Vitros Anti-HBc Reagent Pack For the in vitro qualitative detection of total antibody (IgG and IgM) to hepatitis B core antigen (total anti-HBc) in human adult and pediatric serum and plasma (EDTA and citrate) and neonate serum using the VITROS ECi Immunodiagnostic System. Assay results, in conjunction with other serological and clinical information, may be used for the laboratory diagnosis of individuals with acute or chronic hepatitis B, or recovery from hepatitis B infection. The presence of anti-HBc may be used as an aid in the determination of exposure to HBV infection for individuals prior to HBV vaccination. Vitros Anti-HBc Calibrator For use in the calibration of the Vitros ECi Immunodiagnostic System when used for the in vitro qualitative detection of total antibody (IgG and IgM) to hepatitis B core antigen (total anti-HBc) in human adult and pediatric serum and plasma (EDTA and citrate) and neonate serum using Vitros Anti-HBc Reagent Packs.
Device Story
Competitive immunoassay for total anti-HBc detection; utilizes Vitros ECi Immunodiagnostic System. Input: human serum/plasma samples. Process: sample reacts with recombinant HBcAg-coated wells; unbound material washed; HRP-labeled mouse monoclonal anti-HBc conjugate added; bound HRP measured via enhanced chemiluminescence (luminol derivative + peracid salt + electron transfer agent). Output: light signal inversely proportional to anti-HBc concentration. Used in clinical laboratories by technicians. Output interpreted by clinicians alongside other serological markers to diagnose HBV status or assess vaccination eligibility. Benefits: aids in identifying HBV infection, recovery, or exposure status.
Clinical Evidence
Prospective multi-center study (N=1691 US, N=315 India). Compared Vitros Anti-HBc to FDA-licensed reference assays. Population I (US): 92.38% positive percent agreement (PPA), 99.60% negative percent agreement (NPA). Population II (India): 98.56% PPA, 100% NPA. Additional testing included seroconversion panels, cord blood, and cross-reactivity with HAV, HCV, and other conditions. Results demonstrate performance comparable to reference methods.
Indicated for in vitro qualitative detection of total anti-HBc (IgG and IgM) in human adult, pediatric, and neonate serum and plasma (EDTA/citrate) to aid in diagnosis of acute/chronic HBV, recovery from HBV, or determination of prior HBV exposure before vaccination.
Regulatory Classification
Identification
A hepatitis B virus (HBV) antibody assay is identified as an in vitro diagnostic device intended for prescription use in the detection of antibodies to HBV in human serum, plasma, or other matrices, and as a device that aids in the diagnosis of HBV infection in persons with signs and symptoms of hepatitis and in persons at risk for hepatitis B infection. Results from assays may be qualitative or quantitative, such as quantitative anti-HBs. In addition, results from an anti-HBc IgM (IgM antibodies to core antigen) assay indicating the presence of anti-HBc IgM are indicative of recent HBV infection. Anti-HBs (antibodies to surface antigen) assay results may be used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or when vaccination status is unknown. The assay is not intended for screening of blood, plasma, cells, or tissue donors. The assay is intended as an aid in diagnosis in conjunction with clinical findings and other diagnostic procedures.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The labeling required under § 809.10(b) of this chapter must include:
(i) A prominent statement that the assay is not intended for the screening of blood, plasma, cells, or tissue donors.
(ii) A detailed explanation of the principles of operation and procedures for performing the assay.
(iii) A detailed explanation of the interpretation of results.
(iv) Limitations, which must be updated to reflect current clinical practice and disease presentation and management. The limitations must include statements that indicate:
(A) When appropriate, performance characteristics of the assay have not been established in populations of immunocompromised or immunosuppressed patients or other special populations where assay performance may be affected.
(B) Detection of HBV antibodies to a single viral antigen indicates a present or past infection with hepatitis B virus, but does not differentiate between acute, chronic, or resolved infection.
(C) The specimen types for which the device has been cleared, and that use of the assay with specimen types other than those specifically cleared for this device may result in inaccurate assay results.
(D) Diagnosis of hepatitis B infection should not be established on the basis of a single assay result but should be determined by a licensed healthcare professional in conjunction with the clinical presentation, history, and other diagnostic procedures.
(E) A non-reactive assay result may occur early during acute infection, prior to development of a host antibody response to infection, or when analyte levels are below the limit of detection of the assay.
(F) Results obtained with this assay may not be used interchangeably with results obtained with a different manufacturer's assay.
(v) For devices intended for the quantitative detection of HBV antibodies (anti-HBs), in addition to the special controls listed in paragraphs (b)(1) and (2) of this section, labeling required under § 809.10(b) of this chapter must include:
(A) The assay calibrators' traceability to a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard) and the limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), linearity, and precision to define the analytical measuring interval.(B) Performance results of the analytical sensitivity study testing a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard).(2) Design verification and validation must include the following:
(i) Detailed device description, including all parts that make up the device, ancillary reagents required but not provided, an explanation of the device methodology, and design of the antigen(s) and capture antibody(ies) sequences, rationale for the selected epitope(s), degree of amino acid sequence conservation of the target, and the design and composition of all primary, secondary and subsequent standards used for calibration.
(ii) Documentation and characterization (
*e.g.,* supplier, determination of identity, and stability) of all critical reagents (including description of the antigen(s) and capture antibody(ies)), and protocols for maintaining product integrity throughout its labeled shelf life.(iii) Risk analysis and management strategies, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on assay performance.
(iv) Final release criteria to be used for manufactured assay lots with appropriate evidence that lots released at the extremes of the specifications will meet the identified analytical and clinical performance characteristics as well as stability.
(v) Stability studies for reagents must include documentation of an assessment of real-time stability for multiple reagent lots using the indicated specimen types and must use acceptance criteria that ensure that analytical and clinical performance characteristics are met when stability is assigned based on the extremes of the acceptance range.
(vi) All stability protocols, including acceptance criteria.
(vii) When applicable, analytical sensitivity of the assay that is the same or better than that of other cleared or approved assays.
(viii) Analytical performance studies and results for determining the limit of blank (LoB), limit of detection (LoD), cutoff, precision (reproducibility), including lot-to-lot and/or instrument-to-instrument precision, interference, cross reactivity, carryover, hook effect, seroconversion panel testing, matrix equivalency, specimen stability, reagent stability, and cross-genotype antibody detection sensitivity, when appropriate.
(ix) For devices intended for the detection of antibodies for which a standardized reference material (that FDA has determined is appropriate) is available, the analytical sensitivity study and results testing the standardized reference material. Detailed documentation of that study and its results must be provided, including the study protocol, study report, testing results, and all statistical analyses.
(x) For devices with associated software or instrumentation, documentation must include a detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(xi) Detailed documentation of clinical performance testing from a clinical study with an appropriate number of HBV reactive and non-reactive samples in applicable risk categories and conducted in the appropriate settings by the intended users. Performance must be analyzed relative to an FDA cleared or approved HBV antibody assay or a comparator that FDA has determined is appropriate. Additional relevant patient groups must be validated as appropriate. The samples must include prospective (sequential) samples for each identified specimen type and, as appropriate, additional characterized clinical samples. Samples must be sourced from geographically diverse areas.
