P030002S027 · Bausch & Lomb, Inc. · MJP · May 20, 2013 · Ophthalmic
Device Facts
Record ID
P030002S027
Device Name
TRULIGN TORIC POSTERIOR CHAMBER INTRAOCULAR LENS
Applicant
Bausch & Lomb, Inc.
Product Code
MJP · Ophthalmic
Decision Date
May 20, 2013
Decision
APPR
Regulation
21 CFR 886.3600
Device Class
Class 3
Indications for Use
The Trulign™ Toric Posterior Chamber Intraocular Lens is intended for primary implantation in the capsular bag of the eye for the visual correction of aphakia and postoperative refractive astigmatism secondary to removal of a cataractous lens in adult patients with or without presbyopia who desire reduction of residual refractive cylinder with increased spectacle independence and improved uncorrected near, intermediate and distance vision.
Device Story
Trulign™ Toric Posterior Chamber IOL is a modified Crystalens (Model AT-45) featuring a toroidal posterior optic surface to compensate for corneal astigmatism. Available in three cylinder powers (1.25 D, 2.00 D, 2.75 D) and spherical equivalents (+4 D to +33 D). Device is implanted in the capsular bag during cataract surgery. Surgeon uses the proprietary Trulign™ Toric IOL Calculator—inputting preoperative keratometry, incision location, and surgically-induced astigmatism (SIA)—to determine optimal cylinder power and axis orientation. Correct alignment of the flat meridian of the IOL with the steep corneal meridian reduces astigmatism. Benefits include improved uncorrected visual acuity and increased spectacle independence. Potential risks include lens rotation, vaulting, tilt, or secondary surgical intervention for repositioning.
Clinical Evidence
Prospective, randomized, single-masked clinical trial (IDE #G990163) with 229 subjects (227 implanted) in US/Canada. Primary endpoints: percent reduction in cylinder, eyes within 0.50D/1.00D of intended correction, and lens axis misalignment. Results: 85% mean reduction in absolute cylinder for all toric cohorts; 78.4% of eyes within 0.50D and 95.5% within 1.00D of intended correction. 100% of eyes showed ≤5° rotation between consecutive visits. Safety endpoints met; no SSIs related to rotational instability. Comparison to ISO 11979-7 SPE rates showed no significant increase in adverse events.
Technological Characteristics
Silicone optic/body; polyimide haptics. Toric posterior surface. Overall diameter 11.5mm; optic diameter 5.0mm. Index of refraction 1.4301. Sterilization via moist heat. Conforms to ISO 11979-2 (optical), ISO 11979-3 (mechanical), and ANSI Z80.30. Software: Trulign™ Toric IOL Calculator (Major Level of Concern).
Indications for Use
Indicated for adult patients with cataractous lens removal requiring primary implantation in the capsular bag for visual correction of aphakia and postoperative refractive astigmatism, with or without presbyopia, desiring reduced residual refractive cylinder and improved uncorrected near, intermediate, and distance vision.
Regulatory Classification
Identification
An intraocular lens is a device made of materials such as glass or plastic intended to be implanted to replace the natural lens of an eye.
Crystalens UVAM model (Model AO1UV) (P030002/S020)
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
## I. GENERAL INFORMATION
Device Generic Name: Toric Posterior Chamber Intraocular Lens (IOL)
Device Trade Name: Trulign™ Toric Posterior Chamber Intraocular Lens Models AT50T, BL1AT and BL1UT; Trulign™ Toric IOL Calculator
Device Procode: MJP
Applicant's Name and Address:
Bausch + Lomb
30 Enterprise, Suite 450
Aliso Viejo, CA 92656
Date(s) of Panel Recommendation: April 8, 2013
Premarket Approval Application (PMA) Number: P030002/S027
Date of FDA Notice of Approval: May 20, 2013
Expedited: not applicable
The original PMA (PMA P030002) was approved on November 14, 2003 and is indicated for primary implantation in the capsular bag of the eye for the visual correction of aphakia in adult patients in whom a cataractous lens has been removed and is intended to provide near, intermediate, and distance vision without spectacles. The Crystalens™ IOL provides approximately one diopter (D) of monocular accommodation. The SSED to support the indication is available on the CDRH website and is incorporated by reference here. The current supplement was submitted to expand the indication for the Trulign™ Toric Posterior Chamber Intraocular Lens.
## II. INDICATIONS FOR USE
The Trulign™ Toric Posterior Chamber Intraocular Lens is intended for primary implantation in the capsular bag of the eye for the visual correction of aphakia and postoperative refractive astigmatism secondary to removal of a cataractous lens in adult patients with or without presbyopia who desire reduction of residual refractive cylinder with increased spectacle independence and improved uncorrected near, intermediate and distance vision.
## III. CONTRAINDICATIONS
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 1
{1}
None
# IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the Trulign™ Toric IOL labeling.
# V. DEVICE DESCRIPTION
The Trulign™ Toric Intraocular Lens is a modification to the currently approved Crystalens Intraocular Lens (Model AT-45) approved on November 14, 2003 under PMA P030002. The only difference between the Trulign™ Toric IOLs and the currently approved Crystalens IOLs is in the incorporation of a toroidal posterior optic surface. The lens toricity results in a periodic change in refractive power between the flat and steep lens meridians. When the flat meridian of the toric IOL is correctly aligned with the steep corneal meridian, the toric IOL compensates the corneal astigmatism. The Trulign™ Toric, Intraocular Lenses are available in three different cylinder powers (1.25 D, 2.00 D and 2.75 D) and in spherical equivalent powers of +4 D to +33 D.
In addition to the clinically studied models, this supplement also seeks approval to add the Toric posterior surface to the currently approved Crystalens aspheric models (Model AT50AO approved October 23, 2009 under PMA P030002/S021) and Crystalens UVAM model (Model AO1UV approved August 5, 2011 under PMA P030002/S020). The physical characteristics of the Trulign™ Toric Intraocular Lens are shown in Table 1.
Table 1: PHYSICAL CHARACTERISTICS OF TRULIGN™ TORIC INTRAOCULAR LENS
| Models | | | |
| --- | --- | --- | --- |
| Feature | AT50T | BL1AT | BL1UT |
| Power Range | +4.0 D to +10.0 D (1.0 D Increments) SE +10.5 D to +33.0 D (0.5 D Increments) SE SE - Spherical Equivalent | | |
| Cylinder Powers | 1.25 D, 2.00 D and 2.75 D | | |
| Anterior Surface | Spherical with Axis-mark | Aspheric with Axis-mark | |
| Posterior Surface | Toric | Aspheric Toric | |
| Overall Diameter | 11.5mm | | |
| Optic Diameter | 5.0 mm | | |
| Plate Overall | 10.5mm | | |
| Plate Configuration | Rectangular | | |
| Material Body & Plates | Silicone (10% UV Cutoff at 354nm) | | Silicone (10% UV Cutoff at 400nm) |
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 2
{2}
| Loop (haptic) material | Polyimide |
| --- | --- |
| Index of Refraction at 546nm | 1.4301 |
For optimal results, the Trulign™ Toric IOL Calculator will be used to select the appropriate cylinder power of the toric lens. The Trulign™ Toric IOL Calculator will calculate the predicted post-operative corneal astigmatism using pre-operative keratometry, phaco/insertion incision location and expected magnitude of surgically-induced astigmatism inputs from the Surgeon. The calculator will account for surgically induced astigmatism, incision location and the patient’s pre-operative corneal astigmatism, and will determine the Toric IOL cylinder power needed and placement orientation in order to best correct a patient’s expected post-operative corneal astigmatism. For optimal toric IOL calculations, it is recommended that surgeons customize their surgically induced corneal astigmatism values based upon individual surgical technique and past results.
