← Product Code [DIO](/productcode/DIO) · K994165

# ONTRAK TESTSTIK 3 FOR COCAINE/MORPHINE/THC,CAT. 1118595 (K994165)

_Roche Diagnostics Corp. · DIO · Apr 7, 2000 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K994165

## Device Facts

- **Applicant:** Roche Diagnostics Corp.
- **Product Code:** [DIO](/productcode/DIO.md)
- **Decision Date:** Apr 7, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3250
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

OnTrak TesTstik 3 for Cocaine/Morphine/THC is an in vitro diagnostic test intended for professional use for the qualitative detection of cocaine metabolite, morphine, and cannabinoids in uripe. The OnTrak TesTstik 3 for Cocaine/Morphine/THC cutoff levels (300 ng/ml for cocaine and morphine and 50 ng/ml for THC) are based on the Federal Mandatory Guidelines. OnTrak TesTstik 3 for Cocaine/Morphine/THC is not intended for over-the-counter sale. OnTrak TesTstik 3 for Cocaine/Morphine/THC provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytices I rosult. Tras chromatography/mass spectrometry (GC/MS) is the preferred confirmation method.) Cluireal consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

## Device Story

In vitro diagnostic test for professional use; detects cocaine metabolite, morphine, and cannabinoids in urine. Principle: competitive microparticle capture inhibition; competition between drugs in urine and drug conjugates immobilized on membrane. Input: urine sample; Output: qualitative color-based visual result. Used in clinical settings by professionals. Provides preliminary results; requires GC/MS confirmation. Clinical decision-making relies on professional judgment alongside preliminary test results.

## Clinical Evidence

Clinical evaluation using 70 positive and 100 negative urine specimens. Accuracy demonstrated via comparison to automated immunoassay screening and GC/MS confirmation. Agreement: 94% for cocaine, 94% for morphine, 95% for THC. Precision >95% at 150% of cutoff.

## Technological Characteristics

Competitive microparticle capture inhibition assay. Reagents: blue dyed microparticles coated with mouse monoclonal anti-benzoylecgonine, anti-morphine, and anti-cannabinoid antibodies; drug conjugates and mouse monoclonal anti-BSA antibody immobilized on membrane. Qualitative color-based endpoint. Urine sample type.

## Regulatory Identification

A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.

## Special Controls

*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Predicate Devices

- OnTrak TesTstik for Cocaine ([K973075](/device/K973075.md))
- OnTrak TesTstik for Morphine ([K973075](/device/K973075.md))
- OnTrak TesTstik for THC ([K973075](/device/K973075.md))

## Reference Devices

- Abuscreen OnTrak for Cocaine
- Abuscreen OnTrak for Morphine
- Abuscreen OnTrak for THC

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### 510(k) Summary

| Introduction                              | According to the requirements of 21 CFR 807.92, the following information<br>provides sufficient detail to understand the basis for a determination of<br>substantial equivalence. |                              |  |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|--|
| 1) Submitter<br>name, address,<br>contact | Roche Diagnostics Corporation<br>9115 Hague Rd.<br>Indianapolis, IN 46250                                                                                                          |                              |  |
|                                           | Contact Person: Jennifer Tribbett                                                                                                                                                  | Date Prepared: March 6, 2000 |  |

2) Device Name The device name, including both the trade/proprietary name and classification name is provided below.

| Product Name                                      | Classification<br>Name     | Product<br>Code | CFR<br>Classification<br>Name | Predicate<br>Device Name           | Date<br>Predicate<br>Cleared | Predicate<br>510(k)<br>Number |
|---------------------------------------------------|----------------------------|-----------------|-------------------------------|------------------------------------|------------------------------|-------------------------------|
| OnTrak TesTstik<br>3 for Cocaine/<br>Morphine/THC | Cocaine Test<br>System     | 91DIO           | 862.3250                      | OnTrak<br>TesTstik for<br>Cocaine  | 10/9/97                      | K973075                       |
| OnTrak TesTstik<br>3 for Cocaine/<br>Morphine/THC | Morphine Test<br>System    | 91DJJ           | 862.3640                      | OnTrak<br>TesTstik for<br>Morphine | 10/9/97                      | K973075                       |
| OnTrak TesTstik<br>3 for Cocaine/<br>Morphine/THC | Cannabinoid<br>Test System | 91LDJ           | 862.3870                      | OnTrak<br>TesTstik for<br>THC      | 10/9/97                      | K973075                       |

