← Product Code [CAF](/productcode/CAF) · K994352

# SIDE DRAFT NEB-U-MIST NEBULIZER, MODEL 41896 (K994352)

_Hudson Respiratory Care, Inc. · CAF · Mar 22, 2000 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K994352

## Device Facts

- **Applicant:** Hudson Respiratory Care, Inc.
- **Product Code:** [CAF](/productcode/CAF.md)
- **Decision Date:** Mar 22, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5630
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Hudson RCI Side Draft Neb-U-Mist® Nebulizer is a non-sterile, single patient use. Disposable Small Volume Nebulizer device designed to be used for intermittent aerosol therapy to deliver hydrating agents or other prescribed inclutions for inhalation. The Hudson RCI Side Draft Neb-U-Mist® Nebulizers intended to treat bronchial spasms, or airway obstruction, associated with diseases such as asthma, chronic bronchitis, and emphysema. The Side Draft Neb-U-Mist Nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. This device is for single patient use and is not intended for reprocessing.

## Device Story

Small volume nebulizer; uses Bernoulli Effect to transform liquid medication into aerosol particles. Gas stream creates low pressure, drawing liquid up capillary tube; liquid contacts gas stream and shatters into 0.5–5.0 micron particles against a baffle. Larger particles settle and return to liquid state for re-nebulization. Device connects to oxygen source (6-8 LPM flow); oxygen enters through side of lid. Used in clinical or home settings for intermittent therapy; operated by patient or clinician. Output is inhaled mist; facilitates delivery of hydrating agents or prescribed medications to airways. Benefits include effective medication delivery for respiratory conditions.

## Clinical Evidence

Bench testing only. Particle size distribution and residual volume testing performed using Albuterol Sulfate, Metaproternol Sulfate, and Tobramycin. Results demonstrate particle sizes in the 0.5μm–5.0μm range, consistent with the predicate device.

## Technological Characteristics

Small volume nebulizer; material: Polypropylene. Principle: Bernoulli Effect/baffle impingement. Energy source: External gas/oxygen supply (6-8 LPM). Form factor: Jar, lid, capillary tube, and baffle. Single-patient use; non-sterile. No software or electronic components.

## Regulatory Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

## Predicate Devices

- Hudson RCI 6350 Home Care Nebulizer ([K930525](/device/K930525.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 994352

# 510(K) SUMMARY

## Side Draft Neb-U-Mist® Nebulizer Section 20

#### 1.0 Date

December 21, 1999

#### 2.0 Submitter

Hudson Respiratory Care, Inc. 27711 Diaz Road Temecula, California 92590

#### Contact Person 3.0

Jeannie Denning Regulatory Affairs Manager

#### Telephone 4.0

(909) 676-5611, ext. 1232

#### 5.0 Proprietary Device Name

Side Draft Neb-U-Mist® Nebulizer

#### 6.0 Classification Name

Nebulizer

#### 7.0 Common Name

Nebulizer

#### 8.0 Predicate Device

Hudson RCI 6350 Home Care Nebulizer, 510(k) number K930525

{1}------------------------------------------------

### 510(K) SUMMARY Side Draft Neb-U-Mist® Nebulizer

Section 20

#### Device Description 9.0

The Hudson RCI Side Draft Neb-U-Mist® Nebulizer employs the Bernoulli Effect in order to transform the solution of liquid medication into tiny particles that can in order to transform the belation on the Bernoulli Effect - the phenomenon be carried in a gad mixture to with increased stream velocity in a fluid – causes the lowering of lateral pressure around the stream of gas to draw liquid up a the lowering of lateral peaches the top of the capillary tube it comes into contact with the stream of gas, it is then shattered into small particles (approximately 3 to 5 microns). These particles are then forced against a baffle (approximately o to o microne) of the particles that will be delivered to the patient that rouses the larger particles. The larger particles settle inside the Nebulizer by removing the largor particles, returning mist to liquid to repeat nebulization process. The smaller particles are then administered as the patient inhales.

