EX 45-BP, MODEL 330 132; EX 53-BP, MODEL 130 018; EX 60-BP, MODEL 330 133

K994240 · Biotronik, Inc. · DTB · Apr 13, 2000 · Cardiovascular

Device Facts

Record IDK994240
Device NameEX 45-BP, MODEL 330 132; EX 53-BP, MODEL 130 018; EX 60-BP, MODEL 330 133
ApplicantBiotronik, Inc.
Product CodeDTB · Cardiovascular
Decision DateApr 13, 2000
DecisionST
Submission TypeTraditional
Regulation21 CFR 870.3680
Device ClassClass 3
AttributesTherapeutic

Indications for Use

BIOTRONIK's ELOX (EX xx-BP) transvenous, active fixation endocardial leads are indicated for permanent pacing and sensing. Active fixation pacing leads with a bipolar (BP) IS-1 connector configuration are designed for use in conjunction with implantable pulse generators with IS-1 headers. The leads may be used with single or dual chamber pacing systems. The ELOX (EX xx-BP) lead models are intended for placement in either the right atrium or right ventricle.

Device Story

Elox is a straight, bipolar endocardial pacing lead with an electrically active extendable/retractable fixation helix; used for permanent cardiac pacing and sensing. Device implanted in right atrium or right ventricle; connects to pulse generator via 3.2 mm IS-1 bipolar connector. Helix composed of 70% Pt / 30% Ir alloy with fractal Iridium coating; distal sleeve made of Pellethane 2363-75D; conductors consist of quadrifilar MP35N wire in coaxial configuration; insulated with silicone tubing. Physician operates device during implantation procedure. Output is electrical signal transmission between heart tissue and pulse generator; facilitates cardiac rhythm management. Benefits patient by providing reliable pacing/sensing therapy.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and design comparison to legally marketed predicate devices.

Technological Characteristics

Bipolar endocardial pacing lead; extendable/retractable fixation helix (70% Pt / 30% Ir alloy with fractal Iridium coating); Pellethane 2363-75D distal sleeve; quadrifilar MP35N wire conductors; silicone insulation; 3.2 mm IS-1 bipolar connector; available in 45 cm, 53 cm, and 60 cm lengths.

Indications for Use

Indicated for permanent pacing and sensing in patients requiring single or dual chamber pacing systems. Leads are placed in the right atrium or right ventricle and connected to implantable pulse generators with IS-1 headers.

Regulatory Classification

Identification

Temporary pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an external pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator. Permanent pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ APR 1 3 2000 # Elox Active Fixation Endocardial Lead 510(k) Notification ### 1. 510(K) SUMMARY Name and Address of Sponsor: BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, OR 97035 Establishment Registration Number: Device Name: 1028232 Proprietary Names: Elox Leads Class III (21 CFR 870.3680(b)) Classification: Classification Name: Cardiovascular Permanent Pacemaker Electrode Product Code: DTB Date Prepared: December 15, 1999 #### General Description and Predicate Devices: Elox (EX xx-BP) is a straight, bipolar endocardial pacing lead utilizing an electrically active extendable/retractable fixation helix. The extendable/retractable fixation helix is comprised of a 70% Pt / 30% Ir alloy with a fractal Iridium coating. The non-insulating distal sleeve, consisting of an inner and outer sleeve, is composed of Polyurethane (Pellethane 2363-75D). The lead contains two conductors composed of quadrifilar MP35N wire in coaxial configurations and is insulated with silicone tubing. A 3.2 mm IS-1 bipolar connector attaches the lead to the pulse generator. The Elox lead is available in lead lengths of 45 cm, 53 cm, and 60 cm. BIOTRONIK proposes the following leads cleared through 510(k) notifications as predicate devices for the Elox lead: - BIOTRONIK's Retrox bipolar, active fixation, endocardial lead (#K981083, cleared 07/22/98) . - Medtronic's Model 6957 J-58 unipolar, active fixation, endocardial lead (#K843707, cleared 01/08/85) . - BIOTRONIK's FH bipolar, active fixation, endocardial lead (#K910608, cleared 07/30/91) . - BIOTRONIK's Synox bipolar, passive fixation, endocardial lead (#K980869, cleared 09/10/98) . #### Indications for Use: BIOTRONIK's ELOX (EX xx-BP) transvenous, active fixation endocardial leads are indicated for permanent pacing and sensing. Active fixation pacing leads with a bipolar (BP) IS-1 connector configuration are designed for use in conjunction with implantable pulse generators with IS-1 headers. The leads may be used with single or dual chamber pacing systems. The ELOX (EX xx-BP) lead models are intended for placement in either the right atrium or right ventricle. #### Name and Address of Manufacturing Site: BIOTRONIK GmbH & Co. (reg. no. 7010992) Woermannkehre 1, 12359 Berlin, Germany 011-49-30-689-05-304 #### Contact Person(s) and Phone Number: Jon Brumbaugh Director, Regulatory Affairs Phone (888) 345-0374 Fax (503) 635-9936 {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the border. Inside the circle is a stylized image of three horizontal lines curving upwards, resembling a bird in flight. The seal appears to be a logo or emblem representing the U.S. Department of Health and Human Services. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 1 3 2000 Mr. Jon Brumbaugh Biotronik, Inc. 6024 Jean Road Lake Oswego, OR 97035 Re: K994240 Elox Active-Fixation Endocardial Pacing Lead, Model EX 45-BP, EX 53-BP, and EX 60-BP Regulatory Class: III (three) Product Code: DTB March 22, 2000 Dated: Received: March 23, 2000 Dear Mr. Brumbaugh: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ #### Page 2 - Mr. Jon Brumbaugh This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by referenće to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Sincerely yours, Clint M. Ahfor James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## 2. INDICATIONS FOR USE BIOTRONIK's ELOX (EX xx-BP) transvenous, active fixation endocardial leads are indicated for permanent pacing and sensing. Active fixation pacing leads with a bipolar (BP) IS-1 connector configuration are designed for use in conjunction with implantable pulse generators with IS-1 headers. The leads may be used with single or dual chamber pacing systems. The ELOX (EX xx-BP) lead models are intended for placement in either the right atrium or right ventricle. Chito lyn. Ah for Dillard. (Division Sign-Off) Division of Carelovascular, Respiratory, and Neurological Devices 510(k) Namber
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...