← Product Code [LXA](/productcode/LXA) · P020049

# PROCOL VASCULAR BIOPROSTHESIS (P020049)

_LeMaitre Vascular, Inc. · LXA · Jul 29, 2003 · Cardiovascular · APWD_

**Canonical URL:** https://fda-staging.innolitics.com/device/P020049

## Device Facts

- **Applicant:** LeMaitre Vascular, Inc.
- **Product Code:** [LXA](/productcode/LXA.md)
- **Decision Date:** Jul 29, 2003
- **Decision:** APWD
- **Device Class:** Class 3
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The ProCol® Vascular Bioprosthesis is intended for the creation of a bridge graft for vascular access subsequent to at least one previously failed prosthetic access graft.

## Device Story

ProCol® Vascular Bioprosthesis is a glutaraldehyde-fixed bovine mesenteric vein graft used as a bridge for hemodialysis access. The device is implanted surgically by a physician to create a conduit for needle access in patients with prior failed synthetic grafts. It features a 6 mm nominal diameter, 10-40 cm length, and is supplied in buffered saline. The inflow end is beveled for identification and irrigation; a green suture loop marks the outflow end. The device serves as an alternative to synthetic grafts, peritoneal dialysis, or catheter-based access. By providing a biological conduit, it aims to reduce complications like infection and thrombosis compared to synthetic alternatives, thereby extending secondary patency and reducing the frequency of interventions required to maintain access.

## Clinical Evidence

Multi-center, non-randomized clinical study (n=183 ProCol; n=93 concomitant synthetic control; n=128 internal control). Primary endpoint: primary and secondary patency at one year. Results: Primary patency not statistically different. Secondary patency significantly higher for ProCol (66% at 12 months) vs. concomitant control (56%) and internal control (34%) (p=0.0361). ProCol showed significantly lower rates of infection, thrombosis, and seroma compared to synthetic controls. Bench testing included biocompatibility (ISO 10993), mechanical testing (burst strength, suture retention, kink radius), and canine implantation studies confirming patency and healing.

## Technological Characteristics

Bovine mesenteric vein, glutaraldehyde-fixed. Nominal diameter 6 mm; length 10-40 cm. Wall thickness 0.48 mm. Suture retention 1.61 lbs. Kink radius 16 mm. Terminally sterilized via gamma radiation. Supplied in borosilicate glass packaging with buffered physiological saline. No software or electrical components.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

# SUMMARY OF SAFETY AND EFFECTIVENESS DATA

## 1. GENERAL INFORMATION

Device Generic Name: Vascular Access Graft

Device Trade Name: The ProCol® Vascular Bioprosthesis

Applicant’s Name and Address: Hancock Jaffe Laboratories, Inc.
2807 McGaw Avenue
Irvine, CA 92614

PMA Application Number: P020049

Date of Panel Recommendation: None

Date of Notice of Approval to the Applicant: July 29, 2003

## 2. INDICATIONS AND USAGE

The ProCol® Vascular Bioprosthesis is intended for the creation of a bridge graft for vascular access subsequent to at least one previously failed prosthetic access graft.

## 3. CONTRAINDICATIONS

There are no contraindications related to the use of the ProCol® Vascular Bioprosthesis.

## 4. WARNINGS AND PRECAUTIONS

The Warnings and Precautions can be found in the Instructions for Use for this product.

## 5. DEVICE DESCRIPTION

The ProCol® Vascular Bioprosthesis is derived from a continuous segment of bovine mesenteric vein processed with glutaraldehyde. Branches are ligated with surgical suture. The bioprosthesis has a nominal diameter of 6 mm and is manufactured with a minimum length ranging from 10 cm to 40 cm. Flow direction is indicated by a loop of green surgical suture attached to the outflow end of the device. The inflow end of the device is beveled both to identify the inflow end of the vessel and to assist in inserting the irrigation needle during the pre-implantation rinse procedure. Each device is identified by a unique serial number imprinted on a PETG identification tag sutured to the bioprosthesis.

The finished bioprosthesis is supplied on a glass mandril, immersed in buffered physiological saline. Devices are packaged in a borosilicate glass packaging tube with screw type closure and terminally sterilized by exposure to gamma radiation.

