9 | 45.9 | -13.9 [-19.6,-8.3] | <0.0001 |
| Unadjusted | 32.09±18.38 (211) (4.07, 100.00) | 45.61±23.48 (105) (7.29, 100.00) | -13.53 [-18.26, -8.80] | <0.0001 |
### Subgroup Analyses
There were no subgroup analyses performed in the TAXUS ATLAS Small Vessel clinical study.
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 21
28
{21}
# XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION
In accordance with the provisions of section 515(c)(2) of the Act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Circulatory Systems Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
# XII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES
The data in this application support the reasonable assurance of safety and effectiveness of the TAXUS Liberté Atom MR and OTW Coronary Stent Systems when used in accordance with the indications for use. Non-clinical studies, including biocompatibility, pharmacokinetics, engineering testing, coating characterization, chemistry, manufacturing and controls information, and stability testing demonstrate that the stent is adequate and will perform as intended. Clinical study data demonstrate that the TAXUS Liberté Atom MR and OTW Coronary Stent Systems are safe and effective for their intended use. The TAXUS Liberté Atom stent was determined to be statistically non-inferior to the TAXUS Express small vessel stent and statistically superior to the Express bare metal stent with respect to in-segment percent diameter stenosis, the primary study endpoint. In addition, the rates for MACE and MACE components (cardiac death, MI, and TVR) demonstrate that the TAXUS Liberté Atom stent was comparable to the TAXUS Express small vessel stent.
# XIII. CDRH DECISION
CDRH issued an approval order on May 21, 2009. The final conditions of approval cited in the approval order are described below:
The applicant should collect and report to the Agency on an annual basis clinical outcomes through 5 years post-procedure on at least 80% of patients enrolled (excluding those discontinued due to death) from the TAXUS ATLAS Small Vessel study. When appropriate or as requested by FDA, the applicant should submit PMA supplements requesting approval to update their IFU to include these data.
The applicant's manufacturing and sterilization facilities were inspected and were found to be in compliance with the Quality System Regulation (21 CFR 820 and pharmaceutical current Good Manufacturing Practice (cGMP) regulations.
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 22
29
{22}
# XIV. APPROVAL SPECIFICATIONS
Directions for Use: See product labeling.
Hazard to Health from Use of the Product: See Indications, Contraindications, Warnings, Precautions and Adverse Events in the device labeling.
Post Approval Requirements and Restrictions: See approval order.
PMA P060008/S008: FDA Summary of Safety and Effectiveness Data
page 23
30