K994275 · Sims Portex, Inc. · FRN · Feb 22, 2000 · General Hospital
Device Facts
Record ID
K994275
Device Name
SIMS PORTEX ANESTHESIA CATHETER
Applicant
Sims Portex, Inc.
Product Code
FRN · General Hospital
Decision Date
Feb 22, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SIMS Portex Anesthesia Catheter is indicated for the infusion of local anesthetics or narcotics into the intraoperative site for post operative pain management and for regional anesthesia, outside of the epidural space. Roures of administration may include intraoperative or percutaneous. The catheter is not intended for intravenous or intramuscular use.
Device Story
Flexible nylon catheter; closed-end with lateral eyes or open-ended with finished tip; includes depth markings (1-20 cm). Used for infusion of anesthetics, analgesics, or narcotics into intraoperative sites for pain management. Operated by clinicians; may be used with infusion pumps. Provides delivery path for medication; facilitates post-operative pain control. Not for IV or IM use. Provided sterile; may include stylet, catheter connector, or insertion device.
Clinical Evidence
No clinical data. Bench testing only: dimensional characteristics, flow rate, leakage, hub/catheter detachment, tensile strength, elongation, ETO residuals, and biological safety per ISO 10993.
Technological Characteristics
Flexible nylon tubing; 21g (0.033" OD / 0.019" ID); 38-inch nominal length. Features depth markings. Sterile. Evaluated per ISO 10993 for biological safety.
Indications for Use
Indicated for infusion of local anesthetics or narcotics into intraoperative sites for post-operative pain management and regional anesthesia outside the epidural space. Routes include intraoperative or percutaneous. Contraindicated for intravenous or intramuscular use.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
B. Braun Medical Inc., PERIFIX® Epidural Catheter Set, EC20-C (K813186)
Submission Summary (Full Text)
{0}------------------------------------------------
K994275
Image /page/0/Picture/1 description: The image shows the logo for Smiths Industries Medical Systems. The logo is composed of two lines of text. The first line reads "SMITHS INDUSTRIES" in a bold, sans-serif font. The second line reads "Medical Systems" in a serif font.
FEB 2 2 2000
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
#### 510(K) SUMMARY:
### COMPANY INFORMATION:
SIMS Portex Inc 10 Bowman Drive Keene, NH 03431 (603) 352-3812 Contact: Timothy J. Talcott Director of Regulatory Compliance
#### PREPARATION DATE OF SUMMARY:
February 16, 2000
## TRADE NAME:
SIMS Portex Anesthesia Catheter
#### COMMON NAME:
Anesthesia Conduction Catheter
### PRODUCT CLASS/CLASSIFICATION:
Class II, 79 MEB, accessory to 21 CFR 880.5725 (infusion pump)
#### SIMS Portex Inc.
10 Bowman Drive PO Box 0724 Keene NH 03431 USA Telephone: 603-352-3812 Fax: 603-352-3703
{1}------------------------------------------------
#### PREDICATE DEVICE(S):
SIMS Portex Inc., Epidural Catheter; 4910-18, 21g nylon, closed-end, three-eyed This device is marketed under 510(k), K992471.
- B. Braun Medical Inc., PERIFIX® Epidural Catheter Set, EC20-C This device is marketed under 510(k), K813186 and the indications for use are as specified in I-Flow Corporation's 510(k)s, K980558 and K991543.
#### DESCRIPTION:
The SIMS Portex Anesthesia Catheter is made of flexible, nylon tubing. The catheter may be closed-ended with lateral eyes or an open-ended catheter with finished tip. The tip of the catheter is marked. The catheter has a single mark at 5 cm from the tip with 1 cm increments, up to 20 cm. The 10 cm mark is indicated by two marks, 15 cm by three marks, and 20 cm by four marks.
The catheter is available as 21g (O.D. .033"/I.D. .019"). The catheters have a nominal length of 38 inches. The catheters may include a stylete.
The catheters are provided with a catheter connector (K965017) to provide a means of administration of anesthetics, analgesics, and/or narcotics. The catheter may be used in conjunction with an infusion pump. The catheter may be used, outside of the epidural space, for up to 30 days. An insertion device may be provided to aid the placement of the catheter into the introducer needle. They are provided sterile in individual packages or as a component of a regional anesthesia procedure tray (K965017).
#### INDICATIONS FOR USE:
The SIMS Portex Anesthesia Catheter is indicated for the infusion of local anesthetics or narcotics into the intraoperative site for post operative pain management and for regional anesthesia, outside of the epidural space. Roures of administration may include intraoperative or percutaneous. The catheter is not intended for intravenous or intramuscular use.
### TECHNICAL CHARACTERISTICS:
The design of the catheter and materials are identical to our current catherer. The catheter has been evaluated and demonstrated equivalency to the predicate devices for the following dimensional and functional characteristics: ID, OD, length, flow rate, leakage, eye patency, tensile strength, percentage elongation, luer taper, and security of connection of the catheter to the catheter connector.
{2}------------------------------------------------
#### NON-CLINICAL DATA:
Data submitted demonstrates that the catheter performs equivalently to the predicate devices. Data submitted covers; dimensional characteristics, flow rate, leakage, hub/catheter detachment, tensile strength, elongation, ETO residuals, and biological safety per ISO 10993.
## CLINICAL DATA:
Not applicable
### CONCLUSION:
The comparison to the predicate devices demonstrate that the proposed device is safe and effective and is substantially equivalent to the predicate devices.
Very truly yours,
SIMS PORTEX INC.
Timothy J. Fallon
Timothy J. Talcott Director of Regulatory Compliance
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines suggesting movement or connection.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## FEB 2 2 2000
Mr. Timothy J. Talcott Manager, Regulatory Affairs SIMS Portex, Incorporated 10 Bowman Drive P.O. Box 0724 Keene, New Hamphshire 03431
Re : K994275 SIMS Portex Anesthesia Catheter Trade Name: Regulatory Class: II Product Code: FRN Dated: December 17, 1999 December 20, 1999 Received:
Dear Mr. Talcott:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in
{4}------------------------------------------------
Page 2 - Mr. Talcott
the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fdag.gov/cdrh/dsmamain.html".
Sincerely yours,
U. Whitcomb
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
b
510(k) Number (if known): Unknown 13994275
Page 1 of 1
Device Name: Anesthesia Catheter
Indications For Use:
The SIMS Portex Anesthesia Catheter is indicated for the infusion of local anesthetics or narcotics into the intraoperative site for post operative pain management and for regional anesthesia, outside of the epidural space. Routes of administration may include intraoperative or percutaneous. The catheter is not intended for intravenous or intramuscular use.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
510(k) Number.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X OR Over-The-Counter Use **__**
(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices
Controly 994225
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.