## **VI. ALTERNATIVE PRACTICES AND PROCEDURES**
Patients who undergo cataract extraction presently have several non-surgical and surgical alternatives for restoring functional vision of the aphakic eye. Non-surgical options include special cataract glasses or contact lenses. Surgical options such as monofocal, multifocal, simultaneous vision or accommodative IOLs are also available. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
## **VII. MARKETING HISTORY**
The Trulign™ Toric IOL has not been marketed in the United States or any foreign country.
## **VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH**
Potential adverse events and complications accompanying cataract or implant surgery may include, but are not limited to the following: lens subluxation, corneal endothelial damage, non-pigment precipitates, cystoid macular edema, infection (endophthalmitis), retinal detachment, vitreous loss, pupillary block, corneal edema, hypopyon, secondary glaucoma, iris prolapse, vitreous- wick syndrome, uveitis, secondary surgical intervention and pupillary membrane. Secondary surgical interventions include, but are not limited to: lens repositioning (due to decentration, rotation, subluxation, lens vaulting, etc.), lens replacement, vitreous aspirations or iridectomy for pupillary block, wound leak repair and retinal detachment repair.
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 3
{3}
For the specific adverse events that occurred in the clinical studies, please see Section X below.
## IX. SUMMARY OF PRECLINICAL STUDIES
Preclinical studies performed demonstrate the safety and effectiveness of the Trulign™ Toric IOL and the Trulign™ Toric IOL Toric Calculator. The Trulign™ Toric IOL is composed of identical materials to the currently approved Crystalens IOLs. Therefore, no new biocompatibility/toxicological data were provided. Biocompatibility and physical–chemical testing consistent with ISO 10993, Biological Evaluation of Medical Devices, and ISO 11979-5, Ophthalmic Implants - Intraocular Lenses - Part 5: Biocompatibility was provided in the approved PMA P030002 (November 14, 2003) and PMA P030002/S20 on August 5, 2011). The Trulign™ Toric IOL material met the requirements of the International Standards, ISO 11979-5, *Ophthalmic Implants – Intraocular Lenses – Part 5: Biocompatibility*. In addition, sterilization, packaging, shelf life and transport stability testing was performed in P030002 and associated supplements in accordance with AAMI TIR 13: *Principles of Industrial Moist Steam Sterilization*, ISO 17665-Part 1 & 2, *Sterilization of Health Care Products-Moist Heat*, 34/29 USP 2011, *Bacterial Endotoxin Testing*, and ISO 11979-6, *Ophthalmic Implants – Intraocular Lenses – Part 6: Shelf-life and transport stability*. A shelf-life of 5-years was previously established for the Crystalens IOLs.
### A. Laboratory Studies
#### Dimensional, Optical, and Mechanical Testing
To validate the toric modification, dimensional, optical, and mechanical testing was performed on finished, sterilized, Trulign™ Toric IOLs to verify the conformance of the design to the ANSI Standard for Toric IOLs, ANSI Z80.30 *Ophthalmics – Toric Intraocular Lenses*, as well as the International Standards ISO 11979-2 - *Part 2: Optical properties and test methods*, and ISO 11979-3, *Part 3: Mechanical properties and test methods*, and internal specifications. Folding and insertion testing was also performed to verify recovery of lens properties following simulated insertion. The Trulign™ Toric IOLs passed all requirements established in ANSI Z80.30, ISO11979-2, ISO 11979-3 standards, and product specifications. A summary of the results of the dimensional, optical and mechanical testing performed on the Trulign™ Toric IOLs are summarized in **Table 2**.
**Dimensional, Optical and Mechanical Test Requirements Summary**
| Requirement | Result | |
| --- | --- | --- |
| **Optical Requirements** | | |
| Dioptric Power | Power Range (D) | Spherical Acceptance Criteria |
| | 4.0 to 33.0 | ±0.25D |
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 4
{4}
| Cylinder Power | Cylinder SKU range (D) | Cylinder Acceptance Criteria |
| --- | --- | --- |
| | 1.25, 2.00 and 2.75 | ±0.25D |
| Image Quality | Greater than 0.7 modulation unit for MTF @ 70% of design maximum attainable at 100 c/mm. This specification corresponds to the MTF specification described in ISO 11979-2:1999. | |
| Axis Orientation Mark(s) | Combined angular errors of the cylindrical axis mark and any deviation from orthogonality between the meridians of highest and lowest dioptic power within ±5° | |
| Spectral Transmittance | % T=10% at ~354 nm (for BL1AT) % T=10% at ~400 nm (for BL1UT) | |
| Mechanical Requirements | | |
| Overall Diameter | 11.50 ± 0.30mm | |
| Sagitta | ± 0.45mm from nominal | |
| Clear Optic Diameter | 4.50± 0.10mm | |
| Optic Body Diameter | 5.00 ± 0.10mm | |
| Optic Decentration | Mean + 2SD< 0.45 mm (10% clear optic) | |
| Optic Tilt | Mean + 2SD < 5° | |
| Angle of Contact | Angle of Contact mean = 84.88° (within ± 40% of the angle of contact of parent model AT45, at 66.33°) | |
| Compression Force and Decay | Meet the guideline in ISO/TR 22979:2006 as a Level A modification of the parent model AT-45 | |
| Dynamic Fatigue Durability | No breakage or damage after 250,000 cycles of haptic compression | |
| Loop Pull Strength | Loop pull strength mean = 0.74N (≥ 0.25N, per ISO 11979-3) | |
| Surface and Bulk Homogeneity | Essentially free from defects and deviations from intended features of design when inspected under 10x magnification | |
### B. Additional Studies
Software Validation
A software validation was performed for the Trulign™ Toric IOL Calculator according to the procedures described in FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for software Level of Concern: MAJOR. The software for this device was developed under an appropriate software development program. A hazard analysis was performed from both the patient's and user's standpoint and all identified hazards were
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 5
{5}
addressed. These procedures provide the foundation for assuring, to the extent possible, that the software will operate in a manner described in the specifications, and in no other way.
# X. SUMMARY OF PRIMARY CLINICAL STUDY
The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness of subjects undergoing cataract extraction and IOL implantation with the Trulign™ Toric IOL for primary implantation in the capsular bag of the eye for the visual correction of aphakia and postoperative refractive astigmatism secondary to removal of a cataractous lens in adult patients with or without presbyopia who desire reduction of residual refractive cylinder with increased spectacle independence and improved uncorrected near, intermediate and distance vision in the US and Canada under IDE # G990163. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below.
# A. Study Design
Subjects were treated between June 1, 2010 and September 22, 2011. The database for this P030002/S027 reflected data collected through October 15, 2012 and included 229 subjects. There were 9 investigational sites.
The study was a prospective, randomized, single-masked clinical trial to evaluate the safety and effectiveness of the Trulign™ Toric IOL (Models AT50T/AT52T) in reducing postoperative refractive astigmatism in subjects undergoing cataract extraction. IOL implantation was conducted in the US and Canada with study duration of up to 1 year. Subjects in the lowest astigmatic power (1.25D) cohort were randomized to undergo implantation of either the toric test lens or the non-toric spherical control lens (Crystalens Accommodating IOL Models AT50/AT52) in one eye. Subjects in the higher astigmatic power cohorts (2.00D, 2.75D) were implanted with the test lens in one eye.
