K994170 · Biomatrix, Inc. · MRC · Mar 2, 2000 · Ophthalmic
Device Facts
Record ID
K994170
Device Name
HYLASHIELD CL
Applicant
Biomatrix, Inc.
Product Code
MRC · Ophthalmic
Decision Date
Mar 2, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5918
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Hylashield CL Lubricating Eye Drop is intended to lubricate and wet/rewet rigid gas permeable contact lenses.
Device Story
Hylashield CL Lubricating Eye Drop is a sterile, isotonic, preservative-free, elastoviscous ophthalmic solution containing 0.15% hylan fluid, a biocompatible derivative of hyaluronan. Supplied in single-use containers, the device is applied by the patient to rigid gas permeable (RGP) contact lenses to provide lubrication and wetting/rewetting. This process relieves ocular discomfort associated with RGP lens wear. The device functions by coating the lens surface, improving comfort during insertion and while the lens is worn. It is intended for over-the-counter use.
Clinical Evidence
Clinical study compared Hylashield CL to Wet-N-Soak in RGP lens wearers. Endpoints included slit lamp findings, symptoms/complaints, visual acuity, average wear time, and lens replacement. Results showed substantial equivalence in safety and efficacy, with high patient compliance and low adverse event rates in both groups.
Technological Characteristics
Sterile, isotonic, preservative-free, elastoviscous ophthalmic solution. Active ingredient: 0.15% hylan fluid (hyaluronan derivative). Form factor: single-use container. Biocompatibility testing performed per ISO 10993 (referenced as ISO-19033 in text).
Indications for Use
Indicated for lubricating and wetting/rewetting rigid gas permeable contact lenses.
Regulatory Classification
Identification
A rigid gas permeable contact lens care product is a device intended for use in the cleaning, conditioning, rinsing, lubricating/rewetting, or storing of a rigid gas permeable contact lens. This includes all solutions and tablets used together with rigid gas permeable contact lenses.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Predicate Devices
Wet-N-Soak® Rewetting Drops
OPTIMUM by Lobob® Wetting/Rewetting Drops
Submission Summary (Full Text)
{0}------------------------------------------------
MAR - 2 2000
Image /page/0/Picture/1 description: The image shows the logo for Biomatrix. The logo consists of a geometric shape on the left and the word "Biomatrix" on the right. The geometric shape is black and appears to be a stylized cube or block. The word "Biomatrix" is also in black and is in a bold, sans-serif font.
2.
Biomatrix, Inc. 65 Railroad Avenue Ridgefield New Jersey Tel 201 945 9550 Fax 201 945 0363
# K994170
# 510(k) SUMMARY
Hylashield® CL Lubricating Eye Drop hylan fluid, 0.15% Lubricant - Wetting/Rewetting Drop
- SUBMITTER'S NAME 1.
- CONTACT PERSON AT BIOMATRIX, INC. 2.
- DATE THAT 510(k) SUMMARY WAS PREPARED 3.
- NAME OF THE MEDICAL DEVICE (Classification/Common/Proprietary) 4.
- LEGALLY MARKETED DEVICES TO WHICH SUBSTANTIAL EQUIVALENCE IS 5. CLAIMED
- DESCRIPTION OF THE DEVICE 6.
- INTENDED USE OF THE DEVICE 7.
- TECHNOLOGICAL COMPARISON BETWEEN SUBJECT AND PREDICATE 8. DEVICES
- SUMMARY OF PRECLINICAL SAFETY STUDIES AND CONCLUSIONS FROM 9. PRECLINICAL SAFETY STUDIES
- SUMMARY OF PRECLINICAL PERFORMANCE STUDIES 10.
- CLINICAL SAFETY AND EFFICACY STUDY 11.
## SUBMITTER'S NAME 1.
BIOMATRIX, INC.
