P000028 · Medtronic Sofamor Danek USA, Inc. · MAX · Jun 13, 2002 · Orthopedic
Device Facts
Record ID
P000028
Device Name
AFFINITY (TM) CAGE SYSTEM
Applicant
Medtronic Sofamor Danek USA, Inc.
Product Code
MAX · Orthopedic
Decision Date
Jun 13, 2002
Decision
APRL
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic, Expedited Review
Indications for Use
The AFFINITY™ Anterior Cervical Cage System (hereinafter called the AFFINITY™ Cage) is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. AFFINITY™ implants are to be used with autogenous bone graft and implanted via an open, anterior approach.
Device Story
AFFINITY™ Anterior Cervical Cage System consists of hollow, threaded, tapered titanium alloy (Ti-6Al-4V) implants designed for anterior cervical interbody fusion. Implants feature continuous external screw threads for vertebral endplate purchase and transverse holes to facilitate bony ingrowth. The anterior end is open to receive autologous bone graft harvested from the iliac crest; the posterior end is closed. Used in open, anterior surgical approach by experienced surgeons. Device provides structural support and maintains disc height while fusion occurs. Clinical benefit includes stabilization of the cervical spine to address radiculopathy or myelopathy. Healthcare providers use radiographic imaging to confirm fusion and assess clinical success via NDI scores and neurological status. System includes 14 size variations to accommodate patient anatomy.
Clinical Evidence
Prospective, non-randomized, multi-center clinical study (IDE G960201) comparing AFFINITY™ (n=202) to autogenous bone graft control (n=62). Primary endpoint: overall success (fusion, pain/disability improvement, neurological success, no secondary surgery). At 24 months, AFFINITY™ overall success was 68% (95% CI 60-74%) vs 61% (95% CI 48-75%) for control. Fusion rates were 94% vs 86%. Adverse events included neurological complications, neck/arm pain, and spinal events. Bayesian statistical methods used to adjust for demographic/preoperative covariates.
Technological Characteristics
Manufactured from titanium alloy (Ti-6Al-4V, ASTM F136). Design: hollow, threaded, tapered (8° taper) metal cage. Features: continuous external screw threads, transverse holes for bony ingrowth. Dimensions: 14 sizes (minor diameter 6-12mm, length 12-14mm). Energy source: none (mechanical). Connectivity: none. Sterilization: not specified.
Indications for Use
Indicated for skeletally mature patients (age ≥18) with cervical disc disease (intractable radiculopathy/myelopathy, herniated disc, or osteophyte formation) at one level from C2-C3 to C7-T1. Contraindicated in patients with active infection or allergy to titanium/titanium alloy.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Reference Devices
BAK/Cervical Interbody Fusion System
NOVUS LT
NOVUS LC
Spine-Tech BAK
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## I. GENERAL INFORMATION
| DEVICE GENERIC NAME: | Intervertebral Cervical Cage |
| --- | --- |
| DEVICE TRADE NAME: | AFFINITY™ Anterior Cervical Cage System |
| APPLICANT'S NAME: | Medtronic Sofamor Danek USA
1800 Pyramid Place
Memphis, TN 38132 |
| DATE OF PANEL RECOMMENDATION: | None |
| PREMARKET APPROVAL (PMA) APPLICATION NUMBER: | P000028 |
| DATE OF NOTICE OF APPROVAL TO THE APPLICANT: | June 13, 2002 |
## II. INDICATIONS FOR USE
The AFFINITY™ Anterior Cervical Cage System (hereinafter called the AFFINITY™ Cage) is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. AFFINITY™ implants are to be used with autogenous bone graft and implanted via an open, anterior approach.
## III. CONTRAINDICATIONS
The AFFINITY™ Anterior Cervical Cage System should not be implanted in patients with an active infection or with an allergy to titanium or titanium alloy.
## IV. PRECAUTIONS
The AFFINITY™ Cage System should only be used by surgeons who are experienced in cervical interbody fusion procedures and have undergone adequate training with this device. A lack of adequate experience and/or training may lead to a higher incidence of adverse events, such as neurological complications.
AFFINITY™ Anterior Cervical Cage System SSED
Page 1 of 16
{1}
# V. DEVICE DESCRIPTION
The Affinity™ Cage System consists of hollow, threaded, tapered metal devices implanted into the intervertebral disc space. The implants employ the following design characteristics:
- Implants are manufactured from titanium alloy (Ti-6Al-4V, ASTM F136).
- Implants have a continuous screw thread on the outer surface for purchase into the vertebral endplates. The implants are tapered (8° taper).
- There are circular and elongated transverse holes along the length of the implant to allow for bony ingrowth. During implantation, the cage is rotated so that the two elongated holes on opposite sides of the implant are aligned with the vertebral endplates.
