DOGBONE ACFS

K994383 · Theken Surgical · KWQ · Feb 4, 2000 · Orthopedic

Device Facts

Record IDK994383
Device NameDOGBONE ACFS
ApplicantTheken Surgical
Product CodeKWQ · Orthopedic
Decision DateFeb 4, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Dogbone ACFS is indicated for: The Dogbone ACFS is indicated for trauma, deformity (lordosis and kyphosis). oseudoarthrosis, previously failed cervical soine fusion, tumor, degenerative disk disease (DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.), and spinal canal stenosis. The Dogbone ACFS is indicated for stabilizing the spine from C2 to C7. Unicortical or bicortical fixation from the anterior face of the vertebral bodies may be employed using either the variable angle or fixed angle screws. This product may be employed as an internal fixation device during the time interval required for arthrodesis. The Dogbone ACFS is indicated for stabilizing the spine from C2 to C7. Unicortical or bicortical fixation from the anterior face of the vertebral bodies may be employed using either the variable angle or fixed angle screws. This product may be employed as an internal fixation device during the time interval required for anthrodesis.

Device Story

Dogbone ACFS is a titanium alloy anterior cervical plate fixation system; consists of pre-contoured plates and two styles of screws (fixed angle for rigid constructs; variable angle for non-rigid constructs). Eight locking fingers around each screw hole provide mechanical resistance against screw back-out. Used by surgeons in clinical settings to stabilize cervical spine (C2-C7) during arthrodesis. Provides mechanical support to vertebral bodies via unicortical or bicortical fixation; facilitates spinal fusion by maintaining alignment during healing.

Clinical Evidence

Bench testing only. Static and fatigue cantilever axial compression testing and static torsion testing were performed to characterize stiffness, strength, and fatigue life.

Technological Characteristics

Titanium alloy anterior cervical plate system. Features pre-contoured plates and fixed/variable angle screws. Includes locking fingers for screw retention. Mechanical fixation device; no energy source or software.

Indications for Use

Indicated for patients requiring cervical spine stabilization (C2-C7) due to trauma, deformity (lordosis/kyphosis), pseudoarthrosis, failed fusion, tumor, degenerative disc disease, or spinal canal stenosis. Used as internal fixation during arthrodesis.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ FEB 510(k) Summary # Dogbone ACFS 510 (k) Summary - Company: Theken Surgical 1100 Nola Avenue Barberton, Ohio 44203 Trade Name: Dogbone ACFS Classification: KWG 888.3060, Spinal Intervertebral Body Fixation Orthosis. Class II. - The Dogbone ACFS is a titanium alloy anterior cervical plate fixation system. Description: Plates are pre-contoured in two planes, and come in a variety of lengths. Screws arc available in two diameters, several lengths, and two different styles; fixed angle and variable angle. Fixed angle screws are used to build a rigid fixation construct. Variable angle screws are used to build a non-rigid construct. Hybrid constructs combining elements of fixed and variable angle screws may be built also. Eight locking tingers circumferentially disposed around the rim of each screw hole engage the screw heads to provide mechanical resistance against screw back-out. ## Performance Data: #### Non-clinical: Static and fatigue cantilever axial compression testing was performed. Properties of stiffness, strength, and fatigue life were characterized. Static torsion was also examined. ## Intended Use; The Dogbone ACFS is indicated for: The Dogbone ACFS is indicated for trauma, deformity (lordosis and kyphosis). oseudoarthrosis, previously failed cervical soine fusion, tumor, degenerative disk disease (DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.), and spinal canal stenosis. The Dogbone ACFS is indicated for stabilizing the spine from C2 to C7. Unicortical or bicortical fixation from the anterior face of the vertebral bodies may be employed using either the variable angle or fixed angle screws. This product may be employed as an internal fixation device during the time interval required for arthrodesis. The Dogbone ACFS is indicated for stabilizing the spine from C2 to C7. Unicortical or bicortical fixation from the anterior face of the vertebral bodies may be employed using either the variable angle or fixed angle screws. This product may be employed as an internal fixation device during the time interval required for anthrodesis. ## Substantial Equivalence: Synthes (USA) Titanium Locking Plate System (TILPS) (K970048) {1}------------------------------------------------ Public Health Service FEB 4 2000 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Lukas Eisermann Director, Regulatory Affairs Theken Surgical 1100 Nola Avenue Barberton, Ohio 44203 Re: K994383 Trade Name: Dogbone ACFS Regulatory Class: II Product Code: KWQ Dated: December 27, 1999 Received: December 28, 1999 Dear Mr. Eisermann: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {2}------------------------------------------------ Page 2 - Mr. Lukas Eisermann If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Neil K.P. Ogden James E. Dillard II Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 610/k) Number K994383 510 {k} Number {if known): Device Name: Dogbone ACFS - 1. Indications for Use: The Dogbone ACFS is indicated for trauma, deformity (lordosis and kyphosis), pseudoarthrosis, previously failed cervical spine fusion, tumor, degenerative disk disease (DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.), and spinal canal stenosis. The Dogbone ACFS is indicated for stabilizing the spine from C2 to C7. Unicontical or bicortical fixation from the anterior face of the vertebral bodies may be employed using either the variable angle or fixed angle screws. This product may be employed as an internal fixation device during the time interval required for arthrodesis. #### Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1/2/96) DRO for JZD (Division Sign-Off) Division of General Restorative Devices 994393 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...