(3) For any HBV antibody assay intended for quantitative detection of anti-HBV antibodies, the following special controls, in addition to those special controls listed in paragraphs (b)(1) and (2) of this section, also apply:
(i) Detailed documentation of the metrological calibration traceability hierarchy to a standardized reference material that FDA has determined is appropriate.
(ii) Detailed documentation of the following analytical performance studies conducted, as appropriate to the technology, specimen types tested, and intended use of the device, including upper and lower limits of quantitation (UloQ and LloQ, respectively), linearity using clinical samples, and an accuracy study using the recognized international standard material.
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
I. General Information
Device Generic Name: Antibody to Hepatitis B virus core antigen (Anti-HBc) assay and Antibody to Hepatitis B virus core antigen assay calibrator (Anti-HBc assay calibrator)
Device Trade Name: Vitros Immunodiagnostic Products Anti-HBc Reagent Pack and Vitros Immunodiagnostic Products Anti-HBc Calibrator
Name and Address of Applicant: Ortho-Clinical Diagnostics, Inc, 100 Indigo Creek Drive, Rochester NY 14626-5101
Date of Panel Recommendation: None
PMA Number: P030024
Date of Notice of Approval to Applicant: March 4, 2004
II. Contraindications: None
III. Warnings and Precautions
The warnings and precautions can be found in the Vitros Immunodiagnostic Products Anti-HBc Reagent Pack and Vitros Immunodiagnostic Products Anti-HBc Calibrator labeling (Attachment 1).
The Vitros Anti-HBc Calibrator has been validated for use only on the Vitros System with Vitros Immunodiagnostic Products anti-HBc Reagent Packs. Refer to the Vitros Anti-HBc Reagent Pack instructions for use for further details.
IV. Indications for Use
Vitros Anti-HBc Reagent Pack
For the in vitro qualitative detection of total antibody (IgG and IgM) to hepatitis B core antigen (total anti-HBc) in human adult and pediatric serum and plasma (EDTA and citrate) and neonate serum using the VITROS ECi Immunodiagnostic System. Assay results, in conjunction with other serological and clinical information, may be used for the laboratory diagnosis of individuals with acute or chronic hepatitis B, or recovery from hepatitis B infection. The presence of anti-HBc may be used as an aid in the determination of exposure to HBV infection for individuals prior to HBV vaccination.
{1}
Page 2 of 23 – P030024 Summary of Safety and Effectiveness Data
## Vitros Anti-HBc Calibrator
For use in the calibration of the Vitros ECi Immunodiagnostic System when used for the *in vitro* qualitative detection of total antibody (IgG and IgM) to hepatitis B core antigen (total anti-HBc) in human adult and pediatric serum and plasma (EDTA and citrate) and neonate serum using Vitros Anti-HBc Reagent Packs.
## V. Device Description
### Principle of Device Methodology
The Vitros ECi Immunodiagnostic System (*Vitros* Analyzer) allows for the determination of analytes in human samples (for example, serum and plasma). All assays on the *Vitros* Analyzer employ an enhanced chemiluminescence detection reaction. The analyzer is fully automated with a refrigerated on board assay storage system. All standard bar code symbologies are supported by the analyzer, which has a throughput of up to 90 assays per hour. The analyzer also provides menu driven software, which can be accessed, from a high-resolution touchscreen monitor.
The *Vitros* Anti-HBc assay is a competitive immunoassay, which involves the reaction of antibodies to HBc present in the specimen with recombinant hepatitis B core antigen (HBcAg) coated on the wells. Unbound material is removed by washing. Horseradish peroxidase (HRP)-labeled antibody conjugate (mouse monoclonal anti-HBc) is then allowed to react with the remaining exposed recombinant HBcAg on the well surface. Unbound conjugate is removed by washing.
The bound HRP conjugate is measured by a luminescent reaction. A reagent containing luminogenic substrates (a luminol derivative and a peracid salt) and an electron transfer agent is added to the wells. The HRP in the bound conjugate catalyzes the oxidation of the luminol derivative, producing light. The electron transfer agent (a substituted acetanilide) increases the level of light produced and prolongs its emission. The light signals are read by the *Vitros* Analyzer. The amount of HRP conjugate bound is inversely proportional to the concentration of anti-HBc present.
The *Vitros* Anti-HBc Calibrator allows calibration of the *Vitros* Anti-HBc assay. Calibration is lot specific. A master calibration is established for each new reagent lot by performing multiple assays. The lot-specific parameter, the expected calibrator signal and the data that enables the *Vitros* System to calculate the cut-off value are encoded on the lot calibration card. Scanning the lot calibration card loads the encoded data into the *Vitros* System. When the calibrator is processed, the validity of the calibration is assessed against a quality parameter that compares the actual signal of the calibrator with the expected signal. If the calibration is acceptable the cut-off value is calculated and stored for use with any reagent pack of that lot. Recalibration is required after a predetermined calibration interval, or when a different reagent pack lot is loaded.
{2}
Page 3 of 23 – P030024 Summary of Safety and Effectiveness Data
## Kit Configuration and Components
For detection of anti-HBc, the *Vitros* System is comprised of the following:
- *Vitros* Immunodiagnostic Products Anti-HBc Reagent Pack (*Vitros* Anti-HBc Reagent Pack) and *Vitros* Immunodiagnostic Products Anti-HBc Calibrator (*Vitros* Anti-HBc Calibrator) together comprise the *Vitros* Anti-HBc assay.
The *Vitros* Anti-HBc Reagent Pack is composed of 3 reagents:
- Conjugate reagent [HRP-mouse monoclonal anti-HBc in buffer with mouse serum, human plasma and anti-microbial agent (Kathon)]
- Assay reagent [buffer with newborn calf serum, bovine gamma globulins and anti-microbial agent (Kathon)]
- Coated microwells [recombinant HBcAg derived from E. Coli]
The *Vitros* Anti-HBc Calibrator contains:
- Anti-HBc negative human plasma with anti-microbial agent (Kathon). The Calibrator is supplied ready for use.
In addition, the following components are required:
- *Vitros* ECi Immunodiagnostic System (*Vitros* Analyzer) is dedicated instrumentation, cleared by the FDA as an immunodiagnostic analyzer (K962919/S1), which provides automated analysis of the *Vitros* assays.
- *Vitros* Immunodiagnostic Products Signal Reagent and *Vitros* Immunodiagnostic Products Universal Wash Reagent are Common Reagents used in all *Vitros* System assays.
## VI. Alternate Practices and Procedures
Determining the presence of anti-HBc in patients may be achieved by using a variety of commercially available, FDA licensed serological tests. Additionally, when test results are used in conjunction with a physician’s assessment and other laboratory test results, infection with HBV can be identified.
