STARCLOSE VASCULAR CLOSURE SYSTEM

P050007S001 · Abbott Vascular Devices · MGB · Feb 2, 2007 · Cardiovascular

Device Facts

Record IDP050007S001
Device NameSTARCLOSE VASCULAR CLOSURE SYSTEM
ApplicantAbbott Vascular Devices
Product CodeMGB · Cardiovascular
Decision DateFeb 2, 2007
DecisionAPPR
Device ClassClass 3
AttributesTherapeutic

Indications for Use

The StarClose™ Vascular Closure System is indicated for the percutaneous closure of common femoral artery access sites while reducing times to hemostasis and ambulation, in patients who have undergone diagnostic or interventional endovascular catheterization procedures utilizing a 5F or 6F procedural sheath. The StarClose™ Vascular Closure System is indicated for the percutaneous closure of common femoral artery access sites while reducing time to dischargeability in patients who have undergone diagnostic endovascular catheterization procedures utilizing a 5F or 6F procedural sheath.

Device Story

Vascular closure system delivers nitinol clip to common femoral artery access sites post-catheterization. System includes clip applier and exchange system. User positions applier via exchange or introducer sheath; distal vessel locator (collapsible nitinol bands in 'X' shape) expands inside vessel; locator pulled against inner wall to align device; clip deployed to approximate arterial puncture edges; locator retracts for removal. Used in clinical settings by physicians. Reduces time to hemostasis and ambulation compared to manual compression. Benefits include faster patient recovery and discharge.

Clinical Evidence

Pivotal prospective, multi-center, open-label, randomized clinical trial (N=483 total; 275 interventional). Interventional arm compared StarClose (n=184) to standard compression (n=91). Primary safety endpoint: major vascular complications at 30 days (1.1% StarClose vs 1.1% control, p=1.000). Effectiveness: Mean time to hemostasis 7.95 min (StarClose) vs 29.06 min (control), p<0.001. Subset analysis of 86 patients receiving GP IIb/IIIa inhibitors showed no significant difference in complication rates.

Technological Characteristics

Implantable nitinol clip; clip applier with collapsible nitinol vessel locator; 5F/6F exchange system. Mechanical actuation; extravascular closure principle. Standalone device.

Indications for Use

Indicated for percutaneous closure of common femoral artery access sites in patients undergoing diagnostic or interventional endovascular catheterization using 5F or 6F sheaths. Reduces time to hemostasis, ambulation, and dischargeability. No known contraindications.

Submission Summary (Full Text)

