K994210 · Philips Medical Systems, Inc. · LLZ · Feb 4, 2000 · Radiology
Device Facts
Record ID
K994210
Device Name
INTURIS SUITE
Applicant
Philips Medical Systems, Inc.
Product Code
LLZ · Radiology
Decision Date
Feb 4, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Philips Inturis Suite is intended to be used for the display and reading of cardiac angiographic images stored on a compact disk (CD) or an image server. Inturis Suite also allows the user to change the appearance of the images and export selected frames to other programs for printing and further manipulation.
Device Story
Inturis Suite is modular software for import, storage, archive, review, analysis, and database management of digital cardiovascular angiographic images. Inputs include DICOM 3.0 formatted images from CDs or image servers. Software operates on Windows NT PCs or UNIX servers; used in clinical environments (e.g., Cath Lab) by healthcare professionals. Features include image display, digital zoom, window/level adjustment, annotation, and export of frames (bmp, jpg, avi). Output allows clinicians to visualize and manipulate angiographic data to support cardiovascular diagnostic and procedural decision-making.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Software-based image review system. Operates on Windows NT or UNIX platforms. Supports DICOM 3.0 standard for image import/export. Features include digital zoom, window/level adjustment, and annotation. Connectivity via network or CD-ROM. Modular architecture.
Indications for Use
Indicated for the display and reading of cardiac angiographic images for clinical review and analysis. No specific patient population or contraindications defined.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
ImageView Coronary Angiography Display and Review System (K972229)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a logo for Philips. The logo is in black and white and features the word "PHILIPS" at the top. Below the word is a shield-like shape containing a design with two wavy lines and two four-pointed stars. The overall design is simple and recognizable.
# Philips Medical Systems
# 510(K) Summary
K994210
In accordance with the requirements of the Safe Medical Device Act, Philips Medical Systems North America Company herewith submits a 510(K) summary of safety and effectiveness for the following device.
| SUBMITTER NAME / ADDRESS: | Philips Medical Systems North America Company<br>710 Bridgeport Avenue<br>Shelton, CT 06484-0917 |
|---------------------------|--------------------------------------------------------------------------------------------------|
| CONTACT PERSON / TEL NO: | Frank Gianelli<br>Tel No: (203) 926-7729 |
| DATE SUMMARY PREPARED: | December 8, 1999 |
| ESTABLISHMENT NO.: | 1217116 |
| TRADE/PROPRIETARY NAME: | Inturis Suite |
| COMMON/USUAL NAME: | Image Review Software for Angiographic X-ray System |
| CLASSIFICATION NAME: | Angiographic X-ray System<br>(21 CFR 892.1600; Class II; Tier 2; 90-IZI) |
| PREDICATE DEVICE(S): | ImageView Coronary Angiography Display and Review System; Heartware, Inc. |
# DEVICE DESCRIPTION:
Philips Inturis Suite is software comprised of modular software programs designed to perform the necessary functions required for the import, storage, archive, review, analysis, reporting and database management of digital cardiovascular angiographic images in DICOM 3.0 formats within and beyond the Cath Lab domain. Inturis Suite allows multiple users to quickly access and transfer multiple angiographic exams. Inturis Suite software can be installed and run on Windows NT based personnel computers for small applications and on UNIX based servers for high performance applications. The modular software design of Inturis Suite, allows users to tailor their image archive, transfer and communication requirements based on their particular performance needs and the number of cath labs importing images to an Inturis Suite via DICOM. Inturis Suite software will be provided on a CD-Rom which contains the installation program, the applicable executable files, DLL files and a readme file.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the Philips logo. The logo consists of the word "PHILIPS" in bold, uppercase letters, positioned above a shield-like shape. Inside the shield, there is a stylized globe with wavy lines representing water and star-like shapes representing land masses. The logo is black and white.
#### INTENDED USE:
Philips Inturis Suite is intended to be used for the display and reading of cardiac angiographic images stored on a compact disk (CD) or an image server. Inturis Suite also allows the user to change the appearance of the images and export selected frames to other programs for printing and further manipulation.
## SUBSTANTIAL EQUIVALENCE INFORMATION:
Inturis Suite is considered comparable and substantially equivalent to the following predicate device which has been cleared for commercial distribution via its referenced 510(k) submission.
· Device: ImageView Coronary Angiography Display and Review System Manufacturer: HeartWare, Inc. 510(k) No.: K972229
# TECHNOLOGICAL CHARACTERISTICS:
Predicate Device: ImageView Coronary Angiography Display and Review System, HeartWare, Inc. Comparison matrix of main features: Inturis Suite vs. predicate device
| Feature | Philips<br>Inturis Suite | HeartWare<br>ImageView |
|--------------------|--------------------------------------|-------------------------------------|
| Operating system | Windows NT; UNIX | Windows 95; Windows NT |
| Memory requirement | 128 MB (minimum)<br>256 MB (maximum) | 32 MB (minimum)<br>128 MB (maximum) |
| Image source | DICOM 3.0 SCU<br>DICOM 3.0 (CD-rom) | DICOM 3.0 SCU<br>DICOM 3.0 (CD-rom) |
| Display rate | up to 30 frames per sec | up to 30 frames per sec |
| Multiple windows | Yes | Yes |
| Digital zoom | Yes | Yes |
| Window (contrast) | Yes | Yes |
| Level (brightness) | Yes | Yes |
| Annotation | Yes | Yes |
| Image export | bmp, jpg, avi | bmp, gif |
| Network access | Yes | Yes |
## SAFETY INFORMATION:
Inturis Suite introduces no new safety and efficacy issues other than those already identified with the predicate device. The results of the hazard analysis, combined with the appropriate preventive measures taken indicate that the device is of minor level of concern.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings and body. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
4 2000 FEB
Frank Gianelli Senior Regulatory Affairs Specialist Philips Medical Systems 710 Bridgeport Avenue P.O. Box 860 Shelton, CT 06484-0917
Re:
K994210 Inturis Suite Dated: December 8, 1999 Received: December 14, 1999 Regulatory class: II 21 CFR 892.2050/Procode: 90 LLZ
Dear Mr. Gianelli:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act indude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either dass III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verty sych assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdch/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name :
Indications For Use :
Philips Inturis Suite is intended to be used for the display and reading of cardiac angiographic images stored on a compact disk (CD) or an image server. Inturis Suite also allows the user to change the appearance of the images and export selected frames to other programs for printing and further manipulation.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel le. Lyman
(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological Devices 510(k) Number
Prescription Use
( Per 21 CFR 801.109 )
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.