The Implant Sciences 125 Iodine (Model 3500) Brachytherapy Seeds with activities from 0.1 to 5.0 mCi are indicated for temporary or permanent interstitial or intracavity implantation or surface application to treat selected localized tumors. They can be used either as primary treatment (such as for prostate cancer or unresectable tumors) or as treatment for residual disease after excision of primary or recurrent tumors. The 155 lodine Seeds may be used concurrently with or following treatment with other interventions, such as external beam therapy, hyperthermia or chemotherapy. Tumors of the head, neck, lung, pancreas, prostate and other accessible tumors are commonly treated.
Device Story
I-125 brachytherapy seeds (Model 3500) consist of laser-welded titanium capsule containing radioactive iodine source on silica capillary substrate; includes silver radiopaque marker for localization. Seeds implanted interstitially, intracavitarily, or applied to surface by clinicians to deliver localized radiation therapy to tumors. Device provided non-sterile; requires hospital-based sterilization (autoclave or ethylene oxide) prior to use. Radiation emission treats localized tumors (prostate, lung, head/neck, etc.) as primary or adjuvant therapy. Benefit: targeted tumor irradiation while sparing surrounding healthy tissue.
Clinical Evidence
Bench testing only. Performance evaluated against ANSI/HPS N43.6-1997 standards for sealed radioactive sources, including temperature resistance, external pressure, impact, vibration, and puncture resistance. Radiation profile characterization and capsule leak testing performed to confirm equivalence to predicate devices.
Technological Characteristics
Titanium capsule; silica capillary substrate; silver radiopaque marker. Dimensions and materials identical to predicates. Energy source: Iodine-125 radionuclide. Sterilization: Autoclave or ethylene oxide (user-performed). Compliance: ANSI/HPS N43.6-1997.
Indications for Use
Indicated for patients with localized tumors (head, neck, lung, pancreas, prostate, or other accessible sites) requiring temporary or permanent interstitial, intracavity, or surface brachytherapy. Used as primary treatment or for residual disease post-excision; may be used concurrently or sequentially with external beam therapy, hyperthermia, or chemotherapy.
Regulatory Classification
Identification
A radionuclide brachytherapy source is a device that consists of a radionuclide which may be enclosed in a sealed container made of gold, titanium, stainless steel, or platinum and intended for medical purposes to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy.
Special Controls
*Classification.* Class II (special controls). A prostate seeding kit intended for use with a radionuclide brachytherapy source only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Implant Sciences, Inc. I-Plant™ I-125 Brachytherapy Seeds (Model 3000) (K990193)
Medi-Physics, Inc. (Amersham) I-125 Seeds (No. 6711) (K914281)
Submission Summary (Full Text)
{0}------------------------------------------------
K994317
Page 1 of 2
## SUMMARY OF SAFETY AND EFFECTIVENESS
## General Company Information
| Name: | Implant Sciences Corporation |
|------------|--------------------------------------------------|
| Address: | 107 Audubon Road, #5<br>Wakefield, MA 01880-1246 |
| Telephone: | (781) 246 - 0700 |
(781) 246 — 1167
# General Device Information
| Product Name: | Implant Sciences "I-plant" 125 Iodine Brachytherapy Seeds (Model 3500) |
|-----------------|------------------------------------------------------------------------|
| Classification: | Brachytherapy Radionuclide Source, 21 CFR 892.5730 - Class II |
## Predicate Devices
Fax:
lmplant Sciences, Inc. I-Plant™ I-125 Brachytherapy Seeds (Model 3000) [510(k) K990193]
Medi-Physics, Inc. (Amersham) I-125 Seeds (No. 6711) [510(k) K914281]
## Description
The Implant Sciences 125 lodine (Model 3500) Brachytherapy Seeds consist of a laser welded titanium capsule containing a silica capillary tube that serves as a substrate for the radioactive iodine source. The tube is positioned around a silver radiopaque x-ray marker that identifies the source location and orientation. The seeds are provided non-sterile and must be sterilized prior to use.