The effective corneal powers for each of the lens plane cylindrical powers of the test IOLs are shown in Table 3 below:
Table 3: Trulign™ Toric IOL Cylinder Power
| Cylinder Power at IOL Plane (D) | Cylinder Power at Corneal Plane (D) | Range of Predicted Postoperative Corneal Cylinder* (D) |
| --- | --- | --- |
| 1.25 | 0.83 | 0.83 – 1.32 |
| 2.00 | 1.33 | 1.33 – 1.82 |
| 2.75 | 1.83 | 1.83 – 2.50 |
*based upon inclusion criteria and the expectation of 0.50 D incisional effect
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 6
{6}
In order to facilitate toric IOL selection and axis placement, a proprietary Toric Calculator was used to determine the appropriate Trulign™ Toric IOL model and axis of placement for each eye. The calculator accounted for SIA, incision location, and the subject's preoperative corneal astigmatism. In this trial, all cataract incisions were to be placed on the preoperative keratometric steep axis and a fixed SIA value of 0.50 D was used in the Trulign™ Toric IOL Calculator for all study subjects.
For the safety endpoints of best corrected distance visual acuity (BCDVA) and complications/adverse events, study rates for the toric IOLs were compared to ISO SPE rates using a one-sided exact test based on the binomial distribution. For rotational stability, the proportion of the toric eyes with ≤5° of rotation between consecutive visits was reported in accordance with the ANSI Standard for Toric IOLs, Z80.30. Study sample size was based on the ANSI Standard for Toric IOLs, Z80.30.3
Primary effectiveness endpoints were:
- Percent reduction in cylinder, expressed as a percentage of the intended reduction in cylinder, calculated as:
$$\frac{(\text{Postoperative Manifest Refractive Cylinder} - \text{Preoperative Keratometric Cylinder})}{(\text{Intended Postoperative Manifest Refractive Cylinder} - \text{Preoperative Keratometric Cylinder})} * 100$$
- Percent of eyes with "reduction of cylinder" within 0.50 D and within 1.00 D of intended; and
- Lens axis misalignment as determined by a photographic method.
Analysis of the percent reduction in absolute cylinder included continuous summary statistics and 95% confidence intervals (CIs) around the mean. Additionally, a 2-sample t-test assuming unequal variance was performed using the Effectiveness Cohort to test the null hypothesis that the percent reduction of cylinder within the eyes implanted with IOL cylinder power 1.25 D was less than or equal to the percent reduction of cylinder within the eyes implanted with the control IOL. For the endpoint of the percent of eyes with reduction of cylinder within 0.50 D and 1.00 D of target, categorical summary statistics were run, including exact 95% binomial CI around the percent of eyes within 0.50 D and 1.00 D.
Analysis of lens axis misalignment included continuous summaries of the absolute value of the misalignment, the signed value of the misalignment (including a 2-sided 95% tolerance interval, which contains at least 90% of the population), and the proportion of lenses with axis misalignment in the following categories: < 5°, < 10°, < 20°, ≤ 30°, and > 30°.
### 1. Clinical Inclusion and Exclusion Criteria
Enrollment in the Trulign™ Toric study was limited to subjects who met the following inclusion criteria.
- Clinically documented diagnosis of age-related cataract (cortical, nuclear, subcapsular, or a combination) considered amenable to treatment with standard phacoemulsification/ extracapsular cataract extraction
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 7
{7}
- 18 years of age or older and met any applicable local minimum age requirements for IOL implantation following cataract surgery
- Underwent primary IOL implantation for correction of aphakia following continuous curvilinear anterior capsulotomy and phacoemulsification cataract extraction
- Ability to return for scheduled follow-up examinations and had mental capacity to cooperate when undergoing a detailed postoperative exam
- Required spherical lens power of 16.00 D to 27.00 D
- Predicted postoperative corneal astigmatism between 0.83 D and 2.50 D, as determined by the toric calculator
- Potential for BCVA of 20/32 or better in the operated eye (as measured by retinal acuity meter/potential acuity meter testing)
- BCVA equal to or worse than 20/40, with or without a glare source
Subjects were not permitted to enroll in the Trulign™ Toric study if they met any of the following exclusion criteria:
- Anterior segment pathology for which extracapsular phacoemulsification cataract surgery would be contraindicated (e.g., chronic uveitis, iritis, iridocyclitis rubeosis, iridis, clinically significant corneal dystrophy, clinically significant Fuch's dystrophy, clinically significant anterior membrane dystrophy)
- Corneal pathology potentially affecting topography
- Diagnoses of degenerative visual disorders (e.g., macular degeneration or other retinal disorders) predicted to cause future acuity losses to a level of 20/32 or worse
- Conditions associated with increased risk of zonular rupture (zonule rupture during cataract extraction procedure that may affect postoperative rotation, centration, or tilt of the lens)
- Inability to achieve pupil dilation of 5.0 mm
- Any inflammation or edema (swelling) of the cornea, including but not limited to keratitis, keratoconjunctivitis, and keratouveitis
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 8
{8}
- Uncontrolled glaucoma
- Prior retinal detachment
- Diabetic retinopathy (proliferative or nonproliferative)
- Presence of rubella, bilateral congenital, traumatic, or complicated cataract
- Marked microphthalmos or aniridia
- Prior corneal surgery in the planned operative eye
- Irregular corneal astigmatism
- Presence of amblyopia
- Clinically significant retinal pigment or epithelium/macular changes
- Iris or chorioretinal neovascularization
- Presence of optic atrophy
- Chronic use of systemic steroids or immunosuppressive medications
- Concurrent participation in another clinical trial or participation in another clinical trial within 60 days prior to enrollment in this study
- Difference in corneal astigmatism measured with the IOL Master and the topographer greater than 0.50 D using vector analysis
# 2. Follow-up Schedule
Eligible subjects were evaluated preoperatively to obtain a medical history and to establish a baseline for their ocular condition. Subjects in the lowest astigmatic power group were randomized to undergo implantation of either the test or control lens. No corneal or refractive procedures (such as limbal relaxing incisions or astigmatic keratotomies) were permitted at any time during the course of the study.
Postoperatively, subjects underwent complete ophthalmic examinations at regular intervals per the following study visit schedule:
- Preoperative (Form 0A) Day -90 to Day 0
- Operative (Form 0B) Day 0
- Postoperative Visit 1 (Form 1) Day 1-2
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 9
{9}
- Postoperative Visit 2 (Form 2) Day 7-14
- Postoperative Visit 3 (Form 3) Day 30-60
- Postoperative Visit 4 (Form 4) Day 120-180
- Postoperative Visit 5 (Form 5) Day 245-301
- Postoperative Visit 6 (Form 6) Day 330-420
The key timepoints are shown below in the tables summarizing safety and effectiveness.
### 3. Clinical Endpoints
With regards to safety, the study endpoints were preservation of BCVA (distance and near) and incidence of complications and adverse events.
Primary effectiveness endpoints were:
- Percent reduction in cylinder, expressed as a percentage of the intended reduction in cylinder, calculated as:
$$\frac{(\text{Postoperative Manifest Refractive Cylinder} - [\text{Preoperative Keratometric Cylinder}])}{(\text{Intended Postoperative Manifest Refractive Cylinder} - [\text{Preoperative Keratometric Cylinder}])} * 100$$
- Percent of eyes with “reduction of cylinder” within 0.50 D and within 1.00 D of intended; and
- Lens axis misalignment as determined by a photographic method.