- The anterior end is the wider end and is open to receive autologous bone graft and to mate with the implantation instrument. The posterior end of the implant is closed.
- Autologous bone is taken from the iliac crest and packed into the open (anterior) end of the cage prior to insertion.
- The device is available in the following 14 sizes (minor diameter x length):
Table I – Implant Sizes
| Implant Sizes
(minor diameter x length) | Number of
Implants per
Spinal Level | | |
| --- | --- | --- | --- |
| | | One | Two |
| 6 x 12mm | 6 x 14mm | | x |
| 7 x 12mm | 7 x 14mm | x | x |
| 8 x 12mm | 8 x 14mm | x | x |
| 9 x 12mm | 9 x 14mm | x | |
| 10 x 12mm | 10 x 14mm | x | |
| 11 x 12mm | 11 x 14mm | x | |
| 12 x 12mm | 12 x 14mm | x | |
# VI. ALTERNATIVE PRACTICES AND PROCEDURES
Alternative surgical treatments include, but are not limited to, various bone grafting techniques (e.g., Cloward bone dowels, Smith Robinson tri-cortical wedges, and Keystone grafts) sometimes used in conjunction with anterior/anterolateral spinal systems (e.g., plate and screw systems), or posterior spinal systems (e.g., screw/rod, plate systems, posterior wiring systems). In addition, treatment with the BAK/Cervical Interbody Fusion System is an alternative surgical treatment. Non-fusion surgical techniques, such as posterior decompression may also be utilized.
Nonoperative alternative treatments include, but are not limited to, physical therapy, medication, braces, chiropractic care, bed rest, traction, heat, spinal injections, or exercise programs.
AFFINITY™ Anterior Cervical Cage System SSED
Page 2 of 16
{2}
# VII. MARKETING HISTORY
The AFFINITY™ Cage has a marketing history outside the United States, including Europe, Australia, Japan, and Canada. The AFFINITY™ Cage has not been withdrawn from marketing for any reason relating to the safety or effectiveness of the device.
# VIII. ADVERSE EFFECTS OF THE DEVICE ON HEALTH
An investigational device exemptions (IDE) study of the AFFINITY™ Cage System was performed (G960201). A total of 202 AFFINITY™ device patients and 62 control (single level anterior interbody fusion procedure using autogenous bone graft from the iliac crest) patients were enrolled in a multi-center clinical study. In the AFFINITY™ patient group, the most common adverse events were neck and/or arm pain, trauma, subsequent spinal event, gastrointestinal complication, and neurological event. See Table II for a summary of adverse event rates observed in the clinical study; events are listed in alphabetical order. Table III presents the Bayesian statistical comparison of adverse events between the AFFINITY™ device group and the control treatment group. Table IV summarizes the secondary surgical interventions in the AFFINITY™ device and control treatment groups in the 12-Month and 24-Month post-operative intervals. Table IV also presents the Bayesian statistical comparison of secondary surgeries between the AFFINITY™ device group and the control treatment group.
AFFINITY™ Anterior Cervical Cage System SSED
Page 3 of 16
{3}
| TABLE II - ADVERSE EVENTS | | | | | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Adverse Event | Surgery | | Postoperative (1 day to 1 Month) | | 6 Weeks (1 Month to 2 Months) | | 3 Months (2 Months to 5 Months) | | 6 Months (5 Months to 9 Months) | | 12 Months (9 Months to 19 Months) | | 24 Months (or greater) (19 Months to 48 Months) | | Total Adverse Events | |
| | AFFINITY N=202 | Control N=62 | AFFINITY N=202 | Control N=62 | AFFINITY N=200 | Control N=58 | AFFINITY N=191 | Control N=58 | AFFINITY N=186 | Control N=57 | AFFINITY N=178 | Control N=50 | AFFINITY N=174 | Control N=47 | AFFINITY | Control |
| Anatomical/ Technical Difficulty | 1 | 3 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 3 |
| Cancer | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 | 1 | 0 | 0 | 0 | 2 | 0 |
| Cardio/Vascular | 0 | 0 | 3 | 2 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 3 | 1 | 1 | 5 | 6 |
| Carpal Tunnel Syndrome | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 | 1 | 0 | 5 | 0 | 0 | 0 | 7 | 0 |
| Dural Tear | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 |
| Dysphonia/Dysphagia | 0 | 0 | 5 | 4 | 3 | 0 | 0 | 2 | 0 | 0 | 0 | 0 | 0 | 0 | 8 | 6 |
| Gastrointestinal | 0 | 0 | 7 | 0 | 1 | 1 | 0 | 0 | 0 | 0 | 5 | 1 | 3 | 0 | 16 | 2 |