#### Intended Use 10.0

The Hudson RCI Side Draft Neb-U-Mist® Nebulizer is a non-sterile, single patient use. Disposable Small Volume Nebulizer device designed to be used for intermittent aerosol therapy to deliver hydrating agents or other prescribed inclutions for inhalation. The Hudson RCI Side Draft Neb-U-Mist® Nebulizers intended to treat bronchial spasms, or airway obstruction, associated with diseases such as asthma, chronic bronchitis, and emphysema.

#### 11.0 Patient Population

The Hudson RCI Side Draft Neb-U-Mist® Nebulizer is intended for use with pediatric and adult patients.

#### 12.0 Comparison of Technological Characteristics

Hudson RCI Side Draft Neb-U-Mist® Nebulizer is substantially equivalent to the predicate device in design and function, (Hudson RCI Cat. No. 6350 Nebulizer.) Particle Size testing of the mist generated during the nebulization process was performed on both devices utilizing commonly used medications. (Albuterol Sulfate, Metaproternol Sulfate and Tobramicyn.) Test results indicate that the Hudson RCI Side Draft Neb-U-Mist® Nebulizer and the predicate device both generate particle sizes in the desired range of 0.5um - 5.0μm. Residual testing for both the predicate and proposed devices demonstrated that the devices are substantially equivalent.

{2}------------------------------------------------

#### 12.1 Materials

The Hudson RCI Side Draft Neb-U-Mist® Nebulizer and the predicate device are made of similar materials of construction which are common to those used in industry for Nebulizer products. The proposed Hudson RCI Side Draft Neb-U-Mist® Nebulizer device is comprised entirely of Polypropylene, while the predicate device utilizes K-Resin for the Jar, Polystyrene for the jar lid, impingement dome and capillary tube.

#### Device Design 12.2

The Hudson RCI predicate and the proposed device have similar designs. Each product consists of small volume jar, lid, capillary tube and dome on which particles are shattered. Both products function at flows of 6-8 LPM of gas, and both devices are designed with flat exterior bottoms to facilitate steady filling of the jar. Both devices have etched markings to reflect the maximum fill level, although the proposed device indicates the level of fluid in incremental markings and the predicate device reflects Maximum fill. Both devices connect to an oxygen source however, the predicate device is designed with an oxygen connection port at the bottom of the jar, while the proposed device has an oxygen connection port on the side of the lid. The gas enters the predicate device through the bottom of the jar where it hits a dome which shatters the particles into therapeutic size. In the case of the proposed device, the Oxygen enters the device through the cap where it meets with medication that has been drawn up the capillary tube. The medication is forced against a baffle on the lid, where the particles are shattered into therapeutic size.

#### 13.0 Conclusion

Based on the information contained in this 510(k) submission, Hudson RCI has determined that the proposed Hudson RCI Side Draft Neb-U-Mist Nebulizer and the predicate Hudson RCI Cat. no. 6350 Home Care Nebulizer are substanially equivalent.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a circular logo with text around the perimeter and a stylized bird-like symbol in the center. The text around the circle reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The bird-like symbol in the center is composed of three curved lines that resemble wings or feathers.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 2 2000

Ms. Jeannie Denning Hudson Respiratory Care, Inc. 27711 Diaz Road P.O. Box 9020 Temecula, CA 92589-9020

Re: K994352 Side Draft Neb-U-Mist Nebulizer Requlatory Class: II (two) Product Code: 73 CAF Dated: December 21, 1999 December 23, 1999 Received:

Dear Ms. Denning:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in In addition, FDA may publish further announcements regulatory action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

### Page 2 - Ms. Jeannie Denning

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Jean A. Watson, Jr.

James E. Dillard III Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if Known): K994352 Device Name: Hudson RCI Side Draft Neb-U-Mist Nebulizer

## Indications for Use

## Indications for Use

The Side Draft Neb-U-Mist Nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. This device is for single patient use and is not intended for reprocessing.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

vascular, Respiratory, 510(k) Number

Prescription Use
Rev. 21 CFR 801.109

OR

Over-the-Counter Use

Por 21 CFR 801.109

(Optional format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K994352](https://fda-staging.innolitics.com/device/K994352)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