3
{3}
Page 4 of 23 – P030024 Summary of Safety and Effectiveness Data
## VII. Marketing History
Below is a table describing the countries where the Vitros Anti-HBc Reagent Pack and Calibrator are currently available through February 2003.
| Argentina | Australia |
| --- | --- |
| Austria | Belgium |
| Brazil | Canada |
| Chile | Colombia |
| Czech Republic | Denmark |
| France | Germany |
| Greece | Holland |
| India | Indonesia |
| Italy | Japan |
| Korea | Luxembourg |
| Malaysia | Norway |
| Panama | Philippines |
| Poland | Romania |
| Portugal | Russia |
| Saudi Arabia | Singapore |
| Slovak Republic | Slovenia |
| Spain | Switzerland |
| Taiwan | Thailand |
| Turkey | United Kingdom |
| Venezuela | |
The Vitros Immunodiagnostic Products Anti-HBc Reagent Pack and Vitros Immunodiagnostic Products Anti-HBc Calibrator have not been withdrawn from marketing for any reason relating to the safety and effectiveness of the device.
## VIII. Potential Adverse Effects of the Device on Health
Since the Vitros Immunodiagnostic Products Anti-HBc Reagent Pack and Vitros Immunodiagnostic Products Anti-HBc Calibrator are for in vitro diagnostic use, there is no direct adverse effect on the health of the patient.
However, failure of the product to perform as indicated, or human error in use of the product may lead to a false result.
A false reactive result may be considered a patient or public health concern because false reactive results in the diagnostic setting may lead to diagnostic confusion and inappropriate therapy. This is due to the fact that both anti-HBs and anti-HBc tests may be used to determine past exposure to HBV. As appearance of anti-HBs may be delayed after HBsAg clearance, anti-HBc is sometimes the only serological marker for HBV infection. If a false reactive result is obtained for an anti-HBc test, there is a possibility that the patient would be considered to be previously exposed and therefore immune to HBV.
4
{4}
Page 5 of 23 – P030024 Summary of Safety and Effectiveness Data
A false negative result in a diagnostic setting may lead to a patient with HBV going unidentified. Under these circumstances, there is a safety concern for both the patient and the public, since such individuals may be capable of transmitting HBV infection. A false negative result could also lead to an appropriate treatment in a patient with acute hepatitis.
## IX. Summary of Non-Clinical Studies
### Instrumentation
Software and hardware verification testing was performed for the Vitros ECi Immunodiagnostic System (Vitros Analyzer). Appropriate information and study results were furnished demonstrating that the Vitros Analyzer hardware and software, used with the Vitros Immunodiagnostic Products Anti-HBc Reagent Pack and Calibrator functioned as described and had appropriate safeguards in place.
### Analytical Sensitivity
The concentration at the cutoff of the Vitros Anti-HBc assay was confirmed by using two kit lots to assay a commercial anti-HBc (total) sensitivity serum panel that had been calibrated against serial dilutions of the Paul Ehrlich Institute anti-HBc standard (100 U/mL Anti-HBc-IgG #82). A graph of the mean Vitros Anti-HBc assay result versus the factored concentration of each member of the sensitivity panel was used to determine the concentration at the cutoff.
The estimated detectable concentration of anti-HBc at the cut-off (Result = 1.00 signal/cutoff (s/c)) of the Vitros Anti-HBc assay, estimated from the graph was 1.00 PEI Units/mL. The calculated concentration was 0.97 PEI Units/mL.
### Comparison of Fresh Serum/Plasma Samples
To determine the acceptability of using the Vitros Anti-HBc assay for testing serum or plasma specimens, fifty fresh blood samples (25 unspiked and 25 spiked with anti-HBc to give a target result of $0.6 \pm 0.4 \mathrm{~s} / \mathrm{c}$), were collected and aliquoted into a variety of serum and plasma collection tubes. Anticoagulants $\mathbf{K}_2$ EDTA and sodium citrate were evaluated. Testing with the Vitros Anti-HBc assay was conducted on the same day blood was drawn.
For matched negative samples, EDTA plasma compared with serum showed $-0.6\%$ mean difference $(n=25)$ in anti-HBc s/c ratio and citrate plasma compared with serum showed $-1.8\%$ mean difference $(n=25)$ in anti-HBc s/c ratio. The mean s/c ratio results for unspiked samples were equivalent by the Bonferoni test of means. All unspiked samples, with serum and plasma (EDTA and citrate) preparations, were classified correctly as negative in the Vitros Anti-HBc assay. The within sample precision estimates were equivalent for unspiked serum and EDTA and citrated plasma.
{5}
Page 6 of 23 – P030024 Summary of Safety and Effectiveness Data
The s/c ratios for the anti-HBc spiked samples correlated very strongly with the s/c ratios of serum. EDTA and citrate plasma had equivalent minimum and maximum s/c ratios. Although the serum samples had higher estimates of within sample precision than citrate plasma, differences in precision were mostly (92.75%) accounted for by differences in mean s/c ratio results.
The above results indicate that serum and plasma (EDTA and citrate) should be suitable for use in the Vitros Anti-HBc assay.
## Comparison of Stability of Serum/Plasma Samples
Twenty fresh blood samples (10 unspiked and 10 spiked with anti-HBc to give a target result of 0.6 ± 0.4 s/c), were collected and aliquoted into a variety of serum and plasma collection tubes. Anticoagulants K₂ EDTA and sodium citrate were evaluated. Testing with the Vitros Anti-HBc assay was conducted on the same day blood was drawn, and again after 5 and 7 days storage at 2 – 8 °C (36 ° - 46 °F) and after 28 days at -20 °C (-4 °F).
None of the storage conditions tested had clinically significant effects on reactive samples. All anti-HBc spiked samples were qualitatively classified correctly as reactive, regardless of the storage condition. When analyzed statistically, there were five statistically significant positive slopes over time for anti-HBc spiked samples stored at 2 – 8 °C (36 - 46 °F). However, these statistically significant differences over time did not change the qualitative classification of the specimens; therefore they are not clinically significant.
These data show that storage of serum or plasma (EDTA and citrate) samples for up to 5 days at 2 – 8 °C (36 - 46 °F), or 28 days at -20 °C (-4 °F) should not have a significant effect on the test results with the Vitros Anti-HBc assay.
## Potentially Cross-reacting Subgroups
The specificity of the Vitros Anti-HBc assay was evaluated by testing 232 samples from 16 potentially cross-reacting sub-groups. Patient samples from the following sub-groups were tested: HAV, HEV, HCV, non-viral liver disease, autoimmune disease (rheumatoid arthritis and systemic lupus erythrematosis), CMV, EBV, HSV, parvovirus B19 infection, rubella, syphilis, toxoplasmosis, HIV 1/2 antibody positive, HTLV 1/2 antibody positive, and HBV vaccine recipients.
Of the 232 samples tested, 230 were observed to be negative. One autoimmune disease (rheumatoid arthritis) sample was initially reactive in the Vitros Anti-HBc assay, but was negative on repeat determination. One syphilis sample was reactive initially in the Vitros Anti-HBc assay and also reactive on repeat determination.
{6}
Page 7 of 23 – P030024 Summary of Safety and Effectiveness Data
The specificity of the *Vitros* Anti-HBc assay was evaluated further by testing anti-HBc spiked and unspiked samples with an additional spike of *Staphylococcus aureus*, *Escherichia coli* or *Pseudomonas aeruginosa*.
Of the samples that were tested none of the anti-HBc unspiked (negative) samples were found to be false reactive and none of the anti-HBc spiked samples were observed to be false negative in the *Vitros* Anti-HBc assay.