| Graft Site Related | 0 | 0 | 5 | 2 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 5 | 3 |
| Implant Displacement/ Loosening Collapse | 0 | 0 | 0 | 1 | 0 | 2 | 0 | 3 | 0 | 2 | 0 | 1 | 0 | 0 | 0 | 9 |
| Infection | 0 | 0 | 5 | 0 | 1 | 0 | 0 | 0 | 0 | 1 | 1 | 2 | 0 | 2 | 7 | 5 |
| Malpositioned Implant | 0 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 |
| Neck and/or Arm Pain | 0 | 0 | 2 | 0 | 7 | 0 | 5 | 2 | 5 | 1 | 10 | 2 | 6 | 1 | 35 | 6 |
| Neurological Upper Body2 Lower Body3 | 0 | 0 | 1 | 4 | 3 | 1 | 1 | 3 | 0 | 3 | 5 | 2 | 3 | 4 | 13 | 17 |
| Non-Union4 Non-Union Pending5 | 0 | 0 | 0 | 0 | 0 | 0 | | 0 | 3 | 2 | 2 | 1 | 3 | 2 | 8 | 5 |
| Other Pain6 | 0 | 0 | 1 | 0 | 1 | 0 | 1 | 1 | 2 | 0 | 5 | 1 | 3 | 2 | 13 | 4 |
| Respiratory | 0 | 1 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 2 |
| Spinal Event: Cervical Spine Thoracic Spine Lumbar Spine | 0 | 0 | 0 | 0 | 2 | 0 | 3 | 3 | 2 | 0 | 5 | 3 | 3 | 2 | 15 | 8 |
| Thoracic Spine | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 |
| Lumbar Spine | 0 | 0 | 0 | 0 | 0 | 0 | 5 | 0 | 5 | 1 | 2 | 0 | 1 | 0 | 13 | 1 |
| Subsidence | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 |
| Trauma | 0 | 0 | 3 | 0 | 1 | 0 | 10 | 1 | 6 | 1 | 5 | 0 | 5 | 4 | 30 | 6 |
| Urogenital | 0 | 0 | 2 | 4 | 0 | 0 | 0 | 0 | 2 | 0 | 0 | 1 | 1 | 0 | 5 | 5 |
| Vascular Intra-op | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 |
| Other7 | 0 | 0 | 3 | 2 | 1 | 0 | 0 | 1 | 0 | 1 | 6 | 1 | 6 | 4 | 16 | 9 |
1 Control = Single level anterior interbody fusion procedure using autogenous bone graft from the iliac crest.
2 Neurological adverse events that affected the upper body, i.e., arms, neck, etc.
3 Neurological adverse events that affected the lower body, i.e., legs, feet, etc.
4 Non-union adverse events that have resulted in a second surgery.
5 Non-union adverse events that have not resulted in a second surgery.
6 "Other pain" consists of pain that is not related to the surgery or the treatment area. Examples are bursitis, knee pain, back pain, migraine headaches.
7 The "Other" adverse event category consists of the following adverse events reported in the clinical trial: allergy/rash, allergic reaction to chemotherapy, chemotherapy side effects, cholecystectomy, diabetes, elevated temperature, fibromyalgia, hardware removal, hearing loss and cataracts, hepatomegaly, Horner's Syndrome, joint crepitus, low B12 and folate, malpositioned cervical plate, narcotic addiction, psychological disorder, and toothache.
The most noteworthy adverse events in the AFFINITY™ device group were neurological complications and spinal events. A total of 15 upper and lower body neurological events occurred in 15 patients in the AFFINITY™ device group. These events included: 9 events of tingling and/or numbness in arms or hands either with or without associated pain; 2 cases of new myelopathy; 1 event producing leg numbness symptoms; 1 case of hand cramping; 1 Morton's neuroma of the foot; and 1 median nerve entrapment which was not carpal tunnel.
AFFINITY™ Anterior Cervical Cage System SSED
Page 4 of 16
{4}
A total of 29 spinal events occurred in 27 patients in the AFFINITY™ device group. These events included the following: 6 cervical spondyloses, 4 cases of herniated nucleus pulposus in the cervical spine; 3 cervical degenerative disc disease; 1 cervical arthritis; 1 bone spur; 1 thoracic herniated nucleus pulposus and 13 lumbar associated events, such as degenerative disc disease.
In addition, there were 29 patients in the AFFINITY™ device group who had 35 reports of neck and/or arm pain. Of the 14 events reported between surgery discharge and 6 months postoperatively, 7 involved neck pain including muscle cramps or strains, 6 involved shoulder or arm pain including rotator cuff injuries, and 1 involved hand pain. Of the 10 events occurring between 6 and 12 months postoperatively, 5 involved neck and arm pain, 4 involved shoulder pain including 1 rotator cuff tendonitis, and 1 involved cervical muscle pain and headache. Eleven events occurred at least 12 months after the initial surgery. Of these, 3 involved shoulder pain, 4 involved neck and/or arm pain, 1 involved arm pain associated with fatigue, 1 involved joint pain in neck, shoulders, back, and hands, 1 involved elbow pain, and 1 involved thoracic pain.
