K994384 · Bioject, Inc. · KZE · Jun 22, 2000 · General Hospital
Device Facts
Record ID
K994384
Device Name
CLICKER
Applicant
Bioject, Inc.
Product Code
KZE · General Hospital
Decision Date
Jun 22, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5430
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Clicker™ Needle-Free Self-Injection Device for Personal Use with Saizen® [somatropin (rDNA origin) for injection].
Device Story
Clicker™ is a non-electrically powered, spring-loaded needle-free jet injector designed for subcutaneous administration of Saizen® (somatropin). The device uses a sterile plastic Stem Tip and Clear View Nozzle to deliver medication via high-pressure stream through the skin. Intended for personal use by patients or caregivers. Operation involves winding the device, attaching the vial via a connector, and triggering the spring-driven injection. The device replaces traditional needle-and-syringe delivery, aiming to reduce pain, needle-stick injury risk, and environmental waste. Clinical studies demonstrated bioequivalence in GH and IGF-1 levels compared to conventional needle injection. Users reported higher satisfaction, convenience, and preference for the jet injector, despite initial nervousness in pediatric populations.
Clinical Evidence
Clinical evidence includes a randomized, single-dose, two-way crossover pharmacokinetic/pharmacodynamic study in 21 healthy adults (ages 18-40) and an acceptability study in 29 children (ages 12-18) with Type I diabetes. Pharmacokinetic endpoints (AUC, Cmax, tmax, t1/2) for GH and IGF-1 met bioequivalence criteria (90% CI within 80-125%). Questionnaires showed significant preference for the jet injector over needles regarding pain, ease of use, and convenience. No serious adverse events reported.
Technological Characteristics
Non-electrically powered, spring-actuated jet injector. Components include a sterile plastic Stem Tip and Clear View Nozzle. Dimensions/form factor: handheld. Connectivity: none. Sterilization: not specified. Software: none.
Indications for Use
Indicated for long-term treatment of children with growth failure due to inadequate endogenous growth hormone secretion. Contraindicated in patients who are visually impaired, have neuromuscular/arthritic conditions hindering device operation, cannot understand/follow safety procedures, bruise/bleed easily, take anticoagulants, or are unwilling to comply with device/accessory replacement procedures. Also contraindicated where Saizen® is contraindicated.
Regulatory Classification
Identification
A nonelectrically powered fluid injector is a nonelectrically powered device used by a health care provider to give a hypodermic injection by means of a narrow, high velocity jet of fluid which can penetrate the surface of the skin and deliver the fluid to the body. It may be used for mass inoculations.
Predicate Devices
Medi-Jector Needle-Free Bio-Tropin™ Drug Delivery System (K960285)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the text "K994384 1 of 6" in a handwritten style. The text appears to be a label or identifier, possibly indicating a document or item number. The phrase "1 of 6" suggests that this is the first page or item in a series of six.
510(k) Summary (As required by 21 CFR 807.92(a))
ﻪ
#### A. Submitter Information
Bioject, Inc. 7620 SW Bridgeport Road Portland, Oregon 97224
Phone: 800-683-7221 ext. 424 Fax: 503-624-9002 Email: nancy@bioject.com Contact: Nancy J. Gertlar Manager, QA/RA December 17, 1999 Date:
B Device Information Trade/Proprietary Name:
Common Name:
Classification Name:
Predicate Device(s):
Device Description:
Intended Use:
- C Comparison of Required Technological Characteristics:
Jet Injector, Non-Electrically Powered Fluid Injector
Needle Free Injector, Jet Injector
- . Medi-Jector Needle-Free Bio-Tropin™ Drug Delivery System K960285
- . Vitajet K962625
Clicker™
Clicker™ Needle-Free Self-Injection Device for Personal Use with Saizen® [somatropin (rDNA origin) for injection]. Needle Free Injector, Jet Injector
Clicker™ Needle-Free Self-Injection Device for Personal Use with Saizen® [somatropin (rDNA origin) for injection].
