AT HOME DRUG TEST, MODEL 9073

K994219 · Phamatech · DIO · Dec 30, 1999 · Clinical Toxicology

Device Facts

Record IDK994219
Device NameAT HOME DRUG TEST, MODEL 9073
ApplicantPhamatech
Product CodeDIO · Clinical Toxicology
Decision DateDec 30, 1999
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.3250
Device ClassClass 2

Indications for Use

The At Home Drug Test is a rapid, qualitative immunoassay for the detection of the target drugs/drug metabolites in urine. The cut-off concentration for this test is as follows: Cocaine (benzoylecgonine): 300 ng/ml. This assay is intended for use in the home to assist in the prevention of drug abuse. This kit is designed to incorporate a mechanism for anonymous confirmation testing to be performed at a certified laboratory.

Device Story

At Home Drug Test (Model 9073) is a lateral flow, one-step immunoassay for qualitative detection of cocaine metabolites in urine. Device utilizes visual color sandwich immunochemical technology; specific antibody-drug-antibody complexes form in presence of analyte. Intended for home use by lay individuals to assist in drug abuse prevention. Kit includes mechanism for anonymous confirmation testing at certified laboratory. User applies urine sample to test device; visual interpretation of color bands indicates presence or absence of drug above 300 ng/ml cutoff. Provides preliminary screening results; intended to facilitate subsequent confirmatory testing if required.

Clinical Evidence

Clinical sample correlation study and blind labeled spiked study performed. Correlation >98% compared to Behring EMIT II and GC/MS. Professional user accuracy >97%. Consumer study (n=4715) showed 96.65% overall accuracy; consumer interpretation of cocaine test showed 97.24% accuracy (917/943 correct).

Technological Characteristics

Lateral flow, one-step visual color sandwich immunoassay. Qualitative detection of benzoylecgonine at 300 ng/ml cutoff. Standalone, non-electronic, disposable test kit.

Indications for Use

Indicated for home use by lay users for the qualitative detection of cocaine (benzoylecgonine) in urine at a 300 ng/ml cutoff. Includes a mechanism for anonymous laboratory confirmation.

Regulatory Classification

Identification

A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.

Special Controls

*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510 (k) SUMMARY AS REQUIRED BY SECTION 807.92(C) At Home Drug Test (Model 9073) ldentification: Description: Immunoassay for the qualitative detection of Cocaine (benzoylecgonine) in urine Name Of Manufacturer: Phamatech 9265 Activity Road #112 San Diego, California 92126, USA Intended Use: The At Home Drug Test is a rapid, qualitative immunoassay for the detection of the target drugs/drug metabolites in urine. The cut-off concentration for this test is as follows: Cocaine (benzoylecgonine): 300 ng/ml. This assay is intended for use in the home to assist in the prevention of drug abuse. This kit is designed to incorporate a mechanism for anonymous confirmation testing to be performed at a certified laboratory. The At Home Drug Test, like many commercially available drug screening test Technology: kits, qualitatively measures the presence of target drugs or their metabolites by visual color sandwich one step immunoassay technology. Examples of such predicate devices include the QuickScreen At Home Drug Test and the Applied Biotech SureStep Test (San Diego, CA 92121). All of the above devices rely on the basic immunochemical sandwich assay principle of recognition and formation of specific antibody / target drug / antibody / complexes. Performance: The product performance characteristics of the At Home Drug Test were evaluated in a clinical sample correlation study and a blind labeled spiked study. The results of these studies demonstrate the Phamatech At Home Drug Test to be substantially equivalent to the reported performance characteristics of other commercially available tests for the qualitative detection of the stated target drugs in urine. Correlation studies, using clinical specimens, produced a >98% correlation when compared to the Behring EMIT II (Cupertino, CA 95014) and GC/MS methodology. Clinical studies. performed at two independent laboratories, were also performed. In them the Phamatech QuickScreen" exhibited excellent overall accuracy (>97%) in the hands of professional users. A comsumer study was also performed. The At Home Druq Test exhibited excellent overall accuracy (4557/4715 correct or 96.65%) in the hands of lay users. Consumer interpretation of the Cocaine test in this study showed accuracy to be 917/943 or 97.24%. For the reasons mentioned above, it may be concluded that the Conclusion: Phamatech At Home Drug Test Model 9073 is substantially equivalent to a variety of detection tests currently in commercial distribution and is safe in the hands of the lay user. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol. The logo is black and white. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 DEC 3 0 1999 Mr. Carl A. Mongiovi Vice President of Operations Phamatech 9265 Activity Road Suite #112 San Diego, California 92126 Re: K994219 > Trade Name: At Home Drug Test (Model 9073) Regulatory Class: II Product Code: DIO Dated: August 25, 1999 Received: December 15, 1999 Dear Mr. Mongiovi: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failurs to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ ## Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrl/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## INDICATIONS FOR USE Applicant: Phamatech 510 (k) Number (if known): K994219 Device Name: At Home Drug Test (Model 9073) Indications for Use: A home drug screening, and if needed, confirmation service. This kit provides a preliminary result for the detection/presence of the following drugs of abuse in urine: Cocaine (benzoylecgonine). Dean Cooper (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K994219 ## PLEASE DO NOT WRITE BELOW THIS LINE Concurrence of the CDRH Office of Device Evaluation (ODE) Division Sign-off Division of Clinical Laboratory Devices 510 (k) Number: Over the Counter:_ Prescription Use: OR Per 21 CFR 801.109
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...