Summary of Safety and Effectiveness Data
P020049

{1}

Each unit is supplied inside an individual corrugated carton with polyethylene foam liner and sealed with PVC shrink wrap. The device has a shelf life of 54 months.

## 6. ALTERNATE PRACTICES AND PROCEDURES

For a patient population consisting of at least one already failed synthetic access graft, alternative practices to achieve dialysis would include placement of another synthetic graft to create another site for needle dialysis, or resorting to peritoneal or catheter based dialysis. Creation of an access site using the patient’s native vein is another alternative.

## 7. MARKETING HISTORY

The ProCol® Vascular Bioprosthesis has had a CE mark approval since June 1998. Approximately 4,000 devices have been shipped to ten European countries since October 1992 in diameters ranging from 3 to 6 mm. In countries outside of the U.S., the device is used to establish vascular access, accomplish peripheral vascular reconstruction, or to reconstruct one or more coronary arteries. Distribution has been to the following countries: The United Kingdom, Switzerland, Italy, France, Germany, The Netherlands, Austria, Spain, Turkey, and Greece. The ProCol® Vascular Bioprosthesis has not been withdrawn from marketing for any reason relating to safety or effectiveness of the device.

## 8. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH

The following adverse events may be associated with use of a vascular access graft: needle stick damage resulting in bleeding and/or pseudoaneurysms, hemorrhage, anastomotic aneurysms, steal, patient sensitivity to device materials, graft dilatation, thrombosis/occlusion of graft, infection, pain, swelling of affected limb, embolic events, stenosis, slow wound healing, failure to achieve access, events associated with an invasive surgical procedure and possibly death.

## 9. SUMMARY OF PRE-CLINICAL STUDIES

The ProCol® Vascular Bioprosthesis for vascular access was subjected to a comprehensive pre-clinical testing program. Testing methods were in accordance with the FDA draft “Guidance for the Preparation of Research and Marketing Applications for Vascular Graft Prostheses”.

### 9.1 BIOCOMPATIBILITY, IMMUNOLOGY, AND TOXICOLOGY STUDIES

Selected toxicity and biocompatibility studies were conducted for the bioprosthesis. Testing was conducted on finished irradiated product of the same dimensions and using the same manufacturing processes as the current model but with a fabric cover. The

Summary of Safety and Effectiveness Data
P020049
Page 2
9

{2}

current model would not be more toxic or mutagenic than the tested device which includes the presence of the additional component (polyester). The testing was conducted in accordance with Good Laboratory Practices per 21CFR 58 and ISO 10993. The ProCol® bioprosthesis is classified by ISO 10993 as an implantable device, blood contact, C - long-term.

Results of the testing performed are summarized in Table 1. As there is a long history of use of glutaraldehyde fixed bovine tissue, and there was a history of use with the ProCol® device outside of the U.S., the sponsor did not conduct many of the recommended biocompatibility tests, such as sensitization, irritation, systemic toxicity, thrombogenicity, chronic toxicity or carcinogenicity. The results of the testing performed and other animal studies (described in the next section) suggested a biocompatible device.

TABLE 1: BIOCOMPATIBILITY STUDIES

|  TEST | TEST METHOD/EXTRACT | RESULT  |
| --- | --- | --- |
|  Cytotoxicity – MEM Elution Method | 60 sq cm test article placed in 20 mls MEM supplemented with 5% bovine serum, extracted at 37° C for 24 hours. MEM aliquot as negative control. Latex as positive control. | Intermediate toxicity after 72 hours was comparable to a control product.  |
|  Genotoxicity - Ames Test for Mutagenicity | Ames et. al (1975), modified to permit utilization of test article extracts (saline extract, incubated at 37° C for minimum 48 hours). | Test article was non-mutagenic.  |
|  Hemolysis – Direct Contact Method | 1 gram test article placed in 5 ml 0.9% Sodium Chloride USP. 5 ml water and 0.1 ml blood as positive control. 5 ml 0.9 % Sodium Chloride USP and 0.1 ml blood as a negative control. | Test article was non-hemolytic.  |
|  Sensitization, Irritation, Systemic toxicity, Sub-chronic toxicity, Implantation, Coagulation | See animal and clinical study results. | No positive indications.  |

## 9.2 ANIMAL STUDIES

Objective: To verify functionality and healing characteristics of the ProCol® bioprosthesis in a canine model.

