Arabesk Flow® is a flowable, light-curing and radiopaque glass ceramic microhybrid composite. It is suitable for restorations in both anterior and posterior applications. Arabesk® Flow is intended for: - fillings requiring minimally invasive preparation - fillings for small cavities - extended fissure sealing - substitute fillings of cavities with undercuts - Class III-V fillings including V-shaped defects and cervical caries - repair of fillings and veneers - luting of translucent prosthetic pieces (e.g., procelain-only crowns)
Device Story
Arabesk Flow is a light-curing, radiopaque glass ceramic microhybrid composite dental filling material. Composition includes 69% inorganic fillers (silica, barium/strontium borosilicate glass, 0.7um average size) and 30% dimethacrylates (BIS-GMA, urethane dimethacrylate, triethyleneglycol dimethacrylate). Supplied in bulk syringes or single-use applicator-tip capsules. Used by dentists in clinical settings for restorative procedures. Material applied to tooth cavity; low viscosity ensures wetting and adhesion; polymerized using halogen light. Provides high-strength, color-stable, polishable restorations. Benefits include improved cavity adaptation, reduced air entrapment, and efficient filling process.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing of performance properties including viscosity, compressive strength, polymerization shrinkage, color stability, and resistance to abrasion compared to predicate devices.
Indicated for dental restorations in anterior and posterior teeth, including minimally invasive preparations, small cavities, extended fissure sealing, cavity undercuts, Class III-V fillings, cervical caries, repair of fillings/veneers, and luting of translucent prosthetic pieces.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
7 2000 FEB
Image /page/0/Picture/1 description: The image shows a logo with the word "VOCO" in large, stylized letters above the number "K994304". The letters of "VOCO" are designed with geometric shapes, with the "V" resembling an inverted triangle, the "O" as a solid circle, the "C" as a partial circle, and the second "O" as another solid circle. The number "K994304" is written in a handwritten style, and there is a horizontal line below the number.
VOCO GmbH · P.O.B. / Postfach 767 · D-27457 Cuxhaven (Germany)
Anton-Flettner-Straße 1-3 D-27472 Cuxhaven (Germany)
Telefon: (0 47 21) 7 19 - 0 intl.: +49 (47 21) 7 19-0
Telefax: (0 47 21) 7 19 - 1 09 intl.: +49 (47 21) 7 19-1 40
Internet: http://www.voco.de e-mail: info@voco.de
Ihr Zeichen yr. ref.
ﻟﺴﻴ
Ihre Nachricht vom dtd.
Unser Zeicher our ref.
ﻟــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
Datum date
# 510(k) SUMMARY
### ARABESK® FLOW (76 EBF)
- 1. SUBMITTER'S NAME
- 2. CONTACT PERSON for VOCO GmbH
- 3. DATE THAT 510(k) SUMMARY WAS PREPARED
- 4. NAME OF THE MEDICAL DEVICE (Classification / Common / Proprietary)
- 5. LEGALLY MARKETED DEVICES TO WHICH SUBSTANTIAL EQUIVALENCE IS CLAIMED
- 6. DESCRIPTION OF THE DEVICE
- 7. INTENDED USE OF THE DEVICE
- 8. TECHNOLOGICAL COMPARISON BETWEEN SUBJECT AND PREDICATE DEVICES
- SUMMARY OF PRECLINICAL PERFORMANCE STUDIES AND CONCLUSIONS FROM 9. PRECLINICAL PERFORMANCE STUDIES
SUBMITTER'S NAME VOCO GmbH Anton-Flettner-Str. 1-3 27472 Cuxhaven GERMANY 011-49 47 21 719 0 Tel: Fax: 011-49 47 21 719 140
{1}------------------------------------------------
القرآن
な ﻬ
page 2 of 4
U.S. REGULATORY CONTACT PERSON FOR VOCO GmbH
Evan Dick, Ph.D. E.G. Dick & Associates 7527 Westmoreland Avenue St. Louis, MO 63105
Tel: (314) 721-0112 (314) 721-7591 Fax:
DATE THAT 510/k) SUMMARY WAS PREPARED
December 21, 1999
| <b>4. NAME OF THE MEDICAL DEVICE</b> | |
|--------------------------------------|------------------------------------------------------|
| Classification name | Material, tooth shade, resin<br>(Dental 76 EBF) |
| Common / usual name | Flowable, light-curing composite<br>filling material |
| Proprietary names | ARABESK® FLOW |
LEGALLY MARKETED DEVICES TO WHICH SUBSTANTIAL EQUIVALENCE IS CLAIMED
- Tetric Flow (K964285, Ivoclar)
- · Flow Line (K990756, Kulzer)
{2}------------------------------------------------
#### DESCRIPTION OF THE DEVICE
Arabesk Flow is a light-curing and radiopaque glass ceramic microhybrid composite. It is suited for fillings in both anterior and posterior applications. Arabesk Flow's low viscosity results in excellent wetting and adhesion to the tooth. Arabesk Flow produces a high strength and color stable restoration that can be polished to a high gloss. Arabesk Flow is polymerized (cured) by halogen light. Arabesk Flow conforms well to cavity surfaces, helps to avoid trapped air, and speeds cavity filling.