{0} # Summary of Safety and Effectiveness Data # I. General Information Device Generic Name: Vascular Hemostasis Device Device Trade Name: StarClose™ Vascular Closure System Applicant: Abbott Vascular Devices 400 Saginaw Drive Redwood City, California 94063 Premarket Approval Application (PMA) Number: P050007/S001 Date of Panel Recommendation: None Date of Notice of Approval to Applicant: February 2, 2007 On December 21, 2005, the StarClose Vascular Closure System (P050007) was approved for the percutaneous closure of common femoral artery access sites while reducing times to hemostasis, ambulation and dischargeability in patients who have undergone diagnostic endovascular catheterization procedures utilizing a 5 F or 6 F procedural sheath. For more information on the data which supported the original indication, the summary of safety and effectiveness data (SSED) to the original PMA should be referenced. Written requests for copies of the summary of safety and effectiveness data can be obtained from the Dockets Management Branch (HFA-305), Food and Drug Administration, 12420 Parklawn Drive, rm. 1-23, Rockville, MD 20857 and may be found on the FDA CDRH Internet Homepage located at http://www.fda.gov/cdrh/pmapage.html. The sponsor submitted this supplement to expand the indication for use to include the percutaneous closure of common femoral artery access sites while reducing times to hemostasis, and ambulation in patients who have undergone interventional endovascular catheterization procedures utilizing a 5 Fr. or 6 Fr. procedural sheath. # II. Indications for Use The StarClose™ Vascular Closure System is indicated for the percutaneous closure of common femoral artery access sites while reducing times to hemostasis and ambulation, in patients who have undergone diagnostic or interventional endovascular catheterization procedures utilizing a 5F or 6F procedural sheath. The StarClose™ Vascular Closure System is indicated for the percutaneous closure of common femoral artery access sites while reducing time to dischargeability in patients who have undergone diagnostic endovascular catheterization procedures utilizing a 5F or 6F procedural sheath. 8 {1} ### III. Contraindications There are no known contraindications to the use of the StarClose™ Vascular Closure System. ### IV. Warnings and Precautions The Warnings and Precautions can be found in the StarClose™ Vascular Closure System labeling. ### V. Device Description #### A. Materials and Configuration The StarClose™ Vascular Closure System is designed to deliver a nitinol clip to close femoral artery access sites following percutaneous catheterization procedures. The StarClose™ Vascular Closure System consists of the StarClose Clip Applier and a StarClose 6F Exchange System. An implantable Clip is mounted on the Clip Applier, which delivers the Clip through the exchange system or introducer sheath for extravascular closure of access sites. The StarClose™ Vascular Closure System can also be used with the StarClose™ 6F Introducer Set, which is packaged and sold separately. #### B. Principles of Operation for the StarClose™ Vascular Closure System: At the end of the endovascular diagnostic or interventional procedure the user ensures placement of either a StarClose™ Exchange Sheath or StarClose™ Introducer Sheath at the access site. Either sheath is used to introduce and position the StarClose™ Clip Applier. The distal end of the Clip Applier features a vessel locator. The locator is designed as collapsible nitinol bands that extend into an “X” shape. The expanded locator is gently pulled until it meets with the inner surface of the vessel wall. After aligning the clip applier, the clip is deployed, drawing the edges of the arterial puncture together. The vessel locator simultaneously retracts so that the device may be removed. ### VI. Alternative Practices and Procedures Alternative practices for achieving hemostasis of the femoral artery puncture site post-catheterization include manual compression, mechanical compression, collagen-based hemostasis devices, and percutaneous delivery of sutures to the femoral artery access site. Pressure dressings and sandbags are routinely used in combination with compression methods to control oozing. P050007/S1 SSED Page 2 of 11 2 9 {2} # VII. Marketing History The StarClose™ Vascular Closure System has been marketed in the United States for the same intended use for the diagnostic patient population under P050007. Refer to P050007 SSED for foreign marketing information. The