## Intended Use
The Implant Sciences 125 Iodine (Model 3500) Brachytherapy Seeds with activities from 0.1 to 5.0 mCi are indicated for temporary or permanent interstitial or intracavity implantation or surface application to treat selected localized tumors. They can be used either as primary treatment (such as for prostate cancer or unresectable tumors) or as treatment for residual disease after excision of primary or recurrent tumors. The 155 lodine Seeds may be used concurrently with or following treatment with other interventions, such as external beam therapy, hyperthermia or chemotherapy. Tumors of the head, neck, lung, pancreas, prostate and other accessible tumors are commonly treated.
{1}------------------------------------------------
K994317
Page 2 of 2
## Technological Characteristics Versus Predicate Devices
The Implant Sciences 155 lodine Brachytherapy Seeds use the same titanium capsule material and are the same physical dimensions (length and diameter) as the predicate devices. The range of apparent activity levels is similar to that of the predicate devices. The results of safety testing in accordance with the ANSI Standard for Sealed Radioactive Sources (ANSI/HPS N43.6 - 1997) are consistent with the test results for the predicate devices.
### Substantial Equivalence
This 510(k) Notice supports the position that the Implant Sciences 155 lodine (Model 3500) Brachytherapy Seeds are substantially equivalent in design and function to the Implant Sciences (Model 3000) 125 lodine Brachytherapy Seed [510(k) K990193] and the Medi-Physics, Inc. (Amersham) I-125 Seeds (No. 6711) [510(k) K914281], which are brachytherapy radionuclides that have previously been cleared for marketing under the Premarket Notification requlations. These devices have been placed in the same classification category (21 CFR 892.5730) as the ISC Model 3500 seeds; and each is indicated for the same clinical application.
The 510(k) Notice contains summaries of in vitro studies which were conducted to evaluate the safety, and appropriateness of the ISC Model 3500 Brachytherapy Seeds. Data are presented which demonstrate that the Model 3500 seeds satisfy performance requirements for temperature resistance, external pressure resistance, impact resistance, vibration integrity, and puncture resistance as specified by ANSI Standard N43.6-1997. These test results confirm that the ISC 125 lodine (Model 3500) Brachytherapy Seeds meet the requirements for radioactive sources in its class and are equivalent to other currently marketed radionuclide brachytherapy sources. The 510(k) Notice also describes further standardized studies which describe the radiation profile of the sources. These profiles are consistent with the data generated from other lodine-125 brachytherapy seeds with similar physical geometries. In addition, capsule leak testing was carried out in accordance with ANSI standards. The results of these studies also support the equivalence of the ISC Model 3500 device to the predicate devices.
Both the Model 3500 device and the predicate devices are provided non-sterile and must be sterilized by autoclave or ethylene oxide at the hospital prior to use. The manufacturing and cleaning processes for the seeds introduce no new chemical or biological entities to the surface of the device.
Implant Sciences believes that the information provided establishes that similar predicate devices have been used historically for the same types of clinical applications as the Implant Sciences 145 lodine (Model 3500) Brachytherapy Seeds. The materials from which the Implant Sciences device is fabricated have an established history of use in medical applications, and the specific materials used by Implant Sciences have been thoroughly tested in accordance with applicable guidelines.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 1 2000
Howard L. Schrayer Implant Sciences Corp. 107 Audubon Road #5 Wakefield, MA 01880-1246 Re:
K994317 I-plant Model 3500 (125-Iodine Brachytherapy Seeds) December 21, 1999 Dated: Received: December 22, 1999 Regulatory class: II 21 CFR 892.5730/Procode: 90 KXK
Dear Mr. Schrayer:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce pror to May 26, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalion assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notfication submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classfication for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours,
Daniel G. Schultz, M.D Captain. USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{3}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
K994317
Device Name: Implant Sciences 125 lodine Brachytherapy Seeds - (Model 3500)
Indications For Use:
The Implant Sciences 125 lodine Brachytherapy Seeds with activities from 0.1 to 5.0 mCi are indicated for temporary or permanent interstitial or intracavity implantation or sufface application to treat selected localized tumors. They can be used either as primary treatment (such as for prostate cancer or unresectable tumors) or as treatment for residual treatment (each as is is in the 135 lodine Seeds may be 135 lodine Seeds may be used concurrently with or following treatment with other interventions, such as external beam therapy, hyperthermia or chemotherapy. Tumors of the head, neck, lung, pancreas, prostate and other accessible tumors are commonly treated.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
**Over-The-Counter Use**
David C. Bergman
(Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological Devic 510(k) Number
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.