In addition to the 35 reports of neck/arm pain, Table II includes 15 patients who reported hand pain. Of the 50 patients reporting neck/arm/hand pain symptoms, 35 of the complaints could be attributed to the operative or adjacent levels. Of the 50 patients complaining of postoperative neck, arm and hand symptoms, 10 were considered neck pain failures and 7 were considered arm pain failures according to the success/failure criteria.
Table III presents the Bayesian statistical comparison of adverse events between the AFFINITY™ device group and the control treatment group.
AFFINITY™ Anterior Cervical Cage System SSED
Page 5 of 16
13
{5}
| Table III - Bayesian Comparison of Adverse Events | | |
| --- | --- | --- |
| Adverse Event | There is a 95% Probability that adverse event rates will fall between the following ranges | |
| | AFFINITY™ Device | Control |
| Anatomical/Technical Difficulty | 0% to 3% | 2% to 13% |
| Cancer | 0% to 4% | 0% to 6% |
| Cardio/Vascular | 2% to 32% | 6% to 52% |
| Carpal Tunnel | 2% to 7% | 0% to 6% |
| Dural Tear | 0% to 3% | 0% to 6% |
| Dysphonia/Dysphagia | 3% to 11% | 2% to 15% |
| Gastrointestinal | 6% to 15% | 1% to 9% |
| Graft Site Related | 2% to 47% | 3% to 54% |
| Implant Collapse/ Displacement/Loosening | 0% to 2% | 8% to 25% |
| Infection | 2% to 10% | 0% to 12% |
| Malpositioned Implant | 0% to 3% | 0% to 6% |
| Neck and/or Arm Pain | 12% to 23% | 3% to 16% |
| Neurological | 4% to 12% | 26% to 53% |
| Non-Union (Outcome Pending) | 1% to 7% | 2% to 21% |
| Other Pain | 3% to 11% | 2% to 19% |
| Respiratory | 0% to 2% | 1% to 11% |
| Spinal Event | 10% to 23% | 7% to 27% |
| Subsidence | 0% to 3% | 0% to 6% |
| Trauma | 5% to 47% | 6% to 35% |
| Urogenital | 1% to 7% | 3% to 23% |
| Vascular Intraop | 0% to 2% | 0% to 9% |
| Other Adverse Event | 5% to 13% | 5% to 24% |
| Any Adverse Event | 46% to 60% | 55% to 81% |
Some of the adverse events led to surgical interventions subsequent to the clinical trial surgery. These surgical interventions can be classified as revisions, removals, supplemental fixations, reoperations, and other (see footnotes below Table IV for an explanation of these terms). Table IV summarizes the secondary surgical interventions in the AFFINITY™ device and control treatment groups in the 12-Month and 24-Month post-operative intervals. Table IV also presents the Bayesian statistical comparison of secondary surgeries between the AFFINITY™ device group and the control treatment group.
AFFINITY™ Anterior Cervical Cage System SSED
Page 6 of 16
{6}
| Table IV - Secondary Surgical Procedures | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Clinical Comparison of Secondary Surgeries | | | | | | | Bayesian Statistical Comparison of Second Surgeries | |
| Type of Secondary Surgical Procedure | Up to 12 Months (1 day to 19 Months) | | 24 Months or Later (19 Months to 48 Months) | | Total Events^{4} | | There is a 95% Probability that Second Surgery rates will fall between the following ranges | |
| Type of Secondary Surgical Procedure^{3} | AFFINITY™
N=178 | Control^{1}
N=50 | AFFINITY™
N=174 | Control
N=47 | AFFINITY™ | Control | AFFINITY™
Cage System | Control |
| Revision | 3 | 5 | 0 | 2 | 3 | 7 | 1% to 31% | 6% to 58% |
| Removal | 2 | 0 | 3 | 0 | 5 | 0 | 1% to 6% | 0% to 6% |
| Supplemental Fixation | 5 | 0 | 1 | 0 | 6 | 0 | 1% to 6% | 0% to 6% |
| Reoperation | 1 | 2 | 1 | 0 | 2 | 2 | 0% to 4% | 1% to 11% |
| Other^{2} | 27 | 5 | 4 | 3 | 31 | 8 | 9% to 19% | 8% to 30% |
1 Control = Single level anterior interbody fusion procedure using autogenous bone graft from the iliac crest.
2 Other Second Surgery is any surgical procedure not classified as a revision, removal, supplemental fixation, or a reoperation such as surgeries for hernias, rotator cuff tears, lumbar adverse events, carpal tunnel syndrome, cervical adverse events that occurred at a different level, etc.