StarClose™ Vascular Closure System has not been withdrawn from marketing for any reason relating to the safety and effectiveness of the device. # VIII. Potential Adverse Effects of the Device on Health The use of the StarClose™ Vascular Closure System in diagnostic catheterization patients was evaluated in a pivotal, prospective, multi-center, open-label, randomized clinical study involving 208 diagnostic patients and 275 interventional patients (483 total randomized patients) enrolled at 17 United States clinical centers. The first randomized patient was enrolled on 3/15/04 and enrollment in the interventional arm of the study was completed on 11/11/04. In the interventional arm the StarClose device was compared to standard compression (SC) methods following cardiac and peripheral vascular catheterization procedures utilizing 5F and 6F sheath sizes. The interventional patients were randomized using a 2:1 scheme (StarClose device vs. SC control). Of the 275 interventional patients, 184 patients (66.9%) were randomized to the StarClose device and 91 patients (33.1%) were randomized to SC. A subset of 86 patients in the interventional arm received GP IIb/IIIa inhibitors. Of the 86 interventional patients, 56 patients (65.1%) were randomized to the StarClose device and 30 patients (34.9%) were randomized to SC. All primary analyses comparing the 2 randomized groups were based on an intent-to-treat (ITT) analysis in which patients were assigned to the treatment group to which they were randomized. The 30-day safety and effectiveness results for the interventional subjects assigned to the StarClose VCS compared favorably to the control group. For all subjects within each treatment group, the major vascular complications rate was 2/184 (1.1%) for the StarClose VCS group and 1/91 (1.1%) for the control group (p=1.000). The total number of major events in the StarClose group was 3 due to 1 subject who had 2 major vascular complications. The numbers and percentages of major and minor complications for the interventional patients in the clinical study are shown in Table 1 and Table 2. This interventional arm was further categorized into 86 subjects receiving glycoprotein (GP) IIb/IIIa inhibitors and 189 subjects not receiving GP IIb/IIIa inhibitors during their procedures. For the group of subjects in whom GP IIb/IIIa inhibitors was administered, the major and minor vascular complication rates were 3.6% (3/56) and 8.9% (5/56), respectively, for subjects receiving StarClose, and 0.0% (0/30) and 13.3% (4/30), respectively, for control subjects. The differences between the treatment arms were non-significant with p=0.540 for the major vascular complication rate and p=0.713 for the minor vascular complication rate. P050007/S1 SSED Page 3 of 11 3 10 {3} Table 1: Major and Minor Complications Through 30 Days – Interventional ITT Patients | Description of Event | CLIP Device (N=184) | Standard Compression (N=91) | All Patients (N=275) | Difference [95% C.I.] | P-value | | --- | --- | --- | --- | --- | --- | | Major Vascular Complications (Composite) | 1.1% (2/184) | 1.1% (1/91) | 1.1% (3/275) | -0.0% [-4.9%, 2.9%] | 1.000 | | Vascular Injury Requiring Repair | 0.5% (1/184) | 0.0% (0/91) | 0.4% (1/275) | 0.5% [-3.5%, 3.0%] | 1.000 | | Surgery* | 0.5% (1/184) | 0.0% (0/91) | 0.4% (1/275) | 0.5% [-3.5%, 3.0%] | 1.000 | | Angioplasty | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | | Other Percutaneous Procedure | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | | New Ipsilateral Lower Extremity Ischemia | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | | Access Site-related Bleeding Requiring Transfusion*/**/*** | 1.1% (2/184) | 1.1% (1/91) | 1.1% (3/275) | -0.0% [-4.9%, 2.9%] | 1.000 | | Antibiotics or Prolonged Hospitalization | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | | Access Site-related Nerve Injury Requiring Intervention | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | | **Complications** | | | | | | | Death | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | | Minor Vascular Complications (Composite) | 4.3% (8/184) | 9.9% (9/91) | 6.2% (17/275) | -5.5% [-13.7%, 0.6%] | 0.107 | | Pseudoaneurysm** | 0.0% (0/184) | 1.1% (1/91) | 0.4% (1/275) | -1.1% [-6.0%, 1.1%] | 0.331 | | Arteriovenous Fistula | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | | Hematoma (>=6 cm)*** | 4.3% (8/184) | 7.7% (7/91) | 5.5% (15/275) | -3.3% [-11.0%, 2.3%] | 0.268 | | Late access site-related bleeding | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | | Transient lower extremity ischemia | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | | Ipsilateral deep vein thrombosis | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | | Access site-related nerve injury w/o intervention | 0.0% (0/184) | 1.1% (1/91) | 0.4% (1/275) | -1.1% [-6.0%, 1.1%] | 0.331 | | Access site-related vessel injury | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | | Access site wound dehiscence | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | | Ultrasound Compression or Thrombin Injection**/*** | 0.0% (0/184) | 1.1% (1/91) | 0.4% (1/275) | -1.1% [-6.0%, 1.1%] | 0.331 | | Re-bleeding at time of first ambulation, req.