The secondary effectiveness endpoints were intermediate visual acuity with distance correction (DCIVA), near visual acuity with distance correction (DCNVA), with and without the minimal reading add, best corrected distance visual acuity (BDCVA), uncorrected distance (UCDVA), intermediate (UCIVA), and near visual acuity (UCNVA), absolute rotation since implantation, and rotational stability
### B. Accountability of PMA Cohort
At the time of database lock, of the 229 subjects enrolled (227 implanted) in PMA study, 212 subjects (92.6%) were available for analysis at the completion of the study, the 330-420 days postoperative visit. Refer to Table 4 below for subject accountability by each scheduled visit.
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 10
{10}
**Table 4: Accountability of Subjects at Each Form Visit (All Enrolled)**
| Subject Status | Preop Exam n (%) | Op Report n (%) | Form 1 n (%) | Form 2 n (%) | Form 3 n (%) | Form 4 n (%) | Form 5 n (%) | Form 6 n (%) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Available for Analysis | 229 (100.0) | 229 (100.0) | 227 (99.1) | 223 (97.4) | 219 (95.6) | 219 (95.6) | 215 (93.9) | 212 (92.6) |
| Discontinued | 0 | 0 | 2 (0.9) | 3 (1.3) | 3 (1.3) | 3 (1.3) | 5 (2.2) | 10 (4.4) |
| Missing at scheduled visit but seen later | 0 | 0 | 0 | 3 (1.3) | 7 (3.1) | 6 (2.6) | 5 (2.2) | 2 (0.9) |
| Not seen but accounted for | 0 | 0 | 0 | 0 | 0 | 0 | 2 (0.9) | 1 (0.4) |
| Lost to follow-up | 0 | 0 | 0 | 0 | 0 | 1 (0.4) | 2 (0.9) | 4 (1.7) |
| Active | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Percent Accountability | | | (100.0) | (98.7) | (96.9) | (96.9) | (96.0) | (96.8) |
Notes: Percentages are based on the number of enrolled subjects.
Percent accountability = 100 x Available for Analysis / (Enrolled - Discontinued - Active)
The figure below displays a breakdown of the analysis populations used in this study:
All Enrolled Cohort, Safety Cohort, Repositioned Lens Cohort, Effectiveness Cohort, Best Case Cohort, Not Best Case Cohort, and Consistent Cohort.

*A total of 41 subjects who reported 49 ocular or corneal pathologies were included in the Not Best Case Cohort.
Notes: n=Control/Toric 1.25 D/Toric 2.00 D/Toric 2.75 D IOL.
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 11
{11}
### C. Study Population Demographics and Baseline Parameters
The demographics (**Table 5**) of the study population are typical for an IOL study performed in the US. In total, 106 subjects were male, and 123 subjects were female. The mean age for subjects in the Control Cohort was 69.8 years and the mean age for subjects in the All Toric Cohort was 70.1 years. The age range for all subjects was 47 to 89 years. With regard to gender, the majority for both cohorts were female, with 55.3% and 52.9% in the Control and All Toric Cohorts, respectively.
**Table 5: Demographics (All Enrolled)**
| | Control IOL (N=76) | Toric IOL 1.25 D (N=82) | Toric IOL 2.00 D (N=47) | Toric IOL 2.75 D (N=24) | All Toric IOL (N=153) |
| --- | --- | --- | --- | --- | --- |
| **Age (years)** | | | | | |
| Total Non-Missing | 76 | 82 | 47 | 24 | 153 |
| Mean (SD) | 69.8 (9.2) | 69.9 (8.8) | 70.4 (8.4) | 70.4 (10.8) | 70.1 (9.0) |
| Median | 71.0 | 70.0 | 70.0 | 73.0 | 70.0 |
| Min, Max | 47, 89 | 48, 88 | 52, 89 | 51, 86 | 48, 89 |
| < 60 | 13 (17.1%) | 8 (9.8%) | 3 (6.4%) | 5 (20.8%) | 16 (10.5%) |
| 60 to 69 | 20 (26.3%) | 32 (39.0%) | 20 (42.6%) | 5 (20.8%) | 57 (37.3%) |
| 70 to 79 | 34 (44.7%) | 27 (32.9%) | 15 (31.9%) | 7 (29.2%) | 49 (32.0%) |
| ≥ 80 | 9 (11.8%) | 15 (18.3%) | 9 (19.1%) | 7 (29.2%) | 31 (20.3%) |
| **Gender** | | | | | |
| Total Non-Missing | 76 | 82 | 47 | 24 | 153 |
| Male | 34 (44.7%) | 35 (42.7%) | 27 (57.4%) | 10 (41.7%) | 72 (47.1%) |
| Female | 42 (55.3%) | 47 (57.3%) | 20 (42.6%) | 14 (58.3%) | 81 (52.9%) |
| **Operative Eye** | | | | | |
| Total Non-Missing | 76 | 82 | 47 | 24 | 153 |
| OD | 42 (55.3%) | 37 (45.1%) | 25 (53.2%) | 12 (50.0%) | 74 (48.4%) |
| OS | 34 (44.7%) | 45 (54.9%) | 22 (46.8%) | 12 (50.0%) | 79 (51.6%) |
### D. Safety and Effectiveness Results
#### 1. Safety Results
The analysis of safety was based on the Safety cohort of 227 implanted subjects and 2 not implanted subjects who underwent surgery for the implantation of a study lens (either test or control). The key safety outcomes for this study are presented in tables 2 and 3. Adverse effects are reported in **Tables 6 – 7** below.
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 12
{12}
# **Adverse effects that occurred in the PMA clinical study:**
**Table 6: ISO 11979-7 SAFETY AND PERFORMANCE ENDPOINTS (SPE) ADVERSE EVENTS REPORTED AT EACH POSTOPERATIVE VISIT, IMPLANTED SUBJECTS (SAFETY, CONTROL IOL)**
| Adverse Event | Unscheduled n/N (%) | Form 1 n/N (%) | Form 2 n/N (%) | Form 3 n/N (%) | Form 4 n/N (%) | Form 5 n/N (%) | Form 6 n/N (%) | Cumulative n/N (%) | ISO SPE (%) | p-value |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cumulative | | | | | | | | | | |
| Endophthalmitis | 0/44 | 0/76 | 0/75 | 0/74 | 0/72 | 0/70 | 0/69 | 0/76 | 0.1 | >0.999 |
| Hypopyon | 0/44 | 0/76 | 0/75 | 0/74 | 0/72 | 0/70 | 0/69 | 0/76 | 0.3 | >0.999 |
| Lens Dislocated From Posterior Chamber | 0/44 | 0/76 | 0/75 | 0/74 | 0/72 | 0/70 | 0/69 | 0/76 | 0.1 | >0.999 |
| Macular Edema | 0/44 | 0/76 | 0/75 | 1/74 (1.4) | 1/73 (1.4) | 0/70 | 0/69 | 1/76 (1.3) | 3.0 | 0.901 |
| Pupillary Block | 0/44 | 0/76 | 0/75 | 0/74 | 0/72 | 0/70 | 0/69 | 0/76 | 0.1 | >0.999 |