65 Railroad Avenue
Ridgefield, New Jersey 07657 USA
Tel: (201) 945-9550
Fax: (201) 945-0363
# U.S. REGULATORY CONTACT PERSON FOR BIOMATRIX, INC. 2.
Nadine M. Qashu BIOMATRIX, INC. 65 Railroad Avenue Ridgefield, New Jersey 07657 USA Tel: (201) 945-9550, ext. 202 Fax: (201) 945-0363
{1}------------------------------------------------
## 3. DATE THAT 510(k) SUMMARY WAS PREPARED
November 22, 1999
| 4. NAME OF THE MEDICAL DEVICE | |
|-------------------------------|---------------------------------------------------------------------|
| Classification name | Accessories to contact lenses - cleaning and wetting agents (86LPN) |
| Common / usual name | Lubricant - Wetting/Rewetting Drop |
| Proprietary name | Hylashield® CL Lubricating Eye Drop<br>hylan fluid, 0.15% |
LEGALLY MARKETED DEVICES TO WHICH SUBSTANTIAL 5. EQUIVALENCE IS CLAIMED Wet-N-Soak® Rewetting Drops For use with rigid gas permeable contact lenses as a rewetting drop OPTIMUM by Lobob® Wetting/Rewetting Drops For use with fluoro-silicone acrylate, silicone acrylate and hard contact lenses as an in-the-eye lubricant and wetting/rewetting drop
# 6. DESCRIPTION OF THE DEVICE
Hylashield® CL Lubricating Eye Drop is a device that has not been previously cleared by FDA for different intended uses.
Hylashield® CL Lubricating Eye Drop is a sterile, isotonic, preservative-free, clear, elastoviscous, ophthalmic solution presented in a single-use container. It contains hylan fluid, a viscous, elastic and biocompatible substance derived from the natural polysaccharide, hyaluronan. It lubricates and wets/rewets your lenses, thereby providing prolonged relief from symptoms of ocular discomfort associated with the use of rigid gas permeable contact lenses.
{2}------------------------------------------------
## 7. INTENDED USE OF THE DEVICE
Hylashield CL Lubricating Eye Drop is intended to lubricate and wet/rewet rigid gas permeable contact lenses.
# 8. TECHNOLOGICAL COMPARISON BETWEEN SUBJECT AND PREDICATE DEVICES
Allergan markets Wet-N-Soak® Rewetting Drops and Lobob Laboratories markets OPTIMUM by Lobob® Wetting and Rewetting Drop for use with rigid gas permeable contact lenses. These predicate devices are indicated for relief of eye discomfort caused by wearing rigid gas permeable contact lenses. Hylashield® CL Lubricating Eye Drop is substantially equivalent to these products in that it has a similar intended use and indication.
The material from which Hylashield® CL Lubricating Eye Drop is made is different from the two predicate devices. Hylashield® CL Lubricating Eye Drop is composed of hylan fluid, whereas Wet-N- Soak® Rewetting Drops and OPTIMUM by Lobob® Wetting and Rewetting Drops are composed of hydroxyethylcellulose. OPTIMUM by Lobob® Wetting and Rewetting Drop, in addition, contains polyvinylpyrrolidone and polyvinyl alcohol. Hylan has been shown to be compatible with ocular structures and raises no new safety and efficacy issues. Also, a preservative is added to the two predicate devices, due to the fact that they are not supplied in a single-use container. Since Hylashield® CL Lubricating Eye Drop is supplied in a sterile, single-use container, it does not contain the preservatives that are used in the two predicate devices.
Hylashield® CL Lubricating Eye Drop is the same as Wet-N-Soak® Rewetting Drops and OPTIMUM by Lobob® Wetting/Rewetting Drop in that they are all supplied in the form of a sterile eye drop and serve as lubricants and wetting/rewetting drops for rigid gas permeable contact lenses. They are all made from materials which have demonstrated biocompatiblity.
In conclusion, Hylashield® CL Lubricating Eye Drop has the same intended use as the marketed devices and differs only in the material from which it is made. All three devices are used to relieve eye discomfort caused by wearing rigid gas permeable contact lenses by wetting the lenses prior to insertion and by rewetting and lubricating the lenses while they are on the eve.