>30 min. for re-hemostasis | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | | Localized access site infection treated with IM or oral antibiotics | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | | UADE | 0.0% (0/184) | 0.0% (0/91) | 0.0% (0/275) | 0.0% [-4.1%, 2.0%] | -- | Numbers are % (events/sample size). 95% Confidence Interval of difference 0.0% was provided using Newcombe approach. P-value was provided using Fisher's Exact Test. *One subject had 2 major vascular events (Surgery and Access site-related bleeding requiring transfusion) and 1 minor vascular event (Hematoma >=6cm). **One subject had 1 major vascular event (Access site-related bleeding requiring transfusion) and 3 minor vascular events (Pseudoaneurysm and 2 occurrences of Ultrasound Compression or Thrombin injection). ***One subject had 1 major vascular event (Access site-related bleeding requiring transfusion) and 1 minor vascular event (Hematoma >=6cm) ***The protocol states ultrasound compression and thrombin injection were classified as major vascular complications, however, these have since been reclassified as minor vascular complications per the FDA. P050007/S1 SSED Page 4 of 11 4 11 {4} Table 2: Major and Minor Complications Through 30 Days – Interventional ITT Patients Receiving Glycoprotein IIb/IIIa Inhibitors | Description of Event | StarClose (N=56) | Standard Compression (N=30) | All Subjects (N=86) | Difference [95% C.I.] | P-value | | --- | --- | --- | --- | --- | --- | | Major Vascular Complications (Composite) | 3.6% (2/56) | 0.0% (0/30) | 2.3% (2/86) | 3.6% [-8.1%, 12.1%] | 0.540 | | Vascular Injury Requiring Repair | 1.8% (1/56) | 0.0% (0/30) | 1.2% (1/86) | 1.8% [-9.7%, 9.4%] | 1.000 | | Surgery* | 1.8% (1/56) | 0.0% (0/30) | 1.2% (1/86) | 1.8% [-9.7%, 9.4%] | 1.000 | | Angioplasty | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | Other Percutaneous Procedure | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | New Ipsilateral Lower Extremity Ischemia | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | Access Site-related Bleeding Requiring Transfusion*/** | 3.6% (2/56) | 0.0% (0/30) | 2.3% (2/86) | 3.6% [-8.1%, 12.1%] | 0.540 | | Antibiotics or Prolonged Hospitalization | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | Access Site-related Nerve Injury Requiring Intervention | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | **Complications** | | | | | | | Death | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | Minor Vascular Complications (Composite) | 8.9% (5/56) | 13.3% (4/30) | 10.5% (9/86) | 4.4% [-18.7%, 9.9%] | 0.713 | | Pseudoaneurysm | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | Arteriovenous Fistula | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | Hematoma (>=6 cm)** | 8.9% (5/56) | 13.3% (4/30) | 10.5% (9/86) | 4.4% [-18.7%, 9.9%] | 0.713 | | Late access site-related bleeding | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | Transient lower extremity ischemia | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | Ipsilateral deep vein thrombosis | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | Access site-related nerve injury w/o intervention | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | Access site-related vessel injury | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | Access site wound dehiscence | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | Ultrasound Compression or Thrombin Injection*** | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | Re-bleeding at time of first ambulation req.>30 min. for re-hemostasis | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | Localized access site infection treated with IM or oral antibiotics | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | | UADE | 0.0% (0/56) | 0.0% (0/30) | 0.0% (0/86) | 0.0% [-11.4%, 6.4%] | .. | Numbers are % (events/sample size). 