This submission changes the labeling of the Vitajet 3, (Innova) (K962625) to allow the device to be used for needle-free subcutaneous administration of Saizen® [somatropin (rDNAorigin) for injection].
There are no other significant changes to the Vitajet 3 (Innova) in device design or function.
153
{1}------------------------------------------------
#### D Summary and Conclusion of Nonclinical and Clinical Tests:
# GROWTH HORMONE DELIVERY USING A NEEDLE-FREE JET-INJECTOR
A number of clinical and laboratory studies were completed to prepare for an FDA submission for Saizen to be used with a needle-free device. These included laboratory studies designed to address the ability of Saizen to remain intact after administration through the jet-injector and measurement of any Saizen , which may adsorb to the plastic component parts. In addition, procedures were developed to measure and identify any microorganisms that may develop after seven daily injections with the same disposable plastic nozzle and vial connector.
## LABORATORY STUDIES
The goals of the laboratory studies were to evaluate potential shearing and fragmentation of Saizen® and the interaction of growth hormone with the various plastic device components. In addition, a Bioburden study was performed to evaluate microorganism growth in and on the disposable clear nozzle as well as in the vial itself and on the vial connector. Four protocols were completed to reach the above goals.
## SHEAR STRESS TESTING
Clicker" utilizes a spring to induce a high-pressure injection of growth hormone through the skin. Protocol P-00026-01 was conducted to determine the effect of shear stress on Saizen®. A total of two lots of the 5.0 mg vial Saizen® and one lot of 8.8 mg vial Saizen® were tested. Technical report R-00055-02 summarizes the results. Results were within assay variability for high-pressure liquid chromatography (HPLC) analysis, physical tests and pH. Therefore, shear stress caused by Clicker™ did not physically alter the structure of Saizen®.
# CHEMICAL COMPATIBILITY TESTING
### CLEAR VIEW NOZZLE STUDY
Clicker" utilizes a sterile plastic Stem Tip and Clear View Nozzle to perform a highpressure injection of growth hormone through the skin. Protocol P-00028-01 was performed to assess interaction of growth hormone with the plastic components of the needle-free jet-injector. Two lots of the 5.0 mg vial Saizen® and one lot of 8.8 mg vial Saison® were tested. Technical report R-00056-02 summarizes the results. Results were within assay variability for HPLC analysis, physical tests and pH. Therefore, the Clear View Nozzle and Stem Tip are suitable for use with Saizen®
## VIAL CONNECTOR STUDY
Protocol P-00025-01 examined the Vial Connector needed on the vial of growth hormone to facilitate drawing up the material into the injector. The growth hormone
{2}------------------------------------------------
samples were observed at seven and fourteen days for particulates by holding up to the light, checking against black paper and checking against white paper. The samples were also tested for pH and assessed for purity by SE-HPLC. Technical report R-00058-02 summarizes these results. Results demonstrated within- assay variability for HPLC analysis, physical tests and pH. Therefore, the Vial Connector is suitable for use with Saizen®.
### BIOBURDEN STUDY
Protocol P-00027-02 investigates the effect of repeated use of Clicker" on the number and pathogenicity of microorganisms noted on the various component parts. The components (plastic Stem Tip and Clear View Nozzle) can be used for up to seven days, while the reconstituted vial of Saizen® can be used for up to 14 days. This study was conducted to determine the amount and type of microorganism growth on the injector components, in the vial and in the injected drug after 14 days of use.
Multiple drug lots were tested with 12 jet-injectors. Technical report R-00057-01 summarizes the results. The liquid contents injected from the nozzle showed no microorganism growth in any sample, while the nozzle surface itself showed growth in only one sample. The Stem Tip exhibited growth in two samples, but did not show growth on the same, more concentrated filtration samples. One culture grew a nonpathogenic Staphylococcus species and another grew a mold (Aspergillus glaucus) and non-pathogenic bacteria (Bacillus lichenformis). The vial contents exhibited growth on only one sample. This single isolate was a non-pathogenic mold (Aspergillus versicolor). The vial adapter surface showed multiple occurrences of skin contaminants (non- pathogenic Staphylococcus species) only. All recovered isolates were determined not to be Escherichia coli, Pseudomonas aeauginosa, Salmonella or Staphylococcus aureus by screening on selective agar.