{3}------------------------------------------------
# 9. SUMMARY OF PRECLINICAL SAFETY STUDIES AND CONCLUSIONS FROM PRECLINICAL SAFETY STUDIES Extensive in vivo animal and human use of hyaluronan and hylans have shown these products to be highly compatible for contact lens wearers. The studies contained in section A, ISO Biocompatibility Tests, were performed under and comply with the regulations outlined in 21 CFR Part 58 for Good Laboratory Practices. A. ISO BIOCOMPATIBILITY TESTS - Performed under GLP Ref: International Standard ISO - 19033 (Biological Evaluation of Medical Devices) 1. SHORT TERM BIOLOGICAL TESTS 1.1 Irritation Tests BXR 15202-F- Intracutaneous toxicity study in rabbit [hylan fluid]-GLP Ocular Irritation: BXR 50002-Eye Irritation Testing Protocol: Biomatrix Hylashield® 14-day Repeat Dose with Modified Hackett-McDonald Scoring-GLP 1.2 Sensitization and Immunogenicity Assays BXR-10501-Dermal sensitization study (A Maximization Test) of hylan solution in the guinea pig -GLP . 1.3 Cytotoxicity BXR 15201-F- In vitro cytotoxicity study (MEM Elution Method) L-929 mouse fibroblast cell line [hylan fluid]-GLP 1.4 Acute Systemic Toxicity BXR 15201-F Systemic toxicity study in mice [hylan fluid] (USP)-GLP 1.5 Hemocompatibility and Hemolysis BXR 15200-F In vitro hemolysis study (direct contact method) [hylan fluid]-GLP 1.6 Pyrogenicity BXR 15302-F - Rabbit pyrogen test of hylan fluid (USP) 1.7 Implantation BXR 15204-F – USP muscle implantation study (with histopathology) in the rabbit (7 days) [hylan fluid]-GLP BXR 15203-F - Muscle implantation study (with histopathology) in the rabbit (30 days) [hylan fluid] - GLP 1.8 Mutagenicity BXR 10201-F - Ames mutagenicity test of hylan solution-GLP 2. LONG TERM BIOLOGICAL TESTS 2.1 Subchronic Toxicity BXR 15303-2-F -- Subchronic Two-Week Intraperitoneal Toxicity Study on Hylan A in Male Guinea Pigs [Effect of repeated intraperitoneal injections of hylan fluid (9.53 mg/ml) on blood
BXR 15305-2-F - Subchronic Two-Week Intraperitoneal Toxicity Study on Hylan A Female Guinea Pigs [Effect of repeated intraperitoneal injections of hylan fluid (10mg/ml) on blood parameters and histology of selected tissue - GLP
parameters and histology of selected tissues in male guinea pigs] - GLP
{4}------------------------------------------------
# 9. SUMMARY OF PRECLINICAL SAFETY STUDIES AND CONCLUSIONS FROM PRECLINICAL SAFETY STUDIES (continued)
BXR 15209-F - Subchronic One-Month Intra-Arterial Toxicity Study on Hylan A in Male Rabbits [Effect of intra arterial injections of hylan fluid (1 mg/ml, 3 mg/kg) on hematology and blood chemistry, blood hyaluronan content, and histology of selected tissue in male rabbits] - GLP
BXR 15303-F - Subchronic One-Month Intra-Articular Toxicity Study on Hylan A in Male Rabbits [Effect of repeated intra-articular application of hylan fluid (3.54 mg/ml) on joint biochemistry , hematology, and blood chemistry in rabbits] - GLP
#### 2.2 Reproduction Studies
BXR 12243-F – Effect of Intra-ocular Implantation of Hylan Fluid, Hylan Gel, Hyaluronan and Hylan G-F 20 on the Reproductive Capacity of Owl Monkeys: A Retrospective Study.
#### 3. PHARMACOKINETICS
BXR 15211-F - Clearance of hylan fluid after intra-articular injection into rabbit knees as Determined by radiotracer studies