95% Confidence Interval of difference 0.0% was provided using Newcombe approach. P-value was provided using Fisher's Exact Test. *One StarClose subject 16-301 had 2 major vascular events (Surgery and Access site-related bleeding requiring transfusion) and 1 minor vascular event (Hematoma >=6cm). **One StarClose subject 03-313 had 1 major vascular event (Access site-related bleeding requiring transfusion) and 1 minor vascular event (Hematoma >=6cm)/ ***The protocol states ultrasound compression and thrombin injection were classified as major vascular complications, however, these have since been reclassified as minor vascular complications per the FDA. P050007/S1 SSED Page 5 of 11 5 12 {5} # IX. Summary of Preclinical Studies Please refer to the preclinical data in the SSED for the original PMA (P050007). No additional preclinical studies were required for this supplement. # X. Clinical Studies # A. StarClose™ Vascular Closure System U.S. IDE Multi-Center, Randomized Clinical Trial The use of the StarClose Vascular Closure System in diagnostic and interventional catheterization patients was evaluated in a pivotal, prospective, multi-center, open-label, randomized clinical study involving 208 diagnostic patients and 275 interventional patients (483 total randomized patients) enrolled at 17 United States clinical centers. The first randomized patient was enrolled on 3/15/04. Enrollment in the interventional arm of the study was completed on 11/11/04. In the interventional arm the StarClose device was compared to standard compression (SC) methods following cardiac and peripheral vascular catheterization procedures utilizing 5F and 6F sheath sizes. The interventional patients were randomized using a 2:1 scheme (StarClose device vs. SC control). Of the 275 interventional patients, 184 patients (66.9%) were randomized to the StarClose device and 91 patients (33.1%) were randomized to SC. A subset of 86 patients in the interventional arm received GP IIb/IIIa inhibitors. Of the 86 interventional patients, 56 patients (65.1%) were randomized to the StarClose device and 30 patients (34.9%) were randomized to SC. All primary analyses comparing the 2 randomized groups were based on an intent-to-treat (ITT) analysis in which patients were assigned to the treatment group to which they were randomized. The randomized interventional patients in the study had to meet general inclusion criteria, general exclusion criteria, access site exclusion criteria (including some criteria evaluated via limited femoral artery angiogram), and procedural exclusion criteria. The interventional patients consisted of 80.4% men and ranged in age from 39 to 81. The interventional patients who were randomized to the StarClose device were asked to ambulate 4 hours after the interventional procedure was complete, and the interventional patients who were randomized to SC were ambulated according to institutional standards and guidelines. The primary safety endpoint for the study was the combined rate of major complications within 30 ± 7 days following the catheterization procedure. The secondary safety endpoint for the study was the combined rate of minor complications within 30 ± 7 days following the catheterization procedure. The null hypothesis for safety was that the StarClose Vascular Closure System had a primary safety endpoint rate exceeding that of the standard of care (standard P050007/S1 SSED Page 6 of 11 6 13 {6} compression) by delta. The alternative hypothesis was that the StarClose Vascular Closure System had a primary safety endpoint rate less than that of standard compression or exceeding that of standard compression by no more than delta; i.e., $$H_0: \pi_{IC} > \pi_{SC} + \delta$$ $$H_a: \pi_{IC} \leq \pi_{SC} + \delta$$ where $\pi_{IC}$ was the primary endpoint rate estimated for the StarClose Vascular Closure System and $\pi_{SC}$ was the primary endpoint rate estimated for the standard of care (standard compression). For the interventional patients, the StarClose device also demonstrated safety. By Day 30, a combined total of 2 (1.1%) major complications was reported for the randomized interventional patients who received the StarClose device, and a combined total of 1 (1.1%) major complications was reported for the randomized interventional patients who received SC. For the interventional patients, the rates of minor complications were low between the 2 randomized treatment groups. Of the 17 minor vascular complications noted, 8 occurred in the StarClose device group and 9 minor complications