Methods: Thirteen animals received 6 mm bioprostheses as either descending aorta to vena cava shunts (Group 1, n=8), or descending aorta to iliac artery shunts (Group 2,

Summary of Safety and Effectiveness Data
P020049

{3}

n=5). Four Group 1 animals were implanted with an earlier device model featuring a cloth covered graft. Angiography was performed at several postoperative intervals. Grafts were examined by physical examination and blood flow via cannulation at explant. Following explant, in situ examination, gross anatomical evaluation, and histological evaluation of the explanted graft was performed.

Results: The results are summarized in Table 2.

TABLE 2: ANIMAL STUDIES

|  ID # | Location | Date Implanted | Date of Sacrifice | Reason for sacrifice | Graft condition at sacrifice | Result  |
| --- | --- | --- | --- | --- | --- | --- |
|  F45 | Descending aorta to vena cava | 11/8/83 | 5/29/84 | ≥6 months post implant | Patent | No significant inflammatory or foreign body response to graft  |
|  F85 | Descending aorta to vena cava | 11/16/83 | 11/20/83 | Animal experienced non-graft related complications of surgery. | N/A - considered lost from study | N/A – Less than 6 months  |
|  F118 | Descending aorta to vena cava | 1/25/84 | 1/27/84 | Animal expired | N/A – considered lost from study | N/A – Less than 6 months  |
|  F124 | Descending aorta to vena cava | 1/30/84 | 10/19/84 | ≥6 months post implant | Patent | No significant inflammatory or foreign body response to graft  |
|  F202 | Descending aorta to vena cava | 11/28/83 | 6/21/84 | ≥6 months post implant | Patent | No significant inflammatory or foreign body response to graft  |
|  F227 | Descending aorta to vena cava | 4/4/84 | 3/14/85 | ≥6 months post implant | Patent | No significant inflammatory or foreign body response to graft  |
|  F244 | Descending aorta to vena cava | 12/6/83 | 6/21/84 | ≥6 months post implant | Patent | No significant inflammatory or foreign body response to graft  |
|  F254 | Descending aorta to vena cava | 2/21/84 | 10/19/84 | ≥6 months post implant | Patent | No significant inflammatory or foreign body response to graft  |
|  D240 | Descending aorta to iliac artery | 4/23/84 | 6/12/84 | Graft occluded | Occluded – technique related to postoperative leg position causing kinking | N/A – Less than 6 months  |
|  D248 | Descending aorta to iliac artery | 4/23/84 | 6/12/84 | Graft occluded | Occluded – technique related to postoperative leg position causing kinking | N/A – Less than 6 months  |
|  D269 | Descending aorta to iliac artery | 7/23/84 | 3/14/85 | ≥6 months post implant | Patent | No significant inflammatory or foreign body response to graft  |
|  D274 | Descending aorta to iliac artery | 7/13/84 | 10/10/85 | ≥6 months post implant | Patent | No significant inflammatory or foreign body response to graft  |
|  D185 | Descending aorta to iliac artery | 7/31/85 | 11/13/86 | ≥6 months post implant | Patent | No significant inflammatory or foreign body response to graft  |

Summary of Safety and Effectiveness Data
P020049

{4}

9.3 MECHANICAL TESTING

A summary of the mechanical testing conducted on the ProCol® Bioprosthesis is presented in Table 3. Testing was in substantive compliance with AAMI/ISO 7 198 (1998). The data demonstrates the mechanical properties of the ProCol® Bioprosthesis are suitable for its intended use for vascular access.