## Interfering Substances
The potentially interfering effects of hemoglobin, bilirubin and triolein were evaluated using samples from 10 blood donors. The results (test results at each level of interferent) demonstrate that hemoglobin (up to 500 mg/dL), bilirubin (up to 20 mg/dL) and triolein (up to 3000 mg/dL), should have no effect on result classification.
Samples spiked with anti-HBc to give a target result of 0.7 – 0.9 s/c were observed to remain reactive at all levels tested with each potential interferent. Similarly no interference was observed in samples not spiked with anti-HBc (negative).
## Stability
*Vitros* Anti-HBc Reagent Packs, Calibrator and Controls that were subjected to a period of simulated transport to mimic effects of shipment were tested at various time points up to 26 weeks after storage at 2 – 8 °C (36 – 46 °F). All results obtained were within acceptability limits, and overall no trends were evident.
In addition, a commercially obtained performance panel was tested using transported, stored materials at week 0 and week 26. Materials stored for 26 weeks yielded results that indicated no change in the qualitative classification of the samples from the classifications obtained at the initial time point.
These data supports the storage of the *Vitros* Anti-HBc Reagent Pack, Calibrator and Controls for 26 weeks at 2 – 8 °C (36 – 46 °F).
## Open On-Board Storage for the *Vitros* Anti-HBc Reagent Pack
*Vitros* Anti-HBc Reagent Packs that were subjected to a period of simulated transport to mimic effects of shipment were opened and placed in an environmental chamber (4 – 8 °C, ≤ 40% humidity) for a period of 8 weeks to simulate the storage on board the *Vitros* Analyzer. These Reagent Packs were tested at various time points within the 8 week time period. In addition, a single transported, opened Reagent Pack from each kit lot was removed from the chamber on 6 different occasions, and brought to room temperature over the 8 week period to simulate typical customer usage. Results of testing were within acceptability limits and overall no trend was observed between Reagent Packs stored at 2 – 8 °C (36 – 46 °F) and freshly opened, and Reagent Packs stored opened on board for 8 weeks. These data supports the on board storage of Reagent Packs for up to 8 weeks.
{7}
Page 8 of 23 – P030024 Summary of Safety and Effectiveness Data
## Open Off-Board Storage for Vitros Anti-HBc Calibrators
*Vitros* Anti-HBc Calibrators that were subjected to a period of simulated transport to mimic effects of shipment were opened, pooled, sub-aliquoted and stored at 2 - 8 °C (36 – 46 °F) and -20 °C (-4 °F) for 13 weeks. Results of testing these calibrators at various time points up to 13 weeks indicated no observable trends and met all acceptance criteria.
The data supports the storage of the calibrators at 2 – 8 °C (36 – 46 °F) and -20°C (-4°F) after opening for up to 13 weeks (with no more than 1 freeze-thaw cycle).
## Vitros Universal Wash Reagent Study
*Vitros* Anti-HBc Reagent Packs, calibrators, and controls that were subjected to a period of simulated transport to mimic the effects of shipment were tested with 3 lots of *Vitros* Universal Wash Reagent at weeks 0 and 26 to determine the effect of aged *Vitros* Universal Wash Reagent.
The data show that the performance of the *Vitros* Anti-HBc assay is acceptable when used with *Vitros* Universal Wash Reagent that is either fresh or up to 26 weeks old.
## Vitros Signal Reagent Study
*Vitros* Anti-HBc Reagent Packs, calibrators, and controls that were subjected to a period of simulated transport to mimic the effects of shipment were tested with 3 lots of *Vitros* Signal Reagent to determine the effect of aged *Vitros* Signal Reagent.
The data show that the performance of the *Vitros* Anti-HBc assay is acceptable when used with *Vitros* Signal Reagent that is either fresh or 6 months old.
## Temperature Stressing Study (-20 °C / -4 °F)
*Vitros* Anti-HBc assay Reagent Packs and calibrators were subjected to 2 freeze/thaw cycles and the performance compared with Reagent Packs and calibrators stored at 2 - 8 °C (36 – 46 °F). Although freezing and thawing the *Vitros* Anti-HBc assay had no adverse effect on calibration quality parameters or control results, as a precaution it is recommended that the *Vitros* Anti-HBc assay is not frozen.
## Temperature Stressing Study (30 °C and 37 °C / 86 °F and 99 °F)
*Vitros* Anti-HBc Reagent Packs and calibrators were subjected to 5 days at 30 °C (86 °F) or 1 day at 37 °C (99 °F) and the performance compared with Reagent Packs and calibrators stored at 2 – 8 °C (36 – 46 °F).
{8}
Page 9 of 23 – P030024 Summary of Safety and Effectiveness Data
Exposing Vitros Anti-HBc Reagent Packs and calibrators to a temperature of 30 °C (86 °F) for 5 days or 37 °C (99 °F) for 1 day had no adverse effect on calibration quality parameters or control results. Exposure of the Reagent Packs or calibrators up to these temperatures for the times stated should not significantly compromise the performance of the Vitros Anti-HBc assay.
## Microbiology
Vitros Anti-HBc reagents are formulated with anti-microbial agents (Bronidox and Kathon) that provide protection against adventitious contamination by microorganisms. Evaluation of the microbial load of each reagent (Assay Reagent, Conjugate Reagent and Calibrator) post-dispensing and at 26 weeks demonstrated that the total aerobic count is generally on the order of ≤10 CFU/mL. In addition, the levels of preservative in each reagent were determined over a period of at least 26 weeks. Results for the Assay Reagent, Conjugate Reagent and Calibrator demonstrated that the preservative concentrations were above the minimum inhibitory concentration throughout 26 weeks of testing.
A study conducted according to US Pharmacopoeia (USP) 23/NF 18, general chapter 51, assessed the ability of the reagents to withstand or control microbial contamination. Results indicate that the preservative systems for Assay Reagent and Calibrator met the requirements of the USP 23 at 0 and 26 weeks. The Conjugate Reagent met the USP 23 requirement at week 0 for all challenge microorganisms and at week 26 for all but carnobacterium divergens.
{9}
Page 10 of 23 – P030024 Summary of Safety and Effectiveness Data
## Precision and Reproducibility
Precision was evaluated on a different *Vitros* ECi Immunodiagnostic System at three external sites, using one reagent pack and calibrator kit lot. At least two replicates each of a three member panel were assayed on a single occasion per day on 20 different days. The data shown in the table below were rounded following all calculations.
| Clinical Site | Mean Vitros aHBc S/C (Ratio) | Within Day * | | Between Day † | | Total ‡ | | No. Obs. | No. Days |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | CV (%) | SD | CV (%) | SD | CV (%) | | |
| Site 1 | 3.52 | 0.091 | 2.6 | 0.093 | 2.6 | 0.130 | 3.7 | 40 | 20 |
| | 0.34 | 0.026 | 7.7 | 0.019 | 5.6 | 0.032 | 9.5 | 40 | 20 |
| | 0.95 | 0.030 | 3.2 | 0.060 | 6.4 | 0.067 | 7.1 | 40 | 20 |
| | | | | | | | | | |
| Site 2 | 3.50 | 0.064 | 1.8 | 0.039 | 1.1 | 0.075 | 2.1 | 40 | 20 |
| | 0.40 | 0.037 | 9.2 | 0.056 | 14.1 | 0.067 | 16.9 | 40 | 20 |
| | 1.10 | 0.082 | 7.4 | 0.062 | 5.6 | 0.103 | 9.3 | 40 | 20 |
| | | | | | | | | | |
| Site 3 | 3.43 | 0.071 | 2.1 | 0.089 | 2.6 | 0.113 | 3.3 | 40 | 20 |
| | 0.31 | 0.014 | 4.3 | 0.023 | 7.4 | 0.027 | 8.6 | 40 | 20 |
| | 0.90 | 0.027 | 3.0 | 0.025 | 2.8 | 0.037 | 4.1 | 40 | 20 |
* Within Day: Variability of the assay performance from replicate to replicate.