3) Predicate We claim substantial equivalence to the currently marketed Roche device Diagnostics OnTrak TesTstik for Cocaine (K973075), OnTrak TesTstik for Morphine (K973075) and OnTrak TesTstik for THC (K973075).

4) Device The OnTrak TesTstik 3 for Cocaine/Morphine/THC is an in vitro diagnostic Description test intended for professional use in the qualitative detection of cocainc metabolite, morphine and cannabinoids in urine at or above a cutoff of 300 ng/ml (cocaine and morphine) and 50 ng/ml (THC).

> Like the predicate TesTstik devices, the OnTrak TesTstik 3 is also based on the principles of microparticle capture inhibition. The test relies on the competition between drugs, which may be present in the urine being tested, and drug conjugates immobilized on the membrane.

{1}------------------------------------------------

| 5. Technology<br>Characteristics | Tables 1, 2, 3 and 4 outline the technological characteristics (methodologies) of the OnTrak TesTstik 3 for Cocaine/Morphine/THC in comparison to the predicate devices.                                                                                                                                                                                  |  |  |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| 6. Substantial<br>Equivalence    | Tables 1, 2, 3 and 4 also provide the significant characteristics relied upon for a determination of substantial equivalence. This information concludes that the performance of the TesTstik 3 for Cocaine/Morphine/THC is substantially equivalent to the currently marketed OnTrak TesTstik Cocaine, OnTrak TesTstik Morphine and OnTrak TesTstik THC. |  |  |

### TABLE 1

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

| Item                               | OnTrak TesTstik 3 for<br>Cocaine/Morphine/THC<br>New Device                                                                                                  | OnTrak TesTstik<br>for Cocaine<br>Predicate                                                            | OnTrak TesTstik<br>for Morphine<br>Predicate                                                 | OnTrak TesTstik<br>for THC<br>Predicate                                                            |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Methodology                        | Competitive microparticle<br>capture inhibition                                                                                                              | Same                                                                                                   | Same                                                                                         | Same                                                                                               |
| Measurement                        | Qualitative                                                                                                                                                  | Same                                                                                                   | Same                                                                                         | Same                                                                                               |
| Sample Type                        | Urine                                                                                                                                                        | Same                                                                                                   | Same                                                                                         | Same                                                                                               |
| Endpoint read                      | Color                                                                                                                                                        | Same                                                                                                   | Same                                                                                         | Same                                                                                               |
| Cutoff                             | 300 ng/ml (cocaine)<br>300ng/ml (morphine)<br>50 ng/ml (THC)                                                                                                 | Same<br>300 ng/ml (cocaine)                                                                            | Same<br>300 ng/ml (morphine)                                                                 | Same<br>50 ng/ml (THC)                                                                             |
| Reagent<br>(active<br>ingredients) | • Blue dyed microparticles<br>coated with mouse<br>monoclonal anti-<br>benzoylecgonine antibody,<br>anti-morphine antibody and<br>anti-cannabinoid antibody. | • Blue dyed<br>microparticles coated<br>with mouse<br>monoclonal anti-<br>benzoylecgonine<br>antibody. | • Blue dyed<br>microparticles coated<br>with mouse<br>monoclonal anti-<br>morphine antibody. | • Blue dyed<br>microparticles coated<br>with mouse<br>monoclonal anti-<br>cannabinoid<br>antibody. |
|                                    | • Drug conjugates<br>immobilized on a membrane                                                                                                               | • Drug conjugates<br>immobilized on a<br>membrane                                                      | • Drug conjugates<br>immobilized on a<br>membrane                                            | • Drug conjugates<br>immobilized on a<br>membrane                                                  |
|                                    | • Mouse monoclonal anti-<br>BSA antibody immobilized<br>on membrane                                                                                          | • Mouse monoclonal<br>anti-BSA antibody<br>immobilized on a<br>membrane                                | • Mouse monoclonal<br>anti-BSA antibody<br>immobilized on a<br>membrane                      | • Mouse monoclonal<br>anti-BSA antibody<br>immobilized on a<br>membrane                            |
| Performance:<br>Precision          | >95% confidence at 150%<br>cutoff                                                                                                                            | Same                                                                                                   | Same                                                                                         | Same                                                                                               |