3 Revision: A procedure that adjusts or in any way modifies the original implant configuration.
Removal: A procedure at the involved level that removes one or more components of the original implant configuration without replacement with the same type of trial device.
Supplemental Fixation: A procedure at the involved level in which additional cervical fixation devices that are not approved as part of the protocol are placed.
Reoperation: Any surgical procedure at the involved level that is not classified as a Removal, Revision, or Supplemental Fixation, such as a procedure for wound drainage of the graft site.
Other: Any surgical procedure not classified as a revision, removal, supplemental fixation, or a reoperation, such as surgeries for hernias, rotator cuff tears, lumbar adverse events, carpal tunnel syndrome, cervical adverse events that occurred at a different level, etc.
4 Some patients experienced more than one second surgery.
Of the 47 patients in the AFFINITY™ device group who required a second surgery, 8 had surgery for non-union, 7 had surgery for neck and/or arm pain, 5 had surgery to treat a lumbar condition, 5 required surgery due to trauma, and 3 had surgery to treat carpal tunnel syndrome. Most of the second surgeries occurred up to and including the 12-Month post-operative interval.
IX. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
The following is a list of potential adverse events which may occur with cervical interbody fusion surgery with the AFFINITY™ Cage System. Some of these adverse events have been previously reported in the adverse events table.
- Bending, breakage, loosening, and/or migration of components
- Foreign body (allergic) reaction
AFFINITY™ Anterior Cervical Cage System SSED
Page 7 of 16
{7}
- Tissue or nerve damage
- Post-operative change in spinal curvature, loss of correction, height, and/or reduction
- Infection
- Dural tears
- Neurological system compromise
- Dysphagia/dysphonia
- Scar formation
- Bone fracture
- Non-union (or pseudarthrosis), delayed union, mal-union
- Cessation of any potential growth of the operated portion of the spine. Loss of spinal mobility or function
- Graft donor site complications
- Damage to blood vessels and cardiovascular system compromise
- Gastrointestinal complications
- Damage to internal organs and connective tissue
- Development of respiratory problems
- Incisional complications
- Change in mental status
- Death
Note: Additional surgery may be necessary to correct some of these potential adverse events.
## X. SUMMARY OF PRECLINICAL STUDIES
### Laboratory Studies
Table V summarizes the laboratory studies performed on the AFFINITY™ Cage. A Finite Element Analysis (FEA) was conducted and confirms that, of the
AFFINITY™ Anterior Cervical Cage System SSED
Page 8 of 16
{8}
7-12mm diameter implants, the 7mm device is the worst case. Note: The surgical protocol limits implantation of the 6mm implant to 2 per level.
Table V – Laboratory Studies
| Study | Results/Conclusions |
| --- | --- |
| Static Axial Compressive Yield Load, 7mm implant (n=5)
Samples were loaded at 60mm/min and the ultimate compressive loads were measured. | 34,660 ± 699 N
Therefore, the device should withstand anticipated loads in the cervical spine.
NOTE: Medtronic Sofamor Danek estimated the compressive strength of cervical bone graft constructs based on strengths published in the literature. White’s^{1} research reports the strength of cervical bone constructs as ranging from 2,312 to 3,028 N; Wittenberg’s^{2} research determined the strength to range from 789 to 5,070 N. |
| Compressive Fatigue Strength, 7mm implant (n=10)
Samples were tested at 10Hz until failure or run-out to 5 million cycles. | Implant survived 5 million cycles at 1,730 N.
Therefore, the device should withstand anticipated loads in the cervical spine.
NOTE: According to White and Panjabi^{3}, the maximum compressive load in the cervical intervertebral disc is 74 N. |
| Insertion Torque, 9 x 14mm implant (n=3)
Human cadaver spines were potted from C2-T1 in PMMA and implants were placed at the C4-C5 level. The implant size tested was based on the size which best fit the disc height and vertebral body depth. | 1.15 ± 0.63 Nm
This result compares favorably with the insertion torques of 3 lumbar cages:
• NOVUS LT: 0.95 Nm
• NOVUS LC: 0.87 Nm
• Spine-Tech BAK: 0.74 Nm |
| Push-Out Strength, 7 x 12mm implant (n=5)
Testing was conducted in accordance with ASTM Draft Standard F04.25.02.02. Each cage was placed between two foam blocks with a 100N pre-load. | 347 ± 42 N
This is higher than the push-out strength measured for a 14 x 20mm bone dowel (i.e., 199 ± 50 N) when tested in the same manner. |
## XI. SUMMARY OF CLINICAL STUDIES
### A. Objectives
The objective of the clinical study was to assess the safety and effectiveness of the AFFINITY™ Anterior Cervical Cage System in the treatment of cervical disc disease.