TABLE 3: MECHANICAL TESTING

|  DESCRIPTION OF TEST | SPECIMEN DESCRIPTION | RESULT  |
| --- | --- | --- |
|  Visual Inspection | Finished device (Crosslinked vein with dissection and ligation processes completed). Inspection conducted prior to packaging and irradiation. | Product is 100% inspected. Device must conform to acceptance criteria documented in inspection procedures.  |
|  Leakage | Finished, irradiated device – identical to current models | Mean leakage per cm graft length per minute: 0.6 ml/cm graft length/min. (SD: 0.4 ml)  |
|  Pressurized Burst Strength | Finished, irradiated device – identical to current models | Mean burst: 381 mmHg (50.8 kPa) Standard Deviation: 59 mmHg (7.9 kPa)  |
|  Determination of Usable Length | The length of each presterile product is measured at a load of zero as a condition of release. | Product is 100% inspected. The measured length must be at least as long as the specified length.  |
|  Pressurized Internal Diameter | Finished product - irradiated device – fabric removed prior to testing making specimen identical to current model. | Mean internal diameter: 6.0 mm (Standard Deviation: 0.29 mm).  |
|  Wall Thickness | Finished product - irradiated device – fabric removed prior to testing making specimen identical to current model | Mean wall thickness: 0.48 mm (Standard deviation: 0.09 mm)  |
|  Suture Retention | Finished product - irradiated device – fabric removed prior to testing making specimen identical to current model | Mean suture pullout: 1.61 lbs. (730g) Standard deviation. 0.33 lbs (150g).  |
|  Kink Radius | Finished product - irradiated device – identical to current model | Kink Radius: 16 mm Standard Deviation: 6.0 mm  |
|  Compliance | Finished product – irradiated – with fabric cover. At physiological pressure the device is not restricted by the fabric cover which has a larger diameter (8.5 mm). Therefore, results of this study are applicable to the device with or without the fabric cover. | Plot prepared to display compliance showed superior compliance to control devices.  |

Summary of Safety and Effectiveness Data
P020049

{5}

Summary of Safety and Effectiveness Data
P020049

## 9.4 ADDITIONAL STUDIES

This device contains no software or electrical components.

## 10. SUMMARY OF CLINICAL STUDIES

### 10.1 OBJECTIVES

This clinical study compared primary and secondary patency rates at one year and the overall incidence of postoperative complications for patients receiving the ProCol® Vascular Bioprosthesis as a vascular access graft to patients receiving standard synthetic vascular access grafts. All patients had at least one prior failed synthetic vascular access graft.

### 10.2 STUDY DESIGN

This multi-center (seven U.S. hospitals), non-randomized clinical study compared the access patency of the ProCol® Vascular Bioprosthesis (n=183) to the results of a concomitant control cohort subsequent to the loss of a previously placed synthetic vascular graft. The concomitant control cohort (n=93) were those subjects in whom vascular access was established with a synthetic access graft during the same implanting period as the treatment arm, by surgeons in the same center but not participating in the trial at the time of implant. Additionally, when the implant and abandonment dates were available for the ProCol® patient’s prior failed graft (n=128), the secondary patency of the prior graft (internal control cohort) was compared to the secondary patency of the ProCol® cohort.

Efficacy was determined by primary and secondary patency of the ProCol® Vascular Bioprosthesis as compared to the concomitant control cohort and the internal control cohort (for secondary patency only). Primary patency was defined as any event that caused a loss of graft patency or required an intervention to the lumen of the graft. Secondary patency was the cumulative graft survival time from graft placement until the graft was considered no longer salvageable and deemed abandoned. Safety was measured by comparing the frequency of anticipated and unanticipated events of the ProCol® Vascular Bioprosthesis to the concomitant control cohort.

### 10.3 EVALUATION OF GENDER BIAS

Study inclusion and exclusion criteria were designed and the study was conducted in a manner to avoid gender bias in the subject population. A primary selection criterion in the study was identification of subjects with a prior failed synthetic access graft.

Page 6
13

{6}

Table 4 displays the percentage of females to males enrolled in the study. Overall, more females were enrolled than males. This finding reflects the higher percentage of women than men having a prosthetic graft in the general population. However, the gender makeup was comparable between ProCol® and the concomitant control cohort.

## 10.4 DEMOGRAPHIC DATA

TABLE 4: PATIENT CHARACTERISTICS

|  Characteristic | ProCol® | Concomitant Control^{1} | p-value^{2}  |
| --- | --- | --- | --- |
|  Female | 59.6% | 59.1% | 1  |
|  Hypertension | 85.8% | 90.2% | 0.342  |
|  Diabetes | 44.8% | 54.3% | 0.160  |
|  Hypercoagulation | 20.2% | 8.7% | 0.015  |
|  African American | 71.3% | 75.0% | 0.221  |
|  Number prior grafts | 2.08 + 1.13 | 1.63 + 0.94 | <0.001  |

1. Concomitant control = commercially available synthetic graft: ePTFE (n=90), silicone (n=1), polyetherurethaneurea (n=2).
2. Bold indicates statistical significance.

