K994198 · Toray Industries (America), Inc. · KDI · Aug 25, 2000 · Gastroenterology, Urology
Device Facts
Record ID
K994198
Device Name
HEMOFEEL-CH, MODELS CH-0.35L, CH-0.6L, CH-1.0L
Applicant
Toray Industries (America), Inc.
Product Code
KDI · Gastroenterology, Urology
Decision Date
Aug 25, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hemofeel™-CH hemofilter is indicated for use in the prevention or treatment of fluid overload, electrolyte and acid/base imbalances in cases of acute renal failure or toxemia with oliguria or anuria. It may also be used when removal of excess fluid is indicated, such as patients in pulmonary edema or congestive heart failure refractory to diuretic therapy.
Device Story
Hemofeel™-CH Hemofilter is a single-use extracorporeal device for continuous hemofiltration. Blood enters through an arterial port, flows through polymethylmethacrylate (PMMA) hollow fibers, and exits through a venous port. The device filters blood to remove toxins and excess fluid. It is used in clinical settings for critically ill patients; operated by healthcare professionals. Physician selects filter size based on patient weight and condition. The device assists in managing fluid and electrolyte balance, providing therapeutic benefit for renal failure and refractory heart failure patients.
Clinical Evidence
Clinical evaluation conducted in eight anuric, critically ill patients. Continuous hemofiltration demonstrated utility in managing fluid and electrolyte balance in acute renal failure. No complications related to device use were reported. Supporting bench testing included ultrafiltration coefficient, water permeability, pressure drop, hemolysis, and shelf-life stability testing, all meeting acceptance criteria.
Technological Characteristics
Polymethylmethacrylate (PMMA) hollow fiber membrane; polystyrene casing and headers. Gamma-ray sterilized. Single-use. Dimensions and surface area vary by model size. No electronic components or software.
Indications for Use
Indicated for patients with acute renal failure, toxemia, oliguria, or anuria requiring management of fluid overload, electrolyte, or acid/base imbalances. Also indicated for fluid removal in patients with pulmonary edema or congestive heart failure refractory to diuretic therapy.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
# AUG 2 5 2000
## 510(k) SUMMARY K994198
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
#### Submitter's Name: Toray Industries (America), Inc.
600 Third Avenue, 5th floor New York, NY 10016-1921 Telephone: (212) 697-8150
Contact person: Mr. Hidehiko Okubo, Director, Medical and Pharmaceutical
Date of Summary: May 31, 2000
Device Name: Hemofeel™-CH Hemofilter
Device Classification Name: High permeability hemodialysis system (78 KDI); 21 CFR, Part 876.5860
Legally Marketed Devices to which Equivalence is Claimed: The legally marketed predicate devices are the Renal Systems. Inc. Renaflo Hemofiter (K854011), determined to be substantially equivalent to a legally marketed (preAmendment) device on December 5, 1985; and the Toray Industries Filtryzer™ Model BK-U (K935471), determined to be substantially equivalent to a legally marketed (preAmendment) device on May 19, 1995.
Device Description: The Hemofeel-CH Hemofilter consists of highly permeable polymethylmethacrylate (PMMA) hollow fibers housed in a polystyrene casing and secured at each end with polystyrene headers. Blood enters the device through the arterial inlet port, passes through the hollow fibers and exits at the opposite end through the venous outlet port. The Hemofeel-CH hemofilter is provided in a range of sizes based on the surface area of the hollow fiber membrane. Hemofilter size should be selected by the physician according to the patient's body weight and medical condition. The device is gamma-ray sterilized and intended for single use only.
Intended Use: The Hemofeel™-CH hemofilter is indicated for use in the prevention or treatment of fluid overload, electrolyte and acid/base imbalances in cases of acute renal failure or toxemia with oliguria or anuria. It may also be used when removal of excess fluid is indicated, such as patients in pulmonary edema or congestive heart failure refractory to diuretic therapy.
{1}------------------------------------------------
99499
Descriptive Summary of Technological Characteristics and Those of Predicate Devices: The features and capabilities of the Hemofeel are identical to those of the Renaflo Hemofilter. The two devices are similar in dimensions. For example, the two devices have essentially identical inner diameter, casing length, fiber length, membrane surface area, and blood volume. These similarities will logically result in substantially equivalent performance characteristics. Both the Hemofeel and the Filtryzer employ a PMMA membrane to filter the blood and remove toxins.
## Performance Data:
In Vitro Testing: Ultrafiltration coefficient testing, water permeability and pressure drop testing were conducted on the Hemofeel-CH hemofilter. All samples met the acceptance criteria. Hemolysis testing was also carried out on the device, with no hemolysis seen. These test results establish that the Hemofeel-CH hemofilter possesses performance characteristics that make it acceptable for its intended use.
Shelf life testing was conducted to establish the validity of the labeled three-year shelf life of the device. The results of the testing establish that biological, sterility, leak and water permeability characteristics are maintained at the end of the labeled shelf life.
Clinical: The Hemofeel-CH hemofitter was evaluated in the clinical setting in eight anuric critically ill patients. Continuous hemofiltration was found to be useful for the management of fluid and electrolyte balance in acute renal failure. There were no complications related to the use of the Hemofeel hemofilter. The data and information from the clinical study indicate that the Hemofeel is safe and effective for use in continuous hemofiltration.
Conclusion: The information and data provided in this 510(k) Notification establish that the Hemofeel-CH hemofilter is substantially equivalent to the legally marketed predicate devices.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is in black and white.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 5 2000
Toray Industries (America), Inc. c/o Ms. Lisa S. Jones, R.A.C. Devices for the Future 540 College Street Bellaire, TX 77401-5010
Re: K994198 Hemofeel™ - CH Hemofilter Dated: May 31, 2000 Received: June 1, 2000 Regulatory Class: II 21 CFR §876.5860/Procode: 78 KDI
#### Dear Ms. Jones:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the incluse for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce pror to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, Inerefore, market the device, subject to the general controls provisions of the Act. The general contrôls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against mistranding and adulteration.
If your device is classified (see above) into either class III (Premarket-Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices; General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notfication submission does not affect any obligation you micht have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours,
Daniel G. Schultz, M.D.
Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{3}------------------------------------------------
May 31, 2000
Page 1 of 1
## 510(k) Number: K994198
Device Name: Hemofeel™-CH Hemofilter
Indications for Use: The Hemofeel™-CH hemofilter is indicated for use in the prevention or treatment of fluid overload, electrolyte and acid/base imbalances in cases of acute renal failure or toxemia with oliguria or anuria. It may also be used when removal of excess fluid is indicated, such as patients in pulmonary edema or congestive heart failure refractory to diuretic therapy.
(Concurrence of CDRH, Office of Device Evaluation (ODE)
Yhind A. Sypm
---
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices 9941 510(k) Number .
Prescription Use
(Per 21 CFR 801.109) \$\underline{\hspace{1cm}} \checkmark\$
OR
Over-the-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.