| Retinal Detachment | 0/44 | 0/76 | 0/75 | 0/74 | 0/72 | 0/70 | 0/69 | 0/76 | 0.3 | >0.999 |
| Secondary Surgical Intervention | 0/44 | 0/76 | 1/75 (1.3) | 0/74 | 0/72 | 1/70 (1.4) | 0/69 | 2*/76 (2.6) | 0.8 | 0.124 |
| Persistent | | | | | | | | | | |
| Corneal Edema | | | | | | | | 0/69 | 0.3 | >0.999 |
| Iritis | | | | | | | | 0/69 | 0.3 | >0.999 |
| Macular Edema | | | | | | | | 0/69 | 0.5 | >0.999 |
| Raised IOP Requiring Treatment | | | | | | | | 0/69 | 0.4 | >0.999 |
| *Reason for secondary surgical intervention (SSI): 1) IOL reposition due to IOL malposition; 2) IOL Exchange. | | | | | | | | | | |
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 13
{13}
**Table 7: ISO 11979-7 SPE ADVERSE EVENTS REPORTED AT EACH POSTOPERATIVE VISIT, IMPLANTED SUBJECTS (SAFETY, ALL TORIC IOL)**
| Adverse Event | Unscheduled n/N (%) | Form 1 n/N (%) | Form 2 n/N (%) | Form 3 n/N (%) | Form 4 n/N (%) | Form 5 n/N (%) | Form 6 n/N (%) | Cumulative n/N (%) | ISO SPE (%) | p-value |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cumulative | | | | | | | | | | |
| Endophthalmitis | 0/77 | 0/151 | 0/148 | 0/145 | 0/147 | 0/145 | 0/143 | 0/151 | 0.1 | >0.999 |
| Hypopyon | 0/77 | 0/151 | 0/148 | 0/145 | 0/147 | 0/145 | 0/143 | 0/151 | 0.3 | >0.999 |
| Lens Dislocated From Posterior Chamber | 0/77 | 0/151 | 0/148 | 0/145 | 0/147 | 0/145 | 0/143 | 0/151 | 0.1 | >0.999 |
| Macular Edema | 0/77 | 0/151 | 0/148 | 1/145 (0.7) | 1/147 (0.7) | 0/145 | 0/143 | 1/151 (0.7) | 3.0 | 0.990 |
| Pupillary Block | 0/77 | 0/151 | 0/148 | 0/145 | 0/147 | 0/145 | 0/143 | 0/151 | 0.1 | >0.999 |
| Retinal Detachment | 0/77 | 0/151 | 0/148 | 0/145 | 0/147 | 0/145 | 0/143 | 0/151 | 0.3 | >0.999 |
| Secondary Surgical Intervention | 0/77 | 0/151 | 0/148 | 0/145 | 1/147 (0.7) | 0/145 | 0/143 | 1*/151 (0.7) | 0.8 | 0.703 |
| Persistent | | | | | | | | | | |
| Corneal Edema | | | | | | | | 0/143 | 0.3 | >0.999 |
| Iritis | | | | | | | | 0/143 | 0.3 | >0.999 |
| Macular Edema | | | | | | | | 0/143 | 0.5 | >0.999 |
| Raised IOP Requiring Treatment | | | | | | | | 0/143 | 0.4 | >0.999 |
| *Reason for SSI: Reposition of IOL (2.00 D Cohort) not related to lens axis misalignment or rotation. | | | | | | | | | | |
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 14
{14}
## 2. Effectiveness Results
The analysis of effectiveness was based on the 210 subjects who were implanted with a study lens (either Test or Control), whose lens was not repositioned, and who had no major protocol deviations. Key effectiveness outcomes are presented in tables 1 to 18.
The results achieved by 227 subjects followed postoperatively for six months provide data to support the conclusion that eyes implanted with a Trulign™ Toric IOL following cataract extraction achieve visual correction of aphakia and astigmatism.
The data support a reduction in absolute cylinder, rotational stability of the lens, and improvement of uncorrected visual acuity at distance following implantation of Trulign™ Toric IOL.
One of the primary effectiveness endpoints was the mean percent reduction in absolute cylinder. The percent reduction in absolute cylinder is defined as the difference between the postoperative magnitude of the subjective Manifest Refractive cylinder (converted to the corneal plane) and the preoperative magnitude of the keratometric cylinder, divided by the intended reduction in cylinder expressed as a percentage. The intended reduction in cylinder is the difference between the “intended” magnitude of the postoperative Manifest Refractive cylinder (converted to the corneal plane) and the magnitude of the preoperative keratometric cylinder. When comparing the mean (SD) for the Control Cohort 46.5% (43.8%) to that of the 1.25 toric Cohort 79.9% (31.8%), a statistically significant difference (p < 0.001) was demonstrated. The mean (SD) percent reduction of cylinder for the 2.00 toric Cohort was 88.0% (27.1%); for the 2.75 toric Cohort it was 97.4% (19.2%), and for the all toric cohort it was 85.0% (29.3%).
The percent of eyes within 0.50D and 1.00D of intended correction for the Trulign™ Toric Posterior Chamber Intraocular Lens was 78.4% and 95.5%, respectively. The minimum preoperative keratometric cylinder treated was 1.33 D. The results are shown in Table 8. Preservation of best-corrected visual acuity show 98.0% and 100.0% of eyes implanted with a toric lens reported a VA of 20/40 or better at six months at distance and near, respectively. All visual acuity results are presented in Tables 9 - 17. The rotational stability of the toric lens was demonstrated in a cohort of subjects across the Form 3 to Form 4 (1-2 months to 3-6 months) postoperative intervals. One hundred percent (122/122 eyes) of subjects demonstrated less than or equal to 5 degrees of rotation between
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 15
{15}
Consecutive Visits. Additionally, 96.1% of eyes exhibited rotation of less than or equal to 5 degrees between the day of surgery and the Form 4 visit, demonstrating rotational stability in the early postoperative period. The results are shown in **Table 18**. No eyes (0%, 0/20) in the highest available cylinder correction (2.75D) reported significant visual disturbances through Form 4. The results are shown in **Table 19**.
The data provided below includes all subjects enrolled in Clinical Trial 650, including 4 eyes with the control lens AT52, and 6 eyes with the test lens AT52T. Please note, the AT52T is not approved.