Arabesk Flow contains 69% inorganic fillers (highly pure silica and barium/strontium borosilicate glass particles - average size 0.7um) and 30% dimethacrylates (BIS-GMA, urethane dimethacrylate, triethyleneglycol dimethacryrlate).
Arabesk Flow is available in bulk dispensing syringes (3gm each) and in single-use applicator-tip capsules (0.25 gm each).
INTENDED USE OF THE DEVICE
Arabesk Flow is a composite filling material for both anterior and posterior applications, as well as for inlay techniques. Indications for use are:
- · Fillings requiring minimally invasive preparation
- · Fillings for small cavities
- · Extended fissure sealing
- · Substitute fillings of cavities with undercuts
- · Class III-V fillings including V-shaped defects and cervical caries
- · Repair of fillings and veneers
- Luting of translucent prosthetic pieces (e.g., procelain-only crowns)
{3}------------------------------------------------
#### TECHNOLOGICAL COMPARISON BETWEEN SUBJECT AND PREDICATE DEVICES
Arabesk Flow, Tetric Flow (K964285), and Flow Line (K990756) are all flowable, light-curing, glass ceramic composite filling materials for use in both anterior and posterior sites.
Arabesk Flow, Tetric Flow, and Flow Line are all composed of inorganic silicates (~60-70%, w/w/), dimethacrylates (~30-40%, w/w/), photoinitiators and pigments (~1%. w/w). The subject and predicate devices are substantially alike with respect to such performance properties as viscosity, compressive strength, polymerization shrinkage, color stability, and resistance to abrasion.
Arabesk Flow is formulated from chemical components that are commonly associated with currently marketed dental composite materials.
The chemistry of Arabesk Flow raises no new issues or questions that effect safety, effectiveness, or biocompatibility for a dental composite product.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is an emblem featuring three stylized human figures or profiles, possibly representing health, family, and community.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
7 2000 FEB
Evan G. Dick, Ph.D. E.G. Dick & Associates c/o 7527 Westmorland Avenue St. Louis, Missouri 63105
Re: K994304 Material, tooth shade, resin (Dental 76 EBF) Trade Name: Requlatory Class: II Product Code: EBF December 21, 1999 Dated: Received: December 21, 1999
Dear Mr. Dick:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any
{5}------------------------------------------------
Page 2 - Mr. Dick
obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director
Directfor Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K994304
Arabesk® Flow Device Name:
Indications For Use:
Arabesk Flow® is a flowable, light-curing and radiopaque glass ceramic microhybrid composite. It is suitable for restorations in both anterior and posterior applications. Arabesk® Flow is intended for:
- · fillings requiring minimally invasive preparation
- · fillings for small cavities
- · extended fissure sealing
- · substitute fillings of cavities with undercuts
- Class III-V fillings including V-shaped defects and cervical caries
- · repair of fillings and veneers
- · luting of translucent prosthetic pieces (e.g., procelain-only crowns)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (per 21 CFR 801.109) | <div> <span></span> </div> |
|---------------------------------------|----------------------------|
|---------------------------------------|----------------------------|
OR
| Over-The-Counter Use | <div> <span></span> </div> |
|----------------------|----------------------------|
|----------------------|----------------------------|
| (Division Sign-Off) | <div> <span></span> </div> |
|---------------------------------------------------------------------|----------------------------|
| Division of Dental, Infection Control, and General Hospital Devices | <div> <span></span> </div> |
| 510(k) Number | <div> <span></span> </div> |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.