### B. Study Design
The clinical study for the AFFINITY™ Anterior Cervical Cage System compared AFFINITY™ implants to a single-level anterior interbody fusion procedure using autogenous bone graft from the iliac crest. The multi-center, prospective, non-randomized, controlled investigation was designed as an equivalence trial. Treatment patients were enrolled at separate sites than control patients. The effectiveness measures selected for this investigation evaluated whether the implanted disc level was fused, whether there was an improvement in neck pain and disability, whether neurological status was maintained or improved, and whether there were improvements in a patient's general health. Safety information was measured by an analysis of reported adverse events and second surgeries.
AFFINITY™ Anterior Cervical Cage System SSED
Page 9 of 16
{9}
AFFINITY™ Anterior Cervical Cage System SSED
Page 10 of 16
# C. Inclusion and Exclusion Criteria
Patients were enrolled in this study according to the following inclusion/exclusion criteria:
## Inclusion criteria
- Symptomatic cervical disc disease, as defined by intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression which is documented by diagnostic imaging findings
- Single level involvement from C2-C3 disc to C7-T1 disc
- Unresponsive to 6 weeks conservative, nonoperative treatment, or has the presence of progressive symptoms or signs of nerve root or spinal cord compression in face on continued nonoperative management
- Age ≥18 years
## Exclusion criteria
- Previous surgical intervention at the involved level
- Severe osteopenia, osteoporosis, osteomalacia, or metabolic bone disease
- Overt or active spinal and/or systemic infection
- Instability greater than 3.5mm translation, or 11° of angular motion
- Condition that required postoperative medications that interfere with fusion, such as steroids
- Known metal allergy
- Mental incompetence
- Prisoner status
- Pregnancy
- Alcohol or drug abuse
# D. Patient Assessments
Patient follow-up examinations were performed preoperatively, perioperatively and postoperatively at 6 weeks, 3 months, 6 months, 12 months, 24 months, and biennially after the 24-month time point. The effectiveness variables included assessment of fusion, pain/disability status according to the Neck Disability Index (NDI), neck pain, arm pain, neurological status, general health status, disc height status, and overall success.
## Outcomes Assessed and Success Criteria:
- **Fusion** was assessed by independent review of lateral flexion/extension (F/E) and anterior/posterior (A/P) radiographs. **Fusion success** was
{10}
defined as $\leq 4^{\circ}$ of motion on lateral F/E radiographs and no evidence of radiolucency $>2\mathrm{mm}$ covering more than half of either the superior or inferior surface of the implant or graft. Also, patients having secondary surgeries due to nonunions were considered fusion failures.
- **Pain/disability status** were assessed using the Neck Disability Index (NDI). **Pain/disability success** was based on the postoperative NDI score being better than the preoperative score by at least 15 points if the preoperative score were at least 30 points or by at least $50\%$ if the preoperative score were less than 30 points. If the preoperative NDI score were zero, the postoperative score also had to be zero for success.
- **Neck pain** was rated on a 5-point scale. **Neck pain success** was defined as maintenance or improvement in the score.
- **Arm pain** was rated on a 5-point scale. **Arm pain success** was defined as maintenance or improvement in the score.
- **Neurologic status** included an assessment of motor function, sensory, reflexes, and the foraminal compression reproducing pain. **Neurological status success** was defined as maintenance or improvement in the status.
- **General health** was assessed with the Medical Outcomes Study 36-Item Short Form Health Survey (SF-36). **General health success** was defined as maintenance or improvement in the SF-36 subscores, the physical component score (PCS), and mental component score (MCS).
- **Disc height** measurements were taken from radiographs. **Disc height success** was defined as no more than a 2mm decrease in the anterior or posterior post-operative disc heights relative to the pre-operative measurements.
- **Overall success** was defined as a patient demonstrating fusion, a successful pain and disability outcome, neurological success as well as the patient not having a secondary surgery classified as a revision, removal or supplemental fixation.
## E. Demographic Data
A total of 202 patients were entered in the AFFINITY™ device clinical trial. A total of 62 control patients were also entered into the clinical trial from two U.S. studies and two U.K. studies.
Demographic information pertaining to the patients participating in these clinical trials is presented in Table VI.
AFFINITY™ Anterior Cervical Cage System SSED
Page 11 of 16
{11}
| TABLE VI – DEMOGRAPHIC INFORMATION | | |
| --- | --- | --- |
| | AFFINITY™ Device
N=202 | Control
N=62 |
| Age (yr.)
Mean [Range] | 44.5 [27-76] | 50.1 [27-85] |
| Weight (lbs.)