{2}------------------------------------------------

## 510K Summary -Continued-

### Table 2

| Item                                | OnTrak TesTstik 3 for<br>Cocaine/Morphine/THC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | OnTrak TesTstik for Cocaine<br>Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cocaine<br>Performance:<br>Accuracy | OnTrak TesTstik 3 for<br>Cocaine/Morphine/THC was<br>evaluated using clinical<br>specimens screened by an<br>automated immunoassay and<br>confirmed positive by GC/MS.<br>Seventy (70) specimens,<br>including near cutoff specimens,<br>positive for cocaine were<br>evaluated on OnTrak TesTstik 3.<br><br>One hundred urine specimens<br>were obtained from a clinical<br>laboratory and screened negative<br>by automated immunoassays<br>relative to a 300 ng/ml cutoff for<br>cocaine. A portion (15%) of the<br>negative specimens were also<br>confirmed negative for cocaine<br>by GC/MS. Clinical correlation<br>for all positive (including near<br>cutoff samples) and negative<br>specimens was demonstrated as<br>94% agreement. | OnTrak TesTstik for Cocaine<br>was evaluated using specimens<br>screened by an automated<br>immunoassay and confirmed<br>positive by GC/MS at a 150<br>ng/ml cutoff. Fifty (50) samples<br>positive for cocaine were positive<br>by OnTrak TesTstik (100%).<br><br>One hundred six (106) urine<br>samples, obtained from a clinical<br>laboratory and screened negative<br>by an automated immunoassay<br>relative to a 300 ng/ml cutoff for<br>cocaine were evaluated and<br>found negative using OnTrak<br>TesTstik.<br><br>All positive and negative samples<br>were also assayed by, and<br>compared to, Abuscreen OnTrak<br>for Cocaine. All samples<br>demonstrated 100% agreement<br>between the two assays. |

{3}------------------------------------------------

# 510K Summary -Continued-

| Table 3                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item                                 | OnTrak TesTstik 3 for<br>Cocaine/Morphine/THC                                                                                                                                                                                                                                                                                                                                                                                                                | OnTrak TesTstik for Morphine<br>Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Morphine<br>Performance:<br>Accuracy | OnTrak TesTstik 3 for<br>Cocaine/Morphine/THC was<br>evaluated using clinical<br>specimens screened by an<br>automated immunoassay and<br>confirmed positive by GC/MS.<br>Seventy (70) specimens,<br>including near cutoff specimens,<br>positive for morphine were<br>evaluated on OnTrak TesTstik 3.                                                                                                                                                       | OnTrak TesTstik for Morphine<br>was evaluated using specimens<br>screened by an automated<br>immunoassay and confirmed<br>positive by GC/MS at a 300<br>ng/ml cutoff. Forty nine (49)<br>samples positive for morphine<br>were positive by OnTrak<br>TesTstik (>99%).                                                                                                                                                                                                                                                                                                                       |
|                                      | One hundred urine specimens<br>were obtained from a clinical<br>laboratory and screened negative<br>by automated immunoassays<br>relative to a 300 ng/ml cutoff for<br>morphine. A portion (15%) of<br>the negative specimens were also<br>confirmed negative for opiates<br>(morphine and codeine) by<br>GC/MS. Clinical correlation for<br>all positive (including near cutoff<br>samples) and negative specimens<br>was demonstrated as 94%<br>agreement. | One hundred six (106) urine<br>samples, obtained from a clinical<br>laboratory and screened negative<br>by an automated immunoassay<br>relative to a 300 ng/ml cutoff for<br>morphine were evaluated using<br>OnTrak TesTstik. One hundred<br>five (105) were and negative for<br>morphine using OnTrak<br>TesTstik.<br><br>All positive and negative samples<br>were also assayed by, and<br>compared to, Abuscreen OnTrak<br>for Morphine. One hundred fifty<br>six (156) samples tested by both<br>OnTrak TesTstik and Abuscreen<br>OnTrak for Morphine<br>demonstrated 99.3% agreement. |