occurred in the control. The combined total rates of minor complications at Day 30 were 4.3% for the randomized diagnostic StarClose device patients and 9.9% for the randomized diagnostic SC patients. The primary effectiveness endpoint for the interventional study was time to hemostasis. The secondary effectiveness endpoints were time to ambulation, procedure success at discharge, and device success. Time to hemostasis was defined as the elapsed time between sheath removal and first observed hemostasis. Time to ambulation was defined as the elapsed time between sheath removal and the time when the patient stands and walks at least 20 feet without re-bleeding. Procedure success was defined as the attainment of final hemostasis using any method and freedom from major vascular complications. Device success was defined as the attainment of final hemostasis using the StarClose Vascular Closure System alone or with adjunctive compression $\leq 5$ minutes and freedom from major vascular complications. The effectiveness results for the interventional patients in the clinical study are shown in Table 6, Table 7, and Table 8. P050007/S1 SSED Page 7 of 11 7 14 {7} Table 6: Primary Effectiveness Endpoint – Interventional ITT Patients | Time to Hemostasis (Mins) | CLIP Device (N=184) | Standard Compression (N=91) | All Patients (N=275) | Difference [95% C.I.] | P-value*** | | --- | --- | --- | --- | --- | --- | | Mean ± SD (N) | 7.95 ± 28.22 (182) | 29.06 ± 35.26 (74) | 14.05 ± 31.83 (256) | -14.01 [-16.21, -11.81] | <0.001 | | Median | 0.33 | 19.60 | 1.83 | | | | Range (min, max) | (0.0, 184.2) | (0.0, 245.3) | (0.0, 245.3) | | | Numbers are % (counts/sample size) or Mean ± 1 Standard Deviation. Treatment group comparisons were performed using Fisher's Exact test for categorical variables. For continuous variables, the Shapiro-Wilk test was used to test normality of Time to Hemostasis and Time to Ambulation. The p-value from this test was <0.001, indicating non-normality and skewness of the data distribution. Therefore the Wilcoxon Rank Sum Test was used to calculate the p-value between the two groups. P050007/S1 SSED Page 8 of 11 8 15 {8} Table 7: Secondary Effectiveness Endpoints – Interventional ITT Patients | Interventional | StarClose Device (N=184) | Standard Compression (N=91) | All Subjects (N=275) | Difference [95% C.I.] | P-value | | --- | --- | --- | --- | --- | --- | | Procedure Success* | 98.9% (181/183) | 98.7% (74/75) | 98.8% (255/258) | 0.2% [-2.8%, 3.2%] | 1.000 | | Device Success** | 86.8% (158/182) | N/A | N/A | N/A | N/A | | Time to Hemostasis (Mins)*** | | | | | | | Mean±SD (N) | 7.95±28.22 (182) | 29.06±35.26 (74) | 14.05±31.83 (256) | -21.11 [-29.37, -12.85] | <0.001 | | Range (min, max) | (0.0,184.2) | (0.0,245.3) | (0.0,245.3) | | | | Median | 0.33 | 19.60 | 1.83 | | | | Percentile (5%, 95%) | (0.03,27.92) | (13.47,79.67) | (0.03,66.42) | | | | Time to Ambulation (Mins)*** | | | | | | | Mean±SD (N) | 406.99±282.61 (178) | 466.02±257.23 (90) | 426.82±275.29 (268) | -59.03 [-128.90,10.85] | <0.001 | | Range (min, max) | (129.0,1686.0) | (41.0,1310.0) | (41.0,1686.0) | | | | Median | 278.50 | 389.00 | 305.00 | | | | Percentile (5%, 95%) | (228.00,1075.00) | (235.00,1023.00) | (234.00,1050.00) | | | table1.sas 02SEP05 Numbers are % (counts/sample size) or Mean ± 1 Standard Deviation. Treatment group comparisons were performed using Fisher's Exact test for categorical variables. For continuous variables, the Shapiro-Wilk test was used to test normality of Time to Hemostasis and Time to Ambulation. The p-value from this test was <0.001, indicating non-normality and skewness of the data distribution. Therefore the Wilcoxon Rank Sum Test was used to calculate the p-value between the two groups. * Procedure success was defined as the attainment of final hemostasis using any method and freedom from major vascular complications. StarClose subject 01-315 had time of hemostasis (T6) missing; therefore, Procedure Success was incalculable. Standard compression subjects 01-304, 01-307, 01-311, 01-313, 01-317, 01-321, 01-327, 01-332, 01-333, 01-340, 01-355, 03-305, 04-308, 06-317, 06-330, and 16-310 had the "seconds" field incomplete for time of hemostasis (T6), and therefore procedure success was not calculable. ** Device success was defined as the attainment of final hemostasis using the StarClose VCS alone or with adjunctive compression ≤5 minutes and freedom from major vascular complications. Device success could not be calculated for two (2) Starclose subjects; 01-315 