BXR 15215-F - Clearance of hylan fluid after direct intravascular administration
#### BASIC EXPLORATORY STUDIES (SUPPORTIVE STUDIES) A.
#### SHORT TERM BIOLOGICAL STUDIES 1.
#### 1.1 Acute Systemic Toxicology
BXR 15208-F ~ Effect of single intra-articular injection (0.3 ml) of hylan fluid (5 mg/ml) on joint biochemistry
BXR 15208-2-F – Effect of single large volume (0.5 ml) intra-articular application of hylan fluid (10 mg/ - joint histology/biochemistry)
BXR 15408-F ~ Effect of single intra-articular application of hylan fluid (10 mg/ml) on joint histology in female rabbits
BXR 15311-F - Histology studies of rabbit joint tissues one and four weeks after single Intra-articular administration of 0.3 ml hylan fluid (5 mg/ml)
BXR 15308-F - Effect of single large volume (2.0 ml) intra-articular application of hylan fluid on joint biochemistry
#### 1.2 Hemocompatibility and Hemolysis
BXR 12203-F – Rabbit blood hemolysis test of hylan fluid BXR 12205-F - Rabbit blood hemolysis: hylan fluid extracts
BXR 12216-F - Partial thromboplastin time: effect of hylan fluid BXR 12217-F - Platelet activation studies: in vitro effect of hylan fluid
{5}------------------------------------------------
# SUMMARY OF PRECLINICAL SAFETY STUDIES AND CONCLUSIONS 9. FROM PRECLINICAL SAFETY STUDIES (continued)
#### 1.3 Implantation
BXR 15310-B-F - Implantation of hylan fluid into the intra-vitreal space of primate eyes: short term evaluation
BXR 15310-A-F - Implantation of hylan fluid into the intra-vitreal space of primate eyes: long term evaluation
#### 2. LONG TERM BIOLOGICAL TESTS
# Subchronic Toxicity
BXR 10522-F - Short term toxicity testing of hylan fluid in rats; administration by weekly Intra-peritoneal injection
BXR 15406-F - Effect of repeated intra-articular application of hylan fluid on joint histology of female rabbits
#### 3. Reproduction Studies
BXR 15320-F - Effect of hylan fluid on fertility and pregnancy in horse
## 10. IN VIVO TEST OF EFFICACY
BXR 15501 – Effect of Hylan Fluid on Fluorescein Break-Up-Time (FBUT) and Drying Time in Owl Monkey Eyes
Summaries and conclusions of all the above studies are included in the Biocompatibility Section of this document.
{6}------------------------------------------------
## CLINICAL SAFETY AND EFFICACY STUDY 11.
The safety and efficacy of Hylashield® CL Lubricating Eye Drop as a lubricating and wetting/rewetting agent was compared to one of the predicate devices, Wet-N-Soak®, in minimizing daily discomfort associated with rigid gas permeable (RGP) contact lenses in this study. These products were evaluated in conjunction with the two most commonly used RGP lens types. The analysis of the clinical study data showed the substantial equivalence of Hylashield® CL Lubricating Eye Drop with Wet-N-Soak® with regard to slit lamp findings, symptom/problems/complaints, visual acuity, average wear time and lens replacement. The high patient compliance rate and extremely low rate of adverse events in both treatment groups indicated that Hylashield® CL Lubricating Eye Drop is well tolerated and safe when compared to Wet-N- Soak®.
In conclusion, the data from this report demonstrated that Hylashield® CL Lubricating Eye Drop is substantially equivalent to its predicate device, Wet-N-Soak with respect to safety and effectiveness as a lubricating and wetting/rewetting agent to provide prolonged relief from symptoms of ocular discomfort associated with the use of RGP contact lenses. Wet-N-Soak® has the same intended use as OPTIMUM by Lobob® Wetting and Rewetting Drop.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR - 2 2000
Ms. Nadine M. Qashu Director, Regulatory Affairs Biomatrix, Inc. 65 Railroad Avenue Ridgefield, New Jersey 07657
Re: K994170 Trade Name: Hylashield CL Lubricating Eye Drops Regulatory Class: II Product Code: 86 MRC Dated: December 10, 1999 Received: December 10,1999
Dear Ms. Qashu:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{8}------------------------------------------------
Page 2 - Ms. Nadine M. Qashu
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on.the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A Roerl Rosenthal
A. Ralph Rosenthal, M.D Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# 1. Statement of Indications for Use:
Application: Biomatrix, Inc. _________________________________________________________________________________________________________________________________________________
510(K) Number (if known): K994 170
Device Name: Hylashield® CL Lubricating Eye Drops
Hylashield® CL Lubricating Eye Drops are indicated for Indications for Use: lubricating and wetting/rewetting rigid gas permeable contact lenses.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | | or | Over-the-Counter Use | X |
|-----------------------|--|----|--------------------------|---|
| (per 21 CFR 886.5918) | | | (Optional Format 1-2-96) | |
(Division Sign-Off)
Division of Ophthalmic Devices
| 510(k) Number | K994170 |
|---------------|---------|
|---------------|---------|
l
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.