## . Table 3

{4}------------------------------------------------

# 510K Summary -Continued-

### Table 4

| Item                            | OnTrak TesTstik 3 for<br>Cocaine/Morphine/THC                                                                                                                                                                                                                                                         | OnTrak TesTstik for THC<br>Predicate                                                                                                                                                                                                                    |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| THC<br>Performance:<br>Accuracy |                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                         |
|                                 | OnTrak TesTstik 3 for<br>Cocaine/Morphine/THC was<br>evaluated using clinical<br>specimens screened by an<br>automated immunoassay and<br>confirmed positive by GC/MS.<br>Seventy (70) specimens,<br>including near cutoff specimens,<br>positive for THC were evaluated<br>on OnTrak TesTstik 3.     | OnTrak TesTstik for THC was<br>evaluated using specimens<br>screened by an automated<br>immunoassay and confirmed<br>positive by GC/MS at a 15 ng/ml<br>cutoff. Forty five (45) samples<br>positive for THC were positive by<br>OnTrak TesTstik (100%). |
|                                 | One hundred urine specimens<br>were obtained from a clinical<br>laboratory and screened negative<br>by automated immunoassays<br>relative to a 50 ng/ml cutoff for<br>THC. A portion (15%) of the<br>negative specimens were also<br>confirmed negative for THC by<br>GC/MS. Clinical correlation for | One hundred six (105) urine<br>samples, obtained from a clinical<br>laboratory and screened negative<br>by an automated immunoassay<br>relative to a 50 ng/ml cutoff for<br>THC were evaluated and found<br>negative using OnTrak TesTstik.             |
|                                 | all positive (including near cutoff<br>samples) and negative specimens<br>was demonstrated as 95%<br>agreement.                                                                                                                                                                                       | All positive and negative samples<br>were also assayed by, and<br>compared to, Abuscreen On Trak<br>for THC. All samples<br>demonstrated 100% agreement<br>between the two assays.                                                                      |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR - 7 2000

Ms. Jennifer Tribbett Regulatory Affairs Specialist Roche Diagnostics Corporation 9115 Hague Road P.O. Box 50457 Indianapolis, Indiana 46250-0457

Re: K994165 Trade Name: Ontrak TesTstik™ 3 Regulatory Class: II Product Code: DIO, DJJ, LDJ Dated: March 6, 2000 Received: March 7, 2000

Dear Ms. Tribbett:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{6}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other gencral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

510(k) Number (if known): Device Name: Roche Diagnostics Corporation, OnTrak TesTstik™ 3 for Cocaine/Morphine/THC

Indications for Use:

OnTrak TesTstik 3 for Cocaine/Morphine/THC is an in vitro diagnostic test intended for professional use for the qualitative detection of cocaine metabolite, morphine, and cannabinoids in uripe. The OnTrak TesTstik 3 for Cocaine/Morphine/THC cutoff levels (300 ng/ml for cocaine and morphine and 50 ng/ml for THC) are based on the Federal Mandatory Guidelines. OnTrak TesTstik 3 for Cocaine/Morphine/THC is not intended for over-the-counter sale.

OnTrak TesTstik 3 for Cocaine/Morphine/THC provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytices I rosult. Tras chromatography/mass spectrometry (GC/MS) is the preferred confirmation method.) Cluireal consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Clinical Laboratory evices
510(k) Number 4994165

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

CONFIDENTIAL

---

**Source:** [https://fda-staging.innolitics.com/device/K994165](https://fda-staging.innolitics.com/device/K994165)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).