In summary, there are no clinically significant microorganisms found within the nozzle or within the Saizen® vial over a period of 14 days. Only minor expected surface contaminants were seen on the outer surface of the Vial Connector, Stem Tip and Clear View nozzle.
#### CLINICAL STUDIES
Clinical studies included a randomized crossover pharmacokinetic, pharmacodynamic study comparing a conventional needle injection to Clicker™. A second study evaluated the acceptability of the needle-free device following a single jet-injection compared to needle injection in 29 children (ages 12-18), with Type I diabetes mellitus. The children with diabetes mellitus were taking daily or more frequent insulin needle injections for a period of one to five years. Two separate clinical studies were completed on 21 adults and 29 children.
# PHARMACOKINETIC AND PHARNACODYNAMIC CYNAMIC STUDIES IN ADULTS
A randomized, single-dose, two-way crossover relative bioavailability study of Saizen® administered subcutaneously by needle or needle-free device in normal
{3}------------------------------------------------
healthy adult subjects (ages 18- 40 years) was completed to determine bioequivalence between the conventional subcutaneous needle injection of GH and GH administered using a needle-free device. Statistical bioequivalence analyses were based on 21 subjects. In the clinical pharmacokinetic study, the measured GH levels from subjects using the needle injection and the needle-free device were similar during the entire 24 hours of blood monitoring. (Figure 1)
Image /page/3/Figure/2 description: The image is a graph titled "Figure 1" that shows the serum concentration of growth hormone over a 24-hour period. The x-axis represents the sampling time in hours, ranging from 0 to 24. The y-axis represents the serum concentration (geometric mean) of growth hormone. Two lines are plotted on the graph, representing two different treatments: a jet injector and a syringe.
The maximum concentration of GH and the peak time for the maximum GH concentration were also not statistically different. The data for AUC, AUCtiast) and Cmax were log transformed prior to analysis. The data for tmax and t1/2 were assessed for adherence to assumptions. For AUC 138, the 90 % confidence interval for the ratio of test to reference expressed, as a percentage was (0.95, 1.12) falling within the (80%, 125%) interval required for bioequivalence as specified in the protocol. For Cmax the 90 % confidence interval for the ratio of test to reference expressed as a percentage was (0.81, 1.20), falling within the (80%, 125%) interval required for bioequivalence as specified in the protocol.
For untransformed tmax the p-values associated with the Shapiro-Wilks tests of normality were not statistically significant for intrasubject error (p = 0.428) and
{4}------------------------------------------------
intersubject error (p = 0.857). The test for formulation differences between the reference formulation and the test formulation twas not statistically significant (p = 0.523). Period effects were also not statistically significant (p = 0.472).
The serum IGF1 values measured every six hours for 24 hours after injection were very similar when comparing both the needle and needle-free device. (Figure 2) After log transforming the data, the p-values for the Shapiro-Wilks test of normality for both the intrasubject and intersubject error were not significant (p = 0.840 for intrasubject error, p = . 0599 for intersubject error).
Image /page/4/Figure/2 description: The image is a line graph titled "Figure 2" that shows the mean serum concentration of IGF-1 over a 26-hour period. The x-axis represents the sampling time in hours, ranging from 0 to 26. The y-axis represents the mean serum concentration of IGF-1, ranging from 100 to 400. Two lines are plotted on the graph, representing two different treatments: Jet Injector and Syringe.
Analog scales were developed to evaluate drug penetration of the skin, bleeding and bruising immediately after injection, thirty minutes after injection and twenty-four hours after injection. Inspections of these count data suggested that there was no statistically significant difference between both treatments (needle vs. needle- free) with respect to the number of subjects injected and events at various time points (0, 30 minutes and 24 hours after injection).
There were no serious adverse events or deaths in this study and no subjects withdrew prematurely from the study. The mild adverse reactions that were reported did not appear to be device related.