**TABLE 8**
PERCENT OF EYES WITH REDUCTION IN CYLINDER WITHIN 0.50 D AND 1.00 D OF INTENDED – FORM 4 (EFFECTIVENESS)
| | Control IOL (N=72) | Toric IOL 1.25 D (N=75) | Toric IOL 2.00 D (N=40) | Toric IOL 2.75 D (N=23) | All Toric IOL (N=138) |
| --- | --- | --- | --- | --- | --- |
| Total Non-Missing | 68 | 74 | 39 | 21 | 134 |
| Within 0.50 D of Intended | 30 (44.1%) | 59 (79.7%) | 31 (79.5%) | 15 (71.4%) | 105 (78.4%) |
| Within 1.00 D of Intended | 49 (72.1%) | 71 (95.9%) | 36 (92.3%) | 21 (100.0%) | 128 (95.5%) |
**TABLE 9**
PRESERVATION OF BCDVA AT EACH EXAMINATION (ALL TORIC IOLS, IMPLANTED SAFETY SUBJECTS)
| | Preop | Form 3 | Form 4 | Form 5 | Form 6 | Unscheduled |
| --- | --- | --- | --- | --- | --- | --- |
| 20/40 or Better | 108 (72.0%) | 143 (99.3%) | 144 (98.0%) | 143 (98.6%) | 141 (98.6%) | 39 (97.5%) |
| Worse than 20/40 | 42 (28.0%) | 1 (0.7%) | 3 (2.0%) | 2 (1.4%) | 2 (1.4%) | 1 (2.5%) |
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 16
{16}
**TABLE 10**
PRESERVATION OF BEST CORRECTED NEAR VISUAL ACUITY (BCNVA) AT EACH EXAMINATION (ALL TORIC IOLS, IMPLANTED SAFETY SUBJECTS)
| | Preop | Form 3 | Form 4 | Form 5 | Form 6 | Unscheduled |
| --- | --- | --- | --- | --- | --- | --- |
| 20/40 or Better | 124 (89.2%) | 141 (98.6%) | 147 (100.0%) | 142 (98.6%) | 142 (99.3%) | 30 (100.0%) |
| Worse than 20/40 | 15 (10.8%) | 2 (1.4%) | 0 | 2 (1.4%) | 1 (0.7%) | 0 |
**TABLE 11**
DISTANCE CORRECTED INTERMEDIAT VISUAL ACUITY (DCIVA) AT 32 INCHES (80 CM) – FORM 4 (EFFECTIVENESS)
| | Control IOL (N=72) | Toric IOL 1.25 D (N=75) | Toric IOL 2.00 D (N=40) | Toric IOL 2.75 D (N=23) | All Toric IOL (N=138) |
| --- | --- | --- | --- | --- | --- |
| DCIVA (logMAR) | | | | | |
| Total Non-Missing | 67 | 74 | 38 | 21 | 133 |
| Mean (SD) | 0.074 (0.142) | 0.060 (0.107) | 0.059 (0.149) | 0.024 (0.139) | 0.054 (0.125) |
| DCIVA (Snellen) | | | | | |
| 20/40 or Better | 64 (95.5%) | 74 (100.0%) | 38 (100.0%) | 21 (100.0%) | 133 (100.0%) |
| Worse than 20/41 | 3 (4.5%) | 0 | 0 | 0 | 0 |
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 17
{17}
**TABLE 12**
DCNVA AT 16 INCHES (40 CM) – FORM 4 (EFFECTIVENESS)
| | Control IOL (N=72) | Toric IOL 1.25 D (N=75) | Toric IOL 2.00 D (N=40) | Toric IOL 2.75 D (N=23) | All Toric IOL (N=138) |
| --- | --- | --- | --- | --- | --- |
| DCNVA (logMAR) | | | | | |
| Total Non-Missing | 67 | 74 | 38 | 21 | 133 |
| Mean (SD) | 0.309 (0.138) | 0.308 (0.130) | 0.310 (0.154) | 0.258 (0.172) | 0.301 (0.144) |
| DCNVA (Snellen) | | | | | |
| 20/40 or Better | 42 (62.7%) | 47 (63.5%) | 22 (57.9%) | 16 (76.2%) | 85 (63.9%) |
| Worse than 20/41 | 25 (37.3%) | 27 (36.5%) | 16 (42.1%) | 5 (23.8%) | 48 (36.1%) |
**TABLE 13**
DCNVA AT 16 INCHES (40 CM) WITH ADD – FORM 4 (EFFECTIVENESS)
| | Control IOL (N=72) | Toric IOL 1.25 D (N=75) | Toric IOL 2.00 D (N=40) | Toric IOL 2.75 D (N=23) | All Toric IOL (N=138) |
| --- | --- | --- | --- | --- | --- |
| DCNVA with Add (logMAR) | | | | | |
| Total Non-Missing | 68 | 74 | 39 | 21 | 134 |
| Mean (SD) | 0.045 (0.072) | 0.039 (0.075) | 0.031 (0.066) | 0.038 (0.074) | 0.036 (0.072) |
| DCNVA with Add (Snellen) | | | | | |
| 20/40 or Better | 68 (100.0%) | 74 (100.0%) | 39 (100.0%) | 21 (100.0%) | 134 (100.0%) |
| Add (D) | | | | | |
| Mean (SD) | 1.599 (0.575) | 1.473 (0.451) | 1.423 (0.494) | 1.405 (0.599) | 1.448 (0.486) |
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 18
{18}
**TABLE 14**
# BCDVA WITHOUT GLARE – FORM 4 (EFFECTIVENESS)
| | Control IOL (N=72) | Toric IOL 1.25 D (N=75) | Toric IOL 2.00 D (N=40) | Toric IOL 2.75 D (N=23) | All Toric IOL (N=138) |
| --- | --- | --- | --- | --- | --- |
| BCDVA (logMAR) | | | | | |
| Total Non-Missing | 68 | 74 | 39 | 21 | 134 |
| Mean (SD) | 0.012 (0.094) | 0.003 (0.073) | -0.003 (0.076) | 0.019 (0.180) | 0.004 (0.097) |
| BCDVA (Snellen) | | | | | |
| 20/40 or Better | 68 (100.0%) | 74 (100.0%) | 39 (100.0%) | 20 (95.2%) | 133 (99.3%) |
| Worse than 20/41 | 0 | 0 | 0 | 1 (4.8%) | 1 (0.7%) |
**TABLE 15**
# UCDVA – FORM 4 (EFFECTIVENESS)
| | Control IOL (N=72) | Toric IOL 1.25 D (N=75) | Toric IOL 2.00 D (N=40) | Toric IOL 2.75 D (N=23) | All Toric IOL (N=138) |
| --- | --- | --- | --- | --- | --- |
| UCDVA (logMAR) | | | | | |
| Total Non-Missing | 68 | 74 | 39 | 21 | 134 |
| Mean (SD) | 0.189 (0.181) | 0.099 (0.140) | 0.082 (0.119) | 0.091 (0.130) | 0.093 (0.132) |
| UCDVA (Snellen) | | | | | |
| 20/40 or Better | 51 (75.0%) | 71 (95.9%) | 39 (100.0%) | 21 (100.0%) | 131 (97.8%) |
| Worse than 20/41 | 17 (25.0%) | 3 (4.1%) | 0 | 0 | 3 (2.2%) |
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 19
{19}
TABLE 16
UCIVA – FORM 4 (EFFECTIVENESS)
| | Control IOL (N=72) | Toric IOL 1.25 D (N=75) | Toric IOL 2.00 D (N=40) | Toric IOL 2.75 D (N=23) | All Toric IOL (N=138) |
| --- | --- | --- | --- | --- | --- |
| UCIVA (logMAR) | | | | | |
| Total Non-Missing | 68 | 74 | 39 | 21 | 134 |
| Mean (SD) | 0.069 (0.153) | 0.044 (0.116) | 0.058 (0.155) | 0.004 (0.117) | 0.042 (0.129) |
| UCIVA (Snellen) | | | | | |
| 20/40 or Better | 64 (94.1%) | 73 (98.6%) | 37 (94.9%) | 21 (100.0%) | 131 (97.8%) |
| Worse than 20/41 | 4 (5.9%) | 1 (1.4%) | 2 (5.1%) | 0 | 3 (2.2%) |
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 20
{20}
TABLE 17
UCNVA – FORM 4 (EFFECTIVENESS)
| | Control IOL (N=72) | Toric IOL 1.25 D (N=75) | Toric IOL 2.00 D (N=40) | Toric IOL 2.75 D (N=23) | All Toric IOL (N=138) |
| --- | --- | --- | --- | --- | --- |
| UCNVA (logMAR) | | | | | |
| Total Non-Missing | 68 | 74 | 39 | 21 | 134 |
| Mean (SD) | 0.286 (0.137) | 0.284 (0.155) | 0.310 (0.142) | 0.268 (0.149) | 0.289 (0.150) |
| UCNVA (Snellen) | | | | | |
| 20/40 or Better | 49 (72.1%) | 53 (71.6%) | 24 (61.5%) | 17 (81.0%) | 94 (70.1%) |
| Worse than 20/41 | 19 (27.9%) | 21 (28.4%) | 15 (38.5%) | 4 (19.0%) | 40 (29.9%) |
TABLE 18
ROTATIONAL STABILITY BETWEEN CONSECUTIVE VISITS – FORM 4 (CONSISTENT)
| | Toric IOL 1.25 D (N=69) | Toric IOL 2.00 D (N=34) | Toric IOL 2.75 D (N=19) | All Toric IOL (N=122) |
| --- | --- | --- | --- | --- |
| Absolute Value of Rotation (°) | | | | |
| Total Non-Missing | 69 | 34 | 19 | 122 |
| Mean (SD) | 1.074 (0.966) | 1.166 (0.906) | 1.537 (1.406) | 1.172 (1.034) |
| Lenses Rotating ≤5° Since Last Visit | 69 (100.0%) | 34 (100.0%) | 19 (100.0%) | 122 (100.0%) |
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 21
{21}
# **TABLE 19**
Subjects Experiencing One or More Significant Visual Disturbances – Form 4
(Effectiveness, Subjects Whose Lens Was Not Repositioned)
| | Control IOL (N=72) n(%) | Toric IOL 1.25 D (N=75) n(%) | Toric IOL 2.00 D (N=40) n(%) | Toric IOL 2.75 D (N=23) n(%) | All Toric IOL (N=138) n(%) |
| --- | --- | --- | --- | --- | --- |
| Total Non-Missing | 66 | 74 | 39 | 20 | 133 |
| Significant Visual Disturbance | 6 (9.1) | 0 | 1 (2.6) | 0 | 1 (0.8) |
| No Significant Visual Disturbance | 60 (90.9) | 74 (100.0) | 38 (97.4) | 20 (100.0) | 132 (99.2) |
# 3. Subgroup Analyses
The following preoperative characteristics were evaluated for potential association with outcomes device effectiveness across the range of preoperative corneal astigmatism (Tables 20-24). Tables 25-26 provide information on the predictive error of the Trulign™ Toric IOL calculator. Table 27 shows the dioptric change in cylinder (D) by IOL cylinder power including control.
