K994279 · Pure Water Solutions, Inc. · FIP · Dec 22, 2000 · Gastroenterology, Urology
Device Facts
Record ID
K994279
Device Name
PURE WATER-CDS (CENTRAL DIALYSIS SOLUTION)
Applicant
Pure Water Solutions, Inc.
Product Code
FIP · Gastroenterology, Urology
Decision Date
Dec 22, 2000
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 876.5665
Device Class
Class 2
Indications for Use
This device is intended to remove organic and inorganic substances and microbial contaminants from water that is used to dilute dialysis concentrate to form dialysate, and to produce purified water for dialyzer reprocessing and equipment rinse and disinfection.
Device Story
PURE WATER-CDS is a water purification system for hemodialysis clinics. It processes influent water through multi-media filtration, water softening, and granular activated carbon to remove suspended solids, hardness, chlorine, and chloramines. The purifier phase utilizes reverse osmosis (RO) and/or portable-exchange deionization (DI). Post-treatment circulation includes storage tanks, stainless steel repressurization pumps, ultraviolet sterilization, and absolute final filters (0.05-0.2 micron). Auxiliary components include booster pumps, blending valves, and remote alarms. The system is operated and maintained by clinic technicians. Output is purified water meeting AAMI standards for hemodialysis. The system ensures water quality via temperature-compensated conductivity monitoring. Clean-in-place functionality uses paracetic acid for RO membranes and ozone for the circulation phase. The device benefits patients by providing high-quality, contaminant-free water for dialysis procedures.
Clinical Evidence
bench testing only
Technological Characteristics
Materials: stainless steel, polypropylene, FDA 21 CFR 173.25 compliant resins. Components: multi-media filters, water softeners, granular activated carbon, RO units, DI exchange, UV sterilization, absolute filters (0.05-0.2 micron). Monitoring: temperature-compensated conductivity instrument. Disinfection: paracetic acid (RO), ozone (circulation). System is modular, assembled from existing cleared technologies.
Indications for Use
Indicated for use in hemodialysis clinics to provide purified water for diluting dialysis concentrate, dialyzer reprocessing, and equipment rinsing/disinfection by removing organic, inorganic, and microbial contaminants.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
DEC 22 2000
## 510 (k) Summary
| Submitter: | PURE WATER Solutions, Inc.<br>14550 East Easter Avenue, Suite 800<br>Englewood, Colorado 80112<br>Phone 303/693-7610, Fax 303/693-4479<br>Contact Name: Dean S. Lewis | |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Classification Name: | Water Purification System for Hemodialysis | |
| Device Name: | PURE WATER-CDS (Central Dialysis Solution) | |
| Predicate Device: | Mar Cor Services, Inc. #K945559 | |
The PURE WATER Solutions' Water Purification System for Description of System: Hemodialysis is designed to provide water of the required quality to meet AAMI standards for riohodialysto is actigate will remove organic, inorganic and microbial contaminants from the influent water supplied to Hemodialysis clinics. This system comes in various sizes to provide the proper volume of water for each individual clinic. It is easily used and maintained by clinic technicians.
The water treatment phase or pretreatment includes multi-media filtration to reduce suspended solids, water softening to reduce hardness and granular activated carbon filters to remove chlorine and chloramines. If finer filtration is required, cartridges with appropriate housings may be used.
The purifier phase includes reverse osmosis (RO) and or portable-exchange deionization (DI). The primary method will be reverse osmosis using an existing FDA cleared device. DI may be selected for use either independently or in conjunction with RO depending upon the influent water analysis, the available space and the demand required. If DI is selected, then 0.1 micron, absolute, final filters will be installed downstream of the DI and a temperaturecompensated, conductivity instrument will be used to monitor the quality.
The circulation phase or post treatment may consist of an RO storage tank with an appropriate 0.2 micron, absolute, vent filter, a stainless steel repressurization pump, ultraviolet sterilization, and 0.05 micron, absolute, final filter.
Auxiliary devices such as booster pumps, temperature blending valves, pressure gauges, sample ports, fittings, piping, level control switches and remote alarms are used as needed throughout the system.
000000
{1}------------------------------------------------
K994279
p. 2/2
A clean-in-place system consisting of an inert plastic or stainless steel pump and a solution tank is used to clean RO membranes. A paracetic acid disinfection product may be used to chemically disinfect the RO while ozone is used to disinfect the circulation phase.
All components used in the system are components with Safety and Effectiveness: existing safety records. Materials of construction consist of stainless steel, polypropylene, existing survey records. Water treatment resins are standard resins used in potable and high poryonifications and conform to FDA standards 21 CFR 173.25 and filters are validated for low media migration and with demonstrated microbial retentiveness. All of these materials 10 i meet and do not contribute contaminants to treated water systems. The PURE WATERare mere and is assembled using existing technologies, an RO unit with existing FDA marketing clearance and personnel with certification to Water Quality Association, Lisle, Illinois, standards of system design, maintenance and treatment.
Conclusion: PURE WATER Solutions, Inc. has designed a water treatment system for Hemodialysis. We believe this system is equivalent to an FDA approved system currently on the market, namely, the predicate device from Mar Cor Service, Inc. 510(k) #K945559. The the mailed, amazerials that are safe and effective and methods of design, construction and installation sanctioned by current standards in industry. We therefore request that PURE WATER Solutions' PURE WATER-CDS be granted substantial equivalence.
000007
Page 2
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES * USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread, facing to the right.
## Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 22 2000
Mr. Dean S. Lewis President Pure Water Solutions, Inc. 14550 East Easter Avenue Suite 800 ENGLEWOOD CO 80112 Re: K994279
Pure Water - CDS (Central Dialysis Solution) Dated: October 1, 2000 Received: October 6, 2000 Regulatory Class: II 21 CFR §876.5665/Procode: 78 FIP
Dear Mr. Lewis:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21.CFR Part 820) and that, through periodic QS inspections, the Food and Drug, Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours,
Daniel G. Schultz, M.D.
Captain, USPHS Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure (s)
{3}------------------------------------------------
| Page | of |
|------|----|
|------|----|
K994279 510(k) Number
PURE WATER-CDS Device Name
Indications For Use:
This device is intended to remove organic and inorganic substances and microbial contaminants from water that is used to dilute dialysis concentrate to form dialysate, and to produce purified water for dialyzer reprocessing and equipment rinse and disinfection.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Sezman
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Device 510(k) Number
(Optional Format 3-10-98)
Prescription Use (Per 21 CFR 801.109)
000452
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.