## 10.5 DATA ANALYSIS AND RESULTS

The total implant time represented in this study was 188.16 years (ProCol® bioprosthesis) and 91.68 years (concomitant control).

**Patency Rates:** The primary patency was not statistically different between the two cohorts. The secondary patency for the ProCol® Vascular Bioprosthesis was significantly higher per log rank (p=0.0361) than both the concomitant control and internal control cohorts. The Kaplan-Meier patency rates for primary and secondary patency are presented in Table 5.

TABLE 5: KAPLAN-MEIER GRAFT SURVIVAL RATES FOR HEMODIALYSIS ACCESS

|  Study Cohort | Total (n) | Primary Patency |   |   | Secondary Patency  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  6 mo. | 12 mo. | 24 mo. | 6 mo. | 12 mo. | 24 mo.  |   |
|  Concomitant control | Synthetic redo graft | 93 | 44% | 28% | 16% | 70% | 56% | 45%  |
|  Internal control | Study patient's prior synthetic graft | 128 | --- | --- |  | 48% | 34% | 18%  |
|  ProCol® Cohort^{1} | Redo grafts | 183 | 45% | 36% | 22% | 75% | 66% | 60%  |
|  |   |   |   |   |   |   |   |   |
|  DOQI-goal^{2} Both first time and redo arteriovenous access grafts |   | --- | --- | --- |  |  | 70% |   |

1. Secondary patency significant at Log-rank p&lt;0.0361, for all grafts in intention to treat analysis.
2. DOQI-goal stated in National Kidney Foundation-Dialysis Outcomes Quality Initiative Guideline 36.

Summary of Safety and Effectiveness Data
P020049

{7}

Complication Rates: Complications observed in the study are described in Table 6. The linearized rate of moderate events was lower in the ProCol® cohort with a rate of 1.2 events/graft year compared to a rate of 2.0 events/graft year in the concomitant control group. The rates of infection and thrombosis were significantly lower with the ProCol® Vascular Bioprosthesis than with the concomitant control, as was the rate for seroma. The rate for technical complication was significantly higher in the ProCol® cohort. Interventions to restore or maintain patency were performed at a rate of 1.3743/year (concomitant control) compared to 0.9726/year for the ProCol cohort (p=0.0031).

TABLE 6: COMPLICATIONS REQUIRING AN INTERVENTION OR LOSS OF GRAFT PATENCY

|  Complication | ProCol® (n=183 grafts) |   |   | Control (n=93 grafts) |   |   | Cox² P-value  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  # Event | # Grafts with Event | Event Rate¹ (#/graft year) | # Event | # Grafts with Event | Event Rate¹ (#/graft year)  |   |
|  Subject Related  |   |   |   |   |   |   |   |
|  Bleeding | 3 | 3 | 0.0159 | 4 | 2 | 0.0436 | 0.1767  |
|  Steal | 3 | 3 | 0.0159 | 2 | 2 | 0.0218 | 0.6963  |
|  Swelling | 20 | 12 | 0.1063 | 9 | 4 | 0.0982 | 0.8609  |
|  Wound Healing | 7 | 7 | 0.0372 | 0 | 0 | 0.0000 | 0.0327  |
|  Total Subject Related | 3 |  | 0.02 | 6 |  | 0.07 |   |
|  Technique Related  |   |   |   |   |   |   |   |
|  Invasive Radiology | 4 | 4 | 0.0213 | 3 | 3 | 0.0327 | 0.5504  |
|  Dialysis Cannulation Trauma | 19 | 18 | 0.1010 | 8 | 6 | 0.0873 | 0.7456  |
|  Technical Complication³ | 10 | 10 | 0.0531 | 0 | 0 | 0.00 | 0.0086  |
|  Total Technique | 21 |  | 0.14 | 4 |  | 0.05 |   |
|  Graft Involved  |   |   |   |   |   |   |   |
|  High Venous Pressure | 13 | 10 | 0.09 | 6 | 6 | 0.07 | 0.686  |
|  Low Flow | 4 | 4 | 0.03 | 0 | 0 | 0.00 | 0.100  |
|  Dilatation | 6 | 6 | 0.0319 | 0 | 0 | 0.00 | 0.0515  |
|  Infection | 10 | 8 | 0.0531 | 18 | 15 | 0.1963 | 0.006  |
|  Kinking | 3 | 3 | 0.0159 | 1 | 1 | 0.0109 | 0.7958  |
|  Pseudoaneurysm | 3 | 3 | 0.0159 | 5 | 3 | 0.0545 | 0.0814  |
|  Seroma | 0 | 0 | 0.0000 | 6 | 4 | 0.0654 | 0.0004  |
|  Thrombosis | 133 | 93 | 0.7068 | 124 | 59 | 1.3525 | <.0001  |
|  Total Graft Involved | 161 | --- | 1.06 | 148 | --- | 1.71 |   |
|  Total Moderate Events | 221 | 128 | 1.1745 | 180 | 74 | 1.9633 | <0.0001  |
|  Repeat Thrombosis⁴ | 36 | 25 | 0.9726 | 65 | 31 | 1.3743 | 0.0031  |