# **TABLE 20**
Percentage Reduction of Cylinder (ANSI) Stratified by Predicted Postoperative Corneal Cylinder
| Predicted Postoperative Corneal Cylinder (D) | Percentage Reduction of Cylinder (ANSI) | |
| --- | --- | --- |
| | Toric IOL 1.25 D (n) mean | Sphere IOL (n) mean |
| 0.83 to 1.07 | (45) 77.52 | (51) 44.59 |
| 1.08 to 1.32 | (29) 83.55 | (17) 52.23 |
Notes: Percentage reduction of cylinder (ANSI) = percentage of intended reduction achieved from preoperative keratometric cylinder to postoperative manifest cylinder
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 22
{22}
# **TABLE 21**
Percent Reduction of Preoperative Corneal Cylinder Stratified By Preoperative Keratometric Cylinder
| | Percent Reduction of Cylinder | |
| --- | --- | --- |
| Preoperative Keratometric Cylinder (D) | Toric IOL 1.25 D (n) mean | Sphere IOL (n) mean |
| 1.33 to 1.57 | (43) 71.12 | (50) 41.23 |
| 1.58 to 1.82 | (29) 65.50 | (17) 41.33 |
Notes: Percent reduction of cylinder = (preoperative keratometric cylinder – postoperative refractive cylinder) / preoperative keratometric cylinder
# **TABLE 22**
Refractive cylinder stratified by preoperative keratometric cylinder
| | Magnitude of Refractive Cylinder (D) | |
| --- | --- | --- |
| Preoperative Keratometric Cylinder (D) | Toric IOL 1.25 D n mean | Sphere IOL n mean |
| 1.33 to 1.57 | 43 0.41 | 50 0.86 |
| 1.58 to 1.82 | 29 0.59 | 17 1.01 |
# **TABLE 23**
logMAR uncorrected distance visual acuity (UCDVA) stratified by preoperative keratometric cylinder
| | logMAR UCDVA | |
| --- | --- | --- |
| Preoperative Keratometric Cylinder (D) | Toric IOL 1.25 D n mean | Sphere IOL n mean |
| 1.33 to 1.57 | 43 0.10 | 50 0.20 |
| 1.58 to 1.82 | 29 0.11 | 17 0.16 |
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 23
{23}
# **TABLE 24**
Change in absolute cylinder stratified by preoperative keratometric cylinder
| Preoperative Keratometric Cylinder (D) | Change in Absolute Cylinder (D) | Toric IOL 1.25 D | Sphere IOL |
| --- | --- | --- | --- |
| 1.33 to 1.57 | Total Non-Missing | 43 | 50 |
| | Mean (SD) | -1.020 (0.407) | -0.573 (0.609) |
| | Median | -1.010 | -0.610 |
| | Min, Max | -1.52, 0.14 | -1.56, 1.23 |
| | < 0.50 D | 43 100.0% | 48 96.0% |
| | > 0.50 D | 0 | 2 4.0% |
| | ≤± 0.50 D | 6 14.0% | 21 42.0% |
| | >-0.50 D | 6 14.0% | 23 46.0% |
| | <-0.50 D | 37 86.0% | 27 54.0% |
| | <-0.75 D | 33 76.7% | 22 44.0% |
| | <-1.00 D | 22 51.2% | 11 22.0% |
| | Missing | 0 | 0 |
| 1.58 to 1.82 | Total Non-Missing | 29 | 17 |
| | Mean (SD) | -1.104 (0.454) | -0.679 (0.564) |
| | Median | -1.100 | -0.780 |
| | Min, Max | -1.81, 0.16 | -1.69, 0.49 |
| | < 0.50 D | 29 100.0% | 17 100.0% |
| | > 0.50 D | 0 | 0 |
| | ≤± 0.50 D | 3 10.3% | 6 35.3% |
| | >-0.50 D | 3 10.3% | 6 35.3% |
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 24
{24}
| Preoperative Keratometric Cylinder (D) | Change in Absolute Cylinder (D) | Toric IOL 1.25 D | Sphere IOL |
| --- | --- | --- | --- |
| 1.58 to 1.82 (Continued) | Total Non-Missing | 29 | 17 |
| | <-0.50 D | 26 89.7% | 11 64.7% |
| | <-0.75 D | 25 86.2% | 9 52.9% |
| | <-1.00 D | 20 69.0% | 4 23.5% |
| | Missing | 0 | 0 |
| | | | |
Notes: Change in absolute cylinder = |postoperative refractive cylinder| - |preoperative keratometric cylinder|
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 25
{25}
# **TABLE 25**
Error in the predicted postoperative keratometric astigmatism at 120 to 180 days after surgery
| | Randomized Eyes (N=150) |
| --- | --- |
| | |
| Signed Bias versus Predicted (D) | |
| Total Non-Missing | 141 |
| Mean (SD) | 0.132 (0.398) |
| Median | 0.120 |
| Min, Max | -0.78, 1.24 |
| | |
| Absolute Error (D) | |
| Total Non-Missing | 141 |
| Mean (SD) | 0.335 (0.250) |
| Median | 0.270 |
| Min, Max | 0.01, 1.24 |
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 26
{26}
# **TABLE 26**
Error in the predicted postoperative keratometric steep axis at 120 to 180 days after surgery
| | Randomized Eyes (N=150) |
| --- | --- |
| Signed Bias versus Predicted (degrees) | |
| Total Non-Missing | 141 |
| Mean (SD) | -0.1 (17.9) |
| Median | -1.0 |
| Min, Max | -49, 88 |
| | |
| Absolute Error (degrees) | |
| Total Non-Missing | 141 |
| Mean (SD) | 11.7 (13.6) |
| Median | 7.0 |
| Min, Max | 0, 88 |
# **TABLE 27**
Dioptric change in cylinder (D) by IOL cylinder power including control
| | Sphere | Toric 1.25D | Toric 2.00D | Toric 2.75D |
| --- | --- | --- | --- | --- |
| N | 68 | 74 | 39 | 21 |
| Mean (SD) | -0.61 (0.59) | -1.06 (0.42) | -1.61 (0.50) | -2.27 (0.45) |
| Median | -0.64 | -1.07 | -1.73 | -2.37 |
| Min, Max | -1.69, 1.23 | -1.81, 0.16 | -2.11, 0.46 | -2.93, -1.34 |
Notes: Dioptric change in cylinder = |postoperative manifest cylinder| - |preoperative keratometric cylinder|
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 27
{27}
### **E. Financial Disclosure**
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 9 investigators, of which none were full-time or part-time employees of the applicant and 3 investigators had disclosable financial interests/arrangements as defined in 21 CFR 54.2(a), (b), (c) and (f) and described below:
- Significant payment of other sorts: 3 investigators
The applicant has adequately disclosed the financial interest/arrangements with clinical investigators. Statistical analyses were conducted by FDA to determine whether the financial interests/arrangements had any impact on the clinical study outcome. The information provided does not raise any questions about the reliability of the data.