† Between Days: Variability of the assay performance from day to day.
‡ Total: Variability of the assay performance combining the effects of within day and between days.
Reproducibility was evaluated incorporating between site and between lot variations. The study was performed at three external sites using three reagent lots. At least three replicates each of a four member panel were assayed on a single occasion per day on six different days. The between site, between lot, and total precision estimates (CV (%)) were derived from a variance component analysis. The data shown in the table below were rounded following all calculations.
/0
{10}
Page 11 of 23 – P030024 Summary of Safety and Effectiveness Data
| Mean Vitros Anti-HBc S/C (Ratio) | Between Site * | | Between Lot † | | Total ‡ | | No. Obs. |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | CV (%) | SD | CV (%) | SD | CV (%) | |
| 3.21 | 0.000 | 0.0 | 0.071 | 2.2 | 0.159 | 5.0 | 162 |
| 1.19 | 0.121 | 10.2 | 0.047 | 4.0 | 0.155 | 13.0 | 162 |
| 1.10 | 0.121 | 11.0 | 0.059 | 5.4 | 0.157 | 14.3 | 162 |
| 0.25 | 0.090 | 35.9 | 0.017 | 7.0 | 0.102 | 40.7 | 162 |
* Between site: Variability of the assay performance from site to site.
† Between lot: Variability of the assay performance from lot to lot, calculated using data across all sites.
‡ Total: Variability of the assay incorporating factors of site, lot and day.
## Calibration Interval
The performance of the Vitros Anti-HBc assay within and beyond one instrument calibration interval (28 days) was evaluated at three sites by testing a three member sera panel with one kit lot. One panel member was close to the Vitros Anti-HBc assay cut-off. Additional testing was performed on Days 29 and 30 of the calibration cycle to show that the Analyzer would still yield valid results beyond the end of a 28-day cycle. Two replicates of each panel member were assayed per day at each clinical site. Appropriate instrument calibration was performed and verified on Day 0 of the study, and the testing was performed for a total of 20 study days over a 28-day period.
Least squares regression analyses were performed within site and across sites. For analyses within site, although the slopes were statistically significant for two panel members in at least one site, the changes in s/c ratios over the entire testing period were either not clinically relevant, i.e., did not change the qualitative result interpretation, or were too small to have any clinical implications. For analyses across sites, the mean slope was not statistically significant for any of the panel members.
The Vitros Anti-HBc assay demonstrated adequate performance throughout the entire calibration interval, and continued to perform successfully two days beyond the expiration of calibration, as per the study design.
## Seroconversion Panels
Six commercially available HBV seroconversion panels were tested. The Vitros and reference anti-HBc assays’ results are summarized below. The table lists the first bleed of each panel that tested reactive with the Vitros and the reference assays as well as the difference between the two assays in identifying the first reactive panel member by number of days.
{11}
Page 12 of 23 – P030024 Summary of Safety and Effectiveness Data
| Days to Reactive Anti-HBc Result | | | | | |
| --- | --- | --- | --- | --- | --- |
| Panel ID | Reference Anti-HBc Assay | | Vitros Anti-HBc Assay | | Difference in Days to Anti-HBc Reactive Result |
| | - * | + ** | - * | + ** | Reference - Vitros |
| 6278 | 26 | 33 | 33 | 37 | -4 |
| 6281 | 36 | 41 | 36 | 41 | 0 |
| PHM935A | 50 | 66 | 50 | 66 | 0 |
| RP009 | 13 | 29 | 13 | 29 | 0 |
| RP016 | 24 | 56 | 24 | 56 | 0 |
| RP017 | 43 | 65 | 43 | 65 | 0 |
* Post bleed day of last nonreactive result, usually denotes previous bleed from first reactive.
** Post bleed day of first reactive result.
Based on the results of the testing, the Vitros Anti-HBc assay demonstrated agreement with the time of anti-HBc comparable to the reference anti-HBc assay.
## Cord Blood Testing
A total of 20 cord blood samples were tested in the Vitros Anti-HBc assay.
In testing the cord blood samples, 1 out of 20 samples was found to give a repeat reactive result in the Vitros Anti-HBc assay. This repeat reactive sample was also repeat reactive in the reference method.
Duplicate aliquots from 10 cord blood samples were spiked with either anti-HBc positive or negative plasma. Recovery of anti-HBc from serum and cord blood was assessed by calculating the percent shift of results [(negative spike-reactive spike)/negative spike]. The mean percent shift for serum was 75.8%. The mean percent shift for cord blood was 76.2%. It appears that anti-HBc is recovered from cord blood to the same extent as in serum.
The above information shows that cord blood should be an acceptable sample type for use with the Vitros Anti-HBc assay. Due to the low sample numbers and a single matrix being tested OCD has stated in the indications for use statement that serum may only be used and placed a warning in the Interpretations of Results stating that low level results should not be repeated. The warning implies that another specimen should be obtained and tested.
12
{12}
Page 13 of 23 – P030024 Summary of Safety and Effectiveness Data
## X. Summary of Clinical Studies
A multi-center prospective study was conducted to evaluate the clinical performance of the Vitros Anti-HBc assay among individuals with signs or symptoms or biochemical manifestations (elevated liver function tests) of hepatitis and those at high risk of hepatitis infection due to lifestyle, behavior, occupation, or known exposure events. Specimens were obtained from 1691 subjects prospectively enrolled at three geographically separated collection sites within the United States (Population I) located in Miami, FL (37.0%), Dallas, TX (28.1%) and Chicago, IL (34.9%). Specimens were also obtained from 315 subjects prospectively enrolled in an area in India with a high prevalence of viral hepatitis (Population II). Statistical testing performed to evaluate the homogeneity of the distribution of Vitros Anti-HBc s/c values across the four collection sites indicated that the data from Population I and Population II could not be pooled for statistical analysis.
The HBV disease classification for each subject was determined by a single point serological assessment using a hepatitis marker profile consisting of reference assays (previously licensed or approved by the FDA) for the detection of HBsAg, HBeAg, anti-HBc, anti-HBc IgM, anti-HBe, and anti-HBs (quantitative). The reference assays' procedures were adhered to during the clinical laboratory study.
The subjects in Population I were Caucasian (24.9%), African American (44.1%), Hispanic (22.4%) and Asian (3.7%), with the remaining 4.9% represented by other ethnic groups. The group was 52.4% male and 47.6% female, and ranged in age from 5 to 89 years. Testing of these specimens with the Vitros Anti-HBc assay occurred at diagnostic laboratories located in Miami, FL (37.0%), Port Jefferson, NY (34.9%) and Minneapolis MN, (28.1%). Agreement of the Vitros Anti-HBc assay was assessed relative to the reference anti-HBc assay and HBV disease classification using serum samples from the 1691 subjects in Population I.