Mean [Range] | 178.6 [97-280] | 172.1 [115-270] |
| Height (in.)
Mean [Range] | 67.5 [59-76] | 67.1 [60-73.5] |
| Sex - Freq. (%)
Male | 105 (52.0%) | 35 (56.5%) |
| Female | 97 (48.0%) | 27 (43.5%) |
| Tobacco used - Freq. (%)
Yes | 73 (36.1%) | 30 (48.4%) |
| No | 129 (63.9%) | 32 (51.6%) |
| Workers Comp. - Freq. (%)
Yes | 44 (21.8%) | 4 (6.5%) |
| No | 158 (78.2%) | 58 (93.5%) |
| Taking Preop. Medication for Pain - Freq. (%)
Yes | 148 (73.3%) | 38 (61.3%) |
| No | 54 (26.7%) | 24 (38.7%) |
| Previous Back Surgery – Freq. (%)
Yes | 14 (6.9%) | 1 (1.6%) |
| No | 188 (93.1%) | 61 (98.4%) |
Comparisons between the AFFINITY™ treatment group and the control group demographics demonstrated some differences that potentially favored the AFFINITY™ patients. Medtronic Sofamor Danek addressed these potential differences in their statistical analyses. All preoperative variables were considered as covariate candidates and the five most important ones, gender, preoperative work status, tobacco use, neurological compression test (F.C.T.) reaction, and whether a patient had preoperative radicular symptoms, were incorporated into covariate analyses of the outcome parameters, thereby adjusting the posterior probabilities in accordance with their influence. Consequently, based on this statistical methodology, most important prognostic differences between the two treatment groups for demographic and preoperative information have been taken into account in assessing the outcome parameters.
## F. Patient Accountability
The database was closed for analysis as of April 5, 2001. The patient accountability data are summarized in Table VII.
AFFINITY™ Anterior Cervical Cage System SSED
Page 12 of 16
{12}
| Table VII –Patient Accountability | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Treatment Group | Preoperative x/n (%) | Surgery/ Discharge x/n (%) | 6 Weeks x/n (%) | 3 Months x/n (%) | 6 Months x/n (%) | 12 Months x/n (%) | 24 Months x/n (%) |
| AFFINITY™ Device | 202/202 (100) | 202/202 (100) | 200/202 (99.0) | 191/202 (94.6) | 186/201 (92.5) | 178/197 (90.4) | 174/192 (90.6) |
| Control | 62/62 (100) | 62/62 (100) | 58/59 (98.3) | 58/59 (98.3) | 57/59 (96.6) | 50/55 (90.9) | 47/50 (94) |
## G. Data Analyses and Results
The results of the clinical study were evaluated using Bayesian statistical methods. All patients involved in the clinical trial of the AFFINITY™ Anterior Cervical Cage System and the control group studies were enrolled under the same inclusion/exclusion criteria. To substantiate the comparability of the two groups, a logistic regression analysis was performed which examined the relationship of all demographic, preoperative medical conditions and preoperative measurements of effectiveness variables on the overall success results. All preoperative variables were considered as covariate candidates and the five most influential ones (gender, preoperative work status, tobacco use, neurological compression test (F.C.T.) reaction, and whether a patient had preoperative radicular symptoms) were incorporated into covariate analyses of the outcome parameters, thereby adjusting the posterior probabilities in accordance with their influence. Consequently, based on this statistical methodology, the most influential prognostic differences between the two treatment groups for demographic and preoperative information were taken into account in assessing the outcome parameters.
A small fraction of the patients did not have their 24-month postoperative evaluations when the results were analyzed. Their 24-month results were predicted from their 12-month outcomes and the relationship established from patients that had both 12 and 24-month evaluations.
## 1. Effectiveness Analysis
As previously stated, the effectiveness analysis included assessment of fusion at the involved level, pain/disability status, neck pain, arm pain, neurological status, general health status, disc height status, and overall success. In some cases, only partial data were available (i.e., not all of the outcome measures were obtained for all patients at all follow-up points). In these cases, all available outcomes were summarized in the analyses. Therefore, the number of patients included in the assessment of the outcomes varies slightly due to missing data. The effectiveness analyses involved the comparison of the AFFINITY™ device group to the control group.
