← Product Code [HFW](/productcode/HFW) · K994367

# UMBILICAL CORD CLAMP (K994367)

_Medical Sales Co. · HFW · Mar 17, 2000 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K994367

## Device Facts

- **Applicant:** Medical Sales Co.
- **Product Code:** [HFW](/productcode/HFW.md)
- **Decision Date:** Mar 17, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.4530
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Pediatric

## Indications for Use

The device is a sterile, single-use, disposable umbilical cord clamp, used to clamp over the umbilical cord of a newborn at delivery. The device is used to hold the cord securely and prevent blood loss and seepage as the cord dries and shrinks after the birth, prior to and after cutting.

## Device Story

Umbilical cord clamp; injection-molded plastic; blue; single-use; disposable. Used at delivery to clamp newborn umbilical cord; holds cord securely to prevent blood loss/seepage during drying/shrinking process; used prior to and after cutting. Operated by clinicians in delivery settings. Device provides mechanical clamping force.

## Clinical Evidence

Bench testing only. Biocompatibility testing (cytotoxicity, hemolysis) performed by an independent lab with no issues noted.

## Technological Characteristics

Injection-molded plastic; single-use; sterile (100% EtO gas). Mechanical clamping mechanism.

## Regulatory Identification

An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.

## Special Controls

*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.

## Predicate Devices

- [K990737](/device/K990737.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## MEDICAL SALES COMPANY PRODUCT DEVELOPMENT DIVISION

MAR 1 7 2000

## PREMARKET NOTIFICATION

## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

Date Prepared: December 15, 1999

| 01. Submitter (and specification developer): | Medical Sales Company<br>P O Box 540314<br>101 N. 700 West<br>North Salt Lake UT 84054     |
|----------------------------------------------|--------------------------------------------------------------------------------------------|
| 02. Packager:                                | Medical Techniques LC<br>125 N. 400 West, Suite C<br>North Salt Lake UT 84054              |
| 03. Sterilizer:                              | Griffith Micro Sciences -- SLC Div.<br>5725 W. Harold Gatty Rd.<br>Salt Lake City UT 84116 |
| 04. Contact Person:                          | Tom Haslam, Sr., Owner                                                                     |
| Telephone:                                   | 800-292-4056                                                                               |
| 05. Trade Name:                              | Umbilical Cord Clamp                                                                       |
| 06. Common Name:                             | Umbilical Cord Clamp                                                                       |
| 07. Classification Name:                     | Clamp, Umbilical                                                                           |
| 08. Product Code:                            | 85 HFW                                                                                     |
| Device Class:                                | Class II                                                                                   |
| 09. Performance Standards:                   | No performance standards have been officially adopted by the FDA.                          |
| 09. Predicate Device(s):                     | K990737, et al, submissions, information attached, which were determined to be SE          |

510(k): UMBILICAL CLAMP Section 4 - 13063 South Davis Blvd * Bountiful, Utah * Telephone 801-295-6023 * Fax 801-296-5963

{1}------------------------------------------------

<4444567  

142 324

## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

- 10. Description: Umbilical cord clamp, injection-molded plastic, blue, singleuse, disposable, packaged, sterile (sterilized by 100% EtO gas).
- 11. Intended Use: The device is a sterile, single-use, disposable umbilical cord clamp, used to clamp over the umbilical cord of a newborn at delivery. The device is used to hold the cord securely and prevent blood loss and seepage as the cord dries and shrinks after the birth, prior to and after cutting.
- 12. Summary of the Technological Characteristics of This Device Compared to the Predicate(s):

There are no substantive changes in materials, basic configuration and function, method of manufacture or packaging, or sterilization between this device and its predicates.

- 13. Summary of Nonclinical Tests and Results:
The device has been tested by an independent lab for biocompatability (cyctotoxicity, hemolysis) with no problems noted. The clamps will be subjected to QC upon receipt, during/after manufacture, and prior to release.

- 14. Conclusion:
The function and use of this device is no different than that of the predicates, as well as similare devices, marketed since pre-amendment.

Signed: Thomas Pilla Dated: 12-16-99

Thomas J. Haslam, Owner MEDICAL SALES COMPANY

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 7 2000

Medical Sales Co. c/o Mr. John E. Lincoln J.E. Lincoln and Associates P. O. Box 154 Tooele, UT 84074

Re: K994367 Umbilical Cord Clamp Dated: February 17, 2000 Received: February 22, 2000 Regulatory Class: Il

21 CFR §884.4530/Procode: 85 HFW

Dear Mr. Lincoln:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 26, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predication for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitiled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-5597 or at its internet address "http://www.fda.qov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{3}------------------------------------------------

Kag436 510(k) Number (if known):

Umbilical Cord Clamp Device Name:

Indications For Use:

Sterile, single-use, disposable umbilical cord clamp, used to clamp over the umbilical cord of a newborn at delivery. The device is used to hold the cord securely and prevent blood loss and seepage as the cord dries and shrinks after the birth, prior to and after cutting.

(PLRASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGES IF NEWDEND)

Concurrenos of CDRH, Office of Device Evaluation (ODE)

Prescription Use_ XX (Per 21 CFR 801.109)

OR
Over-The-Counter Use

(Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological Devi 510(k) Number

A

(Optional Formul 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K994367](https://fda-staging.innolitics.com/device/K994367)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