The subjects in Population II were Asian Indian (100%). The group was 73.0% male and 27.0% female, and ranged in age from 18 to 90 years. Testing of these specimens with the Vitros Anti-HBc assay occurred at diagnostic laboratories located in Miami, FL (33.0%), Minneapolis MN, (32.4%) and Los Angeles, CA (34.6%). Agreement of the Vitros Anti-HBc assay was assessed relative to a reference anti-HBc assay and HBV disease classification using serum samples from the 315 subjects in Population II.
## Results by Specimen Classification
The data were analyzed following the assignment of HBV disease classifications based upon the positive (+) / negative (-) patterns for the six HBV serological reference markers. The table below summarizes how these classifications were derived. There were 28 unique reference marker profiles observed among the subjects in Populations I and II (24 unique patterns in Population I and 18 unique patterns in Population II) during the Vitros Anti-HBc clinical study.
{13}
Page 14 of 23 – P030024 Summary of Safety and Effectiveness Data
# HBV Reference Marker Profiles and HBV Disease Classification
| Reference HBsAg^{1,2} | Reference HBeAg | Reference IgM aHBc | Reference Total aHBc | Reference aHBe | Reference aHBs ≥10 mIU/mL | HBV Disease Classification |
| --- | --- | --- | --- | --- | --- | --- |
| + | + | + | + | + | – | Acute |
| + | + | + | + | – | – | Acute |
| + | – | + | + | + | + | Acute |
| + | – | + | + | + | – | Acute |
| + | – | + | + | – | – | Acute |
| + | – | – | – | – | – | Acute |
| + | + | – | + | + | – | Chronic |
| + | + | – | + | – | + | Chronic |
| + | + | – | + | – | – | Chronic |
| + | – | – | + | + | + | Chronic |
| + | – | – | + | + | – | Chronic |
| – | – | + | + | + | + | Early Recovery |
| – | – | + | + | + | – | Early Recovery |
| – | – | + | + | – | + | Early Recovery |
| – | – | + | + | – | – | Early Recovery |
| – | – | – | + | + | – | Early Recovery |
| – | – | – | + | + | + | Recovery |
| – | – | – | + | – | + | Recovered |
| – | – | – | + | – | – | Recovered |
| – | – | – | – | – | + | HBV Vaccine Response |
| – | – | – | – | – | – | Not Previously Infected with HBV |
| + | + | – | – | + | + | Uninterpretable |
| + | – | – | – | – | + | Uninterpretable |
| – | + | – | + | – | – | Uninterpretable |
| – | + | – | – | – | + | Uninterpretable |
| – | + | – | – | – | – | Uninterpretable |
1. Positive = Reference HBsAg assay reactive and confirmed by neutralization.
2. Negative = Reference HBsAg assay negative or not confirmed by neutralization.
14
{14}
Page 15 of 23 – P030024 Summary of Safety and Effectiveness Data
## Comparison of Results
The table below compares the *Vitros Anti-HBc* results with the reference anti-HBc results by specimen classification for the subjects in Population I.
### Comparison of *Vitros Anti-HBc* Results with Reference Anti-HBc Results by HBV Disease Classification - Population I (N=1691)
| | Reference Anti-HBc Result | | | | |
| --- | --- | --- | --- | --- | --- |
| | Reactive | | Negative | | |
| HBV Disease Classification | Vitros Anti-HBc Result | | Vitros Anti-HBc Result | | |
| | Reactive | Negative * | Reactive | Negative | Total |
| Acute | 8 | 0 | 0 | 9 | 17 |
| Chronic | 40 | 3 | 0 | 0 | 43 |
| Early Recovery | 46 | 1 | 0 | 0 | 47 |
| Recovery | 138 | 0 | 0 | 0 | 138 |
| Recovered | 168 | 28 | 0 | 0 | 196 |
| HBV Vaccine Response | 0 | 0 | 0 | 169 | 169 |
| Not Previously Infected with HBV | 0 | 0 | 5 ** | 1069 | 1074 |
| Uninterpretable | 0 | 1 | 0 | 6 | 7 |
| Overall | 400 | 33 | 5 | 1253 | 1691 |
* These samples were tested with a second FDA approved anti-HBc assay with the following results:
Chronic: 2/3 negative,
Early recovery: 1/1 indeterminate,
Recovered: 18/28 negative; 1/28 indeterminate,
Uninterpretable: 1/1 negative,
Overall: 23/33 (69.7%) negative or indeterminate.
** These samples were tested with a second FDA approved anti-HBc assay with the following results:
2/5 (40%) positive.
The table below compares the *Vitros Anti-HBc* results with the reference anti-HBc results by specimen classification for the subjects in Population II.
15
{15}
Page 16 of 23 – P030024 Summary of Safety and Effectiveness Data
## Comparison of Vitros Anti-HBc Results With Reference Anti-HBc Results by HBV Disease Classification – Population II (N=315)
| | Reference Anti-HBc Result | | | | |
| --- | --- | --- | --- | --- | --- |
| | Reactive | | Negative | | |
| | Vitros Anti-HBc Result | | Vitros Anti-HBc Result | | |
| HBV Disease Classification | Reactive | Negative * | Reactive | Negative | Total |
| Acute | 86 | 2 | 0 | 16 | 104 |
| Chronic | 184 | 1 | 0 | 0 | 185 |
| Early Recovery | 1 | 0 | 0 | 0 | 1 |
| Recovery | 0 | 0 | 0 | 0 | 0 |
| Recovered | 2 | 1 | 0 | 0 | 3 |
| HBV Vaccine Response | 0 | 0 | 0 | 3 | 3 |
| Not Previously Infected with HBV | 0 | 0 | 0 | 17 | 17 |
| Uninterpretable | 0 | 0 | 0 | 2 | 2 |
| Overall | 273 | 4 | 0 | 38 | 315 |
* Zero of three samples were negative with a second FDA approved anti-HBc assay. One sample (Chronic) was not tested (insufficient volume.)
## Percent Agreement
Positive and negative percent agreement between the Vitros Anti-HBc assay and the reference anti-HBc assay were calculated for subjects in Population I (N=1691) with various HBV disease classifications, and for the overall study population. The table below summarizes these calculations and provides the upper and lower 95% exact confidence intervals.