AFFINITY™ Anterior Cervical Cage System SSED
Page 13 of 16
{13}
The adjusted posterior means of success probabilities for the primary effectiveness parameters, including overall success, at 24 months postoperative can be found in Table VIII.
| Table VIII – Posterior Means (95% HPD Credible Intervals) of Success Probabilities for Primary Effectiveness Variables | | |
| --- | --- | --- |
| | 24 Months | |
| | AFFINITY™ Device
Success Rate (Range^{1}) | Control
Success Rate (Range^{1}) |
| Overall Success^{2} | 68% (60% to 74%) | 61% (48% to 75%) |
| Fusion | 94% (63% to 97%) | 86% (68% to 99%) |
| NDI Pain/Disability Improvement | 75% (68% to 81%) | 75% (62% to 87%) |
| Neurological Status
Maintenance or Improvement | 96% (87% to 100%) | 78% (45% to 92%) |
1 There is a 95% probability that success rates will fall between the ranges listed.
2 See Section XI.D. for the definitions of success used in this study.
In Table VIII, neurological success is defined as success in 3 of the 4 subsections (sensory, motor, reflex, and foraminal compression test) as per the protocol. If neurological success were redefined to require successes in 4 of 4 subsections, 13 of 171 AFFINITY™ device patients and 16 of 45 control patients would not be a neurological success. Of the 13 AFFINITY™ patients, there were ten patients with reflex deficits and three with sensory deficits. Eight of these deficits are associated with the operative or adjacent levels.
## 2. Safety Analysis
Safety analyses included all patients regardless of the completeness of their follow-up data or length of follow-up. Table II summarizes all adverse events which occurred in the AFFINITY™ device and control patients. A majority of the adverse event rates and secondary surgery rates were comparable between the groups.
## 3. Effectiveness Analysis—Intent-To-Treat
An "intent-to-treat" analysis of the AFFINITY™ group was also performed. For this analysis, secondary surgery failures, deaths, and missing observations due to other causes resulted in missing observations for the outcome variables and therefore were included in the denominators of the calculated rates, i.e., considered as "failures". By treating these unobserved data as treatment failures, the clinical outcome rates in the intent-to-treat analysis naturally will be lower than the rates reported in the actual observed clinical data. Table IX provides the results for the intent-to-treat analysis.
AFFINITY™ Anterior Cervical Cage System SSED
Page 14 of 16
{14}
| Table IX - Intent-to-Treat Analysis for AFFINITY™ Device Deaths, Secondary Surgery Failures, and Missing Observations Are Considered as Failures and Are Included in the Denominator of the Rates | |
| --- | --- |
| | 24 Month Rates |
| Fusion | 77.7% (157/202) |
| NDI Pain/Disability Improvement | 67.3% (136/202) |
| Neurological Status Maintenance or Improvement | 83.7% (169/202) |
| Overall Success | 60.9% (123/202) |
## XII. CONCLUSIONS DRAWN FROM THE STUDIES
Overall success (i.e., fusion, a successful pain and disability outcome, neurological success as well as the patient not having a secondary surgery classified as a revision, removal or supplemental fixation) was the primary endpoint for the clinical trial and it is the parameter on which the success of the clinical trial is determined. The overall success rate for the AFFINITY™ device group was found to be at least statistically equivalent to the autograft control group rate.
The AFFINITY™ device was found to be at least as safe as the control treatment. A majority of the adverse event rates and secondary surgery rates were comparable between the groups.
The results of the clinical study provide reasonable assurance that the AFFINITY™ Anterior Cervical Cage System is safe and effective for the indicated patient population.
## XIII. PANEL RECOMMENDATION
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Orthopedic and Rehabilitation Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicated information previously reviewed by this panel.
## XIV. CDRH DECISION
FDA issued an approval order on June 13, 2002. The applicant's manufacturing facility was inspected and found to be in compliance with the Quality System Regulation (21 CFR 820).
The PMA for the AFFINITY™ Cage was granted expedited review status on August 31, 2000, because the device potentially represented a clinically meaningful advantage over existing technology.
AFFINITY™ Anterior Cervical Cage System SSED
Page 15 of 16
{15}
# XV. APPROVAL SPECIFICATIONS
Directions for Use: See product labeling.
Hazard to Health from Use of the Device: See Indications, Contraindications, Precautions, and Adverse Events in the labeling.
Post Approval Requirements and Restrictions: See Approval Order.
# XVI. REFERENCES
1. "An Experimental Study of the Immediate Load Bearing Capacity of Three Surgical Constructions for Anterior Spine Fusions", White III, A.A., Jupiter, J., Southwick, W.O., Panjabi, M.M., CORR, No. 91, pp. 21-28, March-April 1973.
2. "Compressive Strength of Autogenous and Allogenous Bone Grafts for Thoracolumbar and Cervical Spine Fusion", Wittenberg, R.H., Moeller, J., Shea, M., and White III, A.A., Spine, Vol. 15, No. 10, pp. 1073-1078. 1990.
3. Clinical Biomechanics of the Spine, 2nd Edition, White III, A.A. and Panjabi, M.M., J.B. Lippincott Company, Philadelphia, 1990.
AFFINITY™ Anterior Cervical Cage System SSED
Page 16 of 16
24
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.