The adult questionnaire (ages 18-40 years) indicated a trend toward less pain with the jet-injector (p = 0.093). However, the response to specific questions relating to "ease of use" (p = 0.017), " convenience" (p = 0.004), "pleasantness" (p = 0.011), " less
{5}------------------------------------------------
apprehension" (p = 0.0110 and "overall preference" (p = 0.053) showed a highly significant preference for the needle-free jet-injector over the conventional needle injection.
# QUESTIONNAIRES IN CHILDREN WITH TYPE I DIABETES MELLITUS
Twenty-nine adolescents with one to five years of insulin dependent diabetes mellitus were asked to compare a single, self-administered needle-free dose of saline with the previous six months of needle injections of insulin. Questionnaires were administered shortly after their needle-free injection. The questionnaires in the group of 29 adolescents with Type I diabetes mellitus (ages 12-18 years) demonstrated highly significant differences relating to both "pain" and "unpleasantness" favoring the needle-free jet-injector over needle injections. The subjects chose the needle as the method that would hurt more by a count of 22 to 7. This confirmed the level of pain data and was again highly significant, p < 0.01. The needle-free device was chosen as more pleasant to use by a 22 to 7 ratio (p < 0.01). However, in this population, there was no significant difference between the needle-free device and the needle injections in regard to "convenience" or "ease of use". The adolescent group also found that the first needle-free jet-injection made them significantly more nervous than their familiar needle injection.
#### SUMMARY
In summary, both clinical studies suggest that the needle-free device is easier to use, convenient, more pleasant and generally less painful than conventional needle injection. The adult study demonstrated a clear preference for the needle-free device over needle injection. Measurements of GH and IGF-1 hormone levels demonstrated bioequivalence between needle and needle-free delivery of growth hormone.
The Center for Disease Control in the United States clearly favors the use of needlefree injections to reduce the incidence of Hepatitis B, C, and HIV infection as well as to diminish the pain from needle injection. This new technology is rapidly replacing conventional needle therapy. In recent years, a number of safety features have been devised for the needle and syringe to reduce occupational hazards and associated costs. Needle-free injection is an alternative method that does not require a needle and its wide spread use should reduce the incidence of contagious infections related to needles and at the same time reduce the environmental burden that needle disposal poses on a nation.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, represented by three curved lines that suggest wings and a head.
Public Health Service
# JUN 2 2 2000
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Nancy J. Gertlar Manager, QA/RA Bioject, Incorporated 7620 SW Bridgeport Road Portland, Oregon 97224
Re: K994384 Clicker Trade Name: Regulatory Class: II Product Code: KZE Dated: April 11, 2000 Received: April 12, 2000
Dear Ms. Gertlar:
We have reviewed your Section 510(k) notification of intent to we have reviewed referenced above and we have determined the market the device fally equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate use stated in the encrobate, co as researchmont date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, rood, brug, and occmeblect to the general controls provisions The general controls provisions of the Act of the Act. of the nee? Incents for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{7}------------------------------------------------
## Page 2 - Ms. Gertlar
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fdag.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Page 1 of 1
510(k) Number (if known): new submission
Device Name: Clicker™
Indications for Use:
- This product is indicated for use with Saizen® [somatropin (rDNAorigin) for . injection] for the long-term treatment of children with growth failure due to inadequate secretion of endogenous growth hormone.
Contraindications:
This product is not recommended for patients:
- Who are visually impaired, .
- Who have neuromuscular or arthritic conditions which would make winding the . Clicker™ difficult,
- Who are not able to understand and follow the procedure for safe use of the . device,
- Who bruise or bleed easily, or are taking anti-coagulant medication (blood . thinners), or any other medication or therapy which may contribute to excess bleeding or bruising after injections,
- Who are not willing to fully comply with the procedures of use of the device and . with the recommended frequency for replacement of the disposable accessories,
- Where Saizen® [somatropin (rDNAorigin) for injection] is contraindicated for . treatment of that patient.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Patricia Cucenito
Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
ાર
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.