¹ Event Rate = Events / Graft Year. Total time: ProCol® = 152.53 years; Control = 86.76 years.
² Cox F-Test: compares total number of events. Bold indicates statistical significance.
³ Technical events: Anastomosed to known diseased vessel (3); graft physically compressed (3); sizing (1); implant technique (1); and radiology infiltrated graft (1).
⁴ Repeat Thrombosis: Total Time- ProCol=33.55 years; Control=25.76 years. Events in thrombosis tally.

Summary of Safety and Effectiveness Data
P020049

{8}

TABLE 7: RELATIVE RISK OF EVENTS INVOLVING AN INTERVENTION OR LOSS OF PATENCY IN A HIGH-RISK POPULATION

|  Event | ProCol® Event Rate
#/graft year
(Confidence interval²) |   | Control Event Rate
#/graft year
(Confidence interval) |   | Relative Risk¹
(ProCol® = 1)  |
| --- | --- | --- | --- | --- | --- |
|  Infection | 0.053 | (0.087) | 0.196 | (0.285) | 3.70  |
|  Thrombosis | 0.707 | (0.813) | 1.353 | (1.564) | 1.91  |
|  Total Complications | 1.175 | (1.310) | 1.963 | (2.216) | 1.67  |
|  Total Interventions | 0.973 | (1.097) | 1.374 | (1.588) | 1.41  |
|  Abandonment: Infection | 0.027 | (0.052) | 0.120 | (0.192) | 4.51  |
|  Abandonment: Thrombosis | 0.128 | (0.176) | 0.295 | (0.400) | 2.31  |
|  Graft Abandonment (all) | 0.324 | (0.398) | 0.502 | (0.635) | 1.55  |

¹ Relative risk: the occurrence of events observed in the concomitant control divided by the rate observed in the ProCol® cohort and suggests the higher frequency the event will occur.
² 95% Upper confidence limit for linearized rate.

## 11. CONCLUSIONS DRAWN FROM STUDIES

The data established that the ProCol® Vascular Bioprosthesis is suitable for its intended use for vascular access subsequent to one or more previously failed prosthetic access graft(s). The ProCol® Vascular Bioprosthesis was successfully used for vascular access in a high-risk population. Secondary patency results achieved with ProCol® Vascular Bioprosthesis approached the National Kidney Foundation-Dialysis Outcomes Quality Initiative goal of 70% at 12 months with lower complication and intervention rates than a concurrent concomitant control cohort implanted with synthetic vascular grafts.

## 12. PANEL RECOMMENDATION

In accordance with the provisions of section 515©(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Circulatory System Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel and the FDA.

## 13. CDRH DECISION

The results from the preclinical and clinical data provide reasonable assurance the ProCol® device is safe and effective for patients who already have at least on failed prosthetic access graft. The preclinical data demonstrate that the device is biocompatible, and possesses the mechanical and physical strength necessary for an access indication. The applicant’s manufacturing facility was inspected and found to be in compliance with the Quality System Regulation (21 CFR 820). CDRH issued an approval order on July 29, 2003.

Summary of Safety and Effectiveness Data
P020049

{9}

# 14. APPROVAL SPECIFICATIONS

Directions for use: See the labeling

Postapproval Requirements and Restrictions: See approval order

Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions and Adverse Events in the labeling.

Summary of Safety and Effectiveness Data
P020049
Page 10
17

---

**Source:** [https://fda-staging.innolitics.com/device/P020049](https://fda-staging.innolitics.com/device/P020049)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a PMA](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