{16}
Page 17 of 23 – P030024 Summary of Safety and Effectiveness Data
# Positive and Negative Percent Agreement between the Vitros Anti-HBc and Reference Anti-HBc Assays in Population I
| HBV Disease Classification | Positive Percent Agreement (%) | 95% Exact Confidence Interval | Negative Percent Agreement (%) | 95% Exact Confidence Interval |
| --- | --- | --- | --- | --- |
| Overall | 400/433 (92.38%) | 89.46, 94.70 | 1253/1258 (99.60%) | 99.07, 99.87 |
| Acute | 8/8 (100%) | 63.06, 100 | 9/9 (100%) | 66.37, 100 |
| Chronic | 40/43 (93.02%) | 80.94, 98.54 | 0/0 (N/A) | N/A |
| Early Recovery | 46/47 (97.87%) | 88.71, 99.95 | 0/0 (N/A) | N/A |
| Recovery | 138/138 (100%) | 97.36, 100 | 0/0 (N/A) | N/A |
| Recovered | 168/196 (85.71%) | 80.02, 90.29 | 0/0 (N/A) | N/A |
| HBV Vaccine Response | 0/0 (N/A) | N/A | 169/169 (100%) | 97.84, 100 |
| Not Previously Infected with HBV | 0/0 (N/A) | N/A | 1069/1074 (99.53%) | 98.92, 99.85 |
| Uninterpretable | 0/1 (0%) | N/A | 6/6 (100%) | 54.07, 100 |
The positive percent agreement with the reference anti-HBc assay was determined by dividing the number of reactive Vitros Anti-HBc results by the total number of subjects reactive with the reference anti-HBc assay. As a result of this study, the overall positive percent agreement of the Vitros Anti-HBc assay with the reference anti-HBc assay in Population I was estimated to be 92.38% (400/433, with a 95% exact confidence interval of 89.46, 94.70).
The negative percent agreement with the reference anti-HBc assay was determined by dividing the number of negative Vitros Anti-HBc results by the total number of subjects negative with the reference anti-HBc assay. As a result of this study, the overall negative percent agreement of the Vitros Anti-HBc assay with the reference anti-HBc assay in Population I was estimated to be 99.60% (1253/1258, with a 95% exact confidence interval of 99.07, 99.87).
Positive and negative percent agreement between the Vitros Anti-HBc assay and the reference anti-HBc assay were also calculated for subjects in Population II (N=315) with various HBV disease classifications, and for the overall study population. The table below summarizes these calculations and provides the upper and lower 95% exact confidence intervals.
{17}
Page 18 of 23 – P030024 Summary of Safety and Effectiveness Data
## Positive and Negative Percent Agreement between the Vitros Anti-HBc and Reference Anti-HBc Assay in Population II
| HBV Disease Classification | Positive Percent Agreement (%) | 95% Exact Confidence Interval | Negative Percent Agreement (%) | 95% Exact Confidence Interval |
| --- | --- | --- | --- | --- |
| Overall | 273/277 (98.56%) | 96.34, 99.61 | 38/38 (100%) | 90.75, 100 |
| Acute | 86/88 (97.73%) | 92.03, 99.72 | 16/16 (100%) | 79.41, 100 |
| Chronic | 184/185 (99.46%) | 97.03, 99.99 | 0/0 (N/A) | N/A |
| Early Recovery | 1/1 (100%) | 2.5, 100 | 0/0 (N/A) | N/A |
| Recovered | 2/3 (66.67%) | 9.43, 99.16 | 0/0 (N/A) | N/A |
| HBV Vaccine Response | 0/0 (N/A) | N/A | 3/3 (100%) | 29.24, 100 |
| Not Previously Infected with HBV | 0/0 (N/A) | N/A | 17/17 (100%) | 80.49, 100 |
| Uninterpretable | 0/0 (N/A) | N/A | 2/2 (100%) | 15.81, 100 |
The positive percent agreement with the reference anti-HBc assay was determined by dividing the number of reactive Vitros Anti-HBc results by the total number of subjects reactive with the reference anti-HBc assay. As a result of this study, the overall positive percent agreement of the Vitros Anti-HBc assay with the reference anti-HBc assay in Population II was estimated to be 98.56% (273/277, with a 95% exact confidence interval of 96.34, 99.61).
The negative percent agreement with the reference anti-HBc assay was determined by dividing the number of negative Vitros Anti-HBc results by the total number of subjects negative with the reference anti-HBc assay. As a result of this study, the overall negative percent agreement of the Vitros Anti-HBc assay with the reference anti-HBc assay in Population II was estimated to be 100% (38/38, with a 95% exact confidence interval of 90.75, 100).
## Percent Agreement of the Vitros Anti-HBc Assay With Clinical Status for Subjects With Clinically Diagnosed Acute or Chronic HBV Infection
The performance of the Vitros Anti-HBc assay was further evaluated among archived serum samples from subjects with clinical and laboratory documentation of acute or chronic (HBsAg present for ≥6 months) HBV infection. The table below summarizes the performance of the Vitros Anti-HBc assay in samples from subjects with documented acute or chronic HBV infection.
18
{18}
Page 19 of 23 – P030024 Summary of Safety and Effectiveness Data
## Overall Clinical Performance of the *Vitros Anti-HBc* Assay in Samples From Subjects With Clinically Documented Acute or Chronic HBV Infection
| HBV Infection | Number of Samples | Number (%) of Vitros Anti-HBc Reactive Samples | 95% Exact Confidence Interval |
| --- | --- | --- | --- |
| Acute | 8 | 8 (100.0) | 63.06, 100 |
| Chronic | 76 | 75 (98.7) | 92.89, 99.97 |
| Total | 84 | 83 (98.8) | 93.54, 99.97 |
For the information contained above, percent agreement for the *Vitros Anti-HBc* Assay is generally greater than 90% (lower 95% confidence interval). When percent agreement is less than 90% (95% CI lower bounds) it is believed that the lower percent agreement may be due to the low numbers of specimens tested in each disease classification or differences between the two assays, e.g., antigen or detector antibody differences. The above information shows that the *Vitros Anti-HBc* Assay has similar performance compared to a reference anti-HBc assay in the HBV disease categories listed. It is believed that with the above performance the *Vitros Anti-HBc* Assay will furnish useful and meaningful results when used as indicated.
## Clinical Performance of the *Vitros Anti-HBc* Assay in Pre-Vaccination Samples
Serum samples obtained from 41 individuals immediately prior to HBV vaccination were tested with the *Vitros* and reference anti-HBc assays. The results are shown below for both assays.
### *Vitros* and Reference Anti-HBc Results in Pre-Vaccination Samples (N=41)
| Test Result | Reference Anti-HBc Assay | Vitros Anti-HBc Assay |
| --- | --- | --- |
| Initially Negative | 37 | 41 |
| Initially Reactive | 4 | NA |
| Repeatedly Reactive | 0 | NA |
| Total Negative Results | 41 | 41 |
NA = Not applicable
## Potentially Cross-Reacting Subgroups
Samples with evidence of hepatitis A virus infection (HAV) or hepatitis C virus infection (HCV) were identified in a population of 1691 samples prospectively collected from subjects in the U.S with signs or symptoms of, or at risk for, viral hepatitis (Population I). The tables below compare *Vitros* Anti-HBc results with reference anti-HBc results according to the HBV disease classifications assigned to the study subjects.