had time of hemostasis (T6) missing and 15-307 had time procedural sheath removed (T2) missing. *** Time to hemostasis was defined as the elapsed time between sheath removal and first observed hemostasis. The time to hemostasis was calculated by subtracting CRF IVC 007, Q. 11.7 (time Introducer sheath removed) from IVC 007, Q.11.8 (time hemostasis first observed) for device subjects, or calculated by subtracting CRF IVC 008, Q. 12.1 (time procedural sheath removed) from IVC 008, Q.12.2 (time hemostasis first observed) for control subjects. StarClose subjects 02-117, 04-303, 04-305, 04-307, 10-305 and control subject 02-304 had time to hemostasis equal to '0'. These values have been queried and confirmed by the investigators. StarClose subjects with incalculable values for TTH were 15-307 and 01-315. Standard compression subjects with incalculable TTH values were: 01-304, 01-307, 01-311, 01-313, 01-317, 01-321, 01-327, 01-332, 01-333, 01-340, 01-355, 03-305, 04-308, 06-317, 06-320, 06-330, and 16-310. *** Time to ambulation was defined as the elapsed time between sheath removal and time when the subject stood and walked at least 20 feet without re-bleeding. The time to ambulation was calculated by subtracting IVC 007 Q.11.7 (time Introducer sheath removed) for device subjects or IVC 008 Q. 12.1 (time procedural sheath removed) for control subjects, from CRF IVC 011, Q.1.8 (time of first ambulation). StarClose subjects with incalculable time to ambulation values were: 01-315, 02-319, 02-335, 04-304, 06-308, and 15-307. The standard compression subject with an incalculable value for time to ambulation was 01-317. P050007/S1 SSED Page 9 of 11 9 16 {9} Table 8: Effectiveness Results by Post-Procedure Time Interval for Interventional ITT Patients Table 8: Effectiveness Results by Post Procedure Time Interval for Interventional ITT Patients | Percentage of Patients Achieving Hemostasis within Time Interval | | < 5 min | < 10 min | < 15 min | < 30 min | < 60 min | ≤ 120 min | > 120 min | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Star Close * | 83.2% (153/184) | 89.1% (164/184) | 90.8% (167/184) | 94.0% (173/184) | 94.6% (174/184) | 96.2% (177/184) | | | | | | Standard Compression ** | 2.7% (2/74) | 4.1% (3/74) | 10.8% (8/74) | 77.0% (57/74) | 90.5% (67/74) | 97.3% (72/74) | | | | | Percentage of Patients Ambulating Within Time Interval | | < 1 hr | < 2 hrs | < 3 hrs | < 4 hrs | < 6 hrs | < 12 hrs | < 18 hrs | ≤ 20 hrs | > 20 hrs | | | Star Close # | 0% (0/184) | 0% (0/184) | 2.2% (4/184) | 5.4% (10/184) | 66.3% (122/184) | 84.8% (156/184) | 92.4% (170/184) | 92.9% (171/184) | 96.7% (178/184) | | | Standard Compression ## | 2.2% (2/91) | 2.2% (2/91) | 2.2% (2/91) | 5.5% (5/91) | 38.5% (35/91) | 83.5% (76/91) | 96.7% (88/91) | 97.8% (89/91) | 98.9% (90/91) | *Table 1 on page 12 of the submission notes there is no data for subjects 01-315 and 15-307, resulting in only 182 StarClose subjects with actual times recorded. ** Note 3 of Table 21 on page 59 of the submission states that 17 subjects had no time to hemostasis recorded, leaving 74 Compression subjects (blue) with actual times recorded. Values in orange include all subjects. # Six (6) subjects had no time to ambulation noted due to the absence of one or more data points needed to calculate TTA. ## One (1) subject had no time to ambulation noted due to the absence of one or more data points needed to calculate TTA. ### B. Repuncture Through StarClose and Reclosure Please refer to the clinical section of the original SSED for repuncture information. ### XI. Conclusions Drawn from Studies Results of the biocompatibility testing, in vitro bench testing, animal studies, cadaver study and clinical investigations provide valid scientific evidence and reasonable assurance that the StarClose™ Vascular Closure System is safe and effective when used in accordance with its Instructions for Use. ### XII. Panel Recommendation In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Circulatory System Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by the panel. P050007/S1 SSED Page 10 of 11 10 17 {10} ### XIII. CDRH Decision FDA issued a PMA approval order to Abbott Vascular Devices on February 2, 2007. ### XIV. Approval Specifications A. Instructions for Use: See the labeling. B. Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events sections of the labeling. C. Post Approval Requirements and Restrictions: See approval order. P050007/S1 SSED Page 11 of 11 11 18
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...