## **XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION**
### **A. Panel Meeting Recommendation**
At an advisory meeting held on April 8, 2013. The Ophthalmic Devices Advisory Panel voted 10/0/2 that there is reasonable assurance the Trulign™ Toric IOL is safe, 10/1/1 that there is reasonable assurance that the device is effective, and 10/1/1 that the benefits of the device do outweigh the risks in subjects who meet the criteria specified in the proposed indication. The panel consensus was that the evidence supported device safety and the effectiveness of the lens in providing astigmatic correction and providing improved intermediate vision, as well as some benefit to near vision. However, in panel discussions regarding the accommodative ability of the lens, including published literature and clinical trial data, the panel did not believe that the data available supported the claim that this IOL provides approximately one diopter of accommodation.
### **B. FDA's Post-Panel Action**
The panel discussed the evidence concerning the accommodative ability of the lens including published literature and clinical trial data. They did not believe that the data available supported the claim that this IOL provides approximately one diopter of accommodation. Therefore, statements related to 'accommodation' were removed from the labeling.
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 28
{28}
## **XII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES**
### **A. Effectiveness Conclusions**
Significant reduction of astigmatism was achieved for all models. The Trulign™ Toric IOL astigmatic correction combined with the incisional effects of the cataract surgery provided on average about 85% of the intended astigmatic correction.
For the randomized portion of the trial (subjects qualifying for the lowest cylinder power correction), the mean percent reduction in astigmatism (compared to the intended) was ~80% for the toric IOL arm compared to 46% for the control spherical IOL arm. It is important to note that part of this reduction was due to incisional effects of the procedure.
The percent of eyes within 0.50D and 1.00D of intended correction for the Trulign™ Toric Posterior Chamber Intraocular Lens was 78.4% and 95.5%, respectively.
Approximately 96% of eyes exhibited rotation of less than or equal to 5 degrees between the day of surgery and the 6 month visit.
In general, good best-corrected distance visual acuities and good uncorrected distance visual acuities were seen in all Trulign™ IOL cylinder powers studied.
### **B. Safety Conclusions**
The risks of the device are based on nonclinical laboratory and/or animal studies as well as data collected in a clinical study conducted to support PMA approval as described above.
Potential AEs and complications accompanying cataract and IOL implantation surgery may include, but are not limited to the following: corneal endothelial damage, infection (endophthalmitis), inflammation, retinal detachment, vitritis, cystoid macular edema, corneal edema, pupillary block, iris prolapse, hypopyon, elevated intraocular pressure (IOP) requiring treatment, and secondary surgical intervention. Because of the design of the Trulign™ Toric IOL, the device can vault forward and get stuck in a forward or tilted position.
The main serious AEs specifically related to use of a toric device, is the potential for secondary surgery to correct IOL misalignment (rotation). In addition, while not considered an AE, there is risk of a poor uncorrected acuity outcome, if there is significant IOL rotation. The secondary surgery generally provides improved uncorrected visual acuity in the event of IOL rotation. However, it is not required for ocular health or acceptable vision, as the eye can generally achieve good acuity with spectacle correction.
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 29
{29}
Visual distortion is a theoretical potential visual symptom related to toric IOL rotation. It can vary in perceived severity and is most likely to be of limited duration. If very severe and persistent, a patient might request secondary surgery to remove the IOL (a more serious adverse event), but this result was not seen. In addition, because of the small optic, there is potential for visual symptoms such as glare or halos from stray light.
Potential SSIs can include, but are not limited to, lens repositioning (due to vaulting, decentration, rotation, subluxation, etc.), lens replacement, vitreous aspirations or iridectomy for pupillary block, wound leak repair, retinal detachment repair, corneal transplant, and lens replacement due to refractive error or severe inflammation.
All safety endpoints of Study 650 were met. The Trulign™ Toric IOL demonstrated preservation of BCDVA and BCNVA, and the incidences of complications and AEs reported for the All Toric population did not exceed the ISO 11979-7 SPE rates of historical controls or those reported with the parent Crystalens IOL, Model AT45 (PMA P030002). No SSIs were related to rotational instability.
### C. Benefit-Risk Conclusions
The probable benefits of the device are also based on data collected in a clinical study conducted to support PMA approval as described above. The results from Study 650 demonstrate that the Trulign™ Toric IOL is a safe and effective option for cataract patients with astigmatism who, with the counsel of their surgeon, select the Trulign™ Toric IOL. This may reduce the need for patients to pursue spectacle, contact lens, or other refractive correction. These findings support a favorable risk-benefit ratio for visual correction of aphakia and astigmatism with the Trulign™ Toric IOL.
In conclusion, given the available information above, the data support that for primary implantation in the capsular bag of the eye for the visual correction of aphakia and postoperative refractive astigmatism secondary to removal of a cataractous lens in adult patients with or without presbyopia who desire reduction of residual refractive cylinder with increased spectacle independence and improved uncorrected near, intermediate and distance vision, the probable benefits outweigh the probable risks.
### D. Overall Conclusions
The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the indications for use. The data and information provided in PMA P030002/S027 support the safety and effectiveness of the Trulign™ Toric IOL. Under controlled clinical investigation (Study 650), the Trulign™ Toric IOL met or exceeded the safety and effectiveness endpoints established for the trial. The Trulign™ Toric IOL demonstrated rotational stability with favorable effectiveness and safety compared with the Control IOL and
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 30
{30}
the historical control (ISO 11979-7 SPE) population. The Trulign™ Toric IOL provides correction of astigmatism in a single procedure. The Trulign™ Toric IOL provides distance and intermediate vision, and improved near vision. These findings support the safe and effective use of the Trulign™ Toric IOL in the proposed patient population.
### XIII. CDRH DECISION
CDRH issued an approval order on May 20, 2013. The final conditions of approval cited in the approval order are described below.
*Trulign™ Toric IOL New Enrollment Study:* The study will be a single arm, multi-site, prospective study with a sample size of 500 eyes at the end of 3 year follow-up. The specific question to be answered is the incidence of IOL vaulting (i.e., position change such as clinically significant anterior vaulting, clinically significant tilt, and secondary surgical intervention related to such vaulting) up to three years post implant. Follow-up assessments will occur at 1 day, 1 week, 1 month, 6 month, 1 year, and then every year until the 3rd year of follow-up. Data collection will include observations, symptoms, diagnosis, treatment, sequelae, and resolution.
### XIV. APPROVAL SPECIFICATIONS
Directions for use: See device labeling.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling.
Post-approval Requirements and Restrictions: See approval order.
PMA P030002/S027: FDA Summary of Safety and Effectiveness Data
Page 31
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.