19
{19}
Page 20 of 23 – P030024 Summary of Safety and Effectiveness Data
## Comparison of Vitros and Reference Anti-HBc Results and HBV Disease Classification among Anti-HAV IgM Reactive Study Subjects - Population I (N=7)
| | Reference Anti-HBc Result | | | | |
| --- | --- | --- | --- | --- | --- |
| | Reactive | | Negative | | |
| HBV Disease Classification | Vitros Anti-HBc Result | | Vitros Anti-HBc Result | | |
| | Reactive | Negative | Reactive | Negative | Total |
| Acute | 0 | 0 | 0 | 0 | 0 |
| Chronic | 0 | 0 | 0 | 0 | 0 |
| Early Recovery | 0 | 0 | 0 | 0 | 0 |
| Recovery | 0 | 0 | 0 | 0 | 0 |
| Recovered | 2 | 0 | 0 | 0 | 2 |
| HBV Vaccine Response | 0 | 0 | 0 | 0 | 0 |
| Not Previously Infected with HBV | 0 | 0 | 0 | 5 | 5 |
| Uninterpretable | 0 | 0 | 0 | 0 | 0 |
| Overall | 2 | 0 | 0 | 5 | 7 |
## Comparison of Vitros and Reference Anti-HBc Results and HBV Disease Classification among Anti-HCV Reactive Study Subjects - Population I (N=353)
| | Reference Anti-HBc Result | | | | |
| --- | --- | --- | --- | --- | --- |
| | Reactive | | Negative | | |
| | Vitros Anti-HBc Result | | Vitros Anti-HBc Result | | |
| HBV Disease Classification | Reactive | Negative | Reactive | Negative | Total |
| Acute | 1 | 0 | 0 | 3 | 4 |
| Chronic | 8 | 1 | 0 | 0 | 9 |
| Early Recovery | 25 | 0 | 0 | 0 | 25 |
| Recovery | 43 | 0 | 0 | 0 | 43 |
| Recovered | 92 | 8 | 0 | 0 | 100 |
| HBV Vaccine Response | 0 | 0 | 0 | 22 | 22 |
| Not Previously Infected with HBV | 0 | 0 | 1 | 147 | 148 |
| Uninterpretable | 0 | 1 | 0 | 1 | 2 |
| Overall | 169 | 10 | 1 | 173 | 353 |
Samples with evidence of hepatitis A virus infection (HAV) or hepatitis C virus infection (HCV) were identified in a population of 315 samples prospectively collected from subjects in an area in India with a high prevalence of viral hepatitis (Population II). The tables below compare Vitros Anti-HBc results with reference anti-HBc results according to the HBV disease classifications assigned to the study subjects.
20
{20}
Page 21 of 23 – P030024 Summary of Safety and Effectiveness Data
## Comparison of Vitros and Reference Anti-HBc Results and HBV Disease Classification among Anti-HAV IgM Reactive Study Subjects - Population II (N=29)
| | Reference Anti-HBc Result | | | | |
| --- | --- | --- | --- | --- | --- |
| | Reactive | | Negative | | |
| HBV Disease Classification | Vitros Anti-HBc Result | | Vitros Anti-HBc Result | | |
| | Reactive | Negative | Reactive | Negative | Total |
| Acute | 10 | 1 | 0 | 7 | 18 |
| Chronic | 0 | 1 | 0 | 0 | 1 |
| Early Recovery | 0 | 0 | 0 | 0 | 0 |
| Recovery | 0 | 0 | 0 | 0 | 0 |
| Recovered | 0 | 0 | 0 | 0 | 0 |
| HBV Vaccine Response | 0 | 0 | 0 | 3 | 3 |
| Not Previously Infected with HBV | 0 | 0 | 0 | 6 | 6 |
| Uninterpretable | 0 | 0 | 0 | 1 | 1 |
| Overall | 10 | 2 | 0 | 17 | 29 |
## Comparison of Vitros and Reference Anti-HBc Results and HBV Disease Classification among Anti-HCV Reactive Study Subjects - Population II (N=90)
| | Reference Anti-HBc Result | | | | |
| --- | --- | --- | --- | --- | --- |
| | Reactive | | Negative | | |
| HBV Disease Classification | Vitros Anti-HBc Result | | Vitros Anti-HBc Result | | |
| | Reactive | Negative | Reactive | Negative | Total |
| Acute | 58 | 0 | 0 | 0 | 58 |
| Chronic | 32 | 0 | 0 | 0 | 32 |
| Early Recovery | 0 | 0 | 0 | 0 | 0 |
| Recovery | 0 | 0 | 0 | 0 | 0 |
| Recovered | 0 | 0 | 0 | 0 | 0 |
| HBV Vaccine Response | 0 | 0 | 0 | 0 | 0 |
| Not Previously Infected with HBV | 0 | 0 | 0 | 0 | 0 |
| Uninterpretable | 0 | 0 | 0 | 0 | 0 |
| Overall | 90 | 0 | 0 | 0 | 90 |
{21}
Page 22 of 23 – P030024 Summary of Safety and Effectiveness Data
## XI. Conclusions Drawn From Studies
These data demonstrate acceptable performance is obtained with the *Vitros Anti-HBc* assay when testing specimens collected in serum and plasma (EDTA and citrate).
The *Vitros Anti-HBc Reagent Pack* and Calibrator can be stored for up to 26 weeks at $2 - 8^{\circ}\mathrm{C}$ $(36 - 46^{\circ}\mathrm{F})$. After opening, the Reagent Pack can be stored on-board the *Vitros Analyzer* $(4 - 8^{\circ}\mathrm{C} (39 - 46^{\circ}\mathrm{F}), \leq 40\%$ relative humidity) for up to 8 weeks, and the Calibrator stored for up to 13 weeks at $2 - 8^{\circ}\mathrm{C}$ $(36 - 46^{\circ}\mathrm{F})$ or $-20^{\circ}\mathrm{C}$ $(-4^{\circ}\mathrm{F})$ (with no more than one freeze-thaw cycle).
The preservative systems that the *Vitros Anti-HBc* assay reagents are formulated with have been shown to meet USP 23 requirements at 26 weeks for Assay Reagent and Calibrator. The Conjugate Reagent meets the USP 23 requirements at 26 weeks with the exception of *carnobacterium divergens*.
The *Vitros Anti-HBc* assay demonstrated adequate precision estimates for within day and between day for each site as well as across all sites, and between replicate, between day, between site and between lots when these variables were introduced.
The *Vitros Anti-HBc* assay has been shown to perform adequately over a 28-day calibration interval.
The *Vitros Anti-HBc* assay agrees with a reference anti-HBc assay in five out of six seroconversion panels. One of the six panels was detected as being repeatedly reactive one bleed later than determined by the reference assay.
Cord blood samples did not interfere with the *Vitros Anti-HBc* assay, and anti-HBc is recovered from cord blood to the same extent as in serum.
The results of the clinical laboratory studies provide reasonable assurance the *Vitros Anti-HBc* assay, when used according to the provided instructions for use, is safe and effective for the laboratory diagnosis of individuals with acute or chronic hepatitis B, or recovery from hepatitis B infection. The presence of anti-HBc may be used as an aid in the determination of exposure to HBV infection for individuals prior to HBV vaccination.
## XII. Panel Recommendation
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Microbiology Advisory Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
## XIII. CDRH Decision
FDA issued an approval order on March 4, 2004.
{22}
Page 23 of 23 – P030024 Summary of Safety and Effectiveness Data
The applicant’s manufacturing facility was inspected on June 14, 2002 and found to be in compliance with the Quality Systems Regulation (21 CFR 820).
## XIV. Approval Specifications
Directions for use: See the labeling.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, precautions and Adverse Events in the labeling.
Postapproval Requirements and Restrictions: See approval order.
23
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.